You are on page 1of 6

Specifications & more

HumaCount 80TS
> System Specifications
> Reagents
> Accessories
> Consumables

System Overview REF 16420/80


Analyzer type Automatic hematology Analyzer
Reagent system Closed
# of Parameters 22 parameter, 3-part differential of WBC
Throughput 80 samples / hour
Barcode reader External 2D barcode reader (optional)
Reaction / reading system Impedance method (2 chambers)
LED 540 nm for HGB measurement
Languages available English, French, German, Indonesian,
Italian, Portuguese, Russian, Spanish, Czech, Polish
User interface Color capacity touch screen, 800 x 480 -pixels,
LED back light
HumaCount 80TS
Automatic Hematology System


System Specifications
Tests Parameters WBC, LYM, MID, GRA, LYM%, MID%, GRA%, HGB, RBC, HCT, MCV, MCH,
MCHC, RDWcv, RDWsd, PLT, PCT, MPV, PDWcv, PDWsd, P-LCC*, P-LCR*
Calibration 3 and 7 measurement automatic
Manual (factors) calibration of WBC, HGB, RBC, PLT, MCV (or HCT), RDW, MPV
Independent calibration of pre-diluted mode
Veterinary mode Yes
Self check Yes
Flags / messages Interpretative messages, (diagnostic flags)
Warning flags


Sample Sample type Whole blood, diluted blood
Sample vessels Open tube system with automatic sample rotor
Standard EDTA tubes
Sample positions 1
Sampling volume 25 µl of whole blood or in pre-diluted mode
Dilution Yes, internal


Reagent Reagent type Diluent
Lyse
Cleaner

Liquid Handling Liquid transportation Dilutor pumps, vacuum


Liquid level sensors Yes, internal


Reading Light source Green LED for HGB
Wavelength pre-installed 540 nm
Detector Light to frequency converter
Aperture diameter 80 μm (RBC/PLT)
100 μm (WBC)

* research only

2 of 6

Data Processing Memory for QC, records including flags and histograms
Memory capacity 10.000 results, including RBC, PLT WBC 3-part histograms & full patient data
Reports for Result, QC, self check, calibration
Quality control module Monthly report and 24 month overview Levey-Jennings charts
Separate QC database (2x3 levels), 24 lots
Max. number of control levels 24
Warnings Range, linearity, errors
Printer Built-in thermal printer, 58 mm wide roll paper, full report with histograms
External Printers via USB using printer language PCL-3.
LIS HL7
Connectors 4x USB, 1x USB-B, Ethernet


General Physical dimensions (W x D x H) Instrument without any components: 32 x 26 x 36.5 cm
Space required for routine use: 52 x 46 x 50 cm
Packaging: 50 x 41 x 59 cm
Weight: Gross: 12.5 kg, net: 12 kg
Electrical requirements External auto-ranging power unit for 100...120 or
200...240 Vac, 50–60Hz, 60 VA
Environmental Operating: temperature: 15…30°C, humidity: < 80 % non-condensing
Storage: temperature: 15…30°C, humidity: < 80 % non-condensing
Altitude: up to 3,500 m (with special software)
Wash / waste tank Waste Container 20 L

Characteristics

Parameter: Accuracy Reproducibility Carry-over sample Test range Unit


(CV) to sample
WBC 3% 3% <1% 4.0-20.0 103/μl
RBC 3% 2% <1% 4.0-15.0 106/μl 
HCT 3% 3% <1% 25.0-50.0 %
MCV 2% 1% N/A 60-100 fl
HGB 2% 2% <1% 9-16 g/dl
PLT 5% 5% <3% or <20 20-2000 103/μl

3 of 6
HumaCount 80TS
Automatic Hematology System

Scope of Supply Unit/Size REF


HumaCount 80 TS
1 16420/80
Packlist
External Power Supply and Cable 1 16420/168
Power Safety Socket 1 18969
Adapter EU to USA 1 18967
Adapter EU to UK 1 18968
Reagent Tubing Kit 1 16420/622
1 x Diluent Tube (green)
1 x Lyse Tube (yellow)
1 x Cleaner Tube (blue)
2 x Waste Tube (red)

Reagent Tubing Kit 1 x Cleaning Tube Kit
4 x Caps for Reagent Containers
Thermal Printer Paper (2 rolls) 1 16420/621
Tube Adapter I 1 16420/180
Tube Adapter II 1 16420/181
Tube Adapter III 1 16420/323
User Manual 1 16420/801
Waste Container 20 L + Box 1 ……….

Obligatory Items Service Kits REF


Need to be ordered with the first instrument:
HumaCount 80TS Starter Spare Part Kit (Sufficient for up to 5 instruments) 16420/260


Need to be ordered with each instrument:
Lubrication Kit (for HumaCount, HumaCount Plus, HumaCount 30TS & 80TS, HumaCount 5L) 16420/12


System Reagents REF
HC-Diluent (20 L) 17400/11
HC-Lyse CF (2 x 1 L) (Cyanid-free reagent) 17400/22
HC-Cleaner (1 L) 17400/31
HC-Regular (enzymatic cleaner) (50 ml) 17400/55
HC-Acute (hypochlorite cleaner) (50 ml) 17400/56

4 of 6
Obligatory Items
Controls (Strongly recommended) REF
HC-CONTROL (3 x 2.5 ml) 17400/40
HC-Calibrator (1 x 2.0 ml) 17400/50
As with all other blood control products on the market, the hematology controls and calibrators have
a limited shelf life. As a result, controls and calibrators are available only on the basis of a standing
order in accordance with a fixed delivery schedule. Controls and Calibrators are delivered as seperate shipment
under cooling conditions.
HC-CONTROL (Level 1, 2, 3)


Accessories REF
UPS - Power Supply Unit 0.8 KVA / 230 VAC 18961
Use of an online UPS with this instrument is obgligatory. If this item is purchased locally, the specifications
must be approved by HUMAN technical service prior to the HumaCount 80TS delivery.

Optional Items Consumables REF


Thermal Printer Paper (57 mm) 5 pcs. 18144/5

Accessories REF
Solar Panel 18965/100
Portable Battery System (possible battery operation for at least 6 hours) 18965/220
2D Barcode Reader 16430/11
Tube Adapter I 16420/180
Tube Adapter II 16420/181
Tube Adapter III 16420/323

Tube Adapter I Tube Adapter II Tube Adapter III

5 of 6
16420/80/2020-07 © 2020 HUMAN

Legal statement
While Human is making every effort to include accurate and up-to-date information, we make = this item is classified as dangerous goods

no representations or warranties, express or implied, as to the accuracy or completeness of the

information provided in this document and disclaim any liability for the use of it.

HUMAN Gesellschaft für Biochemica und Diagnostica mbH


Max-Planck-Ring 21 · 65205 Wiesbaden · Germany
Tel. +49 6122-9988-0 · Fax +49 6122-9988-100 · e-mail: human@human.de · www.human.de