Professional Documents
Culture Documents
This system is versatile device to care acute and sub-acute respiratory failure patient.
HFT700 have Fulfill hardware for Oxygen therapy with Turbine, Oxygen blender, Humidifier,
heated circuit system, FiO2, SpO2 and Respiration monitoring.
Display Monitor 5.0" Color TFT(800 x 480), Touch Screen
2. Detail Specification
□ Control Settings
HF
Flow (Inspiratory flow rate) 1 ~ 60 LPM
RR detection OFF
Internal sensor
CPAP
CPAP (continuous positive airway pressure) 4 ~ 20 cmH2O ±(1.7 + 4 % of the set value) cmH2O
bi-Level[S/T]
IPAP 4 ~ 40 cmH2O ±(1.7 + 4 % of the set value) cmH2O
Common
FiO2 % High pressure 21, 25 ~ 100 %, Electronic controlled blender (built-in system)
Accuracy : ± 5 %
Low flow Device does not support oxygen concentration control. (Only FiO2 monitoring)
□ Display Data
Parameters Setting parameters/patient status parameters, Alarm status , Humidifier status
Alarm Patient alarm Open Circuit, APNEA, Check Cannula, Leakage High, Total Leakage high
System alarm AC Loss, Check Blower, O2 inlet pressure High, Blower temp High,
Chamber Temp. Sensor is Fail, Air Temp. High, Airway Temp Low
PIP(Peak Inspiration Pressure) High : IPAP(CPAP) Setting +1 ~ 43 cmH2O, Low : 0 ~ IPAP(CPAP) setting-1
FiO2 % High : Low setting value +1 ~ 100 % , Low : 18 ~ High setting value -1
RR(Respiration Rate) High : Low setting value +1 ~ 150 /min, Low : 0 ~ High setting value -1
Airway Temperature High : Low setting value +1 ~ 44℃, Low : 0 ~ High setting value -1
SpO2 High : Low setting value +1 ~ 100 %, Low : 0 ~ High setting value -1
PR(Pulse Rate) High : Low setting value +1 ~ 250 / min, Low : 0 ~ High setting value -1
V delivery High : Low setting value +1 ~ 2500 / ml, Low : 50 ~ High setting value -1
□ Humidifier
Humidification level 1~5
Advanced humidification control system based on set level, flow rate, room
Humidification capacity AH 20 ~ 40 mg / L
□ Key, Indicators
Lock Touch Screen Lock
□ Electrical Requirements
Power AC : 100~240 V 50/60 Hz Free voltage , Fuse Capacity 250 V / 2.5A
□ Communication
RS232 Labview program connection using PC
Wi-Fi CMS
□ Physical Characteristics
Weight 4.5 Kg
Dimensions (With 8.4 External Power Supply) 204 x 338 x 254 (Width x Depth x Height) mm
□ Language
English / Korean
□ Environmental Requirements
Atmospheric condition 700 ~ 1060 mbar
Altitude Up ~ 3000 m
Humidity 0 ~ 95 % (Non-condensing)
Operating Temperature 5 ~ 40 °C