You are on page 1of 4

HFT700 GLOBAL向

Document No.: WQ32-20-L003 (Rev.3.6)


Date: June 22. 2021

Specifications Model : HFT700


1. Features and Short form Specifications

This system is versatile device to care acute and sub-acute respiratory failure patient.
HFT700 have Fulfill hardware for Oxygen therapy with Turbine, Oxygen blender, Humidifier,
heated circuit system, FiO2, SpO2 and Respiration monitoring.
Display Monitor 5.0" Color TFT(800 x 480), Touch Screen

Therapy Modes HF mode, CPAP, Bi-Level mode

Vital Sign Functions SpO2

Option MASIMO SpO2 module kit, Wifi, Bluetooth, TSF

External Power Supply(MPS100)

Cart(Available installation UPS and O2 Cylinder)

2. Detail Specification
□ Control Settings
HF
Flow (Inspiratory flow rate) 1 ~ 60 LPM

RR Sensitivity x0.2 ~ x3.0

RR detection OFF

Internal sensor

Respiratory detection Line (With an external RR detection accessory)

CPAP
CPAP (continuous positive airway pressure) 4 ~ 20 cmH2O ±(1.7 + 4 % of the set value) cmH2O

Pressure Assist + (PA(+)) OFF, 1 ~ 3 cmH2O

Pressure Assist - (PA(-)) OFF, -3 ~ -1 cmH2O

Trigger Level 3 ~ 20 lpm

Rise time Fast(0.2 sec), Medium(0.3 sec), Slow(0.4 sec)

Auto Start ON, OFF

bi-Level[S/T]
IPAP 4 ~ 40 cmH2O ±(1.7 + 4 % of the set value) cmH2O

EPAP 4 ~ 20 cmH2O ±(1.7 + 4 % of the set value) cmH2O

Pressure Assist - (PA(-)) OFF, -3 ~ -1 cmH2O

Respiration Rate 2 ~ 60 bpm

Inspiratory Time(Ti) 0.3 ~ 3.0 seconds

Trigger Level 3 ~ 20 lpm

Trigger Type S/T, T, S

Rise time Fast(0.2 sec), Medium(0.3 sec), Slow(0.4 sec)

Auto Start ON, OFF

Common
FiO2 % High pressure 21, 25 ~ 100 %, Electronic controlled blender (built-in system)

Accuracy : ± 5 %

Low flow Device does not support oxygen concentration control. (Only FiO2 monitoring)

□ Display Data
Parameters Setting parameters/patient status parameters, Alarm status , Humidifier status

Graphic Waveforms Pressure-Time & Flow-Time(Only on Bi-Level Mode)

Alarm Patient alarm Open Circuit, APNEA, Check Cannula, Leakage High, Total Leakage high

SpO2 High/Low, PI High/Low, PR High/Low, RR High/Low, FiO2 High/Low,

Flow High/Low, PIP High/Low, PEEP High/Low,

SpO₂ Probe off Patient, V delivery High/Low, Insufficient Leak Alarm.

MSF-730-027(Rev,01) 1/4 MEKICS Co., Ltd.


HFT700 GLOBAL向
Document No.: WQ32-20-L003 (Rev.3.6)
Date: June 22. 2021

System alarm AC Loss, Check Blower, O2 inlet pressure High, Blower temp High,

O2 Supply Fail, No Chamber, Heat-Plate Fail, Heat-Plate Temp. Sensor Fail

Heat-Wire is Disconnected, Airway Temp. Sensor is Fail,

Chamber Temp. Sensor is Fail, Air Temp. High, Airway Temp Low

Information alarm Check SpO₂ Probe

□ Alarm Setting Parameters (MAIN)


Flow (Inspiratory flow rate) High : Low setting value +1 ~ 65 lpm , Low : 0 ~ High setting value -1

PIP(Peak Inspiration Pressure) High : IPAP(CPAP) Setting +1 ~ 43 cmH2O, Low : 0 ~ IPAP(CPAP) setting-1

High : IPAP(bi-level) Setting +1 ~ 50 cmH2O, Low : 0 ~ IPAP(bi-level) setting-1

PEEP/CPAP High : EPAP(CPAP) Setting+1 ~ 22 cmH2O, Low : 0 ~ EPAP(CPAP) setting-1

FiO2 % High : Low setting value +1 ~ 100 % , Low : 18 ~ High setting value -1

RR(Respiration Rate) High : Low setting value +1 ~ 150 /min, Low : 0 ~ High setting value -1

Airway Temperature High : Low setting value +1 ~ 44℃, Low : 0 ~ High setting value -1

Apnea Off , 10 ~ 60 sec (adjustable increment : 1 sec)

SpO2 High : Low setting value +1 ~ 100 %, Low : 0 ~ High setting value -1

PI High : Low setting value +1 ~ 19 %, Low : 0.03 ~ High setting value -1

PR(Pulse Rate) High : Low setting value +1 ~ 250 / min, Low : 0 ~ High setting value -1

High Leak. High : 10 ~ 70 lpm

V delivery High : Low setting value +1 ~ 2500 / ml, Low : 50 ~ High setting value -1

□ Humidifier
Humidification level 1~5

Advanced humidification control system based on set level, flow rate, room

temperature and humidity

Dewpoint temperature HF mode 34 ~ 37 ℃ , CPAP & Bi-Level mode : OFF, 31 ~ 37 ℃

Humidification capacity AH 20 ~ 40 mg / L

(Depending on the operating environment, flow setting and humidification level)

Temperature control using 3-channel temperature sensor (Max. 95 ℃)

Auto-Off electrical switch (at 105 ℃)

□ Key, Indicators
Lock Touch Screen Lock

Alarm silence Mute audible alarms for 2 minutes

□ Electrical Requirements
Power AC : 100~240 V 50/60 Hz Free voltage , Fuse Capacity 250 V / 2.5A

Power Consumption Max 350 VA (with humidifier)

External Battery Li-ion Battery (DC: 10.89 V, 10,400 mAh)

Battery charging time is approximately 5 hours. 

□ Communication
RS232 Labview program connection using PC

Wi-Fi CMS

□ Physical Characteristics
Weight 4.5 Kg

Dimensions 204 x 338 x 204 (Width x Depth x Height) mm

Weight (With 8.4 External Power Supply) 7.5 Kg

Dimensions (With 8.4 External Power Supply) 204 x 338 x 254 (Width x Depth x Height) mm

□ Data save & S/W Upgrade


Micro SD card Main S/W upgrade, Data save

RS232 PNEU S/W upgrade using PC and dedicated JIG

□ Language
English / Korean

MSF-730-027(Rev,01) 2/4 MEKICS Co., Ltd.


HFT700 GLOBAL向
Document No.: WQ32-20-L003 (Rev.3.6)
Date: June 22. 2021

□ Oxygen inlet port


High pressure O2 DISS male connector , Input pressure range : 35 ~ 90 psi (241.3 ~ 620.5 kPa)

Low flow O2 Input Pressure : < 8.7 Psi (60 kPa)

MSF-730-027(Rev,01) 3/4 MEKICS Co., Ltd.


HFT700 GLOBAL向
Document No.: WQ32-20-L003 (Rev.3.6)
Date: June 22. 2021

□ Environmental Requirements
Atmospheric condition 700 ~ 1060 mbar

Altitude Up ~ 3000 m

Storage Temperature -25 ~ 70 °C

Humidity 0 ~ 95 % (Non-condensing)

Operating Temperature 5 ~ 40 °C

Humidity 10 ~ 90 % (Non-condensing)

MSF-730-027(Rev,01) 4/4 MEKICS Co., Ltd.