You are on page 1of 8

Medical device software under

the EU MDR
By Leo Hovestadt, MSc

The EU Medical Device Regulation (MDR) has been published with new medical
device software (MDSW) requirements. Qualification guidance to determine if
software is MDSW was combined with guidance for MDSW classification.
However, EU Medical Device Directive (MDD) guidance for clinical evaluation
should have been replaced but remains in use. The change to the MDR has
introduced new problems for clinical evaluation, so guidance has been
introduced for equivalence, legacy devices, and the clinical evaluation
assessment report (CEAR) of the notified body and for the clinical evaluation
of MDSW.

Introduction
Manufacturers who think their software as a medical device (SaMD) technical
file will satisfy the MDR1 are in for a surprise. Approaches to MDSW under the
MDR differ significantly from other regulations and are generally more strict.

For instance, an MDSW can be part of a hardware device, which is driven or


influenced by the software, and in that case, the MDSW gets the highest
classification of either the hardware or the MDSW. SaMD is standalone software

©2021 Regulatory Affairs Professionals Society

January 2021
and thus, cannot contain hardware in the way MDR MDSW can. The MDR
guidance contains a step called Qualification to determine whether software is
MDSW. There is no such step for SaMD. MDSW is often confused with SaMD,
resulting in the MDSW being assigned an incorrect, lower classification.

The implications of an incorrect classification can be significant. For example,


skin cancer apps for early detection of melanoma and other types of skin cancer
were gaining popularity during the development of the MDR. To address
concerns about software in these apps and similar products, the MDR Rule 11
was developed. However, the rule was not sufficiently comprehensive, with the
result that some apps received a higher classification than warranted, at times
as high as class III. As a result, many devices that were safely on the market for a
long time with an MDD or active implantable MDD CE-marked device had to do
clinical investigations before market release under the MDR because of the class
III classification.

It took a long time to develop guidance with a reasonable approach for Rule 11,
which met the new regulations in the MDR. The new Medical Device
Coordination Group (MDCG) guidance for the qualification and classification of
MDSW2 software solved the problem but was very complex. Many consultants
tried to explain the MDR by using their experience with the MDD,3 Medical
Devices Documents (MEDDEV),4 or International Medical Device Regulators
Forum (IMDRF)5 guidance, but they created confusion. The new guidance for
MDSW has many new requirement for qualification, 2 classification,2 clinical
evaluation,6-10 and cybersecurity.11,12 It is important for MDSW manufacturers
who want to place their devices on the EU market to carefully read all new
guidance on MDSW.

Qualification
Qualification is the activity that determines whether MDSW is covered under
the MDR. The qualification method is mentioned in guidance MDCG 2019-11.2
Examples of qualification will be given in future MDCG guidance on borderline
and classifications, which is currently being developed.

It is most important that the guidance’s qualification workflow is followed


(Figure 1, p. 3).

From the flow depicted in Figure 1, we can see that software accessories at step
2a or software driving or influencing a medical device at step 2b are covered by
the MDR. Accessories are classified in their own right according to implementing
Rule 3.2, and software driving or influencing a medical device are classified
together with the medical device according to MDR Annex VIII implementing
Rule 3.3. So at step 2a, Rule 11 has to be applied, and via step 2b, Rule 11 does
not have to be applied, which can be a significant difference.

January 2021
In step 5, there is a check to ensure that the software’s intended purpose meets
the definition of a medical device under MDR Article 2 definition (Step 1),
making it MDSW.

Classification
Classification according to MDR Annex VIII and MDCG 2019-112 is the activity
that determines the risk class of a medical device. The classification rules are
mentioned in MDR Annex VIII and guidance MDCG 2019-11 on classification.
The risk class indicates what is required to obtain the Conformité Européenne
(CE) certification of the MDR. (The CE mark indicates the manufacturer takes
responsibility for the product’s compliance with EU requirements.) The
classification rules for active devices apply for MDSW, because software under
the MDR is an active device. All applicable classification rules have to be applied,
and the rule with the highest classification determines the final classification of
the device.

MDR Annex VIII, Rule 11 most often gives the highest classification. Not reading
the guidance, however, again comes at a high price, as shown in the Table (p. 4).

January 2021
The text of Rule 11 is repeated in this table. For example, according to Rule 11,
all MDSW treating or diagnosing cancer (which is critical) is class III. However,
under the MDD, in general, this is class IIb or lower, and sometimes even class I.

To avoid being assigned a classification that is too high, one should use the
guidance MDCG 2019-11 and MDR Annex VIII implementation Rule 3.3. The
MDCG 2019-11 guidance is based on the IMDRF SaMD working group’s N125
guidance, which is the source for Table 1. The IMDRF guidance recognizes that
most software has an indirect influence on treatment or diagnosis and that
therefore, the classification should be lower. So software that drives clinical
management (see the middle column in the table) or software that informs
clinical management (right column), should have a lower risk class. An example
coming from the IMDRF guidance is radiation therapy treatment planning. This
software is driving clinical management of radiation treatment delivery for
cancer, which is critical. Applying MDCG 2019-11 puts this software in the
middle column in the top row, and thus the classification is class IIb. It should be
noted that the IMDRF SaMD working group’s N12 document5 contains a
mistake, which is explained in the note from MDCG 2019-11 Annex III. (The N12
document mistakes are in 7.3 Criteria for Category II for i and iii; 7.4 Category III
for i and ii examples.)

In Table 1, the software in the left column is often part of treating and
diagnosing hardware. Here, implementing Rule 3.3a is important. It says that
software driving a medical device or influencing the use of a medical device,
should fall within the same class as the medical device, avoiding the problem of
Rule 11 for hardware containing software.

MEDDEV 2.7.1 rev 4 clinical evaluation


The MEDDEV 2.7.1 rev 4 clinical evaluation6 guidance was written for the MDD,
the predecessor of the MDR. This guidance was not written with MDSW in mind.

January 2021
Notified bodies expect that clinical evaluation follows this guidance closely, in
combination with the MDR requirements. The notified body uses the CEAR
template7 for this review. The manufacturer is well advised to study the CEAR in
detail, which can be found in the MDCG 2020-10/1 CEAR template.7

The MEDDEV clinical evaluation guidance is currently the best that is available,
until it is upgraded to the MDR. The most important gaps for MDR requirements
in the MEDDEV clinical evaluation include the following omissions:

• No references to the MDR, such as articles, annexes, and general safety


and performance requirements (GSPR) sections.
• No discussion about MDR Article 61(10) that technical data can replace
clinical data under certain conditions.
• Missing MDR requirements for:
o Clinical evaluation plan,
‒ Identifying GSPRs requiring (clinical) data
‒ Defining which level of clinical evidence is required;
o Clinical investigation for class IIb implantables, and class III devices;
o Postmarket surveillance plan, postmarket clinical follow-up plan,
and clinical development plan;
o Common specifications;
o Stricter requirements for clinical data; and
o Qualified assessment of sufficient level of clinical evidence.

Clinical evidence
Clinical evidence, according to the MDR, has to be based on a very strict
definition of clinical data from an original or equivalent device, with clinical data
coming mainly from clinical investigations. In addition, the quality of the
available postmarket surveillance clinical data not considered good enough. This
was problematic for a lot of medical devices with “sufficient clinical evidence”
but where the clinical data did not meet the MDR requirements or the
expectations of the reviewers. It was especially problematic for MDSW. Three
guidance documents were developed to circumvent the problems:
• MDCG 2020-6 guidance on sufficient clinical evidence for legacy
devices.8 This guidance is not specific for MDSW but is helpful for the
transfer of clinical evidence from the MDD to the MDR. In addition,
Appendix III shows a useful overview of clinical evidence sources.
• MDCG 2020-5 guidance on clinical evaluation.9 This guidance provides
an additional and more realistic explanation of software equivalency.
• MDCG 2020-1 guidance on clinical (MDR) or performance (IVDR)
evaluation of medical device software.10 This guidance fills the gaps for
MDSW clinical evaluation and explains how to create clinical evidence.
The guidance requires clinical evidence for technical performance (for
tool-based MDSW without direct clinical benefits) and, where
applicable, clinical evidence for a valid clinic association and for clinical
performance.

Clinical evaluation is required, according to the MDR Article 61(1), so this has to
be created in addition to MDSW clinical evaluation. However, there is no

January 2021
explanation anywhere for how the two clinical evaluations should be combined.
Figure 2 outlines a way this can be done.

In Figure 2, both clinical evaluations require that acceptance criteria have to be


based on state-of-the-art, current practice of medicine (see left column). Clinical
data has to be collected, appraised, and analyzed (middle column). For MDR
Article 61(1), clinical evaluation has to be based on clinical data from the device
to be released or from an equivalent device. That restriction does not apply to
MDSW clinical evaluation, which can use more data sources. Based on MDR
Article 61(10), justifications can be written when there is MDSW clinical
evidence in addition to MDR Article 61(1) clinical evidence, so that there is
sufficient clinical evidence. Article 61(10) cannot be used for class III MDSW.

For MDSW clinical evaluations, for example, curated clinical data can be used
from real-world data sources. The data still has to be appraised to guarantee its
quality and applicability for the evaluated MDSW. Finally, the clinical evidence
has to be assessed (right column). The assessment uses the acceptance criteria
from the first box. Article 61(1) requires clinical evidence for safety (including
for no unacceptable side-effects), clinical performance, and a positive benefit-
risk ratio. For the MDSW, clinical evidence for a valid clinical association,
technical, and clinical performance is required.

Cybersecurity guidance
The MDR contains a limited set of cybersecurity requirements, such as MDR
Annex I: 14.2.(d), 17.2, 17.4, 18.8, and 23.4ab. When the MDR came into effect
in May 2017, the WannaCry ransomware attacked hospitals around the world.
In response to that event, MDCG 2019-16 guidance on cybersecurity11 was
immediately developed to protect MDSW against cyberattacks. New releases of
this guidance are expected soon, along with other requirements for
cybersecurity patches.

This guidance also contains the manufacturer’s cybersecurity requirements to


inform the hospital asset owner and system integrator how the MDSW can be
protected. The manufacturer’s disclosure statement, or the Medical Device
Security (MDS2) form, was developed for this purpose and is used to exchange

January 2021
essential cybersecurity information between the manufacturer and the hospital.
If the MDSW handles patient data, then the General Data Protection
Regulation,12 which requires the patient data to be protected, also applies.

Conclusion
The MDCG guidance for the MDR should be studied carefully because it contains
solutions for problems and additional requirements for acquiring the MDR CE
mark. The guidance related to clinical evaluation of MDSW, equivalence, legacy
devices, and the CEAR contains solutions for performing clinical evaluation. The
guidance for MDSW qualification and classification has a section on
classification Rule 11. Applying the guidance for Rule 11 often avoids a higher
risk classification. The cybersecurity guidance contains requirements that
cannot be found in the MDR.

Abbreviations
CE, Conformité Européenne; CEAR, clinical evaluation assessment report; IMDRF, International
Medical Device Regulators Forum; MDD, Medical Device Directive; MDCG, Medical Device
Coordination Group; MDR, [EU] Medical Device Regulation; MDSW, medical device software;
MEDDEV, Medical Devices Documents; SaMD, software as a medical device.

References
1. European Commission. EUR-Lex website. Regulation (EU) 2017/745 of the European
Parliament and of the Council of 5 April 2017 on medical devices, amending Directive
2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and
repealing Council Directives 90/385/EEC and 93/42/EEC. https://eur-
lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32017R0745&from=EN. Current as
of 24 April 2020. Accessed 8 December 2020.
2. European Commission. MDCG 2019-11 – Guidance on qualification and classification of
software in Regulation (EU) 2017/745-MDR and Regulation (EU) 2017/746-IVDR.
https://ec.europa.eu/health/sites/health/files/md_sector/docs/md_mdcg_2019_11_gui
dance_qualification_classification_software_en.pdf. Dated October 2019. Accessed 8
December 2020
3. EUR-Lex website. Council directive 93/42/EEC of 14 June 1993 concerning medical
devices. https://eur-
lex.europa.eu/LexUriServ/LexUriServ.do?uri=CONSLEG:1993L0042:20071011:en:PDF.
Accessed 8 December 2020.
4. European Commission. MEDDEV 2.1/6 – Guidelines on the qualification and
classification of stand alone software used in healthcare within the regulatory
framework of medical devices.
https://ec.europa.eu/docsroom/documents/17921/attachments/1/translations/en/ren
ditions/pdf. Dated July 2016. Accessed 8 December 2020.
5. International Medical Device Regulators Forum. IMDRF SaMD working group (N12).
Software as a medical device: Possible framework for risk categorization and
corresponding considerations. http://www.imdrf.org/docs/imdrf/final/technical/imdrf-
tech-140918-samd-framework-risk-categorization-141013.pdf. Dated 18 September
2014. Accessed 8 December 2020.
6. European Commission. Clinical evaluation: A guide for manufacturers and notified
bodies under Directives 93/42/EEC and 90/385/EEC. MEDDEV 2.7/1 revision 4.
https://ec.europa.eu/docsroom/documents/17522/attachments/1/translations/en/ren
ditions/native. Dated June 2016. Accessed 8 December 2020.
7. European Commission. MDCG 2020-13 – Clinical evaluation assessment report
template.
https://ec.europa.eu/health/sites/health/files/md_sector/docs/mdcg_clinical_evaluatio
ntemplate_en.pdf. Dated July 2020. Accessed 8 December 2020.
8. European Commission. MDCG 2020-6 – Regulation (EU) 2017/745: Clinical evidence
needed for medical devices previously CE marked under Directives 93/42/EEC or
90/385/EEC.

January 2021
https://ec.europa.eu/health/sites/health/files/md_sector/docs/md_mdcg_2020_6_guid
ance_sufficient_clinical_evidence_en.pdf. Dated April 2020. Accessed 8 December 2020.
9. European Commission. MDCG 2020-5 – Clinical evaluation – Equivalence.
https://ec.europa.eu/health/sites/health/files/md_sector/docs/md_mdcg_2020_5_guid
ance_clinical_evaluation_equivalence_en.pdf. Dated April 2020. Accessed 8 December
2020.
10. European Commission. MDCG 2020-1 – Guidance on clinical evaluation (MDR)/
performance evaluation (IVDR) of medical device software.
https://ec.europa.eu/docsroom/documents/40323/attachments/1/translations/en/ren
ditions/native. Dated March 2020. Acessed 8 December 2020.
11. European Commission. MDCG 2019-16 rev.1 – Guidance on cybersecurity for medical
devices.
https://ec.europa.eu/health/sites/health/files/md_sector/docs/md_cybersecurity_en.p
df. Dated July 2020. Accessed 8 December 2020.
12. EUR-Lex website. Regulation (EU) 2016/679 of the European Parliament and of the
Council of 27 April 2016 on the protection of natural persons with regard to the
processing of personal data and on the free movement of such data, and repealing
Directive 95/46/EC (General Data Protection Regulation). https://eur-
lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32016R0679&from=EN. Accessed
8 December 2020.

About the author


Leo JC Hovestadt, MSc, is responsible for EU governmental affairs at Elekta, a radiation therapy
company based in The Netherlands. He is involved in European and international work groups for
writing guidance on clinical evaluation and investigation. Hovestadt has more than 25 years of
experience with high-risk medical devices at Elekta and Medtronic. He can be contacted at
leo.hovestadt@elekta.com.

Citation Hovestadt LJC. Medical device software under the EU MDR. REGULATORY FOCUS. January
2021. Regulatory Affairs Professionals Society.

January 2021

You might also like