Professional Documents
Culture Documents
net/publication/234936131
CITATIONS READS
6 434
2 authors:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Gerard VAN Dalen on 27 May 2014.
Introduction
Stabilisation of air bubbles in liquid food systems is notoriously difficult. They have to
be stable during production or preparation. Examples of relative stable aerated
products are bread, ice cream and mousses in which the gas bubbles are entrapped
within the solid continuous phase. However liquid or soft solid foams are
thermodynamically unstable resulting in an increase in bubble size (coarsening) in
time and eventually to complete loss of air. These products are subject to
destabilisation processes such as coalescence and Ostwald ripening (diffusion of air
from small to large bubbles). Stabilisation of aerated soft solid oil in water (o/w)
emulsion offers a huge challenge.
A better and comprehensive insight into mechanisms responsible for the formation
and stabilization of foam structures is required. Quantitative methods are needed for
the characterization of aerated products and therefore suitable for the derivation of
substantiated structure-function relationships. The amount of gas and the bubble size
distribution of these products have to be followed in time. This paper shows how X-
ray microtomography (µCT)1,2 can be used for the 3D visualisation and quantitative
analysis of bubbles in aerated emulsions during storage.
Method
Model emulsions were imaged using a Skyscan 1172 desktop µCT system. The used
image acquisition parameters are listed in Table 1. Tomographic reconstructions
were performed on a cluster computer (HP 220c server Blade with quad core Xeon
CPU with 16 cores in total) using Nrecon (v 1.6.3.2). A smoothing factor of 4, beam
hardening correction of 50% and ring artefact reduction of 20 was used. The features
in the stacks of the 2D µCT images were identified and measured using an image
analysis toolbox (DIPlib from the Delft University of Technology). For visualisation in
3-D space, isosurface rendering was used (Avizo 6.2 from the Visualization Sciences
Group). Images of time series were aligned in Avizo using affine registration. Before
registration a coarse manual alignment was performed using marks on the sample
holder wall (small hole visible in Figure 14 ). The quantitative analysis of the bubble
size distribution will be reported in a separate paper3.
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 22
Figure 1 Projection image showing the total scanned area (obtained by stitching
multiple vertical overlapping scans).
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 23
Figure 3 Influence of sample holder on the storage of 2.5 – 3.5 gram water (w) and
an emulsion (o/w) in sample holders with an inner diameter of 11mm.
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 24
Results
Representative cross sections of µCT images of an aerated o/w emulsion with 25%
oil imaged at a pixel size of 2.0, 4.0 and 8.4 μm using sample holders with an inner
diameter of 6, 11 and 26 mm are shown in Figure 1. The air bubbles are clearly
visible within the fat/protein/water matrix by their high grey value (low absorption
coefficient). These images show a broad distribution of bubble sizes including very
small and very large bubbles. A high magnification (pixel size of 2.0 μm) reveals the
presence of clusters of very small bubbles. However at this magnification the field of
view is limited, missing the very large bubbles. Selection of the size of a sample
holder will always be a compromise between obtaining maximum detail (resolution)
and being representative for the total sample. For accurate analysis of the total range
of bubble sizes in the sample shown in Figure 1, images could be made using
sample holders with an inner diameter of 6 and 26 mm. Larger sample holders will
also result in less sampling artefacts and will be less influenced by environmental
conditions during storage, representing more the actual shelf life.
Figure 1: µCT images of an aerated o/w emulsion imaged in sample holders with an
inner diameter of 6, 11 and 26 mm, scanned with a pixel resolution of 2.0,
4.0 and 8.4 μm, respectively. Bottom: enlarged view of the area indicated
by the yellow box.
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 25
more beam hardening (Figure 4). Standard beam hardening correction5 during
tomographic reconstruction (square function in NRecon) reduced the cupping effect
but could not remove cupping completely. A better correction was obtained using a
3th order polynomial function (Figure 5B, C & D).
i The images and graphs were scaled relative to the average grey level in the centre. The
profiles show the vertical average over 50 pixels along a line shown in the images.
ii Transmission through the centre of the sample.
iii The relative standard deviation (RSD) of the mean grey level (matrix – air) over an area of
200*200 pixels.
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 26
main drawback of using a filter is that it degrades the X-ray signal at all energies to
some degrees, leading to greater image noise. Even when using a higher voltage
(59>95kV) and a higher acquisition (exposure) time (589>1767ms), a much higher
noise level is observed (Figure 7A). The longer acquisition time leads to an increase
in total scan time from 1.5 to 2.5 hours (scan with a rotation step of 0.20o/180o and
frame averaging of 4). So this will not be a preferred option, especially because
multiple scans (oversized scan) are needed to image the total sample holder.
100.0 25.0
A 80.0 25.0
B 90.0 Cupping
70.0
80.0 % RSD 20.0
20.0
60.0
cupping, %
70.0
cupping, %
% RSD
% RSD
40.0 50.0
C D
3e-order coefficient:
2.0
0.0 – 0.3 – 1.0
(0=100% BHC)
relative grey level
1.5
11 mm
1.0
0.5
0.0
0 500 1000 1500 2000 2500 3000 3500
profile, pixels
A B
iv
The cupping is calculated as a percentage of the average grey value at the border of the
sample relative to the average grey value at the centre.
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 27
A B C
0% - 50% - 100% beam
hardening correction
11 mm
Beam hardening can also be reduced using post processing of the horizontal cross
sections obtained after tomographic reconstruction. Two options were tested, using a
cross sectional average of the total stack of slices of 1: the sample (same dataset)
and 2: a physical phantom (separate dataset). The phantom should be uniform and
comparable density, size and shape, imaged using identical acquisition parameters.
As phantom an emulsion without air was used. For both methods the extent of the
beam hardening has to be same for each slice. For an aerated emulsion the beam
hardening effect is not significantly influenced by the vertical position.
The result of the first correction method is shown in Figure 8-1. The cross sectional
average was obtained by averaging a selection of about 40 slices, equally divided
over the total stack. These slices were first aligned and clipped at about 90% of the
average grey level of the matrix in the centre of the sample holder to reduce the
contribution of the low grey value of the bubbles. The cross sectional average was
blurred using a Gaussian filter with a sigma of 6 pixels and stretched between 1 and
99% percentile. After angular averaging this image was used to correct the total
stack of slices of the sample: each slice was divided by the correction image and
stretched between 1 and 99%.
The result of the second correction method using a non aerated emulsion as
phantom is shown in Figure 8-2. The procedure is comparable to the first method.
However in this case no clipping was used and the corrected image was aligned with
the image of the sample. The final results of both methods are more or less
comparable. Only method 2 also corrects for some residual ring artefacts (no blurring
used as in method 1).
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 28
Profile, pixels
Figure 8 Correction for beam hardening using cross sectional averages of the total
stack of slices of the sample (top) and a non aerated emulsion (bottom).
The images were stretched between 1-99% percentile (BHC=50%).
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 29
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 30
Figure 10 Relative grey level profiles through an air bubble with a diameter of 17 μm
(FWHM) within a reconstructed image of an o/w emulsion in sample
holders with an inner diameter of 11 mm showing the influence of
smoothing during reconstruction. Grey levels were scaled between matrix
(1) and air outside the sample holder (0). Image pixels size = 4 μm.
The stability of aerated o/w emulsions was tested by imaging for each sample exactly
the same sub sample in time. For this purpose sub samples were imaged using
Perspex sample holders with an inner diameter of 6 and 11mm (Figure 2B). The
sample holders were stored at ambient conditions and imaged during 16 weeks.
Important microstructural parameters of aerated products are the air content and the
bubble size distribution. For this study the volume weighted mean diameter D[4, 3],
and the surface area weighted mean diameter D[3, 2] were measured.
Examples are shown in Figure 12 and Figure 13. The gas bubbles of this sample
were not stable during storage. The number of air bubbles per volume was reduced
and larger air bubbles were formed in time. For large bubbles coalescence was
observed (bubbles combine into larger ones). For small bubbles Ostwald ripening
was observed (Figure 14 and Figure 15). The air phase is transferred from small to
large bubbles due to their different Laplace pressures. It is controlled by the solubility
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 31
of air in the emulsion. At locations were the bubbles disappeared a higher absorption
was observed (Figure 16). This can be due to the migration of water (higher
absorption than oil/water).
Beside a decrease of the total volume of gas bubbles in time, also a decrease of the
emulsion volume was observed (Figure 13). This was caused by evaporation (see
discussion about sample holders in the method section).
Figure 12 μCT images of horizontal (middle) and vertical (bottom) cross sections of
an aerated o/w emulsion during 16 weeks storage in a sample holder with
an inner diameter of 11mm. The top row shows the 3D visualisation of the
bubbles (box size = 13.8mmx13.8mmx20.4mm, pixel size = 8.0 μm).
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 32
6 mm
1 day 3 days 1 week 1 month
Figure 14 μCT images of horizontal cross sections of an aerated o/w emulsion
during 1 month storage in a sample holder with an inner diameter of 6mm.
Enlarged views (2.4mm * 4.8mm) at different storage times with outlines
of bubbles detected at T0 in red.
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven
P a g e | 33
xz
xy
1 day
xz
3 days xy
Figure 16 μCT images of horizontal and vertical cross sections (2.0mm * 2.0mm) of
an aerated o/w emulsion at 1 and 3 days after preparation.
Conclusion
X-ray microtomography allows in-situ observation and analysis of bubbles during
storage of aerated emulsions. Various coarsening mechanism can be identified and
investigated, namely coalescence and Ostwald ripening.
References:
G. van Dalen e.a., μCT imaging of aerated emulsions, SkyScan user meeting, April 13-15, 2011, Leuven