You are on page 1of 2

FORMAT SPESIFIKASI

I DATA ADMINISTRASI (*
JENIS PRODUK : CONTINUOUS VENTILATOR
NAMA PRODUK : MEK Intensive Care Portable Ventilator
MODEL/TIPE PRODUK : MTV 1000
NO. REGISTRASI DAN MASA BERLAKU : KEMENKES RI AKL 20403310085/ 15 Mei 2024
NAMA SOLE AGENT : PT. Permana Putra Mandiri
ALAMAT SOLE AGENT : JL. Taman Mini Indonesia Indah (TMII)
Pintu II atas No.43 Cipayung
Jakarta Timur - Indonesia
PHONE/ FAX. / E-MAIL : ±62-21 841 5976/ ±62-21 840 7865/ info@ptppm.id
NAMA PABRIK : MEKICS CO., LTD., Korea
ALAMAT PABRIK : Industrial Zone 861-10 Taegye-Dong, Chunchon-Si
Gangwon-Do-Korea 200-944

II DATA SPESIFIKASI (*
INFORMASI PRODUK PRODUK

COUNTRY OF ORIGIN
FACTORY OF ORIGIN
OUTSOURCE EXTERNAL MANUFACTURE (OEM)
PRODUCT'S REGISTRATION (AKL/AKD)
PRODUCT'S CERTIFICATION
FDA Clearance
CE Mark (MDD)
Certificate ISO 13485
Certificate ....
Short Form Specifications Transportable Ventilator [ICU(Child to Adult)]
7" Color TFT, Touch Screen & Knob
Mode Dual Limb PACV, PSIMV,VACV, VSIMV, PRVC, AUTO
Mode Single Limb PACV, PSIMV,VACV, VSIMV, Spont+/CPAP, PRVC, AUTO, O2 Stream
Optional SpO2, EtCO2
(* informasi produk di isi sesuai dengan yang dimiliki oleh produk
tersebut

Detalil Specifications
Ventilator Data (Setting Parameters)
Predicted Body weight 10 to 150kg (Volume)
1 to 150kg (Pressure)
Tidal Volume 50 to 2500 ml
Inspiratory pressure 5 to 80 cmH2O
Pressure support 0 to 60 cmH2O
Respiratory rate 2 to 120 bpm
Inspiratory time 0,2 to 9,9 sec
Enable inhalation trigger(En-sense) 10 to 80% of set tidal volume
Exhalation trigger Sensitivity(Ex-sense) 10 to 80% of exhaltion time
Inspiratory Pressure limit 8 to 80 cmH2O in PRVC and O2 Stream
Trigger sensitivity : Pressure trigger OFF, 0.5 to 20 cmH2O
Trigger sensitivity : Flow trigger OFF, 0.5 to 20 lpm
FiO2% 21% to 100%
BTPS OFF, MANUAL DRY, MANUAL HUMIDI
MASK ON / OFF
Display Data
Parameters Setting parameters/patient status parameters/Alarm status / I:E ratio
Graphic Waveform Pressure-Time, Flow-Time, Volume-Time
Alarm Airway pressure , Tidal volume , Minute volume, Respiratory rate
O2%, Airway leak, Circuit open, Battery indicator
O2 supply pressure failure, Ventilation in-operation
Alarm Setting Parameters (MAIN)
High tidal volume (Expiratory volume) OFF, 20 to 2500 ml
Low tidal volume (Expiratory volume) OFF, 10 to 2400 ml
High minute volume (Expiratory minute volume) 0.5 to 50 lpm
Low minute volume (Expiratory minute volume) 0.0 to 49 lpm
High respiration rate 2 to 150 bpm
Low respiration rate 0 to 148 bpm
Low peak airway pressure OFF, 0 to 110 cmH20
High O2 % OFF, 0 to 100 %
Low O2 % OFF, 0 to 80 %
Airway leak OFF, 10 to 500 ml
O2 supply pressure Fail
Ventilator in-operation
(Option, SPO2)
High SpO2 OFF, 52 to 100 %
Low SpO2 51 to 100 %
High pulse rate (SpO2) 26 to 250 bpm
Low pulse rate (SpO2) 25 to 249 bpm
(Option, EtCO2) 0.0 to 10.0 %
High EtCO2 OFF, 0.0 to 9.9 %
Low EtCO2 0.0 to 10.0 %
High inspired CO2 OFF, 0.0 to 9.9 %
High respiration rate (EtCO2) 2 to 149 Rpm
Low respiration rate (EtCO2)
Key, Indicators
Inspiration hold Measures patient's lung compliance and
Elasticity, Time constant
Key lock Key & LCD Touch Screen Lock
Graph freeze
Alarm silence Mutes audible alarms for 2 minutes
Alarm reset Clears visual indicators and messages
Electrical Requirements
Power AC : 100-230 VAC, 50/60 Hz (Free voltage)
DC : 10.8V 8.8A, Internal battery
Provides Max. 180 minutes of normal operation

Physical Characteristics
Weight 6.8 kg (main unit)
25 x 15.6 x 37.1 (Width x Depth x Height) cm
Dimensions

Environmental Requirements
Atmospheric condition 700 to 1060 mbar (10.2 ~ 15.4 psi)
Altitude -443 ~ 3048 m (-1350 ~ 10,000ft )
Height Up to 6560 m (20,000 ft )
Conditions of use 10 to 40 ℃ (50 ~ 104 ℉)
Relative humidity 10 to 85 %
Pneumatic Requirements
Oxygen inlet supply pressure 2.46 to 6.32 kgf/cm2 / 2.4 to 6.2 bar (35 to 90 psig)

Aksesoris
Test Lung Adult
Selang O2 High Pressure
Breating Circuit (Disposible)

INFORMASI SPESIFIKASI TAMBAHAN LAIN YANG ADA


(*Spesifikasi di isi sesuai dengan produk anda sendiri
Tidak semua kolom harus diisi, disesuaikan dengan spesifikasi
produknya

III INFORMASI LAINNYA


Garansi : Garansi Unit 1 Tahun
Training : ........................................................................
Spare Part : ........................................................................
Dan Lain-Lain : ........................................................................