Professional Documents
Culture Documents
Nellcor
TM
OxiMax N-65
Handheld Pulse Oximeter
OxiMax TM
Technology
COVIDIEN, COVIDIEN with logo, the Covidien logo and positive results for life are U.S. and
internationally registered trademarks of Covidien AG. Other brands are trademarks of a
Covidien company. ©2011 Covidien.
U.S. Patents: 5,485,847; 5,676,141; 5,743,263; 6,035,223;6,226,539; 6,411,833;
6,463,310; 6,591,123; 6,708,049; 7,016,715; 7,039,538; 7,120,479;7,120,480; 7,142,142;
7,162,288; 7,190,985;7,194,293; 7,209,774; 7,212,847; 7,400,919.
To obtain information about a warranty, if any, contact Covidien Technical Services at
1.800.635.5267 or your local representative.
Purchase of this instrument confers no express or implied license under any Covidien patent
to use the instrument with any handheld pulse oximeter that is not manufactured or licensed
by Covidien.
Table of Contents
Chapter 1. Safety Information ..................................................................... 7
Overview.................................................................................................. 7
Warnings ................................................................................................. 7
Cautions................................................................................................... 9
iv Service Manual
Safety Testing Standards ...................................................................... 64
Performance Tests................................................................................. 64
Power-On-Self-Test ......................................................................... 64
Operational Setup................................................................................. 67
Overall Performance Check .................................................................. 67
Operation with a Live Subject........................................................ 68
Pulse Oximetry Functional Tests........................................................... 69
Overview.......................................................................................... 69
Test Setup ........................................................................................ 70
BPM Test .......................................................................................... 72
SpO2 Test ......................................................................................... 74
Modulation Level Test .................................................................... 75
Light Level Test ............................................................................... 81
Service Manual v
Page Left Intentionally Blank
vi Service Manual
1 Safety Information
Overview
This section contains safety information for service personnel who will
service the Nellcor™ OxiMax™ N-65™ handheld pulse oximeter. Only
qualified service personnel should service this product. Read the
Operator's Manual carefully and thoroughly understand the operation
of the oximeter.
Warnings
WARNING
The OxiMax pulse oximetry sensor extrapolates from the date and time
provided by the Nellcor OxiMax N-65 handheld pulse oximeter when
recording the sensor event record to the sensor. The accuracy of the
7
Safety Information
WARNING
Explosion hazard—Do not use the Nellcor OxiMax N-65 handheld pulse
oximeter in the presence of flammable anesthetics.
WARNING
Do not spray, pour, or spill any liquid on the pulse oximeter, its
accessories, connectors, switches, or openings in the enclosure as this
may damage to the oximeter.
WARNING
The LCD panel contains toxic chemicals. Do not touch broken LCD
panels. Physical contact with a broken LCD panel can result in
transmission or ingestion of toxic substances.
WARNING
Do not silence or decrease the volume of the OxiMax N-65 handheld
pulse oximeter’s audible alarm if patient safety could be compromised.
WARNING
The use of accessories, sensors, and cables other than those specified
may result in increased emission and/or create invalid readings of the
Nellcor OxiMax N-65 handheld pulse oximeter.
WARNING
To ensure accurate performance and prevent device failure, do not
subject the Nellcor OxiMax N-65 handheld pulse oximeter to extreme
moisture, such as direct exposure to rain. Such exposure may cause
inaccurate performance or device failure.
WARNING
Do not lift the Nellcor OxiMax N-65 handheld pulse oximeter by the
sensor or extension cable. The cable may disconnect, potentially drop
ppi ng the pulse oximeter on a patient or damaging surface.
8 Service Manual
Cautions
WARNING
As with all medical equipment, carefully route patient cabling to reduce
the possibility of patient entanglement or strangulation.
Cautions
Caution
Dispose of batteries in accordance with local requirements and regulations.
Caution
Observe ESD (electrostatic discharge) precautions when working within the
unit.
Caution
Observe ESD (electrostatic discharge) precautions when disassembling and
reassembling the Nellcor OxiMax N-65 handheld pulse oximeter and when
handling any of the components of the monitor.
Caution
When reassembling the Nellcor OxiMax N-65 handheld pulse oximeter,
tighten the screws that hold the cases together to a maximum of ten
inch-pounds. Over-tightening could strip out the screw holes in the front
case, rendering it unusable.
Service Manual 9
Page Left Intentionally Blank
10 Service Manual
2 Introduction
Overview
The Service Manual contains information for servicing the Nellcor™
OxiMax™ N-65™ handheld pulse oximeter. Only qualified service
personnel should service this product. Before servicing the oximeter,
thoroughly read the Operator's Manual for safe operation. The latest
version of this manual is available on the Internet at:
http://www.nellcor.com/serv/manuals.aspx
Related Documents
To perform test and troubleshooting procedures and to understand
the principles of operation and circuit analysis sections of this manual,
you must know how to operate the Nellcor OxiMax N-65 handheld
pulse oximeter. Refer to the Operator's Manual. To understand the
various OxiMax pulse oximetry sensors that work with the monitor,
refer to the individual OxiMax pulse oximetry sensor's Instructions For
Use.
The latest version of the Operator’s Manual and the Service Manual are
posted on the Internet at:
http://www.nellcor.com/serv/manuals.aspx
11
Introduction
Intended Use
WARNING
The Nellcor OxiMax N-65 handheld pulse oximeter is a prescription
device to be operated only by trained personnel. The oximeter is for
attended monitoring only.
The Nellcor OxiMax N-65 handheld pulse oximeter is indicated for
continuous or spot check monitoring of functional arterial oxygen
saturation (SpO2) and pulse rate of adult, pediatric and neonatal
patients in hospital, hospital-type facilities, transport and mobile
environments as well as in the homecare environment.
Note:
Intended typical usage may be defined to include the following for the
Nellcor OxiMax N-65 handheld pulse oximeter:
• Hospital use typically covers such areas as general care floors (GCFs), operating rooms,
special procedure areas, intensive and critical care areas within the hospital and in
hospital-type facilities. Hospital-type facilities include physician office-based facilities,
sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
• Intra-hospital transport includes transport of a patient within the hospital or hospital-type
facility.
• Homecare use involves a lay person (parent or other similar non-critical caregiver) in the
home environment.
Oximeter Components
Front Panel
This section identifies the symbols, controls, displays, and indicators
on the Nellcor OxiMax N-65 handheld pulse oximeter.
Refer to the Operator’s Manual for a complete description of all keys,
symbols, controls, displays and indicators.
The front panel provides the user with multiple methods not only of
obtaining patient data, but also oximeter and sensor status.
12 Service Manual
Oximeter Components
Note:
Verify oximeter functionality during monitoring of patient data via the
regular updates to patient values and blip bar movement.
Figure 1. Front Panel Keys
Service Manual 13
Introduction
1 %SpO2Value
2 %SpO2 Upper and Lower Limit indicator
3 Pulse Amplitude indicator (Blip Bar)
4 Time Colon time/date field separator
5 Adjust Time Mode indicator
6 Adjust Volume Mode indicator
7 Set Limit Mode indicator
8 Pulse Rate (BPM) Upper and Lower Limit indicator
9 Pulse Rate Value (BPM)
10 Print indicator
11 Sensor Off Patient indicator
12 Interference indicator
13 Sensor Disconnected indicator
14 Pulse Beep Tone Volume Adjust indicator
15 Pulse Search indicator
16 Alarm Volume Adjust indicator
17 Data indicator
18 Data In-Sensor indicator
19 Alarm Silence indicator
20 Low Battery indicator
User Interface
This section describes the control keys on the front panel of the
oximeter. Activate by pressing the appropriate key. Pressing any key
besides the POWER key when the oximeter is turned on results in
14 Service Manual
Oximeter Components
either a valid or invalid key tone. If a tone fails to emit when you press
a key, contact qualified service personnel.
Figure 3. OxiMax N-65 Handheld Pulse Oximeter User Interface
ALARM SILENCE Key—Use to silence all audible alarms for the alarm
silence duration period. After silencing an alarm, press the key again
to reactivate the alarm. Alarms that occur throughout the Power-On
Self-Test (POST) cannot be silenced.
Press and hold the ALARM SILENCE key for three seconds to display the
Alarm Silence Duration menu.
ADJUST DOWN Key—Use to decrease variable parameters of the
oximeter or the pulse beep volume. Press and hold the key for more
than three seconds to repeat the decrement.
DATA (Record/Print) Key—Use to store currently shown SpO2 and
pulse rate values, known as snap-shot data. Use this same key to print
the stored snap-shot data, such as a single-event report, snap-shot,
and/or sensor-event data by pressing the DATA key in print mode.
MENU Key—Press repeatedly during normal operation to display
oximeter parameters.
• Upper and lower SpO2 alarm limits
• Upper and lower Pulse Rate (BPM) alarm limits
• Alarm volume
• Pulse beep volume
• Data print mode
Caution
The oximeter returns to normal operation and incorporates the selected
parameter value after approximately 30 seconds of inactivity. See Menu
Structure, page 57 for additional information about setting parameters.
Service Manual 15
Introduction
Alarms and alarm indicators are enabled when parameters are set and
display in the event of an alarm condition.
• Sensor Disconnected
• Pulse Search
• Interference
• Low Battery
• Alarm Silence
Press the MENU key during Power-On Self-Test (POST) until the Time/
Date set displays. Repeatedly pressing the MENU key allows users to
set time/date parameters.
• Hour
• Minute
• Day
• Month
• Year
After setting time/date parameters, users may return to the default
display.
BACKLIGHT Key—Press to toggle the backlight ON or OFF. After
activation, the backlight remains on for approximately ten minutes.
ADJUST UP Key—While in menu mode, press repeatedly to increase a
parameter by one increment. Press and hold the key for more than
three seconds to repeat the increment.
Press the ADJUST UP key during normal operation while monitoring to
increase the pulse beep volume.
Note:
Oximeter functions listed below are prerequisites for normal operation.
• Powered on
• Sensor connected to the oximeter
• Sensor applied to a patient
• Oximeter reporting patient’s %SpO2 (oxygen saturation percentage) and pulse rate
readings (BPM)
• No error conditions exist
16 Service Manual
Oximeter Components
Service Manual 17
Introduction
18 Service Manual
Oximeter Components
WARNING
Pressing ALARM SILENCE will keep ALL alarms from sounding for the
alarm silence period.
Data Indicator—Displays during the store snap-shot data or data
print mode and flashes when snap-shot data prints.
Pulse Search Indicator—Lights continuously prior to initial
acquisition of a pulse signal and during prolonged challenging
monitoring conditions. The pulse search indicator flashes throughout
a loss-of-pulse signal.
Interference Indicator—Lights whenever the oximeter algorithm
detects degraded incoming signal quality. An intermittently lit
Interference Indicator is common during patient monitoring, and
indicates the oximeter algorithm dynamically adjusts the amount of
data required for measuring SpO2 and Pulse Rate. When the indicator
is lit continuously, the oximeter algorithm is extending the amount of
data required for measuring SpO2 and Pulse Rate. In this case, fidelity
in tracking rapid changes in these values may be reduced.
Note:
Degradation can be caused by ambient light, poor sensor placement,
electrical noise, electro-surgical interference, patient activity, or other
causes.
Print Indicator—Displays during the print mode and flashes when
the oximeter prints single-event, snap-shot, or patient alarm events
data stored in the sensor.
Service Manual 19
Introduction
Rear Panel
Figure 4. OxiMax N-65 Handheld Pulse Oximeter Rear Panel
20 Service Manual
Carton Label Symbols
Rx prescription only
Fragile
Humidity limitation
Temperature limitation
Service Manual 21
Page Left Intentionally Blank
22 Service Manual
3 Product Specifications
Overview
This section contains Nellcor™ OxiMax™ N-65™ handheld pulse
oximeter physical and operational specifications. Ensure all product
requirements are met prior to installation of the oximeter.
Physical Characteristics
Weight 0.62 lbs. (0.28 kg)
Dimensions 2.875 in. x 6.25 in. x 1.375 in.
(7.3 cm x 15.9 cm x 3.5 cm)
Electrical Requirements
Batteries
Type 4 AA alkaline
Voltage 6 Volts DC (with 4 AA batteries)
Typical Battery life Alkaline 15 hours
Lithium 40 hours
Environmental Conditions
Operation
Temperature 5 ºC to 40 ºC (41 ºF to 104 ºF)
Altitude -390 m to 3,012 m (-1,254 ft. to 9,882 ft.)
Atmospheric Pressure 70 kPa to 106 kPa (20.6 in. Hg to 31.3 in. Hg)
Relative Humidity 15% to 95% non-condensing
23
Product Specifications
Performance Specifications
Measurement Range
SpO2 1% to 100%
Pulse Rate 20 to 250 beats per minute (bpm)
Perfusion Range 0.03% to 20%
24 Service Manual
Performance Specifications
Pulse Rate
1
Subjects used to validate SpO2 measurement accuracies were healthy and recruited from the local population.
Comprised of both men and women, subjects spanned a range of skin pigmentations and ranged in age from 18-50
years old. Accuracy specifications are based on controlled hypoxia studies with healthy non-smoking adult volunteers
over the specified saturation SpO2 range(s). Pulse oximeter SpO2 readings were compared to SaO2 values of drawn
blood samples measured by hemoximetry. All accuracies are expressed as ±1 SD. Pulse oximeter equipment
measurements are statistically distributed; about two-thirds of pulse oximeter measurements can be expected to fall
in this accuracy (ARMS) range. Because scatter and bias of pulse oximeter SpO2 and blood SaO2 comparison
commonly increase as the saturation decreases, and accuracy specifications are calculated from data spanning the
stated range, different accuracy values may result when describing partially overlapping ranges.
2
Clinical functionality of the MAX-N has been demonstrated on a population of hospitalized neonate patients. The
observed SpO2 accuracy was 2.5% in a study of 42 patients with ages of 1 to 23 days, weight from 750 to 4,100 grams,
and 63 observations made spanning a range of 85% to 99% SaO2.
3
Clinical functionality has been demonstrated in a population of hospitalized neonate patients. The observed SpO2
accuracy was 3.0% in a study of 57 patients with ages of 24 to 40 weeks, weight from 710 to 5,000 grams, and 185
observations made spanning a range of 63% to 99% SaO2.
4
Specification applies to N-65 oximeter performance. Reading accuracy in the presence of low perfusion (detected IR
pulse modulation amplitude 0.03% - 1.5%) was validated using signals supplied by a patient simulator. SpO2 and
pulse rate values were varied across the monitoring range over a range of weak signal conditions and compared to the
known true saturation and pulse rate of the input signals.
Service Manual 25
Product Specifications
26 Service Manual
Audible Indicators
Audible Indicators
Table 3. Audible Indicator Parameters and Values
Priority 1
Priority 2
Priority 3
Service Manual 27
Product Specifications
Priority 4
Priority 5
Priority 6
28 Service Manual
Audible Indicators
Priority 7
Priority 9
Service Manual 29
Product Specifications
Priority 10
Priority 11
Product Compliance
Product Standards for Compliance
ISO 9919: 2005
EN ISO 9919: 2005
RoHS EU Directive 2011/65/EU
30 Service Manual
Manufacturer’s Declaration
Manufacturer’s Declaration
Basics
WARNING
The use of accessories, sensors, and cables other than those specified
may result in inaccurate readings of the Nellcor OxiMax N-65 handheld
pulse oximeter and increased emission or decreased immunity of the
oximeter.
The Nellcor OxiMax N-65 handheld pulse oximeter is suitable for
prescription use only in the specified electromagnetic environments.
Use the unit in accordance with the electromagnetic environments
described in this section.
Note:
The OXIMAX N-65 is RoHS compliant for products manufactured as of July 2014.
Service Manual 31
Product Specifications
Voltage fluctuations/
flicker emission
IEC 61000-3-3: 2005
Electromagnetic Immunity
Note:
Use the unit in accordance with the electromagnetic environments
described in this section.
Table 5. Electromagnetic Immunity Testing
32 Service Manual
Manufacturer’s Declaration
d = 0.2 P
20 V/m (transport)
3 V/ma
800 MHz 3 V/m (non-transport)a
2.5 GHz d = 2.3 P
d = 0.4 P
20 V/m (transport)
a
Per ISO 9919:2005, compliance level is 20 V/m when used with the transport cover and the included ferrite bead on
the sensor cable.
Service Manual 33
Product Specifications
These guidelines may not apply in all situations. Absorption and reflection
from structures, objects, and people affect electromagnetic propagation.
34 Service Manual
Safety Tests
Sensors
Sensor Cables
Note:
The sensors and sensor cables are RoHS compliant for products
manufactured as of July 2014.
Safety Tests
Leakage Current
The following table displays the maximum patient leakage current
allowed.
Table 9. Patient Applied and Patient Isolation Risk Current
IEC 60601-1
Condition UL 60601-1
Normal 100 μA
Service Manual 35
Page Left Intentionally Blank
36 Service Manual
4 Theory of Operations
Overview
This section explains the theory behind Nellcor™ OxiMax™ N-65™
handheld pulse oximeter operations.
37
Theory of Operations
Automatic Calibration
Because light absorption by hemoglobin is wavelength dependent
and because the mean wavelength of LEDs varies, an oximeter must
identify the mean wavelength of the OxiMax sensor's red LED to
accurately measure SpO2.
During monitoring, the oximeter’s software selects coefficients
appropriate for the wavelength of that individual OxiMax sensor's red
LED; these coefficients are then used to determine SpO2.
Additionally, to compensate for differences in tissue thickness, the
light intensity of the OxiMax sensor's LEDs adjusts automatically.
38 Service Manual
OxiMax™ Pulse Oximetry Sensor Technology
Service Manual 39
Theory of Operations
40 Service Manual
5 Product Overview
Overview
The Nellcor™ OxiMax™ N-65™ handheld pulse oximeter relies on
unique oximetry technology and design in providing hospitals,
clinicians, and caregivers accurate, timely data.
41
Product Overview
42 Service Manual
System Board Components
Service Manual 43
Product Overview
Keypad
The user interface PCB keypad contains seven keys. These keys allow
the user to setup and control of the oximeter. The keys are:
• Alarm Silence
• Adjust Down
• Data
• Menu
• Backlight
• Adjust Up
• Power
The signals from the keys pass directly through the SpO2 PCB, then
applied to the user interface PCB CPU. The POWER key signal is applied
to the SpO2 power supply, turning the oximeter on or off.
LCD
The liquid crystal display (LCD) is driven by the CPU. The LCD displays
the patient’s %SpO2 and pulse rate, as well as status and function
indicators. See User Interface, page 14.
Audio
The audio circuit drives a transducer to provide the oximeter tones.
Real-Time Clock
The real-time clock and non-volatile random access memory (RAM)
chip maintain time and date used in the monitor. These circuits utilize
a gold electrolytic super-capacitor for power when the oximeter is
turned off.
IrDA
The IrDA (Infrared Data Association) is an industry standard for
infrared communications. The IrDA receives information from the CPU,
converts it, and sends it to an external printer. The external printer
communicates with the IrDA. Refer to the printer manual for more
information.
44 Service Manual
System Board Components
Backlight
The backlight illuminates the display for better viewing.
Service Manual 45
Page Left Intentionally Blank
46 Service Manual
6 Set-up
Overview
This section provides information on setting up the Nellcor™ OxiMax™
N-65™ handheld pulse oximeter prior to first time usage.
WARNING
To ensure patient safety, do not place the Nellcor OxiMax N-65
handheld pulse oximeter in any position that might tip or fall on the
patient.
WARNING
As with all medical equipment, carefully route cables to reduce the
possibility of patient entanglement or strangulation.
WARNING
Ensure the speaker is clear of any obstruction and speaker holes are not
covered. Failure to do so could result in an inaudible alarm tone.
WARNING
Disconnect the oximeter and OxiMax pulse oximetry sensor from the
patient throughout magnetic resonance imaging (MRI) scanning.
Objects containing metal can become dangerous projectiles when
subjected to the strong magnetic fields created by MRI equipment.
Also, induced current could potentially cause burns.
WARNING
To ensure accurate performance and prevent device failure, do not
subject the Nellcor OxiMax N-65 handheld pulse oximeter to extreme
moisture, such as direct exposure to rain. Such exposure may cause
inaccurate performance or device failure.
WARNING
Do not use an Nellcor OxiMax N-65 handheld pulse oximeter, OxiMax
pulse oximetry sensor, cables, or connectors that appear damaged.
47
Set-up
WARNING
Do not lift the Nellcor OxiMax N-65 handheld pulse oximeter by the
sensor or extension cable. The cable may disconnect, potentially
dropping the pulse oximeter on a patient or a damaging surface.
WARNING
The Nellcor OxiMax N-65 handheld pulse oximeter is not defibrillator-
proof. It may remain attached to the patient during defibrillation or
during use of an electrosurgical unit, however readings may be
inaccurate during the defibrillation and shortly thereafter.
WARNING
Use only Nellcor-approved OxiMax pulse oximetry sensors when
connecting to the OxiMax sensor connector. Connecting any other
sensor influences the accuracy of sensor data, which may lead to
adverse results.
WARNING
Use only the Nellcor DEC-4 or DEC-8 extension cable with the Nellcor
OxiMax N-65 handheld pulse oximeter. Use of another pulse oximetry
cable will have an adverse effect on performance. Do not attach any
cable intended for computer use to the sensor port.
48 Service Manual
Using Battery Power
Battery Installation
To install batteries
1. Turn OFF the power.
2. Pull the battery compartment latch downward toward the bottom
of the oximeter, and remove the battery access door.
3. Properly orient and install four AA size batteries.
Service Manual 49
Set-up
50 Service Manual
Using Battery Power
The LOW BATTERY indicator flashes and a low priority alarm sounds
when approximately 15 minutes of operation remains available.
Replace the batteries.
Figure 9. Display with Low Battery Indicator
Service Manual 51
Page Left Intentionally Blank
52 Service Manual
7 Operation
Overview
Default parameters for the Nellcor™ OxiMax™ N-65™ handheld pulse
oximeter are set at the factory. Institutional parameters may be set on
an individual basis by the clinician.
Institutional parameters remain in effect until power off. At power off,
the oximeter returns to the factory default parameters.
Table 10. Parameter Ranges and Defaults
Factory
Parameter Ranges/Selections Defaults
Alarm Silence Duration Alarms 30, 60, 90, 120 seconds, 30 secs
OFF
Alarm Volume 1 to 10 10
53
Operation
Caution
During the Power-On-Self-Test (POST) immediately after power-up, confirm
all display segments and indicators display and the oximeter speaker sounds
a one-second tone.
Note:
Users must install the four AA batteries, keep a sensor plugged into the
oximeter, and leave the unit ON for 20 minutes before the oximeter retains
time and date information. This is only necessary the first time the time/date
is set. If batteries are removed after being installed for five hours, the
oximeter retains time and date information for approximately 20 minutes.
Note:
Continue to monitor the display for the low battery indicator during power-
on self test.
At every power up, verify the oximeter works properly and is safe to
use. Power-On Self-Test (POST) takes approximately ten seconds to
complete. At power on, POST automatically tests the oximeter
circuitry and functions.
54 Service Manual
Monitoring Oximeter Power
Service Manual 55
Operation
Note:
The POST pass tone also functions as an audible confirmation that the
speaker performs properly.
Note:
Continue to monitor the display for the low battery indicator during power-
on self test.
56 Service Manual
Menu Structure
Menu Structure
This section contains quick reference information on how to set
oximeter parameters and limits using the menu structure.
See Set Menu Parameters, page 57, to change oximeter default
parameters.
• Upper and lower SpO2 limits
• Upper and lower BPM limits
• Alarm volume
• Pulse beep volume
• Data printing
See Set Time/Date Parameters, page 58, to set the time and date.
WARNING
Do not silence the OxiMax N-65 audible alarm, or decrease its volume, if
patient safety could be compromised.
Service Manual 57
Operation
WARNING
The OxiMax sensor uses the date and time provided by the OxiMax N-65
when the patient alarm events are recorded to the memory chip in the
sensor. The accuracy of the date/time used for patient alarm events
depends on the date/time set in the oximeter.
58 Service Manual
Menu Structure
Note:
To set the date and time, press the MENU key during POST.
# of
Presses Parameter Press Function
1 Hour Adjust 1 to 23
2 Minute Adjust 1 to 59
3 Day Adjust 1 to 31
4 Month Adjust 1 to 12
Service Manual 59
Page Left Intentionally Blank
60 Service Manual
8 Oximeter Preventive
Maintenance
Overview
This section describes the steps required to maintain, service, and
properly clean the Nellcor™ OxiMax™ N-65™ handheld pulse oximeter.
Follow local governing ordinance and recycling instructions regarding
the disposal or recycling of the oximeter and its accessories.
Cleaning
WARNING
Do not spray, pour, or spill any liquid on the Nellcor OxiMax N-65
handheld pulse oximeter, its accessories, connectors, switches, or
openings in the enclosure.
For surface cleaning and disinfecting the oximeter, follow your
institution's procedures or the recommended actions below.
• Surface cleaning—Use a soft cloth dampened with either a commercial,
nonabrasive cleaner or a solution of 70 percent alcohol in water, and
lightly wiping the surfaces of the oximeter.
• Disinfection—Use a soft cloth saturated with a ten percent solution of
chlorine bleach in tap water, lightly wiping the surfaces of the oximeter.
Before cleaning an OxiMax™ pulse oximetry sensor, read the
Instructions For Use enclosed with the sensor. Each sensor model has
cleaning instructions specific to that sensor. Follow the OxiMax pulse
oximetry sensor cleaning and disinfecting procedures in the particular
sensor's Instructions for Use.
61
Oximeter Preventive Maintenance
62 Service Manual
9 Performance Verification
Overview
This section provides information to trained service technicians on
setting institutional defaults, verifying Nellcor™ OxiMax™ N-65™
handheld pulse oximeter performance, and accessing service
functions.
WARNING
Only qualified service personnel should open the oximeter housing,
remove and replace components, or make adjustments. If your medical
facility does not have a qualified service technician, please contact
Nellcor Technical Services or your local Nellcor representative.
Performance Verification
Overview
This section discusses the tests used to verify performance following
repairs or during routine maintenance. All tests can be performed
without removing the Nellcor OxiMax N-65 handheld pulse oximeter
cover. Perform all tests before the battery charge and battery
performance checks, then perform both battery checks as the last
operation before returning the pulse oximeter to the caregiver.
If the oximeter fails to perform as specified in any test, correct the
problem before returning the oximeter to the caregiver.
63
Performance Verification
Required Equipment
Equipment Description
Performance Tests
The Power-On-Self-Test (POST) performance test verifies that the
Nellcor OxiMax N-65 handheld pulse oximeter is ready for patient
monitoring.
Power-On-Self-Test
Discussion
WARNING
If you do not hear the POST pass tone, do not use the oximeter.
WARNING
Ensure that the speaker is clear of any obstructions and that the speaker
holes are not covered. Failure to do so could result in an inaudible alarm
tone.
64 Service Manual
Performance Tests
Caution
If any indicator or display element does not light when the oximeter is
turned on, do not use it. Instead, contact qualified service personnel, your
local Nellcor representative, or Nellcor Technical Services.
Caution
During POST (immediately after power-up), confirm that all segments and
indicators display, and the oximeter speaker sounds a one-second tone.
Note:
In addition to serving as the POST pass verification, the POST pass tone also
functions as audible confirmation that the speaker is performing properly. If
the speaker does not function, the alarm warning sounds cannot be heard.
Note:
The OxiMax N-65 automatically starts POST, which tests the oximeter
circuitry and functions.
Before using the oximeter, you must verify that it is working properly
and is safe to use. Proper working conditions are verified each time
the monitor is turned on as described in the following procedure. The
POST verification procedure takes approximately ten seconds to
complete.
When the oximeter is turned on, the backlight remains on. The display
displays the following sequence, in order, as POST takes place:
• All display graphics are shown for three seconds and the backlight is
turned on.
• The display goes blank (all display elements off) for one second.
• The software version number displays for three seconds as a three digit
number in the right number field (with leading zeros if the version
number is less than 100) and two dashes in the left number field.
• The current time of day is shown in 24-hour format.
• Successful completion of POST is announced by a POST pass tone. A
failed POST is announced by a high-priority alarm tone.
Note:
Physiological conditions, medical procedures, or external agents that may
interfere with the oximeter’s ability to detect and display accurate
measurements include: dysfunctional hemoglobin, arterial dyes, low
perfusion, dark pigment, and externally applied coloring agents, such as nail
polish, dye, or pigmented cream.
Service Manual 65
Performance Verification
Note:
Do not connect a sensor the oximeter for this procedure. The backlight
remains on during POST. All display numbers and indicators appear for three
seconds.
1. Perform POST, following the procedure listed in Power-On-Self-Test,
page 64, writing down the displayed current software version to
have available when requesting technical assistance from Nellcor
Technical Services. Upon successful completion of POST, the
monitor sounds a one-second tone to indicate it has passed the
test.
2. Ensure the time is properly set. If it has never been set, do not
continue prior to setting the date and time following the
procedure listed in Set Time/Date Parameters, page 58.
Note:
Current time displays in a 24-hour format.
3. Press the ALARM SILENCE key. The ALARM SILENCE indicator displays.
Figure 16. Sensor Disconnected with Alarm Silence Indicator at Lower Left
66 Service Manual
Operational Setup
4. Press the ALARM SILENCE key. The ALARM SILENCE indicator does not
display.
Figure 17. Sensor Disconnected Indicator, no Alarm Silence
Operational Setup
The operational setup procedure sets the time and date into the
oximeter. See Set Time/Date Parameters, page 58, or Set Menu
Parameters, page 57, for individual parameter settings. The oximeter
returns to the factory default settings when it is powered on, so set
individual parameter settings at power up.
Service Manual 67
Performance Verification
68 Service Manual
Pulse Oximetry Functional Tests
Overview
The SRC-MAX functional tester enables qualified technicians to
functionally test Nellcor OxiMax technology- based pulse oximeters
and OEM OxiMax technology-based oximeters. After attaching the
Service Manual 69
Performance Verification
Test Description
Test Setup This procedure establishes the baseline for all the other tests.
Test Setup must be performed before performing any or all of
the SRC-MAX tests.
BPM Test The test procedure simulates an OxiMax sensor attached to a
patient indicating 60 bpm and 200 bpm. The test sets up Test
1 for 60 bpm.
SpO2 Test The test procedure simulates an OxiMax sensor attached to a
patient, indicating 75 percent blood oxygen saturation and 90
percent blood oxygen saturation. The test sets up Test 2 for
75 percent blood oxygen saturation.
Modulation Level The test procedure simulates an OxiMax sensor attached to a
Test patient indicating low and high pulse strength. The test sets
up Test 3 for low pulse strength.
Light Level Test The test procedure simulates an OxiMax pulse oximetry
sensor attached to a patient indicating low and high light level
passing through the patient at the sensor site. The test sets up
Test 4 for low light level.
Note:
The SRC-MAX selectable indicator LEDs may extinguish if there is a delay in
proceeding through the above tests. This is normal operation in order to
increase the battery time.
Note:
Pressing a key on the SRC-MAX during the test procedures may be
requested, changing a certain parameter. If the SRC-MAX LEDs are not lit,
press the key twice. Pressing the key once causes the indicators to relight
and pressing twice initiates the change.
Test Setup
Note:
Use a DEC-4 extension cable between the SRC-MAX and the monitor.
You must complete Test Setup before performing any other pulse
oximetry functional tests. Ensure all initial procedures for oximeter
setup are complete, following Power-On-Self-Test, page 64.
70 Service Manual
Pulse Oximetry Functional Tests
Service Manual 71
Performance Verification
SRC-MAX panel lights and in a few seconds the oximeter tracks the
simulator.
BPM Test
1. Press the SRC-MAX PULSE RATE selection key. The SRC-MAX PULSE
RATE 200 LED lights. The oximeter BPM increases until stabilizing at
200 BPM. The test pass criteria is 197 to 203 BPM inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 200 bpm (pass criteria is 197 to 203 bpm inclusive)
• Displays pulse amplitude indicator at low level modulation
72 Service Manual
Pulse Oximetry Functional Tests
2. Press the SRC-MAX PULSE RATE select key. The SRC-MAX PULSE RATE
60 LED lights. The oximeter registers BPM decreases until
stabilizing at 60 BPM. (The test pass criteria is 57 to 63 BPM
inclusive.)
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 60 BPM (test pass criteria is 57 to 63 bpm inclusive)
• Displays pulse amplitude indicator at low level modulation
Service Manual 73
Performance Verification
SpO2 Test
1. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 90 LED
lights. The oximeter displays two sets of dashes [ - - ] until the
%SpO2stabilizes at a value between 88 to 92 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds no audible alarm
• Displays 90 %SpO2 (pass criteria is 88 to 92 %SpO2 inclusive)
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
Figure 23. SpO2 Test Increase to 90 SpO2
2. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 75 LED
lights. The oximeter displays two sets of dashes [ - - ] until
stabilizing at a value between 73 to 77 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds SpO2 audio alarm
• Displays 75 %SpO2 (pass criteria is 73 to 77 %SpO2 inclusive)
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
74 Service Manual
Pulse Oximetry Functional Tests
Service Manual 75
Performance Verification
2. Press the SRC-MAX PULSE RATE selection key. The SRC-MAX PULSE
RATE 200 LED lights. The oximeter BPM increases until stabilizing at
200 BPM. The test pass criteria is 197 to 203 BPM inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 200 bpm (pass criteria is 197 to 203 bpm inclusive)
• Displays pulse amplitude indicator at high level modulation
Figure 25. Mod Test with Increase to 200 BPM at High Modulation
3. Press the SRC-MAX PULSE RATE select key. The SRC-MAX PULSE RATE
60 LED lights. The oximeter registers BPM decreases until
stabilizing at 60 BPM. (The test pass criteria is 57 to 63 BPM
inclusive.)
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 60 BPM (test pass criteria is 57 to 63 bpm inclusive)
• Displays pulse amplitude indicator at high level modulation
76 Service Manual
Pulse Oximetry Functional Tests
4. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 90 LED
lights. The oximeter displays two sets of dashes [ - - ] until the
%SpO2stabilizes at a value between 88 to 92 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds no audible alarm
• Displays 90 %SpO2 (pass criteria is 88 to 92 %SpO2 inclusive)
• Displays 60 BPM
• Displays pulse amplitude indicator at high level modulation
Figure 26. Mod Test with Increase to 90 SpO2 at High Modulation
5. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 75 LED
lights. The oximeter displays two sets of dashes [ - - ] until
stabilizing at a value between 73 to 77 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds SpO2 audio alarm
• Displays %SpO2 value between 73 to 77 inclusive
• Displays 60 BPM
• Displays pulse amplitude indicator at high level modulation
Service Manual 77
Performance Verification
78 Service Manual
Pulse Oximetry Functional Tests
7. Press the SRC-MAX PULSE RATE selection key. The SRC-MAX PULSE
RATE 200 LED lights. The oximeter BPM increases until stabilizing at
200 BPM. The test pass criteria is 197 to 203 BPM inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 200 bpm (pass criteria is 197 to 203 bpm inclusive)
• Displays pulse amplitude indicator at low level modulation
Figure 28. Mod Test Increase to 200 BPM at Low Modulation
8. Press the SRC-MAX PULSE RATE select key. The SRC-MAX PULSE RATE
60 LED lights. The oximeter registers BPM decreases until
stabilizing at 60 BPM. (The test pass criteria is 57 to 63 BPM
inclusive.)
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 60 BPM (test pass criteria is 57 to 63 bpm inclusive)
• Displays pulse amplitude indicator at low level modulation
Service Manual 79
Performance Verification
9. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 90 LED
lights. The oximeter displays two sets of dashes [ - - ] until the
%SpO2 stabilizes at a value between 88 to 92 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds no audible alarm
• Displays 90 %SpO2 (pass criteria is 88 to 92 %SpO2 inclusive)
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
Figure 29. Mod Test Increase to 90 SpO2 at Low Modulation
10.Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 75 LED
lights. The oximeter displays two sets of dashes [ - - ] until
stabilizing at a value between 73 to 77 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds SpO2 audio alarm
• Displays %SpO2 value between 73 to 77 inclusive
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
80 Service Manual
Pulse Oximetry Functional Tests
Service Manual 81
Performance Verification
2. Press the SRC-MAX PULSE RATE selection key. The SRC-MAX PULSE
RATE 200 LED lights. The oximeter BPM increases until stabilizing at
200 BPM. The test pass criteria is 197 to 203 BPM inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 200 bpm (pass criteria is 197 to 203 bpm inclusive)
• Displays pulse amplitude indicator at low level modulation
Figure 31. Light Level Test Increase to 200 BPM with HIgh Light Level
3. Press the SRC-MAX PULSE RATE select key. The SRC-MAX PULSE RATE
60 LED lights. The oximeter registers BPM decreases until
stabilizing at 60 BPM. (The test pass criteria is 57 to 63 BPM
inclusive.)
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 60 BPM (test pass criteria is 57 to 63 bpm inclusive)
• Displays pulse amplitude indicator at low level modulation
82 Service Manual
Pulse Oximetry Functional Tests
4. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 90 LED
lights. The oximeter displays two sets of dashes [ - - ] until the
%SpO2 stabilizes at a value between 88 to 92 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds no audible alarm
• Displays 90 %SpO2 (pass criteria is 88 to 92 %SpO2 inclusive)
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
Figure 32. Light Level Test Increase to 90 SpO2 with HIgh Light Level
5. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 75 LED
lights. The oximeter displays two sets of dashes [ - - ] until
stabilizing at a value between 73 to 77 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds SpO2 audio alarm
• Displays %SpO2 value between 73 to 77 inclusive
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
6. Press the SRC-MAX LIGHT LEVEL selection key. The SRC-MAX LIGHT
LEVEL LED lights. The oximeter pulse blip bar initially increases in
amplitude and then reduces and stabilizes at 3-5 bars.
Service Manual 83
Performance Verification
8. Press the SRC-MAX PULSE RATE select key. The SRC-MAX PULSE RATE
60 LED lights. The oximeter registers BPM decreases until
stabilizing at 60 BPM. (The test pass criteria is 57 to 63 BPM
inclusive.)
Ensure the oximeter exhibits the following behaviors.
• Sounds BPM audio alarm
• Displays 75 %SpO2
• Displays 60 BPM (test pass criteria is 57 to 63 bpm inclusive)
• Displays pulse amplitude indicator at low level modulation
84 Service Manual
Pulse Oximetry Functional Tests
9. Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 90 LED
lights. The oximeter displays two sets of dashes [ - - ] until the
%SpO2 stabilizes at a value between 88 to 92 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds no audible alarm
• Displays 90 %SpO2 (pass criteria is 88 to 92 %SpO2 inclusive)
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
Figure 34. Light Level Test Increase to 90 SpO2 with Low Light Level
10.Press the SRC-MAX %SpO2 select key. The SRC-MAX %SpO2 75 LED
lights. The oximeter displays two sets of dashes [ - - ] until
stabilizing at a value between 73 to 77 inclusive.
Ensure the oximeter exhibits the following behaviors.
• Sounds SpO2 audio alarm
• Displays %SpO2 value between 73 to 77 inclusive
• Displays 60 BPM
• Displays pulse amplitude indicator at low level modulation
11.Disconnect all equipment and turn off the oximeter.
Service Manual 85
Page Left Intentionally Blank
86 Service Manual
10 Troubleshooting
Overview
This section explains how to troubleshoot common problems while
using the Nellcor™ OxiMax™ N-65™ handheld pulse oximeter. This
section also includes possible oximeter difficulties, along with
probable causes, recommended actions to correct the difficulty, and
how to obtain technical help and support.
WARNING
Only qualified service personnel should open the oximeter housing,
remove and replace components, or make adjustments. If your medical
facility does not have qualified service personnel, contact Nellcor
Technical Services or your local Nellcor representative.
WARNING
Do not spray, pour, or spill any liquid on the Nellcor OxiMax N-65
handheld pulse oximeter, its accessories, connectors, switches, or
openings in the enclosure.
87
Troubleshooting
Technical Services
When encountering a problem while using the Nellcor OxiMax N-65
handheld pulse oximeter which cannot be resolved through a visual
inspection, refer to Table 15: Troubleshooting Guide, on page 89.
Locate the condition and follow corrective actions in the order
presented. For a symptom not listed in Table 15, contact Nellcor
Technical Services Department at 1.800.635.5267, or your local Nellcor
representative.
If an error code is shown on the oximeter front panel display, turn the
oximeter off, wait ten seconds, then turn it back on. This clears
recoverable errors encountered by the oximeter. If the error code
persists, refer to Table 16: Error Codes, on page 93 for an explanation of
the error code. Take the action specified.
Figure 35. Typical Error Code Display
88 Service Manual
Help and Support
The oximeter does not turn on after pressing the POWER key.
The POWER key not fully depressed. Fully press the POWER key.
The batteries are at, or near, a voltage Install new batteries. Refer to Battery
too low for the oximeter to operate. Installation, page 49.
Replacing the front case does not fix Replace the SpO2 PCB. See SpO2 PCB or
the problem. User Interface PCB Replacement, page 107.
Replacing the SpO2 PCB does not fix Replace User Interface PCB. See SpO2 PCB
the problem. or User Interface PCB Replacement, page
107. If this does not fix problem, return
oximeter to Nellcor.
Service Manual 89
Troubleshooting
One or more keys on the front panel keypad does not work.
Replacing the front case does not fix Replace SpO2 PCB. page 107, SpO2 PCB or
the problem. User Interface PCB Replacement.
Replacing the SpO2 PCB does not fix Replace User Interface PCB. See page 107,
the problem. SpO2 PCB or User Interface PCB
Replacement. If this does not fix problem,
return oximeter to Nellcor.
A User Interface PCB component has Inspect User Interface PCB for cracking,
failed. burning, or damage. If damage is found,
replace User Interface PCB. See SpO2 PCB
or User Interface PCB Replacement, page
107.
The User Interface PCB is defective. If damage is found, replace User Interface
PCB. See SpO2 PCB or User Interface PCB
Replacement, page 107.
Pulse rate and %SpO2 value does not display when the oximeter is turned on.
90 Service Manual
Help and Support
The User Interface PCB is defective. Replace User Interface PCB. See SpO2 PCB
or User Interface PCB Replacement, page
107.
The SpO2 PCB is defective. Replace SpO2 PCB. See SpO2 PCB or User
Interface PCB Replacement, page 107.
The display backlight does not come on when pressing the BACKLIGHT key.
Replacing the SpO2 PCB does not fix Replace SpO2 PCB. See page 107, SpO2
the problem. PCB or User Interface PCB Replacement.
Replacing the SpO2 PCB does not fix Replace User Interface PCB. See page 107,
the problem. SpO2 PCB or User Interface PCB
Replacement. If this does not fix the
problem, return oximeter to Nellcor.
The BACKLIGHT LEDs on the User Replace User Interface PCB. See page 107,
Interface PCB have failed. SpO2 PCB or User Interface PCB
Replacement.
The oximeter shuts off when you press the Backlight key.
The batteries are, at or near, a voltage Install new batteries. See Battery
too low for the oximeter to operate. Installation, page 49.
Service Manual 91
Troubleshooting
The IrDA between the printer and the Align printer and oximeter. See the
oximeter is not properly aligned. Operator’s Manual for instructions.
The printer batteries are at, or near, a Install new batteries in printer. See printer
voltage too low to operate. documentation.
Replacing the Front Case does not fix Replace User Interface PCB. See page 107,
the problem. SpO2 PCB or User Interface PCB
Replacement. If this does not fix problem,
return oximeter to Nellcor.
Error Codes
An error code displays when the oximeter detects a non-correctable
failure. Table 16 provides a list of error codes for the oximeter.
However, turn the POWER off and on once (cycling power) to remove
the error code if one of the following occurs:
• Oximeter sounds a low priority alarm
• Measurements stop
• “Er” is shown in the %SpO2 numeric display area
• Error code is shown in the right numeric display
If the error code remains, see Error Codes, page 93.
92 Service Manual
Error Codes
Scope of
Action
Clinician
Return
Repair
Error
Code Action Explanation
Service Manual 93
Troubleshooting
Scope of
Action
Clinician
Return
Repair
Error
Code Action Explanation
94 Service Manual
Error Codes
Scope of
Action
Clinician
Return
Repair
Error
Code Action Explanation
271 X Replace SpO2 board. SpO2 back end reports not ready/
not initialized.
273 1, 3 Restart the oximeter. SpO2 back end reports date out-
2 Set the time and of-range error.
date.
Replace UI board.
Service Manual 95
Troubleshooting
Scope of
Action
Clinician
Return
Repair
Error
Code Action Explanation
96 Service Manual
Error Codes
Scope of
Action
Clinician
Return
Repair
Error
Code Action Explanation
Service Manual 97
Troubleshooting
Scope of
Action
Clinician
Return
Repair
Error
Code Action Explanation
98 Service Manual
11 Ordering Oximeter Spare
Parts
Overview
This section describes how to order spare parts for the Nellcor™
OxiMax™ N-65™ handheld pulse oximeter. Item numbers correspond
to the call-out numbers in Figure 36, Exploded View on page 100.
This manual is available on the Internet at:
http://www.nellcor.com/
99
Ordering Oximeter Spare Parts
Reference Description
Spare Parts
2 SpO2 PCB
5 Battery cover
-- LCD assembly
Accessories
-- Transport Boot
Overview
WARNING
Only qualified service personnel should open the oximeter housing,
remove and replace components, or make adjustments. If your medical
facility does not have qualified service personnel, contact Nellcor
Technical Services or your local Nellcor representative.
Caution
Observe ESD (electrostatic discharge) precautions when disassembling and
reassembling the oximeter and when handling any components of it.
Nellcor™ OxiMax™ N-65™ handheld pulse oximeter major component
parts include all PCBs and subassemblies.
• Top case with front panel keypad
• Bottom case
• Battery cover
• User Interface PCB
• SpO2 PCB
Collect the following tools prior to disassembly of the oximeter.
• Number one (medium) phillips-head screwdriver
• Torque wrench, 4-in lbs. (10 cm/kg)
Only the qualified service technician may disassemble the oximeter to
its major component parts. The supported replacement level for the
Nellcor OxiMax N-65 handheld pulse oximeter is to the printed circuit
board (PCB) and major subassembly level. After isolating the problem
to a suspected PCB, follow the procedures in page 103, Repairing the
Oximeter, then proceed to replace the faulty PCB with a known good
PCB. Verify the symptom disappears and ensure the oximeter passes
all performance tests. If the symptom persists, swap the replacement
PCB with the suspected malfunctioning PCB (the original PCB installed
when you started troubleshooting) and continue troubleshooting.
103
Repairing the Oximeter
Note:
Some spare parts have a business reply card attached. After receiving the
spare parts, please complete and return the business reply card.
Note:
The OXIMAX N-65 is RoHS compliant for products manufactured as of July
2014. The RoHS compliant spare parts, including the SpO2 PCBA and User
Interface PCBA, are compatible with OXIMAX N-65 Handheld Pulse
Oximeters manufactured prior to this date.
Caution
A flex cable runs between the oximeter top and bottom case assemblies. Use
care when separating the cases. Do not apply stress to the flex cable.
Caution
Failure to unlock connector J4 on the SpO2 PCB before removing the flex
cable could damage the flex circuit.
3. Unlock connector J4 on the SpO2 PCB.
1 — SpO2 PCB
2 — J4 lock
3 — Lock direction
4 — Unlock direction
Caution
Do not force the User Interface PCB and SpO2 PCB apart. This may damage
the boards.
5. Separate the User Interface PCB from the SpO2 PCB using the
following steps.
a. Grasp the User Interface PCB in one hand and the SpO2 PCB in the
other.
b. Rotate the ends of the two PCBs until they separate at the connection
points.
c. Gently work the boards apart.
Figure 42. Separating the User Interface PCB and SpO2 PCB
10. Install the User Interface PCB and SpO2 PCB into the bottom case,
aligning battery springs properly in their slots.
Note:
The longer screws install at the top of the case.
11. Align the top and bottom cases and install the four screws.
1 — SpO2 PCB
2 — J4 lock
3 — Lock direction
4 — Unlock direction
6. Align the top and bottom cases and install the four screws.
3. Lift the SpO2 PCB and User Interface from the case.
Note:
The longer screws install at the top of the case.
6. Place the front and rear cases together and install the four screws.
2. Insert the new battery cover in the bottom case assembly and
close until the latch snaps in place.
Overview
This section focuses specifically on Nellcor™ OxiMax™ N-65™ handheld
pulse oximeter printed circuit board (PCB) design and schematics.
The Nellcor OxiMax N-65 handheld pulse oximeter consists of two
printed circuit boards (PCB), the user interface PCB and the SpO2 PCB.
These two components provide support for data exchange and input
from the clinician, technician, or caregiver during use.
For system component descriptions, see Schematics, page 119.
• User Interface Printed Circuit Board (Sheets 1 through 3)
• SpO2 Printed Circuit Board (Sheets 1 through 7)
119
Schematics
Power key 16
A Current, Leakage 35
Accessories
Internet address 11 D
Alarm, Volume 57
Date
As 9 Set the day 59
Set the month 59
B Set the year 59
Description
Battery Front panel 12
Critically low battery indication 51 Front panel symbols 14
Low battery, 15-minute warning 50 Pulse oximeter product 11
Using battery power 48 Rear panel symbols 14
Battery cover, replacement 116 Disassembly
Block diagram 42 Required tools 103
Disinfecting the pulse oximeter 61
C Display
%SpO2 value 17
Cable, extension 64 Pulse rate (BPM) value 18
Calibration, not required 61 Display elements, pulse oximeter 17
Carton label symbols, see Symbol 21 Dyes, arterial 65
Case
Bottom case assembly removal 113
Top case assembly removal 111 E
Caution EMC standards 31
Battery disposal 9 Entanglement, warning 9
ESD precautions 9 Error codes 92
Maximum screw tightness 9 Exploded view 101
Cautions, general 7
External agents 65
Cleaning
Disinfection 61
Surface cleaning 61 F
Cleaning the pulse oximeter 61
Front panel, symbols 14
Conditions, Performance issues 65
Functional checks, see Performance veri-
Controls
fication 62
Adjust down key 15
Adjust up key 16
Alarm silence key 15 H
Backlight key 16
Data (record/print) key 15 Help and support, documentation 87
Menu key 15 Hemoglobin, dysfunctional 65
©2011 Covidien.
Covidien llc,
15 Hampshire Street, Mansfield, MA 02048 USA.
Covidien Ireland Limited,
IDA Business & Technology Park,Tullamore.