Professional Documents
Culture Documents
net/publication/329254652
CITATIONS READS
0 1,320
4 authors, including:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Regis Dias on 28 April 2021.
Fabrício Gonçalves Torres1, Felipe Santiago Apolinário1, Regis Renato Dias1 and
Tiago Lopes Santos1
1
Instituto de Pesquisas Tecnológicas do Estado de São Paulo - IPT, Brazil
fabrigt@ipt.br
1 Introduction
characteristic. On the other hand, biphasic defibrillation alternates the direction of the
pulses, completing one cycle in approximately 12 milliseconds.
Like any electronic equipment, a defibrillator can cause failures over time, includ-
ing applying energy levels with errors greater than the tolerances determined by the
standards, such as ABNT NBR IEC 60601 (2016). Therefore, these devices must be
tested regularly by means of a defibrillator analyzer. In turn, a defibrillator analyzer
must be calibrated periodically, by means of standards that can guarantee the chain of
metrological traceability of the measurements.
According to [2], to comply with the standards ABNT NBR IEC 60601-2-4: 2005
and ANSI / AAMI DF80: 2003 the energy measurement of a defibrillator analyzer
must have a tolerance of ± 5 % of the measured energy or ± 1 J, whichever is greater,
and therefore it is recommended that the measuring system used to perform the cali-
bration has lower uncertainties than the recommended tolerances.
In the literature, several calibration methodologies for defibrillator analyzers can
be found. In [3], it is recommended that the calibration of the analyzer be performed
on each block of the equipment separately: internal attenuator, filter and A/D convert-
er block. However, in order to perform it, it is necessary to open the equipment and
identify its main parts. This practice becomes difficult to perform by calibration la-
boratories since there is a need to have information about equipment that is not nor-
mally available by manufacturers. Another relevant issue is about the fact of the cali-
bration being performed under different circumstances of the use of the item under
calibration may present risks in the reliability of the results obtained. Finally, we don't
recommend opening physically any instrument to calibrate it.
In [4], there are at least two ways of calibrating the energy range of a defibrillator
analyzer: by simultaneously measuring the current and voltage applied by a defibrilla-
tor, or by measuring the resistance of the load and the voltage applied. The second
case was the method adopted in this work. According to [4], the difference between
the two methods is of the order of 0.05 %, a negligible value as compared with the
other uncertainty components of the calibration set.
3
The method adopted for this work consisted in measuring the resistance of the de-
fibrillator analyzer by means of a high accuracy resistance meter and in the plot of the
voltage curve as a function of time by means of a high voltage probe coupled to a
digital oscilloscope. A commercial defibrillator of the monopolar type was used as a
transfer device.
From the plotted curve, the square voltage values were integrated, according to
equation 1.
∆
=
= ∑ × (1)
Where,
Vt = peak voltage measured by the probe coupled to an oscilloscope
R = resistance of the defibrillator analyzer
Ti and TF = initial and final time of the plotted curve (record lenght)
∆T = quantization interval of the plotted curve
The systematic and random error values obtained in the characterization was as-
signed as components of uncertainty and included in the final calibration uncertainty
of the defibrillator analyzer.
3 Estimation of Uncertainty
The table 1 summarizes the distribution assignments made for each input quantity of
the measurement model.
Where,
uE = combined standard uncertainty of the energy measurement
uT = standard uncertainty due to probe and oscilloscope set for time interval
uR = standard uncertainty due to load resistance
uV = standard uncertainty due to probe nonlinearity and probe and oscilloscope set
for measuring voltage
The uR uncertainty was obtained by measuring the internal resistance of the defib-
rillator analyzer under calibration using a digital multimeter. The uncertainty compo-
nents taken into consideration for the uR calculation were: uncertainty and resolution
of the standard multimeter and variability from four resistance readings.
For the uT uncertainty, the component was considered due to time base of the oscil-
loscope on the sampling rate of the oscilloscope (1 MS/s) resulting in the time interval
of 1 µs.
The uncertainty uV was calculated by means of the following uncertainty compo-
nents: nonlinearity (described in item 2.1) and characterization of the probe and oscil-
loscope set (described in item 2.2).
6
The expanded uncertainty was obtained by multiplying the combined standard un-
certainty with the coverage factor to a confidence level of approximately 95 %.
4 Results
The results of the calibration of the defibrillator analyzer are shown in figure 3. For
each point of the graph, there was its respective expanded uncertainty of the meas-
urement represented by means of the vertical bars. The dotted lines of red color repre-
sent the tolerance of ± 5 % of the measured energy or ± 1 J, whichever was greater,
complied with the standards ABNT NBR IEC 60601-2-4: 2005 and ANSI / AAMI
DF80: 2003.
Table 2 contains the most detailed information on calibrated points. For each point,
the following information was reported: measurement error, expanded uncertainty
(U), coverage factor (k) and degrees of freedom (νeff).
According to table 2, it was verified that the expanded uncertainty of the measure-
ment was proportionally higher for the smaller values. This feature was due to the
resolution of the defibrillator analyzer under calibration, which ends up contributing
significantly to the low energy values. For the higher values, the uncertainty of the
standards set is the most significant portion for the expanded uncertainty.
5 Conclusion
This work presented a method of calibration of the energy range of defibrillator ana-
lyzers that could be feasible its application to calibration laboratories, and that al-
lowed obtaining, in its results, low enough uncertainties to meet the determinations of
electromedical standards.
According to the results presented in this work, it can be verified that the proposed
calibration method was adequate, since the uncertainty obtained resulted in a value 10
times lower than the tolerance for this equipment. Although it is not the scope of the
work, it can also be verified that the method can be used to calibrate any type of
waveform (monophasic or biphasic, for example), since the developed system is ver-
satile enough to perform the measurement of the energy for any waveform.
Perhaps the only disadvantage of the proposed calibration method could be the
need to use a defibrillator as the transfer device. However, it is verified that it is the
most reliable means to carry out the calibration, since it uses a signal with the same
characteristics whose defibrillator analyzer must be able to measure with the best
accuracy. Calibrating the equipment by applying another type of waveform may mask
some type of error that is detectable only with the waveform of a defibrillator.
References
1. Fornazier, C., Melchior, S. C., Buss, G., Trindade, E., Pereira, A. A., Barbieri, D. X., Sil-
va, J. E. L., Holsbach, L. R., Perlato, M. T., Glowacki, L. A., Almeida, A. P., Cruz, C. J.
F.: Abordagem de Vigilância Sanitária de Produtos para Saúde Comercializados no Brasil:
Desfibrilador Externo. Boletim Informativo de Tecnovigilância, Nº 01, Brasília, (2011).
2. Moraes, J. C. T. B., Oliveira, C. A.: Especificações mínimas aceitáveis de um analisador
de desfibriladores e cardioversores. Brazilian Journal of Biomedical Engineering, Volume
28, Number 2, pp. 179-189 (2012).
3. Oliveira, C. A., Moraes, J. C. T. B.: Metodologia de avaliação metrológica da medição de
energia para analisadores de desfibriladores e cardioversores. Brazilian Journal of Biome-
dical Engineering, Volume 28, Number 3, pp. 205-216 (2012).
4. Wei, Z., Haiming, S., Feipeng, L., Bo, L., Jiandong, D.: Energy calibration technique of
defibrillator analyzer. IEEE 12th International Conference on Electronic Measurement &
Instruments (2015).