Professional Documents
Culture Documents
Standardisation
Methods
AAC811S
2022
1
Chapter Content
1 Analytical Standards
3 Standard Addition
4 Internal Standards
5 Blank Corrections
2
Standardisation: the process of determining the relationship between the
signal and the amount of analyte in a sample.
(ACS Committee on Environmental Improvement “Guidelines for Data Acquisition and Data Quality Evaluation in
Environmental Chemistry,” Anal. Chem. 1980, 52, 2242–2249)
(3.1)
3
1. Analytical Standards
4
• Preparing a standard often requires additional reagents that are not
primary standards or secondary standards.
For ex., a suitable solvent, and additional reagents may be needed to
adjust the standard’s matrix.
These solvents and reagents are potential sources of additional analyte,
which may produce a determinate error in the standardisation.
5
Typical packaging labels for reagent grade chemicals.
Label (a) provides the manufacturer’s assay for the reagent, NaBr.
K is flagged with (*) because its assay exceeds the limits established by ACS.
Label (b) does not provide an assay for impurities, but indicates that the reagent meets
ACS specifications. An assay for the reagent, NaHCO3 is provided.
6
Preparing Standard Solutions
7
Calibrating the Signal (Stotal)
Examples:
When the signal is a mass, we determine Stotal using an analytical balance.
To calibrate the balance’s signal we use a reference weight that meets
standards established by a governing agency (e.g. NIST or ASTM).
Electronic balances often includes an internal calibration weight for routine
calibrations, as well as programs for calibrating with external weights.
If Sstd = kA Cstd
11
• When a calibration curve is a
straight-line, the slope of the line
gives kA.
0.7
Response = dependant variable = y
0.6
0.5
0.4
Abs
0.3
0.2
0.1
0
0 1 2 3 4 5 6 7 8 9
Conc / mg/l
0.3
0.2
y = 0.075x + 0.003
0.1
0
0 1 2 3 4 5 6 7 8 9
Conc / mg/l
15
The student then measured his sample to have an absorbance of 0.418 and
the blank, 0.003. He, therefore, can calculate the concentration using the
obtained calibration curve.
y = 0.0750x + 0.0029
0.7
0.6
0.5
0.4
Abs
0.3
0.2
y = 0.075x + 0.003
0.1
0
0 1 2 3 4 5 6 7 8 9
Conc / mg/l
17
The equation for a normal calibration curve for the quantitative
analysis of Cu2+ is given as follows:
18
Recall: Method of Least Squares
y = mx + c
where m = slope and c = y-intercept
n xi yi xi yi x y x y x
2
m
i i i i i
n x x
c
n xi2 xi 2 i
2
i
2
19
Correlation Coefficient used as a measure of the correlation
between two variables (x and y).
n x iyi x i yi
r
n x i
2
2
x i n y i y i
2 2
r = 1 An exact correlation between the 2 variables
21
• External standardisation allows to analyse a series of samples using a
single calibration curve.
• This is an advantage when analysing many samples.
• Thus, many of the most common quantitative analytical methods use an
external standardisation.
Limitation:
• If we expect that matrix effects are important, then we try to match the
standard’s matrix to that of the sample (matrix matching).
22
Calibration curves for an analyte in the standard’s matrix and in the sample’s
matrix. If the matrix affects kA, as is the case here, then we introduce a
determinate error into our analysis if we use a normal calibration curve.
• If we are unsure of the sample’s matrix, then we must show that matrix
effects are negligible, or use an alternative method of standardisation.
The MATRIX:
24
We can avoid the complication of matching the matrix of the
standards to the matrix of the sample by conducting the
standardization in the sample.
STANDARD ADDITION!
Assumption:
The matrix will have the same effect on the analyte in the
standard as it would on the original analyte in the sample.
25
Single Standard Addition
28
Solution
Begin by making appropriate substitutions into the previous eqn
and solving for CA.
Note that all volumes must be in the same units; thus, first covert
Vstd from 1.00 µL to 1.00 × 10–3 mL.
29
• It also is possible to make a standard addition directly to the
sample, measuring the signal both before and after the spike.
30
Example 2:
A spectrophotometric method for the quantitative analysis of Pb2+
in blood yields an Ssamp of 0.712 for a 5.00 mL sample of blood.
Solution
To determine the concentration of Pb2+ in the original sample of
blood, we make appropriate substitutions into the corresponding
eqn and solve for CA.
31
32
Multiple Standard Additions
33
• At the top is a set of 6 standard
additions for the determination
of Mn2+.
34
How is this best done in practise?
35
The result:
standard
5 mL
sample
36
Example 3
Gold was determined in a waste stream using
voltammetry. The peak height of the current
signal is proportional to concentration.
A standard addition analysis was done by adding specific volumes of
10 ppm Au solution to the sample as shown in the table below.
All solutions were made up to a final volume of 20 ml. The peak
currents obtained from the analyses are also tabulated below.
Calculate the concentration of Au in the original sample.
50
40
30
20
10
0
0 1 2 3 4 5 6
Conc added std / ppm 38
2 - Find the best straight line
n xi yi xi yi x y x y x
2
n x x
m i i i i i
n x x
2 2 c 2 2
i i i i
y = 6.50x + 8.68 39
3 - Extrapolate to the x-axis (y = 0)
y = 6.50x + 8.68 50
Peak current / A
40
30
Y=0
20
Thus, X = CA y = 6.50x + 8.68
= -8.68/6.50 10
R2 = 1.00
= 1.33 ppm 0
(Absolute value) -2 -10 0 2 4 6
CA (Diluted) x 20 ml = CA (Original) X 10 ml
1.33 ppm x 20 ml = CA (Original) X 10 ml
CA (Original) = 1.33 ppm X (20/10)
41
Using a Standard Addition to Identify Matrix Effects
How?
42
4. INTERNAL STANDARDS
The signal from the analyte is compared to the signal from the
internal standard when determining the concentration of analyte
present.
43
Because the analyte and the internal standard in any sample or standard
receive the same treatment, the ratio of their signals is unaffected by any
lack of reproducibility in the procedure.
Assumption:
If the internal standard signal increases by 10% for the same solution from
one run to the other, it is most likely that the signal from the sample also
increases by 10%.
44
Say analyte (X) and internal standard (S) have the same
concentration in solution.
The signal height for X may be 1.5 times greater than that for S.
IX IS
F
4 [X] [S]
3
X
S
2
1 [X] and [S]: concentrations of
analyte and standard after they
0 have been mixed together.
0 1 2 3 4 5 6
The response factor (F) is 1.5 times greater for X than for S.
45
Example: In a preliminary experiment, a solution containing 0.083 7 M X and
0.066 6 M S gave peak areas of Ix = 423 and Is = 347 (areas are measured in
arbitrary units). To analyse the unknown, 10.0 mL of 0.146 M S were added
to 10.0 mL of unknown, and the mixture was diluted to 25.0 mL in a
volumetric flask. This mixture gave a chromatogram, for which Ix = 553 and
Is = 582. Find the concentration of X in the unknown.
IX IS
Solution: 1st use the standard mixture to find
the response factor
F
[X] [S]
46
Using the known
IX IS
response factor, we find F
the concentration of [X] [S]
unknown in the mixture:
Because X was diluted from 10.0 to 25.0 mL when the mixture with S
was prepared, the original concentration of X in the unknown was
= 0.143 M
47
Every sample should be analyzed with an internal standard (IS)!!!
•In GC analysis with non-MS detector: if target is, for ex. tri-chlorophenol,
use tri-bromophenol or di-chlorophenol.
48
• From previous slide, we assume that samples have little or no Th or As.
• Solid samples can use a naturally occurring element as IS, provided that
you know the concentration in each sample.
49
Advantages:
• Fluctuations are monitored in each sample/calibration / blank.
Disadvantages:
• Assume that behavior of IS is the same as the analyte.
51
5. Blank Corrections
• So far, we have assumed the use of a suitable reagent blank to correct for
signals arising from sources other than the analyte.
Case Study
About 200 analytical chemists were asked to evaluate a data set
consisting of a normal calibration curve, a separate analyte-free blank,
and 3 samples of different size but drawn from the same source.
53
• That all 4 methods give a different result for CA shows the importance of
choosing a proper blank, but does not tell us which blank is correct.
• Because they all fail to predict the same CA for each sample, none of these
blank corrections properly accounts for an underlying constant source of
determinate error.
54
• To correct for a constant method error, a blank must account for signals from
any reagents and solvents used in the analysis, as well as any bias resulting
from interactions between the analyte and the sample’s matrix.
• Both the calibration blank and the reagent blank compensate for signals from
reagents and solvents.
• Any difference in their values is due to indeterminate errors in preparing and
analyzing the standards.
• Unfortunately, neither a calibration blank nor a reagent blank can correct for
a bias resulting from an interaction between the analyte and the sample’s
matrix.
• To be effective, the blank must include both the sample’s matrix and the
analyte and, consequently, must be determined using the sample itself.
The regression line for the 3 samples is: Ssamp = 0.009844 × Wsamp + 0.185
giving a true blank correction of 0.185 with 3 identical values for CA.
55
• Total Youden Blank is not common in analytical work.
• As long we can ignore any constant bias due to interactions between the
analyte and the sample’s matrix, which is often the case, the accuracy of
an analytical method will not suffer.
56
• Standard blank
Blank used to monitor polyatomic ion interferences, gas peaks, and
contamination from reagents; used for background subtraction.
• Procedural blank
Blank used to monitor contamination acquired during all stages of
sample preparation; grinding, digestion, acidification, powdering, etc.
Outlier tests:
Option 3. Statistical “proof”: the best approach, but needs to be carried out
carefully in order to avoid false negatives and positives.
58
DETECTION LIMITS
All instrumental methods have a degree of noise associated with the
measurement. limits the amount of analyte that can be detected.
Detection limit (DL or LOD): the lowest concentration level that can be
determined to be statistically different from the analyte blank.
LOD = 3*STDEVblank
Limit of Quantification (QL or LOQ): the smallest (minimum) concentration of
a measurand that can be reliably (i.e. with a high degree of confidence)
measured by an analytical procedure: LOQ = 10*STDEVblank
59
A ‘good’ analytical method will:
1. provide the means to calculate an accurate background level
2. allow for correction of instrument drift
3. use Internal standardisation to monitor matrix effects
4. provide some method for monitoring/ correcting interferences
5. Use a proper calibration strategy
60
6. Isotope Dilution (ID)
111Cd
111Cd
61
ID Scheme
‘‘ideal ’’ Internal
Standard
23,1 80
25
20 16,87 60
13,18
15 9,97 40
10 6,87
20 4,92 0,66 0,73 0,11
5 0,15 0 1,63
0 0
196 198 199 200 201 202 204 196 198 199 200 201 202 204 198 199 200 201 202 204
isotopes isotopes isotopes
A known amount of the tracer, called ’’spike’’, with an abnormal isotopic composition
of the element to be analysed is added to the sample, preferably before any
chemical treatment.
The isotopic ratio between the tracer and the element is known; we measure the
isotopic ratio in the spiked sample with a mass spectrometry method. The element
concentration can then be deduced from the isotopic ratio and the amount of the
added spike.
63
64
• The amount of spike is selected so that the resulting ratio between
spiked isotope and unspiked isotope is near unity – maximizes precision
• Typically use the most abundant isotope as the reference -- maximizes
sensitivity
• Disadvantages:
66
Hg Enriched Isotope Standards