Service Note

SD730-424END*D
Vol. 1
Table of contents Service Note
SD730-424END*D

System overview 3
General service work (Optional software) 121
System appearance 4
General service work <Preset backup> 123
General block diagram 10
General service work <IQ report> 125
Signal path for each mode 11
General service work <Log file collection> 126
Main panel 14
General service work <Task manager> 129
Hardware turning point 15
General service work <Screen Capture> 130
Cable connection (SSA-790A: AplioXG) 17
Environment Variables 132
Cable connection (SSA-780A: AplioMX) 25
TCS calibration 135
Cable connection (SSA-680A: XarioXG) 31
Main panel test 137
Cable connection (SSA-660A: Xario) 40
Mandatory work for HW replacement 139
Option (Software: AplioXG and AplioMX) 55
Databases 141
Option (Software: XarioXG and Xario) 59
Network installation 151
DICOM option specification 62
DICOM configuration 153
Optional software installation 63
DICOM software composition 161
Option (Hardware: AplioXG and AplioMX) 66
DICOM configuration (Advanced) 165
Option (Hardware: XarioXG and Xario) 70
DICOM SR 209
UIMV option 74
Basic operation <Getting Start and shutdown> 233
UIVS-001A (DVD-VSR) 91
Basic operation <Security Mode> 237
Option (Peripherals) 99
Basic operation <New Patient> 238
Applicable media 100
Basic operation <New Patient: DICOM tag linkage> 240
Version up record <AplioXG> 102
Basic operation <Examination screen> 243
Version up record <AplioMX> 105
Basic operation <De - Identify> 246
Version up record <XarioXG> 107
Basic operation <Store / Print image> 247
Version up record <Xario V6 and later> 110
Basic operation <Review Stored Image> 252
Version up record <Xario V5.5 and before> 113
Basic operation <Archive Stored image> 253
General service work 116
Basic operation <Delete Archived> 257
General service work <Service Menu> 117
General service work <Password list> 118

1
Service Note
SD730-424END*D

System overview

2
System overview Service Note
SD730-424END*D

SSA-790A SSA-780A SSA-680A SSA-660A

External dimensions: mm
Model name LCD Size Mass: kg (approx.)
Width Height Depth
SSA-790A 19 inch 580 1,428 to 1,805 1,132 to 1,217 190
SSA-780A 19 inch 533 1,425 to 1,782 898 130
SSA-680A 19 inch 540 1,427 to 1,784 898 143
15 inch 540 1,402 to 1,638 814 133
SSA-660A
17 inch 540 1,375 to 1,610 814 136
3
System appearance (SSA-790A: AplioXG) Service Note
SD730-424END*D

4
System appearance (SSA-780A: AplioMX) Service Note
SD730-424END*D

5
System appearance (XarioXG: SSA-680A) Service Note
SD730-424END*D

6
System appearance (Xario: SSA-660A LCD model) Service Note
SD730-424END*D

7
System appearance (Xario: SSA-660A CRT model) Service Note
SD730-424END*D

8
Service Note
SD730-424END*D

Block diagram

9
General block diagram Service Note
SD730-424END*D

CRT: Xario (17”)


When CW LCD: AplioXG (19”)
2 Option:
When Pencil or Pencil
Pencil CB PWB Only for AplioXG XarioXG (19”, 17”)
2 CW Board and AplioMX Xario (17”, 15”)

96 IP PWB Monitor
TX0 PWB IQDAS
Stored to HDD
SVGA
192 Transmit
192
96 TX1 PWB Stored RAW data (800x600)
Probe A Transmit (Temporally)

TI PWB
Transducer
RC PWB
Probe B
192
Interface
64
When CW Receiver BE PWB IO PWB
Control Back End In/Out I/F
192 192
VCR
Probe C RX PWB
Receive 128 VP PWB Power ON / OFF
Control signal
Other than CW Video
(Operating / TCS)
Processor
MC PWB Standard Configuration: Main
Motor AplioXG, AplioMX, XarioXG
Motor Control for Option:
Panel
Overlay
4D /MD-TEE Control Xario Image
TCS Image

CPU
PWB E-IDE
or
Power Supply USB SATA

AplioXG (4 units) AplioMX ATAPI > USB E-IDE:


HDD AXG V2.2 and
1. Trans Unit XarioXG before
DVD/CD XXG V1.2 and before
2. PSA: Power Supply Analog Xario Driver Xario V6.2 and before
Standard 3. PSD: Power Supply Digital One unit
On and before V5 Xario system is
Option 4. PHV: Power High Voltage (OP) only support CD drive.
10
Signal path for each mode Service Note
SD730-424END*D

2D/CDI/PW When CW
CRT: Xario (17”)
LCD: AplioXG (19”)
2 When Pencil or Pencil Option:
Pencil CB PWB Only for AplioXG XarioXG (19”, 17”)
2 CW Board and AplioMX Xario (17”, 15”)

96 IP PWB Monitor
TX0 PWB IQDAS
Stored to HDD
SVGA
192 Transmit
192
96 TX1 PWB Stored RAW data (800x600)
Probe A Transmit (Temporally)

TI PWB
Transducer
RC PWB
Probe B
192
Interface
64
When CW Receiver BE PWB IO PWB
Control Back End In/Out I/F
192 192
VCR
Probe C RX PWB
Receive 128 VP PWB Power ON / OFF
Control signal
Other than CW Video
(Operating / TCS)
Processor
MC PWB Standard Configuration: Main
Motor AplioXG, AplioMX, XarioXG
Motor Control for Option:
Panel
Overlay
4D /MD-TEE Control Xario Image
TCS Image

CPU
PWB E-IDE
or
Power Supply USB SATA

AplioXG (4 units) AplioMX ATAPI > USB E-IDE:


HDD AXG V2.2 and
1. Trans Unit XarioXG before
DVD/CD XXG V1.2 and before
2. PSA: Power Supply Analog Xario Driver Xario V6.2 and before
Standard 3. PSD: Power Supply Digital One unit
On and before V5 Xario system is
Option 4. PHV: Power High Voltage (OP) only support CD drive.
11
Signal path for each mode Service Note
SD730-424END*D

CW(Sector) When CW
CRT: Xario (17”)
LCD: AplioXG (19”)
2 When Pencil or Pencil Option:
Pencil CB PWB Only for AplioXG XarioXG (19”, 17”)
2 CW Board and AplioMX Xario (17”, 15”)

32 IP PWB Monitor
128 TX0 PWB IQDAS
Stored to HDD
SVGA
Control
(TX:64 64 Transmit
RX:64)
32 TX1 PWB 2 x TX
Stored RAW data (800x600)
Probe A Transmit (Temporally)

TI PWB
Transducer
RC PWB
Probe B
192
Interface
64
When CW Receiver BE PWB IO PWB
Control Back End In/Out I/F
192 64
VCR
Probe C RX PWB
Receive 128 VP PWB Power ON / OFF
Control signal
Other than CW Video
(Operating / TCS)
Processor
MC PWB Standard Configuration: Main
Motor AplioXG, AplioMX, XarioXG
Motor Control for Option:
Panel
Overlay
4D /MD-TEE Control Xario Image
TCS Image

CPU
PWB E-IDE
or
Power Supply USB SATA

AplioXG (4 units) AplioMX ATAPI > USB E-IDE:


HDD AXG V2.2 and
1. Trans Unit XarioXG before
DVD/CD XXG V1.2 and before
2. PSA: Power Supply Analog Xario Driver Xario V6.2 and before
Standard 3. PSD: Power Supply Digital One unit
On and before V5 Xario system is
Option 4. PHV: Power High Voltage (OP) only support CD drive.
12
Signal path for each mode Service Note
SD730-424END*D

CW(Pencil) When CW
CRT: Xario (17”)
LCD: AplioXG (19”)
2 When Pencil or Pencil Option:
Pencil CB PWB Only for AplioXG XarioXG (19”, 17”)
2 CW Board and AplioMX Xario (17”, 15”)

96 IP PWB Monitor
TX0 PWB IQDAS
Stored to HDD
SVGA
192 Transmit
192
96 TX1 PWB Stored RAW data (800x600)
Probe A Transmit (Temporally)

TI PWB
Transducer
RC PWB
Probe B
192
Interface
64
When CW Receiver BE PWB IO PWB
Control Back End In/Out I/F
192 192
VCR
Probe C RX PWB
Receive 128 VP PWB Power ON / OFF
Control signal
Other than CW Video
(Operating / TCS)
Processor
MC PWB Standard Configuration: Main
Motor AplioXG, AplioMX, XarioXG
Motor Control for Option:
Panel
Overlay
4D /MD-TEE Control Xario Image
TCS Image

CPU
PWB E-IDE
or
Power Supply USB SATA

AplioXG (4 units) AplioMX ATAPI > USB E-IDE:


HDD AXG V2.2 and
1. Trans Unit XarioXG before
DVD/CD XXG V1.2 and before
2. PSA: Power Supply Analog Xario Driver Xario V6.2 and before
Standard 3. PSD: Power Supply Digital One unit
On and before V5 Xario system is
Option 4. PHV: Power High Voltage (OP) only support CD drive.
13
Main panel Service Note
SD730-424END*D

System Unit Common design for all system

USB HUB(4Port)

Full Trackball Touch Panel


Keyboard LCD LCD Control
A/D Board
USB ASSP

Panel CPU

Power/ SW Rotary encoder Buzzer Set/


Power Standby/ /STC VR Next/
SW LED
SW Battery LED Wheel SW
14
Hardware turning point Service Note
SD730-424END*D

We have several turning point based on hardware point of view.


SSA-790A SSA-780A SSA-680A SSA-660A
Totally different V6.0 and later
N/A N/A N/A
hard ware UIUG-660A
V3.0 and later V2.0 and later V7.0 and later
CPU change *1 N/A
UIUG-793A UIUG-682A UIUG-667A

IO changed *2 N/A V2.0 and later V4.0 and later V9.0 and later

AplioMX
N/A V2.0 and later V4.0 and later V9.0 and later
BP integration
CPU change V5.21*R404 V2.21*R404 V4.21*R404 V9.21*R404
2nd phase *3 and later and later and later and later

Note:
*1: HDD I/F also changed E-IDE to SATA.
*2: UIIO-780A for support analog out put for external monitor.
For china model UIIO-780A is standard configuration.
*3: Graphic board changed. Standard graphic card is not support 4D function.

15
Service Note
SD730-424END*D

Cable connections <SSA-790A>

16
Cable connection (SSA-790A: AplioXG) Service Note
SD730-424END*D

No. Drawing number Connection points No. Drawing number Connection points
1 CM30-35353 PSA cooling fan cable 27 CM30-35476 DVD/CD power cable

2 CM30-35483 Front OEM cooling fan cable 28 CM30-35624 DVD/CD signal cable
CM30-35354 Front OEM cooling fan connection cable 29 CM30-35474 Footswitch cable
3
CM30-35683 PSA DC power cable 30 CM30-35357 PHYSIO to IO cable
4
CM30-35462 PSA control cable 31 CM30-35480 Mecha 4D transducer/pencil transducer cable
5
CM30-35359 IO rear panel USB cable (V2.20 or earlier)
6 CM30-35453 PSD DC power cable 32
CM30-36600 IO rear panel USB cable (V3.00 or later)
7 CM30-35463 PSD control cable
33 CM30-35498 Composite video cable
8 CM30-35680 PSD output cable
34 CM30-35499 Y/C cable for connecting the SVHS VCR
9 CM30-35455 PSA output cable
35 CM30-35497 Ethernet cable
10 CM30-33022 PSA DC power cable (VTX)
36 CM30-35398 Speaker cable
11 CM30-35456 PSA DC power cable (P3.3AQ)
CM30-32988 MA_LAN cable (V2.20 or earlier)
12 CM30-35457 PSA DC power cable (P5AQ) 37
CM30-35520 MA_LAN cable (V3.00 or later)
13 CM30-35684 HV power cable CM30-35289 USB cable for header (V2.20 or earlier)
CM30-35399 HV DC cable 38
14 CM30-36368 USB cable for header (V3.00 or later)
15 CM30-35661 PSH control output cable CM30-35356 USB hookup cable
39
16 CM30-35662 PSH control cable 40 CM30-33062 B&W printer data cable (USB-USB)
17 CM30-35344 System cooling fan cable 41 CM30-33063 Color printer data cable (USB-USB)
18 CM30-35454 PSI control cable 42 CM30-33052 Audio IO to VCR cable
19 CM30-35346 LCD AC power cable 43 CM30-33053 Audio VCR to IO cable
20 CM30-35461 Panel power cable 44 CM30-33054 Y/C OUT: IO to VCR IN cable
21 CM30-35348 LCD cable 45 CM30-33055 Y/C playback: VCR OUT to IO cable

22 CM30-35484 Control panel IO power D-sub cable 46 CM30-33058 VCR remote control: Panasonic: RS232C cable
CM30-35489 Control panel IO RGB cable 47 CM30-33076 VCR remote control: Mitsubishi: RS232C cable
23
CM30-35351 Control panel IO USB data cable 48 CM30-33081 VCR remote control: Sony: RS232C cable
24
CM30-35471 HDD data cable (V2.20 or earlier) 49 BSM31-2587 AC power cable (100 V)
25 50 BSM31-2588 AC power cable (200 V)
CM30-36645 HDD data cable (V3.00 or later)
51 CM30-36580 Front panel USB cable
CM30-35472 HDD power cable (V2.20 or earlier)
26 52 CM30-36367 USB cable for header
CM30-36691 HDD power cable (V3.00 or later)
17
Cable connection (SSA-790A: AplioXG on and later V3.00) Service Note
SD730-424END*D

18
Cable connection (SSA-790A: AplioXG on and later V3.00) Service Note
SD730-424END*D

IO PWB
J8: Y/C IN J7: Y/C OUT J9: Main Panel
J50: Main panel J51: Rear Panel (From VCR) (To VCR) (Display TCS image)
J22: External monitor (Control signal) (USB2)

J17:DVD VSR (Video)


J21: Y/C Video
(to Rear Panel) S3: Video format select for J22
(Left: SVGA,Right: NTSC/PAL)
J1: LCD
Monitor
(800x600)
J18: B/W printer

J20: Composite
Video J15: Color printer
(to Rear Panel) J11: COM1
J12: COM2 Control VCR

J2: Main panel

J25: To rear panel


(iAssist dongle)
J23: Speaker

J27:DVD drive

J13:DVD VSR J10: UJUR

J28: Front USB panel


J19: Rear Panel
(LAN)
J5/J6: Audio IN L/R J4/J3: Audio OUT L/R
(From VSR/VCR) (To VSR/VCR)

: No to use

19
Cable connection (SSA-790A: AplioXG on and later V3.00) Service Note
SD730-424END*D

RP

J171: LAN (Ethernet)


J175: Composite Video

J175: Y/C Video

USB1: iAssist

USB2

20
Cable connection (SSA-790A: AplioXG on and before V2.20) Service Note
SD730-424END*D

21
Cable connection (SSA-790A: AplioXG on and before V2.20) Service Note
SD730-424END*D

IO PWB
J8: Y/C IN J7: Y/C OUT J9: Main Panel
J22: External monitor J50: Main panel (From VCR) (To VCR) (Display TCS image)
(Control signal)

J17:DVD VSR (Video)


J21: Y/C Video
(to Rear Panel) S3: Video format select for J22
(Left: SVGA,Right: NTSC/PAL)
J1: LCD
Monitor
(800x600)
J18: B/W printer

J20: Composite
Video J15: Color printer
(to Rear Panel) J11: COM1
J12: COM2 Control VCR

J2: Main panel

J25: To rear panel


(iAssist dongle)
J23: Speaker

J27:DVD drive

J13:DVD VSR J10: UJUR

J16: Rear Panel USB


J19: Rear Panel
(LAN)
J5/J6: Audio IN L/R J4/J3: Audio OUT L/R
(From VSR/VCR) (To VSR/VCR)

: No to use

22
Cable connection (SSA-790A: AplioXG on and before V2.20) Service Note
SD730-424END*D

RP

J172: USB upper


USB Lower (iAssist)

J171: LAN (Ethernet)


J175: Composite Video

J175: Y/C Video

23
Service Note
SD730-424END*D

Cable connections <SSA-780A>

24
Cable connection (SSA-780A: AplioMX) Service Note
SD730-424END*D

Drawing number Connection points Drawing number Connection points


1 CM30-33052 GR1 Audio IO to VCR 28 CM30-33020 GR1 PSU Control
2 CM30-33053 GR1 Audio VCR to IO 29 CM30-37879 GR1 SYSTEM Cooling Fan
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 30 BSM31-2587 GR1 AC POWER CABLE (100 V)
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 31 BSM31-2588 GR1 AC POWER CABLE (200 V)
VCR RemoteControl: MITSUBISHI: 32 BSM31-5917 GR1 AC POWER CABLE (200 V CHINA)
5 CM30-33076 GR1 9pin-25pin Cross 33 CM30-33026 GR1 Status LED
VCR RemoteControl: SONY: 34 BSM31-4617 JELLY WARMER (AC100V)
6 CM30-33081 GR2 9pin-9pin Cross 35 BSM31-5903 JELLY WARMER (AC200V)
VCR RemoteControl: JVC: 36 BSM31-6195 JELLY WARMER (AC100V JAPAN)
7 CM30-33081 GR3 9pin-9pin Cross 37 PM30-32700 GR1 Pencil Probe Panel - TI Board
8 CM30-38184 GR1 OEM AC Triple Cable (B/W,OEM) 38 CM30-38249 GR1 OEM Twin AC CABLE
9 CM30-38006 GR1 B&W Printer Data (USB-USB) 39 CM30-36581 GR1 Front USB
10 CM30-38007 GR1 ColorPrinter Data (USB-USB)
11 CM30-35966 GR1 MonitorVideo
12 CM30-35967 GR1 Monitor AC Power
13 CM30-33051 GR1 ControlPanel-IO Power D-Sub CABLE
14 CM30-35976 GR1 ControlPanel-IO USB DATA CABLE
15 CM30-33056 GR1 ControlPanel-IO RGB CABLE
16 CM30-33028 GR1 ControlPanel DC POWER CABLE
17 PM30-37900 GR1 FOOT-SW CABLE
18 CM30-37003 GR1 HDD Power Cable
19 CM30-36645 GR1 HDD Data Cable
20 CM30-35924 GR1 DVD Power Cable
21 CM30-36100 GR1 DVD USB Data Cable
22 CM30-33057 GR1 PHYSIO to IO
23 CM30-33021 GR1 PSA DC POWER CABLE
24 CM30-33024 GR1 PSA DC POWER CABLE (P5AQ)
25 CM30-33023 GR1 PSA DC POWER CABLE (P3.3AQ)
26 CM30-33022 GR1 PSA DC POWER CABLE (VTX0, 1)
27 CM30-33025 GR1 PSD DC POWER CABLE

25
Cable connection (SSA-780A: AplioMX on and later V2.00) Service Note
SD730-424END*D

26
Cable connection (SSA-780A: AplioMX on and later V2.00) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)

J5/J6: Audio IN L/R


(From VSR/VCR) J19: LAN
J17: Digital video

J10: UJUR

J27:DVD drive

J2: Main panel


J11: COM1
Control VCR J12: COM2
J28: Front J21: Y/C Video
USB panel

J15: Color J20: Composite


J1: LCD printer Video
Monitor
(800x600)
J18: B/W
printer

S3: Video format select for J22


(Left: NTSC/PAL, Right: SVGA)

J9: Main Panel J50: Main panel


(Display TCS image) (Control signal) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)
J51: Card / Barcode
: No to use reader
(Only Japanese market)

: UIIO required. Note: UIIO is standard configuration for China model (/P1 and /P2).
27
Cable connection (SSA-780A: AplioMX on and before V1.20) Service Note
SD730-424END*D

28
Cable connection (SSA-780A: AplioMX on and before V1.20) Service Note
SD730-424END*D

J4/J3: Audio OUT L/R

IO PWB (To VSR/VCR)

J5/J6: Audio IN L/R


J17: Digital video (From VSR/VCR)
J19: LAN

J10: UJUR

J27:DVD drive

J2: Main panel


J11: COM1
Control VCR
J12: COM2 J28: Front
USB panel

J15: Color
J1: LCD printer
Monitor
(800x600)
J18: B/W
printer

J9: Main Panel J50: Main panel J22: External monitor


(Display TCS image) J7: Y/C OUT J8: Y/C IN (Control signal)
(To VCR) (From VCR)

J51: Rear Panel


(Card / Barcode reader,
: No to use only Japanese market)

29
Service Note
SD730-424END*D

Cable connections <SSA-680A>

30
Cable connection (SSA-680A: XarioXG from V2.00 to V4.00 system) Service Note
SD730-424END*D

SSA-680A from V2.00 to V4.00 system


Drawing number Connection points Drawing number Connection points
(CRT) (CRT)
1 CM30-33052 GR1 Audio IO to VCR 25 CM30-33024 GR1 PSA DC Power Cable (P5AQ)
2 CM30-33053 GR1 Audio VCR to IO 26 CM30-33023 GR1 PSA DC Power Cable (P3.3AQ)
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 27 CM30-33022 GR1 PSA DC Power Cable (VTX0, 1)
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 28 CM30-33025 GR1 PSD DC Power Cable
5 CM30-33058 GR1 VCR RemoteControl: 29 CM30-33020 GR1 PSU Control
Panasonic: 9 pin - 9 pin Straight 30 CM30-33031 GR1 System Cooling Fan
6 CM30-33076 GR1 VCR RemoteControl: 31 BM31-2587 GR1 AC Power Cable (100 V)
MITSUBISHI: 9 pin - 25 pin Cross
32 BM31-2588 GR1 AC Power Cable (200 V)
7 CM30-33081 GR1 VCR RemoteControl:
33 BM31-5917 GR1 AC Power Cable (200 V, China)
SONY: 9 pin - 9 pin Cross
34 CM30-33026 GR1 Status LED
8 CM30-33734 GR1 OEM AC Cable (VIDEO/MO)
35 BSM31-3234E Battery Pack
9 CM30-35977 GR1 B&W Printer Data (USB-USB)
36 BSM31-4617 Gel Warmer (AC 100 V)
10 CM30-35978 GR1 ColorPrinter Data (USB-USB)
37 BSM31-5903 Gel Warmer (AC 200 V)
11 CM30-33732 GR1 OEM AC Cable (B/W,C-Printer)
38 CM30-33029 GR1 OEM Outlet
12 CM30-35966 GR1 MonitorVideo
39 CM30-33046 GR1 Pencil Probe Panel - TI Board
13 CM30-35967 GR1 Monitor AC Power
14 CM30-33051 GR1 ControlPanel-IO Power D-Sub Cable
15 CM30-35976 GR1 ControlPanel-IO USB Data Cable
16 CM30-33056 GR1 ControlPanel-IO RGB Cable
17 CM30-33028 GR1 ControlPanel DC Power Cable
18 CM30-33045 GR1 Foot-SW Cable
19 CM30-37003 GR1 HDD Power Cable
20 CM30-36645 GR1 HDD Data Cable
21 CM30-35924 GR1 DVD Power Cable
22 CM30-36100 GR1 DVD USB Data Cable
23 CM30-33057 GR1 PHYSIO to IO
24 CM30-33021 GR1 PSA DC Power Cable
31
Cable connection (SSA-680A: XarioXG on and later V4.00) Service Note
SD730-424END*D

32
Cable connection (SSA-680A: XarioXG on and later V4.00) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)

J5/J6: Audio IN L/R


J17: Digital video (From VSR/VCR) J19: LAN
for VSR (UIVS-001A)
J10: UJUR

J27:DVD drive

J13:DVD VSR

J25: iAssist dongle

J2: Main panel


J11: COM1
Control VCR J12: COM2
J28: Front J21: Y/C Video
USB panel

J15: Color J20: Composite


J1: LCD printer Video
Monitor
(800x600)
J18: B/W
printer

S3: Video format select for J22


(Left: NTSC/PAL, Right: SVGA)

J9: Main Panel J50: Main panel


(Display TCS image) (Control signal) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)
J51: Card / Barcode
reader
: No to use
(Only Japanese market)

: UIIO required. : These connecters could be access without disassembling the system.
Note: UIIO is standard configuration for China model (/P1 and /P2). Open small cover on right side of the system. 33
Cable connection (SSA-680A: XarioXG from V2.00 till V3.20) Service Note
SD730-424END*D

34
Cable connection (SSA-680A: XarioXG from V2.00 till V3.20) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)
J17: Digital video
for VSR (UIVS-001A) J5/J6: Audio IN L/R
(From VSR/VCR) J19: LAN

J10: UJUR

J27:DVD drive

J13:DVD VSR

J25: iAssist dongle

J11: COM1
Control VCR J2: Main panel

J12: COM2 J21: Y/C Video

J1: LCD J15: Color


Monitor printer
(800x600)
J18: B/W
printer J20: Composite
Video

S3: Video format select for J22


(Left: NTSC/PAL, Right: SVGA)

J9: Main Panel J50: Main panel


(Display TCS image) (Control signal) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)

: No to use J51: Card / Barcode


reader
(Only Japanese market)
: These connecters could be access without disassembling the system.
Open small cover on right side of the system. 35
Cable connection (SSA-680A: XarioXG on and before V1.20) Service Note
SD730-424END*D

On and before V1.20 system


Drawing number Connection points Drawing number Connection points
(CRT) (CRT)
1 CM30-33052 GR1 Audio IO to VCR 25 CM30-33024 GR1 PSA DC Power Cable (P5AQ)
2 CM30-33053 GR1 Audio VCR to IO 26 CM30-33023 GR1 PSA DC Power Cable (P3.3AQ)
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 27 CM30-33022 GR1 PSA DC Power Cable (VTX0, 1)
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 28 CM30-33025 GR1 PSD DC Power Cable
5 CM30-33058 GR1 VCR RemoteControl: 29 CM30-33020 GR1 PSU Control
Panasonic: 9 pin - 9 pin Straight 30 CM30-33031 GR1 System Cooling Fan
6 CM30-33076 GR1 VCR RemoteControl: 31 BM31-2587 GR1 AC Power Cable (100 V)
MITSUBISHI: 9 pin - 25 pin Cross
32 BM31-2588 GR1 AC Power Cable (200 V)
7 CM30-33081 GR1 VCR RemoteControl:
33 BM31-5917 GR1 AC Power Cable (200 V, China)
SONY: 9 pin - 9 pin Cross
34 CM30-33026 GR1 Status LED
8 CM30-33734 GR1 OEM AC Cable (VIDEO/MO)
35 BSM31-3234E Battery Pack
9 CM30-35977 GR1 B&W Printer Data (USB-USB)
36 BSM31-4617 Gel Warmer (AC 100 V)
10 CM30-35978 GR1 ColorPrinter Data (USB-USB)
37 BSM31-5903 Gel Warmer (AC 200 V)
11 CM30-33732 GR1 OEM AC Cable (B/W,C-Printer)
38 CM30-33029 GR1 OEM Outlet
12 CM30-35966 GR1 MonitorVideo
39 CM30-33046 GR1 Pencil Probe Panel - TI Board
13 CM30-35967 GR1 Monitor AC Power
14 CM30-33051 GR1 ControlPanel-IO Power D-Sub Cable
15 CM30-35976 GR1 ControlPanel-IO USB Data Cable
16 CM30-33056 GR1 ControlPanel-IO RGB Cable
17 CM30-33028 GR1 ControlPanel DC Power Cable
18 CM30-33045 GR1 Foot-SW Cable
19 CM30-33042 GR1 HDD Power Cable
20 CM30-33043 GR1 HDD Data Cable
21 CM30-35924 GR1 DVD Power Cable
22 CM30-36100 GR1 DVD USB Data Cable
23 CM30-33057 GR1 PHYSIO to IO
24 CM30-33021 GR1 PSA DC Power Cable
36
Cable connection (SSA-680A: XarioXG on and before V1.20) Service Note
SD730-424END*D

37
Cable connection (SSA-680A: XarioXG on and before V1.20) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)
J17: Digital video
for VSR (UIVS-001A) J5/J6: Audio IN L/R J16: Card / Barcode reader
(From VSR/VCR) (Only Japanese market) J19: LAN

J10: UJUR

J27:DVD drive

J13:DVD VSR

J25: iAssist dongle

J11: COM1
Control VCR J2: Main panel

J12: COM2 J21: Y/C Video

J1: LCD J15: Color


Monitor printer
(800x600)
J18: B/W
printer J20: Composite
Video

S3: Video format select for J22


(Left: NTSC/PAL, Right: SVGA)

J9: Main Panel J50: Main panel


(Display TCS image) (Control signal) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)

: No to use

: These connecters could be access without disassembling the system.


Open small cover on right side of the system. 38
Service Note
SD730-424END*D

Cable connections <SSA-660A>

39
Cable connection (SSA-660A CRT: Xario CRT from V7.00 till V9.00) Service Note
SD730-424END*D

From V7.00 till V9.00 system


Drawing number Connection points Drawing number Connection points
(CRT) (CRT)
1 CM30-33052 GR1 Audio IO to VCR 25 CM30-33024 GR1 PSA DC Power Cable (P5AQ)
2 CM30-33053 GR1 Audio VCR to IO 26 CM30-33023 GR1 PSA DC Power Cable (P3.3AQ)
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 27 CM30-33022 GR1 PSA DC Power Cable (VTX0, 1)
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 28 CM30-33025 GR1 PSD DC Power Cable
5 CM30-33058 GR1 VCR RemoteControl: 29 CM30-33020 GR1 PSU Control
Panasonic: 9 pin - 9 pin Straight 30 CM30-33031 GR1 System Cooling Fan
6 CM30-33076 GR1 VCR RemoteControl: 31 BM31-2587 GR1 AC Power Cable (100 V)
MITSUBISHI: 9 pin - 25 pin Cross
32 BM31-2588 GR1 AC Power Cable (200 V)
7 CM30-33081 GR1 VCR RemoteControl:
33 BM31-5917 GR1 AC Power Cable (200 V, China)
SONY: 9 pin - 9 pin Cross
34 CM30-33026 GR1 Status LED
8 CM30-33734 GR1 OEM AC Cable (VIDEO/MO)
35 BSM31-3234E Battery Pack
9 CM30-35977 GR1 B&W Printer Data (USB-USB)
36 BSM31-4617 Gel Warmer (AC 100 V)
10 CM30-35978 GR1 ColorPrinter Data (USB-USB)
37 BSM31-5903 Gel Warmer (AC 200 V)
11 CM30-33732 GR1 OEM AC Cable (B/W,C-Printer)
38 CM30-33029 GR1 OEM Outlet
12 CM30-33050 GR1 MonitorVideo
39 CM30-33046 GR1 Pencil Probe Panel - TI Board
13 CM30-33027 GR1 Monitor AC Power
14 CM30-33051 GR1 ControlPanel-IO Power D-Sub Cable
15 CM30-35976 GR1 ControlPanel-IO USB Data Cable Note:
16 CM30-33056 GR1 ControlPanel-IO RGB Cable The CRT model SSA-660A has been discontinued since Apr 2008 production.
But one of China model still using CRT model.
17 CM30-33028 GR1 ControlPanel DC Power Cable
18 CM30-33045 GR1 Foot-SW Cable
19 CM30-37003 GR1 HDD Power Cable
20 CM30-36645 GR1 HDD Data Cable
21 CM30-35924 GR1 DVD Power Cable
22 CM30-36100 GR1 DVD USB Data Cable
23 CM30-33057 GR1 PHYSIO to IO
24 CM30-33021 GR1 PSA DC Power Cable
40
Cable connection (SSA-660A LCD: Xario LCD from V7.00 till V9.00) Service Note
SD730-424END*D

From V7.00 till V9.00 system


Drawing number Connection points Drawing number Connection points
(CRT) (CRT)
1 CM30-33052 GR1 Audio IO to VCR 25 CM30-33024 GR1 PSA DC Power Cable (P5AQ)
2 CM30-33053 GR1 Audio VCR to IO 26 CM30-33023 GR1 PSA DC Power Cable (P3.3AQ)
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 27 CM30-33022 GR1 PSA DC Power Cable (VTX0, 1)
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 28 CM30-33025 GR1 PSD DC Power Cable
5 CM30-33058 GR1 VCR RemoteControl: 29 CM30-33020 GR1 PSU Control
Panasonic: 9 pin - 9 pin Straight 30 CM30-33031 GR1 System Cooling Fan
6 CM30-33076 GR1 VCR RemoteControl: 31 BM31-2587 GR1 AC Power Cable (100 V)
MITSUBISHI: 9 pin - 25 pin Cross
32 BM31-2588 GR1 AC Power Cable (200 V)
7 CM30-33081 GR1 VCR RemoteControl:
33 BM31-5917 GR1 AC Power Cable (200 V, China)
SONY: 9 pin - 9 pin Cross
34 CM30-33026 GR1 Status LED
8 CM30-33734 GR1 OEM AC Cable (VIDEO/MO)
35 BSM31-3234E Battery Pack
9 CM30-35977 GR1 B&W Printer Data (USB-USB)
36 BSM31-4617 Gel Warmer (AC 100 V)
10 CM30-35978 GR1 ColorPrinter Data (USB-USB)
37 BSM31-5903 Gel Warmer (AC 200 V)
11 CM30-33732 GR1 OEM AC Cable (B/W,C-Printer)
38 CM30-33029 GR1 OEM Outlet
12 CM30-33082 GR1 MonitorVideo
39 CM30-33046 GR1 Pencil Probe Panel - TI Board
13 CM30-34332 GR1 Monitor AC Power
14 CM30-33051 GR1 ControlPanel-IO Power D-Sub Cable
15 CM30-35976 GR1 ControlPanel-IO USB Data Cable
16 CM30-33056 GR1 ControlPanel-IO RGB Cable
17 CM30-33028 GR1 ControlPanel DC Power Cable
18 CM30-33045 GR1 Foot-SW Cable
19 CM30-37003 GR1 HDD Power Cable
20 CM30-36645 GR1 HDD Data Cable
21 CM30-35924 GR1 DVD Power Cable
22 CM30-36100 GR1 DVD USB Data Cable
23 CM30-33057 GR1 PHYSIO to IO
24 CM30-33021 GR1 PSA DC Power Cable
41
Cable connection (SSA-660A: Xario, on and later V9.00) Service Note
SD730-424END*D

42
Cable connection (SSA-660A: Xario, on and later V9.00) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)

J5/J6: Audio IN L/R


(From VSR/VCR) J19: LAN
J17: Digital video

J10: UJUR

J27:DVD drive

J25: iAssist dongle

J2: Main panel


J11: COM1
Control VCR J12: COM2
J28: Front J21: Y/C Video
USB panel

J15: Color J20: Composite


J1: LCD printer Video
Monitor
(800x600)
J18: B/W
printer

S3: Video format select for J22


(Left: NTSC/PAL, Right: SVGA)

J9: Main Panel J50: Main panel


(Display TCS image) (Control signal) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)
J51: Card / Barcode
reader
: No to use
(Only Japanese market)

: UIIO required. : These connecters could be access without disassembling the system.
Note: UIIO is standard configuration for China model (/P1 and /P2). Open small cover on right side of the system. 43
Cable connection (SSA-660A: Xario, From V7.00 till V8.20) Service Note
SD730-424END*D

44
Cable connection (SSA-660A: Xario, From V7.00 till V8.20) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)

J17: Digital video


J5/J6: Audio IN L/R
(From VSR/VCR) J19: LAN

J10: UJUR

J27:DVD drive

J25: iAssist dongle

J11: COM1
Control VCR J2: Main panel

J12: COM2 J21: Y/C Video

J1: LCD J15: Color


Monitor printer
(800x600)
J18: B/W
printer J20: Composite
Video

S3: Video format select for J22


(Left: NTSC/PAL, Right: SVGA)

J9: Main Panel J50: Main panel


(Display TCS image) (Control signal) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)

: No to use J51: Card / Barcode


reader
(Only Japanese market)
: These connecters could be access without disassembling the system.
Open small cover on right side of the system. 45
Cable connection (SSA-660A CRT: Xario CRT from V6.00 till V6.20) Service Note
SD730-424END*D

From V6.00 till V6.20 system


Drawing number Connection points Drawing number Connection points
(CRT) (CRT)
1 CM30-33052 GR1 Audio IO to VCR 25 CM30-33024 GR1 PSA DC Power Cable (P5AQ)
2 CM30-33053 GR1 Audio VCR to IO 26 CM30-33023 GR1 PSA DC Power Cable (P3.3AQ)
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 27 CM30-33022 GR1 PSA DC Power Cable (VTX0, 1)
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 28 CM30-33025 GR1 PSD DC Power Cable
5 CM30-33058 GR1 VCR RemoteControl: 29 CM30-33020 GR1 PSU Control
Panasonic: 9 pin - 9 pin Straight 30 CM30-33031 GR1 System Cooling Fan
6 CM30-33076 GR1 VCR RemoteControl: 31 BM31-2587 GR1 AC Power Cable (100 V)
MITSUBISHI: 9 pin - 25 pin Cross
32 BM31-2588 GR1 AC Power Cable (200 V)
7 CM30-33081 GR1 VCR RemoteControl:
33 BM31-5917 GR1 AC Power Cable (200 V, China)
SONY: 9 pin - 9 pin Cross
34 CM30-33026 GR1 Status LED
8 CM30-33734 GR1 OEM AC Cable (VIDEO/MO)
35 BSM31-3234E Battery Pack
9 CM30-35977 GR1 B&W Printer Data (USB-USB)
36 BSM31-4617 Gel Warmer (AC 100 V)
10 CM30-35978 GR1 ColorPrinter Data (USB-USB)
37 BSM31-5903 Gel Warmer (AC 200 V)
11 CM30-33732 GR1 OEM AC Cable (B/W,C-Printer)
38 CM30-33029 GR1 OEM Outlet
12 CM30-33050 GR1 MonitorVideo
39 CM30-33046 GR1 Pencil Probe Panel - TI Board
13 CM30-33027 GR1 Monitor AC Power
14 CM30-33051 GR1 ControlPanel-IO Power D-Sub Cable
15 CM30-35976 GR1 ControlPanel-IO USB Data Cable
16 CM30-33056 GR1 ControlPanel-IO RGB Cable
17 CM30-33028 GR1 ControlPanel DC Power Cable
18 CM30-33045 GR1 Foot-SW Cable
19 CM30-33042 GR1 HDD Power Cable
20 CM30-33043 GR1 HDD Data Cable
21 CM30-35924 GR1 DVD Power Cable
22 CM30-36100 GR1 DVD USB Data Cable
23 CM30-33057 GR1 PHYSIO to IO
24 CM30-33021 GR1 PSA DC Power Cable
46
Cable connection (SSA-660A LCD: Xario LCD from V6.00 till V6.20) Service Note
SD730-424END*D

From V6.00 till V6.20 system


Drawing number Connection points Drawing number Connection points
(CRT) (CRT)
1 CM30-33052 GR1 Audio IO to VCR 25 CM30-33024 GR1 PSA DC Power Cable (P5AQ)
2 CM30-33053 GR1 Audio VCR to IO 26 CM30-33023 GR1 PSA DC Power Cable (P3.3AQ)
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 27 CM30-33022 GR1 PSA DC Power Cable (VTX0, 1)
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 28 CM30-33025 GR1 PSD DC Power Cable
5 CM30-33058 GR1 VCR RemoteControl: 29 CM30-33020 GR1 PSU Control
Panasonic: 9 pin - 9 pin Straight 30 CM30-33031 GR1 System Cooling Fan
6 CM30-33076 GR1 VCR RemoteControl: 31 BM31-2587 GR1 AC Power Cable (100 V)
MITSUBISHI: 9 pin - 25 pin Cross
32 BM31-2588 GR1 AC Power Cable (200 V)
7 CM30-33081 GR1 VCR RemoteControl:
33 BM31-5917 GR1 AC Power Cable (200 V, China)
SONY: 9 pin - 9 pin Cross
34 CM30-33026 GR1 Status LED
8 CM30-33734 GR1 OEM AC Cable (VIDEO/MO)
35 BSM31-3234E Battery Pack
9 CM30-35977 GR1 B&W Printer Data (USB-USB)
36 BSM31-4617 Gel Warmer (AC 100 V)
10 CM30-35978 GR1 ColorPrinter Data (USB-USB)
37 BSM31-5903 Gel Warmer (AC 200 V)
11 CM30-33732 GR1 OEM AC Cable (B/W,C-Printer)
38 CM30-33029 GR1 OEM Outlet
12 CM30-33082 GR1 MonitorVideo
39 CM30-33046 GR1 Pencil Probe Panel - TI Board
13 CM30-34332 GR1 Monitor AC Power
14 CM30-33051 GR1 ControlPanel-IO Power D-Sub Cable
15 CM30-35976 GR1 ControlPanel-IO USB Data Cable
16 CM30-33056 GR1 ControlPanel-IO RGB Cable
17 CM30-33028 GR1 ControlPanel DC Power Cable
18 CM30-33045 GR1 Foot-SW Cable
19 CM30-33042 GR1 HDD Power Cable
20 CM30-33043 GR1 HDD Data Cable
21 CM30-35924 GR1 DVD Power Cable
22 CM30-36100 GR1 DVD USB Data Cable
23 CM30-33057 GR1 PHYSIO to IO
24 CM30-33021 GR1 PSA DC Power Cable
47
Cable connection (SSA-660A: Xario, From V6.00 till V6.20) Service Note
SD730-424END*D

48
Cable connection (SSA-660A: Xario, From V6.00 till V6.20) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)

J17: Digital video


J5/J6: Audio IN L/R J16: Card / Barcode reader
(From VSR/VCR) (Only Japanese market) J19: LAN

J10: UJUR

J27:DVD drive

J25: iAssist dongle

J11: COM1
Control VCR J2: Main panel

J12: COM2 J21: Y/C Video

J1: LCD J15: Color


Monitor printer
(800x600)
J18: B/W
printer J20: Composite
Video

S3: Video format select for J22


(Left: NTSC/PAL, Right: SVGA)

J9: Main Panel J50: Main panel


(Display TCS image) (Control signal) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)

: No to use

: These connecters could be access without disassembling the system.


Open small cover on right side of the system. 49
Cable connection (SSA-660A CRT: Xario CRT on and before V5.50) Service Note
SD730-424END*D

On and before V5.50 system


Drawing number Connection points Drawing number Connection points
(CRT) (CRT)
1 CM30-33052 GR1 Audio IO to VCR 25 CM30-33023 GR1 PSA DC Power Cable (P3.3AQ)
2 CM30-33053 GR1 Audio VCR to IO 26 CM30-33022 GR1 PSA DC Power Cable (VTX0, 1)
3 CM30-33054 GR1 Y/C OUT: IO to VCRin 27 CM30-33025 GR1 PSD DC Power Cable
4 CM30-33055 GR1 Y/C PlayBack: VCRout to IO 28 CM30-33020 GR1 PSU Control
5 CM30-33058 GR1 VCR RemoteControl: 29 CM30-33031 GR1 System Cooling Fan
Panasonic: 9 pin - 9 pin Straight 30 BM31-2587 GR1 AC Power Cable (100 V)
6 CM30-33076 GR1 VCR RemoteControl: 31 BM31-2588 GR1 AC Power Cable (200 V)
MITSUBISHI: 9 pin - 25 pin Cross
32 CM30-33026 GR1 Status LED
7 CM30-33081 GR1 VCR RemoteControl:
33 BSM31-3234E Battery Pack
SONY: 9 pin - 9 pin Cross
34 BSM31-4617 Gel Warmer (AC 100 V)
8 CM30-33734 GR1 OEM AC Cable (VIDEO/MO)
35 BSM31-5903 Gel Warmer (AC 200 V)
9 CM30-33062 GR1 B&W Printer Data (USB-USB)
36 CM30-33029 GR1 OEM Outlet
10 CM30-33063 GR1 ColorPrinter Data (USB-USB)
37 CM30-33046 GR1 Pencil Probe Panel - TI Board
11 CM30-33732 GR1 OEM AC Cable (B/W,C-Printer)
38 CM30-33079 GR1 USB HOOKUP CABLE
12 CM30-33050 GR1 MonitorVideo
39 CM30-33079 GR2 USB HOOKUP CABLE
13 CM30-33027 GR1 Monitor AC Power
14 CM30-33051 GR1 ControlPanel-IO Power D-Sub Cable
15 CM30-33061 GR1 ControlPanel-IO USB Data Cable
16 CM30-33056 GR1 ControlPanel-IO RGB Cable
17 CM30-33028 GR1 ControlPanel DC Power Cable
18 CM30-33045 GR1 Foot-SW Cable
19 CM30-33042 GR1 HDD Power Cable
20 CM30-33043 GR1 HDD Data Cable
21 CM30-33040 GR1 DVD Power Cable
22 CM30-33057 GR1 PHYSIO to IO
23 CM30-33021 GR1 PSA DC Power Cable
24 CM30-33024 GR1 PSA DC Power Cable (P5AQ)
50
Cable connection (SSA-660A: Xario, on and before V5.50) Service Note
SD730-424END*D

51
Cable connection (SSA-660A: Xario, on and before V5.50) Service Note
SD730-424END*D

IO PWB J4/J3: Audio OUT L/R


(To VSR/VCR)
J16: Card / Barcode reader
(Only Japanese market)
J17: Digital video
J5/J6: Audio IN L/R
(From VSR/VCR) J19: LAN

J10: UJUR

J25: iAssist dongle

J12:
COM2 J2: Main panel
J11: COM1
Control VCR J27: To UP PWB
J14: Main panel (J32A)
(Control signal)

J1: CRT
Monitor J20: Composite
(800x600) Video
J18: To UP PWB
(J33A)

J21: Y/C Video


S3: Video format select for J22
(Left: SVGA, Right: NTSC/PAL)

J9: Main Panel


(Display TCS image) J22: External monitor
J7: Y/C OUT J8: Y/C IN
(To VCR) (From VCR)

: No to use

: These connecters could be access without disassembling the system.


Open small cover on right side of the system. 52
Cable connection (SSA-660A: Xario, on and before V5.50) Service Note
SD730-424END*D

UP PWB

J33A: To IO PWB
(J18)

J33A

J32B: To IO PWB
(J27)

J32B J32A

J31B: Color J31A: B/W


printer J31B J31A printer

: No to use

53
Service Note
SD730-424END*D

Optional software list for AplioXG and AplioMX

54
Option (Software: AplioXG and AplioMX 1) Service Note
SD730-424END*D

Applicable version
No Option name Model name Remarks
AplioXG AplioMX
Eliminates noise components and improves the
1 Precision Imaging USPI-790A V4.0- V1.0-
visualization of tissues.
2 IASSIST kit USIA-790A V1.1- V1.0- Protocol assistant system.
Radiology contrast
3 USHR-790A V1.1- N/A CHI for radiology. VRI and MFI are included.
Harmonic Imaging
Cardiology contrast CHI for cardiology.
4 USHC-790A V1.1- N/A
Harmonic Imaging RSI is included. UJUR-790B is needed.
Contrast Harmonic Imaging USHI-
5 N/A V1.0- Kit for Radiology and cardiology.
(CHI) 770A/XA
Contrast Micro Flow
6 USCM-780A N/A V1.0- The USHI-770A/XA is required.
Imaging (MFI)
This kit corrects images in MFI mode to compensate for
7 MFI Image Stabilizer USIS-790A V4.0- V1.0- the effects of respiratory motion or unsteady
application of the transducer, providing stable images.
8 CHI-Q USCQ-790A V1.1- V1.0- UIIP-790A and USHI-770A/XA are required.
9 TDI-Q USTQ-790A V1.1- V1.0- UIIP-790A is required.
10 Stress Echo USSE-790A V1.1- V1.0- For Stress Echo. UJUR-790B / 780B is needed.
11 Fusion 3D USFR-790A V1.1- V1.0- Freehand Fusion 3D.
B/W images can be obtained with a wider field of view
12 Panoramic View USPV-790A V1.1- V1.0-
by moving the transducer in the lateral direction.
13 Dynamic Micro Slice USMS-790A V1.1- N/A For BX/AX transducers. UIHV-790A is required./A

55
Option (Software: AplioXG and AplioMX 2) Service Note
SD730-424END*D

Applicable version
No Option name Model name Remarks
AplioXG AplioMX
Automated contour Tracking for LV measurement.
USAT-790A V1.1- N/A
UJUR-790B is needed.
14 ACT ACT automatically traces
USAT-780A N/A V2.1- the inner membrane of the left ventricle and measures
the area, volume, ejection fraction, etc.
4D transducer and UIMV-791A or UIMV-780A is
15 STIC USST-790A V2.2- V1.0-
required
Enables color vascular display in Volume Imaging.
16 Volume color USMC-780A N/A V1.0- 4D transducer and UIMV-791A or UIMV-780A is
required.
17 MicroPure USMP-790A V2.01- V1.0-
18 Elastography USEL-790A V3.0- V1.0- UIIP-790A is required.
19 Measurement Registration USMR-790A V1.1- V1.0- Software for external PCs.
SCU: Verification, Storage and Print
20 DICOM USDI-790A V1.1- V1.0-
SCP: Verification
SCU: Verification, Storage, Print, MWM, MPPS, Q/ R
21 DICOM Advanced USDI-792C V1.1- V1.0- SCP: Verification, Storage
Storage SCU supports SR (Structured Report) object.

56
Option (Software: AplioXG and AplioMX 2) Service Note
SD730-424END*D

Applicable version
No Option name Model name Remarks
AplioXG AplioMX
The time required for contrast medium to reach the
22 Parametric MFI USPM-790A V5.0- N/A target region is indicated by color-coded display.
USHR-790A is required.
The border between the lumen and intima and the
border between the media and external are then traced
23 Auto-IMT USAI-790A V5.2- V2.2-
automatically, and the thickness of the intima-media
complex is displayed.

57
Service Note
SD730-424END*D

Optional software list for XarioXG and Xario

58
Option (Software: XarioXG and Xario 1) Service Note
SD730-424END*D

Applicable version
No Option name Model name Xario V6 Xario V5 Remarks
XarioXG
and later and before
This kit enables Automated Contour Tracking
1 ACT USAT-770A/XA V1.0- V6.0- V1.0-
for LV measurement.
2 Stress Echo USSE-770A/XA V1.0- V6.0- V1.0- This kit enables Stress Echo examination.
Differential THI to realize superior resolution
3 D-THI USDT-680A V1.0- N/A N/A
even in deeper region.
Speckle reduction and wide field of view for
4 ApliPure Plus USPP-680A V1.0- N/A N/A convex and
linear transducers in ApliPure mode.

*1 This kit eliminates noise components and


5 Precision Imaging USPI-680A V3.0- V9.0- N/A
improves the visualization of tissues.
B/w images can be obtained with a wider field
6 Panoramic View USPV-770A/XA V1.0- V6.0- V1.0- of view by moving the transducer in the lateral
direction.
7 IASSIST software USIA-770A/XA V1.0- V6.0- V1.0- Protocol assistant system "IASSIST".
Contrast Harmonic features for radiology and
8 CHI USHI-770A/XA V1.0- V6.0- V1.0-
cardiology.
permits small blood vessels flowing into a
tumor or the area around the tumor to be
9 Micro Flow Imaging USCM-680A V1.0- N/A N/A observed in examinations employing contrast
medium. The USHI-770A/XA is required
separately.
*1: Precision Imaging option supported on and later V9.0 Xario which producing on and later Sep. 2010.
Also V7 through V9 (producing on and before Aug. 2010) also supports the option but we need apply USUG-670A which producing on and later Sep. 2010.
59
Option (Software: XarioXG and Xario 2) Service Note
SD730-424END*D

Applicable version
No Option name Model name Xario V6 Xario V5 Remarks
XarioXG
and later and before
This kit corrects images in MFI mode to
compensate for the effects of respiratory
10 Image Stabilizer USIS-680A V3.0- N/A N/A
motion or unsteady application of the
transducer, providing stable images.
MicroPure Software for PLT-1204AT, PLT-
11 MicroPure USMP-680A V3.0- N/A N/A 1204BT, and PLT-805AT. ApliPure Plus
(USPP-680A) is needed.
A 3D image combining B/w image and color
12 Fusion 3D USFR-770A/XA V1.0- V6.0- V1.0- image can be reconstructed using a standard
transducer.

13 Basic DICOM USDI-770A/XA V1.0- V6.0- V1.0-

14 Advanced DICOM USDI-772B/XA V1.0- V6.0- V1.0-

User measurement This software is used to program calculation


15 USMR-680A V1.0- N/A N/A formulas on an external PC and install them
registration into the Xario XG.
This kit enables Spatio-Temporal Imaging
16 STIC USST-680A V1.2- V6.2- N/A correlation (STIC). UIMV-661A and PVT-
575MV are required.
Advanced
17 USDF-770A/XA STD V6.0- V1.0- Advanced DYNAMIC FLOW imaging.
DYNAMIC FLOW

60
Option (Software: XarioXG and Xario 3) Service Note
SD730-424END*D

Applicable version
No Option name Model name Xario V6 Xario V5 Remarks
XarioXG
and later and before

ApliPure Migration
18 USMG-660A N/A V6.0- V1.0- This kit enables the use of ApliPure on Xario.
kit

This kit enables speckle noise on B/W images


to be reduced and contrast resolution and
19 ApliPure USAP-770A/XA STD V6.0- V1.0- uniformity to be increased.
USMG-660A is required.
Automatically adjusts STC/GAIN and/or
20 QuickScan USQI-770A/XA STD V6.0- V1.0- Doppler zero shift/V.range through one-touch
operation.
Supports Trapezoid Scan for sector and linear
21 Trapezoid Scan USTS-770A/XA STD V6.0- V1.0-
transducers (with some exceptions).
User measurement Software for programming user measurement
22 USMR-770A/XA N/A V6.0- V1.0-
registration formulas on a PC and registering them in Xario
The border between the lumen and intima and
the border between the media and external are
23 Auto-IMT USAI-790A V4.2- V9.2- N/A
then traced automatically, and the thickness of
the intima-media complex is displayed.

61
DICOM option specification Service Note
SD730-424END*D

DICOM option functionality list


USDI-790A USDI-792C Remarks
No option
USDI-770A/XA USDI-772B/XA
1 2 3 4 5 1 2 3 4 5 1 2 3 4 5
Transfer Still/Cine/Toshiba private data to storage server N N N N N Y Y Y Y Y Y Y Y Y Y
Transfer DICOM SR data to storage server
N N N N N N N N N N Y Y Y Y Y
(Basic Text, Private)
Transfer DICOM SR data to storage server
N N N N N N N N N N Y Y Y Y N
(Enhanced)
Transfer DICOM SR data to storage server On and later V5.2 AplioXG,
(Comprehensive) N N N N N N N N N N Y Y Y Y N V2.2 AplioMX, V4.2
XarioXG and V9.2 Xario
Store Still/Cine/Toshiba private data to CD-R by DICOM Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y
Store Still/Cine/Toshiba private data to DVD+R by DICOM On and before V5.5 Xario
Y Y Y Y N Y Y Y Y N Y Y Y Y N
support only CD-R.
Query /Retrieve (Q/R SCU) N N N N N N N N N N Y Y Y Y Y
Storage Commitment (Storage Commitment SCU) N N N N N N N N N N Y Y Y Y Y
Print DICOM images (Print SCU) N N N N N Y Y Y Y Y Y Y Y Y Y
Get worklist (MWM SCU) N N N N N N N N N N Y Y Y Y Y
Send PPS (MPPS SCU) N N N N N N N N N N Y Y Y Y Y

Note!:
1: AplioXG (SSA-790A)
2: AplioMX (SSA-780A)
3: XarioXG (SSA-680A)
4: Xario V6 and later (SSA-660A)
5: Xario V5.5 and before (SSA-660A)
62
Optional software installation Service Note
SD730-424END*D

Install procedure
Advanced DICOM option should install after Basic DICOM install.
Step Operation
1 [Menu] > [Maintenance] > Press [Ctrl], [Shift] and [Next]. Enter password
2 Click [HW/SW] > [Option Install]
3 Insert optional CD to CD / DVD drive
4 Click [Install]
5 Enter optional software by 8 digits
(Incase if you have 10 digits serial number, please convert into 8 digit with encoding rule)
6 If you have several option return step 3 and do step 4 and 5.
7 Click [Home] > [Exit], system ask you to system shutting down. Click [OK].
8 Restart the system then confirm the option working or not.

Uninstall procedure
If you click [Uninstall], system uninstall it without confirmation.
Please carefully.
Step Operation
1 [Menu] > [Maintenance] > Press [Ctrl], [Shift] and [Next]. Enter password
2 Click [HW/SW] > [Option Install]
3 Select option which you want to uninstall
4 Click [Uninstall]
5 If you have several option return step 3 and do step 4.
6 Click [Home] > [Exit], system ask you to system shutting down. Click [OK].
7 Restart the system then confirm the option working or not. 63
Optional software installation Service Note
SD730-424END*D

Option serial number encoding


In case if you have 10 digits option software serial number, please cnvert
the serial number into 8 digits.
For example if option serial number is “”99B0542001”.

<1> <2> <3> <4>


99B 05 4 2001
V V V V
B 5 54 2001
<1>: First 3digits of serial number (99B) only leave alphabet. (Leave B)
<2>: Next 2 digits of serial number (05) only leave last digit. (Leave 5)
<3>: Next 1 digit of serial number (4) add 50 incase of the digit is number (0-9)
Incase the digit is alphabet (X, Y, Z), convert it;
X for 60
Y for 61
Z for 62
<4>: last 4 digits leave it no conversion.

64
Service Note
SD730-424END*D

Optional hardware list for AplioXG and AplioMX

65
Option (Hardware: AplioXG and AplioMX) Service Note
SD730-424END*D

Applicable version
No Option name Model name Remarks
AplioXG AplioMX
UICW-790A V1.1- N/A Continuous-Wave Doppler Unit for AplioXG
1 CW Unit
UICW-780A N/A V1.0- Continuous-Wave Doppler Unit for AplioMX
UJUR-790A V1.1- N/A For General country,. Including ECG cable. AplioXG
UJUR-791A V1.1- N/A For USA, Including ECG cable. AplioXG
UJUR-790B V2.2- N/A For General country,. Including ECG cable. AplioXG
2 Reference signal unit
UJUR-791B V2.2- N/A For USA, Including ECG cable. AplioXG
UJUR-780B N/A V1.0- For General country,. Including ECG cable. AplioMX

UJUR-781B N/A V1.0- For USA, Including ECG cable. AplioMX


PCG and Pulse transducer. A respiration sensor is included in
3 Sensor kit for UJUR UJUR-772A V1.1- V1.0-
the UJUR-772A but cannot be used with AplioXG, AplioMX.
UZGW-001A V1.1- V1.0- For 100V – 120V region.
UZGW-002A V1.1- V1.0- For 220V – 240V region.
4 Gel warmer UZGW-005A V1.1- V1.0- For Japan.
Oct. Oct. AXG: xxI10X5964 and later
UZGW-007A
2010 - 2010 - AMX: xxE10X3034 and later
UZFS002A V1.1- N/A
5 Foot switch
UZFS-003A V5.0- V1.0-
UAEH-002A V2.0- V2.21- For PET-511BTM and PET-512MC
6 TEE hangar kit
UAEH-770A V1.1- V1.0- For PET-510MA/MB

66
Option (Hardware: AplioXG and AplioMX) Service Note
SD730-424END*D

Applicable version
No Option name Model name Remarks
AplioXG AplioMX
Required for TDI-Q, CHI-Q, and Elastography.
7 IP board kit UIIP-790A V1.1- V1.0- Elastography supported on and later V3.0 AplioXG and all of
version AplioMX
Same function as UIMV-792A.
V5.21
UIMV-793A N/A This kit including graphic card which necessary V5.21*R404
*R404
and later system
Same function as UIMV-792A.
UIMV-792A V3.0- N/A This kit excluding graphic card which is unnecessary on and
later V3 AplioXG.
4D kit for PVT-575MV, PVT-382MV, PLT-1204MV, PVT-
8 4D kit UIMV-791A V2.0- N/A 681MV, and PVT-375MV.
This kit is also required to use the PET-511BTM.
UIMV-790A V1.1- N/A 4D kit for PVT-575MV.
Same function as UIMV-780A.
V2.21
UIMV-781A N/A This kit including graphic card which necessary V2.21*R404
*R404
and later system
4D kit for PVT-575MV, PVT-382MV, PLT-1204MV, PVT-
UIMV-780A N/A V1.0-
681MV, and PVT-375MV.
9 HV power kit UIHV-790A V1.1- N/A For BX/AX transducers. USMS-790A is required.

10 IASSIST controller kit UIIA-791A V1.1- N/A Consists of the IASSIST controller and mounting kit.

67
Option (Hardware: AplioXG and AplioMX) Service Note
SD730-424END*D

Applicable version
No Option name Model name Remarks
AplioXG AplioMX
DVD Video Streaming Recorder. Software for viewing and
11 DVD VSR UIVS-001A V2.0- N/A
editing recorded dynamic images on an external PC.
12 DVD VSR mounting kit UZRI-790A V2.0- N/A Kit for mounting UIVS-001A.
13 Peripheral mounting kit UZRI-780A N/A V1.0- Mounting VCR, DVD recorder AND Color printer.
B/W printer mounting
14 UZRI-781A N/A V1.0- Install B/W printer at 90 degrees.
kit
15 Peripheral mounting kit UZRI-782A N/A V2.0- Mounting VCR, DVD recorder OR Color printer.
Add Y/C, composite and RGB analog output terminals. (IO
16 IO Panel kit UIIO-780A N/A V2.0-
PWB)
17 17-inch LCD UILM-001A V1.0- V1.0- LCD replacement kit to 17 inch
When the system is shut down in Standby mode and then
moved with the power cable disconnected, Standby status
Oct. 2010
18 Battery support kit UEBT-780A N/A production
can be maintained.
SSA-780A systems manufactured in October 2010 or later.
(S/N 99E*******, 17B******* or later)

Note: UIIO is standard configuration for Chinese market (model name /P1 and /P2).

68
Service Note
SD730-424END*D

Optional hardware list for XarioXG and Xario

69
Option (Hardware: XarioXG and Xario) Service Note
SD730-424END*D

Applicable version
No Option name Model name Xario V6 Xario V5 Remarks
XarioXG
and later and before
1 CW Unit UICW-660A V1.0- V6.0- V1.0- Standard configuration
UJUR-660A V1.0- V6.0- V1.0- For General country,. Including ECG cable.

Reference UJUR-661A N/A N/A V2.0- For USA, Including ECG cable.
2
signal unit UJUR-660B V1.2- V6.2- N/A For General country, Including ECG cable.
UJUR-661B V1.0- V6.0- N/A For USA, Including ECG cable.
PCG and Pulse transducer. A respiration
Sensor kit for
3 UJUR-772A V1.0- V6.0- V1.0- sensor is included in the UJUR-772A but
UJUR
cannot be used with XarioXG, Xaio.
UZGW-001A V1.0- V6.0- V1.0- For 100V – 120V region.
UZGW-002A V1.0- V6.0- V1.0- For 220V – 240V region.
4 Gel warmer UZGW-005A V1.0- V6.0- V1.0- For Japan.
Oct. Oct. XXG: xxH10X4202 and later
UZGW-007A N/A
2010 - 2010 - Xario: xxE10X2753 and later
UZFS002A V1.0- V6.0- V1.0-
5 Foot switch Including foot SW connector which mounting
UZFS-003A V4.0- V9.0- N/A
on front of system.
Cannot used simultaneously with color
6 MO Drive UIMO-660A N/A N/A V1.0-
printer P900D, CP900DW
Video Required when a VCR is connected. UZRI-
7 UIVP-660A STD V6.0- V1.0-
Processor Kit 660A required.
70
Option (Hardware: XarioXG and Xario) Service Note
SD730-424END*D

Applicable version
No Option name Model name Xario V6 Xario V5 Remarks
XarioXG
and later and before
For PET-510MA/MB (PET-510MB available
UAEH-770A V1.0- V6.0- V1.0-
8 TEE hanger on and later V2.0 SSA-660A)
UAEH002A V4.21- V9.21- N/A For PET-511BTM and PET-512MC
Same function as UIMV-662A.
V4.21 V9.21 This kit including graphic card which
UIMV-663A N/A
*R404 *R404 necessary V4.21*R404 XarioXG or
V9.21*R404 Xario and later system
Same function as UIMV-662A.
This kit excluding graphic card which is
UIMV-662A V2.0- V7.0- N/A
unnecessary on and later V 2 XarioXG (V7
Xario).
9 4D kit 4D kit for PVT-575MV, PVT-375MV and
PVT-681MV.
This kit is also required to use the PET-
UIMV-661A V1.0- V6.0- N/A 511BTM.
(PET-511BTM support on and later V1.1
XarioXG and V6.1 Xario)
For XarioXG also enable to connect PVT-
382MV, PLT-1204MV.

UIMV-660A N/A N/A V3.0- 4D kit for PVT-575MV.

Incase system want to install 4D for on and


10 CPU kit UIRM-660A N/A N/A - V2.0
before V2.0 Xario. (Original version)

71
Option (Hardware: XarioXG and Xario) Service Note
SD730-424END*D

Applicable version
No Option name Model name Xario V6 Xario V5 Remarks
XarioXG
and later and before
IASSIST Consists of the IASSIST controller and
11 UIIA-661A V1.0- V6.0- V1.0-
controller kit mounting kit.
VCR mounting
12 UZRI-660A V1.0- V6.0- V1.0- Kit for mounting a VCR unit
kit
This kit permits a B/W printer to be
mounted on the system at 90º away from
B/W printer the front. (With the printer mounting kit
13 UZRI-661A V1.0- V6.0- V1.0-
mounting kit included in the standard configuration, the
printer is mounted facing toward the front of
the system.)
Dynamic images can be recorded on a DVD
14 DVD VSR UIVS-001A V1.0- N/A N/A
at high quality for up to four hours.
This kit is used to mount the DVD VSR on
UZRI-680A V1.0- N/A N/A
DVD VSR the system.
15
mounting kit This kit is used to mount the DVD VSR
UZRI-681A V1.0- N/A N/A
inside the system main unit.
17 inch LCD monitor kit for CRT model.
16 LCD kit UILM-660A N/A V6.0- V1.0-
This kit released on and later V5.0.
-V6.1 17 inch LCD monitor for old type LCD.
17 LCD kit UILM-661A N/A LCD model V1.0-
- 0833421
This kit include 17 inch IPS type LCD.

Add Y/C, composite and RGB analog output


18 IO Panel kit UIIO-780A V4.0- V9.0- N/A
terminals. (IO PWB)
Note: UIIO required. Note: UIIO is standard configuration for Chinese market (model name /P1 and /P2). 72
Service Note
SD730-424END*D

UIMV option installation

73
FAQ: UIMV-790A, UIMV-791A and UIMV-792A Service Note
SD730-424END*D

[Question]
What is difference between UIMV-790A, UIMV-791A, UIMV-792A and UIMV-793A??
[Answer]
Product information (MPIUS0475EA) describes compatibility as below.

SSA-790A
UIMV-790A UIMV-791A UIMV-792A UIMV-793A
V1.0 – V1.1 OK N/A N/A N/A
V2.0 – V2.2 N/A OK N/A N/A
V3.0 – V5.21*R403 N/A Not recommend OK Not recommend
V5.21*R404 N/A N/A N/A OK

[Additional information]
Kit composition of UIMV-790A, UIMV-791A , UIMV-792A and UIMV-793A.
MC AGP Remarks
UIMV-790A PM30-32736 Gr1 BSM31-6584 (ELSA GLADIAC FX 534 128MB)
UIMV-791A PM30-35066 Gr1 BSM31-6584 (ELSA GLADIAC FX 534 128MB) New MC PWB (MC2)
UIMV-792A PM30-35066 Gr1 N/A New MC PWB (MC2)
This kit NOT include Video card.
UIMV-793A PM30-35066 Gr1 BSM31-9362 (ELSA GLADIAC 785GT V2.0 ) New MC PWB (MC2)

* UIMV-790A (MC: PM30-32736 Gr1) is only support PVT-575MV, other 4D probes are NOT supported.
74
UIMV-79xA and UIMV-780A applicable serial number information Service Note
SD730-424END*D

Regarding UIMV-790A, 791A, 792A


System UIMV-790A UIMV-791A UIMV-792A UIMV-793A
Version V1.1 V2.0 – V2.2 V3.0 - V5.21*R404 -
SSA-790A
Serial number 06Z2001 - 0782792 0782793 - 0873725 0873726 - 99L1166355 -

We should confirm actual system version.


Because of we cannot manage a system has applied upgrade or not.
UIMV-791A can be use on and later version 3 system.
But we will be waste a graphic board which unnecessary for V3 afterward.

75
FAQ: UIMV-780A and UIMV-781A Service Note
SD730-424END*D

[Question]
What is difference between UIMV-780A and UIMV-781A??
[Answer]
Product information (MPIUS0475EA) describes compatibility as below.

SSA-780A
UIMV-780A UIMV-781A
V1.0 – V2.21*R403 OK Not recommend
V2.21*R404 N/A OK

[Additional information]
Kit composition of UIMV-780A and UIMV-781A.
MC AGP Remarks
UIMV-780A PM30-35066 Gr1 N/A New MC PWB (MC2)
This kit NOT include Video card.
UIMV-781A PM30-35066 Gr1 BSM31-9362 (ELSA GLADIAC 785GT V2.0 ) New MC PWB (MC2)

76
UIMV-79xA and UIMV-780A applicable serial number information Service Note
SD730-424END*D

Regarding UIMV-780A
System UIMV-780A UIMV-781A
Version V1.0 – V2.21*R404 -
SSA-780A
Serial number 0992001 - 1164026 -

We should confirm actual system version.


Because of we cannot manage a system has applied upgrade or not.

77
FAQ: UIMV-660A, UIMV-661A and UIMV-662A Service Note
SD730-424END*D

[Question]
What is difference between UIMV-660A, UIMV-661A and UIMV-662A??
[Answer]
Product information (MPIUS0475EA) describes compatibility as below.

SSA-680A
UIMV-660A UIMV-661A UIMV-662A UIMV-663A
V1.0 – V1.2 N/A OK N/A N/A
V2.0 – V4.21*R403 N/A Not recommend OK Not recommend
V4.21*R404 - N/A N/A N/A OK

[Additional information]
Kit composition of UIMV-660A, UIMV-661A and UIMV-662A.

MC AGP Remarks
UIMV-660A PM30-32736 Gr1 BSM31-6584 (ELSA GLADIAC FX 534 128MB)
UIMV-661A PM30-35066 Gr1 BSM31-6584 (ELSA GLADIAC FX 534 128MB) New MC PWB (MC2)
UIMV-662A PM30-35066 Gr1 N/A New MC PWB (MC2)
This kit NOT include Video card.
UIMV-663A PM30-35066 Gr1 BSM31-9362 (ELSA GLADIAC 785GT V2.0 ) New MC PWB (MC2)

* UIMV-790A (MC: PM30-32736 Gr1) is only support PVT-575MV, other 4D probes are NOT supported. 78
UIMV-66xA applicable serial number information Service Note
SD730-424END*D

Regarding UIMV-660A, 661A, 662A


System UIMV-660A UIMV-661A UIMV-662A UIMV-663A
Version N/A V1.0 – V1.2 V2.0 - V4.21*R404
SSA-680A
Serial number N/A 0792001 - 0882623 0892624 - 1164720 -

We should confirm actual system version.


Because of we cannot manage a system has applied upgrade or not.
UIMV-661A can be use on and later V2 XarioXG and V7 Xario system.
But we will be waste a graphic board which unnecessary for V2 XarioXG and V7 Xario afterward.
79
FAQ: UIMV-660A, UIMV-661A and UIMV-662A Service Note
SD730-424END*D

[Question]
What is difference between UIMV-660A, UIMV-661A and UIMV-662A??
[Answer]
Product information (MPIUS0475EA) describes compatibility as below.

SSA-660A
UIMV-660A UIMV-661A UIMV-662A UIMV-663A
V3.0 – V5.0 OK N/A N/A N/A
V6.0 – V6.2 N/A OK N/A N/A
V7.0 – V9.21*R403 N/A Not recommend OK Not recommend
V9.21*R404 - N/A N/A N/A OK
[Additional information]
Kit composition of UIMV-660A, UIMV-661A and UIMV-662A.

MC AGP Remarks
UIMV-660A PM30-32736 Gr1 BSM31-6584 (ELSA GLADIAC FX 534 128MB)
UIMV-661A PM30-35066 Gr1 BSM31-6584 (ELSA GLADIAC FX 534 128MB) New MC PWB (MC2)
UIMV-662A PM30-35066 Gr1 N/A New MC PWB (MC2)
This kit NOT include Video card.
UIMV-663A PM30-35066 Gr1 BSM31-9362 (ELSA GLADIAC 785GT V2.0 ) New MC PWB (MC2)

* UIMV-790A (MC: PM30-32736 Gr1) is only support PVT-575MV, other 4D probes are NOT supported. 80
UIMV-66xA applicable serial number information Service Note
SD730-424END*D

Regarding UIMV-660A, 661A, 662A


System UIMV-660A UIMV-661A UIMV-662A UIMV-663A
Version V3.0 - V5.5 V6.0 - V6.2 V7.0 - V9.21*R404
SSA-660A
Serial number 05X3417 - 07Y2740 07Y2741 - 0894473 08X4474 - 1164104 -

We should confirm actual system version.


Because of we cannot manage a system has applied upgrade or not.
UIMV-661A can be use on and later V2 XarioXG and V7 Xario system.
But we will be waste a graphic board which unnecessary for V2 XarioXG and V7 Xario afterward.
81
UIMV option installation Service Note
SD730-424END*D

UIMV installation differences (Except AplioMX!)


Here check point of UIMV installation.
System serial number and software version
If the system on and later V3.0 AplioXG, V2.0 XarioXG and V7.0 Xario.
We don’t need replace graphic board.
CPU part number (Xario V5.5 and before)
Confirm CPU part number which shown back side of the CPU

Part number Action for CPU board Remarks


PM30-32945 UIRM-660A required On and before 99C05X3416
PM30-34538 Replace graphic board which including UIMV On and later 99C05X3417
No need to replace CPU it self. But graphic board need to be replace.
PM30-34564 No action This CPU should replaced by UIRM-660A.

Graphic board model name or part number (This information is not always present.)
On the graphic board, if there is G55xxxx.
We need replace graphic board.

82
UIMV option installation <UIMV-790A> Service Note
SD730-424END*D

This option can be installed in SSA-790A systems with system software V1.10 system.
UIMV-790A installation V2.0 and later, please install UIMV-791A. Due to UIMV-790A only supports PVT-575MV and
many 4D function which are supported on and later V2.0 system.

Basically video card


needs to replace.

83
UIMV option installation <UIMV-791A> Service Note
SD730-424END*D

SSA-790A systems with V3.00 or later, if you have stock of UIMV-791A


UIMV-791A installation (Basically recommend you to use UIMV-792A for V3 and later SSA-790A)

SSA-790A systems with V2.00 to V2.20

Video card
needs to
replace.

On this procedure we don’t need to


change video card due to the CPU
ASSY has been changed on and later
V3.0 SSA-790A.

84
UIMV option installation <UIMV-792A> Service Note
SD730-424END*D

UIMV-792A installation SSA-790A systems with V3.00 till V5.21*R403

85
UIMV option installation <UIMV-780A> Service Note
SD730-424END*D

UIMV-780A installation SSA-780A V1.0 or later

86
UIMV option installation <UIMV-660A> Service Note
SD730-424END*D

UIMV-660A installation SSA-660A systems with versions on and before V5.50.

CPU need to replace by UIRM-660A,


if the system originally V2.0 and before

Video card needs to replace.

87
UIMV option installation <UIMV-661A> Service Note
SD730-424END*D

SSA-680A systems with V2.00 or later


UIMV-661A installation SSA-660A systems with V7.00 or later

Video card
needs to
replace.

On this procedure we don’t need to


change video card due to the CPU
ASSY has been changed on and later
V3.0 SSA-790A.

SSA-660A V7.00 or later and SSA-680A V2.00 or later system.


SSA-680A systems with versions from 1.00 to 1.20 (if you have stock of UIMV-661A, basically recommend you to use
SSA-660A systems with versions from 6.00 to 6.20 UIMV-662A for V7 and later 660A and V2 and later 680A)

88
UIMV option installation <UIMV-662A> Service Note
SD730-424END*D

SSA-680A systems with V2.00 or later


UIMV-662A installation SSA-660A systems with V7.00 or later

89
Service Note
SD730-424END*D

What is UIVS-001A (DVD-VSR)?

90
UIVS-001A (DVD-VSR) Service Note
SD730-424END*D

UIVS-001A block diagram


USB network adapter
Local Area Connection 3

Main unit (AplioXG, XarioXG) UIVS-001A


Communicated by
OS TCP/IP OS
Windows2000 (E) USB I/F USB I/F Windows XP (J)
Thumbnail image
(Connected
IO PWB) Control Signal
*: 192.168.1.1 *: 192.168.1.254 VSR applications
VSR applications
Communication Recorded image Communication
Control Control, Recording
Display image Writing
DVD Drive
DVI-D I/F DVI-D I/F
OUTPUT Digital image signal INPUT
(Connected
HDD
IO PWB) (Backup for Recorded
image before writing to
media)

*The IP address setting is factory default.


It may change depends on customer network configuration

91
UIVS-001A (DVD-VSR) Service Note
SD730-424END*D

Brief Description of the DVD VSR.


The DVD-VSR is used to record high-quality ultrasound images on DVDs in digital
format. In addition, the unit allows recorded dynamic images to be played back and
edited at the diagnostic ultrasound system. It is also possible to convert images to
Windows files using a PC in which the dedicated software (MIV-Organizer) is
installed.

CAUTION:
1. When recording diagnostic images, confirm that the displayed patient ID and patient name
are correct. Recording diagnostic images with an incorrect patient ID may result in incorrect
diagnosis.
Before performing measurements on a freeze image on the video recording unit, confirm that
the patient ID on the image is correct.

2. In this system, recording is stopped when one of the following operations is performed.
[PIMS] > [VIDEO]

> [CONFIRM Start]

(If recording is required, press .)

> [Shutdown], [Standby], or [Log Off]

92
UIVS-001A (DVD-VSR) Service Note
SD730-424END*D

Recommended Media
Writing to DVD-RAM and DVD+R is supported.
Some DVDs may not be suitable for writing, depending on the
manufacturer. Toshiba has confirmed that the DVD-VSR unit can write
data correctly on the DVDs listed below.

Media Manufacturer Remarks


DVD-RAM Panasonic (3x or 5x speed) With and W/O cartridge
DVD+R RICOH (4x, 8x, or 16x speed)

*:If a media that is not supported by the system is inserted,


the message "Not supported media" is displayed and the
media is ejected.

93
UIVS-001A (DVD-VSR) Service Note
SD730-424END*D

Name and Function of Each Section


Emergency eject hole Power supply button Power supply
(for emergency use only) indicator
Disk tray

CAUTION: Emergency shutdown


using the power supply button
R A M DVD+R
Do not press this button except in an
emergency. The internal software may be
damaged, possibly resulting in failure
during start-up.

Disk/Busy indicator Eject button 2


Eject button 1

Name Function
Disk tray A disk is placed on this tray.
Use this hole to eject the disk tray if the tray cannot be ejected by pressing the eject
Emergency eject hole
button.Please try to use Emergency eject hole after shutting down main system.
Used to turn ON the main power of the unit and to turn OFF the unit in the case of an
Power supply button emergency.The unit is shut down by holding down this button for more than 5 seconds.
Do not use this button under normal conditions.
Power supply indicator This indicator lights when the power of the unit is ON and blinks after the unit is shut down.
Disk/Busy indicator This indicator blinks when the media is being accessed.
Eject button 1 Used to open and close the disk tray when it does not contain a DVD.This button is for
Eject button 2 maintenance only. Pressing this button has no effect.
94
UIVS-001A (DVD-VSR) Service Note
SD730-424END*D

Power indicator
Emergency eject hole Power supply button Power supply
(for emergency use only) indicator
Disk tray

CAUTION: Emergency shutdown


using the power supply button
R A M DVD+R
Do not press this button except in an
emergency. The internal software may be
damaged, possibly resulting in failure
during start-up.

Disk/Busy indicator Eject button 2


Eject button 1

Status Mean
Lit ON UIVS-001A boot-up completed and made connection between main system.
Flashing (Early) UIVS-001A boot-up completed but still not made connection between main system.

UIVS-001A under standby condition.


Flashing (Slow) * While main system goes to Standby or shutdown. System will be command to go standby
for UIVS-001A unit. (This process also executed while STOPZ.JS command)

95
UIVS-001A (DVD-VSR) Service Note
SD730-424END*D

V1.37 before and


VSR Installation 1 firmware medium
V1.37 Confirm VSR include in this option
software version

Remove Network cable Reconfigure VSR settings V1.37 before but


no firmware medium

Install VSR based on Ask firmware CD


UIVS-001A installation manual Functionality check
System belongs
System will not hospital network
See belongs hospital and IP address as VSR firmware
Next slide network like 192.168.x.x update
Confirm system
Confirm IP address of LAN1
software version (Required to connect Network?)
See
Next slide System belongs hospital 192.168.1.x
Confirm IP address
network and IP address NOT of LAN1
Apply FMI-MU995 Other than
like 192.168.x.x
(Replace XML file) 192.168.20.x
192.168.1.x 192.168.20.x
No There are Gateway
VSR IP address change
for LAN1? VSR IP address based on installation manual VSR IP address
Memorize VSR settings change based on But the IP address should change based on
Yes installation manual define other than installation manual
the hospital use *1

Confirm XML file exist under N


D:¥Aplio¥Useer
Delete gateway Y Is the hospital using
setting for LAN 3 192.168.1.x ?
Y
Delete the XML file N

Connect Network cable


and check it Installation complete
1 *1: Confirm that the segment of IP address 96
UIVS-001A (DVD-VSR)
Criteria of this section Service Note
SD730-424END*D

SSA-790A:
V2.00*R1xx or higher
V2.01*R1xx or higher
SSA-790A: V2.20*R1xx or higher
V2.00*R000, to *R004 V3.00*R102 or higher
V2.01*R000, *R001 Or V4.00 and later
V2.20*R000 to *R002
V3.00*R002 SSA-680A:
V1.00*R1xx or higher
SSA-680A: V1.10*R1xx or higher
V1.00*R000, to *R002 V1.20*R1xx or higher
V1.10*R001, *R002 Or V2.00 and later
V1.20*R000 to *R002

97
Service Note
SD730-424END*D

Peripheral devices list

98
Option (Peripherals) Service Note
SD730-424END*D

Applicable version
No Device Model name Aplio Aplio XarioX Xario V6 Xario V5 Remarks
XG MX G and later and before
UP-D895(SONY) N/A N/A N/A N/A V1.0-
UP-D897(SONY) V1.1- V1.0- V1.0- V6.0- V5.0- *3
1 B/W printer
P93D(Mitsubishi) V1.1- N/A V1.0- V6.0- V5.0-
P95D(Mitsubishi) V4.0- V1.0- V4.0- V9.0- N/A
CP900D(Mitsubishi) V1.1- N/A V1.0- V6.0- V1.0- *2

Color CP30D(Mitsubishi) V2.01- V1.0- V1.0- V6.0- N/A


2
printer UP-D23MD(Sony) V1.1- V1.0- V1.0- V6.0- V1.0-
UP-D25MD(Sony) V5.0- V2.0- V4.0- V9.0- N/A
V6.0- V1.0-
3 VCR HS-MD3000-T V1.1- V1.0- V1.0-
*1 *1
V6.0- V1.0- Input Signal: S-Video
LQ-MD800 (Panasonic) V1.1- N/A V1.0-
*1 *1 DVD-RAM
DVD Video Input Signal: S-Video
4 DVO-1000MD (Sony) N/A V1.0- N/A N/A N/A
recorder DVD+RW
V9.0- Input Signal: S-Video
BD-X201M (JVC) N/A V1.0- V4.0- N/A
*1 DVD-RW
*1 UIVP-660A required
*2 Not possible install with UIMO-660A
*3 UP-D897MD possible to install for Xario V3 and V4 with installation kit (PM30-35477).

99
Applicable media (Main system / Peripherals) Service Note
SD730-424END*D

Applicable version
No Device Model name Aplio Xario V6 Xario V5 Recommend media
AplioMX XarioXG
XG and later and before
V6.0- V1.0- Input Signal: S-Video
LQ-MD800 (Panasonic) V1.1- N/A V1.0-
*1 *1 DVD-RAM
DVD Video Input Signal: S-Video
1 DVO-1000MD (Sony) N/A V1.0- N/A N/A N/A
recorder DVD+RW
V9.0- Input Signal: S-Video
BD-X201M (JVC) N/A V1.0- V4.0- N/A
*1 DVD-RW
2 DVD-VSR UIVS-001 V2.0- N/A V1.0- N/A N/A DVD-RAM, DVD+R
Main CD/DVD CD/DVD CD/DVD CD/DVD CD
3 --- CD-R, DVD+R
system V1.1- V1.0- V1.0- V6.0- V1.0-

*1 UIVP-660A required
100
Service Note
SD730-424END*D

Version up records

101
Version up record <AplioXG> Service Note
SD730-424END*D

Applicable serial
No Upgrade kit Version Remarks
number
06Y2001
1 Initial release V1.10 Refer MPIUS0360EA for more detail. Release at Nov. 2006.
(Started)
06Y2001
2 USUG-790A V1.10 > V2.00 to Refer MPIUS0370EA for more detail. Release at end of Aug. 2007.
0782792
06Z2001
V1.10, V2.00 >
3 USUG-791A to Refer MPIUS0383EA for more detail. Release at Dec. 2007.
V2.01
07Z3097
0782793
V2.00, V2.01 >
4 USUS-792A to Refer MPIUS0391EA for more detail. Release at Apr. 2008.
V2.20
0833456
Refer MPIUS0395EA for more detail. Release at Jul. 2008.
07Z3098
V2.01, V2.20 > Hardware upgrade.
5 UIUG-793A to
V3.00 Include CPU-ASSY and front USB port.
0873725
And replace RP (Rear Panel unit). Change USB connector.
0873726
Refer MPIUS0415EA for more detail. Release at May. 2009.
6 USUG-794A V3.00 > V4.00 to
0944622
0873726
V3.00, V4.00 > Refer MPIUS0446EA for more detail. Release at end of May. 2010.
7 USUG-795A to
V5.00
1055705
0873726
V3.00 – V5.00 >
8 USUG-796A to Refer MPIUS0461EA for more detail. Release at Dec. 2010.
V5.20
10Y6078

102
Version up record <AplioXG> Service Note
SD730-424END*D

Applicable serial
No Upgrade kit Version Remarks
number
0873726
V3.00 – V5.20 >
9 USUG-797A to Refer MPIUS0466EA for more detail. Release at Jan. 2011.
V5.21
10Z6101

103
Version up record <AplioXG> Service Note
SD730-424END*D

Table of upgrade path


USUG USUG USUG UIUG- USUG USUG USUG USUG
Ini Remarks
-790A -791A -792A 793A -794A -795A -796A -797A
V1.10 --- OK OK N/A N/A N/A N/A N/A N/A xxB06Y2001 till xxB0782792
V2.00 V2.00 OK OK N/A N/A N/A N/A N/A xxC0782793 till xxC07Z3097
V2.01 N/A V2.01 OK OK N/A N/A N/A N/A xxD07Z3098 till xxD0833456
V2.20 N/A N/A V2.20 OK N/A N/A N/A N/A xxE0843457 till xxE0873725
V3.00 N/A N/A N/A V3.00 OK OK OK OK xxF0873726 till xxF0944622
V4.00 N/A N/A N/A N/A V4.00 OK OK OK xxG0954623 till xxG1055705
V5.00 N/A N/A N/A N/A N/A V5.00 OK OK xxH1055706 till xxI10Y6078
V5.20 N/A N/A N/A N/A N/A N/A V5.20 OK xxJ10Z6079 till xxJ10Z6101
V5.21 N/A N/A N/A N/A N/A N/A N/A V5.21 xxK1116102 till
Up to Up to Up to Up to Up to Up to Up to Up to
Remarks
V2.00 V2.01 V2.20 V3.00 V4.00 V5.00 V5.20 V5.21

H/W upgrade
-CPU / HDD upgrading

S/W upgrade
-So that this kit only
support V3.00 system
Required new CPU / HDD

104
Version up record <AplioMX> Service Note
SD730-424END*D

Applicable serial
No Upgrade kit Version Remarks
number
0992001
1 Initial release V1.00 Refer MPIUS0425EA for more detail. Release at Sep. 2009.
(Started)
0992001
2 USUG-780A V1.00 > V1.20 to Refer MPIUS0431EA for more detail. Release at Nov. 2009.
09Y2124
0992001
V1.00, V1.20 >
3 USUG-781A to Refer MPIUS0446EA for more detail. Release at end of May. 2010.
V2.00
1052543
0992001
V1.00 – V2.00 >
4 USUG-782A to Refer MPIUS0449EA for more detail. Release at Jul. 2010.
V2.10
1072754
0992001
V1.00 – V2.10 >
5 USUG-783A to Refer MPIUS0461EA for more detail. Release at Dec. 2010
V2.20
10Y3244
0992001
V1.00 – V2.20 >
6 USUG-784A to Refer MPIUS0466EA for more detail. Release at Jan. 2011
V2.21
10Z3373
7

105
Version up record <AplioMX> Service Note
SD730-424END*D

Table of upgrade path


USUG USUG USUG USUG USUG
Ini Remarks
-780A -781A -782A -783A -784A
V1.00 --- OK OK OK OK OK xxx0992001 till xxx09Y2123
V1.20 V1.20 OK OK OK OK xxx09Y2125 till xxx1052543
V2.00 N/A V2.00 OK OK OK xxx1052544 till xxx1072753
V2.10 N/A N/A V2.10 OK OK xxx1072754 till xxx10Y3244 We cannot trace
S/N as 2124...
V2.20 N/A N/A N/A V2.20 OK xxx10Z3245 till xxx10Z3373
V2.21 N/A N/A N/A N/A V2.21 xxx1113374 till
Up to Up to Up to Up to Up to
Remarks
V1.20 V2.00 V2.10 V2.20 V2.21

106
Version up record <XarioXG> Service Note
SD730-424END*D

Applicable
No Upgrade kit Version Remarks
serial number
0792001
1 Initial release V1.00 Refer MPIUS0373EA for more detail. Release at Oct. 2007.
(Started)
0792001
2 USUG-680A V1.00 > V1.10 to Refer MPIUS0381EA for more detail. Release at Dec. 2007.
07Y2192
0792001
V1.00, V1.10 >
3 USUG-681A to Refer MPIUS0391EA for more detail. Release at Apr. 2008.
V1.20
0832397
Refer MPIUS0399EA for more detail. Release at Sep. 2008.
07Z2193
V1.10, V1.20 > Hardware upgrade.
4 UIUG-682A to
V2.00 Include CPU-ASSY and front USB port.
0882623
Change USB connection.
0892624
5 USUG-683A V2.00 > V3.00 to Refer MPIUS0422EA for more detail. Release at Jun. 2009.
0953132
0892624
V2.00, V3.00 >
6 USUG-684A to Refer MPIUS0431EA for more detail. Release at Nov. 2009.
V3.20
09Y3593
0892624
V2.00 – V3.20
7 USUG-685A to Refer MPIUS0446EA for more detail. Release at end of May. 2010.
> V4.00
1053940

107
Version up record <XarioXG> Service Note
SD730-424END*D

Applicable
No Upgrade kit Version Remarks
serial number
0892624
V2.00 – V4.00
8 USUG-686A to Refer MPIUS0461EA for more detail. Release at Dec. 2010
> V4.20
10Y4342
0892624
V2.00 – V4.20
9 USUG-687A to Refer MPIUS0466EA for more detail. Release at Jan. 2011
> V4.21
10Z4396

108
Version up record <XaioXG> Service Note
SD730-424END*D

We cannot trace
Table of upgrade path S/N as 2192...

USUG USUG UIUG- USUG USUG USUG USUG USUG


Ini Remarks
-680A -681A 682A -683A -684A -685A -686A -687A
V1.00 --- OK OK N/A N/A N/A N/A N/A N/A xxx0792001 till xxx07Y2191
V1.10 V1.10 OK OK N/A N/A N/A N/A N/A xxx07Z2193 till xxx0832397
V1.20 N/A V1.20 OK N/A N/A N/A N/A N/A xxx0842398 till xxx0882623
V2.00 N/A N/A V2.00 OK OK OK OK OK xxx0892624 till xxx0953132
V3.00 N/A N/A N/A V3.00 OK OK OK OK xxx0963133 till xxx09Y3593
V3.20 N/A N/A N/A N/A V3.20 OK OK OK xxx09Y3594 till xxx1053940
V4.00 N/A N/A N/A N/A N/A V4.00 OK OK xxx1053941 till xxx10Y4342
V4.20 N/A N/A N/A N/A N/A N/A V4.20 OK xxx10Z4343 till xxx10Z4396
V4.21 N/A N/A N/A N/A N/A N/A N/A V4.21 xxx1114397 till
Up to Up to Up to Up to Up to Up to Up to Up to
Remarks
V1.10 V1.20 V2.00 V3.00 V3.20 V4.00 V4.20 V4.21

H/W upgrade
-CPU / HDD upgrading

S/W upgrade
-So that this kit only
support V2.00 system
Required new CPU / HDD

109
Version up record <Xario V6 and later> Service Note
SD730-424END*D

Applicable serial
No Upgrade kit Version Remarks
number
Refer MPIUS0374EA for more detail. Release at end of Oct. 2007.
Initial 07Y2741
1 V6.00 Technical point of view Xario V6 system is NOT same as V5.5 and
release (Started )
before Xario system.
07Y2741 Refer MPIUS0381EA for more detail. Release at end of Dec. 2007.
to But following serial number has shipped with V6.10 software.
2 USUG-665A V6.00 > V6.10 07Y3012 So that these serial number system is NOT necessary this upgrade kit
*Some exception 07Z2956 - 07Z2957, 07Z2965 - 07Z2967, 07Z2982 - 07Z2990,
mentioned on remarks. 07Z2993 - 07Z3001, 07Z3004 - 07Z3007
07Y2741
V6.00, V6.10
3 USUG-666A to Refer MPIUS0391EA for more detail. Release at end of Apr. 2008.
> V6.20
0843642
Refer MPIUS0401EA for more detail. Release at end of Oct. 2008.
07Z3013 Hardware upgrade.Include CPU-ASSY and front USB port.
to Change USB connection.
V6.10, V6.20
4 UIUG-667A 0894473 And following serial number has shipped with V6.10 software.
> V7.00
*Some exception So that these serial number system could be apply this upgrade kit
mentioned on remarks. 07Z2956 - 07Z2957, 07Z2965 - 07Z2967, 07Z2982 - 07Z2990,
07Z2993 - 07Z3001, 07Z3004 - 07Z3007,
08X4474
5 USUG-668A V7.00 > V8.00 to Refer MPIUS0423EA for more detail. Release at end of Jun. 2009.
0965604
08X4474
V7.00, V8.00 Refer MPIUS0431EA for more detail. Release at Nov. 2009.
6 USUG-669A to
> V8.20 LGH09Y6364(15C09Y6362) -
09Y6361

110
Version up record <Xario V6 and later> Service Note
SD730-424END*D

Applicable serial
No Upgrade kit Version Remarks
number
08X4474
V7.00 – V8.20
7 USUG-670A to Refer MPIUS0446EA for more detail. Release at end of May. 2010.
> V9.00
1052054
V7.00 - V9.00 08X4474
8 USUG-671A > to Refer MPIUS0461EA for more detail. Release at Dec. 2010
V9.20 10Y3203
V7.00 - V9.20 08X4474
9 USUG-672A > to Refer MPIUS0466EA for more detail. Release at Jan. 2011
V9.21 1113357

111
Version up record < Xario V6 and later > Service Note
SD730-424END*D

Table of upgrade path


USUG USUG UIUG- USUG USUG USUG USUG USUG
V6 Remarks
-665A -666A 667A -668A -669A -670A -671A -672A
V6.00 --- OK OK N/A N/A N/A N/A N/A N/A xxx07Y2741 till xxx07Y3012 *1
V6.10 V6.10 OK OK N/A N/A N/A N/A N/A xxx07Z3013 till xxx0843642
V6.20 N/A V6.20 OK N/A N/A N/A N/A N/A xxx0843643 till xxx0894473
V7.00 N/A N/A V7.00 OK OK OK OK OK xxx08X4474 till xxx0965604
V8.00 N/A N/A N/A V8.00 OK OK OK OK xxx0965605 till xxx09Y6361
V8.20 N/A N/A N/A N/A V8.20 OK OK OK xxx09Y6362 till xxx1052054
V9.00 N/A N/A N/A N/A N/A V9.00 OK OK xxx1052055 till xxx10Y3203
V9.20 N/A N/A N/A N/A N/A N/A V9.20 OK xxx10Z3204 till xxx1113357
V9.21 N/A N/A N/A N/A N/A N/A N/A V9.21 xxx1113358 till
Up to Up to Up to Up to Up to Up to Up to Up to
Remarks
V6.10 V6.20 V7.00 V8.00 V8.20 V9.00 V9.20 V9.21

*1: But following serial number has shipped with V6.10 software.
So that these serial number system is NOT necessary this upgrade kit
H/W upgrade 07Z2956 - 07Z2957, 07Z2965 - 07Z2967, 07Z2982 - 07Z2990,
-CPU / HDD upgrading 07Z2993 - 07Z3001, 07Z3004 - 07Z3007

S/W upgrade
-So that this kit only
support V7.00 system
Required new CPU / HDD

112
Version up record <Xario V5.5 and before> Service Note
SD730-424END*D

Applicable
No Upgrade kit Version Remarks
serial number
4602001
1 Initial release V1.00 Refer MPIUS0300EA for more detail. Release at Oct. 2004.
(Started)
4602001 to
2 USUG-660A V1.00 > V2.00 Refer MPIUS0319EA for more detail. Release at Mar. 2005.
xxx0522400
4602001 to
3 USUG-661A V1.00 > V3.00 Refer MPIUS0337EA for more detail. Release at Oct. 2005.
99B05X3416
5532401 to
4 USUG-662A V2.00 > V4.00 Refer MPIUS0345EA for more detail. Release at Feb. 2006.
99C0624043
05X3417 to
5 USUG-663A V3.00 > V5.00 Refer MPIUS0352EA for more detail. Release at Jun. 2006.
0654766

This version is not longer available for production, only upgrade.


0654766 to
6 USUG-664A V4.00 > V5.50 Refer MPIUS0396EA for more detail.
07Y2740 This upgrade kit has released at Jul. 2008.
Refer MPIUS0388EA for more detail.
This upgrade kit has released at Mar. 2008.
Hardware upgrade.
V2.00 to V5.50 5522401 to
This kit will be replace whole PWB rack include all PWBs and DVD/CD
> V6.10 07Y2740 drive. At this time we couldn’t convert original user presets.
7 UIUG-660A (Means needs create user presets manually otherwise customer lost
presets,)
Refer MPIUS0404E for more detail.
V2.00 to V5.50 5522401 to
Upgrade kit specification changed. (But installation as same as above.)
> V7.00 07Y2740 This change applied UIUG-660A production is on and later Nov. 2008. 113
Version up record <Xario V5.5 and before> Service Note
SD730-424END*D

*1: Upgrade path different depends on


Table of upgrade path production of UIUG-660A

UIUG-660A *1
USUG USUG UIUG- USUG USUG
Ini Oct. Nov. Remarks
-660A -661A 662A -663A -664A 2008 and 2008 and
before *1 later *1

V1.00 --- OK OK N/A N/A N/A N/A N/A x4603001 till xxx0522400
V2.00 V2.00 OK OK N/A N/A OK OK x5532401 till xxx05X3416
V3.00 N/A V3.00 OK OK N/A OK OK xxx05X3417 till xxx0624043
V4.00 N/A N/A V4.00 OK OK OK OK xxx0624044 till xxx0654766
V5.00 N/A N/A N/A V5.00 OK OK OK xxx0664767 till xxx07X2739

No production version, this version


V5.50 N/A N/A N/A N/A V5.50 OK OK
only available with upgrade.

V6.10 N/A N/A N/A N/A N/A V6.10 N/A Upgrade from V6.1 to V7.0, use UIUG-667A

V7.00 N/A N/A N/A N/A N/A N/A V7.00


Remark Up to Up to Up to Up to Up to Up to Up to
s V2.00 V3.00 V4.00 V5.00 V5.50 V6.10 V7.00

S/W upgrade H/W upgrade


- V5.5 is only available by upgrade. - Replace Rack ASSY including CD/DVD drive
No production. - Preset data will not converting (not support)
114
Service Note
SD730-424END*D

General service work

115
General service work Service Note
SD730-424END*D

General service work


Service menu
Password list
Access key for service menu
Security mode
[Shift] with [<]
Optional software installation
IQ report
Taking log files
Screen Capture
Environment Var.
TCS calibration
Main panel test
Mandatory work for H/W replacement

116
General service work <Service Menu> Service Note
SD730-424END*D

Service Menu
How to access service menu
Press [Menu]
Select “Maintenance”
Press [Ctrl] and [Shift] and [Next] key
Enter the password. (Password will be change every year Jan. 1st.)
Pass word for 2012: 3E90D0A1E8A500 AA1D61
Select the check box for "Set as Default" to keep the license entered in the
Note: 0 means Zero, left field. From next time, the service menu can be accessed by entering the
O means Oman
right side password (8 digits) only.
The screen below is displayed.
When the version information below is displayed, the menu items can be
accessed.

117
General service work <Password list> Service Note
SD730-424END*D

Password list
Access key for service menu <from 2001 to 2013.>
Year Left side password Right side password
2001 8CAADAEC397AAB 69C3E7
2002 FCBCD091543BD4 E84661
2003 CCCED09175530A FB8A61
2004 3D90D021E83E62 040E61
2005 3DD2D021C85E64 BB0261
2006 BD94D021E9136A FA8661
2007 BDD6D021C9F24C 454B61
2008 3DB8D0213C9646 14CE61
2009 4DCAD021747E60 058261
2010 FDBCD02154B24E 444761
2011 CDCED02175C908 DB9961
2012 3E90D0A1E8A500 AA1D61
2013 3ED2D0A1C8C50E 9B1161

Note: 0 means Zero, O means Oman


118
General service work <Password list> Service Note
SD730-424END*D

Password list
Security mode (Called HIPPA.)
Security mode will be activate set Enable for Security mode at;
[Menu] > [System setting] > [Other Setting] > Page 2/2 > Press [Shift] and [<]
If Security Mode set to enable, since then system will ask you ID and PASSWORD when system boot
up completed.
Here is factory default passwords.
You can add user accounts when log in with administrator role.
Role ID Password Remarks
Operator SERVICE SERVICE&
Administrator TOSHIBA TOSHIBA& You can add, edit and delete user accounts.

119
General service work <Password list> Service Note
SD730-424END*D

Password list
[Shift] with [<]
Some of setting menu has hidden menu.
The hidden menu will be displayed with [Shift] and [<].

120
General service work (Optional software) Service Note
SD730-424END*D

Install procedure
Advanced DICOM option should install after Basic DICOM install.
Step Operation
1 [Menu] > [Maintenance] > Press [Ctrl], [Shift] and [Next]. Enter password
2 Click [HW/SW] > [Option Install]
3 Insert optional CD to CD / DVD drive
4 Click [Install]
5 Enter optional software by 8 digits
(Incase if you have 10 digits serial number, please convert into 8 digit with encoding rule)
6 If you have several option return step 3 and do step 4 and 5.
7 Click [Home] > [Exit], system ask you to system shutting down. Click [OK].
8 Restart the system then confirm the option working or not.

Uninstall procedure
If you click [Uninstall], system uninstall it without confirmation.
Please carefully.
Step Operation
1 [Menu] > [Maintenance] > Press [Ctrl], [Shift] and [Next]. Enter password
2 Click [HW/SW] > [Option Install]
3 Select option which you want to uninstall
4 Click [Uninstall]
5 If you have several option return step 3 and do step 4.
6 Click [Home] > [Exit], system ask you to system shutting down. Click [OK].
7 Restart the system then confirm the option working or not. 121
General service work (Optional software) Service Note
SD730-424END*D

Option serial number encoding


In case if you have 10 digits option software serial number, please cnvert
the serial number into 8 digits.
For example if option serial number is “”99B0542001”.

<1> <2> <3> <4>


99B 05 4 2001
V V V V
B 5 54 2001
<1>: First 3digits of serial number (99B) only leave alphabet. (Leave B)
<2>: Next 2 digits of serial number (05) only leave last digit. (Leave 5)
<3>: Next 1 digit of serial number (4) add 50 incase of the digit is number (0-9)
Incase the digit is alphabet (X, Y, Z), convert it;
X for 60
Y for 61
Z for 62
<4>: last 4 digits leave it with no conversion.

122
General service work <Preset backup: for CD/DVD> Service Note
SD730-424END*D

Preset Backup for CD/DVD


1. Go to Service Menu.
2. Click [File Maintenance] > [User Data Backup].
3. Set blank CD or DVD media to CD/DVD drive.
a. When upgrading system, please follow the instruction of upgrade kit installation manual.
4. Select “Backup to DVD/CD” for [Backup/Restore].
5. We don’t need select any value for [Drive & Dir].
6. Confirm that all of backup item ware selected for “Select Backup Category”.
7. Click [Go] then system starting
Preset backup process.
8. Wait a moment until “Backup and
Write to Media is Succeeded” displayed. 4.

5.

6.

7.
123
General service work <Preset backup: for USB memory stick> Service Note
SD730-424END*D

Preset Backup for USB memory stick


We can use USB memory stick instead of CD/DVD on and later V3 AplioXG, V1 AplioMX,
V2 XarioXG and V7 Xario. In this time we should check the USB memory stick with virus
scanning software before connecting the USB memory to system.
1. Go to Service Menu. Click [File Maintenance] > [User Data Backup].
2. Set USB memory stick.
3. Select “Backup” for [Backup/Restore].
4. Select USB memory stick drive letter for [Drive & Dir].
5. Confirm that all of backup item ware
selected for “Select Backup Category”.
6. Click [Go] then system starting 4.
Preset backup process.
7. Wait a moment until “Backup is
5.
Succeeded” displayed.
6.

7.

124
General service work <IQ report> Service Note
SD730-424END*D

IQ report
In case if customer has problem, please ask to customer to take IQ report.
IQ report will be record,
Problem image
System status including system settings, operation log etc.
These record very helpful trouble investigation for us also designer investigation.
IQ report will be taking,
Press IQ report on TCS, [Other] menu.
D:¥Aplio¥TUSService¥Data¥IQReport

Shutdown / Standby system.


D:¥Aplio¥TUSLog¥TusshutdownLog¥yyyymmdd¥IQReport

125
General service work <Log file collection> Service Note
SD730-424END*D

Taking log files (AplioXG: V3.0, AplioMX: Initial, XarioXG: V2.0, Xario: V7.0 and later )
Press [Menu] then pop-up menu displayed.
(If you skip this procedure, you cannot use Set/Next in log collecting menu.)
This procedure will be collect all necessary log files to CD / DVD or USB including IQ reports.
Set USB stick memory or blank CD-R media to CD/DVD drive.
Press IQ report which on [Other] menu, TCS.
Press [Ctrl] and [Alt] and [W] on full keyboard.

Enter tsborca (Case sensitive, so recommended press with [Shift].)


Check for “Write to USB” or “Write to CD”, then click [Collect Start].

*: If you get “IQ Report Error”, please take IQ Report first.

Wait 15 to 30 minutes, CD will be automatically eject when log collection completed.


Click [Ok] for message then click [Quit].
Confirm result of log collection before leaving from hospital. Under TEMP folder we can find 6 files as below.

126
General service work <Log file collection> Service Note
SD730-424END*D

Taking log files (AplioXG: V3.0, AplioMX: Initial, XarioXG: V2.0, Xario: V7.0 and later )

Select whether you will store the log files to CD/DVD or USB memory
stick.

You can split log files as split file size.

If you know problem occurred day, please check hear and define problem
occurred date. System only collect day after specified date as a result log
collection time shorter than normal.

Start log collection


127
General service work <Log file collection> Service Note
SD730-424END*D

Taking log files


Press [Menu] then pop-up menu displayed.
(If you skip this procedure, you cannot use Set/Next in log collecting menu.)
This procedure will be collect all necessary log files to CD / DVD including IQ reports.
Set blank CD-R media to CD/DVD drive.
Press IQ report which on [Other] menu, TCS.
Press [Ctrl] and [Alt] and [W] on full keyboard.

Enter tsborca (Case sensitive, so recommended press with [Shift].)


Check for “Write to CD”, then click [Collect Start].

*: If you get “IQ Report Error”, please take IQ Report first.

Wait 15 to 30 minutes, CD will be automatically eject when log collection completed.


Click [Ok] for message then click [Quit].
Confirm result of log collection before leaving from hospital. Under TEMP folder we can find 6 files as below.

128
General service work <Task manager> Service Note
SD730-424END*D

Task manager
This procedure is only for service purpose.
From task manager, we can launch windows explorer, mspaint and etc…
Also we can execute STOPZ command which means we can shutting down
the system in case we cannot do proper shut down procedure.
After using task manager, please shutdown and reboot the system.
Otherwise, we may get strange problem like echo image is not display etc.
Press [Ctrl] and [Alt] and [Del] key, then screen changed.
Press [ ] (Backslash) then [Task Manager] button will be appeared.
Click [Task Manager] button, then “Task Manager” window will be displayed.
Click “Application” tab.
Click [New Task], this function same as “Run” in start menu.
Enter command as what do you want to get.
Command Result
explorer Launch windows explorer.
cmd Launch command prompt.
stopz.js Execute “STOPZ” command.
mspaint Launch PAINT application. 129
General service work <Screen Capture> Service Note
SD730-424END*D

Screen Capture
We can make screen shot with [Print Screen] key.
After capture the screen, then you can past to MSPAINT application.
* For launching MSPAINT, please refer to Task Manager.
Screen capture result
2nd
monitor

1st
monitor

We can recognize that system using dual monitor for TCS and main monitor.
So that graphic board has 2 monitor connectors. Also dual monitor setting very
important for this system.

130
General service work <Screen Capture> Service Note
SD730-424END*D

Screen Capture
If you take screen capture with echo image, we cannot get echo image
correctly. Due to screen capture function will getting the screen which
generates on CPU board.
CPU board will be create banner (top and bottom of screen), overlay
characters and gray (color) scale bar.
Difference between IQ report and Screen Capture
CPU LCD monitor

Screen Capture

Overlay graphics with


chroma key color

BE
2D
CDI
Processing
Front End
echo signal
PW/CW
+
Image
combining IQ Report

131
Environment Variables Service Note
SD730-424END*D

While upgrading installation manual ask you delete environment


variables what is that?
Environment variables is that sharing a value for other tasks application .
On Aplio / Xario system using many environment variables.

One of the variable that Z_DEV_MACHINE controls designer’s mode.

If set Z_DEV_MACHINE, we can see system information such as build


version etc and we can use several windows short cuts.

132
Environment Variables Service Note
SD730-424END*D

How to set?
Press [Menu] > Select “Maintenance”
Press [Ctrl] and [Shift] and [Next] key and Enter the password.
Click [Command Line] > Execute STOPZ.JS > Wait several minutes
Right click [Start] > Select Windows explorer
Right click [My Computer] > Properties > Click [Advanced] tab
Click bottom of [New...] which belongs “System variables”.
Enter variable name and variable value as below and click [OK]
Variable Name: Z_DEV_MACHINE
Variable Value: 1
Click [OK] twice then close all windows
Restart system
System getting designer’s mode

133
Environment Variables Service Note
SD730-424END*D

How to delete?
Press [Menu] > Select “Maintenance”
Press [Ctrl] and [Shift] and [Next] key and Enter the password.
Click [Command Line] > Execute STOPZ.JS > Wait several minutes
Right click [Start] > Select Windows explorer
Right click [My Computer] > Properties > Click [Advanced] tab
Find Z_DEV_MACHINE on list of System Variables.
Please carefully that should NOT select other variable.
Click [Delete], delete the variable without any confirmation.
Click [OK] twice then close all windows
Restart system
System getting normal mode

134
TCS calibration Service Note
SD730-424END*D

Sometime TCS switch location will not corresponds touch screen


censer on TCS, in this time please try to calibration for TCS.
Go to service menu > Go to [Command Line] > Execute STOPZ.JS
Right click [Start] > Select Windows explorer and execute Calwin.exe
AplioXG
C:¥Drivers¥ControlPanel(USB)¥V1¥V10L10¥Calwin.exe
AplioMX, XarioXG, Xario V6 and later
C:¥Drivers¥ControlPanel(USB)¥V1¥ControlPanelDriverV10L16¥Calwin.exe
Xario V5 and before
C:¥Drivers¥ControlPanel(USB)¥V1-R4¥ControlPanelDriverV10L16¥Calwin.exe
A cross mark (+) appears on TCS. Press 9 times the cross mark.
Each time the position of cross mark move as next page.
Press to move the mark to the last position and then press [Enter].
Calibration is completed and the paddle switch setting screen is displayed.
Adjust direction of puddle switches then press [Enter].
Shutting down and reboot the system. Confirm TCS funciton.

135
TCS calibration Service Note
SD730-424END*D

TCS calibration screen

Puddle switch direction setting screen

136
Main panel test Service Note
SD730-424END*D

AplioXG
Press and hold down 3 x rotary encoder below for 4 seconds

Individual panel test mode running.


1. Press each switches.
Once all LEDs are lit off.
2. In this time color of LED for the switch will be change
if the switch works fine.
(Sequence of LED color: Unlit > Green > Orange >
Unlit again)
3. This individual test could be use except Set / Next,
System Power switch, and wheel switch

137
Main panel test Service Note
SD730-424END*D

AplioMX, XarioXG and Xario (Any version)


Press and hold down 3 x rotary encoder below for 4 seconds

Individual panel test mode running.


1. Press each switches.
Once all LEDs are lit off.
2. In this time color of LED for the switch will be change
if the switch works fine.
(Sequence of LED color: Unlit > Green > Orange >
Unlit again)
3. This individual test could be use except Set / Next,
System Power switch, and wheel switch

138
Mandatory work for HW replacement Service Note
SD730-424END*D

Here is the mandatory work for H/W replacement


Step Command Remarks
1 C:¥SetAplio¥10_StopAplio.bat Execute on Windows explorer from Command Line
2 Shutdown system
3 Replace H/W, reboot the system
4 <Only CPU replacement> Check BIOS timer setting Check CPU replacement work too. (See installation
5 <Only CPU replacement> Check drive letter setting manual of CPU replacement which including service parts)

6 D:¥aplio¥bin¥DrvInstall.bat Start > Windows Explorer


7 C:¥SetAplio¥11_UpdateAplio.bat Start > Windows Explorer
8 C:¥zdt¥Load_FW.bat Start > Windows Explorer
9 C:¥SetAplio¥11_UpdateAplio.bat Start > Windows Explorer
10 Add “Z_DEV_MACHINE” for Environment Variables Start > Settings > Control Panel > System > Advanced
11 Shutdown system and reboot system Also remove all probe from system before reboot
12 C:¥SetAplio¥11_UpdateAplio.bat Start > Windows Explorer
13 C:¥zdt¥rxbf_adc_cal.bat Start > Windows Explorer
14 C:¥zdt¥Scw_Power_Cal.bat Start > Windows Explorer
15 Delete “Z_DEV_MACHINE” from Environment Variables Start > Settings > Control Panel > System > Advanced
16 C:¥SetAplio¥11_UpdateAplio.bat Start > Windows Explorer
17 C:¥SetAplio¥12_RestartAplio.bat Start > Windows Explorer
Shutdown system and reboot system. Connecting probe(s)
18
Then confirm system performance
139
Service Note
SD730-424END*D

Databases

140
Databases Service Note
SD730-424END*D

On this system has 5 databases as below


ArchiveStore
Display of media contents DICOM MO/CD-R/DVD.
Local
This database administrating patient information for patient list.
This database records link to patient images called “Image LOID” also.
SchedulerStore
This database administrating patient information for new patient screen which
listed on bottom of the new patient screen means handling worklist information.
PimsExamArch
This database administrating exporting record such as archive to CD/DVD.
Also administrate OB study related information and History browser
information.
Exchangeboard
Administrate Job status information.
When archive image, printing image data object should be through this DB and
merge data object and DICOM header on this DB.
141
Databases Service Note
SD730-424END*D

Exchangeboard

PimsExamArchive
Local

History

Patient image
Worklist
SchedulerStore

Exchangeboard

ArchiveStore
142
Failure case caused database Service Note
SD730-424END*D

Failed in starting new Exam when press Exam Start.

ArchNet error after sending image

No image display on [Print to...] screen


System automatically shutting down while reboot from “Standby”

143
Databases Service Note
SD730-424END*D

Incase you suggest Database problem...


Cause of the problem in general
Too many job remaining on Job status
DICOM setup mistake (Based on system internal communication)
Delete DICOM setting without deleting remaining jobs for the server

Shutting down the system by force while system processing image object
Please confirm that the system is not processing image object
Make sure DICOM image transfer, archive to
CD/DVD cache, converting

Please confirm that the system is not processing print object


Please confirm Queue on [Print to...] 144
Databases Service Note
SD730-424END*D

Incase you suggest Database problem...


Try
Confirm Job status
If there is so many failed, stopped job delete them
Also confirm why the job status is failed.
If images are reached image server, please confirm DICOM setting
on the system
Clean-up Exchange DB
Patient images are remain even after clean-up.
Recreating Exchange DB
Patient images are remain even after recreating Exchange DB.
Recreate all 5 databases
Required to delete all patient images

145
Databases Service Note
SD730-424END*D

Job status
How to confirm?
Go Patient Browser > Tools > Job status

Basically there is 5 successful job


Delete unnecessary job.
If there is so many failed or Stopped job, we suggest;
If all images are stored on server > DICOM setting is not correct.
If all images are not stored on server > server problem.
Or job which remaining on CD/DVD DICOM cache for waiting burning to media.
Burn DICOM cache to CD/DVD media
146
Databases Service Note
SD730-424END*D

Clean-up Exchange DB
How to do it?
Reboot system
And delete all remaining jobs from job status on patient browser.
Go to service menu
Go to “Command Line” then apply STOPZ.JS.
Wait until windows desktop appeared.
Right click [Start] > [Windows Explorer]
Go to C:¥1-APLI¥BackupRestoreTool.
Then execute ExchangeBoardCleaner.bat
Reboot the system

147
Databases Service Note
SD730-424END*D

Recreating Exchange DB
How to do it?
Reboot system
And delete all remaining jobs from job status on patient browser.
Go to service menu
Go to “Command Line” then apply STOPZ.JS.
Wait until windows desktop appeared.
Click [Start] > [Run] > Enter CMD then click [OK]
A command prompt window appear.
Enter a command below then press [Enter].
ExchBRemove
After display
“Database Entry 5 removed from ID file”
display on command prompt, enter a command
below then press [Enter].
ExchBCreate
Finally enter command below then confirm
DBList
there is
Exchangeboard@”hostname”.
Reboot the system 148
Recreate 5 databases <Similar operation as changing computer name> Service Note
SD730-424END*D

Recreate all 5 databases.


Before doing the operation, we have to make patient image and information
backup otherwise lost image or patient information.
Step Command
1 Go to service menu > Command Line > Execute STOPZ.JS
2 Start > Run > Enter CMD and OK > Execute tus_rmdb.js. Wait 3 minuets. (It is depends on number of images.)
3 Delete all of folder and files under D:¥Database and E:¥Images. (We must keep D:¥Database and E:¥Images folder.)
4 Shutdown. Then reboot the system.
5 Start > Run > Enter CMD and OK > Execute tus_mkdb.js
6 Shutdown. Then reboot the system.
7 Go to service menu > Configuration > Local AEs > Click [Save]
8 Network nodes > Host > Select original computer name then click [Delete]
9 Check “Local Host”. System automatically set new computer name and 192.168.0.2. Then Click [Save].
10 If you have DICOM connection, set new computer name for all of server setting in Network nodes > Remote.
11 Go to [Command line]. Execute tus_ddgw_hostset.bat. * If the bat file is not there, you can skip this step.
12 Go to Advanced DICOM. Confirm new computer name is set for “Basic DICOM Host Name”.
set new computer name for system AE title and DVD drive (Like aplio, F and U) in Advanced DICOM > DICOM
13
Destination.
14 Shutdown. Then reboot the system.

149
Service Note
SD730-424END*D

System installation <Network and DICOM>

150
Network installation Service Note
SD730-424END*D

Setting for network


What is LAC, LAC2 and LAC3
Local Area Connection is for external network connection.
Like DICOM network, measurement result transfer by XML for sheared folder.

Local Area Connection 2 is for internal network connection.


This system database mounted on network.
If system doesn’t have network connection, system cannot recognize the
database correctly.
So that system required this connection, we should NOT set disable this
setting in any case.

Local Area Connection 3 is communicating with DVD-VSR (UIVS-001A)


option. This connection only displayed when you install DVD-VSR.
DVD-VSR unit controlled by TCP/IP.

151
Network installation <Change computer name> Service Note
SD730-424END*D

Here is procedure of changing computer name.


Step Command
1 Go to service menu > Command Line > Execute STOPZ.JS
2 Start > Run > Enter CMD and OK > Execute tus_rmdb.js. Wait 3 minuets. (It is depends on number of images.)
3 Delete all of folder and files under D:¥Database and E:¥Images. (We must keep D:¥Database and E:¥Images folder.)
Change computer name, Right click My computer > Property > Network Identification > Properties.
4
In this procedure system ask you to system restart, it must click [NO].
Start > Run > Enter CMD and OK > Execute tus_chnm.js[]”computer name”
5
*[] means space. “computer name” is defined at step 4.
6 Shutdown. Then reboot the system.
7 Start > Run > Enter CMD and OK > Execute tus_mkdb.js
8 Shutdown. Then reboot the system.
9 Go to service menu > Configuration > Local AEs > Click [Save]
10 Network nodes > Host > Select original computer name then click [Delete]
11 Check “Local Host”. System automatically set new computer name and 192.168.0.2. Then Click [Save].
12 If you have DICOM connection, set new computer name for all of server setting in Network nodes > Remote.
13 Go to [Command line]. Execute tus_ddgw_hostset.bat. * If the bat file is not there, you can skip this step.
14 Go to Advanced DICOM. Confirm new computer name is set for “Basic DICOM Host Name”.
set new computer name for system AE title and DVD drive (Like aplio, F and U) in Advanced DICOM > DICOM
15
Destination.
16 Shutdown. Then reboot the system.

152
DICOM configuration Service Note
SD730-424END*D

Make DICOM configuration with DICOM setup assistant tool.

153
DICOM configuration Service Note
SD730-424END*D

How to use setup assistant 1/6

Click “New” for new configuration


Click “Exit” to close this tool
Click “Load” to loading previous setting
154
DICOM configuration Service Note
SD730-424END*D

How to use setup assistant 2/6

1. Select one of
them depends on
system configuration

2. Click “Next” after select


system configuration

155
DICOM configuration Service Note
SD730-424END*D

How to use setup assistant 3/6 Note: You don’t required change the
IP setting due to system use internal
network address automatically.

1. Set computer name


of ultrasound system
2. Select the system Note: If the system AplioMX or
software version. Xario V6 and later system,
select V2.00, V2.01 (AplioXG)
3. Set ultrasound
system Local AE Title

4. If you required IHE


Application, select
YES. Click “i” icon,
Note: This option can be select if
you can see reference
system has Advanced DICOM option
information of IHE
Application

Note: Click “!” icon,


you can see reference
information regarding
port number for
external devvice.

5. Click “Next” after define


the above items
156
DICOM configuration Service Note
SD730-424END*D

7. Click “Add to list” after define the server information


How to use setup assistant 4/6 As a result AE title of the server listed on Entry list.

1. Set host name of


server. It is you can
define as you like

2. Set the server’s


IP address.

3. Set the server’s


AE Title Note:
“Remove from list”:
4. Set the server
Delete setting
name on the system.
parameter from Entry
This is just the
list.
ultrasound system
internal use, so you
“Update”:
can set as you like.
Click it after modify 8. Click Next when
5. Set the server’s any parameter of you complete defining
Port number selected server. all server parameter.

Note: See next slide also

6. Tick service class(s)


which you want to use
with the server.
Note: Click “Save” for parameter storage service.
This function is not works with Windows7 OS.

157
DICOM configuration Service Note
SD730-424END*D

How to use setup assistant (Detail of service class(s))


Only Basic DICOM Installed

Note:
If a server support several service
classes with same
IP address
IP address
AE Title
AE Title Note: Selectable item of
PortPort
number
number service class is depends
Advanced DICOM Installed on system configuration.
We should tick both of service
classes like this.
(On this case, server supports
Storage and MWM)

158
DICOM configuration Service Note
SD730-424END*D

How to use setup assistant 5/6 (Guide setting procedure 1/2)

3. If complete all items and


click [Save] on the system.
2. Confirm setting. Click “Next” for another
If there are any difference, setting screen.
change it on system.

Note: If there are no change,


but recommend you to click
[Save] anyway on system.
1. Confirm where
should you go!

159
DICOM configuration Service Note
SD730-424END*D

How to use setup assistant 6/6 (Guide setting procedure 2/2)

3. If complete all items and


click [Save] on the system.
Click “Next” for another
2. Confirm setting. setting screen.
If there are any difference,
change it on system.

Note: Confirm this kind


additional explanation also

Note: Check all of items if


there is the Device List
Note: If there are no change, available and there are
but recommend you to click several item listed.
[Save] anyway on system.
1. Confirm where
should you go!

160
DICOM software composition Service Note
SD730-424END*D

We have complicated software composition on this system


Aplio / Xario main system software In case, only USDI-xx0A (Basic DICOM) Installed

DDGW Application
Basic DICOM (Advanced DICOM) MWM Gateway IHE Application Remote activities
Listen port 104 Listen port 2000 Listen port 10500 Listen port 11500

Print SCU Print SCU

Q/R SCU Q/R SCU


Storage SCP Storage SCU Storage SCP Storage SCU
Storage SCU Storage SCP

MWM SCP MWM SCU MWM SCP MWM SCU


MWM SCU MWM SCP

MPPS SCU MPPS SCP

MPPS SCP MPPS SCU

Storage SCP Storage SCU


Storage SCU Storage SCP Storage SCU Storage SCP

Commitment Commitment
SCP SCU
Commitment Commitment
SCU SCP

161
DICOM software composition Service Note
SD730-424END*D

We have complicated software composition on this system


Aplio / Xario main system software In case, only USDI-xx2A (Basic DICOM) Installed

DDGW Application
Basic DICOM (Advanced DICOM) MWM Gateway IHE Application Remote activities
Listen port 104 Listen port 2000 Listen port 10500 Listen port 11500

Print SCU Print SCU

Q/R SCU Q/R SCU


Storage SCP Storage SCU Storage SCP Storage SCU
Storage SCU Storage SCP

MWM SCP MWM SCU MWM SCP MWM SCU


MWM SCU MWM SCP

MPPS SCU MPPS SCP

MPPS SCP MPPS SCU

Storage SCP Storage SCU


Storage SCU Storage SCP Storage SCU Storage SCP

Commitment Commitment
SCP SCU
Commitment Commitment
SCU SCP

162
DICOM software composition Service Note
SD730-424END*D

We have complicated software composition on this system


Aplio / Xario main system software In case, MWM Gateway enabled

DDGW Application
Basic DICOM (Advanced DICOM) MWM Gateway IHE Application Remote activities
Listen port 104 Listen port 2000 Listen port 10500 Listen port 11500

Print SCU Print SCU

Q/R SCU Q/R SCU


Storage SCP Storage SCU Storage SCP Storage SCU
Storage SCU Storage SCP

MWM SCP MWM SCU


MWM SCU MWM SCP MWM SCU MWM SCP

MPPS SCU MPPS SCP

MPPS SCP MPPS SCU

Storage SCP Storage SCU


Storage SCU Storage SCP Storage SCU Storage SCP

Commitment Commitment
SCP SCU
Commitment Commitment
SCU SCP

163
DICOM software composition Service Note
SD730-424END*D

We have complicated software composition on this system


Aplio / Xario main system software In case, IHE application enabled. (Of course MWM enabled)

DDGW Application
Basic DICOM (Advanced DICOM) MWM Gateway IHE Application Remote activities
Listen port 104 Listen port 2000 Listen port 10500 Listen port 11500

Print SCU Print SCU

Q/R SCU Q/R SCU


Storage SCP Storage SCU Storage SCP Storage SCU Storage SCP Storage SCU

MWM SCU MWM SCP MWM SCU MWM SCP MWM SCU MWM SCP

MPPS SCU MPPS SCP MPPS SCU MPPS SCP

Storage SCU Storage SCP Storage SCU Storage SCP Storage SCU Storage SCP

Commitment Commitment Commitment Commitment


SCU SCP SCU SCP

164
DICOM configuration (Advanced) Service Note
SD730-424END*D

General
Additional Script
US => Retired US
MONOCHROME2 Conversion
MWM
MWM Gateway
MWM Query – Advanced
Modality
Exam type selection (Comprehensive)
IHE Application
Support IHE
Multiframe Decompression
DICOM Format Conversion for Toshiba Private data
MWM Option
MWM Query - IHE
DICOM SR
What is DICOM SR object
DICOM SR Setting

165
DICOM configuration (Advanced) Service Note
SD730-424END*D

Function overview (Supported version)


SSA-790A SSA-780A SSA-680A SSA-660A
(AplioXG) (AplioMX) (XarioXG) (Xario)
Additional Script V2.0 V1.0 V1.0 V6.0
General Retired US V1.1 V1.0 V1.0 V1.0
Monochrome2 Conversion V1.1 V1.0 V1.0 V3.0
GW V1.1 V1.0 V1.0 V5.0
MWM Query (Advanced) V1.1 V1.0 V1.0 V5.0
MWM
Modality V1.1 V1.0 V1.0 V5.0
Exam Type Selection (Comprehensive) V1.1 V1.0 V1.0 V5.0
IHE support V1.1 V1.0 V1.0 V6.0
Multiframe Decompression V2.0 V1.0 V1.0 V6.0
IHE
Private data Convert V2.0 V1.0 V1.0 V6.0
MWM Query (IHE) V2.0 V1.0 V1.0 V6.0
Private V1.1 V1.0 V1.0 V1.0
Basic text V1.1 V1.0 V1.0 V1.0
SR Enhanced V1.1 V1.0 V1.0 V6.0
Comprehensive V5.0 V2.0 V4.0 V9.0
DICOM Code set V4.0 V1.0 V3.0 V8.0

Note: If there is any color on the cell, means required specific software version.

166
DICOM configuration (Advanced) Service Note
SD730-424END*D

Function overview (Functionality 1/3)


Description Setting items
Additional Script On and later specific software version on previous page, the system usually set Advanced DICOM –
“USTras.scr” for remote servers. Archives
If the setting if “USTrans.scr” system won’t send Pixel Spacing on stored exam
images. If customer wish to receive the data, set “Blank.scr” for this setting
General
Retired US If storage server only accept Retired US DICOM SOP class, set enable for this Advanced DICOM –
setting (Or customer requested so... ) Archives
Monochrome2 If storage server only accept Monochrome2 (8bit B/W) DICOM image, set enable Advanced DICOM –
Conversion for this setting (Or customer requested so... ) Archives
GW If customer would like to use advanced function for querying worklist (Listed Configuration – DICOM
below on category of MWM GW), or if system required to use IHE function should - HIS/RIS Node –
set to “ON” for this setting HIS/RIS
MWM Query If customer wish to set detail query condition when try to get worklist, use this 1. Required MWM GW.
(Advanced) option. 2. New Patient – Option
MWM Modality If customer wish to set blank for Modality query condition use this option 1. Required MWM GW.
(Or IHE)
2.New Patient – Option
Exam Type Selection If customer wish to automatically select exam type depends on worklist 1. New Patient – Option
(Comprehensive) information, use this option and define Work order setting. 2. Menu – System
Setting - Work Order

Note: Detail of settings, please refer each section of the function.

167
DICOM configuration (Advanced) Service Note
SD730-424END*D

Function overview (Functionality 2/3)


Description Setting items
IHE support If customer wish to use work flow based on IHE. 1. Enable MWM Gateway
Set enable for this function. (Configuration – DICOM - HIS/RIS
Also if customer wish to use function below. Node – HIS/RIS)
2. Set “On” for IHE Application
(Configuration – DICOM - Local
AEs)

Multiframe Cine (multi frame) image basically JPEG lossy (JPEG baseline) data. 1. Enable IHE
Decompression Sometime server won’t accept the type of data, in this time we can 2. Advanced DICOM – Archives
decompressed cine image when sending cine image to server,
IHE system internally stored by JPEG lossy.
Private data Convert System has TOSHIBA ultrasound specific data such as 3D, TDI-Q / 1. Enable IHE
CHI-Q etc. 2. Configuration – DICOM -
Basically sever doesn’t accept them due to TOSHIBA private Network nodes – Remote
format.
But we can insert dummy (Ultrasound image) data on the private
data, as a result server may accept the image. Unfortunately the
private data cannot review on the server. Only dummy data can be
review on the server.
MWM Query (IHE) If customer wish to add, edit worklist. It should set enable. 1. Enable IHE
2. New Patient – Option

Note: Detail of settings, please refer each section of the function.

168
DICOM configuration (Advanced) Service Note
SD730-424END*D

Function overview (Functionality 3/3)


Description Setting items
Private Send measurement result by Toshiba Private 1. Configuration – DICOM - Network nodes –
formatted SR. (Actually XML) Remote
2. Menu – System Setting - Report
Basic text Send measurement result by Basic Text SR object. 1. Configuration – DICOM - Network nodes –
(Actually XML format data) Remote
2. Menu – System Setting - Report
Enhanced Send measurement result by Enhanced SR object. 1. Enable IHE
Required IHE Application. 2. Configuration – DICOM - Network nodes –
SR (Based on template which defined on DICOM) Remote
3. Menu – System Setting - Report
Comprehensive Send measurement result by Comprehensive SR object. 1. Enable IHE
Required IHE Application. 2. Configuration – DICOM - Network nodes –
(Based on template which defined on DICOM) Remote
3. Menu – System Setting - Report
DICOM Code set We can adjust format of the SR object (Enhanced and 1. Set SR setting
Comprehensive) by this function 2. Menu – Exam Type – [Shift ] and [<]

Note: Detail of settings, please refer each section of the function.

169
Service Note
SD730-424END*D

General
1. Additional Script
2. Retired US
3. Monochrome 2 Conversion

170
1. Additional Script Service Note
SD730-424END*D

What is the function?


If customer requests that they need a tag information.
The TAG is (0028,0030) Pixel Spacing.
Old Aplio/Xario system DICOM image including (0028,0030) Pixel Spacing.
On and later Vx.xx system, the tag is TAG including by factory default.
Due to the TAG is not state in DICOM standard for Ultrasound image.
What is the (0028,0030) Pixel Spacing??
sometimes use for referring how much distance between pixel by pixel on DICOM
viewer when doing re-measurement.
For more detail, also see next slide.

When we need this function?


Depends on customer
Trouble shooting.
Incase if customer has very old SSA-770A system and they doing
re-measurement on DICOM viewer referring for Pixel Spacing.
But the customer couldn’t do re-measurement on latest version AplioXG
DICOM image.

171
1. Additional Script Service Note
SD730-424END*D

What is the Pixel spacing???


(0028,0030) Pixel Spacing sometimes use for referring how much distance
between pixel by pixel on DICOM viewer when doing re-measurement.
But the TAG only define one distance information, so that sometime we
could NOT get correct result of re-measurement.
As you can imagine, Ultrasound system may have different distance of image in
a DICOM image.
For instance
In dual mode.
Depth of image left side is 10 cm and right side is 15 cm.
Pixel spacing could NOT present correct value for both image...
In 2D + PW mode
2D and PW mode should have different information such as distance, time and
velocity, pixel spacing couldn’t handle them!
Based on the above problems...
Ultrasound image re-measurement should refer to
“US Region Calibration Module”.

172
1. Additional Script - Setting Service Note
SD730-424END*D

Define additional script in Archives – Advanced DICOM

1. Select the server which you


wan to send Pixel Spacing tag.

2. Then select “Blank.sct”


for “Additional Script”.
Default is “USTrans.scr”.
173
2. Retired US Service Note
SD730-424END*D

What is the function


System stored still image converting into Retired Ultrasound Image object
when sending still image to storage server, if the server only accepts
Retired US image.(Ultrasound Image Storage (retired) SOP Class UID:
1.2.840.10008.5.1.4.1.1.6 )
When we need this function?
Server specification
Verify on server DICOM conformance statement, if the server only support
1.2.840.10008.5.1.4.1.1.6 , we should use this option. Usually our system still
image SOP Class UID is 1.2.840.10008.5.1.4.1.1.6.1.

174
2. Retired US - Setting Service Note
SD730-424END*D

Enable conversion function in Archives – Advanced DICOM

1. Select the server which


you want to send still image
by monochrome.

2. Put check mark here.

175
3. Monochrome2 Conversion Service Note
SD730-424END*D

What is the function


System stored still image converting into 8 bit monochrome (B/W) image
when sending still image to storage server.
(Usually system sending 24 bit RGB image to server.)
When we need this function?
Customer request
If customer would like review a image on image viewer by 8 bit monochrome
image.
If server only accept 8 bit monochrome image.
Notice
One of the limitation is that we cannot retrieve monochrome converted
image from server if server support Query and Retrieve function.
This is limitation of this function.
This function only effect still sending image, no effects for stored image on
system HDD and cine image.

176
3. Monochrome2 Conversion Service Note
SD730-424END*D

Enable conversion function in Archives – Advanced DICOM

1. Select the server which


you want to send still image
by monochrome.

2. Put check mark here.

177
Service Note
SD730-424END*D

MWM
1. MWM Gateway
1-1. MWM Query (Advanced)
1-2. Modality
2. Exam Type Selection (Comprehensive)

178
1. MWM Gateway Service Note
SD730-424END*D

What is the function


This function add several functions for MWM operation.
Also MWM Gateway is foundation of IHE application.
So that if we want to use IHE function MWM must turn ON.
When we need this function?
Customer request
If customer would like to define query items to getting worklist from MWM
server. Basically Aplio/Xario system state Modality as US and Study date as
today. But sometime customer request that they would like to get tomorrow or
yesterday patient etc.
Enable MWM Gateway then select Advanced Query at [Option] in New Patient
screen.
We can define detail query items such as Patient ID, Patient name and range of
the examination date when getting patient information from MWM server with
Advanced Query setting.
This function also required to enable IHE application.
Detail of IHE Application will be explain.
179
1. MWM Gateway - Setting Service Note
SD730-424END*D

Enable MWM Gateway in HIS/RIS – HIS/RIS Node - DICOM

MWM Gateway set to “ON”


We don’t needs MWM setting, but set “ON” for MWM Gateway

180
1. MWM Gateway – 1-1. MWM Query (Advanced) Service Note
SD730-424END*D

MWM Option
New Patient > [Option]

This item will be selectable


if MWM Gateway set to ON

181
1. MWM Gateway – 1-1. MWM Query (Advanced) Service Note
SD730-424END*D

New Patient > Click [Refresh] in new patient.


(or just go to New Patient, depends on Option setting)

We can define following query condition for MWM server:


(0010,0020) Patient ID
(0010,0010) Patient Name
(0008,0050) Accession No.
(0040,0002) Scheduled Procedure Step Start Date
(0040,1001) Requested Procedure ID
(0040,0001) Scheduled Station AE Title 182
1. MWM Gateway – 1-2. Modality Service Note
SD730-424END*D

MWM Option
New Patient > [Option]

We can define “US” or blank on query condition This


This item
item will
will be
be selectable
selectable
The value set below when system request work list if
if MWM
MWM Query
Query setting
setting defined
defined
(0008,0060) Modality “Advanced” or “IHE”.
“Advanced” or “IHE”.

“US” means request for Ultrasound system exam list


Blank means request all of exam list

183
MWM (Meaning of Query condition) Service Note
SD730-424END*D

DICOM communication overview in MWM (Example)


Server database
Associate negotiation request
May I ask you? Patient ID Patient Name Study date
1 A^A 20110330
(Today)
Associate negotiation acknowledge
2 B^B 20110331
YES! (Tomorrow)
3 C^C 20110329
State query condition
4 D^D 20110328

I need...
Study starting: Today and tomorrow

Reply to system (Patient name, ID etc)


Result
I have several patient information.
Patient ID Patient Name Study date (Depends on how many patients matched
based on query condition)
1 A^A 20110330
(Today)
2 B^B 20110331 Reply to system (Patient name, ID etc)
(Tomorrow)

Association release request


Thanks a lot!!
Association release acknowledge
See you!!
184
2. Exam Type Selection (Comprehensive) 1/4 Service Note
SD730-424END*D

What is the function


System automatically select exam type based on translation table between
customer’s exam type description and system’s internal exam type which
called Work Order on our system.

When we need this function?


Customer request
If customer defined typical exam description and customer wish to
automatically set system’s exam type based on the worklist information

185
2. Exam Type Selection (Comprehensive) 2/4 Service Note
SD730-424END*D

This is a example of if the system set to condition as below.


Item Condition
Procedure DICOM / MWM
Exam Type Linkage Scheduled Procedure Step Description
Work Order setting
Work Order Description Exam Type
(Customer’s description) (System interenal)
A-Heart Adult Heart
P-Heart Pediatric Heart
Carotid-R Carotid
Carotid-L Carotid

System (Result of MWM) Server database

Scheduled Exam Type Scheduled


Patient Patient Study Patient Patient Study
Procedure Step (Automatically Procedure Step
ID Name date ID Name date
Description select) Description
20110330 20110330
1 A^A A-Heart Adult Heart 1 A^A A-Heart
(Today) (Today)
20110330 20110330
2 B^B Carotid-R Carotid 2 B^B Carotid-R
(Today) MWM (Today)
20110330 20110330
3 C^C Liver Set blank 3 C^C Liver
(Today) (Today)
20110330 20110330
4 D^D Kidney Set blank 4 D^D Kidney
(Today) (Today)

186
2. Exam Type Selection (Comprehensive) 3/4 Service Note
SD730-424END*D

MWM Option
New Patient > [Option]

Select this one

187
2. Exam Type Selection (Comprehensive) 4/4 Service Note
SD730-424END*D

Choose one of the tag which is key information for


Work order (System Setting > Work Order > Press [Shift] and [<]) exam type automatic selection.
(0008,1030) Study description
(0040,0007) Scheduled Procedure Step Description
(0032,1060) Requested Procedure Description

Select system’s exam type.

Define customer’s exam description.


188
Service Note
SD730-424END*D

IHE Application
1. Support IHE
2. Multiframe Decompression
3. DICOM Format Conversion for Toshiba Private data
4. MWM Option (MWM Query - IHE)

189
IHE Application Service Note
SD730-424END*D

What is the function?


If this function enabled, many additional function will be available.
Support IHE
DICOM format conversion for Toshiba Private data.
Multiframe Decompression
Additional MWM query condition, called “IHE”

When we need this function?


Usually based on customer request if customer has IHE applicable facility.
Multiframe decompression function sometime use for DICOM trouble
shooting.

190
1. Support IHE Service Note
SD730-424END*D

What is the function?


IHE is not standard, kind of state general workflow in hospital.
In the workflow define requirement information based on DICOM and HL7.

And modality vendor state that which workflow (functionality) supported on


the system by “IHE Integration Statement” based on IHE connection test
result (Called Connectathon).
You can down load the statement from;
http://www.toshiba-medical.co.jp/tmd/english/ihe/index.html

As a result if hospital has IHE applicable system and its state same
workflow with our system. Basically DICOM connection successfully
without any problems.

For more detail of IHE visit;


http://www.ihe.net/ (For North America)
http://www.ihe-europe.net/ (For Europe)
191
1. Support IHE – Setting (1/2) Service Note
SD730-424END*D

Enable MWM Gateway in HIS/RIS – HIS/RIS Node - DICOM

MWM Gateway set to “ON”


We don’t needs MWM setting, but set “ON” for MWM Gateway

192
1. Support IHE – Setting (2/2 new version) Service Note
SD730-424END*D

Enable IHE Application setting (New setup screen) in Local AEs

IHE Application set to “ON”


This function will be selectable after MWM Gateway set to “ON”.
Mobile Mode and TCE
This function is one of the key function of IHE workflow.
This screen will be shown on following system and version.
SSA-790A: V5.00 and later
SSA-780A: V2.00 and later
SSA-680A: V4.00 and later
SSA-660A: V9,00 and later
193
1. Support IHE – Setting (2/2 old version) Service Note
SD730-424END*D

Enable IHE Application setting (New setup screen) in Local AEs

IHE Application set to “ON”


This function will be selectable after MWM Gateway set to “ON”.
This screen will be shown on following system and version.
SSA-790A: V2.00-V4.00
SSA-780A: V1.00-V1.20
SSA-680A: V1.00-V3.20
SSA-660A: V6.00-V8.20

194
2. Multiframe Decompression Service Note
SD730-424END*D

What is the function?


If this function enabled, system converting multiframe (Clip) image into
uncompressed (RGB: 24bit color,8bit for each color) format when sending
multiframe image to storage (PACS) server.
When we need this function?
Depends on customer (PACS server engineer request)
Trouble shooting.
Incase if you can send still image to server with “Success” status,
but we cannot send multiframe (Clip) image to server.
Always system show “Network job failed” when transfer clip image.

195
2. Multi-frame decompression setting Service Note
SD730-424END*D

The multiframe decompression setting requireds,


USDI-792A or USDI-772A/XA (Mean, needs Advanced DICOM)
Install Basic DICOM and Advanced DICOM
MWM Gateway
Service menu > Configuration > HIS/RIS nodes > HIS/RIS
IHE Application
Service menu > Configuration > Local AEs
Multiframe Decompression
Service menu > Configuration > Advanced DICOM > Archives
> Select server which you want to send uncompressed cine image
Limitation
This function support 360 frames multiframe cine image.
If Data exceed 360 frames (60fps x 6sec), it is not decompressed.
(Error is displayed in JOB STATUS)

196
2. Multiframe decompression - Frame work Service Note
SD730-424END*D

In case, IHE application enabled. (Of course MWM enabled)


Aplio / Xario main system software

DDGW Application
Basic DICOM (Advanced DICOM) MWM Gateway IHE Application Remote activities
Listen port 104 Listen port 2000 Listen port 10500 Listen port 11500

Print SCU Print SCU

Q/R SCU Q/R SCU


Storage SCP Storage SCU Storage SCP Storage SCU Storage SCP Storage SCU

MWM SCU MWM SCP MWM SCU MWM SCP MWM SCU MWM SCP

MPPS SCU MPPS SCP MPPS SCU MPPS SCP

Storage SCU Storage SCP Storage SCU Storage SCP Storage SCU Storage SCP

*1 Multiframe
decompression

Commitment Commitment Commitment Commitment


SCU SCP SCU SCP

*1: If Multiframe decompression activated in Local AEs and Archives in Advanced DICOM

197
2. Multiframe decompression – Setting (Required enable IHE as well) Service Note
SD730-424END*D

Enable Multiframe Decompression in Archives – Advanced DICOM

1. Select the server which


you wan to send cine image
by uncompressed format.

2. Set check mark for


“Multiframe Decompression”.

198
2. Multiframe decompression – Setting (Required enable IHE as well) Service Note
SD730-424END*D

Enable IHE Application setting (New setup screen) in Local AEs

IHE Application set to “ON”


This function will be selectable after MWM Gateway set to “ON”.
Cine Image Decompression
If the system has this option, should be put check mark.
This screen will be shown on following system and version.
SSA-790A: V5.00 and later
SSA-780A: V2.00 and later
SSA-680A: V4.00 and later
SSA-660A: V9,00 and later
199
3. DICOM Format Conversion Service Note
SD730-424END*D

What is the function?


System has TOSHIBA ultrasound specific data such as 4D, 3D, TDI-Q and
CHI-Q etc. Basically sever doesn’t accept them due to TOSHIBA private
format.
If enabled the function system insert a dummy (Based on Ultrasound image
object) data on the private data, as a result server may accept the image.
Unfortunately the private data cannot review on the server. Only dummy
data can be review on the server.

When we need this function?


Depends on customer
Customer would like to store image on the server
If the system connecting with Ultra Extend
(USWS-900A and USWP-100A option)
Incase we should use the private format conversion for 4D data to reviewing
on the Ultra Extend system.

200
3. DICOM Format Conversion – Setting (Required enable IHE as well) Service Note
SD730-424END*D

Enable Format Conversion in Remote – Network node


Select a server from drop down list of “Logical Name” then scroll down.
1. Select a server which you
want to send converted
private image from Logical
name

3. Define which private data


required to convert.

2. Scroll down the window.


You can find the setting.

Setting parameters Parameter Functionality


System won’t convert the private data when
Don’t Convert
sending the private data to the server.
System insert dummy US still image between
header and private data while sending the
Convert private image to server. As a result the private
(or Translate) image handling as Ultrasound still image.
Converted image could NOT retrieving even
using DICOM Q/R service.
System will be skip to send the private image
Don’t send
object to the server

201
3. DICOM Format Conversion – For instance Service Note
SD730-424END*D

We have several thumbnails for 4D volume data in image list of Ultra Extend (USWS-900A).
What is different between them??

Type1 Type2
This thumbnail will shown. This thumbnail will shown.
The image stored: The image stored:
AplioXG V3.0 and before. AplioXG V3.0 and before.
Image received via: Image received via:
DICOM Storage service. CD / DVD
(Media Storage)

Type3
This thumbnail will shown.
The image stored:
AplioXG V4.0 and later
Image received via:
DICOM Storage service.

202
3. DICOM Format Conversion – For instance Service Note
SD730-424END*D

What kind of the function about “Convert” (or “Translate”) as


DICOM Format setting. (Network nodes > Remote)
We have several types of private formatted data such as Fusion 3D,
Stress Echo, RAW(TDI-Q/CHI-Q) and 4D on AplioXG, XarioXG system. Private
Usually those kind data couldn’t receive general DICOM storage SCP
due to modality (AplioXG) sending with private SOP class.
Header
So we should use “DICOM Format” function.
If we set to “Convert” (or “Translate” depends on system software Private Data
version), system will append dummy pixel data between DICOM header
and private data then send it Ultrasound image storage SOP class. As a Translate
result general DICOM server could be receive Toshiba private data. (If
US Image
you open the data, viewer will show dummy pixel data, we cannot see
private data itself without ACP, 4DV anyway.)
Append dummy pixel data.
Header
4D data stored by
Note!: You cannot retrieve 4D data stored by
V4.0 and later AplioXG Dummy pixel Data
V3.0 and before AplioXG
Translated private data
correctly to Ultrasound
system from server. Private Data
203
3. DICOM Format Conversion – For instance Service Note
SD730-424END*D

Normal PACS
or
Transfer via DICOM

US Image Ultra Extend receive as


Ultrasound image SOP class data.
Append dummy
pixel data.
Header US Image
4D data stored by
V3.0 and before AplioXG Dummy pixel Data Header

Private Data Dummy pixel Data

4D data stored by
V4.0 and later AplioXG Private Data

204
3. DICOM Format Conversion – For instance Service Note
SD730-424END*D

Burning

Only Private data Ultra Extend create a thumbnail.

Header

Private Data

Only Private data

Header

Private Data

205
4. MWM Option (MWM Query - IHE) Service Note
SD730-424END*D

What is the function?


If this function enabled, operator enable editing worklist data such as add
series (study based on same study instance UID) or study as operator like.
For instance;
Provides worklist A. And this study is for cardiac.
But this patient required carotid study also, in this time operator add series
based on same Study instance UID.

When we need this function?


Depends on customer request
If customer frequently has above case studies.

206
4. MWM Option (MWM Query - IHE) - Setting (Required enable IHE as well) Service Note
SD730-424END*D

MWM Option
New Patient > [Option]

This item will be selectable


if IHE Application set to ON.
(MWM G/W set to ON as well.)

207
Service Note
SD730-424END*D

DICOM SR
1. What is DICOM SR object
2. DICOM SR setting
2-1 Job routing
2-2 DICOM setting
2-3 Define report server
2-4 Transfer Method
2-5 DICOM Code set

208
1. What is DICOM SR object? Service Note
SD730-424END*D

The DICOM SR object is an object which transferred by DICOM


Storage service.
So there is not special DICOM setting even SR transfer.
But server side must support following SOP by Storage SCP.
Presentation context
Abstract syntax Transfer syntax AXG AMX XXG Xario
Name UID Name UID
Implicit VR Little Endian 1.2.840.10008.1.2
Enhanced SR 1.2.840.10008.5.1.4.1.1.88.22 V1.1- V1.0- V1.0- V6.0-
Explicit VR Little Endian 1.2.840.10008.1.2.1
Implicit VR Little Endian 1.2.840.10008.1.2
Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 V5.0- V2.0- V4.0- V9.0-
Explicit VR Little Endian 1.2.840.10008.1.2.1
Implicit VR Little Endian 1.2.840.10008.1.2
Basic Text SR 1.2.840.10008.5.1.4.1.1.88.11 V1.1- V1.0- V1.0- V1.0-
Explicit VR Little Endian 1.2.840.10008.1.2.1
Private SR Implicit VR Little Endian 1.2.840.10008.1.2
(Toshiba US Private 1.2.392.200036.9116.7.8.1.1.1 V1.1- V1.0- V1.0- V1.0-
Data Storage ) Explicit VR Little Endian 1.2.840.10008.1.2.1

209
1. What is DICOM SR object? Service Note
SD730-424END*D

DICOM SR object composition.

DICOM SR object

DICOM header info Like patient name, ID, Study instance UID etc...

Like measurement method, value and unit.


SR information. Also image mode (2D, M, doppler etc...), view (4ch, 2ch etc) cardiac cycle point
(Measurement result) information (Diastole / Systole).

210
1. What is DICOM SR object? Service Note
SD730-424END*D

DICOM SR using templates which defined on DICOM part 16.


But the specification does NOT concrete explanation due to
future upgrade or some reason. (This is DICOM!!)
This means, even DICOM SR object we called but there are may
some differences between we and server. NOT same for other
manufacturer etc...

PLEASE CAREFULLY!!
We can send SR object to report server but sometime report
viewer may not display the result or some of the measurement
item may corrupt on the report viewer due to above reason.

This is NOT problem, just each system technical specification


differences.

211
2. DICOM SR Setting Service Note
SD730-424END*D

General job routing for DICOM SR connection


Define DICOM setting
Storage (report) server setup
Activate MWM Gateway
Activate IHE Application
Define type of DICOM SR object
Define Report server
Define measurement data transfer method
Define DICOM code set

212
2. DICOM SR Setting Service Note
SD730-424END*D

General job routing for DICOM SR connection


DICOM SR connection is NOT like usual DICOM installation.
We need some discussion with customer and server engineer.
Next page describes idea of general work flow.

213
2. DICOM SR Setting Service Note
SD730-424END*D

General job routing for DICOM SR connection 1


SR installation request
Confirm test data

XML Decide report DICOM SR


transfer method
Discuss with server engineer

Create shearing holder Choose type of SR object


Account setup Enhanced SR Server or
Confirm role etc... or DICOM code set
Comprehensive SR configuration

OK

Connection check (2nd Level)

Server supported
OK (Based on CS) NG NG
Confirm on
NG report viewer
Required Server Connection check Customer wants to correct
supports SR object (1st Level) measurement item
and / or
OK Customer want correct
report format
measurement procedure
OK Confirm by customer
Decide measurement items with customer
(Including method and procedure) OK Customer want correct
report format
Running test
Do measurement for selected items Customer wants to correct
(Create test data, measurement item
also collect report screen) and / or OK
measurement procedure
Installation complete

1 214
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM storage server setting


DICOM SR is an object on Storage service class.
Please put check mark on “Structured Report” in Remote Device Information
screen of DICOM setup assistant tool as well as Storage service.

DICOM SR connection is NOT like usual DICOM network installation.


We need following information from the next page to SR connection.

215
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM storage server setting


Incase if you wish to use Comprehensive SR or Enhanced SR, we have to
enable IHE Application.
Set “ON” for MWM Gateway in HIS/RIS – HIS/RIS nodes.
Set “ON” for IHE Application in Local AEs.
Recommend you to check “YES” on DICOM setup assistant.

Also put check mark for “SR conversion” incase the system as below.
SSA-790A: V5.00 and later
SSA-780A: V2.00 and later
SSA-680A: V4.00 and later
SSA-660A: V9,00 and later

Choose type of SR object in Remote – Network nodes.

216
2. DICOM SR Setting Service Note
SD730-424END*D

Enable MWM Gateway in HIS/RIS – HIS/RIS Node - DICOM

MWM Gateway set to “ON”


We don’t needs MWM setting, but set “ON” for MWM Gateway

217
2. DICOM SR Setting Service Note
SD730-424END*D

Enable IHE Application setting (New setup screen) in Local AEs

IHE Application set to “ON”


This function will be selectable after MWM Gateway set to “ON”.
SR Conversion set to check mark
This function will be selectable after IHE Application set to “ON”.
This screen will be shown on following system and version.
SSA-790A: V5.00 and later
SSA-780A: V2.00 and later
SSA-680A: V4.00 and later
SSA-660A: V9,00 and later
218
2. DICOM SR Setting Service Note
SD730-424END*D

Enable IHE Application setting (New setup screen) in Local AEs

IHE Application set to “ON”


This function will be selectable after MWM Gateway set to “ON”.
This screen will be shown on following system and version.
SSA-790A: V2.00-V4.00
SSA-780A: V1.00-V1.20
SSA-680A: V1.00-V3.20
SSA-660A: V6.00-V8.20

219
2. DICOM SR Setting Service Note
SD730-424END*D

Choose type of DICOM SR object in Remote – Network node


Select a server from drop down list of “Logical Name” then scroll down.
Setting parameters Parameter Functionality
Comprehensive System send measurement result using DICOM template.
Enhanced System send measurement result using DICOM template.
Server may accept, if server accept Basic text SR.
Basic Text But measurement result states by Toshiba format.
So server may not understand the data.
Toshiba private formatted SR, usually server is not
Toshiba Private
support.

Note: Comprehensive SR and


Enhanced SR required IHE
application. And confirm
supported software version.

2. Scroll down the window.


You can find the setting.

1. Select a server which you


3. Define which private data
want to send converted
required to convert.
private image from Logical
name

220
2. DICOM SR Setting Service Note
SD730-424END*D

Define Report server.


Press [Menu] > System Setting > Output > Press [Shift] and [<]
> Select the Report server

221
2. DICOM SR Setting Service Note
SD730-424END*D

Define measurement data transfer method


Press [Menu] > System Setting > Report > Press [Shift] and [<]
> Select the Report server

Show [Send as DICOM]


switch in Report screen

Send DICOM SR object to report server which


defined in “Output” – System setting when click
[Send as DICOM] switch in Report screen.
System store DICOM SR object on HDD then send
it to Report server.
If there is no selection, system only store DICOM
SR object to data HDD.

222
Couldn’t see the Storage and SR on the same study Service Note
SD730-424END*D

Problem : You can’t see the same study of Storage and SR by the Server.
Cause : The Series Instance UID of the SR and Storage was same on the system.
You can change it by you change the following file.
Countermeasure:
1) Go to Maintenance menu > [Command Line]
2) Please open the following folder.
“D;¥aplio¥config¥TUSIHEApp¥IHEApplicationConfig.ini”
3) Please change the FALSE of "SR_UNIQUE_ID" to TRUE.

[SR_UNIQUE_ID]
SR_UNIQUE_SERIES_INSTANCE_UID=FALSE

SR_UNIQUE_SERIES_INSTANCE_UID=TRUE

223
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM code set (Version depends function)


What is DICOM code set / Mapping file??
Like SR structure converter.

Storage (Report) Server


DICOM Code Set
Original data Mapping file DICOM SR object

DICOM header info

Artida internal SR information.


report format (Measurement result)

DICOM Code Set converts


internal SR structure to Converted SR data transfer by DICOM.
DICOM SR object

224
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM code set – Mapping file selection


Press [Menu] > [System setting] > [Exam Type] > Press [Shift] and [<]
> Click [DICOM Code Set]

This function will be shown on following system and version.


SSA-790A: V5.00 and later
SSA-780A: V2.00 and later
SSA-680A: V4.00 and later
SSA-660A: V9,00 and later

225
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM code set – Mapping file selection


Press [Menu] > [System setting] > [Exam Type] > Press [Shift] and [<]
> Click [DICOM Code Set]

226
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM code set – Mapping file selection


Press [Menu] > [System setting] > [Exam Type] > Press [Shift] and [<]
> Click [DICOM Code Set]
Echo tab
Mapping File Remarks
ECHO_Mapping.xml Default parser
ECHO_Mapping.xmlAplioXGV4-AplioMXV1-V1.2 AXG V4.00/V4.20, AMX V1.00/V1.20, XXG V3.00/V3.20, Xario
V8.00/V8.20 compatible
ECHO_Mapping.xmlAplioXGV5-AplioMXV2 AXG V5.00, AMX V2.00, XXG V4.00, Xario V9.00 compatible

Vascular tab
Mapping File Remarks
VASCULAR_Mapping.xml Default parser
VASCULAR_Mapping.xmlAplioXGV4-AplioMXV1-V1.2 AXG V4.00/V4.20, AMX V1.00/V1.20, XXG V3.00/V3.20, Xario
V8.00/V8.20 compatible
VASCULAR_Mapping.xmlAplioXGV5-AplioMXV2 AXG V5.00, AMX V2.00, XXG V4.00, Xario V9.00 compatible

227
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM code set – Mapping file selection


Press [Menu] > [System setting] > [Exam Type] > Press [Shift] and [<]
> Click [DICOM Code Set]
Radiology tab
Mapping File Remarks
RADIOLOGY_Mapping.xml Default parser
RADIOLOGY_Mapping.xmlAplioXGV4-AplioMXV1-V1.2 AXG V4.00/V4.20, AMX V1.00/V1.20, XXG V3.00/V3.20, Xario
V8.00/V8.20 compatible
RADIOLOGY_Mapping.xmlAplioXGV5-AplioMXV2 AXG V5.00, AMX V2.00, XXG V4.00, Xario V9.00 compatible

OB tab
Mapping File Remarks
OB_Mapping.xml Default parser
OB_Mapping.xmlAplioXGV4-AplioMXV1-V1.2 AXG V4.00/V4.20, AMX V1.00/V1.20, XXG V3.00/V3.20, Xario
V8.00/V8.20 compatible
OB_Mapping.xmlAplioXGV5-AplioMXV2 AXG V5.00, AMX V2.00, XXG V4.00, Xario V9.00 compatible

228
2. DICOM SR Setting Service Note
SD730-424END*D

Why we need change mapping file for SR??


DICOM SR mapping file is define conversion structure between Artida native SR
data and DICOM SR object.
Conversion structure slightly different depends on software version.
Due to some of connectivity problem, data compatibility etc.
If SR connection works fine on certain version of system, we may have some
problem if we upgrade the system due to DICOM SR object structure is NOT same
based on above reason.
In this case, we should change the mapping file as same as original version.
Report server may NOT handle the difference.
SR
Server side needs to reconfigure the SR
DICOM Code Set
structure etc...
Mapping file (V4)
A B C

Original data C-1 C-2

C-2-1 SR
DICOM Code Set V4 SR Object
Mapping file (Default)
Artida internal A B C D
report format

D-1 D-2

DICOM Code Set converts


internal SR structure to D-2-1
Default SR Object
DICOM SR object
229
2. DICOM SR Setting Service Note
SD730-424END*D

Define DICOM code set – Modify DICOM code set


Press [Menu] > [System setting] > [Exam Type] > Press [Shift] and [<]
> Click [DICOM Code Set] > Select a measurement item which you need to edit
> Edit the value of the code.
This operation should be extra care, otherwise we are going to more difficult problems!!

230
2. DICOM SR Setting Service Note
SD730-424END*D

Why we need modify DICOM code set??


In case server accepts specific DICOM SR structure which is not supported any
DICOM mapping file in this system. In this case we may try to modify the DICOM
mapping file by our self
This operation should be extra care, otherwise we are going to more difficult problems!!
Server side requirements Also even try this function some case we cannot solve the problem.
SR

A B C D

D-a D-b

D-b-a’
SR

DICOM Code Set


A B C D
Mapping file (Modified) Report server may NOT handle the difference.

Original data D-a D-b

D-b-a’ SR
Modified structure
Artida internal A B C D
report format

D-1 D-2
DICOM Code Set converts DICOM Code Set
internal SR structure to Mapping file (Default) D-2-1
DICOM SR object Default SR Object
231
Service Note
SD730-424END*D

Basic Operation

232
Basic operation <Getting Start and shutdown> Service Note
SD730-424END*D

Getting start

Power On 1. Connect AC cable to wall socket.


2. Turn on the system breaker which located on the trans unit.
3. Confirm Standby LED lit on.

4. Press then system start up.


Wait 4 minutes to system complete boot up.

Power Off
1. Press .
2. System show “Power Control” dialog.
3. Select “Shutdown” or “Standby”.
4. System going to shutdown process.
Wait 3 minutes to complete shutting down.

233
Basic operation <Getting Start and shutdown> Service Note
SD730-424END*D

Shutting down the system


Press Power switch system show dialog below.
Do not turn OFF the power supply of the system during printing, data saving, or
invoking. Doing so may result in failure of saving or printing or damage to the
data. In addition, doing so during access to media may damage the media.

234
Basic operation <Getting Start and shutdown> Service Note
SD730-424END*D

Shutting down the system


Standby switch is not shown depends on system condition as below.

Standby
Message Meaning
switch
Shutdown is recommended. After starting, it has System is working more than 48 hours with out shutting down.
Appear
passed for a long time. Or system is NOT shutting down more than 20 days.
Shutdown is recommended. There is too much More than 15 times using Standby mode without shutting down.
Appear
number of times of standby.
Disappear Standby is halted. Select Shutdown or Cancel. System is working more than 96 hours with out shutting down.
(Required After starting, it has passed for a long time. Or system is NOT shutting down more than 30 days.
shutdown)
Disappear Standby is halted. Select Shutdown or Cancel. More than 20 times using Standby mode without shutting down.
(Required There is too much number of times of standby.
shutdown)
235
Basic operation <Getting Start and shutdown> Service Note
SD730-424END*D

Shutting down the system


Standby switch is not shown depends on system condition as below.
Customer using TEE probe. System show following message in Power Control
dialog box.
“A MTEE transducer is being used. Standby cannot be selected.”

During system setting operation


"Standby cannot be selected in the current system mode."

During the data registration process


"Data registration is in progress. Standby cannot be selected."

During use of an MTEE transducer


"A MTEE transducer is being used. Standby cannot be selected."

When a caution message is displayed

236
Basic operation <Security Mode> Service Note
SD730-424END*D

Security mode (Called HIPPA.)


Security mode will be activate set Enable for Security mode at;
[Menu] > [System setting] > [Other Setting]
If Security Mode set to enable, since then system will ask you ID and PASSWORD when
system boot up completed.
Here is factory default passwords.
You can add user accounts when log in with administrator role.
Role ID Password Remarks
Operator SERVICE SERVICE&
Administrator TOSHIBA TOSHIBA& You can add, edit and delete user accounts.

237
Basic operation <New Patient> Service Note
SD730-424END*D

When you want to start the examination, we need register patient


information. Other wise we cannot store any image to system.
Press [New Patient] .
Then “Patient Registration” screen display.

Enter patient name, ID and so on.


(In case of emergency, click [Emergency] button.)
Select Exam Type, then click [Confirm Start].
238
Basic operation <New Patient: Limitations> Service Note
SD730-424END*D

The following items cannot be changed once the patient information has been
registered. Confirm that these items are set correctly before starting the examination.
Exam Type
Patient information when "OB", "Fetal Heart", or "Endo-Vaginal" is selected for Exam Type
However, when one of these Exam Types is selected, the OB date information can be entered or
edited during OB measurement using the GA Input function.
It is not possible to enter a patient ID consisting of spaces only. Also, note that a
patient ID with a space at the end is deleted automatically.
The following characters cannot be used for patient IDs or patient names.
"*", "?", "^", "=", "¥"
The maximum number of characters is limited to 64 for the patient ID and patient name
due to DICOM restrictions.
Each of the last name, first name, and MI of a patient name can contain up to 64
characters.
The allowable maximum entry for the patient height is 299.9 cm (9'10"). If an attempt
is made to enter a higher value, "#" is displayed in the field.
The allowable maximum entry for the patient weight is 499.9 kg (1102.0 lb). If an
attempt is made to enter a higher value, "#" is displayed in the field.
If "#" is displayed for any patient information or exam information item, the examination
cannot be started even when [CONFIRM Start] is pressed.
Check and correct the entry for that information item.
239
Basic operation <New Patient: DICOM tag linkage> Service Note
SD730-424END*D

Green: System can be shown the information based on worklist information which from MWM server
Yellow: If we enter the information by manually, the information will includes in the TAG.
Red: System will set Exam Type name to (0008, 1030).
If system use DICOM MWM Information (Conventional) or (Comprehensive) in [Option], New patient.
System will set exam type information based on one of the worklist information listed as below.
(0008,1030) Study Description
(0040,0007)Scheduled Procedure Step Description
(0032,1060)Requested Procedure Description
240
Basic operation <New Patient: DICOM tag linkage> Service Note
SD730-424END*D

Tag Name in DICOM Standard Location in the system Remarks


(0008,0050) Accession Number Accession No.
Institution Name System Preset > General Preset Customer name will be include.
(0008,0080)
> Organization Name
(0008,0090) Referring Physician's Name Ref. Physician
(0008,1030) Study Description Exam Type
(0008,1040) Institutional Department Name Department
(0008,1050) Performing Physician’s Name Physician
(0008,1070) Operator’s Name Operator
(0010,0010) Patient’s Name Last Name^First Name^MI Three parameter combined by ^.
(0010,0020) Patient ID ID
(0010,0030) Patient’s Birth Date Date of Birth
Patient’s Sex Sex M = male
(0010,0040) F = female
O = other
(0010,1020) Patient’s Size Height Unit is m (Meter)
(0010,1030) Patient’s Weight Weight Unit is kg
(0010,4000) Patient Comments Insurance, Comment Combined these 2 parameter
(0032,4000) Study Comments Additional Info Also BSA included here

241
Basic operation <New Patient: DICOM tag linkage> Service Note
SD730-424END*D

Green: System can be shown the information based on


worklist information which from MWM server

Tag Name in DICOM Standard Location in the system Remarks


(0040,0001) Scheduled Station AE Title Scheduled Station AE Title
(0040,0002) Scheduled Procedure Step Start Date Scheduled Procedure Step Start Date
(0040,0003) Scheduled Procedure Step Start Time Scheduled Procedure Step Start Time
(0040,0007) Scheduled Procedure Step Description Scheduled Procedure Step Description
(0032,1060) Requested Procedure Description Requested Procedure Description
(0040,1001) Requested Procedure ID Requested Procedure ID
(0008,0060) Modality Modality
242
Basic operation <Examination screen> Service Note
SD730-424END*D

Examination screen
Patient name and ID Sex of the patient. System date and time.

Hospital name
Operator name Preset name

Probe direction indicator


* We can change the
Gray scale
position at Transducer in
System settings

Horizontal scale
* Depends on version, we can
hide the maker. Other setting of Gain and Dynamic Range
Probe information.
System settings information.
Upper: Probe model
Lower: Current frequency

Frame rate of current


image
IP number and Aplipure
parameter
Vertical scale * Depends on version

Store / Print Free area of Raw memory availability if Track ball function
indicator image data system has UIIP (IP PWB option) indicator
243
Basic operation <Examination screen> Service Note
SD730-424END*D

Related setting (Probe direction icon)

Probe direction indicator


* We can change the
position at Transducer in
System settings

System Settings > Transducer > Press [Shift] and [<]

Anti-Projection Projection

244
Basic operation <Examination screen> Service Note
SD730-424END*D

Related setting (Horizontal scale display)


Note:
AplioXG: V4.0
AplioMX: Initial
Horizontal scale
* Depends on version, we can
XarioXG: V3.0
hide the maker. Other setting of Xario: V8.0
System settings

System Settings > Other Setting > Page 2/2 > Press [Shift] and [<]

245
Basic operation <De - Identify> Service Note
SD730-424END*D

De identification function
System will not shown patient identical information on top of screen

If the function enable, system will


show patient information as a number.
It increase each examination.
But if once disable, the number will
be initialize.

Note:
AplioXG: Initial
AplioMX: Initial
XarioXG: Initial
Xario: V6.0

246
Basic operation <Store / Print image> Service Note
SD730-424END*D

Storing image
Required patient registration
Storing still image
Press “Still Store” in PIMS on TCS.
Press switch which assigned Still store
function on main panel.
(It is depends on user function.)

Storing cine image


Enable “Clip start” also define Quality and Capture rate etc.
Press “Clip store” in PIMS on TCS.
Press switch which assigned Still store
function on main panel.
(It is depends on user function.)

247
Basic operation <Store / Print image> Service Note
SD730-424END*D

Storing image
If customer wish to send image to DICOM image server when press still /
clip store switch. Define printer device in Output, System settings
Press [Menu] > Select “System Settings” > Output > Press [Shift] and [<]
Define DICOM image server for DICOM Server.

In case if DICOM image server connected, and if customer wish to send image while examination.
Set the printer here and define a switch for DICOM storing function or set “SERVER” for [Still Destination]
and / or [Clip Destination]. System send store image depends on these setting.

248
Basic operation <Store / Print image> Service Note
SD730-424END*D

Printing image
Press switch which assigned print (B/W or Color print) function on main
panel. (It is depends on user function.)
Define printer device in Output, System settings
Press [Menu] > Select “System Settings” > Output > Press [Shift] and [<]
Define printer device for Sono Printer [B/W] and Sono Printer [Color]

In case if DICOM printer


connected, and if customer wish
to send image while examination.
Set the printer here and define
a switch for DICOM printing
function

249
Basic operation <Store / Print image> Service Note
SD730-424END*D

Printing image (This screen may different depends on version.)


Printing stored image
Go to PIMS on TCS > Patient Browser.
Select the study and choose image which want to print.
Click [Print] in case printer defined as default printer of the system
In case if you wish use another printer device, click [Print to...].
Select the printer and select proper format then click [Print].
Printer device list

Number of copy
Film division

Select “ALL”!!

Execute printing

Delete this job


Confirm print queue
and result.
Change printer default
setting

250
Basic operation <Store / Print image> Service Note
SD730-424END*D

Confirm user function


Press [Menu] > System Setting > Exam Type
Press [Shift] and [<] > Click the drop down list as Application Preset
Editor. In this list, you can find “User Function” menu .

251
Basic operation <Review Stored Image> Service Note
SD730-424END*D

Review stored image


Stored image reviewing with Exam Review.
We have 2 ways to get the Exam Review.
Go to Patient Browser > Select Image > Click [Apply]

Press [ER] (Exam Review) if the switch available on main panel.


This function will be change depends on user function.
If there is no patient information or Press Patient Browser on PIMS, TCS.

Exam Review

If patient registered and study gas image.

252
Basic operation <Archive Stored image> Service Note
SD730-424END*D

Archive stored image


For DVD/CD (By DICOM format)
Go to Patient Browser (PB on main panel or PIMS > Patient Browser on TCS)
Select study or image which you want to archive, Click [Copy to...]
Select DVD/CD then click [OK]

For DICOM Server


Go to Patient Browser (PB on main panel or PIMS > Patient Browser on TCS)
Select study or image which you want to archive, Click [Copy to...]
Select Server then click [OK], in this time system allow to send image 3
different server by one operation.
* Selectable server should set on Option
in Tools, Patient browser.
253
Basic operation <Archive Stored image> Service Note
SD730-424END*D

Archive stored image


Option settings
Press [PIMS] > Patient Browser > Tools > Option

Data archive destination when execute


[Copy] in Patient Browser.

Define servers which shown a destination list


when execute [Copy to] in Patient Browser

Define clip image format as DICOM or


AVI (Windows converted ).
Usually set as DICOM.

254
Basic operation <Archive Stored image> Service Note
SD730-424END*D

Archive stored image


Option settings
Press [PIMS] > Patient Browser > Tools > Option
Define writing speed for DVD+R and CD-R
Depends on system version, it might not
display on latest system

Define you required Media test before burning


And / or Verify data after burning

Volume label when burning


media

Define column items in


Patient Browser.

255
Basic operation <Archive Stored image> Service Note
SD730-424END*D

Archive stored image


Option settings
Press [PIMS] > Patient Browser > Tools > Option

Define study list display method.


Auto:
If no patient information >
All study will be listed
If patient information >
Only today’s study will be listed
Today:
Only today’s study will be listed
Last one:
Depends on last patient list display method
“Today” or “All List”

256
Basic operation <Delete Archived: 1/2> Service Note
SD730-424END*D

Note:
Delete Archived <Setting> AplioXG: V5.0
System will list up studies which matched this setting. AplioMX: V2.0
Then if operator wish to delete, just execute delete XarioXG: V4.0
operation on the list. Xario: V9.0
Menu > System settings > Other Settings > 2/2 > [Shift] and [,]

Delete Archived Studies:


Enable: Use this function
Disable: Not to use this function

Period of keeping date:


Define how long system keeps the study
This case, system
Target data type:
If there are study which matched setting of
“Period of keeping date” system detect
study which matched on this condition.

All: Any study will be target of this function


Archived: Study which has been archived will
be target of this function.
257
Basic operation <Delete Archived: 2/2> Service Note
SD730-424END*D

Note:
Delete Archived <Operation> AplioXG: V5.0
System will list up studies which matched this setting. AplioMX: V2.0
Then if operator wish to delete, just execute delete XarioXG: V4.0
operation on the list. Xario: V9.0
PIMS on TCS > Patient Browser > Tools > Delete Archived
Select all of the study.

If “Delete Archived Studies” System will make the list of


is set to ON, study which matched on the
This switch will be appeared Delete Archived setting.

Execute deleting for


selected study.