Operation / Service Manual

ENT Workstation
Model : New Millennium Grand _2014.02_Rev.6

Manufactured by
MEDSTAR Co., Ltd.

35-12, Goraedeul-gil, Uiwang -si, Gyeonggi-do,Korea

Tel : 82-31-457-1111 Fax : 82-31-457-2670 Since 1997

━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Homepage : http://www.medstar.co.kr
E - mail : president@medstar.co.kr, trade@medstar.co.kr

ENT Workstation_New Millennium Grand_02.Rev.6 1


Contents

1. Brief..............................................................3
2. Specification.................................................4
3. Structural Lay-Out........................................5~16
4. Operation.....................................................16~28
5. Precaution in Use….....................................29~30
6. Maintenance & Storage................................31~37
7. Plumbing Drawing…………………………..…38
8. Trouble Shooting……………………….…...…39~42
9. Standard Accessories...................................43~44
10. Product Information Sticker..... ...................45
11. Warranty Certificate………………………….46

ENT Workstation_New Millennium Grand_02.Rev.6 2


1. Brief
Thanks for your purchasing this equipment.

1-1. Precaution Before Use


Read the following instruction before use.
Warning ! : This product is only used for the medical purpose by the expert in the
medical field.
The failure to read the instruction carefully and misuse might cause to hurt
the human body and damage the product.

Repeat to read and understand how to use this equipment.

1-2. Summary of New Millennium Grand


-- One of ENT Workstation Models, New Millennium Grand, is designed to diagnose and
treat ENT patients, in a very comfortable way, using the basic parts, for instance, spray,
suction. surgical, instrument tray, cans, medicine bottles, warmer.

In addition, using the optional parts, such as endoscopic visual imaging system, film
view box, patient chair connected to New Millennium Grand, stool, as well as ear
irrigation system, automatic drainage system, ENT doctor can carry out their jobs in a
very efficient way

ENT Workstation_New Millennium Grand_02.Rev.6 3


2. Specification
Item Specifications

Model New Millennium Grand

Oilless Vacuum Pump(250W)


Main Suction System 100l/min., at 680mm Hg
Suction Bottle(3,000cc/1,500cc)
Oilless Vacuum Pump(130W)
Sub Suction System 40l/min. at 650mm Hg
Suction Bottle(1,500cc / 2 EA)
Oilless Air Compressor(250W)
Spray System Stainless Air Tank
100l/min. at 7Kg/cm2

Spray Quantity Straight : 3 EA, Curved : 1 EA

Ventilation System Automatic Operation at Lift-Up

Anti-Fog System Fog / Mist Eliminated on the Tip of Endoscopes

Top Plate Tempered Glass

Local Control Panel All Parts controlled by One Touch Panel

Dimension 1,652(W)x782(D)x800(H)mm

Weight 170Kg

Power Source AC220V, 50~60Hz

Power Consumption 1,500W

Classification Class 1, Type B

License No. 07-148

ENT Workstation_New Millennium Grand_02.Rev.6 4


3. Structural Lay-Out
3-1. Overall Top, Front, Side View

18
19

31

20

21
15 12 13 14 16 17

2 3 4 5 6 7 8 9 10 11

ENT Workstation_New Millennium Grand_02.Rev.6 5


No. Description Function

Made by Tempered Glass, to protect any damage/


1 Top Plate
discoloration from strong liquid medicine.

2 Side Cans Waste materials might be put into them.

3 Spray Inject medication into ENT for treatment.

4 Suction Suck the secretion materials by vacuum pump.

Treat the equilibrium of pressure between ear & nose by


5 Ventilation
this device.

6 Vent Hanger Ventilation Tube hung on the hook.

7 Main Suction Suck the internal secretions from ENT.

8 Main Suction Hanger Main Suction Hung on the hook

9 Local Control Panel All Parts Controlled by One-Touch Button

10 Anti-Fog System Eliminate the mist/fog on the endoscopes in winter.

Rectangular Wastes
11 Put the wastes after treatment.
Tray
Surgical Instrument Put a wide range of surgical instruments into the warm
12
Tray tray.
Lid for Surgical
13 Cover for the surgical instrument tray.
Instrument Tray

14 Cans Put a wide range of surgical instruments, gauze, etc.

Illumination Light, Articulated Arm for Micro Visual


15
Curved Post Pole System, Monitor Arm, Medicine Bottles Tray, attached to
it.
Arm for LS-250, Endoscope Hanger, Arm for Film View
16 Straight Post Pole
Box, fastened to it.

17 Straight Post Pole Arm for Film View Box, Another Monitor Arm, Fastened.

18 Light Socket Supply the power to the illumination light.

ENT Workstation_New Millennium Grand_02.Rev.6 6


19 Earth Protect the electricity from the thunder light.

20 Fuse Protect this equipment form the over current.

Connect the plugs of the parts, monitor, film view box,,


21 Additional Socket
light source, etc.

22 Main Power Switch Turn on / off the main power.

23 Chair Receptacle Connect the plug of the chair.

24 Additional Socket Connect the other parts’ plug of this equipment.

25 Silencer Abate the sound of the motor.

Endoscope Sterilizer
26 Turn on / off the power of the endoscope sterilizer.
Power Button

27 Endoscope Holder Disinfect the used endoscope by disinfectant.

28 Swivel Front Drawer Put the keyboard of the PC on it.

Suction Bottles Peep Take a look at how much level suction bottles are filled
29
Hole with wastes, secretions, etc.
Open / Close the door to take out / put the suction
30 Front Door
bottles on the sliding plate.

31 Main Power Socket Supply the main power to this equipment.

ENT Workstation_New Millennium Grand_02.Rev.6 7


3-2. Inside Top & Front View

13

15 16

ENT Workstation_New Millennium Grand_02.Rev.6 8


No. Description Function

1 Suction Bottle Nominal Volume at 3,000cc

2 Suction Bottle Nominal Volume at 1,500cc

3 Air compressor Motor Forward air to the air tank, driven by Motor

4 Compressor Motor Socket Two Head Motor’s plugged into the socket for air compressor.

5 Main Suction Motor Socket Two Head Suction Motor’s plugged into the socket for main suction.

Single Head Suction Motor’s plugged into the socket for sub-
6 Sub Suction Motor Socket
suction.

Nominal Volume at 250cc to put the Watery Particles from Suction


7 Screw Bottle
Motor.

Nominal Volume at 250cc to put the Watery Particles from Air


8 Screw Bottle
Tank.

9 Sub-Suction Motor Operate the sub-suction by motor.

Lifting up, it automatically sucks, consisting of suction line, roller


10 Sub-Suction System
and micro-switch.

Pressing the knob, medication sprays into ENT, consisting of spray


11 Spray System
line, roller and micro-switch.

12 Main Suction Motor Operate the main suction by motor.

13 Air Tank Air filled by compressor, driven by motor.

Pressure Gauge, Regulator, Set the pressure of suction, spray & indicate its value onto the
14
Filter gauze, while filtering the impure particles of the air.
15 Relay PCB Relay the signal that each part operates.
Get the signal from the local control panel, others & control the
16 Main PCB
whole signals and give the signal to relay PCB.

ENT Workstation_New Millennium Grand_02.Rev.6 9


3-3. Anti-Fog
- Pressing down the switch, it automatically operates for six seconds with warm air

supplied, which is used for eliminating the fog/mist, layered on the tip of endoscopes

in the winter season or in a much temperature gap place causing to make a mist in a

dew point.

3-4. Suction & Ventilation System

1) Main & Sub-Suction System


The New Millennium Grand and Millennium Grand have two suction systems, called
main suction in front of the cabinet, sub-suction system on the top plate.
The other models have a single suction system on the top plate.
Whenever lifting up the suction holder, it automatically operates to suck the
secretions from the ear, nose, while lifting down it onto the hollow round hook, it
automatically stops.

ENT Workstation_New Millennium Grand_02.Rev.6 10


2) Ventilation System
Connect the ventilation tube into its pipe, whenever lifting up its adaptor, the air
automatically supplies, while lifting down on its hook, it automatically stops.
Its system is used for flowing the air between ear & nose.

3-5. X-Ray Film View Box

No. Description Function

1 Arm Set Fasten the film view box onto the straight post pole.

2 Button Pressing it, ENT anatomic chart lifts up, with rolling- up
hung.
Pulling down the bottom bar, ENT anatomic chart lifts
3 ENT Anatomic Chart
down.
Inserting the x-ray film into the top bar, ENT doctor can
4 X-Ray Slim View Box
read what it means.
5 Main Power Switch Turn on / off the main power.

6 Light Intensity Knob Control the intensity of the luminous light.

7 Main Power Cable Supply the main power to the x-ray film view box.

ENT Workstation_New Millennium Grand_02.Rev.6 11


3-6. Microscopic Visual Imaging System

4
8 9 10
5
3
2 13
1

M ic r o s c o p e

11

12

No. Description Function

1. Eye Microscope at 3,5,8 Magnification


2. Microscope Mount Adaptor Adaptor to be able to connect the CCD camera
3. CCD Camera(M-25) CCD Camera(M-25)
4. Knob Adaptor & Ball Joint Connect the microscope head and its hand wheel to it.
5 Light Guide Cable Supply the light to the object seen by microscope head.
6 Articulated Arm Able to move the microscope head at the point wanted by this
arm.
7 Elastic Spring It gives the articulated arm a wide range of elasticity.
8 Mounted Arm Fasten the bolt of bottom into the round nut to fix it on the
mounted arm.
9 Light Source(LS-150) Supply the light to the light cable by which doctors take a look
at the specific parts of ENT.
10 Bolster Bar Insert it into the hole of the mounted arm to fasten it by its bolt.
11 Post Pole Bracket Fasten the mounted arm, inserting it into the hole.
12 Curved Post Pole Support Bar to connect the microscopic visual system.
13 Mounted Arm Knob Insert the articulated arm into its hole and fasten it by knob
Adaptor adaptor.

ENT Workstation_New Millennium Grand_02.Rev.6 12


1) Referring to the fig., above, assemble the parts of microscopic visual imaging system,
properly.
2) Fasten the round bracket so as to assemble the mounted arm and articulated arm.
3) Install the light source, placed on the mounted arm, above.
4) Connect the light guide cable into the socket of light source (LS-150), while the
opposite plug of it places to insert into the hollow of the articulated supporting arm,
above.
5) In addition, connect the CCD camera and its cable, placed outside the articulated
arm, fixed in a neat way by its clamp.
6) The careless manipulation of microscope head might hurt the light guide cable.
7) In case of improper display on the monitor, adjust the TV adaptor.

8) Be sure to place each pin into a corrective position, when connecting the CCD

camera to its cable.

3-7. Illumination Light


1) Usage

The illumination light helps to enlighten the surroundings for ENT doctor to be able to

diagnose and treat patients, effectively and efficiently, while it’s moved in a versatile

radius with its bulb heat emission air cooled.

T E N S IO N R E G U L A T IO N

L IG H T A R M

C O VE R
LAM P

L IG H T C O N N E C T O R

L IG H T H E A D

ENT Workstation_New Millennium Grand_02.Rev.6 13


2) Installation Method

- Holding up the illumination light head part by the hand, slightly forcefully insert its

plug into the hole of the end of long swivel bar, until it appears at the rear bottom

rear point through the curved post pole.

- Connect the plug into the socket.

3-8. Light Source(Halogen Lamp, LS-250)


1) Specification
- Power Source : AC 220V, 50~60Hz
- Power Consumption: 300W
- Halogen Lamp : 24V, 250W
- Light Intensity : Adjustable(0-250W)
- Memory Board : `1, 2, 4 Capture(NTSC / PAL)
- Dimension : 382(W)x302(D)x195(H)mm
- Weight : 11Kg

2) Overall View
- This equipment consists of the two CCD camera socket, two light source sockets,
light intensity control knob, several one touch small switches, to be able to take
the photos by microscopic / endoscopic visual imaging system.
- The memory board inside can save the max. four photos, when taking the fifth
photo, the first photo
automatically deletes.

ENT Workstation_New Millennium Grand_02.Rev.6 14


3) Operation Method

3-1) Touch the "Power".

3-2) Touching the "Video A", it turns on, and if you retouch the"Video A", it turns on
"Video B", and if you retouch the "Video A", it turns on "Video C".
When you touch the "Video B", at the first / second / third time, it also turns on "
Video A" / "Video B" / "Video C" in a serial.
When you touch the "Video C", it appears the same result as the above.
"Video C" is the spare electrical video outlet / receptacle / socket, located at the
right bottom of rear panel in the LS-250, as I told you before.

3-3) If you touch the "Mode", you can select the division mode of display, such as
single / double / quadruple / continuous captured mode.
For instance, if you select the quadruple mode :
Focusing on the first object by endoscope and/or microscope, touch the " Capture"
and it's saved in the memory board of LS-250.

ENT Workstation_New Millennium Grand_02.Rev.6 15


Secondly, retouch the "Capture", and the first photo is released, and focusing on
the second object, and touch the "Capture", it's saved as the second photo.

Thirdly, retouch the "Capture", and the previous photo is released, and taking a
focus on the third object and touch the "Capture", it's saved as the third photo.

Fourthly, retouch the "Capture", and the previous photo is released and finally
taking a focus on the fourth object and touch the "Capture", it's saved as the finally
fourth photo.

And then, if you touch the "Division" in a row, which should be corrected "Retrieval",
you can retrieve the first / second / third / fourth / finally all four photos.

4.Operation
4-1. Local Control Panel System

2 1 3 10

5 4 6 7

① Main Power Switch : Turn on / off the power.


② Chair Control Switch : Whenever pressing the each icon, chair automatically lifts up
/ down, reclines down / up. As long as chair’s electrical cable
is connected to the ENT workstation, exactly.

ENT Workstation_New Millennium Grand_02.Rev.6 16


③ Reset : Whenever pressing the “Reset” button, the chair returns to the initial / original
position.
④ Endoscopic Visual System : Ready for taking photos by Endoscopes, using the CCD
camera, inside.
⑤ LCD Monitor, 19’’ : Whenever pressing the button, it automatically turns on.
⑥ Microscopic Visual System : Ready for taking photos by Microscope, using the CCD
camera, inside.
⑦ X-Ray Film Box: Pressing the button, its fluorescent lamps turn on for reading the X-
ray film.
⑧ Ventilation : Pressing the button, the ventilation function is ready for flowing the air
between the nose and ear, as one of parts to mitigate the otitis.
⑨ Illumination Light : Pressing the button, it turns on to help ENT doctor to diagnose &
treat patients, enlightening the surroundings.
⑩ Anti-Fog Device / Warmer : Pressing the button, the heat emits to eliminate the mist /
fog, layered on the endoscopes, in particular, winter, or
the place, big temperature gap happens.

4-2. Endoscopic Visual Imaging System


- Initialization

① It’s done at the supply of DC 4~15Volt.


② The buzzer sounds after initialization.
③ Start to recognize the signal of input after initialization.
④ Check the input signal, previously set between CVBS and YC, while distinguish
NTSC or PAL and start to operate.
⑤ Initialized in the previously used mode(Single, Double, Quad Display in Monitor)

- Video Signal In / Out


① None on the monitor until the initial video signal inputs.
② Without input, it appears black on the monitor.

ENT Workstation_New Millennium Grand_02.Rev.6 17


Initializing the input of NTSC signal, it’s automatically set NTSC color system and
it’s operated in the previous mode / status.

Ex.) PAL System, Double Mode => NTSC System, Double Mode, Live1
PAL System, Quad Mode => NTSC System, Quad Mode, Live1
NTSC System, Double Mode => NTSC System, Double Mode, Live1
NTSC System, QUAD Mode => NTSC System, Quad Mode, Live1

③ Without input, it appears black on the monitor.


Initializing the input of PAL signal, it’s automatically set PAL color system and it’s
operated in the previous mode / status.

Ex.) PAL System, Double Mode => PAL System, Double Mode, Live1
PAL System, Quad Mode => PAL System, Quad Mode, Live1
NTSC System, Double Mode => PAL System, Double Mode, Live1
NTSC System, QUAD Mode => PAL System, Quad Mode, Live1

- Buzzer
① The buzzer sounds at the initialization of memory board.
② Pressing the any button, the buzzer sounds.
③ Without inputting the signal, even pressing any button, it doesn’t sound.

ENT Workstation_New Millennium Grand_02.Rev.6 18


Single Screen Mode Operation

MENU Key indicates the shift of the display/screen in the input.


MENU Key Ditto
※ Mode Shift : Inputting the menu key on any screen, it changes to the next mode.
※ Capture Image Mode Change : Pressing down the button of image division for
some seconds, you can enter the menu, pressing its system, you can choose one of
“Auto, Frame, and Field”
※ Frame Capture Image : Trembled at the capture, it might tremble from top/left to
bottom/right. In case of non-trembling at the capture, the quality of the frame capture
image exceeds that of the filed capture image.
※ Field Capture Image : Regardless of trembling at the capture, it shows the captured
display with a minute trembling from top to bottom. Its quality lowers that of the frame
capture image.

ENT Workstation_New Millennium Grand_02.Rev.6 19


- Dual Screen Mode Operation

MENU Key indicates the shift of the display/screen in the input.


Foot Switch Ditto
IMAGE DIVISION Ditto
IMAGE DIVISION Ditto, inputting more than one second
※ Mode Initialization : Pressing the menu key at any display, it shifts to the LIVE 1,
initialized.
※ Mode Change : Pressing the menu key on the LIVE 1 of capture, it changes to the
next mode.
※ Capture Image Shift : Pressing the button of image division for some seconds, the
frame image changes to the field image.

ENT Workstation_New Millennium Grand_02.Rev.6 20


- Quad Screen Mode Operation

MENU Key indicates the shift of the display/screen in the input.


Foot Switch Ditto
IMAGE DIVISION Ditto
IMAGE DIVISION Ditto, inputting more than one second
※ Mode Change : Pressing the menu key on the LIVE 1 of capture, it changes to the
next mode.
※ Capture Image Shift : Pressing the button of image division for some seconds, the
frame image changes to the field image.
※ Capture Image Shift : Pressing the button of image division for some seconds, the
frame image changes to the field image.

ENT Workstation_New Millennium Grand_02.Rev.6 21


-PIP Mode Operation

MENU Key indicates the shift of the display/screen in the input.


Foot Switch Ditto
IMAGE DIVISION Ditto
※ Mode Initialization : Pressing the menu key at any display, it shifts to the LIVE 1,
initialized.
※ Mode Change : Pressing the menu key on the LIVE 1 of capture, it changes to the
next mode.
※ Capture Image Replacement & Replay : Pressing the button of image division at
the replay after capture, it can correct the current display or can be replaced by new
captured display. Pressing the button of image division at the live status, the previous
captured image replays.

ENT Workstation_New Millennium Grand_02.Rev.6 22


- Multi- Screen Mode Operation ( Serial Capture)

MENU Key indicates the shift of the display/screen in the input.


Foot Switch Ditto
Foot Switch Ditto, in a long input.
IMAGE DIVISION Ditto
IMAGE DIVISION Ditto, in a long input.
※ Mode Initialization : Pressing the menu key at any display, it shifts to the LIVE 1,
initialized.
※ Mode Change : Pressing the menu key on the LIVE 1 of capture, it changes to the
next mode.
※ Capture Operation Mode : Pressing foot switch at a short time, it’s captured one by
one, pressing it at a long time, it’s automatically captured at the speed, below, to the last
56th image.
Repressing foot switch in an automatic capturing, it immediately stops.
※ Replay Operation Mode : same as the capture operation mode, above.

ENT Workstation_New Millennium Grand_02.Rev.6 23


※ Speed Control on the Automatic Capture: Pressing the button of image division at
the initial screen, you can change the icon and speed control, as shown in the table,
below.

※ Common & Translucent Icon : Common icon means the frame capture, while the
translucent one indicates the field capture.
※Figure in Icon : means the speed of automatic capture which is shown on the table,
above..

- Menu

① Menu Entrance
Pressing the menu key for two seconds, the OSD, above appears with buzzer
sounded.

ENT Workstation_New Millennium Grand_02.Rev.6 24


② Key Operation in Menu
- Freeze Key : Choice and Release of items in Menu.
- Menu Key : Move up items or, increase the value of chosen item.
- Image Division : Move down items or, decrease the value of chosen item.
③ CONTRAST
- Basic value is 50, can change from 1 to 100.
- Ascending the value, dark color becomes more darker, while bright color
becomes more brighter.
④ BRIGHTNESS
- Basic value is 50, can change from 1 to 100.
- Control the brightness of screen.
⑤ COLOR
- Basic value is 50, can change from 1 to 100.
- Control the color of screen
- Setting value 0, it becomes black and white screen.
⑥ TINT
- Basic value is 50, can change from 1 to 100.
- Control the volume of color.
⑦ SHARPNESS
- Basic value is 33, can change from 1 to 100 every 33 base.
- Control the sharpness of screen.
⑧ QUALITY ▶
Enter the sub-menu of quality.
⑨ SYSTEM ▶
Enter the sub-menu of system related to the operation mode and other option.
⑩ DEFAULT
Return to the basic value, which can be changed including QUALITY and
SYSTEM.
⑪ EXIT
Entering exit, it disappears OSD menu and returns to the previous status before
menu.

ENT Workstation_New Millennium Grand_02.Rev.6 25


- Quality Menu

① X-POSITION
- Basic value is 50, can change from 40 to 60.
- Move screen to the left/right side.
② Y-POSITION
- Basic value is 50, can change from 46 to 54.
- Move screen up/down
③ OUTPUT SHARP
- Basic value is 20, can change from 0 to 100, at every 10 base.
- Control the sharpness of screen to the very minute level at the video-out socket.
④ 3DYC (3D Y/C Separation)
- Basic value is 50, can change from 0 to 100.
- Control the extent of 3D COMFILTER.
⑤ 3DNR (3D Noise Reduction)
- Basic value is 50, can change from 0 to 100.
- Control the extent of 3D Noise Reduction.
- The value appeared at the right end-side is the actual one, the noisier, the more
increased its value is.
⑥ AGC (AUTO GAIN CONTROL)
- Basic value is on, when off, input value becomes gain one.

ENT Workstation_New Millennium Grand_02.Rev.6 26


⑦ INPUT GAIN
- Basic value is 100, can change from 36 to 227.
- When inputs at CVBS , it gains.
- When inputs at S-VIDEO, INPUT Y & C gains.
- Control the signal of input gain.
⑧ EXIT
Choosing exit, returns to the previous status before sub-menu of quality.

- System Menu

① CAPTURE
- Set the saving way for the captured screen.
- AUTO : Sensing the movement of screen, in case the screen doesn’t move,
it ‘s captured as frame, vice versa, it’s captured as field.
- Frame : Screen captured as Frame.
- Field : Screen captured as Field.
- Pressing the button of image division for some seconds at the fixed display,
“Frame and Field” can be alternated.
② USB STORE (USB Only)
- AUTO : Capturing screens by foot switch, those are automatically saved in PC.
- MANUAL : Not automatically saving the captured images, it could be saved in a
specific way.

ENT Workstation_New Millennium Grand_02.Rev.6 27


③ OSD-COLOR
- Value of Change : WHITE, W-GRAY, D-GRAY, BLUE, GREEN, RED, YELLOW,
MAGENT, CYAN.
- Basic OSD Color Change
- Basic Value : W-GRAY

④ SPLIT-LINE
- Value of Change : BLACK, D-GRAY, W-GRAY ,WHITE , NONE.
- to split the screen.
- Basic Value : NONE
⑤ BUZZER-TIME
- Basic value is 50, can be changed from 0 to 100, at the increment of one(1)
- Adjust the length of Buzzer.
⑥ BUZZER-TONE
- Basic value is 50, can be changed from 0 to 100, at the increment of one(1)
- Adjust the volume of buzzer.
⑦ MODE-EDIT
- Basic Value : 1.ONE 2.DUAL 3.DDUA 4.QUAD 5.AVI
- Changeable Value : ONE, DUAL, DDUA, QUAD, PIP, MULT, AVI, NONE
- Setting this value in the menu key, mode can be changed.
⑧ EXIT
- Choosing exit, return to the previous status before system menu.

ENT Workstation_New Millennium Grand_02.Rev.6 28


5. Precaution in Use
5-1. Installation & Storage
1) Be sure to install and store this equipment out of water.
2) Paying attention to the air pressure, temperature, humidity, sunlight, ventilation,
be sure to install and store it out of place badly affected by particles, salty, and/or
ionized air.
3) Careful attention to the slope, vibration and impact, in particular, to transportation
should be made to install it.
4) The detachable castors at the bottom of it might be used for evenly keeping the
level of ground with safety.
5) Never install and store it at the chemical storage place or gas-generated spot
6) Install it in consistent with its given voltage, frequency and consumed power.

5-2. Checking before Use


1) Check whether this equipment is in safe and right operation or not.
2) Check whether electrical cables, earth line is connected, exactly, or not.
Caution ! : Be sure to connect the electrical cable into the socket at 220 volt,
which is surely earthed.
3) In case of using the other electrical goods together, be sure to follow the expert’s
instruction.
4) Be sure to use this equipment after fully understanding how to operate and handle it.

5-3. Checking in Use


1) Check the level of suction bottle from time to time.
2) Don’t give excessive force to the parts, particularly, illumination light, even it’s
operated in a very smooth way.
3) Be sure to have a very careful contact to the illumination light in a long period of
use, even it’s made with heat air-cooled duplicate glass, in addition, reducing the
possible damage of its heat by heat protection cover.
4) Be sure to inject clean water into the suction hose for several seconds, so as to
eliminate any possible foreign materials, which might prevent it from sucking the

ENT Workstation_New Millennium Grand_02.Rev.6 29


secretion in an efficient way, at every time after using it, or one or two times a day.
5) Be sure to take the proper measures, for instance, immediate stoppage of the
operation of it to make sure that patients are always in safety in abnormal cases as
well as always paying careful attention to the equipment and patients in the
operation.
6) Be always careful that equipment and other electrical goods/parts mustn’t have a
direct touch with patients.
7) Be sure to use it, properly, while paying careful attention to the other electrical
parts, safely.

5-4. Checking after Use


1) Be sure to return all of local switches to the initial status before use in a reverse
sequential order, and turn off power.
2) Be sure to pull off electrical cords after all of operation stops.
3) Be sure to always clean up this equipment after using it, to make sure that next
patients can be properly diagnosed and treated.
Caution ! : In case of disconnecting electrical cables, disconnecting patient chair’s
electrical cable, and replacing fuses, be sure to disconnect the main
power cable from the socket, at first.
4) Be sure to contact the representative, or dealer, you bought, immediately, in case

it’s out of order with proper indication to parts broken down.

5) Be sure to regularly inspect the equipment and its parts.

6) Never modify the equipment, at your disposal.

7) Be sure to understand “Reference”, “Caution”, “Warning” in the operation manual

and always stick to it.

ENT Workstation_New Millennium Grand_02.Rev.6 30


6. Maintenance & Storage
6-1. Suction Bottle
1) Be sure to inject clean water into the suction bottle after using equipment on a
daily basis.
2) Be sure to empty the suction bottle not so as to raise a level beyond its maximum
one, as shown below.
3) Be sure to put a cresol by 100cc into the suction bottle after cleaning out it.

Nominal Volume
; 3,000cc

Net Volume
; 2,500cc

APPROX.100cc OF
CRESOL WATER

<Suction Bottle Volume>

Warning ! : Be sure to clean out the suction bottle before it’s overflowing into the

sub-suction bottle, which could seriously damage the suction motor.

4) Be sure to tightly close the lid / cover of the suction bottle to avoid any leakage of

air. In case air leaks, it could decrease the suction pressure, which can’t continue

the normal operation.

5) Should waste materials with contaminated water in the suction bottle be

Over-flowed into the suction motor, the followings would happen :

- It suddenly sounds strange.

- The liquid is flowing out through the filter of the equipment.

- The fuse of the suction motor cuts off.

- It smokes in the suction motor.

ENT Workstation_New Millennium Grand_02.Rev.6 31


6-2.Spray

PUSH SWITCH
CURVED.
STRAIGHT TIP

HOSE

BLUE.
BROWN BOTTLE

BOTTLE HOLDER

CAP

SPRAY HOSE

1) Fill the bottle with the proper liquid medicine.


2) Put it into its case / basket and fasten it clockwise with the spray head and spray
knob, tightly.
3) When lifting up the spray, air compressor driven by motor, it’s ready for operation,
and push the knob of spray forward, the medicine automatically atomizes.
4) After completely using the spray, just place it onto its hollow holder, thus the air
compressor driven by motor automatically stops.

ENT Workstation_New Millennium Grand_02.Rev.6 32


6-3. Vacuum Pump, driven by Motor
1) Opening the door of the cabinet, first place the sponge and put the suction motor
on it.
2) The grey hose is connected to the air inlet, the black one to the air outlet and
connect the electrical cable to the socket, as shown below.

Warning ! : If the wastes might overflow into the suction motor from the sub-suction
bottle, the internal oil of the suction motor should be replaced, accordingly.
The wastes are filled into the main suction bottle at 2,500cc, secondly sub-
suction bottle at 1,000cc, and lastly overflows into the suction motor.
The wastes are discharged / leaked out from the filter of the suction motor.

3) If the terrible event, above should happen, it might hurt the suction motor,
seriously. In that case, first, open up the lid of the suction motor, secondly take
out water and lastly replace the oil, taking a look at the oil level gauge.
4) Be sure to regularly clean the filter to make sure that the suction motor keeps
in a good performance.
5) In case the suction motor caused by the overflow of the wastes breaks down,
don’t try to repair / maintain it, at your disposal.
6) Be sure to check the screw bottle at 250cc, since humidity / moisture generates
in the suction motor, and regularly empty it.

ENT Workstation_New Millennium Grand_02.Rev.6 33


6-4. Air Compressor, driven by Motor
1) Compressor driven by motor feed the clean air into the air tank / spray hose.
2) Connect the hose to the compressor motor, as shown below and tightly fasten it
so as not to leak the air.
3) If the air pressure is over 5kg f/ cm2 in the tank, it automatically makes pressure
switch open and discharge the air, and if it is under 1.5kg f/cm2 in the tank, it’s
automatically filled with air.
4) If the pressure switch is out of order, the relief switch automatically operates.
compressor Motor (Oilless Type).
Specification
(1) Rated Voltage : AC 220V, 50/60Hz, (2) Consumption Power : 250W’
(3) Basic Discharge Pressure : 0~2.0 kg f/cm2
(Discharge Pressure at Motor : 9.0 kg f/ cm2)

4 5 6

ENT Workstation_New Millennium Grand_02.Rev.6 34


No. Name Function

1 Anti-Vibration Case Decrease the vibration of the motor.

2 Motor Supply the power to create the air.

Solenoid Valve Turning on the power, its’ automatically closed, while

3 turning off the power, it’s automatically open to drain the

pressurized air onto the line to protect the motor.

4 Check Valve Prevent the reverse flow of the air.

5 Air Line Flow the air, made by compressor, into the air tank.

6 Filter Screen the impure air from the outside.

6-5. Spray & Suction Pressure Regulator


- Air tank regulator panel is installed onto the middle of the inside of the main body, as
follows.

ENT Workstation_New Millennium Grand_02.Rev.6 35


No. Description Function

1 Filter Filter the impure the air from its tank.

2 Suction Regulator (Red) Adjust the suction pressure.

3 Spray Regulator(Black) Adjust the spray pressure.

4 Ventilation Regulator(Green) Adjust the ventilation pressure.

5 Ventilation Gauge Display the ventilation pressure value.

6 Spray Gauge Displays the spray pressure value

7 Suction Gauge Display the suction pressure value

6-6. Air Tank Control System

3 4 5 6

8 7

11 9
10

ENT Workstation_New Millennium Grand_02.Rev.6 36


No. Name Function
1 Air tank The air first filtered is filled into the tank.
2 Safety Relief Valve In case pressure set switch of air tank doesn’t work, to
protect the air tank, it automatically opens to discharge the
air at 6 Kg/cm2, and closes at 3Kg/cm2.
3 Ventilation Solenoid Regulating the ventilation pressure, it activates at its
Valve setting pressure.
4 Spray Solenoid Valve Regulating the spray pressure, it activates at its setting
pressure.
5 Air Outlet The air outlet is connected to the filter of the spray
regulator, as shown in the above.
6 Air Inlet The air, made by compressor+ motor enters the tank.
7 Pressure Gauge Display the air tank pressure.
8 Pressure Set Switch Setting the pressure at 5Kgcm2 / 1.5Kg/cm2, the air
of Air Tank compressor driven by motor stops / starts to operate
respectively.
9 Ball valve When the solenoid valve doesn’t work, it should be
closed, as above, to keep using the spray & ventilation
system.
10 Solenoid Valve When the main power turns off, not using the spray
system, it’s open to discharge the residual air into the line.
11 Screw Bottle, 250cc When the residual air discharges out, its condensed water
is filled into the screw bottle.

ENT Workstation_New Millennium Grand_02.Rev.6 37


7. Plumbing Drawing

ENT Workstation_New Millennium Grand_02.Rev.6 38


8. Trouble Shooting

7-1. In case Main Power doesn’t work


1) Check whether the socket is operating or not.
2) Check whether the main power cable is firmly connected to the socket or not.
3) Check the fuse at 20A, 250V.
4) Check whether the main power on the local control panel is turning on or not.
5) Completely checking the procedure, above, if it doesn’t turn on, contact the dealer /
representative in your country.

7-2. In case Local Control Panel doesn’t work


1) Should not be one-touch control panel operated, turning off the main power switch
for 5 seconds, and return on it and the power of the local control panel.
2) Even it doesn’t work, contact your local representative / dealer.

7-3. In case Non-Suction occurs


1) Check whether the suction motor electrical plug is well connected to the receptacle.
2) Check whether the suction bottle lid is completely covered or not, otherwise,
suction pressure might be leaky.
2) Check whether the suction bottles might be cracked / broken, if so, those should be
replaced by the new ones.
3) Completely clean up the hose inside, which might be sometimes clogged / choked
by human secretions, such as blood, mucous materials, sputum, etc.
In order to break the clogged parts, fill the rectangular instrument tray with hot
water and suck it into the hose inside, and firmly press the parts of the hose by
hands, which are thought to be choked.

7-4. In case Terrible Odor emits from the Suction Part


1) The terrible odor is caused by the human secrets inside its hose or human secrets
stored in the suction bottles.
2) First, fill the rectangular instrument tray with hot water, sometimes, a scanty
detergent to remove the terrible odor.
3) Start-up the suction motor to be able to suck hot water into the hose inside, if hot
water doesn’t flow well inside, believed to be the points to be clogged / choked,
firmly press those parts by your hand.

ENT Workstation_New Millennium Grand_02.Rev.6 39


4) In case of using the detergent, be sure to limit to use a scanty amount, which might
cause to corrode the motor, easily.

Warning ! : - Be sure to completely empty the main suction bottle, on a daily basis.
- Be sure to completely empty the sub-suction bottle, if a scanty residue
remains, it might cause to emit the odor / make germs.
- Be sure to empty the screw bottle, at 250cc, which functions to collect the
condensed water from the motors.
If any secretion materials reversely flow into the screw bottle, it might cause
motors to be out of order.

7-5. In case of Weak Suction Pressure


1) Check whether the lid of the main, sub-suction bottles are fully covered.
2) Check whether the suction pressure gauge indicates the value at 0.88Kg/cm2 or not.
3) Adjust to control the part of the hole onto the suction head.
4) If its pressure still feels weak, it might be the out of order by suction motor.

7-6. In case of Over-Noisy Suction Pressure


1) Check whether the suction motor places in a right position, putting on the sponge.
2) If it still sounds very noisy, it might be the out of order by motor.

7-7. In case of Not Operating Spray


1) Check whether the spray motor is in operation or not.
2) Check whether the spray parts are well connected or not.
3) Check whether the spray bottle is broken or not.
4) Check the pressure gauge at 0.88Kg/cm2, otherwise, increase it by the knob of it
inside, taking out the surgical instrument tray.

7-8. In case of Not Spraying the Medication


1) It might be caused by the solidification of the medication at the spray nozzle, and /
or its head.
2) Disassemble the parts of the spray and soak them into the hot water for several
hours until the solid medication is melted, as shown in the < Fig.1, 2, 3>.
3) Connect one side of the nozzle into the suction head so as to suck not melted
medication completely, as shown in the <Fig. 4> and do the same process on the
head, as shown in the <Fig. 5>.

ENT Workstation_New Millennium Grand_02.Rev.6 40


4) Checking that the clogged medication is completely melted, assemble the parts of
the spray, as shown in the <Fig.6>.

<Fig. 1> <Fig. 2>


<Fig. 3>

<Fig. 4> <Fig. 5> <Fig. 6>

7-9. In case of Weak Pressurized Spray


1) Check whether the spray bottle is firmly fastened by its head or not.
2) Check whether the spray bottle is damaged or not.
3) It might be choked by the medication, so undertake the procedure of 7-8, above.
4) Check the pressure gauge at 2Kg/cm2, otherwise, increase it by the knob of it inside,
taking out the surgical instrument tray.

7-10. In case of Continuous Spray


1) When the knob of the spray is not pulling back in an original position, the medication
Is spraying continuously.
2) In this case, foreign materials are inserted onto the rod of the knob, so taking out
these foreign materials & apply the grease / oil onto it.

7-11. In case of Continuous Motor Operation


1) When the suction & spray put into an original position after using them, if their
motors still continue to operate, it might be caused by the micro-switch.
ENT Workstation_New Millennium Grand_02.Rev.6 41
2) In this case, carry out the following procedure, referring to the photos :

- Open the front door as shown in the <Fig.1>.


- Lifting up the suction and / or spray, check whether the micro-switch relays the
signal, which is ready to start-up the concerned motors, as shown in the <Fig.2,3>
- Lifting down the suction and / or spray, check whether the micro-switch relays the
signal, which automatically stops operating the concerned motors, as shown in the
<Fig.2, 3>.
- If it appears that the micro-switch is operating well, place the suction, and / or
spray lines and reels in a right way without avoiding any twist / mingling, as shown
in the <Fig.4> and open the door, as shown in the <Fig.5>.

<Fig. 1> <Fig. 2> <Fig.3>

<Fig. 4> <Fig. 5>

- If it appears that the micro-switch doesn’t work properly, it should be replaced by the
new one.

ENT Workstation_New Millennium Grand_02.Rev.6 42


9. Standard Accessories
No. Description Q'ty Check Accessories Box No.

1 Main Power Cable 1 1

2 Medicine Bottles Tray 1 1

3 Arm Set for Medicine Bottles Tray 1 1

4 Medicine Bottles (White) 5 1

5 Medicine Bottles (Brown) 4 1

6 Medicine Bottles (Blue) 3 1

7 Ointment Jar (White / Brown) 1/1 1

8 Rectangular Wastes Tray 1 1

9 Support Plate w/ Bolt for No.8 1 1

10 Hanger for No.8 1 1

11 Can, 50 ㎜ 2 1

12 Waste Can, 110 ㎜ 2 1

13 Round Hollow S/S Plate,110 ㎜ 1 1

15 Straight Spray 3 1

ENT Workstation_New Millennium Grand_02.Rev.6 43


No. Description Q'ty Check Accessories Box No.

16 Curved Spray 1 1

17 Fuse, 20A, 30mm 2 1

18 Vent Hose 1 1

19 Vent Tip(#1~4) 1 1

20 Suction Hose 1 1

21 Suction Tip #1~4 1 1

22 Illumination Light Arm Set 1 1

23 Motor Sponge (300 ×200 ×40mm 2 2/3

Single Motor Sponge (220 ×200


24 1 4
×40mm)
25 Two head Suction Motor 1 3

26 One Head Suction Motor 1 4

27 Two Head Comp Motor 1 2

28 Remote Controller 1 1

ENT Workstation_New Millennium Grand_02.Rev.6 44


10. Product Information Sticker

Model : New Millennium Grand Serial No. :

Date of Manufacturing : Weight : 170 Kg

Power Source : AC 220V, 50/60Hz


Power Consumption : 1,500W Most Highly Upgraded ENT Workstation
Classification : Class I, Type B

Caution :
(1) It’s only designed for diagnosing and treating patients in the field of ENT.
(2) It’s only used under the given electrical power conditions, above.
(3) Don’t install the equipment at a slope of more than min. 5 degree on the ground,
otherwise, it might be fallen down / capsized on the ground.
(4) Be sure to always operate the equipment at the normal condition, if
it’s malfunctioned, stop its operation, immediately and contact the buyer / dealer
/ distributor in your country.

MEDSTAR Co., Ltd.


Manufacturer 130-1, Ojeon-dong, Uiwang -si, Gyeonggi-do, Korea,
ZIP Code:437-819
Tel. : +82-31-457-1111, Fax. : +82-31-457-2670

ENT Workstation_New Millennium Grand_02.Rev.6 45


Warranty Certificate
n MEDSTAR Co., Ltd is to certify that this product is guaranteed and warranted for a
period of one(1) year.
n In case of not working properly, please contact the local agent / representative with
its detailed phenomena presented.
n The function, shape, etc, of the product for improvement might be subject to change
without prior notice.

Product Name ENT Workstation Warranty Period of Time


Brand Name New Millennium Grand
Manufacturing No. Serial No.:
Purchase Date MM / DD / YY Until. MM / DD / YY
Purchase Place
Telephone No.

n Complimentary After-Service Scope


The product, which didn’t work properly at the normal operation, can be repaired and
trouble-shot, free of charge, within one year from the date of purchase.

n Charged After-Service Scope

1. In case of working, properly


• When asking for the after-service, at the normal operation, you’ll be charged on the
after-service.
• So, be sure to read the operation manual carefully and get the reliable information
from the seller.

2. In case of not working, properly in misuse


• Misuse by End-User, or out of Order from the voluntary modification, or Disassembly
• Not working properly by Non-Authorized Personnel.
• Not working properly, using the Unsuitable Electricity.
• Out of Order, caused by drop in the Movement.
• When it didn’t work, using the non-authentic spare parts made by other company.

3. Others
• It didn’t work by force majeure.

ENT Workstation_New Millennium Grand_02.Rev.6 46


• Consumables Life ended.

ENT Workstation_New Millennium Grand_02.Rev.6 47