Professional Documents
Culture Documents
medical device 6
6.1 Introduction
Safety of a medical device is the primary concern of any national regulatory organi-
zation. After the risks associated with the medical device have been considered, the
next step is to carry out tests to ensure the safety of the product. The safety tests to be
performed depend on numerous factors such as its intended use, packaging, whether it
is sterilized, whether it is intended for one time use, and more. Safety of the device is
not only restricted to the patient’s safety, it also includes safety of the operators and
anybody who may come into contact with the medical device. Related to the previous
chapter of risk assessment, verification and validation of the product with regards to its
safety needs to be performed once it is manufactured and ready to leave the production
floor. Besides biocompatibility of the product, which needs to be considered where
applicable, other safety considerations include packaging, transportation, and compat-
ibility with other electrical appliances. It is important to appreciate that the range of
medical products is huge with the FDA listing over 2000 devices spread over 19
medical specialties. This is not inclusive of medical devices that are not included
in the list yet. It is not possible to come out with a complete list of safety tests to
be performed given the diversity of medical products. The need for tests also depends
on the risk analysis of the individual product. Figure 6.1 shows some general catego-
ries for safety tests.
Unlike performance and functional tests, safety tests are almost always based on
established standards and procedures. Identification of the relevant standards is impor-
tant to ensure that the design is compliant with regulatory requirements. Figure 6.2
gives an example of how this process can be carried out. The first step starts with
knowing the intended use, the design, its components, and the product movement,
people that come into contact with it and its life cycle. Because it is a medical device,
the next step is to look at relevant medical device standards. These standards often
reference other relevant standards that may or may not be specific to medical appli-
cation. This will often provide sufficient coverage in terms of safety test requirements.
There are numerous safety standards covering almost all conceivable designs, com-
ponents, displays, labeling, and applications. Many of these standards are not written
specifically for any medical device, however, your medical device or its component
may still be covered by them depending on your design and intended use. Thus, for
every design and component that has not been mentioned in the medical device stan-
dards, check whether there are any other related standards that could be applicable.
Each medical device is unique and is dependent on its design and intended use. This
chapter will introduce some of the general major safety considerations for medical
devices such as absence of toxic substance, biocompatibility, electrical safety,
Medical devices. http://dx.doi.org/10.1016/B978-0-08-100289-6.00006-5
© 2015 Elsevier Ltd. All rights reserved.
138 Medical devices
• Biocompatibility
• Contamination
• Mechanical
Absence of toxic Packaging and
Product • Electrical
substance transportation
• Radiation
• Noise
• Thermal
Medical device
Household
appliance Connections Storage and
standards safety standards shelf-life
Information
Software Other relevant
technology Labeling
evaluation industry standards
standards
Storage and
shelf-life
mechanical safety, sterility, and transportation. Details of these tests and other tests
that are more specific can be found in recognized standards and regulatory advisories.
Lead 0.1
Mercury 0.1
Cadmium 0.01
Hexavalent chromium 0.1
Polybrominated biphenyls (PBB) 0.1
Polybrominated diphenyl ethers (PBDE) 0.1
Contact duration
A—Limited <24 h
B—Prolonged 24 h
to 30 days Irritation or Subacute and
C—Permanent intracutaneous Systematic/ subchronic
Category Contact >30 days Cytotoxicity Sensitization reactivity acute toxicity toxicity Genotoxicity
Reproduced from ISO 10993-1:2009 with permission from the International Organization for Standardization (ISO). All rights reserved by ISO.
Table 6.3 Supplementary evaluation test for consideration
Nature of body contact Biological effect
Contact duration
A—limited <24 h
B—Prolonged 24 h to
30 days
C—Permanent Chronic Reproductive/
Category Contact >30 days Implantation Hemocompatibility toxicity Carcinogenicity development Biodegradation
Reproduced from ISO 10993-1:2009 with permission from the International Organization for Standardization (ISO). All rights reserved by ISO.
142 Medical devices
In many cases, after the biological effect to be tested has been determined using
Tables 6.2 and 6.3, specific tests and procedures will still need to be selected. Medical
devices come in many forms and the organ that comes into contact with it will differ
and it is the manufacturer’s responsibility to select appropriate tests. ISO 10993 and its
subparts contain procedures for testing but manufacturers may need to look beyond
those if they are not applicable. Most test procedures described are based on chemical
testing which typically involves extracting any leachable substances from the medical
device and using them for dosing followed by checking on its effect on the animal,
cells, or other agents. Although this standard provides a recommendation on the type
of tests to be performed, additional tests may still be warranted depending on the risk
analysis, especially for higher risk medical devices.
generally not required for absorbable medical devices or medical devices with or with-
out leachable substances if there is sufficient evidence to show that there is no risk to
reproductive development. All test procedures generally follow that from OECD for
testing of chemicals [156–158].
6.7 Implantation
The objective of implantation tests is to investigate the local effects of the medical
device at the implantation site. The history and evolution of the tissue response to
the implant and the effect of any degradation and integration will be characterized.
Tissue response to be determined as part of this test includes:
– Extent of fibrosis and inflammation;
– Tissue degeneration;
– Immune cell response in terms of numbers and distribution;
– Necrosis;
– Tissue alterations such as vascularization, bone formation, fatty infiltration, and granuloma
formation;
– Materials parameters, such as any fragmentation and degradation; and
– Quality and quantity of tissue ingrowth.
Analysis of the tissue response may be based on a scoring system where observed
reaction is given a score [159].
6.8 Hemocompatibility
For medical devices that come into contact with blood, hemocompatibility of the
device needs to be evaluated (e.g., stents, and vascular grafts). The first step is to check
whether tests need to be carried out to determine hemocompatibility. Figure 6.3 shows
the decision tree for making this determination based on ISO 10993-4. If tests need to
be carried out for the medical device, only the parts of the device that come into
144 Medical devices
Start
Blood interaction
No requirements met
Does the device contact circulating
blood directly or indirectly? Yes
Yes Is the material Same chemical
Yes Same Yes Yes Yes Same
same as in composition Same body
manufacturing sterilization
marketed and contact?
process? method?
device/ properties?
No No No No No
Figure 6.3 Decision tree to determine whether testing for interaction with blood is
necessary [160].
Reproduced from ISO 10993-4:2002 with permission from the International Organization for
Standardization (ISO). All rights reserved by ISO.
contact with blood need to be tested. The conditions for hemocompatibility testing
need to be as close to clinical usage as possible. A device designed to contact the blood
in vitro and in vivo shall be tested accordingly.
Depending on the medical device and its intended use, the method or condition
under which it comes into contact with blood may differ. The associate blood inter-
action hazard will also be different as a result. Examples of hazardous blood interac-
tions are thrombosis, coagulation, hemolysis, leukocyte interaction, and platelet
adhesion. A detailed outline of the device examples and corresponding test categories
are listed in ISO 10993-4 (Tables 6.5 and 6.6).
6.9 Biodegradation
Biodegradation of a material needs to be considered even though it is not marketed as
biodegradable. This is an important consideration for polymer products as they may
degrade during fabrication, sterilization, storage, or under environmental influence. In
particular, polymer properties can be expected to change when subjected to radiation ster-
ilization. Release of a substance, intentional or otherwise, may come from chemical reac-
tions, leaching, migration, depolymerization, or physical peeling or scaling. Risk analysis
needs to be carried out in view of the material and its usage life cycle to determine whether
a degradation assessment is necessary. ISO 10993-9 provides a framework for what needs
to be tested and documented. It does not provide specific test methodologies on how it is to
Safety testing of a new medical device
Table 6.5 ISO 10993-4:2002
Complement
Device examples Thrombosis Coagulation Platelets Hematology system
a
As stated in 3.8, thrombosis is an in vivo or ex vivo phenomenon. It is recognized that coagulation and platelet responses are involved in this process. Therefore, it is up to the manufacturer, to
decide if specific testing in the coagulation and platelet test categories are appropriate for their device.
b
Hemolysis testing only.
c
Only for aphaeresis equipment and related procedures.
Reproduced from ISO 10993-4:2002 with permission from the International Organization for Standardization (ISO). All rights reserved by ISO.
145
146
Table 6.6 ISO 10993-4:2002
Complement
Device examples Thrombosis Coagulation Platelets Hematology system
a
As stated in 3.8, thrombosis is an in vivo or ex vivo phenomenon. It is recognized that coagulation and platelet responses are involved in this process. Therefore, it is up to the manufacturer, to
decide if specific testing in the coagulation and platelet test categories are appropriate for their device.
b
Hemolysis testing only.
Medical devices
Reproduced from ISO 10993-4:2002 with permission from the International Organization for Standardization (ISO). All rights reserved by ISO.
Safety testing of a new medical device 147
be done or recommend any acceptable limits as it is dependent on the material and the
product life cycle. Protocol found in ISO 10993-13 may be used to identify and quantify
degradation products of polymeric medical devices. Acceptable limits may be deter-
mined by other tests such as ISO 10993-5 and ISO 10993-10.
Devices which come into contact with blood or cerebrospinal fluid need to be verified
for absence of pyrogen. The maximum level of acceptable endotoxin unit for these
devices is given in Table 6.7. A bacterial endotoxins test or limulus amebocyte
lysate test may be used for this verification based on the following standards:
FDA “Guideline on Validation of the Limulus Amebocyte Lysate Test as an End Product
Endotoxin Test for Human and Animal Parenteral Drugs, Biological Products, and Medical
Devices” (1987)
148 Medical devices
Another concern regarding electrical medical devices is risk associated with EMC.
The electromagnetic energy emitted by the medical device shall not interfere with its
own function or with other devices in the vicinity. A requirement under CE mark is
“Devices must be designed and manufactured in such a way as to minimize the risks of
creating electromagnetic fields which could impair the operation of other devices or
equipment in the usual environment.” Satisfying this requirement may be based on
compliance with IEC 60601-1-2. However, not all the tests mentioned in IEC
60601-1-2 are necessary. The immunity tests for class I products under CE mark
may not be necessary [164] unless interference due to electromagnetic interference
will result in an unacceptable risk. There are more than 10 separate tests for determi-
nation of EMC, such as radio frequency emission, electrostatic discharge, radiated
radio frequency fields, bursts, surges, voltage dips, short interruptions, and voltage
variations on power supply input lines.
a
Arm >120 >120 a
Safety testing of a new medical device 151
accreditation bodies. It is important to check that the laboratory performing the test
and calibration services are accredited.
FAQs
(1) What factors affect safety tests?
Safety tests to be performed depend on numerous factors such as the device’s intended
use, packaging, whether it is sterilized, whether the device is for one time use, and more.
(2) What are some of the safety concerns?
General major safety considerations for medical devices are the absence of toxic sub-
stances, biocompatibility, electrical safety, mechanical safety, sterility, storage condi-
tions, and transportation.
(3) Is there an international standard test for determining the concentration level of the toxic
substance in a material?
Currently, there are no specific international test standards used specifically for deter-
mining the concentration level of the toxic substance in a material. X-ray fluorescence
(XRF) is often used for this purpose as it is nondestructive. More accurate techniques such
as high performance liquid chromatography (HPLC) or gas chromatography/mass spec-
trometry (GC/MS) may be used for the detection of chromium and bromine compounds.
(4) How can we determine which tests need to be performed on the medical device?
The tests to be performed depend on how the medical device is intended to be used.
ISO 10993 biological evaluations of medical devices, recognized by most major national
regulatory bodies including the FDA and CE mark as the standard for selecting the bio-
logical tests necessary for assessing the safety of a medical device, can be referred to. To
facilitate manufacturers in determining the tests to be performed, ISO 10993-1 has listed
the tests to be performed based on the area of contact between the medical device and the
patient and its duration of contact.
(5) Which is the ISO for cytotoxicity test? Which is the ISO for sensitization and irritation test?
Which is the ISO for systematic/acute, subacute, subchronic, and chronic toxicity test?
ISO 10993-5 covers the in vitro cytotoxicity test procedures that are applicable to all
medical devices that come into contact with the body.
ISO 10993-10 covers the in vitro sensitization and irritation test procedures that are
applicable to all medical devices that come into contact with the body.
Procedures for systematic/acute, subacute, subchronic, and chronic toxicity tests are
described in ISO 10993-11, although the main method described is based on dosing a
chemical/leachable substance.
(6) What are the procedures for genotoxicity, carcinogenicity, and reproductive and develop-
ment toxicity tests?
All test procedures generally follow that from OECD476, OECD451, OECD421 for
testing of chemicals.
(7) Why should implantation tests be done? Which ISO should be referred to for
implantation tests?
The objective of implantation tests is to investigate the local effects of the medical
device at the site of implantation. ISO 10993-6:2007 can be referred to for more details.
(8) Is the device required to undergo hemocompatibility tests? Which ISO should be referred
to for hemocompatibility test?
For medical devices that come into contact with blood, hemocompatibility of only the
parts of the device that come into contact with the blood need to be tested. The conditions
Safety testing of a new medical device 153
for hemocompatibility testing needs to be as close to clinical usage as possible. Figure 6.3
shows the decision tree for making this determination based on ISO 10993-4.
(9) What are some of the hazardous blood interactions?
Examples of hazardous blood interactions are thrombosis, coagulation, hemolysis,
leukocyte interaction, and platelet adhesion. As such, medical devices that come into
contact with blood should undergo a hemocompatibility test.
(10) Which ISO should be referred to for biodegradation tests and documentation?
ISO 10993-9 provides a framework on what needs to be tested and documented. It does
not provide specific test methodologies on how it is to be done or recommend any accept-
able limits as it is dependent on the material and the product life cycle. Protocol found in
ISO 10993-13 may be used to identify and quantify degradation products of polymeric
medical devices. Acceptable limits may be determined by other tests such as ISO
10993-5 and ISO 10993-10.
(11) What are the recommended sterility standards?
The FDA requires devices labeled as sterile to have a SAL of 10 6 unless it is
only intended for intact skin contact, in which case, a SAL of 10 3 is recommended.
Recognized standards for sterilization are:
ANSI/AAMI/ISO 11135—Medical devices—Validation and routine control of ethyl-
ene oxide sterilization
AAMI/CDV 11137 Sterilization of health care products—Radiation
AAMI/ISO Technical Information Report (TIR) 13409 Sterilization of health care
products—Radiation sterilization.
(12) What are the recognized transport test standards?
Commonly recognized transport test standards are the ISTA and ASTM D4169.
(13) What are the electrical appliances tests for electrical medical devices?
The main reference to safety of electrical medical devices is IEC60601 “Medical
electrical equipment.”
(14) What is protective earth connections?
Protective earth connection is a connection to ground or any protective earth terminal
for protection from electrical shock.