technical note
Mean: 0.0003
SD: 0.00015
Instrument Conditions
3 x SD x mo 3x0.00015 x 22 pg
Lamp EDL IDL = =
0.0044 0.0044
Wavelength 193.7 nm
= 2.3 pg
Slit 0.7 nm
Low Concentration As Signal
BOC 2.0 s
(40 µL of 0.5 µg L-1 =20 pg As)
Read 4.0 s Integrated absorbance data:
Graphite Tube End-capped THGA 0.0043, 0.0043, 0.0044, 0.0044,
0.0048, 0.0038, 0.0042, 0.0042,
Volume 40 µL
0.0044, 0.0040, 0.0042
Modifier 5 µg Pd / Mean: 0.0044
3 µg Mg(NO3)2 SD: 0.00024
Char. Mass 22 pg RSD: 5.6%
Mean: 0.0414
SD: 0.0003
RSD: 0.8%
Comments
• End-capped tube improves sensitivity, characteristic mass 22 pg vs. 40 pg for
standard THGA graphite tubes.
• Solution detection limit for 50-µL sample volume = 0.05 µg L-1.
• Mixed chemical modifier of palladium/magnesium nitrate employed.
For many samples allows pyrolysis temperatures up to 1200°C.