The following relationship must be satisfied in
TT systems: Re: log $25
where: R,is the earth resistance of the esrth plate
Gin ohm;
14,18 the highest nominal operating residual cur-
rent of the RCDs that for protection of the system
Ginamperes).
In group 1 medical locations, the standard re-
uires protection only of the terminal circuits
that supply sockets outlets with a rated current
of up to 32 A, through the use of residual current.
device with |, $ 20 mA, although residual current
‘ype protection of allthe circuits is desirable, In
group 2 medical premises its mandatory forall
circuits that are not powered by a medies! IT Sy
tem to be protected by RCDs with, ¢ 30mA (un-
less those circuits are supplying power to fixed
devices positioned at a height above 2.5 mand
which cannot enter the patient environment),
Some loads, such as uninterruptable power sup-
plies (UPS), personal computers, printers, electro-
‘medical equipment, for example devices for com-
puterized axial tomography or magnetic
resonance (RM) etc. incorporate electronic cir-
cuits which, in the event of an earth fault, cause
currents with continuous components that can
compromise the operation of the normaldiffer-
ential devices of the AC type for protecting power
supply ctrcults (the ingirect contact currents are
not detected by the toroidal transformer,
This is why group 1 and group 2rooms are
obliged, depending on the type of leakage cur-
rent, to use type ARCDs, which can also intervene
with pulsating unidirectional leaking currents or
‘type B RCDs also capable of intervening with uni-
directional pulsating and continuous leakage cur-
rents.
In the case of power supply via three-phase UPS.
the product standard requires protection to be
achieved by means of RCDs of type 8.
‘Aunidirectional pulsating current isa current that
assumes a value no greater than 6 mA for an in-
terval of at least 150° of each period of the rated
frequency (at $0 Hz: 8.33 ms).
‘Types of REDS
Symbol ype “application
~ « ereasing soy
FRA 6 asopesteatins, orentng sony
cornseby mepetertunavectona put
+b) Supplementary equipetential bonding
IEC 60364.7-710 prescribes the implementation
‘of main equipotential connections, at the base of
‘each building, in order to guarantee the equipo-
tentiality of all the extraneous conductive parts
centering the same building, and of supplemen-
tary equipotential connections in the environ-
ments at greatest electrical risk
Group 1 and 2 medical locations are expressly cov-
‘ered by this prescription because the differences
of potential between conductive parts and extra-
neous conductive parts and therefore the cur-
rents that could affect a patient in contact with
such conductive parts are limited to the maxi
mum with the additional equipotential connec-
tions.
Each room for medical use must therefore be
‘equipped with its own equipotential bonding bus
bar to which the electrical devices and all the me-
tallic parts that can close an electrical circuit to
‘arth must be connected, zo that if an indirect
contact of a device (even external to the prem-
ses) occurs all the conductive parts and the ex-
traneous conductive parts assume almost the
same potential instantaneously (no significant
difference of potential between the devices ac-
ccessible to the patient),
For group 2 medical locations, the equipotential
bonding bus bar resistance shall notexceed 0.2 0.