You are on page 1of 2

Vcheck SDMA

The best choice for SDMA testing

1, Stronger correlation with the Gold standard method


The correlation of Vcheck SDMA (R2=0.9624) with HPLC is stronger than that of Analyzer 'C' of
company 'I' (R2=0.8871).

* HPLC (High-Performance Liquid Chromatography): a Gold standard of SDMA

Vcheck SDMA vs. Analyzer 'C' of company 'I' SDMA


120
(Correlation with a Gold standard)

Vcheck SDMA
SDMA concentration (㎍/dl)

100
y = 1.3352x - 6.2373
80
R²= 0.9624
60
Analyzer 'C' of
40 company 'I' SDMA
20 y = 0.5859x + 0.9318
0
R²= 0.8871
0 10 20 30 40 50 60 70 80

HPLC SDMA (㎍/dL)

2, High correlation with company 'I' laboratory


With Vcheck SDMA, you can check the SDMA result values on the spot within 11 minutes.
Vcheck SDMA has a high correlation (96.9%) with company 'I' laboratory.

Correlation with company 'I' laboratory


120

100
SDMA(ug/dl)
(㎍/dl)

80 Vcheck SDMA
60
SDMA

y = 1.345x - 5.5975
R²= 0.9686
Vcheck

40
Vcheck

20

0
0 10 20 30 40 50 60 70 80

IDEXX Lab. SDMA(ug/dl)


Company 'I' laboratory (㎍/dL)
3, Excellent Spike-Recovery results
Spike-Recovery rate of Vcheck SDMA (R2=0.9686) is higher than that of analyzer 'C' of company
'I' (R2=0.8793)

Spike-Recovery Assessment
120
Measured concentration(μg/dL)

100
Vcheck SDMA
80 y = 0.9803x - 1.5463
R²= 0.9686
60

Analyzer 'C' of
40 company 'I' SDMA
y = 0.4435x + 2.2656
20
R²= 0.8793
0
0 10 20 30 40 50 60 70 80 90 100

Expected concentration (μg/dL)

4, High specificity
Vcheck SDMA has been verified to have few cross-reactions with 3 analogues
(NMMA, ADMA, L-arginine), which have similar structures to SDMA (less than 1%).

Vcheck SDMA Cross-reactivity (%)


120
SDMA NMMA
100
100% <1
Cross Reac�vity(%)

80

60

40

20 ADMA L-arginine
0 <1 <1
SDMA NMMA ADMA L-Arginine
Compound

No need to wait for laboratory test results anymore!

Cost effective and Accurate results,


Vcheck SDMA makes it possible.