Note: The prefixes of the PID and checking serial number must contain numbers, letters,
or Chinese characters.
Age:
Used to configure the method of worklist patient age calculation


Configure the field used for barcode searching and matching.


9.10
9.11 Lock Device

9.12



9.13
10
11

11.1



11.2
11.3
11.4
11.5

11.5.1
11.5.2












11.5.3










➢ Minimum Density: Only digits are available (0-65535).
➢ Maximum Density: Only digits are available (0-65535).
➢ Margin Density: Set the density between the film images. Options available: WHITE,
BLACK.
➢ Empty Graph Density: Set the density in the peripheral area. Options available: WHITE,
BLACK.
➢ Image Polarity: Set the default image polarity. Options available: NORMAL,
REVERSE.
➢ Film Target: Options available: PROCESSOR, MAGAZINE, BIN_1, BIN_2, …BIN_8.
➢ Media Type: Options available: BLUE FILM, PAPER, CLEAR FILM, MAMMO CLEAR
FILM, MAMMO BLUE FILM. The default settings is BLUE FILM.
➢ Interpolation Type: Options available: REPLICATE, BILINEAR, CUBIC and NONE.
➢ Printing Border: whether to set the printing border, YES or NO.
➢ Priority level: HIGH, MED or LOW levels are available. MED by default.
➢ Configuration information: The printer can adopt different concentration levels (or left
not configured) according to the conformance configurations of the printer.
➢ Owner identity: Displayed on the workstation control software interface, Mindray by
default. It can be modified as needed.
➢ Film tag: Mindray DX by default. It can be modified as needed. Letters and digits are
accepted as input.
➢ Film size: Set default film size. The available film size can be configured by clicking

to set the default film size.

11.5.4











11.5.5





11.5.6





11.5.7




11.5.8

11.5.9
⚫ Enable remote control failure time: After this function is enabled, the remote control
must be activated and authorized before it can be used. After being activated, the
remote control is authorized and fails based on the configured time (the remote control
can be started only when NFC is configured; otherwise, the remote control fails).
⚫ Head inching speed: Set the trolley movement speed under the control of the remote
control.
⚫ Enable automatic standby: After this function is enabled, the equipment automatically
becomes standby when it is not operated within the set time range.
⚫ Enable head homing standby: After this function is enabled, the equipment
automatically becomes standby when the head returns to its original position.
⚫ Enable contact anti-collision: After this function is enabled, the trolley brakes forcibly
when the bumper at the chassis of the trolley is triggered.
⚫ Enable non-contact anti-collision: After this function is enabled, the equipment
generates an audible and visual alarm when the trolley moves near an obstacle.
⚫ Enable exposure countdown tone: After this function is enabled, the speaker executes
countdown when the exposure is delayed.
⚫ Enable exposure success music: After this function is enabled, the speaker plays the
music after exposure is complete.
⚫ Enable column indicator: After this function is enabled, the column indicator provides a
light prompt.
⚫ Volume: Drag the slider to adjust the volume of the speaker.
⚫ Push hand feeling: including the light, standard, and steady modes.
⚫ Pressure sensor calibration: You can calibrate the pressure sensor.
12


Calibration
progress bar
13

13.1
The maintenance tasks are as follows:

1.

2.

3.
4.

5.


13.2


Appendix A Technical Data

A.1 Safety

A.2 Environmental Conditions


℃ ~ ℃
℃ ~ ℃


℃ ~ ℃


A.3 Software Operating Environment

A.4 Data Interfaces and User Access Control

A.5 Power Supply Conditions

A.6 Trolley Workstation


A.7 Portable Workstation

A.8 X-ray Generator


Secondary maximum inverter frequency:
≥ 460 kHz
Percent ripple:
≤ 1%

The deviation of the voltage value of the X-ray tube is not higher than 5%
for the combination of loading factors used in the clinical protocol.

The deviation of the current value of the X-ray tube is not higher than
±(5%+1 mA) for the combination of loading factors used in the clinical
protocol.

a)
b)

a) The deviation of loading time is not higher than ±(5%+1 ms) for the combination of
loading factors used in the clinical protocol.

a) ( % )
b) Above 0.2mAs: ±(5%+0.2mAs)
( )

A.9 Specifications of the Mobile Rack





A.10 X-ray Tube Assembly

A.11 Collimator





A.12 Specifications of the Source Component


A.13 Detector
A.14 Wall stand
⚫ Lifting travel range of cassette: 350 mm to 1800 mm
⚫ Attenuation equivalent of the front shell: ≤ 1.0 mmAl/100 kV
⚫ The camera stand supports horizontal and vertical placement of the MPX3543R and
EPX3543R flat panel detectors.
⚫ Installation of the ionization chamber is supported.
⚫ Installation of the Grid is supported.

A.15 Mobile Patient Table


⚫ Moveable, pushed or pulled manually
⚫ Bed height: 700 mm±10 mm
⚫ Bed dimensions: 2000 mm x 650 mm x 700 mm (L x W x H)
⚫ Load-bearing capacity: 250 kg
⚫ Others: equipped with casters and locking mechanisms

A.16 Dose Area Product Meter

A.17 Grid
⚫ Model: JPI Grid-1000
⚫ Type: Converging and stationary grids
⚫ Grid density: ≥ 40 l/cm
⚫ Focal length: 1 m/1.3 m/1.8 m
⚫ Grid ratio: ≥ 8:1
⚫ Size: 470 x 450 mm (±5%)

A.18 Other Optional Accessories


Appendix B EMC and Radio

Regulatory Compliance

B.1 EMC




system needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC
information provided below.






P

Note 1 At
Note 2



= = =

If system image distortion occurs, it may be necessary to position system further from sources
of conducted RF noise or to install external power source filter to minimize RF noise to an
acceptable level.
Note 1 At
Note 2 These
B.2 Radio Regulatory Compliance
B.2.1.1 RF parameters

Radio devices 2.4GHz Wi-Fi 5GHz Wi-Fi 434MHz short 13.56MHz RFID
devices devices rang devices devices

Operating
2412-2472MHz 5170-5250MHz 434MHz 13.56MHz
frequency
Modulation
DSSS and OFDM OFDM FSK and LORA ASK
mode
Output power ≤20dBm ≤20dBm ≤10mW ≤10mW

– each frequency or frequency band of reception see above list;


– the preferred frequency or frequency band see above list;
– the bandwidth of the receiving section of the ME EQUIPMENT in those bands is 1MHz.

– see above list, In addition to the requirements of 7.9.3 of the general standard, for ME
EQUIPMENT that includes RF transmitters, the technical description already include each
frequency or frequency band of transmission, the type and frequency haracteristics of the
modulation and the EFFECTIVE RADIATED POWER.“Output power”in above list means
“EFFECTIVE RADIATED POWER”

The radio device used in this product is in compliance with the essential requirements and other
relevant provisions of Directive 2014/53/EU.

B.2.1.2 Radio frequency Radiation Exposure Information

For body worn operation, wireless Flat Panel Detector has been tested and meets the CE RF
exposure guidelines when used with the accessories supplied or those approved for use with
this product. Use of other accessories may not ensure compliance with CE RF exposure
guidelines.
Appendix C Declaration of