You are on page 1of 8

Clinical Chemistry 59:10 Point-of-Care Testing

1506–1513 (2013)

Simple Paper-Based Test for Measuring Blood Hemoglobin


Concentration in Resource-Limited Settings
Xiaoxi Yang,1 Nathaniel Z. Piety,1 Seth M. Vignes,1 Melody S. Benton,2 Julie Kanter,2,3
and Sergey S. Shevkoplyas1*

BACKGROUND: The measurement of hemoglobin con- Hemoglobin (Hb) is the oxygen-transporting protein
centration ([Hb]) is performed routinely as a part of a contained within red blood cells (RBCs). The mass
complete blood cell count to evaluate the oxygen- concentration of Hb ([Hb]) is a measure of the oxygen-
carrying capacity of blood. Devices currently available carrying capacity of blood. The reference ranges of
to physicians and clinical laboratories for measuring [Hb] in blood of adults are 12–16 g/dL in women and
[Hb] are accurate, operate on small samples, and pro- 14 –18 g/dL in men; values ⬍5 g/dL or ⬎20 g/dL are
vide results rapidly, but may be prohibitively expensive considered critical (1 ). A decreased [Hb], referred to as
for resource-limited settings. The unavailability of ac- anemia, may be an indication of chronic blood loss,
curate but inexpensive diagnostic tools often precludes decreased RBC production, or abnormal destruction of
proper diagnosis of anemia in low-income developing RBCs (hemolysis). Hemolytic anemia may be associ-
countries. Therefore, we developed a simple paper- ated with inherited blood disorders such as sickle cell
based assay for measuring [Hb]. disease, but can also be autoimmune in origin or result
from exposure to certain drugs, toxins, infections, or
METHODS: A 20-␮L droplet of a mixture of blood and
transfusion of mismatched blood products (2 ). Other
Drabkin reagent was deposited onto patterned chro-
common causes of anemia include dietary deficiencies
matography paper. The resulting blood stain was digi-
(3, 4 ), cirrhosis, and renal disease (2 ). Anemia reduces
tized with a portable scanner and analyzed. The mean
the oxygen-carrying capacity of blood, straining the
color intensity of the blood stain was used to quantify
[Hb]. We compared the performance of the paper- cardiopulmonary system, which responds by increas-
based Hb assay with a hematology analyzer (compari- ing the heart rate and stroke volume to maintain oxy-
son method) using blood samples from 54 subjects. gen delivery (5 ). Acute-onset anemia may be associ-
ated with fatigue, lethargy, and/or pallor (4 ). Patients
RESULTS: The values of [Hb] measured by the paper- with chronic anemia and having [Hb] ⬍5 g/dL may
based assay and the comparison method were highly experience more serious complications such as angina,
correlated (R2 ⫽ 0.9598); the standard deviation of the congestive heart failure, heart attack, and stroke (5 ).
difference between the two measurements was 0.62 Depending on the clinical status of the patient, a [Hb]
g/dL. The assay was accurate within 1 g/dL 90.7% of the ⬍6 – 8 g/dL is currently recommended as the threshold
time, overestimating [Hb] by ⱖ1 g/dL in 1.9% and trigger for initiating RBC transfusion (6 ). Increased
underestimating [Hb] by ⱖ1 g/dL in 7.4% of the [Hb] may indicate erythrocytosis due to smoking (7 )
subjects. or in response to hypoxia at high altitudes (8 ), polycy-
themia vera (9 ), congenital heart disease (10 ), severe
CONCLUSIONS: This study demonstrates the feasibility of chronic obstructive pulmonary disease (11 ), or severe
the paper-based Hb assay. This simple, low-cost test dehydration (12 ). The increase in blood viscosity asso-
should be useful for diagnosing anemia in resource- ciated with an increased [Hb] may impair oxygen de-
limited settings, particularly in the context of care for livery by reducing the capillary perfusion, which may in
malaria, HIV, and sickle cell disease patients in sub- turn increase the risk of hypoxia, as well as thrombotic
Saharan Africa. and hemorrhagic complications (13 ).
© 2013 American Association for Clinical Chemistry
Because of its importance, [Hb] is measured rou-
tinely as a part of a complete blood cell count. Clinical

1
Department of Biomedical Engineering, Tulane University, New Orleans, LA; 2 Sickle Received February 7, 2013; accepted June 11, 2013.
Cell Center of Southern Louisiana and 3 Department of Pediatrics, Section of Previously published online at DOI: 10.1373/clinchem.2013.204701
4
Hematology/Oncology, Tulane University School of Medicine, New Orleans, LA. Nonstandard abbreviations: Hb, hemoglobin; RBC, red blood cell; POC,
* Address correspondence to this author at: Tulane University, Department of point-of-care; HbCS, Hb color scale; ␮PAD, microfluidic paper-based ana-
Biomedical Engineering, 500 Lindy Boggs Building, New Orleans, LA 70118. lytical device; RGB, red/green/blue; LOD, limit of detection; LOQ, limit of
E-mail shevkop@tulane.edu. quantification.

1506
Paper-Based Hb Test for Resource-Limited Settings

laboratories measure [Hb] with automated hematol- FABRICATION OF THE MICROFLUIDIC PAPER-BASED
ogy analyzers as well as spectrophotometers, blood gas ANALYTICAL DEVICES
analyzers, and stand-alone CO-oximeters (14 ). The The microfluidic paper-based analytical devices
need for rapid and accurate measurement of [Hb] at (␮PADs) were fabricated by a previously published
the bedside has driven the development and imple- method (26, 27 ). Briefly, the pattern of the ␮PADs was
mentation of [Hb] testing on the point-of-care (POC) designed with illustration software (Canvas 11, ACD
platform (15 ). Currently available POC devices are de- Systems International). The devices were printed onto
signed to operate in close proximity to the patient, pro- sheets of chromatography paper (No. 1, Whatman)
vide accurate results (within 1 g/dL of clinical labora- with a solid-ink (wax) printer (Phaser 8560N, Xerox),
tory analyzers), require small volumes of blood, and be and then heated on a hot plate at 150 °C for 3 min.
used by persons without expert training, making the
nearly instantaneous measurement of [Hb] in operat- QUANTIFICATION OF THE BLOOD STAIN COLOR
ing room or emergency care settings a reality (16 –18 ). Blood samples were mixed at 1:10, 1:15, or 1:20 ratios
Notwithstanding their obvious advantages, the (vol/vol) with Drabkin reagent (Ricca Chemical Co.)
relatively high per-test and analyzer costs make the use and incubated at room temperature (20 –25 °C) for 10
of conventional POC devices for measuring [Hb] pro- min. The components of Drabkin reagent lysed red
hibitively expensive in resource-limited settings of low- blood cells and reacted with all forms of Hb except
income developing countries in sub-Saharan Africa, on sulfhemoglobin present in the sample, converting
the Indian subcontinent, and throughout Southeast them to cyanmethemoglobin, a stable brownish-
Asia (19, 20 ). Currently available low-cost alternatives colored compound. A 20-␮L drop of the mixture was
including the “hematocrit/3” method (21 ) and the placed onto the center of each ␮PAD. The resulting
WHO Hb color scale (HbCS) test (22 ) have several blood stain was allowed to dry for 25 min (28 ).
drawbacks that limit their use in the field. In addition Quantification of [Hb] was accomplished by scan-
to being somewhat inaccurate, hematocrit/3 relies on ning the sheets of chromatography paper containing
knowledge of hematocrit and, therefore, requires a arrays of ␮PADs on a portable flatbed scanner (Ca-
centrifuge (21, 23 ). The HbCS test is highly sensitive to noScan LiDE110, Canon USA) and then analyzing the
the ambient lighting conditions, operator bias, and de- images automatically with a custom-coded algorithm
viations from the recommended readout time (22, 24 ). (MATLAB, The Math Works). The quantitative analy-
The lack of confidence in the quality of [Hb] measure- sis of blood stains was based on the red/green/blue
ments may result in clinicians relying exclusively on (RGB) color model values of the digitized images. The
their clinical judgment to prescribe transfusions in color data in the red, green, and blue channels were
resource-limited settings (25 ). Therefore, our goal was extracted from the RGB values of each pixel within a
to develop a simple, low-cost, paper-based Hb assay blood stain. We defined the color intensity of each
that addressed these limitations. channel as 255 – (R, G, or B); e.g., for red channel, color
intensity ⫽ 255 – R. The mean color intensity of the
Methods blood stain for each color channel was calculated from
the color values of all pixels within the blood stain,
BLOOD SAMPLES including the pixels corresponding to the darker ring
The study protocol was approved by Tulane University on the periphery of the stain. The correlation between
Biomedical Institutional Review Board. After written the mean color intensity and the actual [Hb] was eval-
informed consent was obtained, blood samples were uated for all color channels; the green channel showed
collected into 4- or 9-mL Vacutainer tubes (K2EDTA, the best linear fit and was selected to quantify [Hb] in
BD) during routine blood draws (January–April 2013) the blood samples.
from patients of the Pediatric Hematology-Oncology
Clinic (Tulane University Hospital) and the Sickle Cell STATISTICAL ANALYSIS
Center of Southern Louisiana (New Orleans, LA). We Pearson correlation coefficient and regression were
measured standard hematological parameters includ- calculated to directly compare data obtained from the
ing [Hb] using a Medonic M-Series hematology ana- paper-based Hb assay and the Medonic M-Series he-
lyzer (Medonic M16C/M20C, Boule Medical). The Hb matology analyzer (comparison method). Three re-
content of blood samples was adjusted artificially to search associates performed both types of measure-
prepare samples with [Hb] ranging from 2.5 to 25 g/dL ments on the same day. A Bland–Altman plot (29 ) was
by adding autologous plasma (for lower [Hb]) or pel- also constructed, with consideration for limitations, to
leted RBCs (for higher [Hb]) to the original sample. visualize the comparison.

Clinical Chemistry 59:10 (2013) 1507


melting point of wax to allow the wax to reflow and
form hydrophobic barriers through the full thickness
of the paper (Fig. 1B). When designing the pattern of
the ␮PAD (Fig. 1A), we accounted for the natural wid-
ening of the printed lines (to about 1 mm thick) during
the melting process (Fig. 1B).
We performed this paper-based Hb assay by first
mixing a sample of whole blood with Drabkin reagent
and then depositing a 20-␮L droplet of the mixture
onto the ␮PAD and letting the blood stain develop
(Fig. 1C). The droplet spread radially from the center
through the paper substrate toward the periphery of
the ␮PAD due to wicking by capillary action (30, 31 ),
forming the characteristic blood stain pattern (Fig. 1D)
reminiscent of the stains produced by drying coffee
(32 ). We selected the volume of the droplet (20 ␮L)
and size of the ␮PAD (2.8 cm) so that the outermost
margin of the stain could not reach the periphery of the
␮PAD (Fig. 1D).
Fig. 1. Design and fabrication of the paper-based Hb We used a portable flatbed scanner to digitize the
assay. sheets of chromatography paper carrying ␮PADs with
(A), The pattern of the ␮PAD was printed with a solid ink
developed blood stains. We analyzed the images to de-
(wax) printer on chromatography paper. (B), The printed
termine the mean color intensity for the blood stain
pattern was melted to create a 1-mm-wide hydrophobic
contained within each ␮PAD using a simple image
barrier spanning the entire thickness of the paper sub-
processing algorithm developed specifically for this
strate. (C), To perform the assay, a drop of blood mixed
purpose (Fig. 1, D and E). The algorithm used the
with Drabkin reagent was placed onto the center of the
alignment mark on the sheets of paper to determine the
device. The lysate wicked laterally toward the periphery,
location of each ␮PAD (Fig. 1D). The algorithm ap-
coloring the paper faint red (pink). (D), The image analysis
plied a circular binary mask to the image to select the
algorithm automatically detected the blood stain within
area within the hydrophobic border of the ␮PAD, and
each ␮PAD (dashed circle) and extracted the RGB color
then isolated the pixels of the blood stain within the
information for all pixels contained within the blood stain.
circular region by removing pixels below a threshold
(E), The color intensity of the pixels (between dashed lines)
value on the basis of the difference in color intensity
was used to calculate the mean color intensity (solid red
between the paper substrate and the blood stain (Fig.
line) for each blood stain. The values of the mean color
1D, dashed circle). Finally, the algorithm used the color
intensity were used to calculate the [Hb] in the blood
information from all pixels within the blood stain to
samples. au, Arbitrary unit.
calculate the mean color intensity of the blood stain
(Fig. 1E, red solid line). We used the mean color inten-
sity of the stain (Fig. 1E) and a calibration curve (Fig.
2B) to measure [Hb] in the blood sample.
Results All operations of the paper-based Hb assay, in-
cluding sample preparation, placement of the sample
DESIGN AND OPERATION OF THE PAPER-BASED Hb ASSAY onto the ␮PAD, drying of the blood stain, image scan-
Figure 1 illustrates the design and operation of the ning, and the automated image analysis, could be com-
paper-based Hb assay. Each individual ␮PAD com- pleted within 35 min.
prised a 2.8-cm-diameter circle (Fig. 1a) designed to
create a constant area for quantification with an image EFFECT OF DILUTION ON THE MEAN COLOR INTENSITY OF THE
analysis algorithm and prevent potential cross- BLOOD STAIN
contamination of samples. Each sheet of chromatogra- Figure 2 illustrates the effect of dilution on the mean
phy paper contained a 5-by-4 array of ␮PADs and in- color intensity of the stains produced by blood on pa-
cluded an alignment mark to simplify automation of per. To perform these experiments, we prepared a se-
the subsequent image analysis. We fabricated the ries of calibration samples with their [Hb] adjusted to
␮PAD arrays by printing their pattern with a solid-ink 2.5, 5, 10, 15, 20, and 25 g/dL (as described in Methods)
(wax) printer (Fig. 1A). We then used a hot plate to using blood from 3 consenting volunteers. We then
heat the printed chromatography paper above the mixed a small volume of each calibration sample with

1508 Clinical Chemistry 59:10 (2013)


Paper-Based Hb Test for Resource-Limited Settings

Fig. 2. Dependence of the mean color intensity of the blood stain on the dilution ratio.
(A), Whole blood samples were diluted or concentrated with plasma or pelleted RBCs to achieve [Hb] ranging from 2.5 to 25
g/dL. A 20-␮L drop of blood mixed with Drabkin reagent at 1:10, 1:15, and 1:20 ratios (vol/vol) was placed at the center of
each ␮PAD. (B–D), Mean color intensities for the red (B), green (C), and blue (D) channels of the blood stain in a ␮PAD produced
by each sample at each dilution ratio. Data shown as mean (SD) (n ⫽ 3). au, arbitrary unit.

Drabkin reagent at 1:10, 1:15, and 1:20 ratios (vol/vol), on the basis of the mean color intensity of its blood
waited 10 min, and placed a 20-␮L droplet of each mix- stain.
ture onto paper. To determine the limit of detection (LOD) and the
Droplets with markedly different Hb content, due limit of quantification (LOQ) of the assay, we varied
to either a difference in [Hb] or dilution, produced [Hb] of blood samples from 0.5 to 25 g/dL by diluting
blood stains with different color intensities that were whole blood with autologous plasma or concentrating
easily distinguishable visually. As expected, the blood whole blood with packed RBCs (as described in Meth-
stain color intensity increased with increasing [Hb] for ods). We defined the LOD as the lowest [Hb] at which
each mixing ratio (dilution) and decreased with in- the blood stain could be identified and measured using
creasing dilution for each [Hb] value (Fig. 2A). The our image processing algorithm, and the LOQ as the
mean color intensity of each blood stain correlated very lowest [Hb] at which the disagreement with the com-
strongly with its [Hb] (Fig. 2B–D). The combination of parison method (Medonic hematology analyzer) was
1:15 mixing ratio and green channel showed the best within 1 g/dL. We found that the LOD for our paper-
linear fit (mean color intensity ⫽ 2.0986 ⫻ [Hb] ⫹ based Hb assay was 1 g/dL and the LOQ was 2.5 g/dL.
3.7226, R2 ⫽ 0.996). Therefore, we used the 1:15 mix-
ing ratio to prepare samples in all further experiments, RELATIVE ACCURACY AND PRECISION OF THE PAPER-BASED Hb
the mean color intensity in green channel, and the lin- ASSAY
ear equation produced by this graph (Fig. 2C) as the We tested the relative accuracy of the paper-based Hb
calibration curve for estimating the [Hb] of the sample assay by comparing the values of [Hb] measured using

Clinical Chemistry 59:10 (2013) 1509


the diagonal line (Fig. 3A, red dashed line), indicating a
good agreement between the [Hb] estimated from the
mean color intensity of blood stains in our paper-based
Hb assay and the actual [Hb] obtained from the stan-
dard analysis (Fig. 3A). The linear least-squares regres-
sion analysis (Fig. 3A, black solid line) showed a high
degree of correlation between the 2 measurements (y ⫽
1.05x – 0.58, R2 ⫽ 0.9598).
We also constructed the Bland–Altman plot (29 )
to evaluate the agreement between the paper-based Hb
assay and the Medonic hematology analyzer (Fig. 3b).
The standard deviation (SD) of the difference be-
tween the comparison method (hematology ana-
lyzer) and the paper-based Hb assay was 0.62 g/dL.
The limits of agreement between the 2 methods were
⫺1.30 and 1.18 g/dL. The paper-based Hb assay agreed
with the hematology analyzer within 1 g/dL 90.7% of
the time, overestimating [Hb] by ⱖ1 g/dL in 1.9% of
subjects and underestimating [Hb] by ⱖ1 g/dL in 7.4%
of subjects (Fig. 3B).
In a separate experiment, we tested the precision of
the paper-based Hb assay by measuring repeatedly
(n ⫽ 5) the [Hb] for a series of blood samples obtained
from the same subject with the actual [Hb] of each
sample adjusted (as described in Methods) to 2.5, 5, 10,
15, 20, and 25 g/dL. The SD for the [Hb] measurements
performed with different droplets from the same blood
sample was consistently ⬍0.5 g/dL for all values of the
true [Hb] tested: 0.21 g/dL (CV 8.3%) for the sample
with true [Hb] of 2.5 g/dL; 0.19 g/dL (CV 3.9%) for 5
Fig. 3. Comparison of the [Hb] measurements per- g/dL; 0.11 g/dL (CV 1.1%) for 10 g/dL; 0.21 g/dL (CV
formed with the paper-based Hb assay and the Me- 1.4%) for 15 g/dL; 0.34 g/dL (CV 1.7%) for 20 g/dL;
donic hematology analyzer (n ⴝ 54 subjects). and 0.78 g/dL (CV 3.1%) for 25 g/dL. For comparison,
(A), Linear regression analysis of the correlation between the [Hb] measurement range for the Medonic hema-
the [Hb] values determined by a hematology analyzer tology analyzer was 0 –35 g/dL, with CV ⬍1.0% (values
(actual [Hb]) and by the paper-based Hb assay (estimated provided by the manufacturer).
[Hb]). The diagonal (red dashed line) represents a 1:1
correspondence between the actual and the estimated [Hb] EFFECT OF DROPLET VOLUME VARIATION AND LONG-TERM
values. (B), Bland–Altman plots of the agreement between STABILITY OF THE PAPER-BASED Hb ASSAY
the paper-based Hb assay and the hematology analyzer. We measured the effect of variation in the volume of
The solid red line is the difference between the 2 methods the sample droplet, which may occur during the use of
(mean bias); the dashed red lines are limits of agreement the assay in the field, on the [Hb] measurement for
(within 2 SD). The mean [Hb] was 11.50 (2.86) g/dL for the samples with true [Hb] of 5, 10, 15, and 20 g/dL. A 20%
paper-based Hb assay and 11.43 (3.05) g/dL for the Me- variation in the volume of the sample droplet (16 –24
donic hematology analyzer. ␮L) resulted in a ⱕ0.21 g/dL SD (CV 4.3%) of the
estimated [Hb] for samples with true [Hb] ⫽ 5 g/dL;
ⱕ0.66 g/dL (CV 5.6%) for 10 g/dL; ⱕ0.67 g/dL (CV
our assay and the Medonic M-Series hematology ana- 4.2%) for 15 g/dL; and ⱕ0.93 g/dL (CV 4.4%) for
lyzer for blood samples from 54 subjects. Subjects in- 20 g/dL (see Supplemental Fig. 1A, which accompanies
cluded individuals with Hb AA, Hb AS, Hb SC, and Hb the online version of this article at http://www.
SS genotypes; the presence of Hb S in the blood samples clinchem.org/content/vol59/issue10). As expected, the
did not affect the [Hb] measurements (data not size of the blood stain depended on the droplet volume.
shown). Figure 3 illustrates the results of this side-by- For droplets of the same volume, but with different
side comparison. The data points representing these true [Hb] values (5, 10, 15, and 20 g/dL) the normal-
measurements were distributed in close proximity to ized size of the blood stains produced by the droplets

1510 Clinical Chemistry 59:10 (2013)


Paper-Based Hb Test for Resource-Limited Settings

had an SD of ⱕ0.03 (CV 4.7%) for 16 ␮L; ⱕ0.03 (CV Unlike the HbCS, our paper-based Hb assay is not af-
3.9%) for 18 ␮L; ⱕ0.03 (CV 4.5%) for 20 ␮L; ⱕ0.04 fected by changes in ambient light because the flatbed
(CV 4.1%) for 22 ␮L; and ⱕ0.03 (CV 3.7%) for 24 ␮L scanner itself provides consistent illumination, and the
(see online Supplemental Fig. 1B). color can be read at any time within a 24-h period after
Finally, we tested the long-term stability of the collection of the blood stain. The result is not affected
paper-based Hb assay measurements by scanning the by operator bias because the measurement is per-
sheets of paper containing the blood stains 4 times dur- formed completely automatically.
ing the first 20 min, while the stain was still drying, and Very little training is required to perform our
13 times over the next 24 h, for samples with true [Hb] paper-based Hb assay. All a user would need to do is to
of 5, 10, 15, and 20 g/dL. The color of the dried blood mix a finger-prick volume of blood with Drabkin re-
stain remained stable after the initial drying period. agent at the appropriate ratio and deposit a droplet of
We therefore were able to scan and analyze the blood this mixture on paper. The complexity of these opera-
stains at any point over the next 24 h without any tions is about the same as that of the sample prepara-
significant change in the [Hb] measurement (see on- tion steps required from consumers for performing at-
line Supplemental Fig. 2). During this study, the home rapid diagnostic tests for sperm count (35 ) or Hb
room temperature in our laboratory varied from A1c (36 ). All subsequent operations of the paper-based
18 °C to 30 °C, and humidity varied from 20% to Hb assay are completely automated and user-
80%. We did not systematically test the effect of independent, including the scanning and analysis of
room temperature or humidity variations on the the stains. Notwithstanding its simplicity, our paper-
performance of the assay. based Hb assay demonstrated excellent comparison
(SD ⫽ 0.62 g/dL, n ⫽ 54) with the Medonic hematol-
Discussion ogy analyzer. The performance of our assay also com-
pares favorably with conventional POC devices
The measurement of blood [Hb] is one of the most (14, 19, 37 ), particularly in the context of our target
frequently performed laboratory tests within a wide applications: (a) providing the measurement of [Hb]
range of medical settings, from acute trauma care to when no other methods are available, and (b) enabling
chronic disease management (1 ). While the primary public health screening in resource-limited settings
reason for initial [Hb] screening is to diagnose (or rule where the lowest cost, rather than the highest accuracy,
out) anemia, frequent monitoring of [Hb] is often re- is most important.
quired in the context of many conditions to quantify We used a USB-powered flatbed scanner (Canon
blood loss, track disease progression, assess treatment CanoScan LiDE110, $44 on amazon.com as of April 21,
efficacy, or monitor patients during procedures. When 2013) and a laptop computer to digitize and analyze the
reliable diagnostic devices are available, they are often blood stains. Any computer that has a functional USB
not used because they are considered to be prohibi- port and complies with the system requirements of the
tively expensive for the resource-limited settings of de- scanner could in principle be used, including older sal-
veloping countries (33 ). For example, in Kenyan dis- vaged or refurbished models. Importantly, the scanner
trict hospitals, nearly 15% of children with clinical and laptop could be used for many other purposes, not
histories indicative of malaria or anemia did not have exclusively for performing our paper-based Hb assay.
their [Hb] tested (20, 34 ). In this context, a reliable, The primary reason for using these multipurpose,
low-cost assay for measuring [Hb] would improve consumer-grade electronic devices as the capital equip-
the diagnosis of anemia and could help minimize ment part of a low-cost diagnostic assay is the ability to
waste of limited resources on unneeded treatments tap into the already well-established market of con-
and reduce the risk of complications associated with sumer electronics. These devices are much more avail-
misdiagnosis (20, 25, 33 ). The urgent need for such able in resource-limited settings than specialized med-
an assay in sub-Saharan Africa (in areas endemic ical equipment such as a spectrophotometer and can be
with sickle cell disease and malaria) was the primary used for performing the [Hb] measurements without
motivation for this study. any modification of their original function. This ap-
Our paper-based Hb assay measures [Hb] by sim- proach represents an important advantage of our work
ply quantifying the appearance of a stain produced by with respect to recent studies that follow the conven-
the blood sample on paper. Perhaps the most widely tional paradigm of POC development and propose to
used alternative low-cost method for estimating [Hb] measure [Hb] by use of a new, potentially low-cost
in resource-limited settings is the WHO HbCS. To per- electronic device (38, 39 ). Such a device, designed spe-
form this test, the user must visually compare the color cifically for measuring [Hb], would have to be sepa-
of a paper strip stained with blood to a standard at a rately developed, manufactured, distributed, cali-
certain time after the blood was applied to the strip. brated, and maintained. The use of these POC devices,

Clinical Chemistry 59:10 (2013) 1511


particularly the maintenance and/or replacement of ing countries (e.g., sub-Saharan Africa), particularly
broken devices, requires a level of medical and engi- in the context of malaria, HIV, and sickle cell
neering infrastructure that is largely unavailable in disease.
resource-limited settings (40 ).
Another important advantage of our approach is
that the use of existing scanners and laptops, which
may already be present in even the most remote facili- Author Contributions: All authors confirmed they have contributed to
ties because they were purchased by consumers for the intellectual content of this paper and have met the following 3 re-
their personal use, could allow for an effectively “zero- quirements: (a) significant contributions to the conception and design,
cost” analyzer, compared with analyzer costs for con- acquisition of data, or analysis and interpretation of data; (b) drafting
or revising the article for intellectual content; and (c) final approval of
ventional POC devices ranging from $800 to $15 000. the published article.
In addition to the prohibitively high cost of the analyz-
ers, conventional POC devices often require proprie- Authors’ Disclosures or Potential Conflicts of Interest: Upon man-
uscript submission, all authors completed the author disclosure form.
tary disposable supplies with per-test costs ranging Disclosures and/or potential conflicts of interest:
from $0.02 to $3.67 (19 ). In contrast, the components
required to perform our paper-based [Hb] assay are Employment or Leadership: None declared.
Consultant or Advisory Role: J. Kanter, ApoPharma, Adventrix
available generically at an estimated combined cost of Pharmaceuticals.
⬍$0.007 (0.7 US cents) per test. Stock Ownership: None declared.
In summary, we developed and validated a paper- Honoraria: None declared.
based, point-of-care Hb assay for the low-cost assess- Research Funding: This work was supported in part by a 2012 NIH
ment of [Hb]. This assay represents a major step to- Director’s Transformative Research Award (NHLBI R01HL117329,
PI: S.S. Shevkoplyas).
ward effective intraoperative care as well as [Hb] Expert Testimony: None declared.
testing at the bedside or in urgent care settings where Patents: X. Yang, 61/692,994, 61/558,009; N.Z. Piety, 61/692,994; J.
traditional methods of the clinical laboratory are un- Kanter, 61/692,994; S.S. Shevkoplyas, 61/692,994, 61/558,009.
available or prohibitively expensive. The paper- Role of Sponsor: The funding organizations played no role in the
based Hb assay will be useful for diagnosing anemia design of study, choice of enrolled patients, review and interpretation
in resource-limited settings of low-income develop- of data, or preparation or approval of manuscript.

References
1. Pagana KD, Pagana TJ. Mosby’s manual of diag- chronic obstructive pulmonary disease. Part one. 1:83–95.
nostic and laboratory tests. Third edition. Phila- Am J Respir Crit Care Med 1994;150:833–52. 19. Patel KP, Hay GW, Cheteri MK, Holt DW. Hemo-
delphia, PA: Mosby, 2010. 11. Vicari AM, Ponzoni M, Alberetto M, Martani C, globin test result variability and cost analysis of
2. Shander A. Anemia in the critically ill. Crit Care Pontiroli AE, Folli F. Erythrocytosis in a patient eight different analyzers during open heart sur-
Clin 2004;20:159 –78. with chronic obstructive pulmonary disease. gery. J Extra Corpor Technol 2007;39:10 –7.
3. Aspuru K, Villa C, Bermejo F, Herrero P, Lopez SG. Haematologica 1998;83:183– 6. 20. Petti CA, Polage CR, Quinn TC, Ronald AR, Sande
Optimal management of iron deficiency anemia 12. Tefferi A. Polycythemia vera: a comprehensive MA. Laboratory medicine in Africa: a barrier to
due to poor dietary intake. Int J Gen Med 2011; review and clinical recommendations. Mayo Clin effective health care. Clin Infect Dis 2006;42:
4:741–50. Proc 2003;78:174 –94. 377– 82.
4. McLean E, Cogswell M, Egli I, Wojdyla D, de 13. Lowe GD, Lee AJ, Rumley A, Price JF, Fowkes FG. 21. Carneiro IA, Drakeley CJ, Owusu-Agyei S,
Benoist B. Worldwide prevalence of anaemia, Blood viscosity and risk of cardiovascular events: Mmbando B, Chandramohan D. Haemoglobin
WHO vitamin and mineral nutrition information the Edinburgh Artery Study. Br J Haematol 1997; and haematocrit: is the threefold conversion valid
system, 1993–2005. Public Health Nutr 2009;12: 96:168 –73. for assessing anaemia in malaria-endemic set-
444 –54. 14. Myers GJ, Browne J. Point of care hematocrit and tings? Malar J 2007;6:67.
5. Weiskopf RB, Viele MK, Feiner J, Kelley S, Lieber- hemoglobin in cardiac surgery: a review. Perfu- 22. Stott GJ, Lewis SM. A simple and reliable method
man J, Noorani M, et al. Human cardiovascular sion 2007;22:179 – 83. for estimating haemoglobin. Bull World Health
and metabolic response to acute, severe isov- 15. Gehring H, Hornberger C, Dibbelt L, Rothsigkeit Organ 1995;73:369 –73.
olemic anemia. JAMA 1998;279:217–21. A, Gerlach K, Schumacher J, Schmucker P. Accu- 23. Quinto L, Aponte JJ, Menendez C, Sacarlal J, Aide
6. Carson JL, Carless PA, Hebert PC. Outcomes using racy of point-of-care-testing (POCT) for determin- P, Espasa M, et al. Relationship between haemo-
lower vs higher hemoglobin thresholds for red ing hemoglobin concentrations. Acta Anaesthe- globin and haematocrit in the definition of anae-
blood cell transfusion. JAMA 2013;309:83– 4. siol Scand 2002;46:980 – 6. mia. Trop Med Int Health 2006;11:1295–302.
7. Aitchison R, Russell N. Smoking: a major cause of 16. Price CP, Kricka LJ. Improving healthcare acces- 24. Paddle JJ. Evaluation of the haemoglobin colour
polycythemia. J R Soc Med 1988;81:431. sibility through point-of-care technologies. Clin scale and comparison with the Hemocue haemo-
8. Horstman D, Weiskopf R, Jackson RE. Work ca- Chem 2007;53:1665–75. globin assay. Bull World Health Organ 2002;80:
pacity during 3-wk sojourn at 4,300 m: effects of 17. Spielmann N, Mauch J, Madjdpour C, Schmugge 813– 6.
relative polycythemia. J Appl Physiol 1980;49: M, Weiss M, Haas T. Accuracy and precision of 25. Allain JP, Anokwa M, Casbard A, Owusu-Ofori S,
311– 8. hemoglobin point-of-care testing during major Dennis-Antwi J. Sociology and behaviour of west
9. Marchioli R, Finazzi G, Specchia G, Cacciola R, pediatric surgery. Int J Lab Hematol 2012;34:86 – African blood donors: the impact of religion on
Cavazzina R, Cilloni D, et al. Cardiovascular 90. human immunodeficiency virus infection. Vox
events and intensity of treatment in polycythemia 18. Tudos AJ, Besselink GJ, Schasfoort RB. Trends in Sang 2004;87:233– 40.
vera. N Engl J Med 2013;368:22–33. miniaturized total analysis systems for point-of- 26. Yang X, Forouzan O, Brown TP, Shevkoplyas SS.
10. MacNee W. Pathophysiology of cor pulmonale in care testing in clinical chemistry. Lab Chip 2001; Integrated separation of blood plasma from

1512 Clinical Chemistry 59:10 (2013)


Paper-Based Hb Test for Resource-Limited Settings

whole blood for microfluidic paper-based analyt- 32. Deegan RD, Bakajin O, Dupont TF, Huber G, mance criteria. Clin Chem 2010;56:44 –52.
ical devices. Lab Chip 2012;12:274 – 80. Nagel SR, Witten TA. Capillary flow as the cause 37. Gomez-Simon A, Navarro-Nunez L, Perez-
27. Carrilho E, Martinez AW, Whitesides GM. Under- of ring stains from dried liquid drops. Nature Ceballos E, Lozano ML, Candela MJ, Cascales A,
standing wax printing: a simple micropatterning 1997;389:827–9. et al. Evaluation of four rapid methods for hemo-
process for paper-based microfluidics. Anal Chem 33. Bates I, Maitland K. Are laboratory services com- globin screening of whole blood donors in mobile
2009;81:7091–5. ing of age in sub-Saharan Africa? Clin Infect Dis collection settings. Transfus Apher Sci 2007;36:
28. Yang X, Kanter J, Piety NZ, Benton MS, Vignes 2006;42:383– 4. 235– 42.
SM, Shevkoplyas SS. A simple, rapid, low-cost 34. English M, Esamai F, Wasunna A, Were F, Ogutu 38. Bond M, Elguea C, Yan JS, Pawlowski M, Wil-
diagnostic test for sickle cell disease. Lab Chip B, Wamae A, et al. Assessment of inpatient pae- liams J, Wahed A, et al. Chromatography paper
2013;13:1464 –7. diatric care in first referral level hospitals in 13 as a low-cost medium for accurate spectrophoto-
29. Bland JM, Altman DG. Statistical methods for districts in Kenya. Lancet 2004;363:1948 –53. metric assessment of blood hemoglobin concen-
assessing agreement between two methods of 35. Coppola MA, Klotz KL, Kim KA, Cho HY, Kang J, tration. Lab Chip 2013;13:2381– 8.
clinical measurement. Lancet 1986;1:307–10. Shetty J, et al. Spermcheck fertility, an immuno- 39. Kim D-S, Choi J-H, Nam M-H, Yang J-W, Pak JJ,
30. Khan MS, Thouas G, Shen W, Whyte G, Garnier diagnostic home test that detects normozoosper- Seo S. LED and CMOS image sensor based he-
G. Paper diagnostic for instantaneous blood typ- mia and severe oligozoospermia. Hum Reprod moglobin concentration measurement technique.
ing. Anal Chem 2010;82:4158 – 64. 2010;25:853– 61. Sensors Actuators B Chem 2011;157:103–9.
31. Fu E, Ramsey SA, Kauffman P, Lutz B, Yager P. 36. Lenters-Westra E, Slingerland RJ. Six of eight 40. Yager P, Domingo GJ, Gerdes J. Point-of-care
Transport in two-dimensional paper networks. hemoglobin A1c point-of-care instruments do not diagnostics for global health. Annu Rev Biomed
Microfluid Nanofluid 2011;10:29 –35. meet the general accepted analytical perfor- Eng 2008;10:107– 44.

Clinical Chemistry 59:10 (2013) 1513

You might also like