Professional Documents
Culture Documents
A 60
10/15 30 ml ml
/20 ml
B 60 ml50 ml
125 ml
CONTACT INFORMATION
Liebel-Flarsheim Company LLC
1034 S Brentwood Blvd.
Suite 800
Richmond Heights, MO 63117
1-314-376-4901
1-855-266-4944
LF.PMQuality@guerbet.com
___________________________________
___________________________________
___________________________________
___________________________________
___________________________________
___________________________________
818509-A 2017-05-16
iii
FOREWORD
Congratulations on the purchase of your OptiStar® Elite MR Contrast Delivery
I—INTRODUCTION
System. The OptiStar Elite MR Contrast Delivery System represents our effort to
provide a quality product to support better health care throughout the world.
Regardless of how well equipment is designed, misuse or abuse will deny
its owner the expected quality of service. Misuse or abuse may occur
unintentionally because the proper method of operating the equipment is
unknown. Read this manual carefully before operating the OptiStar Elite MR
Contrast Delivery System. Retain this manual for future reference.
Date of Installation / /
Installing Company___________________________________________
Address_____________________________________________________
Phone No. ___________________________________________________
Software version numbers may be accessed via the System Info key on the
Results screen
Console_____________________________________________________
Powerpack___________________________________________________
818509-A 2017-05-16
iv
PATENT INFORMATION
U.S. patents protecting this product are listed at this website:
I—INTRODUCTION
www.guerbet.com/patents
DANGER!
WARNING!
CAUTION!
818509-A 2017-05-16
v
I—INTRODUCTION
Symbol Definition
Power OFF/ON
Start Injection
Stop Injection
Fluid Detect
(01) 1074619000XXXX
(21) 123ABCdefG
UDI (Unique Device Identification) located on rating label
818509-A 2017-05-16
vi
Symbol Definition
I—INTRODUCTION
REF Reference
818509-A 2017-05-16
vii
I—INTRODUCTION
Class I equipment
UL/CSA CLASSIFICATION
OPTISTAR® ELITE MR CONTRAST DELIVERY SYSTEM
(MEDICAL EQUIPMENT)
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
33SL IN ACCORDANCE WITH ANSI/AAMI ES 60601-1:2005,
CAN/CSA-C22.2 No. 60601-1 (2008), IEC 60601-1:2005,
UL 60601-1, IEC 60601-1:1998, A1:1991, A2:1995
818509-A 2017-05-16
viii
CE MARK INFORMATION
I—INTRODUCTION
0123
GUERBET
BP 57400
95943 Roissy CdG Cedex
France
(located at : 15 rue des Vanesses,
93420 Villepinte, France)
Liebel-Flarsheim Company LLC, 2111 E. Galbraith Road, Cincinnati, OH 45237
818509-A 2017-05-16
ix
TABLE OF CONTENTS
I—INTRODUCTION
Contact Information..........................................................................................................................................ii
Foreword..............................................................................................................................................................iii
Recording Serial Numbers and Software Versions................................................................................iii
Patent Information...........................................................................................................................................iv
Meanings of Symbols used in this Manual..............................................................................................iv
Meanings of Symbols used on the Injector..............................................................................................v
Classification in accordance with EN60601............................................................................................vii
Type of protection against electric shock.....................................................................................vii
Degree of protection against electric shock................................................................................vii
Degree of Protection against ingress of water............................................................................vii
Degree of Safety of Application in the presence of a Flammable anaesthetic Mixture
with Air or with Oxygen or Nitrous Oxide.......................................................................vii
Emission Standard.................................................................................................................................vii
UL/CSA Classification......................................................................................................................................vii
CE Mark Information..................................................................................................................................... viii
OVERVIEW OF THE
OPTISTAR ELITE MR CONTRAST DELIVERY SYSTEM................................ 1-1-1
1.1 Indications for Use.............................................................................................................................1-1-1
1.1.1 Contraindications................................................................................................................. 1-1-1
1.2 User Qualifications.............................................................................................................................1-2-1
1.3 Configuration of Unit........................................................................................................................1-3-1
1.4 Safety Features....................................................................................................................................1-4-1
1.4.1 Self-testing Design............................................................................................................... 1-4-1
1.4.2 Patency CheckTM Feature ................................................................................................. 1-4-1
1.4.3 Timing BolusTM Feature....................................................................................................... 1-4-1
1.4.4 Drip Mode Feature............................................................................................................... 1-4-1
1.4.5 Electrically Isolated Syringe.............................................................................................. 1-4-1
1.4.6 Syringe Clarity........................................................................................................................ 1-4-1
1.4.7 Positive Positioning of the Powerhead......................................................................... 1-4-1
1.4.8 Hand Start switch Operation............................................................................................ 1-4-1
1.4.9 Physical Stability of Remote Stand................................................................................. 1-4-2
818509-A 2017-05-16
x
1.5 Specifications......................................................................................................................................1-5-1
1.5.1 Dimensions............................................................................................................................. 1-5-1
I—INTRODUCTION
818509-A 2017-05-16
xi
CONSOLE....................................................................................................... 3-1-1
3.1 Turning the System ON....................................................................................................................3-1-1
I—INTRODUCTION
3.2 Console Display Modes of Operation.........................................................................................3-2-1
3.2.1 Main Screen............................................................................................................................ 3-2-1
3.2.2 Memory Screen, System Info Screen, Results Screen and [Enable] Key............ 3-2-1
3.3 Description of Screens.....................................................................................................................3-3-1
3.3.1 Main Screen............................................................................................................................ 3-3-1
3.3.2 Memory Screen .................................................................................................................... 3-3-5
Retrieving a Protocol from Memory...............................................................................3-3-6
Deleting a Protocol from Memory..................................................................................3-3-7
Storing a Protocol in Memory...........................................................................................3-3-9
Editing a Protocol Name.................................................................................................. 3-3-11
3.3.3 System Information Screen.............................................................................................3-3-13
Selectable Features............................................................................................................ 3-3-13
Information Displayed...................................................................................................... 3-3-14
3.3.4 Injection Results Screen...................................................................................................3-3-16
3.3.5 Enabled Screen.....................................................................................................................3-3-18
3.3.6 Drip Mode Screen...............................................................................................................3-3-19
3.3.7 Accessing the Drip Infusion Progress Screen...........................................................3-3-22
3.3.8 Drip Infusion Completed Screen...................................................................................3-3-24
POWERHEAD................................................................................................ 4-1-1
4.1 Description of Powerhead..............................................................................................................4-1-1
4.1.1 Powerhead Features............................................................................................................ 4-1-1
4.1.2 Description of Powerhead Keys...................................................................................... 4-1-4
4.2 Filling Syringes....................................................................................................................................4-2-1
4.3 Purge Air from Syringe and Tubing/Install Syringe(s)...........................................................4-3-1
4.3.1 Contrast Only Injection...................................................................................................... 4-3-1
4.3.2 Begin injecting contrast followed with a saline flush.............................................. 4-3-4
Using 50 ml / 60 ml Saline Syringe..................................................................................4-3-4
Using 125 ml Saline Syringe..............................................................................................4-3-8
4.3.3 Begin injection with Drip Mode, Patency Check, or Timing Bolus ...................4-3-11
Using 125 ml Saline Syringe........................................................................................... 4-3-13
818509-A 2017-05-16
xii
DELIVERING
AN INJECTION.............................................................................................. 5-1-1
I—INTRODUCTION
818509-A 2017-05-16
1-1-1
1
OVERVIEW OF THE
OPTISTAR ELITE MR
CONTRAST DELIVERY
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
SYSTEM
SYSTEM
The purpose of this manual is to provide instructions for using the OptiStar
Elite MR Contrast Delivery System in order to ensure patient safety. For
safe operation, it is important that this manual be read carefully and its
recommended procedures followed.
1.1.1 CONTRAINDICATIONS
Contraindications for the use of this device are determined by the prescribing
physician at the time of use based upon the contrast media package inserts.
818509-A 2017-05-16
1-1-2
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
SYSTEM
818509-A 2017-05-16
1-2-1
1—OVERVIEW OF THE
the use of this equipment and have read and understand this manual.
CONTRAST DELIVERY
OPTISTAR ELITE MR
Training made available upon installation by the manufacturer and repeat of
SYSTEM
training available upon request.
CAUTION! .
Federal (USA) law restricts this device to sale by or on the order of a physician.
(21CFR 801.109(6)(1).
818509-A 2017-05-16
1-2-2
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
SYSTEM
818509-A 2017-05-16
1-3-1
1—OVERVIEW OF THE
shown in Figure 1-3-1.
CONTRAST DELIVERY
OPTISTAR ELITE MR
The console, filter, and the power supply are not designed to be placed in the
SYSTEM
MRI area and must be set up in a separate viewing room. The power control,
powerhead and remote stand have been specifically designed so that they may
remain in the MRI area throughout the procedure without interference.
NOTE: Refer to section 1.5.5 Environmental for further information concerning
component location.
WARNING!
818509-A 2017-05-16
1-3-2
CONTRAST DELIVERY
OPTISTAR ELITE MR
A hand start switch B , a remote control device which attaches to the rear
SYSTEM
of the console, is included so that the injection process may be started and
stopped from outside the MRI area.
The remote stand D allows easy placement of the powerhead for operator and
patient convenience.
The power system consists of one Power Control E 1 , one Power Supply (with
mounting bracket) E 2 and one EFI/RFI Filter E 3 . Both the Power Control
and the Communication Cable (from the filter to the Power Control) are MR
Conditional. More information about the power system is found in Chapter 2—
Power ON / Power OFF.
818509-A 2017-05-16
1-3-3
MR MR
MR
E3
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
C
SYSTEM
A 60 ml
10/15/2 30 ml
0 ml
B 60 ml50 ml
125 ml
E1
D
Dividing
B E2 Wall
Control and
Equipment Rooms Magnet Room
AC Outlet
818509-A 2017-05-16
1-3-4
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
SYSTEM
818509-A 2017-05-16
1-4-1
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
1.4.1 SELF-TESTING DESIGN
When the OptiStar Elite MR Contrast Delivery System is switched ON, it
SYSTEM
automatically performs a series of power-up tests to monitor the system
status. If a problem is detected, an error code message will appear in the
system display. Also, during the enable process and during an injection, system
functions are constantly checked. If a fault is detected, the system will display
an error code message. Refer to Chapter 6—System Generated Messages and
Alarms for error code messages.
1.4.2 PATENCY CHECKTM FEATURE
Prior to the delivery of the main injection, a Patency Check — an injection of a
small volume of saline — can be performed to determine the integrity of the
I.V. site.
1.4.3 TIMING BOLUSTM FEATURE
Prior to the delivery of the main injection, a Timing Bolus injection — an
injection of a small volume of contrast typically followed by an injection of a
small volume of saline — can be delivered to the patient to determine the
optimal scan delay needed to capture the contrast agent in the area of interest.
1.4.4 DRIP MODE FEATURE
Prior to the delivery of the main injection, a Drip Mode injection — multiple
injections of a small volume of saline — can be delivered to the patient to keep
the fluid path open during patient set up and between injections of contrast.
1.4.5 ELECTRICALLY ISOLATED SYRINGE
All syringes are electrically isolated from any electrical contact with the injector.
1.4.6 SYRINGE CLARITY
Semitransparent syringes are used on the OptiStar Elite MR Contrast Delivery
System. If small air bubbles are present, they can be seen by visual inspection.
1.4.7 POSITIVE POSITIONING OF THE POWERHEAD
A preset friction device in the powerhead holds the syringe in the desired
position during injection.
1.4.8 HAND START SWITCH OPERATION
Use of the hand start switch allows the operator to start and stop injections
from within the view/control room.
818509-A 2017-05-16
1-4-2
the cables over the cable hanger helps keep the cables away from the wheels.
CONTRAST DELIVERY
OPTISTAR ELITE MR
WARNING!
SYSTEM
To prevent operator injury, note that injector cables laying on the floor could
possibly cause a trip hazard.
WARNING!
Maintain a constant hold on the remote stand when moving over door
thresholds, injector cables or other obstacles on the floor to help prevent the
injector from tipping over.
818509-A 2017-05-16
1-5-1
1.5 SPECIFICATIONS
1.5.1 DIMENSIONS
• Console 12.25” W X 8.5” H X 2.5” D
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
(311.15 mm W X 215.9 mm H X 63.5 mm D)
• Console Base 9.5” W X 2.5” H X 7” D
SYSTEM
(241.3 mm W X 63.5 mm H X 177.8 mm D)
• Powerhead 7” W X 21.5” H X 5.5” D
(177.8 mm W X 546.1 mm H X 139.7 mm D)
• Remote Stand 27” Diameter (base) X 47.5” H
(685.8 mm diameter X 1206 mm H)
• Power Control 10” W X 4.25” H X 15.25” D
(254 mm W X 107.95 mm H X 387.35 mm D)
• Power Supply 10” W X 4.25” H X 9.5” D
(254 mm W X 107.95 mm H X 241.3 mm D)
• Filter 4” W X 10.0” H X 3” D
(101.6 mm x 254.0 mm x 76.2 mm)
1.5.2 WEIGHT
• Console 5.8 lb. (2.6 kg)
• Power Control 16 lb. (7.26 kg)
• Powerhead (w/arm and cable) 14.5 lb. (6.57 kg)
• Remote Stand 29 lb. (13.00 kg)
• Power Supply 11.5 lb. (5.21 kg)
• Filter 10 lb. (4.54 kg)
1.5.3 POWER REQUIREMENTS
• Power Supply
Voltage 100-127 / 220-240 V~, 0.48 kW
Frequency 50/60 Hz
Current 4 Amp
Mode of Operation Continuous Operation
1.5.4 ELECTRICAL LEAKAGE
Chassis less than 300 microamps @ 115 VAC
less than 500 microamps @ 230 VAC
WARNING!
818509-A 2017-05-16
1-5-2
1.5.5 ENVIRONMENTAL
• Transport and Storage: -40° to +158° F (-40° to +70° C)
10% to 95% relative humidity
1—OVERVIEW OF THE
CONTRAST DELIVERY
DANGER!
WARNING!
WARNING!
818509-A 2017-05-16
1-5-3
Note, however, the Powerhead Cable Connector (connects into the Power
Control) is MR conditional.
Powerhead Cable Connector: The Powerhead Cable Connector is MR
1—OVERVIEW OF THE
Conditional. During installation and removal, the Powerhead Cable Connector
CONTRAST DELIVERY
OPTISTAR ELITE MR
should be carried with caution when within fields between 1,125 G and 2000G.
MR
The Powerhead Cable Connector should not be carried into fields higher than
SYSTEM
2000G. NOTE: During operation the Powerhead Cable Connector is securely
connected to the Power Control and is in an area less than or equal to 300G.
Power Control: The Power Control is MR Conditional. During installation and
removal, the Power Control should not be carried into fields higher than 2000G.
MR During injector operation, the Power Control must be located in an area with a
field less than or equal to 300G.
Communication Cable Connectors: The Communication Cable Connectors are
MR Conditional. During installation and removal, the communication cable
connectors should be carried with caution when within fields between 470G
MR
and 3,200 G. The Communication Cable Connectors should not be carried into
fields greater than 3,200 G. NOTE: During operation the Communication Cable
Connectors are securely connected to the Power Control and Filter.
Console, Power Supply and Mounting Bracket, Hand Start Switch, and Filter:
The console, Power Supply and mounting bracket, hand start switch, and the
MR Filter are not designed to be placed in the MRI area and must be set up in a
separate viewing room.
CAUTION!
If fluid is spilled on the console or power control, remove the unit from
operation and contact your authorized service personnel.
818509-A 2017-05-16
1-5-4
CONTRAST DELIVERY
OPTISTAR ELITE MR
adapter 802730)
• 15 ml, 20 ml Dotarem prefills (used with supplied DOTAclip adapter 802930)
SYSTEM
• 30 ml
• 60 ml single-use, disposable syringe
• 50 and 125 ml
1.5.7 PROGRAMMABLE FLOW RATE (A-SIDE)
• 60 ml syringe: 0.1 - 10.0 ml/sec
• 10 ml, 15 ml, 20 ml, and 30 ml syringe: 0.1 - 8.0 ml/sec
1.5.8 PROGRAMMABLE FLOW RATE (B-SIDE)
• 50 ml and 125 ml syringe: 0.1 - 8.0 ml/sec
• 60 ml single-use, disposable syringe: 0.1 - 8.0 ml/sec
1.5.9 PROGRAMMABLE VOLUME (A-SIDE)
• 10 ml, 15 ml, and 20 ml syringe: 1 ml to 20 ml
• 30 ml and 60 ml syringe: 1 ml to installed syringe size
1.5.10 PROGRAMMABLE VOLUME (B-SIDE)
• 50 ml, 60 ml and 125 ml syringe: 1 ml to installed syringe size
818509-A 2017-05-16
1-5-5
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
10,15,20 ml 60 ml 20 - 150 PSI 138 - 1035 kPa
SYSTEM
10,15,20 ml 125 ml 20 - 200 PSI 138 - 1380 kPa
30 ml 50 ml 20 - 200 PSI 138 - 1380 kPa
30 ml 60 ml 20 - 150 PSI 138 - 1035 kPa
30 ml 125 ml 20 - 200 PSI 138 - 1380 kPa
60 ml 50 ml 20 - 150 PSI 138 - 1035 kPa
60 ml 60 ml 20 - 150 PSI 138 - 1035 kPa
60 ml 125 ml 20 - 150 PSI 138 - 1035 kPa
NOTE: When a 60 ml syringe is installed into the B side, the maximum pressure
limit for both sides will be 150 PSI (1035 kPa).
NOTE: When a 125 ml syringe is installed into the B side, the maximum
pressure for the B side (only) will be 100 PSI (690 kPa).
1.5.12 PROGRAMMABLE SCAN DELAY
0 - 1:00 (1 minute: 0 seconds) in increments of 1 second
1.5.13 PROGRAMMABLE INJECT DELAY
0 - 5:00 (5 minutes: 0 seconds) in increments of 1 second
1.5.14 PROGRAMMABLE PHASE DELAY
• 0 - 600 seconds in increments of 1 second
• Pause
818509-A 2017-05-16
1-5-6
CONTRAST DELIVERY
OPTISTAR ELITE MR
818509-A 2017-05-16
1-6-1
1.6 CONSUMABLES
WARNING!
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
This device has been approved for use with only the consumables listed. Use
of other consumables could result in patient injury, operator injury, equipment
SYSTEM
damage, and/or injector performance issues (i.e. incorrect volume delivery,
incorrect flow rate delivery, incorrect pressure, etc.).
WARNING
NEPHROGENIC SYSTEMIC FIBROSIS (NSF) RISK INCREASE--Gadolinium-
based contrast agents (GBCAs) increase the risk for NSF among patients
with impaired elimination of the drugs. Avoid use of GBCAs in these patients
unless the diagnostic information is essential and not available with non-
contrasted MRI or other modalities. NSF may result in fatal or debilitating
fibrosis affecting the skin, muscle and internal organs.
• Screen patients for acute kidney injury and other conditions that may
reduce renal function. For patients at risk for chronically reduced renal
function (e.g. age > 60 years, hypertension or diabetes), estimate the
glomerular filtration rate (GFR) through laboratory testing.
818509-A 2017-05-16
1-6-2
CONTRAST DELIVERY
WARNING
OPTISTAR ELITE MR
based contrast agents (GBCAs) increase the risk for NSF among patients
with impaired elimination of the drugs. Avoid use of GBCAs in these patients
unless the diagnostic information is essential and not available with non-
contrasted MRI or other modalities.
NSF may result in fatal or debilitating fibrosis affecting the skin, muscle and
internal organs.
• The risk for NSF appears highest among patients with:
o Chronic, severe kidney disease (GFR < 30 mL/min/1.73m2), or
o Acute kidney injury.
• Screen patients for acute kidney injury and other conditions that may
reduce renal function.
For patients at risk for chronically reduced renal function (for example,
age > 60 years, hypertension or diabetes), estimate the glomerular filtration
rate (GFR) through laboratory testing [see Warnings and Precautions in
Dotarem Package Insert].
• For patients at highest risk for NSF, do not exceed the recommended
DOTAREM dose and allow a sufficient period of time for elimination of
the drug from the body prior to any re- administration [see Warnings and
Precautions in Dotarem Package Insert].
818509-A 2017-05-16
1-6-3
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
It is recommended that catheters, connectors and tubing used with the
OptiStar Elite MR Contrast Delivery System be rated to accommodate pressure
SYSTEM
specification of the installed syringe(s). If such products are rated at a pressure
below specified (as indicated in section 1.5.11), it is the responsibility of the
user to select a pressure limit that is appropriate for the product. Setting of the
pressure limit ([PSI/KPA] key) is performed on the Main screen.
801106: Y-Tubing with Single Check Valve; 152 cm (60”),
(200 psi / 1379 kPa maximum pressure)
801107: Y-Tubing with Single Check Valve; 229 cm (90”),
(200 psi / 1379 kPa maximum pressure)
601195: Coiled Tubing; 152 cm (60”),
(400 psi / 2758 kPa maximum pressure)
844010: Y-Tubing; 152 cm (60”),
(400 psi / 2758 kPa maximum pressure)
844011: Y-Tubing with Single Check Valve; 152 cm (60”),
(400 psi / 2758 kPa maximum pressure)
1.6.3 MULTIPACK DISPOSABLES
801800: 60 mL MR Syringe (2) with Single Check Valve Y-Tubing and Filling
Spikes (200 psi / 1379 kPa maximum pressure)
1.6.4 OPTIMARK (GADOVERSETAMIDE INJECTION) PREFILLED SYRINGES
• 10 ml OptiMARK (gadoversetamide injection) Prefilled Syringe
See warning for OptiMARK on page 1-6-1 and OptiMARK Package Insert for full
prescribing information. Package Insert enclosed with drug.
• 15 ml OptiMARK (gadoversetamide injection) Prefilled Syringe
See warning for OptiMARK on page 1-6-1 and OptiMARK Package Insert for full
prescribing information. Package Insert enclosed with drug.
• 20 ml OptiMARK (gadoversetamide injection) Prefilled Syringe
See warning for OptiMARK on page 1-6-1 and OptiMARK Package Insert for full
prescribing information. Package Insert enclosed with drug.
• 30 ml OptiMARK (gadoversetamide injection) Prefilled Syringe
See warning for OptiMARK on page 1-6-1 and OptiMARK Package Insert for full
prescribing information. Package Insert enclosed with drug.
818509-A 2017-05-16
1-6-4
See warning for Dotarem on page 1-6-2 and Dotarem Package Insert for full
prescribing information. Package insert enclosed with drug.
1.6.6 PREFILLED SALINE SYRINGE
• 50 ml Prefilled Sodium Chloride Injection, USP 0.9% Syringe (availability
dependent upon region)
• 125 ml Prefilled Sodium Chloride Injection, USP 0.9% Syringe (availability
dependent upon region)
818509-A 2017-05-16
1-7-1
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
NOTE: The use of accessories and components not complying with the
equivalent safety requirements of this equipment may lead to a reduced level
SYSTEM
of safety of the resulting system. Consideration relating to the choice shall
include:
• use of the accessory and components in the patient vicinity
• evidence that the safety certification of the accessory and
component has been performed in accordance to the appropriate
EN60601-1 and/or EN60601-1-1 harmonized national standard.
ACCESSORIES
P/N 802545 OptiStar Mobile Mount Kit
P/N 802617 Kit, Power Supple Wall Mount
P/N 802620 Start Switch Assembly, 1.8 m
P/N 802622 Start Switch Assembly, 6 m
P/N 802725 Kit, CPU Stand
P/N 802850 Remote Stand Assembly
COMPONENTS
P/N 802614 Kit, Filter
P/N 802930 DOTAclip Syringe Adapter, 15/20 mL
P/N 802730 OptiMARK Syringe Adapter, 10/15/20 mL (optional)
818509-A 2017-05-16
1-7-2
1—OVERVIEW OF THE
CONTRAST DELIVERY
OPTISTAR ELITE MR
SYSTEM
818509-A 2017-05-16
2-1-1
2—POWER ON/
2.1 TURNING SYSTEM POWER ON
POWER OFF
Refer to Figure 2-1-1
1. At the rear of the Power Supply, place the switch to the ON position 1A .
The green LED will illuminate to indicate power is ON 1 B .
2. At the rear of the Power Control, place the DC Power Switch to the
ON position 2 . The switch illuminates to indicate DC power is supplied to
the system.
3. At the Console, push the System ON/OFF button 3 . The button illumi-
nates to indicate power to the console and powerhead is ON. The injector
will enable a power-up check sequence indicated by flashing the syringe
size indicators located on the powerhead.
WARNING!
WARNING!
818509-A 2017-05-16
2-1-2
1A
J2 J1
O I
1B
2—POWER ON/
POWER OFF
O I
818509-A 2017-05-16
2-2-1
At the Console, push System ON/OFF button to switch the system OFF 3
Both the Power Supply and Power Control can be left ON. This allows for
quicker and easier restarts from the Console.
NOTE: If the Power Control is switched off while the console is still powered
2—POWER ON/
POWER OFF
on, an alarm message will be generated at the Console. Cycle Console power to
clear any alarm message.
818509-A 2017-05-16
2-2-2
2—POWER ON/
POWER OFF
818509-A 2017-05-16
3-1-1
3 CONSOLE
3—CONSOLE
• enter protocol parameters
• save protocols
• delete protocols
• recall protocols
• review achieved parameters of delivered protocols
• review alarm history
CAUTION!
Do not press on the touch screen with sharp or pointed items such as
fingernails, pens or pencils. Using items of this type may cause damage to your
screen, resulting in a nonworking unit. Do not allow objects, such as pens and
pencils to lay on the touch screen.
At the Console, push System ON/OFF key A to turn the Console ON.
NOTE: If no response from the console, refer to Chapter 2—Power ON/Power
OFF.
On power-up, the unit displays the ‘Initializing... Screen’ as shown in
Figure 3-1-2.
818509-A 2017-05-16
3-1-2
A
3—CONSOLE
Initializing . . .
818509-A 2017-05-16
3-2-1
3—CONSOLE
Refer to Figure 3-2-1.
Located along the lower portion of the Main screen are the following
4 keys: [Memory], [System Info] [Results] and [Enable]. These keys allow access
to their respective screens.
Memory Screen — All stored protocols are located within the Memory screen.
Accessing this screen allows the operator to recall, store, rename and delete
protocols. To access the Memory screen, press the [Memory] key A located
on the lower portion of the Main screen. More information about the Memory
screen is contained within this chapter.
System Info Screen — Adjust the Display Brightness and Beeper Volume;
change the Pressure Units, Language, Date and Time; access Alarm History and
the ability to display [Patency Check], [Phase Delay], [Inject Delay] and [Scan
Delay] keys on the Main screen via the System Info screen. Also, additional
access to the Memory screen and the Results screen is located on the System
Info screen. To access the System Info screen, press the [System Info] key B
located on the lower portion of the Main screen. More information about the
System Info screen is contained within this chapter.
Results Screen — All information pertaining to the results of delivered
injections is located within the Results screen. Also, additional access to the
System Info screen is located on the Results screen. To access the Results
screen, press the [Results] key C located on the lower portion of the Main
screen. More information about the Results screen is contained within this
chapter.
Enable D — Pressing the [Enable] key “enables” the injector (displays the
[Start] key) for delivery of a protocol after pressing the [Yes] key to the prompt
“Have you checked for air and patency?”. More information about Delivering
Injections is contained in Chapter 5.
818509-A 2017-05-16
3-2-2
4 Peak PSI
20
Timing Bolus
A B C D
Memory System Info Results Enable
818509-A 2017-05-16
3-3-1
3—CONSOLE
B: Syringe Size C — This information indicates the size of the syringe currently
installed in the B-side of the powerhead. Note that B-side parameters are
indicated by the color purple.
Protocol Name D — The name of the protocol currently displayed on the Main
screen is located in this area of the screen. An asterisk, “*” , located at the end of
the name indicates that the displayed parameters have been modified and no
longer match the originally stored parameters.
Phase E — Within a protocol, up to four phases can be input.
Side F — The letters in this column indicate the side from which the injector
will be delivering. The side for Phase 1 is always set as [A]. Sides for Phase 2
through Phase 4 can toggle among the following symbols: [A], [B], [-]. Pressing
an [A] key will toggle the key to a [B] key, indicating injecting from the B-side.
Pressing a [B] key will toggle the key to a [-] key, indicating no injecting for and
past that phase. Pressing a [-] key will toggle the key to an [A] key, indicating
injecting from the A-side.
Flow G — The values entered in this column indicate rate of delivery of the contrast
medium and saline during each respective phase. Flow is expressed in ml/sec.
Volume H — The values entered in this column indicate the volume of
contrast medium and saline to be delivered during each respective phase.
If the volume value highlights magenta intermittently, this indicates that
the programmed volume exceeds syringe volume. Use the [Max Vol] key to
automatically reset the volume of the corresponding phase to the amount
currently available in the syringe or, manually lower the value until it ceases to
highlight magenta. Volume is expressed in ml.
Max Volume I — Pressing this key inputs the Maximum Volume available in
the syringe into its corresponding volume column.
Duration J — The value in this area indicates the duration of all protocol
phases based on the entered volume and flow rate values and includes
Phase Delay. Duration does not include Timing Bolus. Duration is displayed in
minutes:seconds (00:00).
818509-A 2017-05-16
3-3-2
Change Parameter Values via the Slide Bar P — To change the value of a
parameter, select the desired parameter by touching its key. The key will highlight
Q to indicate it is active and the slide bar will display at the bottom of the screen
as shown in Figure 3-3-2. Touch the slide bar at the value required, or use the left
and right double arrows to decrease or increase the value. Use the left and right
single arrows to decrease or increase the value in smaller increments.
818509-A 2017-05-16
3-3-3
A
CHECK FOR AIR IN SYRINGE AND TUBING!
B A: 60 60ml D AB1
C B: 60 60ml J Duration 00:12
Phase Side Flow Volume Phase Delay
E F G ml/sec H ml K sec
3—CONSOLE
00:00
4 PSI
N
20
O
Timing Bolus
2 B
Q
0.1 1 Max Vol 00:00
Scan Delay
3 - 00:00
4 PSI
20
Timing Bolus
P
P P
.1 4.0 8.0
818509-A 2017-05-16
3-3-4
3 - 20
4 Timing Bolus
INJECTOR IS READY!
A: 60 58ml AAB1
Start Protocol
B: 125 93ml Duration 00:15
Phase Side Flow Volume Phase Delay
ml/sec ml sec
3 - 20
4 Timing Bolus
Drip Disable
818509-A 2017-05-16
3-3-5
3—CONSOLE
A: 60 60ml AAB1
B: 60 60ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
1.6 8
Scan Delay
3 B Max Vol
00:00
4 - Peak PSI
20
Timing Bolus
Memory Current 1
3
A
B
8.0
1.6
13
8
0 2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Page Value
B PSI 20
Main
1 2 3 4 5
818509-A 2017-05-16
3-3-6
is not displayed on Page 1, press the Page Number [2-5] 2 to display the
B
page on which the desired protocol is stored. Once the appropriate page is
displayed, press on protocol’s name key.
3. The protocol is immediately displayed on the screen 3 .
3—CONSOLE
2A AAB1
1 A
B
1.6
1.2
5
6
0
0
2
4
A 2.0 18 0
PSI
Inject
Scan
20
00:00
00:00
3
PSI 20
1 A 1.6 5 0 2 Inject 00:00
A1 3 4 Scan 00:00
PSI 20
1 A 2.6 10 0 2 Inject 00:00
A2 3 4 Scan 00:00
PSI 20
1 A 2.6 10 0 2 A 1.6 3 0 Inject 00:00
AAAB1 3 A 1.3 2 0 4 B 1.6 9 0 Scan 00:00
PSI 20
Memory Current 1
3
A
B
8
1.6
13
8
0 2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Page Value PSI 20
Main
1 2 3 4 5
2B
Figure 3-3-5 Retrieving a Protocol from Memory
818509-A 2017-05-16
3-3-7
3—CONSOLE
5. The protocol is deleted 5 .
818509-A 2017-05-16
3-3-8
A 5
A 6
Memory System Info Results Enable
1
3—CONSOLE
3A 1
Phase Side ml/sec ml sec Phase Side ml/sec ml
A 2.1 4 0 2 B 0.1 1
sec
0 Inject 00:00
AB1 3 4 Scan 00:00
Select the protocol to delete. PSI 20
1 A 2.1 4 0 2 B 2.0 18 0 Inject 00:00
AB2 3 4 Scan 00:00
PSI 20
1 A 1.6 14 0 2 B 2.0 18 0 Inject 00:00
AB3 3 4 Scan 00:00
PSI 20
1 A 1.6 14 0 2 B 2.0 18 0 Inject 00:00
AB4 3 4 Scan 00:00
3B PSI 20
1 A 1.6 5 0 2 A 2.0 18 0 Inject 00:00
AAB1 3 B 1.2 6 0 4 Scan 00:00
PSI 20
1 A 1.6 5 0 2 Inject 00:00
A1 3 4 Scan 00:00
PSI 20
1 A 2.6 10 0 2 Inject 00:00
A2 3 4 Scan 00:00
PSI 20
1 A 2.6 10 0 2 A 1.6 3 0 Inject 00:00
AAAB1 3 A 1.3 2 0 4 B 1.6 9 0 Scan 00:00
PSI 20
Memory Current 1
3
A
B
8.0
1.6
13
8
0 2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Page Value PSI 20
Main
1 2 3 4 5
2
5
Delete Edit Name Store Cancel Delete Edit Name Store Cancel
Phase Side ml/sec ml sec Phase Side ml/sec ml sec Phase Side ml/sec ml sec Phase Side ml/sec ml sec
1 A 2.1 4 0 2 B 0.1 1 0 Inject 00:00 1 A 2.1 4 0 2 B 0.1 1 0 Inject 00:00
AB1 3 4 Scan 00:00 AB1 3 4 Scan 00:00
PSI 20 PSI 20
1 A 2.1 4 0 2 B 2.0 18 0 Inject 00:00 1 A 2.1 4 0 2 B 2.0 18 0 Inject 00:00
AB2 3 4 Scan 00:00 AB2 3 4 Scan 00:00
PSI 20 PSI 20
1 A 1.6 14 0 2 B 2.0 18 0 Inject 00:00 1 A 1.6 14 0 2 B 2.0 18 0 Inject 00:00
AB3 3 4 Scan 00:00 AB3 3 4 Scan 00:00
The selected protocol will be PSI 20 PSI 20
2 18 0 Inject 00:00 2 18 0 Inject 00:00
AB4
1
3
A deleted!
1.6 14 0
4
B 2.0
Scan 00:00 AB4
1
3
A 1.6 14 0
4
B 2.0
Scan 00:00
PSI 20 PSI 20
1 A 1.6 5 0 2 A 2.0 18 0 Inject 00:00 1 A 1.6 5 0 2 A 2.0 18 0 Inject 00:00
AAB1 3Do you
B want
1.2 6to continue?
0 4 Scan 00:00 AAB1 3 B 1.2 6 0 4 Scan 00:00
PSI 20 PSI 20
4
1 A 1.6 5 0 2 Inject 00:00 1 A 1.6 5 0 2 Inject 00:00
A1 3 Yes 4 No Scan 00:00 A1 3 4 Scan 00:00
PSI 20 PSI 20
1 A 2.6 10 0 2 Inject 00:00 1 A 2.6 10 0 2 Inject 00:00
A2 3 4 Scan 00:00 A2 3 4 Scan 00:00
PSI 20 PSI 20
Main Main
1 2 3 4 5 1 2 3 4 5
818509-A 2017-05-16
3-3-9
1. Enter desired parameters on the Main screen. Press the [Memory] key 1 to
access the Memory screen.
2. Note the parameters entered in Step 1 are shown at the bottom of the
Memory screen next to the words ‘Current Value’. Press the desired Page
Number [2-5] onto which to store the protocol 2 .
3. Press the [Store] key 3 .
3—CONSOLE
4. Locate an empty key 4A . Note: If a non-empty key 4 B is selected, the key’s
stored protocol parameters will be overwritten with the parameters entered
in step 1.
5. Press the desired key onto which to store the protocol. The keyboard is dis-
played 5 to allow naming of the protocol.
6. After entering the desired protocol name, press the [Enter] key 6 on the
keyboard. Note that the injector will display the protocol in the next avail-
able empty slot.
7. The protocol parameters and name are stored 7 .
818509-A 2017-05-16
3-3-10
1.6 8
PSI
2 A Max Vol 00:00 1 A 1.6 14 0 2 B 2.0 18 0 Inject
20
00:00
AB3 3 4 Scan 00:00
Scan Delay
B 1.6 8
PSI 20
3 Max Vol
1 2 2.0 18 0 Inject 00:00
00:00 AB4 3
A 1.6 14 0
4
B
Scan 00:00
4 - PSI 20
4B
Peak PSI 1 A 1.6 5 0 2 A 2.0 18 0 Inject 00:00
AAB1 3 B 1.2 6 0 4 Scan 00:00
20 2
PSI
Inject
20
1 A 1.6 5 0 00:00
A1 3 4 Scan 00:00
Timing Bolus PSI 20
4A
1 2 Inject 00:00
3 4 Scan 00:00
PSI 20
.1 4.0 8.0 1 2 Inject 00:00
3 4 Scan 00:00
PSI 20
Memory System Info Results Enable
Memory Current 1
3
A
B
8.0
1.6
13
8
0
0
2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Page Value PSI 20
3—CONSOLE
1 2 1 2 3 4 5 Main
Memory Current 1
3
A
B
8.0
1.6
13
8.0
0 2
4
A 1.6 8 0 Inject
Scan
00:00
00:00 6
Page Value PSI 20
Main
1 2 3 4 5
Memory Current 1
3
A
B
8.0
1.6
13
8
0
0
2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Page Value PSI 20
Main
1 2 3 4 5
818509-A 2017-05-16
3-3-11
3—CONSOLE
clear the displayed name.
5. Press the [Enter] key 6 .
6. The Memory screen, along with the renamed protocol, is displayed 7 .
818509-A 2017-05-16
3-3-12
.1 4.0 8.0
1
3—CONSOLE
3B AAB1
1 A
B
1.6
1.2
5
6
0
0
2
4
A 2.0 18 0
PSI
Inject
Scan
20
00:00
00:00
3
PSI 20
1 A 1.6 5 0 2 Inject 00:00
A1 3 4 Scan 00:00
PSI 20
1 A 2.6 10 0 2 Inject 00:00
A2 3 4 Scan 00:00
PSI 20
1 A 2.6 10 0 2 A 1.6 3 0 Inject 00:00
AAAB1 3 A 1.3 0 4 B 1.6 9 0 Scan 00:00
PSI 20
Memory Current 1
3
A
B
8.0
1.6
13
8
0 2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Page Value PSI 20
Main
1 2 3 4 5
2
4 7
Delete Edit Name Store Cancel Delete Edit Name Store Cancel
Phase Side ml/sec ml sec Phase Side ml/sec ml sec Phase Side ml/sec ml sec Phase Side ml/sec ml sec
1 A 2.1 4 0 2 B 0.1 1 0 Inject 00:00 1 A 2.1 4 0 2 B 0.1 1 0 Inject 00:00
AB1 3 4 Scan 00:00 AB1 3 4 Scan 00:00
PSI 20 PSI 20
1 A 2.1 4 Select
0 protocol
2 B to rename.
2.0 18 0 Inject 00:00 1 A 2.1 4 0 2 B 2.0 18 0 Inject 00:00
AB2 3 4 Scan 00:00 AB2 3 4 Scan 00:00
PSI 20 PSI 20
1 A 1.6 14 CCCD1
0 2 B 2.0 18 0 Inject 00:00 1 A 1.6 14 0 2 B 2.0 18 0 Inject 00:00
AB3 3 4 Scan 00:00 AB3 3 4 Scan 00:00
PSI 20 PSI 20
2 2
AB4
10
3
A 1 1.6 214 30 44
B 5 2.0
6 18 70 Inject
8Scan 900:00
00:00 AB4
1
3
A 1.6 14 0
4
B 2.0 18 0 Inject
Scan
00:00
00:00
PSI 20 PSI 20
1 A 1.6 5 0 2 A 2.0 18 0 Inject 00:00 1 A 1.6 5 0 2 A 2.0 18 0 Inject 00:00
AAB1 3Q B W 1.2 E6 0R 4T Y U I OScan P00:00 AAB1 3 B 1.2 6 0 4 Scan 00:00
PSI 20 PSI 20
1 A 1.6 5 0 2 Inject 00:00 1 A 1.6 5 0 2 Inject 00:00
A1 3A S D F 4G H J K LScan - 00:00 A1 3 4 Scan 00:00
PSI 20 PSI 20
1 A 2.6 10 0 2 Inject 00:00 1 A 2.6 10 0 2 Inject 00:00
A2 3 Z X C V 4 B N M . Scan
\ 00:00 A2 3 4 Scan 00:00
PSI 20 PSI 20
1 A 2.6 10 0 2 A 1.6 3 0 Inject 00:00 1 A 2.6 10 0 2 A 1.6 3 0 Inject 00:00
AAAB1 3 CapsAOn 1.3 Cancel0 4 Space 9<- Enter CCCD1
6
B 1.6 Scan 00:00 3 A 1.3 0 4 B 1.6 9 0 Scan 00:00
5
PSI 20 PSI 20
Memory Current 1
3
A
B
8.0
1.6
13
8.0
0 2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Memory Current 1
3
A
B
8.0
1.6
13
8.0
0 2
4
A 1.6 8 0 Inject
Scan
00:00
00:00
Page Value PSI 20 Page Value PSI 20
Main 5 Main
1 2 3 4 5 1 2 3 4
818509-A 2017-05-16
3-3-13
The System Information screen is accessed by pressing the [System Info] key A
located on the Main screen.
Selectable Features
3—CONSOLE
Beeper Volume D — Selectable among Off, Low, Medium or High
Date and Time F — Press the [Change Date] key F1 and/or the [Change Time]
key F2 to access the change arrows F3 (Refer to Figure 3-3-10). Modify the date
and time as desired. Press [Set Date] key F4 and/or the [Set Time] key F5 to set.
Phase Delay H — Phase Delay is a countdown timer that delays the start of
the next phase. Toggle its key to ON to display the [Phase Delay] key on the
Main screen.
Scan Delay J — Scan Delay is a countdown timer that begins counting when
the Start command is activated and stops when the counter reaches 0 (zero).
The scan start signal will initiate once the timer reaches 0 (zero). Toggle its key
to ON to display the [Scan Delay] key on the Main screen.
818509-A 2017-05-16
3-3-14
Information Displayed
2 - 00:00
Scan Delay
00:00
Peak PSI
20
Timing Bolus
.1 4.0 8.0
System Information
B Display Brightness Low High
C Pressure Units PSI
Italian Portuguese
G Patency Check On
F1 F2 L
Change Date Change Time H Phase Delay On
Service Mode
I Inject Delay On
NOV 04 2004 10:07 M
Alarm History J Scan Delay On
818509-A 2017-05-16
3-3-15
System Information
Display Brightness Low High Pressure Units PSI
3—CONSOLE
Italian Portuguese
Patency Check On
F4 F5
Set Date Set Time Phase Delay On
Service Mode
F3 Inject Delay On
NOV 04 2004 10:07
Alarm History Scan Delay On
Alarm History
M1 Page 1
< >
System Info
818509-A 2017-05-16
3-3-16
Sec J — phase delay, in seconds, before the delivery of the next phase
PSI / KPA K — the achieved pressure. Note: An asterisk (*) located on the right
of the pressure results indicates that the injection was pressure limited. The
affected pressure value is also indicated by a yellow color.
818509-A 2017-05-16
3-3-17
2 - 00:00
Scan Delay
00:00
Peak PSI
20
Timing Bolus
.1 4.0 8.0
3—CONSOLE
B Page 1
Injection Results
C D F G H I J K
No. Date - Time Phase Side ml/sec ml sec Phase Side ml/sec ml sec PSI
1 A 2.6 10 2 Press A 7
1 02 NOV 06 10:17 3 4 Press B 20*
1 A 2.1 7 2 Press A 0
7 02 NOV 06 11:17 3 4 Press B 0
1 A 1.9 7 2 Press A 0
8 02 NOV 06 11:28 3 4 Press B 0
E1 E2
< >
System Info Main
818509-A 2017-05-16
3-3-18
The Enabled screen is displayed after pressing the [Air Check] key
located on the powerhead or by pressing the [Enable] key on the console, then
pressing the [Yes] key* for the prompt “Have you checked for air and patency?”
WARNING!
*The [Air Check] key { } on the powerhead and the [Yes] key on
3—CONSOLE
the console are not to be used in lieu of a visual check by the health care
professional administering the contrast media. It is meant as a reminder to the
operator to perform a visual check of the syringe(s) and connection tubing
prior to the delivery of an injection.
While the Enabled screen is displayed, [Side], [Flow] and [Volume] values can be
changed by the operator. The [Timing Bolus] key is active to allow the operator
to program a Timing Bolus injection.
Drip Disable
818509-A 2017-05-16
3-3-19
The [Drip] key A , available to the operator only after the injector has been
enabled, allows access to the Drip Mode. This mode allows the injector to
deliver a “drip” injection to keep the fluid path open during patient set up and
between injections of contrast.
3—CONSOLE
PATENCY ON ... 10ml VOLUME
A: 60 59ml
Start Protocol
B: 50 49ml Duration 00:07
Phase Side Flow Volume Phase Delay
ml/sec ml sec
Drip Disable
49 Cancel
Start Drip
0.5 1.0 30
818509-A 2017-05-16
3-3-20
Flow A — This value indicates the rate of delivery of the saline solution (B-side
syringe). Flow rate is expressed in ml/s and can be set from 0.5 to 1.0 ml/s.
Volume B — This value indicates the amount of saline delivered during each
drip injection. Volume is expressed in ml and can be set from 0.1 to 3.0 ml.
Interval C — This value indicates the amount of time the injector pauses
between each delivery of drip injections. Interval is expressed in seconds and
3—CONSOLE
[Start Drip] key D — Pressing this key will begin the drip injection and display
the Drip Infusion screen. Refer to Figure 3-3-11.
[Cancel] key E — Pressing this key will exit the Drip Mode screen and return to
the Main screen.
Change Parameter Values via the Slide Bar F — To change the value of a
parameter, select the desired parameter by touching its key. The key will
highlight to indicate it is active and the slide bar will display at the bottom of
the screen. Touch the slide bar at the value required, or use the left and right
single arrows to decrease or increase the value.
818509-A 2017-05-16
3-3-21
H G
Drip Mode B I Drip Volume Drip Time
40 ml 0:20:50
Hr:Min:Sec
49 Cancel E
Start Drip D
3—CONSOLE
Flow Volume Interval
A ml/sec ml
B sec
C
0.5 1.0 30
F F F
.5 0.7 1.0
818509-A 2017-05-16
3-3-22
Pressing the [Start Drip] key A on the Drip Mode screen begins the drip
injection and displays the Drip Infusion screen. The Powerhead ram of syringe
B delivers a drip injection using the operator programmed flow rate and
volume. Once the volume is delivered, the ram will pause until the operator
programmed interval has expired, and then deliver another drip injection. This
cycle will continue until the Drip Volume is delivered.
3—CONSOLE
49 Cancel
Start Drip
A
Flow Volume Interval
ml/sec ml sec
0.5 1.0 30
.5 0.7 1.0
Main
818509-A 2017-05-16
3-3-23
Description of Keys/Information
Refer to Figure 3-3-17.
[Drip] key B — Pressing this key will stop the drip infusion, exit the Drip
Infusion screen and return to the Drip Mode screen.
[Main] key C — Pressing this key will stop the drip infusion, exit the Drip
Infusion screen and display the Enabled Main screen. NOTE: Also, pressing the
[Stop] key on the powerhead will exit the Drip Infusion screen and display
the Main screen. Pressing once on the hand start switch will exit the Drip
3—CONSOLE
Infusion screen and display the Drip Mode screen.
Calculated Drip Volume and Drip Time E — These injector calculated values
are displayed in the upper right-hand corner.
Graphic representation of the volume remaining in the syringe and the end
of the drip injection F — This indicates the volume remaining in the syringe.
Also, the saline drop (located under the syringe symbol) flashes to indicate
delivery of the drip injection.
E
Drip Infusion
B Drip Volume
54 ml
Drip Time
0:19:31
Hr:Min:Sec
Main
C
Figure 3-3-17 Description of Keys and Information on the Drip Infusion Screen
818509-A 2017-05-16
3-3-24
E
Drip Infusion
B Drip Volume Drip Time
3—CONSOLE
0 ml 0:00:00
Hr:Min:Sec
Drip End
9 Stop and
return to:
Main A
COMPLETED
818509-A 2017-05-16
4-1-1
4 POWERHEAD
4—POWERHEAD
4.1.1 POWERHEAD FEATURES
Refer to Figure 4-1-1.
Syringe Holder A-side A — The syringe is placed here. Side A is used for
contrast media and can accommodate a 60 ml and a 30 ml syringe or, with the
OptiMARK adapter, a 10 ml, 15 ml, or 20 ml OptiMARK prefilled syringe. Using
a DOTAclip adapter, Side A can accommodate a 15 or 20 ml Dotarem prefilled
syringe
Syringe Holder B-side B — The 2nd syringe is placed here. Side-B is used for
saline flushing or drip solution and can accommodate a 50 ml, 60 ml or 125 ml
syringe.
Powerhead Rams C — Moveable rams which extend and retract. The ram only
pushes the syringe pushrod. It does not retract the syringe pushrod.
Syringe Size Indicators D — The syringe size indicator (A-side and B-side) is
illuminated dependent upon the installed syringe. Both indicators flash during
start up and when patency blockage occurs.
Dotarem Syringe Adapter (P/N 802930) E 2 - Syringe Holder Side A may use a
15 ml or 20 ml Dotarem contrast prefilled syringe by installing the Dotarem
818509-A 2017-05-16
4-1-2
two part syringe adapter, also known as the DOTAclip. Install the DOTAclip
offset wedge into place over the side A ram face by holding the offset wedge
tab and the powerhead and sliding the wedge over the ram face as shown,
ensuring that it snaps into place. Next install the syringe holder into the A-side
of the powerhead, as shown, ensuring that it snaps into place. To remove the
DOTAclip offset wedge, hold the powerhead and hold the offset wedge by the
tab, lift up on the offset wedge and slide it off of the ram face, as shown.
To remove the DOTAclip syringe adapter hold the syringe holder and lift up, as
shown.
The DOTAclip two part syringe adapter is only for use with Dotarem prefilled
syringes.
4—POWERHEAD
Injection Indicators F — The following chart describes the status of the A-side
F1 and B-side F2 indicator lights.
Enabled ON ON
Patency Check OFF (solid) (solid)
Enabled ON ON 0.25 sec
Patency Check ON (solid) OFF 0.25 sec
Patency Check ON 0.5 sec
OFF
B-side Ram moving OFF 0.5 sec
Pivot Adjustment Knob G — The force of the pivot rotation is factory preset.
Adjustment of the knob allows operator to increase the force.
818509-A 2017-05-16
4-1-3
F1
F2
A-SIDE
G
C
4—POWERHEAD
B-SIDE
A D
C A 60 ml
30 ml
B
50 ml
10/15/20 ml B 60 ml
125 ml
E1 E2
818509-A 2017-05-16
4-1-4
activate Auto-Retract. When the B-side ram retraction is completed, the A-side
ram will repeat the same retract movement. To stop Auto-Retract, simply press
any key on the powerhead.
WARNING!
Keep fingers out of the syringe holder when using Auto-Retract. Fingers placed
in this area while using Auto-Retract may result in operator injury.
[Reverse] key (A-side and B-side) J — Retracts the powerhead ram to the
home position on its respective side.
[Air Check] key { } K — The [Air Check] key or the [Enable] key on the
console must be pressed to enable the injector. Pressing the [Air Check] key
signals to the OptiStar Elite MR Contrast Delivery System that the operator
has checked and removed all air from the syringe(s) and connection
tubing. Once the [Air Check] key is pressed, the [Start] key is active.
WARNING!
[Patency Check Start] key { P } L — The [Patency Check Start] key is only
active once the [Air Check] key has been pressed. Pressing the [Patency Check
Start] key begins a Patency Check injection by moving the ram forward.
[Start] key { } M — The [Start] key is only active once the [Air Check] key has
been pressed. Pressing the [Start] key begins the injection. NOTE: Starting the
injection from the powerhead will prevent the use of the Inject Delay feature.
818509-A 2017-05-16
4-1-5
I J
A-SIDE H P
N
L
K
4—POWERHEAD
B-SIDE
818509-A 2017-05-16
4-1-6
4—POWERHEAD
818509-A 2017-05-16
4-2-1
WARNING!
4—PPOWERHEAD
Whenever the spike cover is removed, take proper precaution not to
contaminate the spike!
3. Retract the pushrod (fill the syringe with air) to the desired fill volume. 2
4. Insert the spike into the bottle of contrast until ONLY the first shoulder of
the spike butts up to the stopper. 3
5. Pressurize the bottle by injecting the air into the bottle through advancing
the pushrod. 4
6. Invert bottle and syringe. Advance or retract the pushrod until all air is
removed from the syringe and replaced by fluid. Retract the pushrod until
bottle is empty or desired fill volume is achieved. 5
7. Re-invert bottle and syringe to allow fluid to settle to the bottom of the
bottle. 6
8. Relieve pressure by inserting spike to second shoulder (venting position) to
allow air to escape. 7
9. Remove filled syringe from bottle.
10. Proceed to section 4.3 Purge Air from Syringe and Tubing.
818509-A 2017-05-16
4-2-2
LY
ON
USE
GLE
SIN
4—POWERHEAD
2 3 4
1/2
1
oz
11/2
oz
MONOJECT
2
oz
cc
First Shoulder
Second Shoulder
Venting Position
5 6 7
Expel
Air
818509-A 2017-05-16
4-3-1
4—POWERHEAD
The following instructions advise purging the air from the syringe by hand. If
the injector rams are to be used in lieu of hand purging, ensure the injector
is NOT ENABLED before proceeding! If the injector is enabled with Patency
Check ON, the B-side [Forward] key P is the start key for the Patency Check
injection.
1. Purge the air from the syringe by hand by advancing the pushrod until all
air is removed.
2. Reinstall the cover onto the spike. Rotate the spike counterclockwise as
shown in Figure 4-3-1 and remove from the end of the syringe. Properly
dispose of spike.
GLE
SIN
.
Figure 4-3-1 Remove the Spike
3. When performing a contrast only injection (A-side only), use straight tub-
ing. Twist off the blue end cap 3 A and insert the tube connection into the
syringe Luer thread and rotate clockwise 3 B . Refer to Figure 4-3-2. NOTE:
Do not use Y-tubing when performing a contrast only injection.
4. To remove air from the tubing, hold the syringe upright, push the syringe
pushrod and observe the contrast advancing through the tubing and out
the end of the tubing. Observe syringe and tubing to ensure all air bub-
bles have been purged.
818509-A 2017-05-16
4-3-2
5. When all air bubbles have been purged from the syringe and tubing, place
the syringe in side A of the powerhead. Ensure syringe flanges 5A remain
square with the syringe holder as shown in Figure 4-3-3 and insert the
syringe firmly into side A . NOTE: 5 B When using a 10 ml, 15 ml, or 20 ml
OptiMARK prefilled syringe, the OptiMARK syringe adapter must be used.
Install the OptiMARK syringe adapter into the A-side of the powerhead, en-
suring that the adapter snaps into place. Then install the syringe firmly into
the adapter. Refer to inset of Figure 4-3-3. NOTE: 5 C When using a 15 ml or
20 ml Dotarem prefilled syringe, the DOTAclip adapter must be used. Install
the DOTAclip offset wedge onto the A-side ram face of the powerhead, en-
suring that it snaps into place. Then install the DOTAclip syringe holder into
the A-side of the powerhead, ensuring that it snaps into place. Next, install
4—POWERHEAD
the Dotarem prefilled syringe firmly into the DOTAclip syringe holder. Refer
to Figure 4-3-3 inset.
Ensure the syringe size indicator on the powerhead and the installed sy-
ringe size match. If a discrepancy occurs, refer to Section 7-2-1 Syringe
Holder, for cleaning instructions regarding the switches.
6. Using the A-side [Forward] key, advance the powerhead ram to come into
contact with the base of the syringe pushrod. NOTE: Failure to advance the
A side ram so that it is in contact with syringe pushrod before starting the
injection protocol may cause the injector to stop and display “No Pressure
Detected! Verify Syringe Installation” message soon after beginning the
injection protocol.
CAUTION!
Ensure the base of the pushrod is positioned concentric to the ram 6 . Failure
to align the pushrod to the powerhead ram could cause damage to both the
syringe and to the powerhead. Refer to Figure 4-3-3.
7. Leave the powerhead in the vertical position (syringe pointed upward) and
the patient end of the connection tubing at same elevation or higher as
the syringe until ready for connection to patient. NOTE: Failure to keep the
patient end of tubing at same elevation or higher as syringe may cause fluid
to leak out end of tubing.
818509-A 2017-05-16
4-3-3
3A
NLY
EO
US
E
GL
SIN
3B
4—POWERHEAD
6
5A
LY
ON
USE
GLE
SIN
5B 5C
818509-A 2017-05-16
4-3-4
The following instructions advise purging the air from the syringe by hand. If
the injector rams are to be used in lieu of hand purging, ensure the injector
is NOT ENABLED before proceeding! If the injector is enabled with Patency
Check ON, the B-side [Forward] key P is the start key for the Patency Check
injection.
4—POWERHEAD
(contrast) pushrod to advance the contrast through the short tubing, past
the Y-intersection, through the long tubing 4 and out the unconnected
C
end. Observe both syringes and all tubing to ensure all air bubbles have been
purged. Refer to Figure 4-3-5.
818509-A 2017-05-16
4-3-5
5. When all air bubbles have been purged from the syringes and tubing, place the
A-side syringe (contrast) in slot A and the B-side syringe (saline) in slot B.
Ensure syringe flanges 5A remain square with the syringe holders as shown
in Figure 4-3-6. NOTE: 5 B When using a 10 ml, 15 ml, or 20 ml OptiMARK
prefilled syringe, the OptiMARK syringe adapter must be used. Install the
OptiMARK syringe adapter into the A-side of the powerhead, ensuring
that the adapter snaps into place. Then install the syringe into the adapter.
Refer to Figure 4-3-6 inset. NOTE: 5 C When using a 15 ml or 20 ml Dotarem
prefilled syringe the DOTAclip adapter must be used. Install the DOTAclip
offset wedge onto the A-side ram face of the powerhead, ensuring that it
snaps into place. Then install the DOTAclip syringe holder into the A-side of
the powerhead, ensuring that it snaps into place. Next, install the Dotarem
4—POWERHEAD
prefilled syringe firmly into the DOTAclip syringe holder. Refer to Figure
4-3-6 inset.
Ensure the syringe size indicator on the powerhead and the installed sy-
ringe size match. If a discrepancy occurs, refer to Section 7-2-1 Syringe
Holder, for cleaning instructions regarding the switches.
6. Using the A-side [Forward] key, advance the powerhead ram to come into
contact with the base of the A-side syringe pushrod. Using the B-side [For-
ward] key, advance the powerhead ram to come into contact with the base
of the B-side syringe pushrod. NOTE: Failure to advance the A side ram so
that it is in contact with syringe pushrod before starting the injection proto-
col may cause the injector to stop and display “No Pressure Detected! Verify
Syringe Installation” message soon after beginning the injection protocol.
CAUTION!
Ensure the base of the pushrod is positioned concentric to the ram 6 . Failure
to align the pushrod to the powerhead ram could cause damage to both the
syringe and to the powerhead. Refer to Figure 4-3-6.
7. Leave the powerhead in the vertical position (syringe pointed upward) and
the patient end of the connection tubing at same elevation or higher as
the syringe until ready for connection to patient. NOTE: Failure to keep the
patient end of tubing at same elevation or higher as syringe may cause fluid
to leak out end of tubing.
818509-A 2017-05-16
4-3-6
3B
3C
3A
A-SIDE
NLY
EO
US
E
GL
SIN
3C
4—POWERHEAD
3A B-SIDE
Figure 4-3-4 Attaching the Y-Tubing to the A-side Syringe and B-side Syringe
4C
4B
4A
B-SIDE A-SIDE
SINGLE
USE ONLY
SINGLE
USE ONLY
.
Figure 4-3-5 Purging Air from the Tubing
818509-A 2017-05-16
4-3-7
5A
4—POWERHEAD
NLY
EO
US
E
GL
SIN
5B 5C
818509-A 2017-05-16
4-3-8
1. Purge the air from the A-side syringe by hand by advancing the pushrod
until all air is removed.
2. Reinstall the cover onto the spike. Rotate the spike counterclockwise as
shown in Figure 4-3-1 and remove from the end of the A-side syringe. Prop-
erly dispose of spike.
4—POWERHEAD
3. To attach the Y-tubing 3 A , twist off the blue end caps 3 B and insert each
tube connection into its respective syringe Luer thread. Ensure the blue
check valve 3 C is located on the A-side syringe (contrast). Rotate the tube
endings clockwise as shown in Figure 4-3-7.
4. Place the A-side syringe (contrast) in slot A and the B-side syringe (125 ml
saline) in slot B. Ensure the flanges of the A-side syringe 4 A remain square
with the syringe holders as shown in Figure 4-3-8. NOTE: 4 When using
B
818509-A 2017-05-16
4-3-9
4—POWERHEAD
8. Leave the powerhead in the vertical position (syringe pointed upward) and
the patient end of the connection tubing at same elevation or higher as
the syringe until ready for connection to patient. NOTE: Failure to keep the
patient end of tubing at same elevation or higher as syringe may cause fluid
to leak out end of tubing.
3B
3C
3A
A-SIDE
NLY
EO
US
E
GL
SIN
3C
3A B-SIDE
Figure 4-3-7 Attaching the Y-Tubing to the A-side Syringe and 125 ml B-side Syringe
818509-A 2017-05-16
4-3-10
4A
4—POWERHEAD
LY
ON
USE
GLE
SIN
4C
4B 4D
818509-A 2017-05-16
4-3-11
4.3.3 BEGIN INJECTION WITH DRIP MODE, PATENCY CHECK, OR TIMING BOLUS
This method primes the tubing past the Y-intersection with saline.
WARNING!
The following instructions advise purging the air from the syringe by hand. If
the injector rams are to be used in lieu of hand purging, ensure the injector
is NOT ENABLED before proceeding! If the injector is enabled with Patency
Check ON, the B-side [Forward] key P is the start key for the Patency Check
injection.
4—POWERHEAD
1. Purge the air from the syringes by hand by advancing the pushrods until all
air is removed.
2. Reinstall the covers onto the spikes. Rotate the spikes counterclockwise as
shown in Figure 4-3-1 and remove from the end of the syringes. Properly
dispose of spikes.
3. To attach the Y-tubing, twist off the blue end caps and insert each tube end
into the respective syringe connector. Ensure the blue check valve is lo-
cated on the A-side syringe (contrast). Rotate the tube endings clockwise as
shown in Figure 4-3-4.
4. Holding both syringes upright, push the A-side syringe (contrast) pushrod
first and observe the contrast advance through the tubing, just past the Y-
intersection. Then, push the B-side syringe (saline) pushrod to advance the
saline through the short tubing, past the Y-intersection, through the long
tubing and out the unconnected end. Observe both syringes and tubing to
ensure all air bubbles have been purged. Refer to Figure 4-3-5.
5. When all air bubbles have been purged from the syringes and tubing, place
the A-side syringe (contrast) in slot A and the B-side syringe (saline) in
slot B. Ensure syringe flanges 5A remain square with the syringe holders
as shown in Figure 4-3-6. NOTE: 5 B When using a 10 ml, 15 ml, or 20 ml
OptiMARK prefilled syringe, the OptiMARK syringe adapter must be used.
Install the OptiMARK syringe adapter into the A-side of the powerhead,
ensuring that the adapter snaps into place. Then install the syringe into
the adapter. Refer to Figure 4-3-6. NOTE: 5 C When using a 15 ml or 20 ml
Dotarem prefilled syringe the DOTAclip adapter must be used. Install the
DOTAclip offset wedge onto the A-side ram face of the powerhead, ensur-
ing that it snaps into place. Then install the DOTAclip syringe holder into
the A-side of the powerhead, ensuring that it snaps into place. Next, install
the Dotarem prefilled syringe firmly into the DOTAclip syringe holder. Refer
to Figure 4-3-6 inset.
Ensure the syringe size indicator on the powerhead and the installed sy-
ringe size match. If a discrepancy occurs, refer to Section 7-2-1 Syringe
Holder, for cleaning instructions regarding the switches
818509-A 2017-05-16
4-3-12
6. Using the A-side [Forward] key, advance the powerhead ram to come into
contact with the base of the A-side syringe pushrod. Using the B-side [For-
ward] key, advance the powerhead ram to come into contact with the base
of the B-side syringe pushrod. NOTE: Failure to advance the A side ram so
that it is in contact with syringe pushrod before starting the injection proto-
col may cause the injector to stop and display “No Pressure Detected! Verify
Syringe Installation” message soon after beginning the injection protocol.
CAUTION!
Ensure the base of the pushrod is positioned concentric to the ram 6 . Failure
to align the pushrod to the powerhead ram could cause damage to both the
4—POWERHEAD
7. Leave the powerhead in the vertical position (syringe pointed upward) and
the patient end of the connection tubing at same elevation or higher as
the syringe until ready for connection to patient. NOTE: Failure to keep the
patient end of tubing at same elevation or higher as syringe may cause fluid
to leak out end of tubing.
818509-A 2017-05-16
4-3-13
1. Purge the air from the A-side syringe by hand by advancing the pushrod
until all air is removed.
2. Reinstall the cover onto the spike. Rotate the spike counterclockwise as
shown in Figure 4-3-1 and remove from the end of the A-side syringe. Prop-
erly dispose of spike.
4—POWERHEAD
3. To attach the Y-tubing 3 A , twist off the blue end caps 3 B and insert each
tube connection into its respective syringe Luer thread. Ensure the blue
check valve 3 C is located on the A-side syringe (contrast). Rotate the tube
endings clockwise as shown in Figure 4-3-7.
Hold the A-side syringe upright and manually push the A-side syringe
pushrod and observe the contrast advance through the short tubing, just
past the Y-intersection,
4. Place the A-side syringe (contrast) in slot A and the B-side syringe (125 ml
saline) in slot B. Ensure the flanges of the A-side syringe 4 A remain square
with the syringe holders as shown in Figure 4-3-8. NOTE: 4 When using
B
818509-A 2017-05-16
4-3-14
6. Using the A-side [Forward] key, advance the powerhead ram to come into
contact with the base of the A-side syringe pushrod. NOTE: Failure to
advance the A side ram so that it is in contact with syringe pushrod before
starting the injection protocol may cause the injector to stop and display
“No Pressure Detected! Verify Syringe Installation” message soon after be-
ginning the injection protocol.
CAUTION!
Ensure the base of the pushrod is positioned concentric to the ram 7 . Failure
to align the pushrod to the powerhead ram could cause damage to both the
syringe and to the powerhead. Refer to Figure 4-3-8.
4—POWERHEAD
7. Leave the powerhead in the vertical position (syringe pointed upward) and
the patient end of the connection tubing at same elevation or higher as the
syringe until ready for connection to patient. NOTE: Failure to keep the
patient end of tubing at same elevation or higher as syringe may cause fluid
to leak out end of tubing.
818509-A 2017-05-16
5-1-1
5 DELIVERING
AN INJECTION
This chapter contains instructions on delivering an injection. Safety instructions
to ensure patient safety are also covered here. Please read this section carefully
prior to delivering an injection.
5—DELIVERING AN
Air entrapped in the delivery system can cause patient injury or death. Always
INJECTION
verify that the syringe and tubing have been properly cleared of air just prior to
starting an injection! The OptiStar Elite MR Contrast Delivery System does not
have the capability to check for air in the syringe and tubing. The operator is
responsible for removing all air from the system.
DANGER!
WARNING!
The OptiStar Elite MR Contrast Delivery System does not have the capability to
prevent or detect an extravasation. Extravasation can be minimized through
the following precautions:
• Use the largest vein possible.
• Use lowest flow rate practical to achieve enhancement.
• Use largest gauge teflon type catheter possible.
• Insure good backflow from catheter.
• Continue to monitor from remote location.
• Instruct patient to notify operator of any abnormal pain, pressure or swelling.
818509-A 2017-05-16
5-1-2
WARNING!
This device has been approved for use with only the consumables listed in
section 1.6 Consumables. Use of other consumables could result in patient
injury, operator injury and/or equipment damage.
WARNING!
REVIEW PARAMETERS
Prior to enabling an injection, review all parameters thoroughly to ensure that
they are correct and appropriate for the procedure.
WARNING!
Prior to injecting, rotate the powerhead down to allow any small remaining air
bubbles to float away from the tip, reducing the possibility of injecting into the
INJECTION
patient.
WARNING!
WARNING!
Position this device and all accessories in a fashion that minimizes risk of
accidentally pulling out patient IV during injection.
818509-A 2017-05-16
5-2-1
5—DELIVERING AN
of the delivery system, take appropriate action to remove all air.
5. If no air is found in the syringe or tubing, connect system to patient.
INJECTION
818509-A 2017-05-16
5-2-2
5—DELIVERING AN
INJECTION
818509-A 2017-05-16
5-3-1
5—DELIVERING AN
by a “*” sign to indicate the stored parameters have been modified.
2. Confirm no air exists in the syringe(s) and tubing. If air is found in any part
INJECTION
of the delivery system, take appropriate action to remove all air before
proceeding.
3. Press the [Air Check] key on the powerhead 3A or press the
[Enable] key 3 B on the console, then press the [Yes] key 4 for the prompt
“Have you checked for air and patency?”. Note the Operator Message Line
indicates “PATENCY ON...10ml VOLUME” and the powerhead’s B-side Injector
Indicator flashes 5 . Rotate the powerhead down prior to delivery of the
Patency Check injection.
4. Press the [P] key P on the powerhead 6 to start the Patency Check
injection. Monitor the site and instruct patient to notify operator of any
abnormal pain, pressure or swelling.
5. Pressing the [Stop] key , or any other key on the powerhead 7 will im-
mediately stop the Patency Check injection. NOTE: At the completion of the
Patency Check injection, pressing the [Start] key on the powerhead will
start the main injection
6. If the injector reaches the programmed pressure limit during the patency
injection, the message “Patency Pressure Limited! Check Fluid Path.” is dis-
played 8 and the syringe indicator lights flash. Check the fluid path for any
obstructions (such as a bent catheter or closed valve), then press the
[OK] key to continue.
7. Multiple Patency Check injections can be delivered only if the syringe
contains sufficient volume. The main injection can be delivered by pressing
either the [Start Protocol] key on the console or the [Start] key on the
powerhead.
818509-A 2017-05-16
5-3-2
Italian Portuguese
4 PSI
Patency Check On
Opti
Memory System Info Results Enable star Memory Main Result
Elite
TM
TM
7 3A B: 60
Phase
60ml
Side Flow Volume
Duration 00:12
Phase Delay
ml/sec ml sec
6 00:00
P
4 PSI
A 20
60
10/15 30 ml ml
INJECTION
/20 ml
B 50 Timing Bolus
60 ml
125 ml ml
3B
CHECK FOR AIR IN SYRINGE AND TUBING!
A: 60 60ml AB1
B: 60 60ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
2 B 0.1 1 forMaxairVoland
Have you checked 00:00
patency? Scan Delay
3 - 00:00
4
4 Yes No
PSI
20
Timing Bolus
1 A 0.1 1 Max Vol 0 Inject Delay 1 A 0.1 1 Max Vol 0 Inject Delay
818509-A 2017-05-161
5-4-1
5—DELIVERING AN
of the delivery system, take appropriate action to remove all air before
proceeding.
INJECTION
3. Press the [Air Check] key on the powerhead 1A or press the
[Enable] key 1 B on the console, followed by pressing the [Yes] key 2 B for
the prompt “Have you checked for air and patency?” Note the powerhead’s
injection indicators illuminate 3 . Rotate the powerhead down.
4. Press the [Drip] key located on the displayed screen. 4
5. Adjust the Drip flow, Drip volume and Drip interval as desired using the
slide bar. 5 Drip volume can only be set to an amount that will not com-
promise the programmed protocol.
6. Press the [Start Drip] key 6A or the [Start] key on the powerhead 6 B to
start the Drip Mode injection.
7. Drip Time and Drip Volume decrements while injecting. The [Stop] key
on the powerhead 7A , the [Stop and return to: Drip] key or the [Stop and
return to: Main] key 7 B on the console are available to immediately stop the
Drip Mode injection.
8. Once the Drip Volume is delivered, the injector emits an audible tone for
two (2) seconds and the screen displays the word “COMPLETED”. Press the
[Main] key to exit the Drip Mode and display the enabled Main screen.
818509-A 2017-05-16
5-4-2
3 A: 60
CHECK FOR AIR IN SYRINGE AND TUBING!
60ml AB1
B: 60 60ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
7A 1A
Scan Delay
3 - 00:00
P 4 PSI
A
10/15
6B 60
30 ml
/20 ml ml
1B
20
Timing Bolus
B 50
60 ml
125 ml ml
1B
CHECK FOR AIR IN SYRINGE AND TUBING!
A: 60 60ml AB1
B: 60 60ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
2 B 0.1 1 forMaxairVoland
Have you checked 00:00
patency? Scan Delay
3 - 00:00
4
2B Yes No
PSI
20
5—DELIVERING AN
Timing Bolus
INJECTION
Scan Delay
3 - 00:00 Start Drip
4 Peak PSI
20 Flow
ml/sec
Volume
ml
Interval
sec
Drip Disable
.5 0.7 1.0
54 Cancel
Stop and
6A Drip End
35
return to:
Start Drip
Drip
818509-A 2017-05-16
5-5-1
5—DELIVERING AN
3. Press the [Air Check] key on the powerhead 1A or press the
INJECTION
[Enable] key 1 B on the console, followed by pressing the [Yes] key 2 B for
the prompt “Have you checked for air and patency?” Note the powerhead’s
injection indicators illuminate 3 . Rotate the powerhead down.
4. Press either the [Start] key on the powerhead 4A , the
[Start Protocol] key on the console 4 or the hand start switch 4 to start
B C
the injection. NOTE: Starting the injection from the powerhead will prevent
the use of the Inject Delay feature.
5. The Progress screen will appear on the console and the injection indicators
flash when their respective ram (A-side or B-side) is moving.
or any other key on the powerhead, with the exception of the [Start] key.
Pressing the Hand Start switch 4 also stops an injection.
C
818509-A 2017-05-16
5-5-2
3 A: 60
CHECK FOR AIR IN SYRINGE AND TUBING!
60ml AAB1
B: 60 54ml Duration 00:12
5A
Phase Side Flow Volume Phase Delay
ml/sec ml sec
1B
CHECK FOR AIR IN SYRINGE AND TUBING!
A: 60 60ml AAB1
B: 60 54ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
2 A 0.1 1 forMaxairVoland
Have you checked Pause
00:00
patency? Scan Delay
3 B 00:00
4 -
2B PSI
20
5—DELIVERING AN
Yes No
Timing Bolus
INJECTION
INJECTOR IS READY!
A: 60 60ml AAB1 4B Start Protocol
B: 60 54ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
1.6 8
Scan Delay
3 B Max Vol
00:00
4 - Peak PSI
20
Timing Bolus
4C
Drip Disable
818509-A 2017-05-16
5-5-3
3 A: 60
CHECK FOR AIR IN SYRINGE AND TUBING!
60ml AAB1
B: 60 54ml Duration 00:12
5A
Phase Side Flow Volume Phase Delay
ml/sec ml sec
1B
CHECK FOR AIR IN SYRINGE AND TUBING!
A: 60 60ml AAB1
B: 60 54ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
2 A 0.1 1 forMaxairVoland
Have you checked Pause
00:00
patency? Scan Delay
3 B 00:00
4 -
2B PSI
20
5—DELIVERING AN
Yes No
Timing Bolus
INJECTION
Memory System Info Results Enable
INJECTOR IS READY!
A: 60 60ml AAB1 4B Start Protocol
B: 60 54ml Duration 00:12
Phase Side Flow Volume Phase Delay
ml/sec ml sec
1.6 8
Scan Delay
3 B Max Vol
00:00
4 - Peak PSI
20
Timing Bolus
4C
Drip Disable
818509-A 2017-05-16
5-5-4
5—DELIVERING AN
INJECTION
818509-A 2017-05-16
5-6-1
To stop an injection before it has been completed, press the [Stop] key
on the powerhead 1 , the [Stop] key on the console 1 , or any other key on
A B
the powerhead (with the exception of the [Start] key). Pressing the hand start
switch 1 C also stops an injection.
Pressing the Console System ON/OFF button 2 will not only immediately stop
the injection, but will also turn off the console and powerhead.
1A
5—DELIVERING AN
INJECTION
P
A 60
10/15 30 ml
/20 ml ml
B 50
60 ml
125 ml ml
818509-A 2017-05-16
5-6-2
818509-A 2017-05-16
5-7-1
5—DELIVERING AN
INJECTION
818509-A 2017-05-16
5-7-2
5—DELIVERING AN
INJECTION
818509-A 2017-05-16
6-1-1
6 SYSTEM GENERATED
MESSAGES AND ALARMS
An asterisk, “*” , A will be placed after the protocol name if the operator
modifies any parameter of a displayed protocol retrieved from memory.
.1 4.0 8.0
818509-A 2017-05-16
6-1-2
Pressure
PSI
Yes No Injecting
Side Flow Volume Timing Bolus
ml/sec ml
Programmed PH1:
Protocol
A 8.0 13
PH2: B 1.6 8
Stop
Page 1
Injection Results
No. Date - Time Phase Side ml/sec ml sec Phase Side ml/sec ml sec
1 A 2.6 10 0 2 Press A 0
1 02 NOV 06 10:17 3 4 Press B 0
3 02 NOV 06 10:35
1
3
A
B
3.5
1.4
5
8
0
0
2
4
A 3.3 4 0 Press A
Press B
0
2 C
1 A 1.6 5 0 2 A 3.4 4 0 Press A 0
4 02 NOV 06 10:42 3 B 1.4 8 0 4 Press B 3
1 A 2.1 7 0 2 Press A 0
7 02 NOV 06 11:17 3 4 Press B 0
1 A 1.9 7 0 2 Press A 0
8 02 NOV 06 11:28 3 4 Press B 0
< >
System Info Main
818509-A 2017-05-16
6-2-1
PH2: A Communications
1.6 8 Max Vol Error 00:00
.1 4.0 8.0
-
Memory Timing Bolus Patency Check Results
818509-A 2017-05-16
6-2-2
Alarm
Number Fault Message
1 Pressure Limit Alarm
2 Over Volume Alarm
3 Over Flow Rate Alarm
4 Motor Timing Alarm
6 Motor Alarm
9 Invalid Powerhead Pressure Measurement Alarm
13 Forward/Reverse Limit Sensor Alarm
14 Motor Control Watchdog Alarm
15 Syringe Size Sensing Alarm
MESSAGES AND ALARMS
6—SYSTEM GENERATED
818509-A 2017-05-16
7-1-1
7 CARE OF UNIT
7—CARE OF UNIT
818509-A 2017-05-16
7-1-2
818509-A 2017-05-16
7-2-1
-Install the syringe adapter in the A-side to verify the 10/15/20 ml indicator
illuminates on the keypad.
- Install a 60 ml syringe in both sides to verify the 60 ml indicator illuminates
on the keypad.
- Install a 30 ml syringe into the A-side to verify the 30 ml indicator
illuminates on the keypad.
- Install a 50 ml syringe into the B-side to verify the 50 ml indicator
illuminates on the keypad.
7—CARE OF UNIT
If indicators do not illuminate properly, clean the switches to ensure proper
operation. Refer to section 7.4.1 Syringe Holder.
818509-A 2017-05-16
7-2-2
818509-A 2017-05-16
7-3-1
7.3 CLEANING
CAUTION!
DO NOT use any solvents or alcohol based cleaners to clean the Powerhead,
Console, Power Control or Power Supply.
WARNING!
Clean the switches and the area around the switches thoroughly to ensure
proper switch operation. Improper switch operation may result in patient injury
or equipment malfunction. Refer to A in Figure 7-3-1 for location of switches.
7—CARE OF UNIT
818509-A 2017-05-16
7-3-2
818509-A 2017-05-16
A-1
A ELECTROMAGNETIC
COMPATIBILITY GUIDANCE
Note: Electromagnetic Compatibility Tables can be found in the Installation,
Service and Parts Manual (Part Number 818507).
The OptiStar Elite MR Contrast Delivery System meets EN 60601-1-2 for Class B
conducted and radiated emissions and EMI immunity.
WARNING!
The OptiStar Elite MR Contrast Delivery System should not be used adjacent to
or stacked with other equipment. If adjacent or stacked use is necessary, the
system should be observed to verify normal operation in the configuration in
which it will be used.
COMPATIBILITY GUIDANCE
A—ELECTROMAGNETIC
WARNING!
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
WARNING!
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the OptiStar Elite MR Contrast Delivery System ,
including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
CAUTION!
Portable and mobile RF communication equipment can affect operations of the
OptiStar Elite MR Contrast Delivery System .
CAUTION!
Maintaining programmed flow rate (unless pressure limited) and delivered
volume are essential performance features of the OptiStar Elite MR Contrast
Delivery System. Interference from other equipment could cause degradation
of these functions, resulting in improper operation.
818509-A 2017-05-16
A-2
NOTE: The EMISSIONS characteristics of this equipment make it suitable for use
in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment
might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating
or re-orienting the equipment.
NOTE: If any anomalies in the injector performance are noticed, identify
devices within the immediate area that are capable of producing
electromagnetic interference and call a qualified service representative.
COMPATIBILITY GUIDANCE
A—ELECTROMAGNETIC
818509-A 2017-05-16
I-1-1
Index
Symbols E
* Asterisk: Protocol Retrieved from Memory... 6-1-1 Editing a Protocol Name 3-3-11
Electrical Leakage 1-5-1
A Electrically Isolated Syringe 1-4-1
Accessories 1-7-1 Enable 3-2-1
A: Syringe Size 3-3-1 Enabled Screen 3-3-18
AutoRetract 4-1-4 Enter Parameters 5-2-1
Environmental 1-5-2
B
F
Begin injecting contrast followed with a saline flush
4-3-4 Filling Syringes 4-2-1
Begin injection with Drip Mode, Patency Check, or Tim- Flow 3-3-1
ing Bolus 4-3-11 H
B: Syringe Size 3-3-1
Hand Start Switch Operation 1-4-1
C
I
Care of Unit 7-1-1
CE Mark Information viii Indications for Use 1-1-1
Change Parameter Values via the Slide Bar 3-3-2, 3-3- Inject Delay 3-3-2
20 Injection Results Screen 3-3-16
Check Patency of I.V. Site 5-3-1 Inject Mode Screen 3-3-1
Classification in accordance with EN60601 vii Inject Results 1-5-6
I—Index
Cleaning 7-3-1
Configuration of Unit 1-3-1 K
Connect Patient 5-2-1 KPA / PSI 3-3-2
CONSOLE 3-1-1
Console Display Modes of Operation 3-2-1 L
Consumables 1-6-1
Contrast Only Injection 4-3-1 Load Syringe(s) 5-2-1
D M
Daily Inspection 7-2-1 Meanings of Symbols used in this Manual iv
Deleting a Protocol from Memory 3-3-7 Meanings of Symbols used on the Injector v
Delivering a Drip Mode Injection 5-4-1 Memory Screen 3-2-1, 3-3-5
Delivering
an Injection 5-1-1 O
Delivering the Main Protocol 5-5-1 Operator Message Line 3-3-1
Description of Powerhead 4-1-1 Overview of the OptiStar Elite 1-1-1
Description of Powerhead Keys 4-1-4
Description of Screens 3-3-1
Dimensions 1-5-1
Drip Infusion Progress Screen 3-3-22, 3-3-24
Description of Keys/Information 3-3-23
Drip Mode Screen 3-3-19
Description of Parameter Keys/Information 3-3-20
Duration 3-3-1
818509-A 2017-05-16
I-1-2
P T
Patency Check Feature 1-4-1 Timing Bolus 3-3-2
Phase 3-3-1 Timing Bolus Feature 1-4-1
Phase Delay 3-3-2 Turning System Power OFF 2-2-1
Physical Stability 1-4-2 Turning System Power ON 2-1-1
Positive Positioning of the Powerhead 1-4-1 Turning the System ON 3-1-1
Powerhead 4-1-1
Powerhead Features 4-1-1 U
Power Requirements 1-5-1 UL/CSA Classification vii
Prematurely Stopping an Injection 5-6-1 User Qualifications 1-2-1
Pressure Limited Injection 6-1-2 Using 125 ml Saline Syringe 4-3-8, 4-3-13
Preventive Maintenance & Service 7-1-1
Program. Drip Injection Parameters (B-Side) 1-5-6 V
Programmable Flow Rate (A-side) 1-5-4
Programmable Flow Rate (B-side) 1-5-4 Volume 3-3-1
Programmable Inject Delay 1-5-5 Volume Key Flashing a Magenta Color 6-1-1
Programmable Pressure Limit (A-side) 1-5-5, 3-3-2
Programmable Scan Delay 1-5-5 W
Programmable Volume (A-side) 1-5-4 Weight 1-5-1
Programmable Volume (B-side) 1-5-4
Protocol Memory 1-5-6
Protocol Name 3-3-1
PSI / KPA 3-3-2
R
Recording Serial Numbers and Software Versions iii
Removing Air from Syringe and Tubing 4-3-1
Restarting an Prematurely Stopped Injection 5-7-1
I—Index
S
Safety Features 1-4-1
Safety Instructions 5-1-1
Scan Delay 3-3-2
Self-testing Design 1-4-1
Service 7-1-1
Side 3-3-1
Specifications 1-5-1
Storing a Protocol in Memory 3-3-9
Syringe Clarity 1-4-1
Syringe Sizes 1-5-4
System Generated Alarms 6-2-1
System Generated Messages/Indications 6-1-1
System Information Screen 3-3-13
818509-A 2017-05-16
NOTES
Liebel-Flarsheim Company LLC
2111 E. GALBRAITH ROAD CINCINNATI, OHIO 45237-1624