Professional Documents
Culture Documents
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
D6886 − 14´1
Pigmented Coatings NOTE 3—Methanol can be used as a first choice as a solvent for all
D3960 Practice for Determining Volatile Organic Compound waterborne coatings. THF can be used for solventborne coatings. Acetone
may also be used for solventborne coatings but should not be used for
(VOC) Content of Paints and Related Coatings waterborne coatings because it may react with ammonia and amines which
D4017 Test Method for Water in Paints and Paint Materials are frequently found in waterborne coatings. Other solvents can be used if
by Karl Fischer Method needed but the choice of solvent should be reported.
D6133 Test Method for Acetone, p-Chlorobenzotrifluoride,
4.2 Direct GC/FID, GC/MS and solid phase microextraction
Methyl Acetate or t-Butyl Acetate Content of Solvent-
/ gas chromatography (SPME/GC) of the coating may be used
borne and Waterborne Paints, Coatings, Resins, and Raw
to facilitate identification of the volatile compounds present in
Materials by Direct Injection Into a Gas Chromatograph
a coating (Note 4). Table X1.1 lists the GC retention times for
D7358 Test Method for Water Content of Paints by Quanti-
tative Calcium Hydride Reaction Test Kit some of the volatile compounds which may be found in low
E177 Practice for Use of the Terms Precision and Bias in volatile organic compound content air-dry coatings and for
ASTM Test Methods several possible internal standards, ordinarily not present in
E691 Practice for Conducting an Interlaboratory Study to coatings, which may be used (Note 4).
Determine the Precision of a Test Method NOTE 4—The analyst should consult MSDS and product data sheets for
information regarding solvents which are expected in a particular coating.
2.2 Other Documents: Additional solvents, not shown on the MSDS or PDS may also be present
EPA Method 24 —Determination of Volatile Matter Content, in the coating. Retention times given in Appendix X1 must be verified for
Water Content, Density, Volume Solids, and Weight Solids each individual instrument.
of Surface Coatings NOTE 5—The accuracy of the volatile organic compound weight
40 CFR 51.100 (s) List of components that EPA has classi- percent determined using Test Method D6886 is dependent on the proper
fied as VOC-exempt identification of the compounds detected in the chromatogram. The
response of the flame ionization detector (FID) used in the GC is
3. Terminology dependent on the compound detected. The accuracy of the determination
requires proper identification (by GC/MS, by retention time, or by
3.1 Acronyms: analyzing the sample on a GC column with a different stationary phase)
3.1.1 EGDE—ethylene glycol diethyl ether and calibration of the GC for the compounds detected.
3.1.2 DB—2-(2-butoxyethoxy)ethanol; Butyl Carbitol;3 di-
ethylene glycol monobutyl ether 5. Significance and Use
3.1.3 EB—2-butoxyethanol; Butyl Cellosolve;4 ethylene 5.1 In using Practice D3960 to measure the volatile organic
glycol monobutyl ether compound content of waterborne coatings, precision can be
3.1.4 EG—ethylene glycol poor for low volatile organic compound content air-dry coat-
ings if the volatile organic weight percent is determined
3.1.5 FID—flame ionization detector indirectly. The present method directly identifies and then
3.1.6 GC—gas chromatograph quantifies the weight percent of individual volatile organic
3.1.7 PG—propylene glycol compounds in air-dry coatings (Note 6). The total volatile
3.1.8 SPME—solid phase microextraction organic weight percent can be obtained by adding the indi-
vidual weight percent values (Note 7).
3.1.9 SPDE—solid phase dynamic extraction
NOTE 6—The present method may be used to speciate solvent-borne
3.1.10 TMPD-IB—2,2,4-trimethypentane-1,3-diol, air-dry coatings. However, since these normally contain high, and often
monoisobutyrate complex, quantities of solvent, precision tends to be better using other
methods contained in Practice D3960, where the volatile fraction is
3.1.11 TMPD-DIB—2,2,4-trimethypentane-1,3-diol, di- determined by a direct weight loss determination.
isobutyrate NOTE 7—Detectable compounds may result from thermal decomposi-
3.1.12 VOC—volatile organic compound used in various air tion in a hot injection port or from reaction with the extraction solvent. If
quality regulations it can be shown that a material is a decomposition product, or is the result
of a reaction with the extraction solvent, then results for that compound
4. Summary of Test Method should be discounted from the volatile measured by Test Method D6886.
2
D6886 − 14´1
6.2 Standard GC/FID and GC/MS Instrument Conditions:
NOTE 8—Some coatings may contain high-boiling components which 8. Column Conditioning
elute from the GC capillary column after the specified run time of 20 min.
It is advisable, therefore, to bake out the column between runs in these 8.1 The capillary columns should be conditioned according
cases. to the manufacturer’s recommendation. The columns may then
be used indefinitely without further conditioning.
7. Reagents and Materials
9. Preparation of Standards
7.1 Purity of Reagents—Reagent grade chemicals shall be
9.1 Prepare a stock mixture of ethylene glycol (EG), pro-
used in all tests. Unless otherwise indicated, all reagents shall
pylene glycol (PG), ethylene glycol monobutyl ether (EB),
conform to the available specifications of the Committee on ethylene glycol diethyl ether (EGDE) [or other suitable internal
Analytical Reagents of the American Chemical Society. Other standard], diethylene glycol monobutyl ether (DB), and 2,2,4-
grades may be used, provided it is first ascertained that the trimethylpentane-1,3-diol monoisobutyrate (TMPD-IB) by
reagent is of sufficiently high purity to permit its use without weighing one or two grams of each into an appropriate vial.
lessening the accuracy of the determination. The weight of each component should be approximately the
7.2 Carrier Gas, helium of 99.995 % or higher purity. same and determined to 0.1 mg. Mix the contents.
9.2 Transfer approximately 100 µL of the stock mixture to a
7.3 Tetrahydrofuran (THF), HPLC grade.
septum-capped vial containing 10 mL of THF or methanol and
7.4 Methanol, HPLC grade. mix the contents (Note 9). This solution will contain each of
the known analytes at a concentration of approximately 2
7.5 Possible internal standards: 1-Propanol, mg/mL.
p-fluorotoluene, cyclohexanol, p-chlorotoluene, ethylene glycol NOTE 9—The solvents EG, PG, EB, DB TMPD-IB are widely used in
diethylether (EGDE). the manufacture of waterborne air-dry coatings and may be expected as
probable components of these coatings.
7.6 Fluorocarbon-faced Septum Vials, 20 mL and 40 mL
9.3 Chromatograph the solution in 9.2 by injecting 1 µL into
capacity.
the PMPS column using the chromatographic conditions given
7.7 Ceramic Beads, 0.5–1.0 mm diameter. in 6.2. Calculate the relative response factors for each of the
3
D6886 − 14´1
analytes relative to the EGDE or other suitable internal where:
standard using the relationship: X = one of several possible volatile compounds in the
AA*MI coating,
RF 5 (1) RF = relative response factor of compound X,
AI*MA
AA = peak area of compound X,
where: MI = weight of internal standard,
RF = relative response factor, AI = peak area of internal standard, and
AA = area of analyte, MC = weight of coating.
MI = weight of internal standard (from 9.1), NOTE 11—If volatile compounds other than those in the standard (9.1)
AI = area of internal standard, and are present in the coating, the identity should be confirmed by FID
MA = weight of analyte (from 9.1). retention time comparison with standard material or by GC/MS and the
relative response factor should be determined as outlined in 9.1 – 9.3.
10. Paint Analysis Commercial 2,2,4-trimethylpentane-1,3-diol monoisobutyrate (TMPD-
IB) may contain small amounts of 2,2,4-trimethylpentane-1,3-diol which
10.1 Analysis of Air-dry Solvent-borne and Waterborne elutes approximately 0.5 minutes before butyl carbitol and 2,2,4-
Coatings by GC/FID: trimethylpentane-1,3-diol diisobutyrate (TMPD-IB) which elutes approxi-
10.1.1 Prepare duplicate samples by pipetting 10 mL of mately 1.5 minutes after 2,2,4-trimethylpentane-1,3-diol monoisobutyrate
methanol (waterborne coatings) or THF (solventborne coat- (TMPD-IB). Acetone and isopropyl alcohol have nearly the same reten-
tion time on a PMPS column and if either is found, their identities should
ings) into a vial containing 3 to 5 g of ceramic beads and close be confirmed and quantitated on a Carbowax5 column or by using GC/MS.
with a fluorcarbon-faced septum cap. Using a dedicated glass Isobutyl alcohol coelutes with the solvent (THF) and must be determined
syringe (25 or 50 microliter capacity), add 10 µL of EGDE or on a different column (Carbowax) or using a different solvent (methanol).
other internal standard and weigh to at least 0.1 mg. This SPME, SPDE and static headspace analysis are especially useful tech-
solution must be analyzed by GC to determine if there are niques for confirming that decomposition products are not being observed.
Small quantities (up to 0.5 %) of acetic acid are sometimes found in
peaks that result from it rather than from the paint sample that coatings containing vinyl acetate resins. The acetic acid is formed as a
is prepared in 10.1.2. decomposition product in the GC inlet and should not be counted as a
10.1.2 Pipette 10 mL of methanol or THF into a 20 or 40 mL volatile organic compound. Some coatings contain additives (for example,
vial containing 3 to 5 g ceramic beads and close with a carbamate ester biocides) that may give decomposition products in the hot
fluorocarbon-faced septum cap. Using a disposable 1 mL inlet of the gas chromatograph. If decomposition products are suspected,
a convenient procedure for determining this is to analyze the coating by
syringe, add approximately 0.6 to 0.8 g of the well-mixed paint static headspace gas chromatograhpy. In using static headspace, a large
through the septum cap and weigh to 0.1 mg (Note 10). Using sample of the coating (15 to 20 g) is internally standardized with 10 mg/g
the dedicated syringe, add 10 microliters of pure EGDE (or of EGDE, approximately 5 mL of ceramic beads are added, and manually
other internal standard) through the septum and weigh the mixed by shaking until the paint/internal standard mixture is homoge-
amount added to at least 0.1 mg. Mix the contents vigorously neous. The static headspace procedure is carried out on 40 to 60 mg of the
internally standardized coating using a 20 mL crimp-cap headspace vial.
by shaking for 1 min. Let the vial stand to permit pigments, if Static headspace conditions are: Oven, 20 to 30 minutes at 130°C; Loop,
any, to settle. 150°C; Transfer Line, 150°C. Chromatographic Conditions: as described
in 6.2. Alternatively, if static headspace, SPME, or SPDE capability are
NOTE 10—The paint should be drawn into the syringe without an
attached syringe needle. Excess paint is wiped from the syringe and the not available, the analysis can be done using a lower inlet temperature as
needle is then attached for paint transfer. The mass of the paint may be long as the selected temperature is high enough to fully volatilize the
determined by either the difference in the weight of the filled and empty suspect compounds. Cool on-column injection can also be used to
syringe or by the difference in the weight of the vial before and after determine if a compound is being generated vial thermal decomposition in
adding paint. the hot GC inlet.
10.1.3 Chromatograph the solution in 10.1.2 by injecting 1 10.2 Overlapping Chromatographic Peaks:
µL into the PMPS capillary column using the standard condi- 10.2.1 A number of organic compounds in solvent-borne
tions described in 6.2. If necessary, adjust the split ratio to give paints containing commercial xylene can overlap. These in-
well-defined chromatographic peaks. Identify the volatile com- clude propylene glycol monomethyl ether acetate overlapping
pounds which elute over a 20 minute run time. An optional, with ethylbenzene and butoxyethanol with o-xylene. Resolu-
late-eluting compound, such as methyl palmitate (retention tion can generally be obtained by simply changing the chro-
time of 18.4 min) may be used to verify column performance matographic heating rate.
and retention times. Note that methyl palmitate is not a marker
to determine volatile organic compound/non-volatile organic 10.3 Coatings Containing Silanes, Siloxanes and Silane-
compound status of eluted compounds. Calculate the weight Siloxane Blends:
fraction of each peak using the relationship: 10.3.1 If the coating contains silanes, siloxanes, and silane-
siloxane blends, approximately 50 mg of solid
~ AA!~ MI!~ 100!
%X 5 (2) p-toluenesulfonic acid should be added to the solution in 10.1.2
~ AI!~ RF!~ MC! thirty minutes prior to gas chromatography. The
p-toluenesulfonic acid catalyzes the hydrolysis of alkoxy
silanes to free alcohol (usually ethanol).
5
Carbowax is a registered trademark of The Dow Chemical Company.
4
D6886 − 14´1
11. Reporting Results 13.1.4 95 % Repeatability Limit (within laboratory)—The
11.1 Prepare a table (as indicated below) which contains within-laboratory coefficient of variation is 2.7 % relative. The
information on each of the volatile organic compounds found. 95 % confidence limit for the difference between two averages
Report the identity of the solvent used. Report the split ratio is 7.5 % of the test result.
used if it deviates from 50:1. 13.1.5 95 % Reproducibility Limit (between laboratories)—
The between-laboratory coefficient of variation is 5.8 % rela-
Volatile Organic Compound GC Retention Time Weight % Found
tive. The 95 % confidence limit for the difference between two
averages is 16.2 % of the test result.
13.2 A second interlaboratory study of total weight percent
Total weight percent of all speciated volatile organic com- volatile organic compounds was conducted in accordance with
pounds =. Practice E691. Eleven laboratories participated in the study,
11.2 List volatile organic compounds that are not identified testing twelve coatings (five containing <2 wt% volatile and
as unknown (UK) and use the relative response factor for seven containing <0.21 wt% volatile). Each of the laboratories
2,2,4-trimethylpentane-1,3-diol monoisobutyrate to calculate analyzed the coatings twice according to protocols specified in
the weight % for these unknowns. Practice E691.
13.2.1 In this study the samples were analyzed “blind,” that
12. Alternate Identification Methods is, the participating laboratories had no knowledge of the
12.1 The use of GC/MS for volatile compound identifica- volatile organic compounds contained in each of the coatings
tion is highly desirable even when quantitation is carried out by and followed Test Method D6886–12.
GC/FID. A convenient procedure is to sample the headspace of 13.2.2 The precision statement was determined through
the coating using an SPME or SPDE followed by thermal statistical examination of 264 results, from a total of eleven
desorption onto any standard capillary column and subsequent laboratories, on twelve coatings. The coatings were designated
mass spectral identification. This technique is especially valu- in the study as:
able for identifying oxygenates, aromatics and other volatile (1) Paints containing <0.21 wt% Volatile — cans 1-16,
organic compounds. 17-32, 49-64, 97-112, 129-144, 145-160, 161-176
(2) Paints containing <2 wt% Volatile — cans 33-48,
13. Precision and Bias6 65-80, 81-96, 113-128, 177-192
13.1 Interlaboratory Studies—Two interlaboratory studies 13.2.3 Precision statistics were calculated for the total
have been carried out (Note 12). weight percent volatile found in each of the five coatings and
13.1.1 An interlaboratory study of the total weight percent are presented in Table X1.3 and Table X1.4. The terms
volatile organic compounds was conducted in accordance with repeatability limit and reproducibility limit are used as speci-
Practice E691 in seven laboratories with five materials, with fied in Practice E177.
each laboratory obtaining three test results for each material. 13.2.4 95 % Repeatability Limit (within laboratory)—The
Five commercial waterborne coatings, (a primer, a flat, a satin, 95 % confidence limit for the difference between two averages
a semi-gloss and a gloss), ranging in weight percent volatile is 28.3 % for test results on paints containing <0.21 wt%
content from 0.25 to 4.50, were analyzed. Each of the volatile and 20.0 % for test results on paints containing <2 wt%
laboratories analyzed the coatings three times according to volatile.
protocols specified in Practice E691. 13.2.5 95 % Reproducibility Limit (between laboratories)—
13.1.2 In this study the participating laboratories knew the The 95 % confidence limit for the difference between two
identities of the volatile organic compounds contained in each averages in 258 % for test results on paints containing <0.21
of the coatings and followed Test Method D6886-03. wt% volatile and 93.9 % for test results on paints containing <2
13.1.3 Precision statistics were calculated for the total wt% volatile.
weight percent volatile organic compounds found in each of 13.3 Bias—Bias has not been determined.
the five coatings and are presented in Table X1.2. The terms NOTE 12—The precision statistics determined for this test method are
repeatability limit and reproducibility limit are used as speci- applicable only to gas chromatography using flame ionization detection.
fied in Practice E177. Statistics using mass spectral detection have not been determined.
14. Keywords
6
Supporting data have been filed at ASTM International Headquarters and may
be obtained by requesting Research Report RR:D01-1178. Contact ASTM Customer 14.1 gas chromatography; speciation; volatile organic com-
Service at service@astm.org. pounds; waterborne coatings
5
D6886 − 14´1
APPENDIXES
(Nonmandatory Information)
6
D6886 − 14´1
TABLE X1.1 Continued
Compound RRF PMPS Carbowax PDMS
p-cymene 2.13 10.84 15.62 10.67
benzyl alcohol 1.65 10.89
N-methylpyrrolidinone 0.72 10.96 12.18 10.94
ethylene glycol butyl
ether acetate 11.25 17.73
diethylene glycol monopropyl 11.85
ether (DP) 11.34 14.8
diethylene glycol monopropyl
ether (DPnP) 0.76 11.79, 11.82
2,2,4-trimethylpentane-1,3-diol
diethylene glycol monobutyl
ether (DB) 0.93 12.12
naphthalene 2.03 12.43
2-(2-ethylhexyl)ethanol 12.51
triethylene glycol 0.58 12.53
dipropylene glycol monobutyl
ether (DPnB) 1.04 12.53, 12.58 18.86 13.41
propylene glycol monophenyl
ether 1.14 12.67
2,2,4-trimethylpentane-1,3-diol
monoisobutyrate (TX) 1.27 13.62, 1374
tetraethylent glycol 0.59 14.54
2,2,4-trimethylpentane-1,3-diol,
diisobutyrate (TXIB) 1.17 15.15
2-ethylhexyl benzoate 1.39 16.05
methyl palmitate 18.40
dibutyl phthalate 1.16 18.50
Mineral Spirits 2.23 9 to 11
Aromatic100 2.10 9.2 to 10.6
TABLE X1.2 Precision Statistics for Low Volatile Content Waterborne Air-Dry Coatings Analyzed by Gas Chromatography
Total Volatile in
Repeatability Standard
Material Weight Percent Reproducibility Standard Deviation Repeatability Limit
Deviation
Average
A 0.25 0.006 0.020 0.03
B 2.85 0.058 0.125 0.16
C 4.04 0.087 0.187 0.24
D 4.49 0.145 0.314 0.41
E 4.50 0.100 0.215 0.28
TABLE X1.3 Precision Statistics for Waterborne Air-Dry Coatings Containing <0.21 wt% Volatile Analyzed by Gas Chromatography
Total Volatile in
Repeatability Standard Reproducibility Standard
Material Weight Percent Repeatability Limit Repeatability Limit
Deviation Deviation
AverageA
Can labeled: 33–48 1.84329 0.047305 0.592178 0.132455 1.658097
Can labeled: 65–80 1.141918 0.128753 0.559168 0.360507 1.565670
Can labeled: 81–96 1.352836 0.084883 0.636784 0.237671 1.782996
Can labeled: 113–128 1.288332 0.053950 0.287157 0.151060 0.804041
Can labeled: 177–192 1.52877 0.194856 0.322530 0.545596 0.903084
Average 0.285458 1.342778
A
The average of the laboratories’ calculated averages.
TABLE X1.4 Precision Statistics for Waterborne Air-Dry Coatings Containing <0.21 wt% Volatile Analyzed by Gas Chromatography
Total Volatile in
Repeatability Standard Reproducibility Standard
Material Weight Percent Repeatability Limit Repeatability Limit
Deviation Deviation
AverageA
Can labeled: 1–16 0.040855 0.007284 0.0050469 0.020395 0.14132
Can labeled: 17–32 0.045027 0.007845 0.038091 0.021965 0.106655
Can labeled: 49–64 0.143773 0.016751 0.176590 0.046903 0.494451
Can labeled: 97–112 0.067545 0.012431 0.056781 0.034808 0.158987
Can labeled: 129–144 0.216077 0.009180 0.155209 0.025705 0.434586
Can labeled: 145–160 0.098409 0.008367 0.092197 0.023427 0.258151
Can labeled: 161–176 0.02973 0.002291 0.016000 0.006414 0.044799
Average 0.02566 0.234134
A
The average of the laboratories’ calculated averages.
7
D6886 − 14´1
X2. SUBSEQUENT TREATMENT OF TEST METHOD D6886 DATA FOR VOLATILE ORGANIC COMPOUND
DETERMINATION
TABLE X2.1 Estimated Uncertainties in Coating Volatile Organic Compound Based on Published Precision Statements Assuming 1200
g/L Paint Density and 50 % Water Content and No Exempt Compounds
Indirect Determinations (EPA Method 24)
Theoretical Coating
Theoretical Volatile Organic Theoretical Volatile Organic
(regulatory)
Compound (g/L) Range using Compound (g/L) Range using
Volatile Organic
Intralaboratory Uncertainties Interlaboratory Uncertainties
Compound/(g/L)
~1 1 ± 54 1 ± 55
~50 50 ± 45 50 ± 129
~100 100 ± 44 100 ± 126
Direct Determination (Test Method D6886-03)
Theoretical Coating
Theoretical Volatile Organic Theoretical Volatile Organic
(regulatory)
Compound (g/L) Range using Compound (g/L) Range using
Volatile Organic
Intralaboratory Uncertainties Interlaboratory Uncertainties
Compound/(g/L)
~50 50 ± 4 50 ± 9
~100 100 ± 8 100 ± 18
Direct Determination (Test Method D6886-12)
Theoretical Coating
Theoretical Volatile Organic Theoretical Volatile Organic
(regulatory)
Compound (g/L) Range using Compound (g/L) Range using
Volatile Organic
Intralaboratory Uncertainties Interlaboratory Uncertainties
Compound/(g/L)
~0 0±1 0 ± 13
~50 50 ± 6 50 ± 36
8
D6886 − 14´1
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/