Owner's Manual

D700 Dental Set

Technical Name: D700 Dental Set


Trade Name: D700 Dental Set
Commercial Model: D700
Directions for use Table of contents
CONGRATULATIONS ! 1 – Identification
You have just purchased an equipment 1.1 – Name and model
built with the highest technology available 1.2 – Product description
1.3 – Physical Principle and fundaments of the
in the market, designed to provide product technology applied for its functioning and
maximum yield, quality and safety. action
1.4 – Accompanying parts and accessories
1.5 – Optional materials, consumption materials
and support materials
Equipament 1.6 – Technical Specifications and Features

2 - Special product storage, preservation and/or


Technical Name: D700 Dental Set handling conditions
Trade Name: D700 Dental Set
Commercial Model: D700 3 - Installation

4 - Procedures before using the product


Manufacturer
5 - Directions for use of the product
CNPJ (Corporate Taxpayer's Roll) nº 5.1 - Chair
55.979.736/0001-45 5.2 – Dental unit
Avenida Presidente Castelo Branco, 2525 5.3 – Auxiliary Unit
Lagoinha 5.4 – Reflector
Ribeirão Preto SP CEP (Zip code) 14095-000
6 - Warnings and/other cautions to be adopted
Phone: + 55 (16) 3512-3700
7 - Additional procedures for reutilization
Legal Responsible
8 – Performance on essential safety requisites
and efficacy of the medical product, and
Leonel Issa Halak - Eng. CREA-SP eventual undesired secondary effects
(Regional Engineering, Architecture and
Agronomy Board/São Paulo) 8.1 – Indication, purpose and intended use of
nº 0600256918 the product
8.2 – Secondary or undesired side effects and
contraindications
ANVISA (National Health Surveillance Agency) 8.3 – Product safety and efficacy
Registration nº 10101130061
9 - Installation or connection to other products
IMPORTANT 10 - Preventive and Corrective Maintenance

For greater safety: 10.1 – Preventive Maintenance


Read and understand all the instructions 10.2 – Corrective Maintenance
contained in this Owner´s Manual before 11 - Cautions in case of change in product
installing or operating this product. functioning

This manual was originally written in 12 - Sensitivity to predictable environmental


Portuguese. conditions in normal use situations

Manual Review: November 2006 13 - Cautions in case of product damage

Owner’s manual – D700 Dental Set 03


is caused by venturi system with compressed air.
1 - Identification The dental unit has compressed air hoses and
connectors to supply the hand pieces (low and high
1.1 - Name and model rotation) and syringe with water and air outlet.
Technical name: D700 Dental Set
Trade Name: D700 Dental Set 1.4 - Accompanying parts and
Commercial model: D700 accessories

2
3

1.2 - Product Description Main parts 6


The chair is a set of parts formed by: Patient Chair, 1 - Reflector
Auxiliary Unit, Reflector and Dental unit. 2 - Auxiliary Unit
These parts are interconnected, the use of any of 3 - Coupled Dental unit
them in other chairs being impossible, that is, the 4 - Chair
chair can only be supplied with all the parts 5 - Activation pedal
described in the Owner’s Manual. 6 - Control box (inbuilt in the chair)
The Chair is upholstered with polyurethane foam Accompanying Accessories
coated with PVC. Its structure is made of carbon
steel with powder painting (electrostatic). It also 01 water reservoir
has injected plastic and metal plate finishes.
The Auxiliary Unit has a removable bowl for
cleaning. It also includes a cup holder and suction 1.5 - Optional materials, consumption
hose. materials and support materials
The reflector is articulated and moves easily, with
light intensity between 19,000 and 22,000 Lux. Optional: Reflector lamp 100W
The Dental Unit has a tray optional to support Additional terminals for dental unit's hand
tools, a pedal for activating hand pieces for high pieces
and low rotation and a rotary support for hand Additional terminals for suction hoses of the
pieces. Comes with a water reservoir. auxiliary unit
Consumption materials: Water and compressed air
1.3 - Physical Principle and fundaments Electricity
of the product technology applied for its Support materials: Directions for Use
functioning and action
Manufacturing items for exclusive use in the company's
The chair is a mechanical system with electric products:
activation that allows raising the chair and All parts and accessories described in this Owner´s Manual
inclining the backrest by means of gear motors. and others not described but belonging to the equipment are
of exclusive use by the company's products.
The reflector uses low-voltage lamps and has
reflection system with light passage filter,

!
The user is entirely responsible for the use of any part,
eliminating the infrared spectrum. accessory or material not specified or set forth in this
The auxiliary unit has suction hoses, whose suction Owner´s Manual.

Owner’s manual – D700 Dental Set 04


1.6 - Technical Specifications and Features

Product classification according to ANVISA (National Health Surveillance Agency)


Category class Category class Class I (low risk)

Power supply
Supply Voltage 110/127/220/240V~
Frequency 50 / 60 Hz
No. of phases Single-phase / Two-phase
Inlet fuse 10A for 110/127 V~ and 5A for 220/240V~
Inlet power 237 VA

Other features

Reflector light intensity 19,000 to 22,000 Lux


Lamp power 55 W (standard) 100 W (optional)
Spray water reservoir capacity 1000ml
Venturi Suction Air Consumption 10.57 gal/min
High Rotation Air Consumption 2.38 gal/min
High Rotation Water Consumption 0.01 gal/min
Syringe Air Consumption 4.49 gal/min
Syringe Water Consumption 0.1 l/min
Packaging type Cardboard
Packaging dimensions Main packaging
0.65 x 0.6 x 1.09 yd

Upholstery package
1.15 x 0.46 x 0.135 m
Net total weight 111 Kg
Gross total weight 126 Kg

Owner’s manual – D700 Dental Set 05


2 - Special product storage, After installation
preservation and/or handling Keeping
- Keep away from rain and direct sunlight.
conditions
Preservation
Before installing - When in use, during cleaning and disinfection of
apparatus surfaces, use only neutral soap or Aplic
Marking in the Package Odonto®.
The information pertaining to cares to be taken - Keep the equipment clean and disinfected for the
during transportation and storage of the product next use.
are indicated through standard symbols directly
- Prevent liquids from being inserted in the
on the package.
equipment.
- Do not use organic solvents like thinner to clean
the equipment.
- Keep the equipment in a clean place.
Fragile: caution during transport and storage.
Note:
See item 8 Additional Procedures for Reutilization and
item 9 Additional Procedures Before Product
Protect against humidity during transport
Utilization.
and storage.

Operation:
- During use, check environmental conditions
Handle the packaging with the side indicated according to the table below:
by the arrows up.

Environmental conditions for operation


Operating room temperature range +10ºC to +35ºC
Maximum Piling determines the maximum
number of boxes that can be piled during Recommended room temperature range +21ºC to +26ºC
transport and storage.
Operating Relative humidity range 30% a 75%
(non-condensing)

700 hPa to 1060 hPa


Determines the temperature limit within Atmospheric pressure range (525 mmHg to 795
which the package must be stored or mmHg)
transported.

Storage
- Keep in a place away from rain and direct
sunlight and in original packaging.
- In case of storage of several boxes of the
packaged equipment, the maximum piling must be
as indicated by the piling symbol on the package.

Transport
- During transport, avoid vibrations and impacts in
the equipment
- Do not let it fall on the ground.
- In case of transport of several boxes of the
packaged equipment, the maximum piling must be
4 boxes as indicated by the piling symbol on the
package.

Owner’s manual – D700 Dental Set 06


3 - Installation 5 - Directions for use of the
product
Product installation
This product may be installed by a Technical Ensure that the installation was performed
Assistance or by the customer using the according to item 3 - Product Installation.
installation manual's instructions.
To request for Authorized Technical Assistance, Before utilizing or reutilizing
contact: Follow the procedure described in item 8
Internet Additional Procedures for Reutilization.
http://www.d700.com.br
5.1 - Chair
Needed infrastructure
1
1 - Headrest
2 - Backrest
- Electric network 2 3 - Seat
3 x 1.5 mm2 power supply cable 4 - Chair Structure
Supply voltage 110/127/220/240 V~ 5 - General Switch
Frequency 50 / 60 Hz 6 - Fuse Holder

- Water network
Water inlet pressure 2.0 6.0 bar 3
Water inlet flow 5 l/min
4
- Sewage network
Sewage connection Ø 40 mm
above the ground 20 mm
Max. sewage flow 5 l/min
Water sewage pipe incline 10 mm per meter. 5

- Compressed air network 6


Air connection R 1/2
above the ground min 40 mm / max. 75 mm Use of the Chair
Air inlet pressure 5.2 7 bar

We recommend that the installation Checklist and


warranty certificate be read carefully, as they contain
important information on the operations and warranty,
the important points of which we highlight: 1
- Install in the circuit breaker room for protection of the
equipment and compressor; 2
- Install water meter in a place of easy access for
maintenance and closing after working hours;
- Provide ground wire installation in the equipment - Turn on the general switch located in the rear
power supply; part of the control box cover

4 - Procedures before using the - Adjust the inclination of the headrest using
device 1 and height of the set's movement as a
product whole using device 2.

Install the equipment as described in the item


“Installation”. Even in the first utilization, the
equipment must be clean and disinfected.

Owner’s manual – D700 Dental Set 07


- Position the head and accommodate the patient, - For the functioning of rotary instruments,
using the pedal controls according to the figure. remove the instrument to be used (A or B) from the
E support and press the pedal (C).
C
A D
A - Set up
B - Sit down
C - Rest down C
D - Rest up B
E - Reflector On/Off

Note:
The controls work only
while they are pressed.
A B
5.2 - Dental unit
2 - For functioning of the syringe, remove it from the
support and press:
1

Button 1 for Water.


Button 2 for Air.
3 Both buttons simultaneously to obtain Spray.
4
5
6 1
2
7

1 - Instrument Tray
2 - Articulated Arm Motion Lock
3 - Syringe
4 - Connection for low rotation
5 - Connection for high rotation or micro motor
6 - Water reservoir
7 - Pedal

Dental unit utilization 5.3 - Auxiliary unit


- Check the reservoir level and supply
if necessary.
1
2

- The arm's motion must be manual.


- After placing in the desired position, tighten the
“A” lock. 3 5

1 - Cup Holder
2 - Bowl
3 - Suction
4 - Unit Body
5 - Debris Separator

Owner’s manual – D700 Dental Set 08


Use of the unit Reflector utilization
- For the suction's functioning, just remove it from
the support. On and Off:
- To on and off the reflector, press the switch in the
pedal.

Suction

- To obtain water in the bowl, press the key in


position “1”.

6 - Warnings and/other cautions


Key
to be adopted
- This equipment must be used only by personal
duly instructed and under direction of qualified
dental team acquainted with risks and benefits
currently known associated with the risk of dental
- To obtain water in the cup, press the key in chairs.
position “2”. - We recommend that the professional instruct the
patient or other personnel with access to the
equipment on the cautions during handling.
- Do not try to repair or assemble defective or non-
operating components or replace them with parts
from another device. Only with the use of original
parts can the original technical specifications and
safety of the device be guaranteed.
- To guarantee electrical safety of the device
throughout its service life, we recommend that the
Key equipment be checked by the Technical Assistance
at regular intervals of at least once a year.
- These sets are not appropriate for operation in
areas with danger of explosion.
- Before using for the first time and/or after long
5.4 - Reflector work stoppages like holidays, clean and disinfect
the equipment; remove air and water from internal
hoses.
1 - Mirror protection
- After each work, we recommend removing and
2 - Mirror
1 sterilizing the drills/mills and rotary instruments
3 - On/Off switch 2 to prevent cross contamination.
4 - Handle - Turn of the general power switch of the set and
5 - Lamp close the air and water meters before leaving the
office.
- Do not expose the equipment directly to sunlight
as this will cause early aging of fairing and/or
5
3
upholstery.
4

Owner’s manual – D700 Dental Set 09


- All parts of the Dental Chair D700 model are only comfortable thanks to the articulation that is
comparable among themselves upon installation. positioned in the acetabulum line;
We point out the non-compatibility of equipment - Auxiliary unit: for auxiliary works like water supply
from other manufacturers. for cup and collection of dejects and activation of
- We recommend that all directions for use be read the cuspidor and suction hoses with reliability and
with special care, since we are placing at your practicality for the professional and assistant;
disposal a refined technology. - Dental unit: for activation and control of the
- Operating errors can cause damages to the syringe, rotary instruments and others, providing
equipment or non-utilization of the resources it the best proximity to the work field;
can provide. - Reflector: for lighting of the work field so as to
- For user safety, do not pour liquids on the dental enable visualization during the procedure, enduring
chair, thus preventing short circuit and damages easy and comfortable access by the professional.
to the equipment.
- The equipment’s rotary instruments were 8.2 - Secondary or undesired side effects
designed for dental treatments only, in works and contra indications
requiring greater pressures and torque, like
prosthesis, we recommend specific instruments. - There are no secondary or undesired side effects.
- The maximum admissible load on the chair is 200
kg. 8.3 - Product safety and efficacy
- The maximum load on the dentist main tray is 1 - The Dental Chair D700 model is completely safe,
kg. provided the safety rules are followed in all the
recommendations described in this Owner´s
7 - Additional procedures for Manual.
reutilization - The equipment was designed and manufacturing
using standard (nontoxic) materials for
medical/dental use and that enable easy asepsis. It
Cleaning of Upholstered or Painted Parts is practical and designed to guarantee full safety.
Use a cloth wet with Aplic Odonto® or a similar - The equipment will not lose its physical or
product in cleaning painted/upholstered parts. dimensional characteristic when maintained
(stored) or preserved as mentioned in item 2 of this
Cleaning and Preserving the Suction Filters
Owner´s Manual.
Clean the suction filters after surgical works or
whenever you notice reduction in suction power. 9 - Installation or connection to
Cleaning the Sink and Sink Drain other products
The sink can be removed for cleaning. The drain
must be cleaned daily. The Dental Chair D700 model is not intended for
use in direct connection with other equipment.
Disinfecting the Unit
Disinfecting the Suction Lines: Suction daily in
each suction hose 1 to 1.5 liters of disinfectant 10 - Preventive and corrective
solution for PVC hoses. maintenance
8 - Performance on essential 10.1 Preventive Maintenance and
safety requisites, efficacy of the Preservation (Owner's responsibility)
medical product, and eventual Preventive maintenance must be conducted every
undesired secondary effects year, with the aim of checking the operation of the
chair's electrical circuits, gear motor, hoses, air and
8.1 - Indication, purpose and intended water connections, etc.
Preventive Maintenance must be conducted by
use of the product Technical Assistance only, which must be
- Chair: for accommodating the patient during the requested as informed in the item on Corrective
dental treatment, allows positioning the patient Maintenance. With it, you can plan your
properly in each treatment stage through the equipment's preventive maintenance, thus
motions. The chair's motions are extremely smooth reducing operational failure and safety risks.
since they are caused by low-voltage and refined
For preservation, keep the equipment clean,
mechanical motors. The backrest's motion is very
disinfected and stored according to item 8.

Owner’s manual – D700 Dental Set 10


10.2 - Corrective Maintenance
When there are damages to one or more parts of the equipment, separate it and identify it with a label stating
that the part(s) is(are) “under maintenance”.

Failure detection and correction


Before calling the Technical Assistance, check if the problem does not fit into one of the items listed below:

Table of identification and correction of problems


No Electrical Function is Executed Probable causes
- Circuit breaker is off
- General switch is off
- Burned fuse

Solutions
- Turn on the circuit breaker
- Turn on the general switch
- Replace the fuse

Refletor Não Liga Probable causes


- On/off selector switch is disabled
- Burned lamp
- Burned fuse

Solutions
- Enable on/off selector switch
- Replace according to lamp
- Replace the fuse

Sugador Não Funciona Probable causes


- Suction filters blocked

Solutions
- Clean filters

All the recommendations for use are found in these directions for use, but if any problem or malfunctioning is
detected and cannot be corrected according to the instructions herein, contact the authorized service network to
correct the problem.

Seek the authorized Service network and always use original parts, since they are the guarantee that your product
will always be in ideal working and safety conditions.

Circuit Diagrams, Parts, Components List and Others


The diagrams and lists of parts, components and other information required for Technical Assistance are found in
the CD-ROM that comes with the product.

Owner’s manual – D700 Dental Set 11


Replacing the Reflector Lamp 13 - Cautions in Case of Product
1. Remove protection “1”. Damage
2. Disconnect the lamp wires.
3. Remove the damaged lamp. To prevent environmental contamination or undue
4. Place the new lamp. use of the Dental Chair D700 model when it is
5, Clean the front protection mirror “2”. damaged, it must be disposed of (according to the
legislation in force) in an appropriate location.

The equipment is warranted for 15 (fifteen)


months counted of the date on the sale invoice
issued to the customer.
1

Note:
The lamp’s service life is approximately 400 hours,
being that 3% of them can burn prematurely
during this period.

11 - Cautions in case of change


in product functioning
If the equipment presents abnormal heating or any
other abnormality, check if the problem is related
to some of the items listed in item 7.2. If the
problem cannot be solved, request for authorized
technical assistance.

12 - Sensitivity to predictable
environmental conditions in
normal use situations

The Dental Chair D700 model was designed to be


insensitive to interferences like magnetic fields,
external electrical influences, electrostatic
discharges, pressure or pressure variation,
provided the equipment is installed, clean,
preserved and operated according to this directions
for use.

Check the appropriate conditions in item 2 -


Special product storage, preservation and/or
handling conditions.

Owner’s manual – D700 Dental Set 12