Medical electric drill saw(WY-ZJ-1 、WY-J-1、WY-Z-1)


Application
Medical electric surgical drill and saw are mainly used in drilling and steel plate screws in trauma surgery.
Made of high-strength biomedical stainless steel, its perfect ergonomic design, infinitely variable speed,
low noise, low vibration and easy to use give the operator and patient a good experience. 4.0mm hollow,
suitable for all brands of needles to pass through, can be high temperature and high-pressure sterilization,
safe and reliable performance reduces the risk of surgery.

Product manual
1. Ergonomic design
2. More durable battery
3. Lightweight design, portable
4. High temperature resistance(135°C. Imported motor, stable performance)
Mini
Cannulated Mini Bone
cannulated
Product name drill(B120 Saw(A-J) saw(C- drill(A1200,
drill(C1200,
0, B500) J) A200)
C500)
Output
130 130 130 130 130
power(w)

Rotating
0-1200/500 0-18000 0-1200/500 0-18000 0-1200/200
speed(rpm)

Noise(db) ≤75 ≤75 ≤75 ≤75 ≤75

Disinfection(℃) ≤150 ≤150 ≤150 ≤150 ≤150

Battery(v) 14.4 14.4 8.4 8.4 14.4


Charging time 50min 50min 50min 50min 50min

Advantages:
1).Our whole power tool can be autoclavable.
2).We can autoclavable for 135 degree,with 1.6Mpa,in this way the bacilli and virus which can
not killed by the suffocating will be killed.
3).We use nickel-metal hydride battery is safe in use.
4).The machine has very low noise.
5).Our products has 1 year warranty period.

中国安徽省合肥市高新区柏堰工业园铭传路 TEL:+86-18556357195
Mingchuan Road ,High-tech Zone ,Hefei ,Anhui, China PC:231202
合肥市五洋医疗器械有限公司
Hefei Wuyang Medical Instrument Co.,ltd

Notes:
1).Sterilize temperature ≤ 135 °C,if the temperature is over 135 °C , the motor will be damaged.
2).Normal bone drill can only hold drill bits less than 6mm,acetabulum reaming drill with big torque,
rpm is around 300,canulate drill is used with nail or K wire ,rpm is around 600.
3).The battery need to be get out of the power tool after surgery.Otherwise the small automatic
discharge of the electric will make the life of the battery short.
4).The battery cannot be sterilized ,it is not autoclavable.
5).The cranial bur must be washed right after the surgery by purger in order to avoid the blood
stick in the bur,then we should make it dry.
6).The battery space in the handpiece of the tool should not have any water,after sterilization.
7).The cranial bur blade is made of medical stainless steel,in order to avoid broken,so it is flexible
and not very hard,it is easy to be attrite,we suggest each blade only use for 1 surgery.
8). The medical electric drill, saw only the equivalent of medical doctors or professional use in
a short time,can not use a long time continuously.
9). Charger power cord should not be traces of oil, solvents and other chemicals, erosion or mechanical
damage, if it is found to be immediately replaced.
10).Users may not disassemble or repair, or responsibility by the user, the Company is not responsible for
repairs. Man-made damage is not covered under warranty, the costs is incurred by the users themselves.
11). The tools need repairing during the warranty period,should be packaged well before transportation.