You are on page 1of 3

ULTRAHANDS

BIOMEDICAL ENGINEERING PLC

LABORATORY SHAKER CALIBRATION AND MAINTENANCE


REPORT CERTIFICATE

Medical center:- K.K WOREDA 03 HEALTH CENTER

Device name :- shaker Device ID: - 008

Model :- VPN - 360 SN :-1051375

Department :- ____________________________ Site temperature :-24

Capacity :- _________________________________ Status :- ________________________

Date of visit :20/05/2023

Next date of visit :- 21/11/2023

Engineer’s Full Name :- Hayt Ali mohmed

This certificate is valid from 20/05/2023 to 21/11/2023.

STATUS CHECKLIST

Type of service Device working description

o Routine maintenance o Orbital angle of shaker …………………………………..


o Installation o Max speed (Rpm) …………………………………………
o Repair o Other ………………………………………….
o Training
o Calibration
o Application
o Other ………………………………………….

er

Immunoassay Analyzer

Tel- +251909545190 +251968594321/ +251907552655 Email- ultrahands1@gmail.com

Addis Ababa, Jomo Kenyatta Street, Bambis Hailegebriel building 3rd-floor office No 308/ለ.
ULTRAHANDS
BIOMEDICAL ENGINEERING PLC

Machine parts Clean Not Damaged Functional


yes No yes No Yes No
External body

Electrical connections

Working Rotor/ motor

Speed Adjusting knob

Time adjusting knob

On/off switch

Appropriate water level (Surface)

SPEED CALIBRATION RESULT USING TACHOMETER

Speed 1:-

Time Set Displayed Measured Tolerance

3 min 100 100 99 1%

5 min 200 200 198 1%

8 min 250 250 249 0.77%

Average :-

If no explain
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________

Engineers comment
___________________________________________________________________________
___________________________________________________________________________
___________________________________________________________________________

Tel- +251909545190 +251968594321/ +251907552655 Email- ultrahands1@gmail.com

Addis Ababa, Jomo Kenyatta Street, Bambis Hailegebriel building 3rd-floor office No 308/ለ.
ULTRAHANDS
BIOMEDICAL ENGINEERING PLC

Customer feedback
it has 80% good water leveling surface. Good working condition.

Name :- ___________________________ Designation :- _______________________

Contact No :- _______________________ Signature :- _____________________

Date :- ____________________

Tel- +251909545190 +251968594321/ +251907552655 Email- ultrahands1@gmail.com