ISOLATED POWER PANELS

THE SAFEST SYSTEM YOU CAN HAVE IN YOUR OPERATING ROOM


THE RECOGNIZED MODEL IDP MODEL IDC
LEADER IN ISOLATED ISOLATED
ISOLATED POWER POWER PANEL POWER CENTER
PG LIFELINK has been designing The IDP Isolated Power Panel repre- The IDC Isolated Power Center pack-
and manufacturing isolated power sents our standard system. All of the ages our basic panel with optional
systems since 1957. Our reputa- basic elements are included: hospital devices such as receptacles and
tion for superior quality and an un- grade isolation transformer, line isola- ground jacks. Typically installed in
surpassed dedication to customer tion monitor, branch circuit distribution, patient care areas such as ICUs, CCUs,
and a room reference grounding bus. recovery areas, cardiac catheterization
service has enabled us to become
recognized as the #1 name in Iso- labs, emergency rooms, dialysis units,
This package is available in ratings and delivery rooms, these units offer a
lated Power Products. from 3 to 15 kVA, with several optional compact solution for providing local-
voltage configurations. The panel may ized protection to a small area.
As a pioneer in the field, we have be installed directly inside the patient
introduced many innovative solu- care area or in a nearby corridor. When
tions over the years. Some of the Load and voltage ratings are the same
panels are not installed locally, remote
distinctive advantages that we offer as for basic IDP Power Panel models.
annunciators are utilized to provide in-
include: Available device accessories include
dication of system status inside the area
being served by the panel, per NEC standard 20A hospital grade duplex
• No-cost engineering assistance requirements. and twist-lock receptacles as well as
for our customers. 30A twist-lock ground jacks. Other
Healthcare facilities are designed to options such as our Operating Room
• Fast response to critical delivery meet the increasing demand for safe, Stereo Systems are also available.
schedules by stocking an exten- reliable power. Our Isolated Power Sys-
sive inventory of standard mod- tems fulfill the safety requirements for The IDC Isolated Power Center’s 16
els. special protection in “wet locations” available branch circuits can feed re-
specified by the NEC. They provide an mote mounted receptacles and perma-
• Back boxes can be shipped dur- additional safeguard against interrup- nently installed equipment as well as
ing rough-in stage. Interiors and tion of power under fault conditions. the panel mounted devices.
trim follow as required to limit
potential damage and on-site
storage issues.

• Several major circuit breaker


brands are available to match fa-
cility standard.

• On-site start-up and mainte-


nance services are available.

• Custom retro-fit for your existing


installation.

• We service our equipment and


provide a 2-year warranty against
defects.

At PG LIFELINK we know that our


customers have requirements that
are not always served with off-the-
shelf items, therefore, we welcome
the opportunity to develop cus-
tom applications when needed.
MODEL DIDP QUALITY COMPONENTS ENSURE
DUPLEX ISOLATED QUALITY PRODUCTS
POWER PANEL PG L L ’ Nowhere is IFE INK S commitment to quality, and above all safety,
more evident than in the components that go into the manufacture of our
Isolated Power Panels.
The DIDP Duplex Isolated Power Panel
provides two complete isolated power These components include:
systems in a single enclosure. This com- The single phase, stacked core ISOLATION TRANSFORMER with elec-
pact design offers a reduced footprint 1 trostatic shield, designed for low leakage and quiet operation. The maxi-
that takes up less valuable wall space. mum leakage current is only 10 microamps for 3 and 5 kVA, 14
The panel may be installed directly in- microamps for 7.5 kVA, 20 microamps for 10 kVA and 25 microamps
side the patient care area or in a nearby for 15 kVA.
corridor. When panels are not installed
Ratings available – 3, 5, 7.5, 10 or 15 kVA.
locally, remote annunciators are uti-
lized to provide indication of system Primary voltage – 120, 208, 220, 240, 277, 347, 480 or 600.
status inside the area being served by Secondary voltage – 120, 208, 220, or 240.
the panel. It also simplifies installation,
saving time and money.
2 The PG LIFE LINK MARK IV LINE ISOLATION MONITOR offers
continuous monitoring with non-switching circuitry. The LIM signals
Duplex panels are typically specified are plainly visible on the front of the panel and are not obscured when
to provide Isolated Power protection the circuit breaker door is opened.
for both Normal and Emergency sys-
In accordance with code requirements, we use a two-pole BOLT ON
tems inside the same area. Internal bar-
riers ensure separation between the
3 CIRCUIT BREAKER PANEL. It houses the primary main circuit breaker
and up to sixteen secondary breakers. U.L. Listed.
two systems.
The LOCKABLE DOOR over the circuit breaker panel features a
These units can also be used to provide 4 Corbin lock, preventing unauthorized operation.
additional load capacity without in-
The FIELD WIRING COMPARTMENT is formed by a baffle and fastened
creasing the total hazard current above
nominal level.
5 to the chassis.
A TERMINAL BLOCK is supplied for the connection of remote signal
6 conductors.
The GROUND BUS is used for connection of all internal and branch
7 ground conductors.
The heavy-duty BACK BOX is constructed of galvanized steel and
8 is designed for either flush or surface mounting.
The CHASSIS mounts inside the back box and supports the isolation
9 transformer, LIM and circuit breaker panel simplifying installation.
A FRONT TRIM PANEL fabricated from #12 gauge, 304 stainless steel
10 polished to a #4 satin finish, comes standard. A continuously hinged
front trim panel is available upon request at an additional charge.

Other features include:


• Low monitor hazard current of • Custom designs available
25 microamps
• LED panel meter
• 5 or 2 milliamp alarm point • LED indicator lights
• Self-testing and self-calibrating • Self-resetting fuses
• Sensitivity to LIM “loss of ground” • Proven reliability
APPLICATION NOTES
9 5
1 ❖ When an isolated power panel or any other
electrical device is installed in an operating
room designated as a flammable anesthetiz-
ing location, the enclosure must be located
6 at least 5 feet above the floor.

❖ Conductors used for branch circuit wiring


should be stranded copper having a cross-
2 linked polyethylene insulation or equivalent
with a dielectric constant of 3.5 or less. Type
XHHW is suitable for this purpose. Wire pull-
3 ing compounds, other than talc, produce an
4 adverse effect upon the dielectric constant
7
of the conductor insulation and are specifi-
cally prohibited by the NEC when pulling
10 8 branch circuit wiring energized from the sec-
ondary of the isolation transformer.

❖ Conduit routings should be as direct and as


PRODUCTS DESIGNED WITH short as possible to minimize leakage current.

SAFETY IN MIND ❖ Metallic conduit is required for all hazard-


ous anesthetizing locations.

PG LIFELINK Isolated Power Panels are designed to provide


ungrounded service for a variety of applications.

When properly maintained, these systems supply uninterrupted ISOLATED POWER


electric power to critical equipment with inherent protection
against micro-shock hazards, even in the event of a single line PANEL INSTALLATION
fault to ground. The intelligent Line Isolation Monitor continu-
ously monitors the integrity of the system and incrementally
predicts the available hazard current were a single line fault to Model IDP Isolated Power Panel, Model IDC Iso-
occur. lated Power Center and Model DIDP Duplex Iso-
lated Power Panel are designed for quick and
UL Listings – 1047 “Isolated Power Systems Equipment” simple field installation. The back box is installed
1022 “Line Isolation Monitors” first, during the “roughing in” phase, allowing
the contractor to mount the box as the walls are
CSA Listed – C22.2 No. 29 being erected. The back box must be located in
accordance with NEC requirements. The
subpanel and front trim are installed during the
NFPA-99 requires isolated power in “final” construction phase.
flammable anesthesia and The transformer, LIM, circuit breakers, and all
“wet locations”, and any area other components are preassembled on the
subpanel and wired at the factory. The contrac-
where the interruption of power tor needs only to mount the subpanel inside the
back box and connect the primary feeder, branch
cannot be tolerated. circuits and signal conductors. Ample space is
provided for field wiring.
ORDERING INFORMATION
PRIMARY SECONDARY SECONDARY CIRCUIT CIRCUIT GROUND
VOLTAGE VOLTAGE CIRCUIT BREAKER BREAKER BACKBOX RECEPTACLE RECEPTACLE JACK SPECIAL
TYPE kVA (VAC) (VAC) BREAKER QTY. SPACES MANUFACTURER MOUNTING FRONT LIM SIZE TYPE QTY. QTY. CODES
IDP 3 A – 110 A – 110 0 0 D– SQUARE-D F – FLUSH V – VENTED A – MARK IV IDP D – DUPLEX 20A 0 0 M – SEC. MAIN
IDC 5 B – 120 B – 120 1 1 G– GE S – SURFACE N – NON-VENTED B – MARK III 02 – 43×22×6 S – SIMPLEX 20A 1 1 H – HINGED FRONT
DIDP 7 C – 208 C – 208 2 2 C– EATON X – SPECIAL 03 – 48×22×6 T – TWISTLOCK 20A 2 2 X – SPECIAL SPEC.
10 D – 220 D – 220 3 3 S– SIEMENS 04 – 54×22×6 Z – OTHER 3 3
Add Codes
15 E – 240 E – 240 4 4 05 – 60×25×10 4 4 As Required
F – 277 5 5 06 – 60×25×12 5 5 In Order
G – 480 6 6 6 6
H – 347 7 7 IDC 7 7
J – 600 8 8 11 – 54×22×6 8
9 9 12 – 60×25×10 For IDC Panels Only
10 10 13 – 60×25×12 (Leave Blank For Others)
11 11
12 12 DIDP
13 13 20 – 70×30×12
14 14 21 – 70×30×14
15 15
16 16
Example: DIDPL10FB142R15FB160GFNA21 Duplex Panel:
Maximum Number of Left: 10kVA, 277/120V, 14 CB, 2 Future
Circuit Breakers + Spaces
Right: 15kVA, 277/120V, 16 CB 0 Future, GE Breakers,
Equals 12 or 16
Flush Mount, Non-Vented, Mark IV LIM
ISOLATED POWER PANEL SIZES
Panel sizes listed to the left are standard for vented
MODEL # kVA SIZE # BRANCH BREAKERS BACK BOX SIZE* FLUSH TRIM SIZE* front panels. For non-vented IDP and IDC units, 7.5
IDP 3, 5 1 – 16 43”H x 22”W x 6”D 44-3/4”H x 23-3/4”W kVA panels are 10” deep, and 10-15 kVA panels are
IDP 7.5 1 – 12 48”H x 22”W x 6”D 49-3/4”H x 23-3/4”W 12”deep. All non-vented DIDP panels are 14” deep.
IDP 7.5 13 – 16 54”H x 22”W x 6”D 55-3/4”H x 23-3/4”W
IDP 10, 15 1 – 12 60”H x 25”W x 10”D 61-3/4”H x 26-3/4”W
IDP 10, 15 13 – 16 60”H x 25”W x 10”D 61-3/4”H x 26-3/4”W CALL TOLL-FREE
IDC 3, 5, 7.5 1 – 16 54”H x 22”W x 6”D 55-3/4”H x 23-3/4”W 800-287-4123
IDC 10, 15 1 – 16 60”H x 25”W x 10”D 61-3/4”H x 26-3/4”W
FOR ASSISTANCE
DIDP 3 – 15 1 – 16 70”H x 30”W x 12”D 71-3/4”H x 31-3/4”W
*DIMENSIONS SUBJECT TO CHANGE WITHOUT NOTICE
WITH CUSTOM ORDERING
FOR CONVENIENCE AND SAFETY, USE GENUINE
PG LIFELINK ACCESSORIES
The following products are built to ensure patient safety, while making your laser and x-ray isolated power systems
easier to operate and monitor. For a complete listing and more details, refer to the Isolated Power Accessories brochure.

SINGLE REMOTE REMOTE ANNUNCIATOR PANEL MASTER REMOTE GROUND MODULES


ANNUNCIATOR PANEL WITH METER ANNUNCIATOR PANELS Model GJP contains ground
Model DRA-1 provides remote Model DRA-1S provides remote Model DMRA monitors multiple jacks, copper equipotential
indication of LIM signals. indication of LIM signals. systems from one location. ground bus and one main lug.

POWER & GROUND MODULES POWER & GROUND MODULES GROUND CORD ASSEMBLIES MASTER GROUND MODULES
Model RRP remote receptacle Model RRP combine receptacles Model GC flexible #10 copper Model GJB provides control
panels fit customer furnished and ground jacks into a single panel conductor encased in green collection points for all ground
electrical gang boxes. with custom back box included. neoprene insulation. conductors in patient areas.

PG LIFELINK
manufactures a complete line of patient safety
equipment including Isolated Power Panels,
Line Isolation Monitors, Surgical Facility Panels,
Operating Room Pedestals, Portable Laser
and X-Ray Panels, Digital Clock/Timers,
Patient Service Consoles, Stereo Systems
for Operating Rooms, X-Ray Film Viewers and
Isolated Power Accessories.
PG LIFELINK
Field checkout, equipment certification 167 GAP WAY • ERLANGER, KY 41018
and staff orientation by a PHONE: 859-283-5900 • FAX: 859-372-6272
PG LIFELINK TOLL-FREE: 800-287-4123
factory representative is available. EMAIL: sales@pglifelink.com • WEBSITE: www.pglifelink.com
A Member of the Halma Group

Call toll-free 800-287-4123


for more information.
IPP/3