Scanning Electron Microscope (SEM)
Description of Module
Subject Name Physics
Paper Name Nanoscience and Technology-II
Module Name/Title Scanning Electron Microscope (SEM)
Module Id

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
Content of this Unit
1. INTRODUCTION
1.1 History
1.2 Why Electron Microscope?
1.3 Scanning Electron Microscope (SEM)
1.4 Modes of Imaging
2. Construction and Working Principle
2.1 Structure of Instrument
2.2 Electron Gun
2.3 Condenser and Objective Lenses
2.4 Sample Stage
2.5 Secondary Electron Detector
2.6 Imaging and Recording of Signal
2.7 Vacuum Process
3. Focusing of SEM
4. Magnification of SEM
5. Elemental Analysis
6. Summary

Learning Outcomes

After studying this module, you shall be able to learn

 The basic principles of SEM operation;


 The origin of images and their formation;
 Procedures for collecting and interpreting images;
 Preparing samples for investigation
 Result of scanning electron microscopy
 Fields that rely on electron microscopy

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
INTRODUCTION

Since the Scanning Electron Microscope (SEM) was first commercialized about 40 years ago, the SEM has
shown a remarkable progress. Now, many types of SEMs are being used, and their performance and functions
are greatly different from each other. To utilize these different SEMs, it is essential to recognize their features,
as well as to understand the reasons for the contrast of SEM images. Thus, this document material is aimed at
helping to understand the basics of the SEM, including the instrument principles, specimen preparation and
elemental analysis.

The combination of higher magnification, larger depth of field, greater resolution, compositional and
crystallographic information makes the SEM one of the most heavily used instruments in academic/national lab
research areas and industry.

1.1 HISTORY

The development of a SEM began a few years after the invention of a TEM by Ruska in 1931. But the
commercialization of the SEM required about 30 years. Figure1 shows the history of the initial development
stage of the SEM.

Fig 1: History of the initial development stage of the SEM.

The first SEM image was obtained by Max Knoll, who in 1935 obtained an image of silicon steel showing
electron channeling contrast. Further pioneering work on the physical principles of the SEM and beam specimen
interactions was performed by Manfred von Ardenne in 1937, who produced a British patent but never made a
practical instrument. The SEM was further developed by Professor Sir Charles Oatley and his postgraduate
student Gary Stewart and was first marketed in 1965 by the Cambridge Instrument Company.

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
1.2 WHY ELECTRON MICROSCOPE?

Electron microscopes are scientific instruments that use a beam of energetic electrons to examine objects on a
very fine scale. Electron microscopes were developed due to the limitations of Light Microscopes which are
limited by the physics of light. This required 10,000x plus magnification which was not possible using current
optical microscopes.

Electron wave is a unique medium that can be used in imaging. By accelerating the electrons into high energy
beam (via high voltage), the wavelength thus created is far shorter than white light. For example, for an electron
beam produced from a 20 kV gun, the wavelength is only 1240.7/20,000 (eV) = 0.06 nm = 0.6 Å, corresponding
to a resolution limit of λ/2 = 0.3 Å --- theoretically, it can be used to image a species as small as 0.3 Å. Most
atoms are in size of 2-3 Å.

1.3 SCANNING ELECTRON MICROSCOPE (SEM)

Fig 2: Scanning Electron Microscope (SEM)

The Scanning Electron Microscope (SEM) is used for observation of specimen surfaces. When the specimen is
irradiated with a fine electron beam (called an electron probe), secondary electrons are emitted from the
specimen surface. Topography of the surface can be observed by two-dimensional scanning of the electron
probe over the surface and acquisition of an image from the detected secondary electrons.

Characteristic Information of sample using SEM


Topography
The surface features of an object or "how it looks", its texture; direct relation between these features and
materials properties
Morphology
The shape and size of the particles making up the object; direct relation between these structures and materials
properties
Composition
The elements and compounds that the object is composed of and the relative amounts of them; direct
relationship between composition and materials properties
Crystallographic Information
How the atoms are arranged in the object; direct relation between these arrangements and material properties

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
1.4 MODES OF IMAGING

The SEM image appears as if we observe an object with the naked eye, we may intuitively understand the
features of the object. However, the SEM image often produces a contrast that is difficult to explain. To fully
understand the contrast of the SEM image, we must understand the principle of the formation of the SEM image.

Fig 3: Emission of various electrons and electromagnetic waves from the sample

Originally microscopy was based on the use of the light microscope and could provide specimen resolutions on
the order of 0.2 microns. To achieve higher resolutions, an electron source is required instead of light as the
illumination source, which allows for resolutions of about 25 Angstroms. The use of electrons not only gives
better resolution but, due to the nature electron beam specimen interactions there are a variety of signals that can
be used to provide information regarding characteristics at and near the surface of a specimen

In scanning electron microscopy visual inspection of the surface of a material utilizes signals of two types,
secondary and backscattered electrons. Secondary and backscattered electrons are constantly being produced
from the surface of the specimen while under the electron beam however they are a result of two separate types
of interaction. Secondary electrons are a result of the inelastic collision and scattering of incident electrons with
specimen electrons. They are generally characterized by possessing energies of less than 50 eV. They are used to
reveal the surface structure of a material with a resolution of ~10 nm or better

Backscattered electrons are a result of an elastic collision and scattering event between incident electrons and
specimen nuclei or electrons. Backscattered electrons can be generated further from the surface of the material
and help to resolve topographical contrast and atomic number contrast with a resolution of >1 micron. While
there are several types of signals that are generated from a specimen under an electron beam the x-ray signal is
typically the only other signal that is used for scanning electron microscopy. The x-ray signal is a result of
recombination interactions between free electrons and positive electron holes that are generated within the
material. The x-ray signal can originate from further down into the surface of the specimen surface and allows
for determination of elemental composition through EDS (energy dispersive x-ray spectroscopy) analysis of
characteristic x-ray signals.

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
2. CONSTRUCTION AND WORKING PRINCIPLE

2.1 Structure of Scanning Electron Microscope (SEM)


The SEM requires an electron optical system to produce an electron probe, a specimen stage to place the
specimen, a secondary-electron detector to collect secondary electrons, an image display unit, and an operation
system to perform various operations. The electron optical system consists of an electron gun, a condenser lens
and an objective lens to produce an electron probe, a scanning coil to scan the electron probe, and other
components. The electron optical system (inside of the microscope column) and a space surrounding the
specimen are kept at vacuum.

Fig 4: Basic construction of a SEM.

2.2 Electron Gun

The electron gun produces an electron beam. this is a thermionic emission gun (TE gun). Thermoelectrons are
emitted from a filament (cathode) made of a thin tungsten wire (about 0.1 mm) by heating the filament at high
temperature (about 2800K). These thermoelectrons are gathered as an electron beam, flowing into the metal
plate (anode) by applying a positive voltage (1 to 30 kV) to the anode. If a hole is made at the center of the
anode, the electron beam flows through this hole. When you place an electrode (called a Wehnelt electrode)
between the cathode and anode and apply a negative voltage to it, you can adjust the current of the electron
beam. At this time, the electron beam is finely focused by the action of the Wehnelt electrode. The finest point
of the beam is called the crossover, and this is regarded as an actual electron source with a diameter of 15 to 20
μm. The TE gun, explained here, is most generally used. An LaB6 single crystal is also used as a cathode, but it
requires a higher vacuum because of its high activity. Other electron guns are the field-emission electron gun
(FE gun) or the Schottky-emission electron gun (SE gun)

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
2.3 Condenser and Objective lenses

An electron microscope generally uses a magnetic lens. When you pass a direct electric current through a coil
wound electric wire, a rotationally-symmetric magnetic field is formed and a lens action is produced on an
electron beam. To make a strong magnetic lens (with a short focal length), it is necessary to increase the density
of the magnetic line. The surroundings of the coil are enclosed by yokes so that part of the magnetic field leaks
from a narrow gap. A portion with a narrow gap, called “polepiece,” is fabricated with a high accuracy. The
main feature of the magnetic lens is that when you change the current passing through the coil, the strength of
the lens is also changed. This is not achieved by an optical lens.

The “aperture” is placed between the condenser lens and objective lens. The “aperture,” made of a thin metal
plate, has a small hole. The electron beam, which passed through the condenser lens, illuminates this aperture-
plate. The aperture allows a part of the electron beam to reach the objective lens. If the excitation of the
condenser lens is increased, the electron beam greatly broadens on the aperture and therefore, the number of the
electrons (amount of probe current) reaching the objective lens is decreased. To the contrary, if the excitation of
the condenser lens is decreased, the el electron beam does not broaden very much and therefore, most of the
electrons pass through the aperture and many electrons reach the objective lens. That is, the adjustment of the
excitation of the condenser lens enables you to change the electron-probe diameter and the probe current.
However, even if the excitation of the condenser lens is infinitely increased, the diameter of the electron probe
does not become infinitely small
The objective lens is used for focusing, and this lens is a very important lens that determines the final diameter
of the electron probe. If the performance of the objective lens is not good, an optimally fine electron probe
cannot be produced despite all of the efforts before the action of the objective lens. Thus, it is crucial to make
the objective lens with the best performance.

2.4 Sample Stage

In general, the specimen is observed at a high magnification in an electron microscope. Thus, a specimen stage,
which stably supports the specimen and moves smoothly, is required. The specimen stage for a SEM can
perform the following movements: horizontal movement, vertical movement, specimen tilting, and rotation. The
horizontal movements are used for the selection of a field of view. While the vertical movement provides the
change of image resolution and the depth of focus.

2.5 Secondary Electron Detector

The secondary electron detector is used for detecting the secondary electrons emitted from the specimen. A
scintillator (fluorescent substance) is coated on the tip of the detector and a high voltage of about 10 kV is
applied to it. The secondary electrons from the specimen are attracted to this high voltage and then generate light
when they hit the scintillator. This light is directed to a photo-multiplier tube (PMT) through a light guide. Then,
the light is converted to electrons, and these electrons are amplified as an electric signal. A supplementary
electrode, called the collector, is placed before the scintillator. In general, in order to help the scintillator acquire
secondary electrons, a few hundred volts is applied to this collector. By changing this voltage, you can control
the number of secondary electrons to be collected. This type of the detector was originally developed by
Everhart and Thornley, so this detector can be called the E-T detector. Many SEMs incorporate this detector in
8

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
the specimen chamber; however, when a SEM is equipped with a strongly excited objective lens for higher
resolution. Secondary electron detector is placed above the objective lens and secondary electrons are detected
by utilizing the lens magnetic fields. This detector is often called the TTL (Through The Lens) detector.

2.6 Imaging and Recording of Signal

The output signals from the secondary electron detector are amplified and then transferred to the display unit.
Since the scanning on the display unit is synchronized with the electron-probe scan, brightness variation, which
depends on the number of the secondary electrons, appears on the monitor screen on the display unit, thus
forming a SEM image. A cathode-ray tube (CRT) was used for many years as a display unit; however in recent
years, a liquid-crystal display (LCD) has been widely used. In general, the scan speed of the electron probe can
be changed in several steps, An extremely fast scan speed is used for observation and a slow scan speed is used
for acquisition or saving of images.

To record an SEM image, in the past, the SEM image appearing on the CRT was photographed with a camera.
But recently, the image has been recorded in a digital format (electronic file). This is because it is now difficult
to get a high-resolution CRT and there are many advantages of electronic file and it is easier to process images
and convenient to send or receive image information. An image format with 1M pixels is generally used for the
electronic file.

2.7 Vacuum Process

The inside of the electron optical system and the specimen chamber must be kept at a high vacuum of 10-3 to
10-4 Pa. Thus, these components are evacuated generally by a diffusion pump. If a user desires an oil-free
environment, a turbo molecular pump may be used. When a SEM incorporates an FE gun (explained later), a
sputter ion pump is used because the FE gun needs an ultrahigh vacuum.

To exchange a specimen, either of two methods is applied. One vents the entire specimen chamber at the time of
specimen exchange. The other uses a specimen pre-evacuation chamber (airlock chamber) while keeping a high
vacuum in the specimen chamber.

3. FOCUSING OF SEM

In the observation of a specimen with a substantial depth, if the focus is adjusted to the top side, the bottom side
may be out of focus. In such a case, if the range between upper and lower image blur is large, it is said that “the
depth of focus is large.” Whereas if the range between upper and lower image blur is small, it is said that “the
depth of focus is small.”, when the electron probe is considerably parallel (aperture angle is small), the image
stays in focus even if the focus is changed by a large amount. Whereas when the electron probe is substantially
angular (aperture angle is large), the image goes out of focus even if the focus is only slightly changed. In the
case of an optical microscope (OM) where the probe scanning is not used for imaging, when the angle subtended
by the objective lens from the specimen (aperture angle) is small, the depth of focus is large. Whereas when this
angle is large, the depth of focus is small. Note that even when the image is blurred, this cannot be seen at a low
magnification. However, when the magnification is increased, the image blur is found to appear. That is, the
depth of focus is changed also by the magnification.

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
Figure 5 shows a difference of the depth of focus between the SEM and OM. Although a stereoscopic
microscope provides an image with a relatively large depth of focus between OMs, you can obtain a much larger
depth of focus wi the SEM. This is because the aperture angle of the electron probe in the SEM is much smaller
than that of the objective lens in the OM. Note that the depth of focus for the SEM is different depending on the
observation conditions.

Fig5: OM image and SEM image of the same field of view.

The OM and SEM images of a fractured surface of a screw. This fractured surface has large irregularity, leading
to a fact that only a small part of the surface is actually focused with the OM. But, due to the large depth of
focus of the SEM, the entire observed surface is in sharp focus

4. MAGNIFICATION OF SEM

Magnification in a SEM can be controlled over a range of up to 6 orders of magnitude from about 10 to 500,000
times. Unlike optical and transmission electron microscopes, image magnification in the SEM is not a function
of the power of the objective lens. SEMs may have condenser and objective lenses, but their function is to focus
the beam to a spot, and not to image the specimen. Provided the electron gun can generate a beam with
sufficiently small diameter, a SEM could in principle work entirely without condenser or objective lenses,
although it might not be very versatile or achieve very high resolution. In a SEM, as in scanning probe
microscopy, magnification results from the ratio of the dimensions of the raster on the specimen and the raster
on the display device. Assuming that the display screen has a fixed size, higher magnification results from
reducing the size of the raster on the specimen, and vice versa. Magnification is therefore controlled by the
current supplied to the scanning coils, or the voltage supplied to the deflector plates, and not by objective lens
power.

10

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
5 ELEMENTAL ANALYSES

Analysis by EDS

The Energy Dispersive X-ray Spectrometer (EDS) is used to analyze characteristic X-ray spectra by measuring
the energies of the X-rays. when the X-rays emitted from the specimen enter the semiconductor detector,
electron-hole pairs are generated whose quantities correspond to the X-ray energy. Measuring these quantities
(electric current) enables you to obtain the values of X-ray energy. The detector is cooled by liquid nitrogen, in
order to reduce the electric noise..

Analysis by WDS

The Wavelength Dispersive X-ray Spectrometer (WDS) is used to analyze characteristic X-ray spectra by
measuring the wavelengths of the X-rays. First, the X-rays emitted from the specimen hit the analyzing crystal,
next, this crystal diffracts the X-rays, and finally the X-rays enter the detector and their wavelengths are
measured. The analyzing crystal and detector must move on a so-called “Rowland circle” with a constant radius.
In order to cover (measure) all of wavelengths, a driving mechanism for multiple analyzing crystals is needed,
requiring a lot of time to acquire all of the corresponding X-ray spectra.

Analysis of Non-conductive Specimen

When analyzing a nonconductive specimen, metal coating is required as in SEM observation. However, we must
use a metal that is different from that may be contained in a specimen. When we want to detect light elements, a
coated film must be thin enough because a thick heavy-metal coated film may prevent the emission of X-rays
from light elements.

If we try to analyze a nonconductive specimen without coating, this specimen is analyzed with a low
accelerating voltage caused by to the charging effect. Thus, some problems may occur; for example,
characteristic X-rays with high excitation energies cannot be detected, and the accuracy of quantitative analysis
is degraded. In addition, positional shift of the incident electron probe may occur when line analysis or X-ray
mapping is performed.

11

Nanoscience and Technology-II


Physics
Scanning Electron Microscope (SEM)
6 SUMMARY

In this module you study

 SEM, has been presented as an powerful imaging technique

 The electrons interact with the atoms that make up the sample producing signals that contain
information about the sample's surface topography, composition and other properties such as electrical
conductivity.

 During SEM inspection, a beam of electrons is focused on a spot volume of the specimen, resulting in
the transfer of energy to the spot.

 To produce the SEM image, the electron beam is swept across the area being inspected, producing many
such signals. These signals are then amplified, analyzed, and translated into images of the topography
being inspected. Finally, the image is shown on a CRT.