Professional Documents
Culture Documents
Referenced documents
Elekta does not supply all the documents that we refer to in this document with the equipment.
Elekta reserves the right to make the decision on which of the documents we supply with the
equipment.
Contact information
0 1 20
TECHNICAL HELP
Contact your local Elekta representative for technical help
Table of Contents
Chapter Title Page
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
— Blank page —
Table 2.3 List of abbreviations and acronyms for MLCi and MLCi2 . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Table 3.8 LED Indicators for the dosimetry PCB in usual mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Table 3.14 Control items for the temperature and pressure subsystem . . . . . . . . . . . . . . . . . . . . . . . . . 63
Table 3.16 Control and monitor control items of the dosimetry system . . . . . . . . . . . . . . . . . . . . . . . . 65
Table 3.17 Control and monitor control PCBs of the dosimetry system . . . . . . . . . . . . . . . . . . . . . . . . . 66
— Blank page —
The IEC safety standards compliance gives verification that the equipment is in compliance with the
applicable IEC safety standards for medical devices.
If necessary, contact Elekta Limited for more information on the applicable safety standards.
The IEC classification of the product identifies the degree of protection that the equipment gives.
WARNING 1.1
Do not use this equipment unless the electrical power supply is connected to a grounded
earth to prevent an electrical shock. If you ignore this warning, you can cause fatal injury.
The conventions for warnings, cautions, and notes give information about the class and the level of
the risk in the warnings, cautions, and notes in the text.
WARNING 1.2
A warning is information about procedures, tasks, or conditions, where the result can be
fatal or serious injury, injury, or clinical mistreatment, if you do not obey the instructions.
CAUTION 1.1
A caution is information about procedures, tasks, or conditions, where the result can be
damage to the equipment, data, or the environment, if you do not obey the instructions.
The labels found on Elekta equipment give information about the necessary precautions that apply to
the operation of the equipment.
Hot surface warning This label is a warning that the item can be
hot to touch.
Do not sit caution This label is a caution not to sit on the part.
Safety label - Follow This label tells you that there is a safety risk
instructions for use and you must refer to the Instructions for
Use before you continue with the task or
operation of the equipment.
General label - Follow This label recommends that you refer to the
operating instructions Instructions for Use before you continue
with the task or operation of the equipment.
The safety precautions for the compatibility of the equipment are the recommended precautions for
the operation of the equipment with parts or accessories not approved by Elekta.
WARNING 1.3
Do not use parts or accessories that are not approved by Elekta. The result can be incorrect
radiation delivery and other safety risks. If you ignore this warning, it can cause fatal injury
or clinical mistreatment.
Only operate the equipment with Elekta supplied or approved, compatible equipment or parts.
Contact Elekta for information about the compatibility of other equipment or parts.
Do not use accessories, transducers, and cables that are not specified by Elekta. They can have an
effect on the performance of the electromagnetic compatibility (EMC), which can increase the
emissions or decrease the immunity of the equipment.
The safety precautions for work with heavy parts are the necessary safety precautions to know to
prevent injury.
WARNING 1.4
Do not lift or move heavy parts unless you obey the applicable Health and Safety regulations
and procedures. If you ignore this warning, it can cause serious injury.
The safety precautions for work with VDUs are the necessary safety precautions to know to prevent
damage to your health.
WARNING 1.5
Do not use a visual display unit (VDU) unless you obey the applicable Health and Safety
regulations and procedures. If you ignore this warning, it can cause injury.
The safety precautions for operation of the equipment are the precautions to prevent incorrect
radiation delivery and other malfunctions, which can cause fatal injury.
WARNING 1.6
Do not operate the equipment unless the QA checks are satisfactorily completed, or if you
know, or think that the equipment is defective or incorrectly calibrated. If you ignore this
warning, you can cause fatal injury, clinical mistreatment, and damage to the equipment.
Only operate the equipment if you are a qualified person with the applicable authority. Such
operation must apply best practice for clinical treatment and obey the laws and regulations in the
jurisdiction in which the equipment is installed.
Before you operate the equipment, make sure that:
• There are no defective parts.
• The equipment calibration is correct.
• The applicable QA checks are satisfactorily completed.
• The user routine checks are satisfactorily completed.
• The planned maintenance schedule is satisfactorily completed.
Refer to the Instructions for Use for information on the user maintenance and the planned
maintenance schedule.
It is the responsibility of the authority that has the control of the equipment to set up a QA
program that gives safe and satisfactory system operation.
1.6.5 Safety precautions for work with safety devices during service
operation
The safety precautions for work with safety devices during service operation are the necessary safety
precautions to know to prevent incorrect radiation delivery and other safety risks, which can cause
fatal injury or clinical mistreatment.
WARNING 1.7
Do not remove, change, or override the switches, interlocks, or other safety devices unless
special instructions in this document tell you to. The result can be incorrect radiation
delivery and other safety risks. When the work is completed, set all the safety devices again
and do a test of them. If you ignore this warning, it can cause fatal injury or clinical
mistreatment.
1.6.6 Safety precautions for work with electrical and mechanical parts
The safety precautions for work with electrical and mechanical parts are the necessary safety
precautions to know to prevent fatal injury.
WARNING 1.8
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, it can cause fatal injury.
WARNING 1.9
Do not remove the covers or cables from the equipment unless special instructions in this
document tell you to. The covers give access to high voltage electric parts. Always install the
covers and cables again before you use the equipment. If you ignore this warning, it can
cause fatal injury.
WARNING 1.10
Do not remove the covers from the equipment unless special instructions in this document
tell you to. The covers give access to parts that move. Always install the covers again before
you use the equipment. If you ignore this warning, it can cause fatal injury.
WARNING 1.11
Be careful when the power supply switch of the equipment is put in the off position. There
continues to be a risk of high voltage electric shock. Some of the customer interfaces can
continue to supply electrical power to the electrical terminals. If you ignore this warning, it
can cause fatal injury.
Only operate the equipment in rooms that are in compliance with the applicable electrical safety
laws and regulations for this type of equipment.
Before you start the maintenance and repair work, or start to clean the equipment, isolate the
electrical power supply and prevent unwanted movement of the equipment.
1.6.7 Safety precautions for work with devices that are sensitive to an
electrostatic discharge
The safety precautions for work with devices that are sensitive to an electrostatic discharge are the
necessary safety precautions to know to prevent an electrostatic discharge.
CAUTION 1.2
Do not touch the PCBs and other electronic parts unless you wear a grounded antistatic wrist
strap. The result can be an electrostatic discharge and a malfunction. If you ignore this
caution, you can cause damage to the equipment.
CAUTION 1.3
Do not touch the parts identified with the ESD caution label. Always obey the ESD safety
precautions when you do work on these parts. If you ignore this caution, you can cause
damage to the equipment.
Semiconductor devices, integrated circuits, and the parts that contain them, are sensitive to ESD.
These devices are known as electrostatic discharge sensitive devices. ESD can cause damage to
these devices and cause them to have a malfunction immediately or some time after the
electrostatic discharge. To prevent the ESD damage to these devices, always wear a grounded
antistatic wrist strap, and obey all the ESD precautions and procedures.
Elekta recommends that all users receive the training about ESD. The training must include the
ESD caution label and these ESD safety precautions.
The safety precautions for the occurrence of fire or explosion are the necessary safety precautions to
know to prevent fatal injury or damage to the equipment.
WARNING 1.12
Do not use the equipment in rooms that contain explosive or flammable gases. High levels of
oxygen, and some anesthetic gases and disinfectant sprays, are explosive or flammable. If
you ignore this warning, it can cause fatal injury and damage to the equipment.
The safety precautions for the electromagnetic compatibility are the necessary safety precautions to
know to prevent an unwanted effect on the operation of the equipment by the radio frequency
emissions.
WARNING 1.13
Do not put portable or mobile radio frequency (RF) communication devices, or other
electrical devices, near the equipment. If you ignore this warning, you can cause fatal injury,
clinical mistreatment, and damage to the equipment.
The Elekta products are in compliance with applicable EMC emission standards. It is possible that
the electromagnetic emissions from the portable and mobile RF communication devices, or other
electrical devices, are more than is permitted by the EMC standards. Such electromagnetic
emissions can have an unwanted effect on the operation of the medical electrical equipment.
WARNING 1.14
Do not use the accessories, transducers, and cables unless they are specified by Elekta. If you
ignore this warning, you can cause fatal injury and clinical mistreatment.
Special precautions for EMC emissions are necessary. See the related documentation on how to do
the installation and setting to work procedures, and to operate the equipment.
See the section Technical Description in the Instructions for Use for information about EMC.
The safety precautions for work with UV devices are the necessary safety precautions to know to
prevent serious injury to your eyes.
The Agility™ BLD uses ultraviolet light. This ultraviolet light is contained by the Agility™ BLD
covers and crosswires screen.
WARNING 1.15
Do not look into the ultraviolet light without protective glasses when the Agility™ BLD covers
or crosswires screen are removed. If you ignore this warning, it can cause serious injury to
your eyes.
WARNING 1.16
Do not look into the laser beams. The laser beams can cause damage to your eyes. If you
ignore this warning, the laser beams can cause serious injury to your eyes.
Before you remove the covers or crosswires screen, switch off the ultraviolet light or use protective
glasses.
Protective glasses must have the UV-A specification OD 6+ @ 190 - 532 nm.
If the Agility™ BLD covers or crosswires screen are damaged, switch off the ultraviolet light and
contact your Elekta representative.
The safety precautions for work in areas of induced radioactivity are the necessary safety precautions
to know to prevent the related risks from induced radioactivity.
When you operate or do work on the equipment, Elekta recommends that you use the applicable
radiation protection features, devices, systems, procedures, and accessories. Always obey the
applicable laws and regulations for induced radioactivity for the jurisdiction in which the
equipment is installed.
The safety precautions are applicable to:
• Maintenance, modifications, and decommissioning
• Removal of parts from the radiation head
• Measurement of the induced radioactivity
• Removal of parts with induced radioactivity.
The safety precautions for maintenance, modification, and decommissioning are the necessary safety
precautions to know to prevent excessive exposure to radioactivity, incorrect radiation delivery and
other malfunctions, which can cause fatal injury, serious injury, or clinical mistreatment.
WARNING 1.17
Do not do maintenance work, or make modifications to the equipment unless you are a
qualified person and have the applicable authority. The result can be incorrect radiation
delivery and other malfunctions. If you ignore this warning, it can cause fatal injury and
clinical mistreatment.
Only do work on, or make modifications to the equipment if you are a qualified person with the
applicable authority. Such work must apply best engineering practice and obey the laws and
regulations in the jurisdiction in which the equipment is installed.
Modifications or work that are not approved by Elekta, and that are not done by a qualified person
with the correct authority, can cause damage to the equipment and cancel the warranty.
WARNING 1.18
Do not touch the flight tube, bending magnet, or other parts near to them until a check for
induced radioactivity is completed. The risk can be excessive exposure to radioactivity. If you
ignore this warning, it can cause serious injury.
1.6.11.2 Safety precautions for the removal of parts from the radiation head
The safety precautions for the removal of parts from the radiation head are the necessary safety
precautions to know to make sure that the level of induced radioactivity is at a safe level before the
removal of parts.
After the operation of the digital accelerator at high energy, the parts in the radiation head can
become radioactive, which is a radiation hazard. Before you do work on, or remove, these parts,
use a Geiger-Müller tube or equivalent radiation detector to measure the level of induced
radioactivity. Use the applicable safety procedures to remove radioactive parts. The maximum
dose rate for the safe removal of radioactive material is 50 µSv/hr (microsievert per hour). Do not
remove the radioactive material from the treatment room until it is correctly prepared for safe
removal.
The safety precautions for the measurement of induced radioactivity are the necessary safety
precautions to know to prevent excessive exposure to radioactivity, which can cause serious injury or
damage to the environment.
Always use two radiation detectors to measure the level of induced radioactivity. Only use
radiation detectors that have an applicable calibration certificate. The period of time since the last
calibration date must be less than 12 months. Do not use the radiation detectors if the period of
time since the last calibration date is more than 12 months.
WARNING 1.19
Do not move radioactive material (by air, sea, or land) unless you obey the applicable laws
and regulations for induced radioactivity and disposal of radioactive material of the
jurisdictions through which the movement occurs. If you ignore this warning, it can cause
serious injury and damage to the environment.
1.6.11.4 Safety precautions for the removal of parts with induced radioactivity
The safety precautions for the removal of parts with induced radioactivity give the maximum safe
levels of induced radioactivity applicable in the United Kingdom.
The removal of parts with induced radioactivity must obey the laws and regulations of the
jurisdiction in which the equipment is installed. The safe levels of radiation applicable in the UK
do not apply in all jurisdictions. If there are no safe levels available for a jurisdiction, contact Elekta
for information on the safe removal of parts with induced radioactivity.
The safety precautions for cleaning and disinfection are the necessary safety precautions to know to
prevent fatal injury or damage to the equipment.
WARNING 1.20
Do not clean the equipment, or use a disinfectant spray, before you isolate the equipment
from the electrical supply. You can get a high voltage electric shock. If you ignore this
warning, it can cause fatal injury.
WARNING 1.21
Do not use a spray that is explosive or flammable. The fumes from such sprays can cause an
explosion or start a fire. If you ignore this warning, it can cause fatal injury and damage to
the equipment.
CAUTION 1.4
Do not let water and other liquids go into the equipment. The results can be electrical short
circuits and metal corrosion. If you ignore this caution, it can cause damage to the
equipment.
CAUTION 1.5
Do not use a spray to clean the medical equipment room. The spray can go into the
equipment. The results can be electrical short circuits and metal corrosion. If you ignore this
caution, it can cause damage to the equipment.
At regular intervals, it is necessary to clean the equipment and equipment room with disinfectants.
This section gives the information on the general procedures. Where special procedures and
precautions are necessary, the related documents give the information on these procedures. The
procedures for the disinfection and cleaning of the equipment and the equipment room must obey
the applicable laws and regulations of the jurisdiction.
The recommended general cleaning procedure gives the instructions for cleaning the equipment.
Clean surfaces with a mild detergent solution. Then, dry the surface with a clean, lint-free cloth.
Do not use a solution, solvent, abrasive detergent, or polish that is corrosive. If you do not know
the properties of the solution, do not use it.
The expected service life is the length of time that the regular servicing and repair procedures can
keep the machine in operation.
WARNING 1.22
The expected service life of this equipment is 15 years of normal use. Because of wear of
parts, the risk of malfunctions can increase after 10 years of operation. Before you continue
operation after 10 years, complete the regular extended life checks. If you ignore this
warning, it can cause fatal injury.
The Elekta definition of 'normal use' for this equipment is the treatment of 50 patients each day for
five days each week of the year.
The level of operation of the equipment can be different to ‘normal use’. This has an effect on the
level of the risk for a malfunction to occur. Elekta can help the user with information on the
necessary precautions to keep the equipment for continued operation, or to replace it.
Before you continue to operate the equipment for more than 10 years, contact Elekta for
information on the necessary regular extended life checks.
The procedure for the disposal of the equipment at end of life to prevent damage to health and the
environment caused by radioactive and hazardous material.
WARNING 1.23
Do not move or discard radioactive and hazardous material unless you are a qualified person
and have an applicable license. Obey the laws and regulations of the jurisdiction for disposal
and movement of these materials. If you ignore this warning, it can cause fatal injury and
damage to the environment.
End of life (EOL) disposal refers to the procedure to remove and discard equipment, or parts of the
equipment, to a condition where it will not be possible to operate the equipment for its intended
use.
Always use procedures that give the best possible protection to the environment when you remove
and discard the equipment. The procedures must obey the laws and regulations of the jurisdictions
in which you discard the equipment. Do not discard Elekta equipment through the domestic waste
systems.
Only an approved facility with an applicable license must remove and discard the equipment and
recycle the material. Do not discard material that is hazardous to health and the environment
together with other material.
Before you remove and discard equipment, contact Elekta for information.
2 Introduction
Section Title Page
— Blank page —
• Fitted as standard
— Not available
Configurations
This document gives the information for all configurations of the equipment. Not all of the
functions in this document are available in all product configurations.
It is possible that you do not have a license for all the configurations available.
The intended function of the equipment gives a description of the functions of the digital accelerator
that help you to deliver radiation to the defined target.
The Elekta range of medical digital linear accelerators is intended to assist in the delivery of
radiation to a defined target volume while sparing surrounding normal tissue and critical organs
from excess radiation. They are intended to be used for single or multiple fractions and, depending
on the options purchased, support static and dynamic beams of radiation.
The contraindications for the equipment is a list of the contraindications that can have an effect on
the operation of the digital accelerator.
There are no contraindications for this equipment.
Examples
If the examples in this document refer to patients, physicists, or hospitals by name, then they are
not the names of real persons or hospitals. If an example uses the name of a real person or hospital,
it is accidental.
Term Definition
Authorized person A skilled or instructed person, who is empowered to execute
defined work.
Qualified person A person recognised by a competent authority as having the
requisite knowledge and training to perform particular duties.
5
2
6
7
Note: The A and B positions in Figure 2.1 are correct with the gantry at 0° only. The A and B positions
rotate with the gantry. Therefore, with the gantry rotated 180°, A and B are opposite.
— Blank page —
Table 2.3 List of abbreviations and acronyms for MLCi and MLCi2
Abbreviation Definition
ADC Analog-to-digital-converter
AI Angiographic image
ASU Automatic setup
BLD Beam limiting device (generic term for Asymmetric
Diaphragms, MLCi, MLCi2, and Beam Modulator™)
BUD Backup diaphragm
CAN Controller area network
CAT Customer acceptance test
CCU Camera control unit
CCV Camera condition voltage
CMM Corrective maintenance manual
DAC Digital-to-analog-converter
DCB Detector control board
DPT Dose pulse transmitter
DROT Diaphragm rotation
FCO Field change order
FKP Function keypad
FPGA Field programmable gate array
FPLA Field programmable logic array
FPLS Field programmable logic sequencer
G-T Axis Indication of an axis at a right angle to the gantry in
the general direction from gun to target
HCU Head control unit
HHC Handheld controller
HME Hardware movement enable
HT High tension (high voltage
ICD Inertia compensation distance
I/O Input/output
IEC International electrotechnical commission
IMRT Intensity modulated radiation therapy
ISO The international organization for standardization
ISO9001 International standard for quality systems
Abbreviation Definition
LED Light emitting diode
MLCi/MLCi2 MultiLeaf collimator 1/MultiLeaf collimator 2
MTU Multiplex terminal unit
MU Monitor unit
mV Megavolts
N/A Not applicable
NEMA National electrical manufacturers’ association (USA)
NPN Negative/positive/negative (transistor)
PCB Printed circuit board
PI Portal image
PRF Pulse repetition frequency
PSU Power supply unit
PSS Patient support system
PVC Polyvinylchloride
QA Quality assurance
RLB Relay B
RT Radiation therapy
RHCA Radiation head control area
RHMD Radiation head motor drive
RTU Remote terminal unit
SSD Source surface distance
TCC Treatment control cabinet
TCS Treatment control system
TTL Transistor-transistor-logic
UMD Universal motor drive
VMAT Volumetric Modulated Arc Therapy
XVI X-ray volume imaging
— Blank page —
During the treatment, the dual circuits measure these parameters and send signals to the control
processor of the TCS. The control processor compares the two signals to make sure that the
indications are in the permitted limits.
If a fault or error occurs, the dosimetry system has a number of safety interlocks. The interlocks
operate to prevent the delivery of radiation.
Refer to the dosimetry system diagrams in the Digital Accelerator System Diagrams Manual for
more information.
For each 64 counts, the dosimetry PCBs send a 1 MU signal to the TCS. The TCS sends an
acknowledged signal to the dosimetry PCBs each time it receives a 1 MU signal.
This is a description of the single and multiple dose treatments with the dosimetry system.
A patient treatment can have one dose, or multiple doses, of radiation. For multiple dose
treatments, each segment of the prescription can have different qualities. For example, a segment
can be delivered at a different energy, or from a different gantry angle. Therefore each dosimetry
PCB has two counters that measure the delivered dose:
• The Segment dose counter is loaded with the segment dose value at the start of a beam
(dose period). The segment dose counter records the delivered dose in each segment of a
multiple dose treatment.
• The Total dose counter is loaded with the total dose value at the start of the treatment.
The total dose counter records the total dose for a treatment.
The two dosimetry PCBs record the total dose and the segment dose for each treatment. The TCS
continuously monitors and measures the pulse counts on the PCBs.
When the segment dose counter is at zero, the dosimetry PCB sends a segment dose complete
signal to the TCS. The TCS sends a signal which causes the radiation to pause until the next
segment is loaded. After the total dose is delivered, the TCS stops the radiation. At the end of a
treatment, the total dose will equal the sum of the segment doses.
Related Links:
Description of the DOS-A and DOS-B PCBs on page 40
Refer to the Corrective Maintenance Reference Manual for the digital accelerator for more
information about the software items.
Related Links:
List of the dosimetry status bits on page 36
List of the typical software command sequence on page 36
List of the dosimerty control bits on page 37
This is a list of the dosimetry status bits of the dosimetry status word, which gives their function and
correct status.
This is a description of the time critical software of the dosimetry system for the digital accelerator.
During treatment, some parts of the control software send signals to the TCS as they occur. The
TCS uses these signals to control the software operations.
An integrator makes the 1 MU pulses which decreases the count of the total dose counter. The
system also sends these pulses through status bit 0, as COUNT pulses, to the TCS. This updates the
value displayed on the software dose counter.
Each 1 MU pulse sets the PCB firmware condition from IDLE to WAIT ACKNOWLEDGE. It goes back
to the IDLE condition when it receives a COUNT ACKNOWLEDGED signal (WR17H) from the TCS.
If a second 1 MU pulse occurs before the COUNT ACKNOWLEDGED signal, the firmware changes
the condition to TREATMENT TERMINATE. This gives the error signal COUNT NOT
ACKNOWLEDGED.
During treatment, the TCS reads the segment dose counter to do a check of the dose rate. The dose
rate value goes to the TCS for display during treatment.
For more information about the types of ion chamber see the Elekta Oncology Products Service
Parts Catalog.
Item Description
Dose plates and servo The two dose plates have of a pair of Mylar® sheets, which have an
plates outer carbon fiber layer and ceramic insulation. Dose plate A uses
one Mylar® sheet as the dose plate for dosimetry channel A. Dose
Plate B uses the other Mylar® sheet as the polarized plate for the
opposite servo plate.
The Mylar® sheets are photo-etched to keep sensitive areas clear.
The plates have circuit paths that go to the contacts in the ceramic
insulation. The two dose plates and the servo plate are open to the
air, therefore software temperature and pressure correction are
applied.
Plate size The dose plates in the ion chamber have areas of different size at
the isocenter of the radiation beam (see Figure 3.2, Figure 3.3 and
Figure 3.4).
Polarized plates There are three polarized plates, which are supplied with the same
polarized voltage. The polarized plates give full coverage of the
collector plate area to give the electric field.
The RHCA (Area 12) supplies the -320 VDC polarized voltage to
the ion chamber through a coaxial cable. The voltage is measured
at the ion chamber. If the voltage is missing, an inhibit prevents the
operation of the digital accelerator.
10.9cm
22.2cm
A D
A B = 5.8 cm
A C = 16.5 cm
A D = 7.3 cm
F
A E = 15.3 cm
E
A F = 18.2 cm
NOT TO SCALE 001651 ©2013 Elekta Limited
This is a description of the types of printed circuit board (PCB) in the dosimetry system of the digital
accelerator and their function.
There are two dosimetry PCBs in the RHCA of the digital accelerator, DOS-A and DOS-B. These
use digital signal processors (DSP) to do their functions.
See Figure 3.5 for a block diagram of the DOS PCB.
001652
There are two dose plates in the ion chamber that operate independently, therefore PCBs DOS-A
and DOS-B operate in dosimetry channels 1 and 2. The TCS monitors the two PCBs respectively.
Each dosimetry PCB receives the current pulse output from the related dose plate in the ion
chamber. The PCB measures and records the total dose and the segment dose for each treatment.
The dosimetry PCBs have fault protection circuits. If these operate, they stop the delivery of
radiation. The fault protection circuits include a handshake between the dosimetry electronics and
the TCS for each 1 MU. If radiation pulses go into the PCB but no counts are output, then a
watchdog circuit causes the beam to stop. The keeps the independence of primary and secondary
dosimetry.
The dosimetry PCBs also gives:
• The adjustment of minor to moderate leakage to extend the life of the ion chamber.
• A continuous self-test function of each PCB so that the PCB can stay serviceable in strict
tolerances.
• A monitor to check that the input signal does not cause an incorrect dose collection.
This is a description of the operation of the system dose measurement for the DOS PCBs.
The dosimetry PCB changes the current from the ion chamber to a voltage and then a digital value.
The PCB reads a sample of the current of a typical radiation pulse approximately 200 times. This
gives an input with a very large, linear, dynamic range.
The radiation is in pulses as is the current from the ion chamber. The digital signal processor
(DSP) uses this to accurately calculate the value of the input when there is no radiation. The DSP
then subtracts this background value from the signal. This adjusts for electronic offsets and leakage
current. If the system increases to the adjustment limit, a status bit is set to prevent the operation
of the digital accelerator. You can view the status on the Dosimetry service page of the TCS.
With electronic offset leakage adjustment, only signals from the radiation pulse go through the
input step. The values from the digital input collect until they increase to the calibration value of
1/64 MU.
When the system increases to the calibration value, the DSP decreases the segment and the beam
counters and sets the input accumulator to a value of zero. The control system does not read the
lower bits of the beam counter. This keeps the current 1 MU resolution. Because the same signals
are used to decrease the two counters, the counter keeps the count as multiple segments are
delivered.
This is a description of the operations of the field programmable gate array control of the dosimetry
PCB.
The dosimetry PCB has a field programmable gate array (FPGA) IC19. This controls the DOS PCB
and also gives the dose measurement. At power up a PROM (IC26) programs the FPGA. When the
program load is successful, LED1 comes on and a CONFIG_DONE signal is applied to the HT
relay output.
The FPGA interfaces with the system bus to send the dose measurements and receive dose
controls. The FPGA also interfaces with all the DOS PCB operations. It has a 20 MHz clock and
has a watchdog timer on the 3.3 V supply rail.
The FPGA also has a check circuit. This looks at the dose count output and makes sure that the
FPGA measures and acknowledges that the dose is the equivalent to the received dose.
The FPGA has a program interface (PL7) and a standard JTAG test connectior (PL6).
IC9 is an amplifier which increases and inverts the terminated signal. The signal is input to an
analog-to-digital converter (ADC) IC27. During the change of the signal, IC5 and IC14 add a DC
offset. This has three functions:
• To move the dose signal, so that the noise is in the range of the ADC
• To adjust the DC errors in the amplifiers or ADC
• To adjust DC errors in the dose signal, caused for example, by ion chamber leakage.
The FPGA (IC19) sends a strobe signal at 1 µs intervals to the ADC to start the dose signal sample.
The ADC sends the sample data as a number of Serial bits (dose_data) to the FPGA, which does
the processing of all the subsequent signals.
The comparator (IC18) senses each dose pulse of more than 100 mV at the dosimetry PCB input
(TP84). The dose output watchdog timer circuit reads the PULSE_PRESENT signal.
The FPGA uses the dose data to calculate the quantity of DC adjustment that is necessary to set the
60 mV DC level into the ADC. An ADC (IC5) and an inverter (IC14) apply the DC adjustment to
the dose signal. If the control loop runs out of the control range because of a circuit fault or too
much of a DC error in the dose signal, the dosimetry PCB terminates the treatment through an
inhibit.
This is a description of the operation of the dose measurement of the DOS PCB.
See Figure 3.7 the dose measurement part of the DOS PCB performs the operations that follow:
• Dose integrator
• Segment dose measurement
• Excess dose measurement
• Signal out of range measurement
• Dummy dose measurement
Function Description
Dose integrator The field programmable gate array (FPGA) has a dose integrator
function that collects the dose data samples from the analog-to-
digital converter (ADC). When the value increases to a calibration
threshold, this gives a count and ADC restarts. The calibration
threshold is defined as the dose of one 1/64 of an MU. This threshold
is set during calibration of the digital accelerator; therefore 1 MU is
equal to 1 cGy at a defined position in a water phantom.
The one sample count is the dose count and occurs 64 times for each
MU. The segment and the total dose counters use the dose count to
record the current dose quantity.
Segment dose At the start of a beam (dose period), the segment dose counter is
measurement loaded with a segment dose value. For each count from the dose
integrator the counter decreases. When the count value decreases to
0, the treatment is terminated until the next segment. The output of
the segment dose counter, signal SEGMENT_DOS_COMPLETE, is
set high (logic 1).
Total dose measurement At the start of a treatment, the total dose counter is loaded with a
total dose value. The counter is then decreased by one for each count
from the dose integrator. Therefore it is necessary to have 64
segment counts before the value of the total dose counter changes by
one.
The total dose counter gives each of the 64 counts as a MU output.
At this point the output of the total dose counter, signal
SEG_DOSE_COUNT, is set low (logic 0).
DOS PCB excess dose Two counters measure the number of segment counts that occur in a
measurement time of 0.262 seconds. If, at this time, one of the two counts is more
than 307 (equivalent to 1099 MU per minute), then the system
senses a dose that is too high. One counter is cleared 0.131 seconds
Function Description
after the other. This makes sure that too high a dose is sensed in
0.262 seconds.
When too much dose is sensed, the treatment terminates by the
removal of power to the HT relay.
DOS-PCB signal out of The field programmable gate array (FPGA) has an out of range
range measurement interlock. This counts the input data samples where the ADC has
saturated. If there are more than 511 of these occurrences during a
treatment, then the treatment is terminated. This prevents errors in
the dose measurement that are caused by large pulses that are in
error or have too much negative noise. These conditions can effect
the accuracy of the measured dose.
DOS-PCB dummy dose Channel 2 uses a set dose level to make sure that the DOS-B PCB
measurement operates correctly with dose signals. This set level is put on the
digital-to-analog converter (DAC) output to simulate a constant
dose signal.
When the set level is applied, the treatment control system (TCS)
measures the dose count and makes sure that it is applicable for that
level of signal. The TCS also makes sure that the DOS-B PCB stops
the delivery of radiation if the DOS-A PCB does not stop the delivery
of radiation.
This is description of the operation of the system bus access and write protection of the DOS PCB.
The system bus has:
• 7-bit address bus
• 16-bit data bus
The read/write control bit R/W goes through the address bus buffer (IC3). This sets a read access
when high (logic 1) and sets a write when low (logic 0).
The field programmable gate array (FPGA) uses the strobe signal STROBE to start bus transfers.
When low (logic 0), the STROBE signal tells the PCB that the address and write data are stable on
the system bus and can be read (see Figure 3.8).
When IC4 receives a STROBE signal, it sends a signal BOARD ACK through the system bus if the
address word is for the dosimetry PCB.
If the data is read from the dosimetry PCB, a signal DATA_VALID is sent through the system bus.
IC 1 and IC 2 are active and read the system bus most of the time. This keeps the dosimetry PCB
on and controls the data to the FPGA. When a read signal is decoded to the dosimetry PCB, the
direction pins (pin 1) are set high. This sets the data in the opposite direction and puts the data on
the system bus. The MTU PCB then reads the data and sets the STROBE high (logic 1).
The dosimetry PCB receives the system reset signal RESET which goes through a buffer IC4 to give
a RESET_IN signal. IC12 uses this to remove the power to the HT relay and terminate treatment
through IC6.
Function Description
Write protection The dosimetry PCB uses the write protection to stop writes, that are in error,
to the PCB during treatment. The PCB sets HT_ENABLE_OUT which
terminates the treatment if it senses a write that is in error.
A write error is set in the status word (bit 12 of read address 10). This tells the
treatment control system (TCS) the error that caused the treatment to be
terminated.
The write protection has three conditions as follows:
• Hardware signal PRF_ENABLE which, when logic high, prevents
some bus access
• Hardware signal CCI which, when logic high, prevents some bus
access
• A software setting that is set by the system software at specified
points during operation to lock out more, unwanted settings of the
dosimetry PCB.
Function Description
System backplane supplies The dosimetry PCB has three power rails from the system
backplane. These are:
• +5 V digital supply
• +15 V analog supply
• -15 V analog supply.
The DOS PCB supplies more voltages as follows:
• IC22 gives a +3.3 V digital power rail for the field
programmable gate array (FPGA) from the +5 V
digital supply.
• IC13 and IC16 supply +5 V and -5 V analog power
rails from the ±15 V power rails. These give a clean
supply for the analog section of the PCB.
• IC10 supplies a 4.096 V which is the reference voltage
from the +15 V power rail. This is for the analog-to-
digital converter (ADC) and the digital-to-analog
converter (DAC).
• IC20 supplies a 4.096 V which is the reference voltage
from the +15 V power rail. This is for the power rail
monitor ADC (IC17). It also supplies a positive offset
for the resistor networks that give the scale for the
negative voltage power rails.
-320V supply A -325 V supplies the ion chamber. The DOS A PCB has a
decreased version of this supply as a service item. This
decreased voltage levels is a nominal -6.86 V (±2% for external
Function Description
tolerances) for the -325 V monitor output. The monitor output
works in the range of -300 to -350 V
Power supply monitors An eight bit analog-to-digital converter ADC (IC17) monitors
the supply rails on the dosimetry PCB. This converter reads a
sample of the decreased power rail voltages in sequence. It then
sends each eight bit digital value as serial data to the field
programmable gate array (FPGA).
The 4V_REF for the rail monitor ADC is monitored. A change
in its value shows up as a change in all the rail voltages.
If a supply rail operates out of tolerance then the treatment
terminates with the removal of the power to the HT interlock
relay through HT_ENABLE_OUT. You can use the DIP
switches and LEDs to find the defective rail. For more
information about the rail monitors see Table 3.5.
3.8.2.7 Operation of the DOS PCB LED indications and switch settings
This is a description of the operation of the dosimetry PCB LEDs that can help when you do
troubleshooting.
Each dosimetry PCBs has eight diagnostic LEDs. You can view them at the PCB edge in the system
rack. The LEDs are as follows:
• LED 1: this comes on for a short time after power up when the PROM on the PCB
programs the field programmable gate array (FPGA).
• LED 2: this flashes each second while the PCB operates correctly
• LEDs 3 to 8: these let you examine the condition of the dosimetry PCB. Different signals
are routed to these LEDs when you set DIP switches 1 to 4 and 8.
Function Description
3.3 V and FPGA The 3.3 V watchdog circuit gives a power on reset if the voltage on the 3.3 V
watchdog circuit digital supply drops to less than 3.08 V (-6.7%). This reset usually occurs after
the supply watchdog circuit senses a fault and terminates treatment as its
tolerance is only -5%.
A timer in the FPGA sends a signal to this watchdog circuit each 1.6 seconds.
If not, a power on reset occurs.
4.6 V and rail This watchdog circuit monitors the voltage on the +5 V digital rail and resets
monitor if the voltage drops to less than 4.6 V (-8%). This reset usually occurs after the
watchdog rail watchdog circuit senses a fault and terminates treatment as its tolerance is
only -5%. This watchdog circuit also does a check for activity on the serial
data from the power rail monitor.
Dose output IC18 senses each dose pulse of more than 100 mV and appears as a
monitoring PULSE_PRESENT signal. A binary counter IC23 counts these pulses. If the
watchdog circuit counter increases to 1024 pulses before the FPGA counts a monitor unit
(MU) signal, then there is a fault. The watchdog circuit resets until it receives
a dose reset signal or a correct MU count.
System bus access This watchdog circuit is in the field-programmable-gate-array (FPGA)
watchdog circuit firmware. The circuit checks for a STROBE signal in 0.213 second periods to
make sure that the system bus is active.
The timer resets at each system bus read or write. If there is no system bus
activity, this watchdog removes HT interlock relay power through
HT_ENABLE_OUT.
This is a description of the functions of the dose pulse transmitter (DPT) PCB of the dosimetry system
in the radiation head control area (RHCA) of the digital accelerator.
The dosimetry system has one DPT PCB that monitors the dose information directly from the
DOS-A PCB in the RHCA. The DPT PCB communicates with the iView™ computer through an
RS232 or RS422 serial link. The DPT PCB is installed in the RHCA.
The DPT PCB sends signals to and receives signals from the iView™ computer (see Figure 3.11).
The primary functions of the DPT PCB are:
• Send and receive the dose information to and from the iView™ computer.
• Sense when the dose rate increases to a preset limit, to transmit a start byte.
• Sense when the dose increases to a preset limit, to transmit a stop byte.
• Sense when the dose rate decreases to a fixed lower limit, to transmit a beam end byte.
• Send a status byte if the PCB receives a diagnostics code from the iView™ computer.
• Supply a self-test sequence if the PCB receives a self-test command from the iView™
computer.
• Measure and transmit the atmospheric pressure to the TCS.
The DPT PCB receives the dose information directly from the DOS-A PCB as a pulse signal. Each
pulse signal is 1/64 of a monitor unit (MU). The DPT sends this information to the iView™
computer through the serial link.
The PCB has a pressure transducer circuit which measures local atmospheric pressure. This lets
the dosimetry system make corrections for open chamber sensitivity to pressure.
An interlock circuit senses normally-open and normally-closed conditions on a 3-pin connector.
Item
Pressure sensor The pressure sensor circuit gives a calibrated pressure output of 7.58
mV (±0.19 mV) for each millibar (mBar). Therefore an atmospheric
pressure (1000 mBar) gives approximately 7.58 V (±1%) when measured
at connector PL2-c29.
Programmable logic A programmable logic device (PLD) (IC6) contains the logic functions
device of the PCB. This includes the counters, machine states and other logic
functions of the PCB. An oscillator (IC3) connects to the PLD to give a
basic frequency of 3.6864 MHz and give the baud rate.
Setting the baud rate A quad dual-in-line switch (SW1) on the PCB lets you configure the
and interlock bypass baud rate and the interlock bypass condition of the PCB (see Table
3.10).
Switches SW1-1 to SW1-3 are set to make sure that the baud rate of the
serial communication port on the iView PC™ and the DPT PCB agree.
This is usually set at 9600 Baud.
Interlock bypass setting
• SW1-4 ON = interlock normally open
• SW1-4 OFF = interlock normally closed.
PROM configuration The PROM (IC8) holds the configuration data of the PLD. The PROM
uploads this data to the PLD at power up. If the PLD senses an error
during upload, the PLD uploads the configuration data again.
RS232/RS422 driver/ The DPT PCB has two serial communication channels, RS232 (IC7) and
receiver RS422 (IC5 and IC10). The communications can only occur on one
channel at a time through the inputs on opto-isolators (IC2 and IC4).
DC isolated area The DC-DC supply (IC1) and opto-isolators (IC2 and IC4) isolate the
DC circuits that connect to the RHCA.
Power supply The DPT PCB gets its +5 V and ±15 V supplies through the backplane.
ON ON OFF 19 200
ON ON ON 3840
The table gives a list and description of the light emitting diodes (LEDs) on the the front panel of the
DPT printed circuit board (PCB). The LEDs are adjacent to connector TU1.
This is a description of the functions of the servo input board (SIB) PCB in the dosimetry system of
the digital accelerator.
The dosimetry system has one SIB PCB in the RHCA of the digital accelerator (see Figure 3.13) .
The SIB PCB uses digital signal processors to do its function. The SIB PCB:
• Receives the current pulses from the signal plates in the ion chamber
• Applies the error correction
• Supplies the servo and monitor signals.
These output signals control and monitor the current through the beam centering coils (2R and
2T), and to control the gun filament current.
The PCB has a self test facility.
See Table 3.11 for information about the SIB PCB.
Item Description
Input amplifiers The SIB PCB has six input plate signals:
• Two from the 2T plates of the ion chamber
• Two from the 2R plates of the ion chamber
• One from the inner hump plate, which monitors the level of
the energy in the center of the chamber
• One from the outer hump plate which monitors the level of
the energy level in the outer parts of the chamber.
The plate signals are very high impedance current sources which give
tear-drop shape current pulses of approximately 150 µs. An amplifier
receives each input plate signal. The amplifier changes the current
source into a low impedance voltage.
Digitizers Each of the low impedance voltages from the input amplifiers is
digitized to supply a 12-bit digital word. The rate is controlled at 1 µs by
the CLOCK output from the oscillator.
Oscillator Oscillator OSC1 is a 20 MHz CMOS clock that times the SIB functions,
Error correction The error correction part of the FPGA receives the six digital data
streams (6 × 20 Mbits/second). The FPGA has a function that gives an
accurate sum of the input plate signals during each pulse. DC offset
errors, from a number of sources, have an effect on each plate signal:
Rectifier Description The PPG PCB sends a BLANK synchronization signal to the SIB PCB as
a 1.25 MHz carrier that continues for 220 µs in all energy pulses. The
energy pulse operates through a range of 400 Hz (each 2.5 ms) to 6 Hz
(each 166 ms). The BLANK pulse starts to time the FPGA
accumulations.
Digital-to-analog The digital data from the FPGA is sent in sequence to six of 12 DACs in
converter one IC (IC11). These give a current output in scale with the plate charge
digital word.
FPGA The outputs from the DACs are supplied to the FPGA and then put on
the system bus (D0 to D16) through buffers.
Gun amplifiers The X-ray gun servo system keeps the correct machine energy. The gun
amplifiers make small adjustments to the nominal gun current if there
are changes in magnetron power and frequency.
The X-ray flattening filter gives a flat X-ray field at the correct energy
setting. If the energy changes because of offsets or other changes in the
system, the X-ray field cannot stay flat:
• If the energy increases, this can cause the dose rate in the
center of the field to increase.
• If the energy decreases, this can cause the dose rate in the
center of the field to decrease.
Item Description
The X-ray gun servo system uses this property to sense if there is an
energy change with the two hump plates. The difference between the
two hump plate inputs causes an error signal. This corrects the direction
of the nominal level of gun current set by the operator.
The operation of the electron gun servo system is almost the same as the
X-ray system. In Electron mode, the inner hump signal is compared to a
DC reference voltage. The difference between these two voltages causes
an error signal, which sets the gun current to its nominal level.
3.8.5 Description of the liquid crystal displays and printed circuit board
This is a description of the LCD PCB which displays the dose count, and the control and
communication status of the dosimetry system.
The LCD PCB displays the dose count and the control and communication status of the dosimetry
system. A 4-digit LCD displays the dose count. Six LEDs show the status of the control and
communications system.
The LCD PCB is installed on the front panel of the TCC (Area 92). The user can easily see the LCD
and the LEDs.
The LCD PCB has a battery which supplies power to the dose count display if the primary
electrical supply goes off. The power failure circuit charges the battery.
The LCD PCB receives the dose information as one pulse for each MU. A 16-bit counter reads and
changes the count to operate four 7-segment displays. If there is a power failure, the last count is
latched into the converter. Before the display can be updated, it is necessary to reset the the latch.
The LCD PCB sends status signals to the TCS if the battery and oscillator are serviceable, and the
dose count is latched. The TCS can read the dose count in two bytes in time.
A 70 Hz waveform modulates the drive to the displays.
The LCD PCB is connected to the LIB.
The control sequence that the LCD PCB does at the start of each treatment is as follows:
• The TCS reads and does a check of the LCD counter against the dose for the previous
treatment.
• The BATTERY TEST signal is set low. After 80 ms the TCS does a check of the BATTERY
OK (board status) signal to make sure that it is low.
• The TCS does a check of the status to make sure that the OSCILLATOR OK and the COUNT
LATCHED signals are set high.
• The RESET DOSE READOUT signal is set low. After 5 ms, the TCS checks to make sure
that the OSCILLATOR OK is high and the COUNT LATCHED signal is low.
• The TCS does a check of the dose counter to make sure that it is at a value of zero.
This is a description of the operation of the -320 V system of the dosimetry system.
See the Elekta Digital Linear Accelerator System Diagrams manual for more information.
The -320 V system has a nominal voltage of -325 V. The monitor circuit has a range from -300 to
-350 V.
An auxiliary PSU in the BLD control area (Area 12) supplies the -320 V for the ion chamber
through a coaxial cable. In the ion chamber the ionizing voltage is supplied, in serial, to each plate.
The voltage is then applied to a potential divider network, where it is decreased to approximately
-6.8 V. The decreased voltage is sent through a coaxial cable to two dosimetry PCBs (PCB12B and
PCB12S) in Area 12.
The dosimetry PCBs monitor the voltage and, if the voltage goes out of limits, sends a signal to
software item i506 D1 supply or i456 D2 supply. The DOS-A PCB (PCB12E) also applies a
positive offset and sends the monitored voltage to the TCS. This gives an analog input to the SCC/
AI12 PCBs. Software item i189 -320V mon monitors this analog voltage, and has defined voltage
limits set during calibration.
The table gives a list of the control items which monitor the power supplies in the -320V system of the
dosimetry system.
i506 Dose channel 1 supply A hardware status bit that shows a fault in the
power supplies to dosimetry channel 1. The
monitored supplies are:
• +15V
• -15V
• +6.3 V ref
• -300V
• +5V
The table gives a list of the printed circuit boards (PCBs) which control the operation of the -320V
system of the dosimetry system.
This is a description of the operation of the temperature and pressure subsystem of the dosimetry
system.
See drawing Elekta Digital Linear Accelerator System Diagrams 4513 330 7048 Sheet 3 for more
information.
The temperature and pressure subsystem has dual circuits that measure the local atmospheric
pressure and temperature at the Beam Limiting Device (see Figure 3.14). Pressure and
temperature sensors give the data inputs for the temperature and pressure subsystem. A pressure
sensor on each of the DPT and SIB PCBs give the atmospheric pressure input. Two temperature
sensors on the collimator assembly give the temperature inputs.
The SCC-A and SCC-B PCBs supply a 10 V reference voltage to the two temperature sensors. The
sensors decrease the voltage to 0 V through resistors on the SIB PCB and the DPT PCB. The
temperature signals go through the SCC and AI12 PCBs to software items i224 and i226.
The DPT and SIB PCBs energize the pressure sensors. The pressure signals go through the SCC/
AI12 PCBs and are applied to software items i558 and i559.
The TCS receives the temperature and pressure data through the SCC and AI12 PCB pairs.
Treatment is not permitted if there is a difference between the temperature or pressure circuits.
During treatment, the delivery of radiation stops if there is a drift in the temperature or pressure
circuit.
The table gives a list of the control items for the temperature and pressure subsystem of the dosimetry
system.
Table 3.14 Control items for the temperature and pressure subsystem
The table gives a list of the control PCBs of the temperature and pressure subsystem of the dosimetry
system.
This is a description of the control items and PCBs that control the operation of the dosimetry system.
See Figure 3.15 for a block diagram of the dosimetry system.
The table gives a list of the items that control and monitor the operation of the dosimetry system.
Table 3.16 Control and monitor control items of the dosimetry system
The table gives a list of the PCBs that control and monitor the operation of the dosimetry system.
Table 3.17 Control and monitor control PCBs of the dosimetry system
The table gives a list of the dosimetry system conditions that can cause an HT relay interlock through
the HT radiation control circuit of the DOS PCB. If the HT relay interlock operates, it stops the
radiation delivery immediately.
— Blank page —
— Blank page —
Item Description
i665 Low Dose Mon This software item checks that the dose rate is not less than 75%
of the possible value. The item collects the dose samples (in
0.001 MU) that are less than 75% of the possible dose rate. The
item terminates if the low dose is more than the 10000 (10 MU)
limit.
i366 Dose Diff This item measures the dose difference which is calculated from
the monitored segment MU. The limits can be set to cause beam
termination. The range is ±100 MU.
i604 Ch1 Overrun This item gives a message to tell the user that:
• after a normal termination by dosimetry channel 1,
dosimetry channel 1 senses a dose count of a
minimum 1 MU during the Channel 2 check.
A system reset is necessary to clear the fault.
i507 D1 Ex.rate These items have hardware status bits which are set when the
i457 D2 Ex.rate dose rate is more than 1099 MU/minute.
i506 D1 supply These items have hardware status bits that show if there is a
i456 D2 supply fault in the dosimetry supply. The supplies examined are:
• +15 V
• -15 V
• +6.3 V ref
• -320 V
• +5 V.
i513 D1 cnt NAK These items have hardware status bit which show two counts on
i463 D2 cnt NAK the beam counter but the software did not acknowledge the first.
i475 D1 Force term The system uses these items to terminate the dose counting
i470 D2 Force term. system. This is usually terminated manually but can also be
from an internal problem such as a processor time out. The
items can only operate after the PRF is stopped.
Item Description
i504 D1 tst lnk These items show the hardware status bits which are set if the
i454 D2 tst lnk test link is installed. If the test link is installed, it is only possible
to operate in test mode.
i512 D1 ill.wrt These items show the hardware status bits that are set if an
i462 D2 ill.wrt illegal write operation occurs.
i484 Plates sum This item shows the sum of all the servo plates from the ion
chamber.
i511 D1 rly off These items show the hardware status bits that are set if the
i461 D2 rly off dosimetry channel is enabled. A fault status or the end of beam
on the hardware counters disable the relay. A dosimetry reset
enables the relay again. These items are active if the dose rate is
too high, for example, >800 MU/min.
1 Make sure that the digital accelerator is in the Ready To Start/HT On machine state.
2 Use the table that follows to do the troubleshooting if the dosimetry channel power supply
inhibit stops the radiation delivery. Do the steps in sequence until the dosimetry channel
inhibit clears.
If Then
The i506 D1 Supply 1 Do a check of the power supplies to the DOS A PCB.
error/inhibit is
displayed. This item 2 If it is necessary replace the defective power supply.
shows a dose channel
hardware status bit that
shows if there is a fault
in a dosimetry channel 1
power supply.
A i456 D2 Supply is 1 Do a check of the power supplies to the DOS B PCB.
displayed. This item
shows a dose channel 2 If it is necessary replace the defective power supply.
hardware status bit that
shows if there is a fault
in a dosimetry channel 2
power supply.
1 Make sure that the digital accelerator is in the Closed, Preparatory, or higher machine state.
2 Use the table that follows to do the troubleshooting if a dosimetry pressure inhibit stops the
radiation delivery. Do the steps in sequence until the dosimetry pressure inhibit clears.
If Then
The i771 Pressure Diff error/inhibit is The pressure difference between the two
displayed. pressure sensors is more than 3 mbar.
1 Calibrate the dosimetry pressure.
The i773 Pressure Change error/inhibit is The dosimetry pressure change is more than
displayed. 3 mbar since the start of treatment.
1 Replace the SIB PCB.
2 Replace the PSB/DPT PCB.
The i558 Pressure 1 or i559 Pressure 2 error/ The dosimetry pressure change is more than
inhibit is displayed. 100 mbar from the mean value.
1 Calibrate the dosimetry pressure.
Related Links:
Calibrating the dosimetry pressure on page 91
1 Make sure that the digital accelerator is in the Closed, Preparatory, or higher machine state.
2 Use the table that follows to do the troubleshooting if a dosimetry temperature inhibit stops
the radiation delivery. Do the steps in sequence until the dosimetry temperature inhibit clears.
If Then
The i770 Dose Temp The temperature difference between the two temperature sensors
Diff error/inhibit is is more than 1 °C.
displayed.
1 Calibrate the dosimetry temperature.
If Then
The i772 Temp Change The dosimetry temperature change is more than 1 °C since the
error/inhibit is start of treatment.
displayed.
1 Calibrate the collimator temperature sensor.
— Blank page —
WARNING 5.1
Record the part numbers of the replacement parts when you replace a part. Do the
applicable test, set up, and calibration procedures for the replacement parts before clinical
operation. If you ignore this warning, you can cause clinical mistreatment.
WARNING 5.2
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, you can cause fatal injury.
Note: Obey the ESD procedures and precautions when you do work in a control area. If you ignore this
caution, you can cause damage to the equipment.
2 At the RHCA (AREA 12), remove the RF EMC screen that gives the control area protection.
3 Remove the screws and nuts that attach the bar, and remove the bar.
CAUTION 5.1
Do not touch the PCBs and other electronic parts unless you wear a grounded antistatic wrist
strap. If you ignore this caution, you can cause damage to the equipment.
4 Put on an antistatic wrist strap.
5 At the applicable DOS PCB, pull out the red levers on the front edge of the PCB.
This releases the PCB from the rear connector on the backplane of the control area.
6 Carefully remove the DOS PCB from the PCB rack.
7 Make sure that the replacement DOS PCB has the correct configuration and part number, see
the Service Parts Catalog for information about the PCB.
8 Carefully push the DOS PCB into its card slot until the edge connector of the PCB touches the
backplane connector.
9 Continue to push the PCB until the connectors fully engage.
10 At the RHCA (AREA 12), install the screws and nuts to attach the bar to the rack.
11 Remove the antistatic wrist strap.
12 Install the RF EMC screen.
13 Connect the electrical supply to the RHCA.
14 Calibrate dose A and dose B.
WARNING 5.3
Record the part numbers of the replacement parts when you replace a part. Do the
applicable test, set up, and calibration procedures for the replacement parts before clinical
operation. If you ignore this warning, you can cause clinical mistreatment.
WARNING 5.4
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, you can cause fatal injury.
Note: Obey the ESD procedures and precautions when you do work in a control area. If you ignore this
caution, you can cause damage to the equipment.
2 At the RHCA (AREA 12), remove the RF EMC screen that gives the control area protection.
CAUTION 5.2
Do not touch the PCBs and other electronic parts unless you wear a grounded antistatic wrist
strap. If you ignore this caution, you can cause damage to the equipment.
3 Remove the screws and nuts that attach the bar, and remove the bar.
4 Put on an antistatic wrist strap.
5 At the applicable SIB PCB pull out the red levers on the front edge of the PCB.
This releases the PCB from the rear connector on the backplane of the control area.
6 Carefully remove the SIB PCB from the PCB rack.
7 Make sure that the replacement SIB PCB has the correct configuration and part number, see
the Service Parts Catalog for information about the PCB.
8 Carefully push the SIB PCB into its card slot until the edge connector of the PCB touches the
backplane connector.
9 Continue to push the PCB until the connectors fully engage.
10 At the RHCA (AREA 12), install the screws and nuts to attach the bar to the rack.
11 Remove the antistatic wrist strap.
12 Install the RF EMC screen.
13 Connect the electrical supply to the RHCA.
14 Calibrate the items that follow:
– 2T and 2R current
– gun current control
– hump gain
– Dose level.
WARNING 5.5
Record the part numbers of the replacement parts when you replace a part. Do the
applicable test, set up, and calibration procedures for the replacement parts before clinical
operation. If you ignore this warning, you can cause clinical mistreatment.
WARNING 5.6
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, you can cause fatal injury.
Note: Obey the ESD procedures and precautions when you do work in a control area. If you ignore this
caution, you can cause damage to the equipment.
2 At the RHCA (AREA 12), remove the RF EMC screen that gives the control area protection.
3 Remove the screws and nuts that attach the bar, and remove the bar.
4 Put on an antistatic wrist strap.
Note: Do not touch the PCBs and other electronic parts unless you wear a grounded antistatic wrist strap.
The result can be an electrostatic discharge and a malfunction. If you ignore this caution, you can
cause damage to the equipment.
5 At the front edge of the HDRE attenuator PCB, pull out the red levers.
This releases the PCB from the rear connector on the backplane of the control area.
6 Carefully remove the HDRE attenuator PCB from the PCB rack.
7 Make sure that the replacement HDRE attenuator PCB has the correct configuration and part
number, see the Service Parts Catalog for information about the PCB.
8 Carefully push the HDRE attenuator PCB into its card slot until the edge connector of the
PCB touches the backplane connector.
9 Continue to push the PCB until the connectors fully engage.
10 At the RHCA (AREA 12), install the screws and nuts to attach the bar to the rack.
11 Remove the antistatic wrist strap.
12 Install the RF EMC screen.
13 Connect the electrical supply to the RHCA.
Hospital
Name of engineer
Machine number Date
i64 2R Ctrl i65 2T Ctrl i327 Gun l i65 2T Ctrl i327 Gun l
Energy Part 4 Servo Part 4 Servo Part 4 Servo Part 4 Servo Part 4 Servo
AUTO AUTO AUTO MANUAL MANUAL
X-LOW
X-MID
X-HIGH
E4
E6
E8
E10
E12
E15
E18
E20
E22
20 From the Beam Control page, set the Gun Servo and 2R Servo to 0 (MANUAL).
21 When the beam is stable, in the Table 5.1, record the Part 4 values of:
– i164 2R I Control
– i327 Gun I Control.
22 Do steps 6 to 21 again for each available X-ray and electron energy.
23 Load the Area Voltages service page.
24 Record the value of Item 189.
Prerequisites
Special tools
• Hoist
• Lifting straps
• Eye bolts
WARNING 5.7
Record the part numbers of the replacement parts when you replace a part. Do the
applicable test, set up, and calibration procedures for the replacement parts before clinical
operation. If you ignore this warning, you can cause clinical mistreatment.
WARNING 5.8
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, you can cause fatal injury.
WARNING 5.9
Record the part numbers of the replacement parts when you replace a part. Do the
applicable test, set up, and calibration procedures for the replacement parts before clinical
operation. If you ignore this warning, you can cause clinical mistreatment.
WARNING 5.10
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, you can cause fatal injury.
CAUTION 5.3
Do not touch the ion chamber with your bare hands, always wear gloves. If you touch the ion
chamber with your bare hands the grease from your fingers can cause damage to the ion
chamber.
1
4
2 5
3 3
Related Links:
Removing the collimator on page 81
WARNING 5.11
Record the part numbers of the replacement parts when you replace a part. Do the
applicable test, set up, and calibration procedures for the replacement parts before clinical
operation. If you ignore this warning, you can cause clinical mistreatment.
WARNING 5.12
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, you can cause fatal injury.
CAUTION 5.5
Do not touch the ion chamber with your bare hands, always wear gloves. If you touch the ion
chamber with your bare hands the grease from your fingers can cause damage to the ion
chamber.
CAUTION 5.6
Do not touch the ion chamber unless you wear a grounded antistatic wrist strap. The result
can be an electrostatic discharge and a malfunction. If you ignore this caution, you can cause
damage to the equipment.
3 Put on the antistatic wrist strap and do a test of the connection to ground.
4 Attach the long bracket (2) and the short bracket (5) to the ion chamber (1) with the four
screws (see Figure 5.2.
1
4
2 5
3 3
Related Links:
Installing the collimator on page 85
Prerequisites
Special tools
• Hoist
• Lifting straps
• Eye bolts
WARNING 5.13
Record the part numbers of the replacement parts when you replace a part. Do the
applicable test, set up, and calibration procedures for the replacement parts before clinical
operation. If you ignore this warning, you can cause clinical mistreatment.
WARNING 5.14
Do not start to do work on the digital accelerator before you isolate the power to the work
area at the interface cabinet. You can get a high voltage electric shock. If you ignore this
warning, you can cause fatal injury.
1 Make sure that the dowel pins, which hold the collimator to the magnet plate, are correctly in
their positions.
2 Attach the lifting straps to the eye bolts.
3 Carefully lift and move the collimator in position for installation.
4 Before you lower the collimator, make sure that all the cables are clear of the installation area.
WARNING 5.15
Do not put your hands below the collimator as it is lowered. If it is necessary, use wood
blocks to hold the weight of the collimator. If you ignore this warning, it can cause serious
injury and damage to equipment.
Note: If it is necessary to lift the collimator during the installation, you must lift it clear of the dowel pins.
This lets you make sure that the location of the dowel pins in the magnet plate is correct. You can
then engage the collimator on the dowel pins, fully lower the collimator, and attach it to the magnet
plate.
5 Slowly and carefully lower the collimator. Make sure that the cables to the bending fine
magnet does not catch in the equipment.
6 Connect all the connectors to the mounting bracket on the side of the collimator.
7 Connect the ground (earth) cables to the PCB on the side of the collimator.
8 Connect the red cable to the bending fine magnet on the side of the gantry arm.
9 Install the lead shielding in the gantry arm.
10 Make sure that the target shift mechanism operates freely.
11 Install the three nuts and washers to attach the collimator to the magnet plate.
12 Install the BLD (see the BLD Corrective Maintenance Manual for more information).
13 Set the gantry to 0°.
14 Do an X-ray to light field test at the isocenter.
15 Compare the results with those recorded in the removal task. Make sure that no large
differences are apparent.
Related Links:
Removing the collimator on page 81
— Blank page —
WARNING 6.1
Do not save the calibration data until you make sure that all the parameters are correct.
When you save the calibration data, the system saves a full Linac Calibration. If you ignore
this warning, it can cause clinical mistreatment.
16 Select Save LINAC calibration.
17 Set i181 to 1 (Auto).
Prerequisites
Test equipment:
• Farmer ionization chamber dosemeter
• 20 x 20 polystyrene phantom
5 Set up the Farmer ionization chamber, corrected for temperature and pressure.
6 Make three exposures in Interlock Group 7.
7 Record the value of each exposure.
8 Calculate the average value of the three exposures.
9 Select the Dose Reference Calculator.
10 In the Delivered Dose boxes, type the number of MU delivered for each dosimetry channel.
11 In the Measured Dose box, type the value calculated in step 8 .
12 Click the Set Ref 1 and Set Ref 2 buttons to set the new reference dose values.
13 Click the Save icon.
14 Select Save Energy Cal.Blocks.
15 Do steps 1 to 14 again for all other X-ray energies.
16 Make sure that there is a minimum of 400 cGy/min and a maximum of 850 cGy/min at each
X-ray energy. For 4MV, the minimum level is 350 MU/min.
The procedure to calibrate the electron dose rate of the digital accelerator.
Prerequisites
Test equipment:
• Water phantom
15 For each Electron energy, deliver three beams in sequence. Make sure that the dosimetry
parameters of each beam are in specification.
Related Links:
Calibrating the dosimetry pressure on page 91
Calibrating the dosimetry temperature on page 91
Prerequisites
Thermometer with an accuracy of ±0.5°C.
WARNING 6.2
Do not save the calibration data until you make sure that all the parameters are correct.
When you save the calibration data, the system saves a full Linac Calibration. If you ignore
this warning, it can cause clinical mistreatment.
11 Click Save LINAC Calibration.
Prerequisites
Test equipment:
• Calibrated barometer with an accuracy of 0.1 mbar.
WARNING 6.3
Do not save the calibration data until you make sure that all the parameters are correct.
When you save the calibration, data, the system saves a full Linac Calibration. If you ignore
this warning, it can cause clinical mistreatment.
12 Select Save LINAC Calibration.
13 Make sure that the recorded values for part 10 on the DPT PCB and the servo input board
(SIB) PCB are not more than 40.
A value of more than 40 can show a sensor that is not stable.
Prerequisites
Test equipment:
• Multimeter
• BNC-T connector (75Ω).
1 Put the BNC-T connector in line with SK12-FF and PL12-FF in the RHCA (Area 12).
2 Connect a multimeter to the other end of the BNC-T connector.
3 Load the Calibration service page.
4 Type 189 into Cal.item 1.
5 Disconnect the BNC-T connector from SK12-FF.
6 Type the value 0 into Cal Value .
7 Connect the BNC-T connector to SK12-FF.
8 Type the measured value into Cal Value as a negative four-digit value.
For example, -3200 for a measured value of 320.0 V.
This is the procedure to make sure that the ion chamber dose leakage is not more than the set limits.
This is the procedure to do a functional test of the dosimetry system to make sure that it is stable.
Software items
Note: Make sure that the digital accelerator is warmed up before you start the calibration of the nominal
dose rate.
Note: See the Digital Accelerator Corrective Maintenance Reference Manual for more information about
the machine Items.
7 On the function keypad (FKP), press the MV START key to start the beam.
8 At the bottom of the Deliver Quick Beam dialog box, make sure that the bar indicator
becomes yellow and a dose rate value appears.
9 On the secondary function toolbar, click Display Service Pages.
The Display Service Pages dialog box appears.
10 Click Defaults.
11 Adjust the machine items to make sure that the digital accelerator operates at the maximum
dose rate.
12 Click Load.
The Service Page Selection dialog box appears.
13 In the Pages Available list, select Calibration.
14 Click OK.
15 Set Cal.item 1 to 44.
16 Set Learn enable to 1.
17 On the secondary function toolbar, click Edit Machine Item Parts.
The Edit Machine Item Parts dialog box appears.
18 Select Item 44 and Part 134.
Item 44 Part 134 automatically adjusts to show the dose rate for the current energy.
19 Make sure that Item 44 Part 134 has a constant value.
Learn enable automatically updates to show 0.
20 Set Cal.item 1 to 0.
21 On the primary function toolbar, click Save .
The Save Linac Database and CalBlock dialog box appears.
22 Click Save Energy Cal. Block.
23 At the Beam Delivery dialog box, click Terminate to stop the beam.
24 Do the steps 1 to 23 for each available energy.
The procedure to calibrate the low dose rate monitor of the digital accelerator.
The digital accelerator has a dose interlock that collects differences between the actual dose rate
and the selected dose rate. The actual dose rate must be in a defined range of the selected dose rate
or the related dose rate interlock stops the beam. Therefore, the accurate setting of the nominal
dose rate for each energy is very important.
The low dose rate monitor records the actual dose rate delivered. The stability timer, Item 665
Part 135 (the units are in 125 ms), lets the dose rate become stable before the low dose rate
monitor starts to record. You can adjust this function for a slow beam start.
Note: For Fast Tuning Magnetrons, the limit of the stability timer is 2.5 seconds (20 units). For Non-Fast
Tuning Magnetrons, the limit of the stability timer is 5.0 seconds (40 units).
If the actual dose rate is less than 75% of the selected dose rate, the system records the difference. If
the difference is 10 MU or more, the beam stops. This can occur on a digital accelerator with a
slow start. If a slow start occurs, the low dose rate monitor can record before it increases to the
selected dose.
Note: Make sure that you set up an MU value that is sufficient to complete this procedure.
Note: Calibration of the nominal dose rate is necessary for the low dose rate monitor to operate correctly.
Related Links:
Calibrating the nominal dose rate on page 94
WARNING 1.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
A warning is information about procedures, tasks, or conditions, where the result can be fatal
or serious injury, injury, or clinical mistreatment, if you do not obey the instructions.
CAUTION 1.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
A caution is information about procedures, tasks, or conditions, where the result can be
damage to the equipment, data, or the environment, if you do not obey the instructions.
WARNING 1.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Do not use parts or accessories that are not approved by Elekta. The result can be incorrect
radiation delivery and other safety risks. If you ignore this warning, it can cause fatal injury or
clinical mistreatment.
WARNING 1.4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Do not lift or move heavy parts unless you obey the applicable Health and Safety regulations
and procedures. If you ignore this warning, it can cause serious injury.
WARNING 1.5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Do not use a visual display unit (VDU) unless you obey the applicable Health and Safety
regulations and procedures. If you ignore this warning, it can cause injury.
WARNING 1.6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Do not operate the equipment unless the QA checks are satisfactorily completed, or if you
know, or think that the equipment is defective or incorrectly calibrated. If you ignore this
warning, you can cause fatal injury, clinical mistreatment, and damage to the equipment.
WARNING 1.7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Do not remove, change, or override the switches, interlocks, or other safety devices unless
special instructions in this document tell you to. The result can be incorrect radiation delivery
and other safety risks. When the work is completed, set all the safety devices again and do a test
of them. If you ignore this warning, it can cause fatal injury or clinical mistreatment.
WARNING 1.8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning, it
can cause fatal injury.
WARNING 1.9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Do not remove the covers or cables from the equipment unless special instructions in this
document tell you to. The covers give access to high voltage electric parts. Always install the
covers and cables again before you use the equipment. If you ignore this warning, it can cause
fatal injury.
WARNING 1.10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Do not remove the covers from the equipment unless special instructions in this document tell
you to. The covers give access to parts that move. Always install the covers again before you use
the equipment. If you ignore this warning, it can cause fatal injury.
WARNING 1.11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Be careful when the power supply switch of the equipment is put in the off position. There
continues to be a risk of high voltage electric shock. Some of the customer interfaces can
continue to supply electrical power to the electrical terminals. If you ignore this warning, it can
cause fatal injury.
CAUTION 1.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Do not touch the PCBs and other electronic parts unless you wear a grounded antistatic wrist
strap. The result can be an electrostatic discharge and a malfunction. If you ignore this caution,
you can cause damage to the equipment.
CAUTION 1.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Do not touch the parts identified with the ESD caution label. Always obey the ESD safety
precautions when you do work on these parts. If you ignore this caution, you can cause damage
to the equipment.
WARNING 1.12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Do not use the equipment in rooms that contain explosive or flammable gases. High levels of
oxygen, and some anesthetic gases and disinfectant sprays, are explosive or flammable. If you
ignore this warning, it can cause fatal injury and damage to the equipment.
WARNING 1.13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Do not put portable or mobile radio frequency (RF) communication devices, or other electrical
devices, near the equipment. If you ignore this warning, you can cause fatal injury, clinical
mistreatment, and damage to the equipment.
WARNING 1.14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Do not use the accessories, transducers, and cables unless they are specified by Elekta. If you
ignore this warning, you can cause fatal injury and clinical mistreatment.
WARNING 1.15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Do not look into the ultraviolet light without protective glasses when the Agility™ BLD covers
or crosswires screen are removed. If you ignore this warning, it can cause serious injury to your
eyes.
WARNING 1.16 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Do not look into the laser beams. The laser beams can cause damage to your eyes. If you ignore
this warning, the laser beams can cause serious injury to your eyes.
WARNING 1.17 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Do not do maintenance work, or make modifications to the equipment unless you are a
qualified person and have the applicable authority. The result can be incorrect radiation
delivery and other malfunctions. If you ignore this warning, it can cause fatal injury and clinical
mistreatment.
WARNING 1.18 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Do not touch the flight tube, bending magnet, or other parts near to them until a check for
induced radioactivity is completed. The risk can be excessive exposure to radioactivity. If you
ignore this warning, it can cause serious injury.
WARNING 1.19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Do not move radioactive material (by air, sea, or land) unless you obey the applicable laws and
regulations for induced radioactivity and disposal of radioactive material of the jurisdictions
through which the movement occurs. If you ignore this warning, it can cause serious injury and
damage to the environment.
WARNING 1.20 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Do not clean the equipment, or use a disinfectant spray, before you isolate the equipment from
the electrical supply. You can get a high voltage electric shock. If you ignore this warning, it can
cause fatal injury.
WARNING 1.21 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Do not use a spray that is explosive or flammable. The fumes from such sprays can cause an
explosion or start a fire. If you ignore this warning, it can cause fatal injury and damage to the
equipment.
CAUTION 1.4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Do not let water and other liquids go into the equipment. The results can be electrical short
circuits and metal corrosion. If you ignore this caution, it can cause damage to the equipment.
CAUTION 1.5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Do not use a spray to clean the medical equipment room. The spray can go into the equipment.
The results can be electrical short circuits and metal corrosion. If you ignore this caution, it can
cause damage to the equipment.
WARNING 1.22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
The expected service life of this equipment is 15 years of normal use. Because of wear of parts,
the risk of malfunctions can increase after 10 years of operation. Before you continue operation
after 10 years, complete the regular extended life checks. If you ignore this warning, it can cause
fatal injury.
WARNING 1.23 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Do not move or discard radioactive and hazardous material unless you are a qualified person
and have an applicable license. Obey the laws and regulations of the jurisdiction for disposal
and movement of these materials. If you ignore this warning, it can cause fatal injury and
damage to the environment.
WARNING 5.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Record the part numbers of the replacement parts when you replace a part. Do the applicable
test, set up, and calibration procedures for the replacement parts before clinical operation. If
you ignore this warning, you can cause clinical mistreatment.
WARNING 5.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning,
you can cause fatal injury.
CAUTION 5.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Do not touch the PCBs and other electronic parts unless you wear a grounded antistatic wrist
strap. If you ignore this caution, you can cause damage to the equipment.
WARNING 5.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Record the part numbers of the replacement parts when you replace a part. Do the applicable
test, set up, and calibration procedures for the replacement parts before clinical operation. If
you ignore this warning, you can cause clinical mistreatment.
WARNING 5.4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning,
you can cause fatal injury.
CAUTION 5.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Do not touch the PCBs and other electronic parts unless you wear a grounded antistatic wrist
strap. If you ignore this caution, you can cause damage to the equipment.
WARNING 5.5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Record the part numbers of the replacement parts when you replace a part. Do the applicable
test, set up, and calibration procedures for the replacement parts before clinical operation. If
you ignore this warning, you can cause clinical mistreatment.
WARNING 5.6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning,
you can cause fatal injury.
WARNING 5.7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Record the part numbers of the replacement parts when you replace a part. Do the applicable
test, set up, and calibration procedures for the replacement parts before clinical operation. If
you ignore this warning, you can cause clinical mistreatment.
WARNING 5.8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning,
you can cause fatal injury.
WARNING 5.9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Record the part numbers of the replacement parts when you replace a part. Do the applicable
test, set up, and calibration procedures for the replacement parts before clinical operation. If
you ignore this warning, you can cause clinical mistreatment.
WARNING 5.10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning,
you can cause fatal injury.
CAUTION 5.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Do not touch the ion chamber with your bare hands, always wear gloves. If you touch the ion
chamber with your bare hands the grease from your fingers can cause damage to the ion
chamber.
CAUTION 5.4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Do not touch the ion chamber unless you wear a grounded antistatic wrist strap. The result can
be an electrostatic discharge and a malfunction. If you ignore this caution, you can cause
damage to the equipment.
WARNING 5.11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Record the part numbers of the replacement parts when you replace a part. Do the applicable
test, set up, and calibration procedures for the replacement parts before clinical operation. If
you ignore this warning, you can cause clinical mistreatment.
WARNING 5.12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning,
you can cause fatal injury.
CAUTION 5.5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Do not touch the ion chamber with your bare hands, always wear gloves. If you touch the ion
chamber with your bare hands the grease from your fingers can cause damage to the ion
chamber.
CAUTION 5.6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Do not touch the ion chamber unless you wear a grounded antistatic wrist strap. The result can
be an electrostatic discharge and a malfunction. If you ignore this caution, you can cause
damage to the equipment.
WARNING 5.13 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Record the part numbers of the replacement parts when you replace a part. Do the applicable
test, set up, and calibration procedures for the replacement parts before clinical operation. If
you ignore this warning, you can cause clinical mistreatment.
WARNING 5.14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Do not start to do work on the digital accelerator before you isolate the power to the work area
at the interface cabinet. You can get a high voltage electric shock. If you ignore this warning,
you can cause fatal injury.
WARNING 5.15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Do not put your hands below the collimator as it is lowered. If it is necessary, use wood blocks
to hold the weight of the collimator. If you ignore this warning, it can cause serious injury and
damage to equipment.
WARNING 6.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Do not save the calibration data until you make sure that all the parameters are correct. When
you save the calibration data, the system saves a full Linac Calibration. If you ignore this
warning, it can cause clinical mistreatment.
WARNING 6.2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Do not save the calibration data until you make sure that all the parameters are correct. When
you save the calibration data, the system saves a full Linac Calibration. If you ignore this
warning, it can cause clinical mistreatment.
WARNING 6.3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Do not save the calibration data until you make sure that all the parameters are correct. When
you save the calibration, data, the system saves a full Linac Calibration. If you ignore this
warning, it can cause clinical mistreatment.
— Blank page —
Elekta AB (publ) North America Europe, Latin America, Africa, Asia Pacific, Hong Kong,
Box 7593, SE-103 93 Stockholm, Sweden Atlanta, USA Middle East & India China
Tel: +46 8 587 254 00 Tel: +1 770 300 9725 Tel: +44 1293 544 422 Tel: +852 2891 2208
Fax: +46 8 587 255 00 Fax: +1 770 448 6338 Fax: +44 1293 654 321 Fax: +852 2575 7133
info@elekta.com info.america@elekta.com info.europe@elekta.com info.asia@elekta.com