bizhub PRESS C7000 Training Program

4 & 10 DAY DENSITY BALANCE ADJUSTMENT LAB

Lab Assignment Objectives:


Students will be able to:
1. Perform the Density Balance Adjustment procedure on the C6000/C7000 with a spectrometer
and applicable software.
2. Understand the various steps involved for measuring and applying Density Balance data, as
well as, understanding the process for assigning the data to specific media.

Estimated Completion Time: 60 Minutes

Before you Begin:


 Load Ledger (11x17), 20 lb in Tray 2.
 Configure the paper settings according to the media type being used.

Precautions/Warnings:
 Use A3 / 11” x 17” sheets for chart output; use the paper specific to the job.
 There are several User’s Guides, as well as, ‘Help’ for the customer through the Control panel.
 This operation requires a spectrophotometer (i1-iSis or i1-Pro). We recommend the i1-iSis.
 Install and use the “Profile Maker Measure Tool” software.
 A dongle is required for using the application software; Profilemaker Pro 5 for advanced
software functions, but this lab should be able to be completed without it.

Page 1 of 7 C7000_Density_Balance_Adjustment_Lab.doc
Density Balance Adjustment Lab

Task 1: C7000 POD User’s Guide Section 5 Adjustment 5.11

1. Press [Adjustment] on the Machine


Status Screen.

2. Press [04 Density Balance


Adjustment] and [01 Density Balance
Chart Output].

3. Select the desired spectrophotometer


and type of chart to be used.

4. Select Type1 under <Measure


Spectrophotometer-1 {i1-iSis}> or
Measure Spectrophotometer -2 {i1-
Pro}

5. Touch [Print Mode].

Page 2 of 7 C7000_Density_Balance_Adjustment_Lab.doc
Density Balance Adjustment Lab

6. Select the screen type by touching


[Quality Adj.].

7. Select the desired screen type under


[Screen].

8. For this lab select Dot1/Dot190.

9. Touch OK (or Close) to close the Quality


Adj. Screen.

Page 3 of 7 C7000_Density_Balance_Adjustment_Lab.doc
Density Balance Adjustment Lab

10. Select the appropriate paper size and


tray for the chart output. (11 X 17 paper)

11. Press Start on the control panel.

This will produce test charts selected in step


4 for scanning in color balance data.

12. Press [Exit Print Mode].

13. Touch [Return] to restore the Density


Balance Adjustment Menu Screen.

14. Prepare the Spectrophotometer i1-iSis


[or i1-Pro] by connecting it to a PC
equipped with a USB flash drive.

15. On the USB flash drive, create the path


[C7000/ADJUST_DATA].

Page 4 of 7 C7000_Density_Balance_Adjustment_Lab.doc
Density Balance Adjustment Lab

16. Launch the Profile Maker Measure Tool


application on the PC (listed under the X-
Rite programs folder).

17. Click Configuring.

18. Make sure to select the appropriate


Instrument (e.g., i1isis, or i1Pro); check
the device settings and make sure that
the device status is OK.

19. Close the Instrument Configuration


window before proceeding.

20. Click on Measuring; select the test chart


reference file that matches the size of the
printed output (e.g. DenBal11x17.txt,
then click Start.

Note: You may have to browse to the folder


where the file is stored by clicking on
Open…

Page 5 of 7 C7000_Density_Balance_Adjustment_Lab.doc
Density Balance Adjustment Lab

21. When the measurement is completed,


click on Close.

22. On the next screen click on Export


Lab...

23. Save the colorimetric data in the


following folder path
[C7000/ADJUST_DATA]. Enter a file
name starting with the ID number printed
on the test chart, followed by an
underscore, followed by your desired
name as a unique identifier for the file.
(e.g. ID No_desired name.txt)

24. Back at the machine, touch [02 Density


Balance Data Reg/Del] on the Density
Balance Adjustment Menu Screen

25. Select the desired row number from the


10 on the list.

Page 6 of 7 C7000_Density_Balance_Adjustment_Lab.doc
Density Balance Adjustment Lab

26. Touch [Measured data load].

A dialog appears to prompt you to connect


the USB flash drive.

27. Connect the USB flash drive stored with


the measured data into the USB port of
the machine.

28. Press [OK].

The file name of the measured data will be


displayed on the Density Balance Adj. Data
Register/Del Screen.

29. Press [OK].

The Density Balance Adj. Data Register/Del


Screen will be restored, with [NEW
PROFILE] displayed in the [Adj. Data
ProfileName] field.

30. Press [Change Name]. Enter the


desired profile name, then press [OK].

31. Press either [Enable] or [Disable] in the


right side area of the screen.

Note: If multiple pieces of adjustment data


are registered for the same screen type,
selecting [Enable] for the desired one will
automatically set the others to [Disable].

32. Press [Return] to restore the Density


Balance Adjustment Menu Screen.

Answer the following:

 How many measurements of Density Balance Data can the user register on the bizhub PRESS
C7000?

This completes the Lab Exercise

Page 7 of 7 C7000_Density_Balance_Adjustment_Lab.doc