Specifications

Converter
Product name Immersion type turbidity/SS meter
Model HU-200TB-IM Industrial-use turbidity/SS converter
Detector type SS-150 Industrial-use turbidity/SS detector
Right angle scattering/ Measuring principle suitable for full range turbidity and SS within 200
Measurement transmission mg/L by Kaorin
method
Transmission Measuring principle suitable for SS over 50 mg/L by Kaorin
Scattering/ Formazin 0 NTU to 4000 NTU
Turbidity
transmission Kaolin 0 to 2000 (Display range: 2001 to 4000)
Measurement range Scattering/
transmission 0 mg/L to 2000 mg/L
SS Kaolin
(Display range: 2001 mg/L to 4000 mg/L)
Transmission
0 NTU to 2000 Within 2 of the reading or 0.5
NTU NTU, whichever is larger
Formazin
2001 NTU to
Scattering/ Within 3 of the reading
Turbidity 4000 NTU
transmission
Within 2 of the reading or 0.5,
0 to 1000
Kaolin whichever is larger
Repeatability 1001 to 2000 Within 5 of the reading
0 mg/L to 1000 Within 2 of the reading or 0.5 mg/
Scattering/ mg/L L, whichever is larger
transmission 1001 mg/L to
SS Kaolin Within 5 of the reading
2000 mg/L
0 mg/L to 2000 Within 5 of the reading or 0.5 mg/
Transmission
mg/L L, whichever is larger
The deviation at the mid-range point
0 NTU to 2000 of the span calibration value is within
NTU 3 of the span calibration value or
Formazin 3 NTU, whichever is larger
The deviation at the mid-range point
2001 NTU to
of the span calibration value is within
4000 NTU
Scattering/ 5 of the span calibration value
Linearity Turbidity
transmission The deviation at the mid-range point
of the span calibration value is within
0 to 1000
3 of the span calibration value or
Kaolin 3, whichever is larger
The deviation at the mid-range point
1001 to 2000 of the span calibration value is within
5 of the span calibration value

55
Data

The deviation at the mid-range point


0 mg/L to 1000 of the span calibration value is within
mg/L 3 of the span calibration value or
Scattering/ 3 mg/L, whichever is larger
transmission
The deviation at the mid-range point
1001 mg/L to
of the span calibration value is within
2000 mg/L
5 of the span calibration value
Linearity SS Kaolin
The deviation at the mid-range point
0 mg/L to 1000 of the span calibration value is within
mg/L 3 of the span calibration value or
Transmission 3 mg/L, whichever is larger
The deviation at the mid-range point
1001 mg/L to
of the span calibration value is within
2000 mg/L
10 of the span calibration value
0.01 (Range: 0 to 10)
Auto decimal point or fixed decimal
Turbidity 0.1 (Range: 10 to 100)
point can be specified
1 (Range: 100 to 4000)
SS 1 Fixed range
Resolution
0.000 to 3.000
Concentration can be obtained by the
Absorbance 0.001 cubic function of the absorbance.
The converted value should be within
3.000.
Number of current outputs 2
Output type 4 mA to 20 mA DC, input/output isolation type
Load resistance 900 (max.)
Transmission output Linearity Within 0.08 mA (limited to the current linearity)
Repeatability Within 0.02 mA (limited to the current repeatability)
Error output Provided with burnout function (3.8 mA or 21 mA)
Transmission holding One of the latest value or preset value can be specified
Number of contact outputs 6
Output type No-voltage output contact
Contact type Relay contact, SPST (1a)
Contact capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Selectable from among the upper limit alarm, lower limit alarm,
during transmission output hold, and cleaning output
R1, R2 (closed at alarm operation, open under normal conditions and in a
Contact function power-off state)
Setting range: Within the measurement range of turbidity or
Contact output absorbance
Delay time: 0 s to 600 s
Contact type Relay contact, SPST (1a)
Contact capacity 30 V DC 1 A (resistance load)
Selectable from among the upper limit alarm, lower limit alarm,
during transmission output hold, and cleaning output.
R3 (closed at alarm operation, open under normal conditions and in a
Contact function power-off state)
Setting range: Within the measurement range of turbidity or
absorbance
Delay time: 0 s to 600 s

56
Data

Contact type Relay contact, SPDT (1c)


Contact capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
The relay coil is energized when the system is normal.
FAIL You can use the NO and NC relay contact.
Contact function Error alarms for values outside the measurement range, self-checks,
Contact output and calibration errors can be set
Delay time: 0 s to 600 s
Contact type Relay contact, SPST (1a)
RANG1,
Contact capacity 30 V DC 1 A (resistance load)
RANG2
Contact function Status output of the transmission output range
Number of contact inputs 4
Contact type Open collector no-voltage “a” contact
ON resistance: 100 max.
Condition Open-circuit voltage: 24 V DC
Contact input Short-circuit current: 12 mA DC max.
EXT1 Hold command
Contact EXT2 Cleaning command The minus terminals of the four
function Selection command from four inputs are connected inside
EXT3, EXT4
ranges using 2-bit contacts
Type RS-485
Communication Two wire type, the power source of the communication is isolated
function Signal form from the power of measurement circuit. (The power source of
transmission and communication are not isolated.)
Zero calibration: By filtered clean water
Calibration method
Span calibration: Turbidity adjustment by span coefficient input
Calibration
Compatible standard
Kaolin, Formazin
substances
Shift correction: Shift correction to the measurement value
Correction Correction method Gain correction: Gain correction by coefficient to the measurement
value
Number of output One (For the water/air jet cleaning assistant)
AC voltage output (Output voltage is the same as the power for the
Output type
converter)
Contact type Relay contact, SPST (1a)
Contact capacity 250 V AC 0.5 A
Cleaning cycle 0.2 h to 168.0 h
Cleaning operation
output Setting Cleaning time 30 s to 600 s
(For additional Hold time 30 s to 600 s
assistant cleaning)
Timer accuracy Within 2 min per month
One of the following four methods to start cleaning is selectable
By the internal timer
By the internal timer or external cleaning command input
Cleaning start condition
By the internal timer while the external permission signal is input
By the input of cleaning trigger (Cleaning input over 2 seconds
triggers off the cleaning)
Sensor check error Sensor error
Self-check
Converter error CPU abnormality, memory abnormality
Operating
20 C to 55 C
temperature range
Operating humidity
Relative humidity: 5 to 90 (no condensation)
range

57
Data

Storage temperature 25 C to 65 C
Rated power supply
100 V to 240 V AC ±10% 50/60 Hz
voltage
Power supply
Power consumption 36 VA max with 100 V AC power supply
Power switch A switch is installed inside the converter.
EMC :EN61326-1*
Class A, Industrial electromagnetic environment
Compatible CE marking Safety :EN61010-1
standards RoHS : EN50581
9. Industrial monitoring and control instruments
FCC rules Part 15 Class A
Installation Outdoor installation type
Installation method 50 A pole-mounted or wall-mounted
Protection class IP65
Material of case Aluminum alloy (coated with epoxy-modified melamine resin)
Structure Material of mounting
SUS304
brackets
Material of hood SUS304 (coated with epoxy-modified melamine resin)
Material of display window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) 155 (H) 115 (D) mm (excluding brackets)
Mainframe: approx. 3.5 kg
Mass
Hood, bracket: approx. 1 kg
: The surge test specified in the EMC Directive for CE Marking shall not apply to the case when a sensor cable,
transmission cable, or contact input cable is extended for 30 m or longer.

Note
Although arrestors (400 V discharge starting voltage) are provided for the contact input and
communication lines, install some optimum surge absorber elements on the connection lines,
depending on the surrounding environment, installation conditions, and the external devices.

58
Data

Detector
Product name Industrial-use turbidity/SS detector
Model SS-150
Measurement principle Right angle scattering/transmission method or transmission method
Light source Infrared LED 860 nm
Detector Silicon photo diode
Window of light path PFA
Cell length 30 mm I.D.
Data transfer RS-485 (communication with converter)
Rubber blades wipe optical wall of the sensor head. (Equipped in the
detector)
Actuated by stepping motor
Cleaning function Wiper will rotate while the cleaning command from the converter is
input.
Wiper will stop at the measuring position after cleaning command is
cancelled.
Sample water temperature 0 C to 45 C (no freezing)
Storage temperature 25 C to 65 C
Sample water pressure 0 MPa to 0.1 MPa
Material of liquid contact PFA, POM, SUS316, FKM, EPDM, PVC, PPO, SUS304
Provided standard cable: 10 m, maximum extension distance: 50 m
Cable length
total 60 m.
Power supply 24 V DC 6 W supplied from HU-200TB-IM converter
External dimensions 60 L250 mm (excluding cables)
Mass Approx. 1.0 kg (excluding cables)
Conformation JIS K0101 ISO7027 calibration with Formazin

59