Professional Documents
Culture Documents
9 - Proper Technique of Sectioning Tissue Samples
9 - Proper Technique of Sectioning Tissue Samples
TECHNIQUES
PROPER TECHNIQUE OF SECTIONING TISSUE SAMPLES
2ND SEMESTER | FINALS | PROF. JOHN MARKE BERNARDO
● The correct knife angle is a property of the type of If the ribbon wraps itself around your finger, the knife
knife and the final bevel, and may be different on blade or anything else:
knives from different manufacturer’s and may change ● The problem is static electricity, which is often present
after re-sharpening. in very dry room.
● To reduce the likelihood of changes in knife angle with ● Let boiling water stand nearby, section in the morning
sharpening, the knife should always be hours or make short ribbons.
professionally sharpened.
PROCEDURES
ERRORS
Preparation of the Block of Tissue for Sectioning
If the ribbon is crooked rather than straight: 1. Remove excess wax with a scalpel of sharp knife as
● The upper and lower edges of the block face are not demonstrated so that the tissue, surrounded by wax
parallel to each other or to the knife edge. stands out from the block.
● The knife edge may be uneven. Try another section 2. Trim the upper and lower edges of the block face so
of the knife edge. that they are parallel.
If a ribbon will not form and each section comes off 3. Trim the lateral surfaces so that the resultant face
separately. resembles a trapezoid. The tissue should be
● The blade edge is dull. Sharpen it. surrounded by about a 2-3 mm frame of wax.
● The knife has the wrong tilt. Adjust it. Preparation of the Microtome
● The upper and lower edges of the paraffin block face 1. The microtome should be clean when you start.
are crumbled or rounded. Retrim with a sharp razor Brush away all discarded paraffin ribbons.
blade. 2. Test the action of the hand wheel. If it is stiff, oil it and
If the sections are compressed or folded: clean out the interior of the mechanism.
● The knife is dull. Sharpen it. 3. If the block holder is advanced toward its maximum,
● The knife angle is set too close to the vertical. retract it fully and reset the mechanism.
Increase the angle. 4. Lock the hand wheel or leave in the position at which
● The knife is clogged up with paraffin. Clean the edge the block holder is in its highest position.
with chloroform. Mounting the Block
● The sections are too thin. Increase the thickness 1. Clamp the specimen into the chuck of the block
setting. holder.
● The paraffin is too soft because the room is warm. 2. Adjust the square face of the paraffin block vertically
Pack the blade with ice cubes. Sometimes holding an and horizontally.
ice cube against the face of the block helps. 3. Make sure the block's upper and lower edges are
● The paraffin is still contaminated with xylene. parallel with each other and parallel with the base of
Re-embed. the microtome.
If the specimen crumbles or falls out of the paraffin 4. The care with which the block is initially adjusted is
section : important to the success of sectioning.
● The tissue is inadequately dehydrated or cleared. Blade Preparation
Dissolve off the paraffin, rehydrate, and clear. 1. Make sure the microtome is clean. Wipe the blade
● The tissue is too hard and compact (e.g., liver). Soak carefully with chloroform or xylene to clean the edge,
it in water or 70 % ethanol to soften it. if used.
● The tissue was in the paraffin oven too long at too 2. Place the microtome blade in the blade holder (or
high a heat. Throw it away. razor blade holder with its clamped razor blade).
● The tissue is too hard for a paraffin matrix (e.g., bone, 3. Carefully check the angle of the blade, the firmness of
cuticle, heavy plant fibers, or xylem). Try celloidin its seating in the holder, and its clearance of the
embedding. paraffin block and block holder. The angle of the
If the ribbon continuously splits or scratches center line of the blade with the face of the block
vertically: should be about 20 degrees. DO NOT CHANGE THE
● Crystals or dirt particles are caught on the knife edge ANGLE
or on the paraffin face. Wipe the knife edge 4. If necessary, readjust the paraffin block.
carefully and clean it with chloroform. 5. Some very careful trimming can be done with a razor
● The knife edge is nicked. Use another section of the blade.
knife or sharpen it.. 6. Watch your fingers- the blade is sharp!
7. The upper and lower edges of the block must be
parallel to each other and to the edge of the knife.
paraffin blocks, cutting both large and small ○ Sensitive tissue constituents to be studied
tissues with ease are damaged or destroyed by heat.
○ Recommended for cutting extremely hard ● Although other microtomes can be modified for cutting
and rough tissue blocks. frozen section, this type will give the best results and
○ It is the most dangerous type of microtome is used almost universally.
due to the movable exposed knife. ● The freezing microtome is equipped with a stage
i. A slow but very steady motion is upon which tissue can be quickly frozen using either
therefore required to manipulate the liquid carbon dioxide, from a cylinder, or a low
instrument. temperature recirculating coolant.
FREEZING MICROTOME ● The cutting action of the freezing microtome differs
from those described previously as in this case the
knife is moved whilst the tissue block remains static,
same as STANDARD sliding microtome.
CRYOSTAT OR COLD MICROTOME
(typically 60 to 100 nanometers) of tissue when the instrument is left unattended or when
embedded in epoxy resin. cleaning the instrument.
● Sections are stained with an aqueous solution of an MICROTOME KNIVES
appropriate heavy metal salt and examined with a ● Trimming and section-cutting are done with a
transmission electron microscope (TEM). microtome knife, which is available in three basic
● Also used with its glass knife or an industrial grade types or shapes:
diamond knife to cut semi-thin sections prior to thin
sectioning.
● These semi-thin sections are generally 0.5 to 1 μm
thick and are mounted on a glass slide and stained to
locate areas of interest under a light microscope prior
to thin sectioning for the TEM.
○ Thin sectioning for the TEM is often done
with a gem quality diamond knife.
CARE OF THE MICROTOME AND SAFETY
MEASURES
● After sectioning, all the accumulated paraffin and MICROTOME KNIVES (ACCORDING TO SHAPES)
small pieces of tissues must be brushed away with a
soft brush and not allowed to stay in the microtome, PLANE-CONCAVE ● 25 mm in length
since this may later interfere with the cutting of tissue KNIFE ● One side of the knife is flat
blocks. while the other is concave.
● After carefully drying the machine and knife holder, ○ Less Concave (flat) -
the parts should be wiped with xylol. Prolonged and Celloidin-Embedded
tissues on a Sliding
continuous application of the painted parts with xylene Microtome
should, however, be avoided since this reagent is ○ More concave -
capable of removing the paint. Paraffin sections on
● Movable portions should be oiled thoroughly to base-sledge, rotary or
prevent rusting rocking microtome.
● The microtome must always be covered when not in
use, to prevent accumulation of dust and other dirt BICONCAVE KNIFE ● 120 mm in length
which may later on interfere with the normal ● Both sides concave
sectioning of tissues. ● For cutting paraffin -
● The microtome should be placed on a stable bench, embedded sections on a
away from air drafts, doorways and passing staff. Any rotary microtome.
air movement from air conditioners or other causes
PLANE-WEDGE KNIFE ● 100 mm in length
can make section handling very difficult.
● Both sides straight
● Always remove the knife or blade before cleaning.
● For frozen sections or for
The knife holder can easily be removed to facilitate
cutting extremely hard
access for cleaning. No fluid must enter the inside of
and tough specimens
the instrument during cleaning.
embedded in paraffin
● When cleaning the blade avoid dragging anything
blocks, using a base
along the cutting edge. Even cellulose fibers can
sledge type or sliding
cause damage to the blade.
microtome.
● Have the instrument inspected at least once a year by
a qualified service technician. ● Plane-wedge and plane-concave knives are usually
Safety Measures provided with backs, to maintain the correct bevel
● It is very important that staff are not distracted when angle throughout honing. Detachable handles may be
using the microtome because of the risks of injury attached to the knife during sharpening.
from extremely sharp blades. It is preferable to have OTHER DETAILS ABOUT MICROTOME KNIVES
non-slip flooring in the vicinity of microtomes because, ● BEVEL ANGLE - angle formed between cutting
inevitably, wax fragments will find their way onto the edges
floor where they can produce a slippery surface. ○ About 27° to 32°
● Use forceps or brush instead of fingers to pick up ○ Cutting facet (bevel) found on the tapered
sections or wax fragments from blade or block face. edge of all knives, the sides of which are
● Use hand wheel lock when changing blocks. The more acutely inclined towards each other
knife or blade should be removed from the microtome
than the side proper, forming the actual of the guard to start moving the blade
cutting edge of all knives. laterally out of the clamp.
○ Such angle is maintained for each knife by ○ It can then be grasped with forceps (not
means of a slide-on back, a spring-loaded fingers) and safely removed.
semi-circular metal sheet slipped on to the ● Used blades should be disposed of appropriately in a
knife with one or more plane surfaces “sharps” container or into the “used blades” slot in the
(plane-wedge or plane-concave) to hold the base of the blade dispenser.
cutting edge at a constant, correct angle ● Never place a knife or blade on the bench or in a box
during the process of honing and stropping. with the cutting edge facing up.
○ Each knife should have its own HONING AND STROPPING
corresponding back which should not be ● Badly nicked knives with blunted ends have to
interchanged with another, to keep the bevel undergo sharpening in order to ensure optimum
angle. sectioning of tissue blocks and prevent gross
● A good cutting edge should be made of good quality irregularities on the tissue sections.
steel. ● Jagged edges, if not corrected, will produce tears or
○ Too soft cutting edges - dull easily striations in tissue sections. Sharpening of the knife
○ Too hard cutting edges - likely to produce involves two stages:
nicks or jagged edges and irregularities on HONING (HARD SHARPENING)
the knife edge, producing tears/striations on ● Honing involves the removal of gross nicks on the
the tissue sections knife edge (Coarse Honing) to remove blemishes,
○ A good cutting edge must be able to cut and grinding the cutting edge of the knife on a
good sections from a paraffin wax block stone (Honing Proper) to acquire an even edge.
about 2-3 microns thick, without any ● The degree of sharpness is proportional to the
serration noted on examination. fineness of the abrasive used in sharpening.
● SAFETY RAZOR BLADES ● This procedure makes use of a hone, a natural
○ May be used for partially calcified sharpening stone, or hard grinding surface (e.g.
materials, paraffin, and frozen sections. carborundum), which serves to remove nicks and
○ Readily replaced when dull, and produce irregularities on the knife edges.
similarly good tissue sections as those cut ● Several types of hones may be used
with microtome knives. 1. Belgium Yellow - For manual sharpening
○ They are, however, unsatisfactory for when cutting edge has been rendered blunt
sections less than 10 μm. or nicked. This type usually gives the best
● CUTTING ANGLE result.
○ Theoretically, the perfect and optimum 2. Arkansas - Gives more polishing effect than
cutting angle is obtained when the sides of the Belgium Yellow.
the wedge knife are inclined at an angle of 3. Fine Carborundum -
about 15°, causing maximum penetration a. Much coarser than the first two
of the tissues and minimizing distortion. types
● CLEARANCE ANGLE b. used only for badly nicked knives
○ To prevent uneven sections, or alternate thin followed by either one of the first
and thick sections, the knife should be two knife sharpeners.
inclined with a 5-10° clearance angle from ● The surface of the hone is wiped clean with a soft
the cutting plane so that the cutting facet will cloth moistened with xylene in order to remove the
not compress the block during the scattered small particles of stones and metal.
process of cutting. ● It is then covered with a thin film of Mineral and
● The cutting edge must be thinner than the section Clove Oil, Xylene, Liquid Paraffin or Soapy Water
being cut. for lubrication.
○ A good cutting edge must be sharp enough ● Honing Technique
to cut good sections from a paraffin wax ○ The knife is fitted to its corresponding back,
block at 4 μm thick without causing placed on one end of the hone, and with the
serrations. cutting knife edge first, the "heel" (handle
● The knife or blade should be removed from the end) is drawn obliquely or diagonally
microtome when the instrument is left unattended towards the operator on the stone until
or when cleaning the instrument. the "toe" (head portion) is reached.
○ This is best done by unclamping the blade,
then using the blade ejector on the left side
○ The knife is then turned over, and the other time-saving and produce well sharpened
surface is again drawn forward, EDGE knives with uniform bevels.
FIRST, with a HEEL TO TOE direction. ● PRECAUTIONS DURING HONING
○ Hone is placed on non-skid surface. A damp ○ The hone should be long enough (about 8" x
cloth may be used-to prevent movement of 3") to allow the whole length of the knife
the hone. Light lubricating oil or soapy water edge to be sharpened in a single stroke and
is used for lubrication. wide enough to sufficiently support and
● Step-by-Step Process in Honing prevent the rocking of the knife.
1. The knife complete with handle and backing sheath is ○ The hone should be lubricated with warm
laid on the Hone with the cutting edge facing away soapy water or fine oil before using. It is then
from the operator and the heel roughly at the center of washed, preferably with water, to remove all
the nearest end of hone. metal particles that may have been collected
2. The knife is held with thumb on the back and during the process. The washing fluid used
forefinger on the front surface. The knife is pushed must flow rapidly enough so that the metal
forward diagonally from heel to toe to the other end of chips are removed between strokes and a
the hone, turned over on its back and moved across clean hone is presented every time.
the hone until the heel is in the center with the cutting ○ The pressure on the knife should be gentle
edge leading and then brought back diagonally. and steady to keep it from rocking. The
3. It is then turned across the hone to its original number of strokes usually amounts to 20-30
position. Such sequence forms a double stroke, with a times in each direction, depending upon the
knife held obliquely, taking the same precaution to condition of the knife. Badly nicked knives
hone the entire length of the knife. require greater and longer honing than less
4. Honing is then continued until all the teeth in the knife irregular knives.
edge have been eradicated. ○ The hone should be cleaned before, during,
a. In the case of the Minot or plane-wedge and after use. A black film that develops in
knife, the knife is turned over so as to the hone usually is imparted by the knife that
sharpen the other surface every 10-20 is being sharpened, and should be brushed
strokes. out with a good nailbrush in running water,
b. For plane-concave knives, only the which may either be plain or soapy, until the
concave surface should be rubbed on the hone is thoroughly cleaned. After its use, the
Hone. hone must be washed with warm soapy
5. A flat circular glass plate with finely powdered water, dried, and kept in a box to protect it
aluminum oxide made into paste with water (used as from dust while it is not in use.
an abrasive) may be used for grinding and removing ○ After honing, wipe off the oil or soap from the
nicks. knife with xylene. Then strop it thoroughly
a. Diamantine may also be used for final STROPPING
polishing. ● Process whereby the "burr" formed during honing
6. The plate glass is usually 1/4 to 3/8 inch thick, about is removed and the cutting edge of the knife is
14 inches long and 1-2 inches wider than the length of polished.
the knife blade to be sharpened. ● The purpose of stropping is to polish and sharpen the
7. Due to the plate's relatively greater width, the knife cutting edge, while that of honing is to remove the
blade does not have to be held obliquely, but is irregularities from the knife.
pushed and pulled forward and backward at right ● If the knife has become dull and blunt, but is free from
angles to the transverse diameter of the plate. nicks or teeth, it is usually only necessary to strop it.
● Mechanical Honing with machines For delicate work, the knife is stropped before every
○ May make use of a vibrating frosted glass object is sectioned.
plate or a wheel driven by an electrical ● PADDLE STROP
motor. ○ Preferred option: Made from the best quality
○ The knife is pressed against the flat side of a horse leather firmly attached to a solid back,
rotating glass wheel which is being driven by in order to prevent sagging
a mechanical device. ● The procedure is the reverse of honing.
○ Approximately 30 double strokes are given ○ The knife is first fitted with its appropriate
each side of the knife to which very gentle knife back, then laid obliquely on the strop
pressure is applied. and with the cutting edge behind, EDGE
○ The use of knife sharpening machines, LAST is pushed backward and drawn
although quite expensive, is usually forward in a TOE TO HEEL direction.
○ Around 40-120 double strokes are usually ○ Wax must not be allowed to come in contact
required. with the strop. With an applicator, the used
○ In the case of plane-wedge or Minot knives, knife blade should be washed and flushed
the knife is turned around at the end of each with xylene. The knife is then dried off, by
stroke so as to sharpen each surface wiping the knife on a soft paper or cloth
alternately. (NEVER wipe the paper or cloth on the
○ For planeconcave knives, only the concave knife). The procedure is again repeated with
surface should be stropped. fresh xylene and a fresh sheet of paper or
● PRECAUTIONS DURING HONING both, until all the wax has been removed.
○ The knife should always be wiped clean with BLADES/KNIVES
a soft cloth before and after a series of ● DISPOSABLE BLADES
stropping strokes and before changing from ○ Sharpening (honing) and polishing
a coarse to a fine strop to remove particles (stropping) are no longer common practice in
which may have been taken off the knife. most modern laboratories because of the
After stropping is satisfactorily completed, availability of disposable knives that are
the knife edge is then oiled or greased to cheaper to use than conventional steel
prevent it from rusting. Then, the knife is knives.
kept covered in a suspension box to prevent ○ They have a sharp cutting edge that can cut
the settling of dust and grit on its surface, 2-4 μm thick sections with ease.
causing damage to the knife edge. The knife ○ Some microtome manufacturers have also
should not be allowed to rest on its sides now incorporated a disposable blade holder
since this may also damage the cutting in place of a knife holder.
edge. ○ Magnetic knives are also now available that
○ Pressure during the first stropping strokes can attach to some blade holders and are
should be quite light, since the natural particularly suitable for use in the cryostat.
compressibility of the leather is what actually ● GLASS KNIVES
does the work. Only a gentle pressure ○ Glass knives are generally used for
should be applied while the knife is held trimming and semi-thin sectioning of
steady on the strop, since a slip may cut the tissue blocks for electron microscopy.
strop and damage the cutting edge. ○ They are prepared from commercially
○ Speed in stropping should be avoided. One available 40 x 2.5 cm. plate glass strips
full second should be allowed for each stroke that have been washed with detergent,
to avoid injury to the strop and the knife. rinsed in distilled water and alcohol, and
○ Leather strops are usually dry and require dried with lint-free paper.
oiling before they are used. Strops are ○ Cleaned strips are clamped into a knife
usually treated with vegetable oil (e.g. castor maker, scored with a tungsten carbide
oil) applied into the back of the strop, NOT wheel, cracked to form 25 x 25 mm square
the surface. The strop should not be used for pieces, and further broken into two triangular
at least 24-48 hours after treatment. Too shaped knives using even pressure.
much oil will make the stropping surface ○ Glass knives should be prepared and stored
slippery and will render the procedure in dust-free boxes with lids, just before use,
unsatisfactory. To remove excessive oil from to avoid contamination.
the strop, its surface is scraped with a blunt ● DIAMOND KNIVES
instrument, e.g. the back of the knife. ○ Diamond knives are used to cut any type of
○ Mineral oil is not recommended and resin block for electron microscopy.
should NEVER come in contact with a ○ When supplied by manufacturers, they are
strop since it will tend to blister and destroy already mounted in a metal block designed
the leather. One drop of mineral oil will spoil to fit directly into the knife holder of the
the polish of that area, and produce a ultrathin microtome.
permanent blemish on the strop. ○ Diamond knives are brittle and expensive,
○ Stropping surfaces should be firm and not but very durable, and the cutting edge must
loose, to prevent the turning of the knife's be kept clean to make it cut longer and to
edge. Hence, strops are usually mounted on avoid damage during sectioning.
a wooden canvass and covered with a flat OTHER EQUIPMENT
pad to prevent them from sagging. ● Waterbath - The thermostatically controlled type is
preferable, but if this is unavailable, water from a hot
● Sections are cut between 4-6μ in thickness for routine ● Debris adhering to upper or lower edges of the block,
histologic procedures, after the block has been fixed or the back of the blade, can make it difficult to obtain
and secured to the block holder. cohesive ribbons and cause the ribbon to lift off the
● The micrometer gauge is set to the required thickness blade on the upstroke.
and the knife is positioned in such a way that the ○ If debris is present clear it away, re-chill the
center of the blade is in line with the block and the block and start again.
knife has been securely clamped in place . ● Sections are removed in ribbons of ten to allow easy
● The actual thickness of the first couple of sections in a location of serial sections. The sections are then
ribbon may be thicker than indicated because of floated out on a water bath set at 45-50°C,
thermal expansion when cutting a cold paraffin block. approximately 6-10°C lower than the melting point of
● Using the microtome handle, try to cut in a slow and the wax used for embedding the tissue.
consistent manner - don’t start and stop while the ● This is to flatten the sections and prepare them for
blade is cutting a block as this may produce horizontal mounting onto the slides.
lines across the block and the sections (and very ● Folds and creases may be removed by stretching the
slight changes in thickness). sections gently with a pair of dissecting needles or
● Sectioning is generally improved when the specimen forceps.
and the wax are well matched in hardness. It is for ● Bubbles may be teased out from beneath the sections
this reason that most paraffin blocks must be cold by means of the same needle.
when sections are cut. ● Sections should not be left on the water bath for a
● The actual method used to chill the block is long time (30 seconds will be enough) to avoid undue
● important. expansion and distortion of tissue.
● COLD WAX ● A section is selected for staining and picked up onto a
○ provides better support for the harder clean slide in a vertical position.
elements in a specimen allowing thinner ● The slide is immersed in the water bath in a near
sections to be obtained. vertical position as close as possible to the section.
○ Place the blocks on a cold plate or a cold ● When the slide touches the section, it is lifted
wet surface for a few minutes (such as the vertically out of water and drained.
surface of melting ice). ● Sections may also be flattened out by placing them on
○ Water penetrates a small distance into the a slide which has been flooded with 20% alcohol,
block face, swelling tissues and making them producing convection currents which will serve to
more amenable to cutting. remove the creases in the tissue within a few
■ This is particularly important to seconds.
over-dehydrated, dry or crumbly ● Sections are very easily damaged when dislodging
tissues. wrinkles or bubbles with brush or forceps.
● Incomplete sections are discarded. Complete ribbons ● Examine each section as it floats on the water surface
are picked up at once with a camel hair brush, or a as imperfections can be readily seen.
pair of forceps. ● Leave the section on the water surface just long
● Tissues which tend to crumble (e.g. blood clots, bone enough for it to flatten.
marrow) or do not form a smooth flat surface can be ● Overexpansion can spoil the morphology in
sectioned with ease, by exhaling gently into the block susceptible sections.
surface while the section is being cut slowly, to reduce ● FLOATATION
the effects of static electricity. ○ should expand the section to its original
● Successive sections will usually stick edge-to-edge dimensions and ensure that it is completely
due to local pressure with each cutting stroke, thereby flat. The temperature will need to be 5 - 9
forming a ribbon. ̊C below the melting point of the wax.
● Generally a slow, uniform cutting stroke produces the Make sure the water is clean and free of
best results and the least compression. Do not stop bubbles.
and restart during a cutting stroke as this will produce ■ To promote efficient drainage and to
bands of different thickness across the section. prevent the section from slipping
● The practice of gently breathing on the face of a down the slide, remove slides
chilled block immediately before cutting each section, vertically from the water.
is common in some laboratories. ■ After floating and mounting the
● The application of warm, moist breath tends to make paraffin sections from each block,
sections more cohesive, but it also causes thermal use lint-free Kleenex or Kim-wipe to
expansion thus making the section thicker. thoroughly wipe clean the surface
of the water and the edges of the
flotation bath to prevent floaters or directly into sterile microfuge tubes ready for nucleic
cross-contamination. acid extraction.
○ The mounted section is then placed in a 70 ● The extraction buffer can be added directly to the
degrees Celsius paraffin oven for 20 minutes microfuge tube in order to preserve the molecular
or until water droplets are no longer visible integrity of the sample.
on the slides. ● When cutting sections for DNA or RNA extraction, all
○ Besides the paraffin oven which is instruments and equipment must be pre-cleaned and
maintained at a temperature of 2-5°C above wiped down with RNAse-away before and between
the melting point of the paraffin used, each specimen.
small thermostatically controlled incubators ● Gloves must be worn. Molecular grade water must be
may be used, regulated at 37°C, and at used for floating sections for RNA extraction.
45-55°C, for enzyme digestion, chemical FAULTS OBSERVED DURING SECTIONING
extraction, metallic impregnation and These are the 23 common faults seen during sectioning:
enzyme localization techniques. 1. Section failed to form ribbons
○ Hot plates are not recommended because 2. Sections roll up on cutting so that they adhere and get
they can cause overheating and there is a broken against the knife edge
risk of dust falling onto the section during the 3. Ribbon is curved, crooked, or uneven instead of
drying period. straight
4. Sections are compressed, wrinkled or jammed
○ Excessive heat can cause droplets of water
5. Sections are squashed (width of each section is less
underneath a section to boil and this will than that of the block)
cause damage. 6. A hole is formed in the section
○ Proper drying ensures that sections are 7. Sections of unequal thickness are produced
completely dehydrated, free of heat damage, 8. Sections adhere to the knife or other parts of the
flat and unlikely to lift during staining. microtome
9. Ribbon is split or lengthwise vertical scratches are
○ Drain excess water from beneath the section
seen on sections
before drying. Dry sections for between 5 10. Sections are lifted from the knife on upstrokes
and 30 minutes. 11. Resistance is felt on the lower part of the section
○ Some delicate specimens will produce best during cutting
results when dried at 37 degrees C for a 12. Horizontal or parallel lines or furrows across the
longer time (severalhours to overnight). section ("chatters") are seen
● Metal racks with 25-slide divisions are used to store 13. Section cut is sometimes thin, sometimes thick
14. Knife makes a hard metallic scraping or ringing sound
the mounted sections during the drying process which
on backstroke, when section is cut
usually takes about 5 minutes in the heated oven. 15. Frozen tissue crumbles and comes off the block
Once dry, the whole rack of slides can be taken for holder when cut
manual batch staining or placed on an automated 16. Frozen tissue chips into fragments when cut
staining machine. Staining of serial sections should 17. Ribbons are crooked
never be attempted unless they are completely dried. 18. Sections are too thick
Overheating should be avoided because it will distort 19. On trimming, tissue smells of clearing agent
20. Tissue is opaque, section cutting is difficult due to
the tissue and melt some of the structures like
presence of alcohol
collagen. 21. Tissue shrinks away from wax when trimmed
● Slides must always be grease- and dust-free and 22. On trimming, wax appears crystalline
stored and handled correctly. 23. Paraffin block, after cooling, is moist and crumbles
● If staining is to include antigen retrieval (IHC), enzyme
pretreatment (ISH), or prolonged incubation steps, Presented below are their reasons as to why it happens and
their corresponding remedies.
charged slides or an adhesive must be used.
● Some special stains, particularly those that employ PROBLEM REASONS REMEDY
alkaline reagents, can also cause sections to lift.
● Extended storage (usually more than 3 days) of Surface and
unstained formalin-fixed paraffin embedded slides Edges are not Re-trim the block
parallel
should be avoided as this may result in the loss of 1.
antigens. SECTION Horizontal
● While not established, vacuum sealing and FAILED TO surface Re-adjust and
refrigeration may help preserve some unstable FORM of the block is not re-
antigens. RIBBONS parallel to the orient the block
● For nucleic acid extraction sections, allow the knife
individual sections to roll up naturally and place them
● Casting and Hardening of Celloidin distinctive cut is made on one corner for
○ When infiltration is complete, the block has orientation.
to be cast and hardened. ○ The back is trimmed flat and the block is
○ Paper boats have the advantage in that they placed into 70% ethanol until ready to be
may be cut off if the paper does not peel sectioned.
away easily. ○ The blocks are trimmed in the same manner
○ Some thick celloidin is poured into the as in paraffin blocks, but they do not require
bottom of a boat, then the tissue is oriented, hardening by chilling before cutting.
and more celloidin poured in to cover the ● Celloidin Sectioning
tissue. ○ Tissues embedded in celloidin are usually
○ After filling the boat with thick celloidin, it is sectioned with a sliding microtome
placed under a bell jar with a base that ■ Block is mounted to a holding
ensures air is excluded. platform facing upwards and some
○ Each day the top of the jar is lifted a little thick celloidin is placed onto the
for a few minutes so that evaporated holder, positioning the block so that
ethanol-ether can escape. it will meet the knife as wanted, and
○ More solvent will evaporate from the block the assembly is left until the
each day, evaporating the solvent off slowly attachment is firm.
and ensuring that the celloidin thickens ○ The knife is held at a significant slant so that
and hardens evenly throughout the most of the blade edge is used during the
tissue. cutting stroke, and is quite long, often in
■ Evaporating it too fast will result in excess of 25 cm.
the outside of the block ○ The face of the block is lubricated with 70%
becoming hard while the inside ethanol and the knife drawn across the top of
is still soft. the block at a strong slant, shaving off a
● The slow method of hardening the block allows the section, which is immediately removed and
increasing concentration of celloidin to get into the placed in 70% ethanol.
block and give additional support to the tissue. ○ The surface of the block is then re-lubricated
● The embedding process takes time and is complete for the next cutting stroke.
when the block is sufficiently hard, often judged by ● To avoid dehydration and shrinkage, section cutting is
pressing it with a finger nail without leaving an usually done wet, which means that the block is
impression. Hardening of the celloidin block may be lubricated with a fluid, usually 60-70% ethanol, and is
hastened by placing a small open container of not allowed to dry out.
chloroform under the bell jar. ○ This makes section cutting somewhat messy
○ The chloroform will saturate the atmosphere and quite a bit slower than the dry sectioning
and harden the celloidin without further used with paraffin.
evaporation. ● Celloidin sections do not come off in ribbons and tend
● Celloidin Hardening & Shrinkage to roll up during cutting, and moistening the block and
○ As the block hardens, the celloidin will section with alcohol by means of a camel hair brush
shrink. will serve to flatten the sections on the knife.
○ If at any time the celloidin shrinks enough to ● After cutting the sections, they are immediately
expose the tissue, more celloidin should be collected into 70% alcohol instead of being mounted
poured in to cover it. on to glass slides.
○ At the end of the hardening process there ● They are then stored in the same solution in jars with
should be sufficient celloidin to allow for tightly fitting lids, and finally mounted on to slides after
trimming the back of the block flat so that it they have been stained.
may act as a base for glueing the block to a ● They are usually stained free floating and put on
wooden holder for sectioning. slides at the same time as the coverslip is applied.
○ Once hardened, the block is removed from ● This makes it difficult to prepare serial sections as
the paper boat, preferably by peeling, but it each section must be stored in individual,
can be cut away with a sharp blade if appropriately numbered containers.
necessary. ● Small batches of 5 or more sections may be
● Celloidin Block Trimming prepared, stored in the same alcohol that was used
○ Block is then trimmed, leaving about 3-5 mm for lubrication, and never be allowed to become dry.
of celloidin all around the tissue, then a