P/N: 19695-A
Rev. 4/18/2016
Contact Info:
2 39
Contents
4 Before You Start
4 Introduction
5 Safety Information
12 Getting Started
13 Package Contents
14 Device Components
15 Installation
15 • Tool Holder
15 • Waste Container
15 • HydraFacial® Handpiece
16 • HydroPeel® Tip
16 • Specialty Vial
17 • Skin Solutions
18 • Optional Components
20 Operating Instructions
20 • On/Off, Standby, and Sleep Modes
- This page is intentionally left blank - 21 • Touchscreen
21 • Adjust Fluid Flow
21 • Adjust Vacuum Strength
22 • Activate LED Lights (optional)
22 • Skin Imaging Camera (optional)
23 System Cleaning & Preventive Maintenance
23 • After Each Service
23 • Weekly Cleaning
24 • HydraFacial® Handpiece
26 • Waste Container
26 • Touchscreen
26 • Removable Front Door
27 • Wet Diamond Aggressive Abrasion (Medical Use Only)
28 • Mechanical Lymphatic Drainage Therapy Handpieces (optional)
28 • LED Light Therapy Handpieces (optional)
28 • Skin Imaging Camera (optional)
29 Troubleshooting
32 Limited Warranty
34 Appendix: Client Questionnaire & Consent Form
38 3
INTRODUCTION
Thank you for purchasing the HydraFacial MD® Elite™ System, a revolu-
tionary breakthrough in aesthetic technology! The HydraFacial® is the
next generation of exfoliation and skin rejuvenation done simultaneous-
ly. The process removes dead skin cells and impurities while simultane-
ously Vortex-Fusing™ the underlying new skin with cleansing, hydrating
and moisturizing skin solutions and antioxidants. The procedure is
soothing, refreshing, non-irritating and very effective. You can have the
utmost confidence in this leading-edge system manufactured in the USA
by Edge Systems LLC.
Environmental Conditions:
Temperature: Normal (10-40º C)
Relative Humidty: 30-75% RH
Altitude: 700-1060 hPa
4 37
SAFETY INFORMATION
PRODUCT MARKING SYMBOLS
Symbol Description
Attention: Refer to accompanying documentation.
Aufmerksamkeit: Beziehen Sie sich begleitend auf Dokumentation.
Attention : Consultez la documentation d’accompagnement.
Separate collection for electronic.
Trennen Sie Sammlung für elektronisch.
Collecte sélective pour l’électronique.
Direct Current
Gleichstrom
Courant continu
Alternating Current
Wechselstrom
- This page is intentionally left blank - Courant alternatif
ON (Power)
EIN (Stromversorgung)
ALLUME (Alimentation)
OFF (Power)
AUS (Stromversorgung)
ETEINT (Alimentation)
STANDBY (Power)
STANDBY (Stromversorgung)
PAUSE (Alimentation)
Operating Instructions
Bedienungsanleitung
Mode d'emploi
Fuse
Sicherung
Fusible
Pushing Prohibited
Drücken ist untersagt
Interdiction de pousser
Certified by CSA Group
Zertifiziert von der CSA-Gruppe
Certifié par CSA Group
Connect to supply mains with protective earth (ground)
Zum Versorgungsnetz durch eine Schutzerdung anschließen
Connectez au secteur, avec protection mise à la terre
36 5
CAUTION AND MARKING SYMBOLS Specify your areas of concern (i.e. eyes, forehead, etc.) _________________
_______________________________________________________________________
Symbol Description _______________________________________________________________________
Calls attention to a procedure, practice, or condition that could SECTION 2: CLIENT CONSENT FORM
possibly cause bodily injury or death.
(Initial each acknowledgement line below)
Ruft Aufmerksamkeit zu einem Verfahren, Praxis, oder Bedingung,
die vielleicht körperliche Verletzung oder Tod verursachen könnte.
Une attention particulière doit être portée à une procédure, pratique 1. I acknowledge that my skin might experience temporary irritation,
ou condition qui pourrait causer des blessures corporelles ou la tightness, or redness, which usually dissipates within 72 hours de-
mort. pending on skin sensitivity. _____(initial here)
Calls attention to a procedure, practice, or condition that could 2. I acknowledge that if I fail to use a minimal sunscreen (SPF 30) and
possibly cause damage to equipment or permanent loss of data. follow the direction for use, I am more susceptible to sunburn, sun
Ruft Aufmerksamkeit zu einem Verfahren, Praxis, oder Bedingung, damage & hyperpigmentation. I should avoid excessive sun exposure,
die vielleicht Schaden an Ausrüstungen oder permanentem Verlust especially between 10am - 2pm. _____(initial here)
der Daten verursachen könnte. 3. I have disclosed my history of allergies above and I acknowledge that
Une attention particulière doit être portée à une procédure, pratique if I am allergic to one or more of the ingredients in the products used,
ou condition qui pourrait causer des dommages à l’appareil ou la I may experience an allergic reaction. _____(initial here)
perte permanente de données.
4. I hereby agree to have the treatment performed and agree to follow
Do Not Re-use all pre and post treatment instructions. _____(initial here)
Nicht wieder verwenden 5. I acknowledge that I should avoid use of aggressive exfoliation,
Ne pas réutiliser waxing, and products containing acids that are not part of the rec-
Manufactured by ommended take-home regimen in the treated areas for minimum 2
Hergestellt von weeks pre and post treatment. _____(initial here)
Fabriqué par 6. I acknowledge that I should avoid use of Retin-A type products for a
period of time recommended by my physician and/or skincare practi-
Keep dry
Bleib trocken tioner pre and post the treatment. _____(initial here)
Garder au sec 7. I acknowledge that I have answered all questions truthfully and
completely. _____(initial here)
8. I release Edge Systems, the ______________ (Aesthetician/Doctor),
management and staff of_______________ (Clinic/Office) from any and
all liability associated with any injuries and/or current orfuture condi-
tions resulting from the skincare procedures or products. _____(initial
here)
9. I consent to the use of my before, during and after facial procedure
photographs for education, promotion or advertising purposes. My
name will not be used to identify these photographs without my writ-
ten approval. _____(initial here)
By signing below, I certify that I have read and fully understood the con-
tents of this consent form, and that the information I provided above are
complete, accurate, and up-to-date to my knowledge.
6 35
11. The client should have a minimum sunscreen SPF 30 applied after What is Not Covered:
the procedure and should use sunscreen on an ongoing basis. Use of THIS WARRANTY DOES NOT COVER ANY EQUIPMENT WHICH HAS BEEN
Edge Systems’ UV Smart® Daily sunscreen is highly recommended. DAMAGED BY ACCIDENT, MISUSE, ABUSE, MODIFICATION, WHICH HAS
BEEN USED IN VIOLATION OF THE SYSTEM INSTRUCTIONS OR FOR ANY
12. Empty the waste container after each service according to your PURPOSE OTHER THAN ONE FOR WHICH IT WAS MANUFACTURED, OR
waste handling protocol. Follow the system cleaning instructions in CAUSED BY UNAUTHORIZED REPAIR OR USE OF UNAUTHORIZED PARTS,
this guide to clean your machine and handpiece after each INCLUDING, WITHOUT LIMITATION, DAMAGE TO THE SYSTEM ARISING
procedure. FROM USE OF SKIN SOLUTIONS OR TIPS PURCHASED OR OTHERWISE
OBTAINED FROM A NONEDGE SYSTEMS APPROVED SUPPLIER. THIS WAR-
13. Only Edge Systems skin solutions should be suctioned through the RANTY DOES NOT COVER FREIGHT DAMAGE; OR ANY DAMAGE CAUSED
HydraFacial® handpiece. Use of other, non-approved skin solutions BY ACTS OF GOD OR THIRD PARTIES NOT WITHIN THE CONTROL OF
will void the warranty and clog the machine. Always use solution EDGE SYSTEMS, SUCH AS DAMAGE CAUSED BY POWER SURGES OR LIGHT-
when performing abrasion. NING. THIS WARRANTY DOES NOT COVER THE COSTS OF INSTALLATION,
ADJUSTMENT OF USER CONTROLS, INITIAL TECHNICAL ADJUSTMENT
14. Keep the device in accordance to recommended environmental (SET-UP) AND ROUTINE USER REQUIRED MAINTENANCE. THIS WARRANTY
conditions indicated on page 5 of this manual. Liquid should only be DOES NOT EXTEND TO ANY MACHINERY, APPLIANCES OR PROPERTY
suctioned through the HydraFacial® handpiece. OF THE CUSTOMER USED IN CONJUNCTION WITH OR CONNECTED TO
THE EQUIPMENT. THIS WARRANTY IS THE EXCLUSIVE REMEDY AGAINST
15. Not recommended for severe acne scarring. EDGE SYSTEMS AND NO OTHER REMEDY (INCLUDING BUT NOT LIMITED
TO INCIDENTAL OR CONSEQUENTIAL DAMAGES FOR INJURY TO PERSON
OR PROPERTY, LOST PROFITS, LOST SALES OR ANY OTHER INCIDENTAL
Equipment not suitable for use in the presence of a flammable anesthet- OR CONSEQUENTIAL LOSS) SHALL BE AVAILABLE. THIS WARRANTY IS IN
ics mixture with air or with oxygen or Nitrous Oxide. LIEU OF ANY OTHER WARRANTY, EXPRESS OR IMPLIED, INCLUDING BUT
NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
Ausrüstungen die nicht sind passend für Gebrauch in der Anwesenheit FITNESS FOR A PARTICULAR PURPOSE. ANY REPRESENTATIONS OR PROM-
von einer brennbaren Narkosenmischung mit Luft oder mit Sauerstoff ISES INCONSISTENT WITH OR IN ADDITION TO THIS LIMITED WARRANTY
oder Stickstoffoxyd. ARE UNAUTHORIZED AND SHALL NOT BE BINDING UPON EDGE SYSTEMS.
PLEASE NOTE THAT SOME STATES DO NOT ALLOW LIMITATIONS ON HOW
L’appareil n’est pas destiné à être utilisé en présence d’un mélange anes- LONG AN IMPLIED WARRANTY LASTS OR DO NOT ALLOW THE EXCLUSION
thésique inflammable avec de l’air, de l’oxygène ou de l’oxyde nitreux. OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE
ABOVE LIMITATIONS MAY NOT APPLY TO YOU.
Unauthorized user modifications will void warranty. Edge Systems LLC Edge Systems LLC
will not be responsible for any injuries sustained due to unauthorized 2277 Redondo Avenue
equipment modifications or use of parts not specified by Edge Systems Signal Hill, CA 90755
LLC. Equipment returned to Edge Systems LLC with unsafe modifications Ph: 1-562-597-0102 or 800-603-4996
will be returned to their original operating condition at the customer’s Fax: 1-562-597-0148
expense.
8 33
32 9
Cet appareil a été conçu conformément aux limites de l’organisme de Make sure the light handpiece is
Improper
réglementation en matière d’interférence électromagnétique. Toutefois, connected to the proper color-coded
connection
rien ne garantit l’absence d’interférences dans une installation particu- receptacle.
lière. Si cet appareil provoque des interférences, détectables en allumant LED Light Make sure that you pressed the
et en éteignant l’appareil, celles-ci pourront possiblement être corrigées Handpiece button for red light or blue light
en accomplissant une ou plusieurs des actions suivantes : Not Working Unit is not on the touchscreen to activate the
-Déplacer l’appareil par rapport au récepteur; activated light handpiece. Also make sure you
-Éloigner l’appareil du récepteur; press the button on the back of the
-Brancher l’appareil sur une autre prise de courant pour que l’appareil handpiece.
et le récepteur soient branchés sur différents circuits.
LED Light
Refer to LED Light Therapy Handpiece User Guide part
Handpiece
number 19692 for troubleshooting solutions.
Not Working
To prevent electric shock, do not remove the instrument cover. There
are no user serviceable parts inside. Routine maintenance or cleaning Check that the cable end connector
of internal parts is not necessary. Avoid the use of cleaning agents or Camera Not Unit is not on
is securely inserted into the port.
chemicals on the instrument. Some chemicals may damage plastic parts Working
or lettering. Any external cleaning should be done with a clean, dry or Camera Disconnect the camera, and then
slightly damp cloth. Any replacement cables and parts should be ac- Not Taking Connection issue reconnect the camera to the unit.
quired directly from Edge Systems LLC. Photos Reboot the system.
Elektroschock zu verhindern, nimmt die Instrumentdecke nicht heraus. Unable to Back up photos on a regular basis
Es gibt keine vom Benutzer haltbaren Teile hinein. Routinewartung oder Create New with the 32GB USB flash drive pro-
Out of storage
reinigend von inneren Teilen ist nicht notwendig. Vermeiden Sie den Ge- Member Pro- vided with the Camera kit option.
memory
brauch, um Agenten oder Chemikalien auf dem Instrument zu reinigen. file or Store Go to Settings and select “Photo
Einige Chemikalien dürfen Kunststoffteile oder Beschriftung beschädi- Photos Backup/Restore.”
gen. Irgendein äußerlicher Reinigen sollte mit einem sauberen gemacht Currently
werden, trocknet oder dämpft etwas Tuch. Irgendeine Ersetzung kabelt Stored Restoring photos
Ensure current photos in the system
und Teile sollten direkt von Edge Systems LLC erlangt werden. Photos Are overrides current
are backed up before restoring previ-
Erased After photos stored in
ous photos to the system.
Restoring the system.
Photos
Note: Circuit diagrams & component part lists are available upon
request for service technicians.
10 31
Problem Possible Cause Solution Pour prévenir les chocs électriques, n’enlevez pas le couvercle de
l’instrument. Ce dernier ne contient aucune pièce pouvant être réparée
Check lines for any kinks or leaks. par l’utilisateur. L’entretien ou le nettoyage périodique des pièces in-
Leak in the line
Repuncture bottle if needed. ternes n’est pas nécessaire. Évitez l’utilisation de produits nettoyants ou
Too many solu- Make sure you only have a maximum chimiques sur l’instrument. Certains produits chimiques peuvent endom-
tion stations are of two skin solution stations on the mager les pièces de plastique ou le lettrage. Tout nettoyage externe doit
on “ON” position. être fait avec un chiffon propre, sec ou légèrement humide. Tout câble et
toute pièce de rechange doivent être achetés directement auprès d’Edge
If using only one station at a time,
Systems LLC.
turn current station off and try
another station, if this corrects the
problem please check on the previ-
Faulty station
ous station that internal bottle seal
is completely broken and pick up
assembly (metal portion) is fully
submerged.
Make sure that you are holding the
Operator handpiece correctly and not work-
technique ing against gravity or blocking the
Low or No tubing.
Fluid Flow
Do not let the waste container
fill above 450ml. If the container
becomes overfull, a built-in float will
Waste container
occlude the vacuum opening and
is full
skin solution flow through the hand-
piece will cease. Empty the waste
container after each service.
Flush all stations thoroughly. Make
Clogging sure the machine is cleaned after
each service.
Make sure that the gray
cartridge is properly
Gray cartridge inserted into the hand-
is loose piece. The arrow on the
cartridge should line up
with the lock symbol.
Make sure the flow control valve on
the handpiece is not open too far.
The default setting for
Flow control
Too Much the flow control valve
valve is opened
Fluid Flow is 50% open. To de-
too far
crease fluid flow, turn
the flow control valve
towards the - sign.
30 11
12 29
Camera Handpiece
Camera Holder
28 13
Side Panel:
Port for Mechanical
Charging USB
Lymphatic Drainage
Ports For Cordless
Handpieces
LED Handpieces
Toggle Switches
for Solutions
Manifold Buttons for
Solution Bottles
Removable
Acrylic Door
5. For maximum sterility, abrasive pads, tip, and pad holder are
autoclavable following manufacturer's instructions.
14 27
Other INSTALLATION
WASTE CONTAINER TOOL HOLDER
1. Wash, dry, and wipe/spray germicide over waste canister and lid. 1. Loosen thumbscrews first.
Leave open overnight to air-dry. 2. Attach the tool holder, top first onto the accessory rail.
Note: Always remove the waste container before transporting the 3. Tighten the thumbscrews to lock the handpiece holder in place.
unit.
TOUCHSCREEN
26 15
5. Insert the gray cartridge into the handpiece. Align the arrow on the 5. When the brush and air cleaning are completed, make sure all the
cartridge with the unlock symbol on the handpiece. Push firmly until blue toggle switches are in the OFF position. Turn the gray toggle
it snaps into place, then twist the cartridge up until the arrow aligns switch “SYSTEM RINSE” up to the ON position and the vacuum set-
with the lock symbol. ting on “26”. Press on the cleaning cap and flush through the hand-
piece and tubing for 1 minute.
HYDROPEEL® TIP
SPECIALTY VIAL:
1. To access specialty vials, remove the gray cartridge from the hand-
piece by twisting until the arrow aligns with the unlock symbol, then
pull the cartridge straight out.
2. Insert the vial into the handpiece. Push firmly until it snaps into place.
Note: Keep all the toggle switches off (see below) while using spe-
cialty vials.
16 25
HydraFacial® Handpiece
The HydraFacial® handpiece tubing chambers must be cleaned and dis-
infected using the cleaning brush every week to prevent any bacterial
contamination.
1. Open the front door. Remove the cap of the skin solution bottle and
1. Ensure all the blue toggles switches are in the use the puncture tool to pierce the seal inside the bottle.
OFF position. Turn gray toggle switch "SYS- 2. Press and hold each manifold button as you insert each bottle into
TEM RINSE" up to the ON position and vacuum the manifold, and then release the button.
setting on "26". Attach cleaning cap and flush Note: The Rinseaway™ System Cleaning Solution should be installed
through the handpiece and tubing for 1 minute in the manifold with the gray toggle switch labeled “SYSTEM RINSE”.
2. Detach the handpiece from all tubing. Gently
twist the knobs counter-clockwise to detach
each tube from the handpiece.
3. Insert the cleaning brush into the vacuum
(smaller opening) on the face of the handpiece. Push it in gently all
the way, while rotating in either direction. Repeat with the opening
on the back with the blue dot. A few drops of disinfectant may be
dropped onto the “face” of the hand piece (held in an upright posi-
tion) allowing the disinfectant to run into the handpiece opening
before inserting the cleaning brush.
Note: 70% isopropyl alcohol is not suitable as a disinfectant.
3. The toggle switches control which skin solution bottle is being used.
When a switch is in the ON position, fluid will be drawn from that
bottle. When switched to the OFF position, no fluid will be drawn
from the corresponding bottle.
Vacuum Note: Always keep the “SYSTEM RINSE” toggle switch (gray) off while
Opening performing procedures.
"Face" of the handpiece
24 17
1. Place the diamond pad inside the tip and push it in. Skin Solution Stations And Tubing
2. Place the tip on the HydraFacial® handpiece and press-in until it is
fully seated. AFTER EACH SERVICE
WEEKLY CLEANING
1. Refer to LED Light Therapy Handpiece User Guide part number 19692
for details on installation and operation.
2. USB ports for cordless LED Handpieces are located within each of the
LED Light holders. Plug the LED handpieces into the appropriate hold-
ACTIVATE SKIN IMAGING CAMERA
ers in order to recharge.
1. To operate the camera, go to the Home screen
SKIN IMAGING CAMERA
and select “Camera”.
2. A prompt will appear requesting you to create a
1. Attach the tool holder, top first, onto the accessory rail. Tighten
new profile or select a profile already in the system.
the thumbscrews to lock the tool holder in place.
3. In the member screen, either create a new profile
2. Insert the cable end connector from the camera into the port located
or search for a member by name, date of birth or telephone number.
on the right side of the unit. Push in and twist until it is completely
4. After creating a new profile or selecting the
locked.
member’s profile from the system, the “Photo Capture” button should
3. Place the camera on the tool holder when not in use.
turn from gray to blue.
5. To capture a photo, place the chin rest against the
chin. There are three options for capturing a photo:
• In the member profile screen, select the “Photo Capture” button or
• On the camera, press the red button or
• Use the foot switch to capture a photo. To install the foot switch,
plug the USB cable into one of the USB ports located on the right side
of the unit.
22 19
TURN ON THE SYSTEM 1. Lightly tap the buttons displayed on the screen
with your fingertip (not fingernail). This will
1. Verify the switch on the power entry module cause the unit to perform a variety of functions.
located on the back of the unit is in the OFF ( o ) If the button does not seem to work, verify that
position. Connect the power cord to the power en- you have actually pressed on the button and did
try module and connect the plug into a grounded not miss it.
wall outlet. Switch the power entry module switch 2. There may be a 0.5 second delay when you
to the ON ( I ) position. press the buttons on the touchscreen, so do not
2. The Illuminated Power Switch will light up blue. press the same button repeatedly if it does not
Press the blue Illuminated Power Switch. The respond immediately.
machine should begin its boot up cycle at this 3. Activate the modality you will be using on the
point. When the HydraFacial® page appears on the touchscreen by pressing the button.
screen, select “Here” to go to the main page. Note: The touchscreen has been preset for opti-
mal performance with the HydraFacial® program.
Thus, you cannot and will not need to adjust the
STANDBY / SLEEP MODE monitor.
1. Press the Illuminated Power Switch and a “System ADJUST FLUID FLOW
Cleaning Reminder” will appear on the screen.
2. If you have not cleaned the skin solution stations, 1. Turn the flow control valve towards the +
press NO and the System Cleaning Protocol screen sign to increase fluid flow. Turn towards the -
will appear. sign to decrease fluid flow.
Note: You must clean the skin solution stations
before you can turn the system off or switch to ADJUST VACUUM STRENGTH
the standby mode. Always turn to the standby/
sleep mode at the end of each day. 1. To adjust vacuum settings, simply press the +
3. If you have cleaned the skin solution stations, button to increase the vacuum or – button to
press YES and the system will go into standby decrease the vacuum strength.
mode. 2. On the screen, select 'O' to turn vacuum on or off
20 21