You are on page 1of 7

Procedure for Ferrite Testing

FERRITE CONTENT PROCEDURE

Document No. TP-FT

Issue 1

Issue Date: 10.06.2020

WRITTEN/APROVED BY: DATE: SIGNED:

Shaun Martin 10.06.2020


PCN Level 3 (302525)

AUTHORISED BY: DATE: SIGNED:

Steve Gouldson QHSE Manager


10.06.2020 Steve Gouldson

CLIENT REVIEW: DATE: SIGNED:

TP-FT Issue 1
Page 1 of 7
Uncontrolled when Printed
Procedure for Ferrite Testing

AMENDMENT RECORD SHEET


PAGE ISSUE DETAILS OF CHANGE DATE AUTHORISED
ALL 1 First Issue 10.06.2020 S Martin

TP-FT Issue 1
Page 2 of 7
Uncontrolled when Printed
Procedure for Ferrite Testing

CONTENTS

1. SCOPE
2. PERSONNEL
3. EQUIPMENT
4. GENERAL REQUIREMENTS
5. SAFETY PRECAUTIONS
6. SET UP PROCEDURE/CALIBRATION
7. SURFACE CONDITION
8. OPERATING PROCEDURE
9. TESTING
10. DETECTION LIMITS
11. REPORTING

TP-FT Issue 1
Page 3 of 7
Uncontrolled when Printed
Procedure for Ferrite Testing

1.0 SCOPE

This procedure covers the calibration and operating methods for the Fisher Feritscope MP30,
used for the determination of Ferrite Content of Stainless Steels, Clad materials and
Duplex Stainless Steels.

Determination of the ferrite content is based on a magnetic method, which makes use of the
difference in magnetic permeability, as a function of the percentage ferrite present.

2.0 PERSONNEL

All personnel shall have received ‘in house training’ in the use of the equipment, due to
there being no Nationally recognised qualification scheme.

3.0 EQUIPMENT

The following equipment will be used:


• Fisher Feritscope
• Operating Manual/Instructions
• General Purpose probe EGAB 1.3-FE
• Reference Calibration blocks

4.0 GENERAL REQUIREMENTS

All work shall be carried out to this procedure and in accordance with the Health and Safety at
Work Act 1974, and any local requirements.

5.0 SAFETY PRECAUTIONS

Personal Protective Equipment is to be worn as specified by the Company Health & Safety
Policy.

6.0 SET UP PROCEDURE/CALIBRATION

Calibrate the equipment by verifying the instruments readout against the supplied sets of
reference Standards figures.

If after calibration the readout differs by more than 5%, the instrument must be recalibrated
(see manufacturers handbook).

TP-FT Issue 1
Page 4 of 7
Uncontrolled when Printed
Procedure for Ferrite Testing

7.0 SURFACE CONDITION


Ferrite checks shall be performed only in areas which are clean and free from paint, rust or
other surface contaminates which may reduce the accuracy of the test, surface profiles
must also be taken into account.

8.0 OPERATING PROCEDURE

This test procedure shall be followed for all testing conducted using the Fisher
Feritscope MP30, after the above set up procedure. All parts under examination shall be
within the range of -10°c to 40°c.

9.0 TESTING

Connect the probe to the top of the unit, and secure by rotating the knurled knob. Press
the ON/OFF button once and wait for the unit to complete its self -test. (always
ensure the probe is connected prior to turn on)

Select application number 1 (APPL No), the unit is now ready for use.

Place the probe vertically on to the surface of the item to be tested.

After the measurement has been taken, an audible ‘beep’ will be heard, lift the probe
from the surface (minimum of 10mm)

Take another reading by repeating the above step until the required amount of
readings have been taken.

The data stored in the instrument memory can be displayed at any time by pressing
the ‘BATCH-RES’ key.

(Erroneous measurements can be deleted by pressing the ‘DEL’ key after the
measurement has been taken)

Information such as the mean value, standard deviation etc… can be viewed by
pressing the ‘BATCH-RES’ key a number of times as listed below

1 MEAN Mean value of stored data


2 MIN Lowest value obtained
3 MAX Highest value obtained
4 S Standard deviation of stored data
5 N Number of samples in batch of data.

A new batch of data is automatically started after the results have been reviewed
(Using the ‘FINAL-RES’ key shows results of all batches at once. The memory is
automatically cleared when a new measurement is taken after viewing these results)

The instrument is now ready for measuring a different area.

When all testing is completed and logged, press the ‘ON/OFF’ key once the unit will
now turn off and all memorised data will be cleared.
TP-FT Issue 1
Page 5 of 7
Uncontrolled when Printed
Procedure for Ferrite Testing

10.0 DETECTION LIMITS

A correction factor should be used on the displayed value if one or more of the following is
violated.
(correction factors can be found in the manufacturers handbook)

1 Curvature of surface exceeds <50mm diameter.


2 Limited specimen area <2mm diameter.
3 Clad ferritic steels/ferromagnetic substrate <2mm clad thickness.
4 Clad austenitic steels <1.5mm clad thickness.

11.0 REPORTING

All results shall be recorded on a standard Alliance report format and as a minimum
shall contain the following:
• Customer
• Project
• Date of test
• Items under examination
• Procedure reference
• Material specification
• Equipment details
• Name and signature of technician

See report sample below

TP-FT Issue 1
Page 6 of 7
Uncontrolled when Printed
Procedure for Ferrite Testing

TP-FT Issue 1
Page 7 of 7
Uncontrolled when Printed

You might also like