Professional Documents
Culture Documents
A 1 OF 11
This procedure describes how to add CI functionality to the GCT mass spectrometer.
Parts required
Table 1: Parts required
Tools/materials required
• 2.5-mm Allen wrench
• ½-inch open-end wrench
• 2 x 5/16-inch open-end wrenches
• 2 x 7/16-inch open-end wrenches
• 7-mm open-end wrench
• Large adjustable wrench
• Tube-cutter for 1/8-inch stainless steel tubing (required for adding 1/8-inch plug valve)
Procedure
The procedure to add CI functionality to GCT instruments consists of nine sections:
1. Benchmark sensitivity test
2. Installing the CI upgrade kit (if necessary)
3. Connecting to the CI gas flow-metering valve
4. Installing the CI gas cylinder and plug valve
5. Leak checking the system
6. Installing the GC interface
7. Installing the CI source
8. Performing specifications
9. Training the customer
NOTE: Test the instrument before performing the upgrade. Do not perform the upgrade
until the benchmark tests have been completed.
NOTE: After the upgrade has been performed, repeat the benchmark tests to make
sure that the levels achieved are the same.
NOTE: The measured signal-to-noise ratio for hexachlorobenzene, obtained from the
mass chromatogram of 283.8102 Da must be >10:1 peak-to-peak.
Sensitivity
NOTE: The upgrade is NOT required if a stainless steel capillary is already installed
into the outer source (Figure 1). Continue with the ‘Installing the CI gas cylinder
and plug valve’ section of the procedure.
NOTE: The upgrade MUST be performed if a blank nut is in place of the stainless steel
capillary on the outer source.
Stainless steel
capillary
8. Pass the long length of the CI capillary (M960298BD1) through the 1/16 inch nut (6070826),
ferrule set (6070130) and the 1/16 inch bulkhead from the inside face of the source lid as
shown in Figures 2 and 3.
9. Insert the end of the CI capillary into the brass block on the outer source, next to the
reference capillary line (Figure 3).
NOTE: The CI capillary is an improved design and should already be bent into shape.
10. Finger-tighten plus 1¾ turns the 1/16 inch nut to secure the CI capillary.
NOTE: Contact Global Services Support (+44 (0)161 435 4490) for assistance when
installing the CI valve assembly kit.
1. Use the 2.5-mm Allen wrench to remove the three cap head screws securing the flow valve
bracket to the front of the instrument.
NOTE: The valve fittings are either 1/16 inch Swagelok or 1/16 inch Valco.
2. Connect the PEEK tube capillary (6346003) to the rear of the flow valve (Figure 6) and tighten
the nut and ferrule using the 7/16 inch wrench.
NOTE: The flow valve side port should already be connected to the CI gas steel
capillary coming from the bottom of the instrument.
3. Feed the PEEK tube capillary underneath the reference reservoir and connect to the
CI capillary fitted to the outer source, using the 1/16 inch union (6070202). See Figure 7.
4. Use the two 5/16 inch wrenches to tighten the 1/16 inch union.
1/16 inch
union
5. Ensure the EI inner and outer sources and the GC interface blank flange are fitted securely.
6. Loosen the black knob on top of the reference reservoir.
7. Pump down the source.
8. Enter the following temperature settings:
• Source temperature 180 °C
• Reference reservoir 20 °C
• Reference re-entrant 150 °C
NOTE: After 30 minutes pumping, the source pressure should be in the region of
1.0 E-6 on the Tune page.
NOTE: If 1.0 E-6 source pressure is not reached there may be a gross leak. Before
proceeding, vent the source and check all the connections made in the
procedure.
WARNING: This test is designed to detect gross leaks within the instrument's CI
system and prevent the release of gases at levels where they may become
hazardous.
WARNING: It does not prove that the system is "leak tight" and may not detect leaks
that will affect performance. It will not detect leaks in the customer’s CI
gas delivery system (bottles, regulators, wall tubing), which is their
responsibility.
6. Ensure that the black CI flow needle valve on the front of the instrument is closed.
7. Open the Vacuum Monitor in the Tune page and ensure that the Backing Pressure is offscale
(below 1.0 E-2 mbar).
8. Switch to CI mode and select CI Purge on the Inlet tab of the Tune page
NOTE: You will hear valves click on the back of the instrument. The backing pressure
will normally rise and come down to about 2 or 3 E-4 within 30 seconds. The CI
purge will time out after 45 to 70 seconds. When this happens, it is normal to
hear a clicking noise and see the backing pressure rise briefly before falling again.
NOTE: The backing pressure should come down to 1.0 E-2 mbar within 30 seconds.
10. Keep purging the CI gas line until it times out again.
11. Deselect the CI Purge and wait five minutes.
12. Select CI Purge again.
NOTE: The pressure should not exceed 2.0 E-2 mbar or not change at all. This
indicates a leak tight external CI gas system. When the purge is de-selected,
it is normal for the backing pressure to rise and fall again.
NOTE: If a leak is detected during this test, use the Swagelok 1/8 inch blanks (male
and female included in the kit), to isolate areas of the internal and external CI
gas plumbing and identify the source of the leak.
NOTE: It is recommended that all minor leaks are checked and eliminated in EI mode,
using an easily ionized gas with a known mass (such as argon at m/z 40, or
some spray dusters) around all O-rings, seals and ferrules.
NOTE: For further assistance contact Global Service Support on +44 (0)161 435 4490.
Performance specifications
Upgrade confirmation
Signed Date
I certify that the instrument upgrade above has been successfully completed.
Engineer Date