Professional Documents
Culture Documents
by
Introduction
The measurement of propulsive power is required at least once in the life of
a ship, and quite often the measurement of power or torque is made regularly as
part of a general study of performance. This paper describes the experience
gained on one type of torsiorimeter, namely that using wire or foil resistance
strain gauges.
In 1957, Ship Division of the National Physical Laboratory started upon a
research programme involving the need for a torsionmeter to measure torque on small
diameter (k in) shafts. Because of the small diameter involved, (making most
other systems impracticable) together with the general strain gauge experience of
the Division, it was decided to use a strain gauge system. Since that date a
number of shaft systems have been gauged, all the while building up a considerable
background knowledge. Critical examination of the results has led the author to
believe that a strain gauge torsionmeter is as good as any system for all practical
applications and has a number of advantages over most.
Jr
T q GQ
1
where T
J
=
=
Torque
Polar moment of inertia
q = Sh.ar stress
r = Radius
G = Shear modulus
Q = Angle of twist
1 = Length
this may be rewritten as
qJ GQJ
T
r
2GeJ GQJ
T where e = direct strain.
r 1
Assiuning J,l and r are known constants, then a chosen system must measure
either:-
Method (1) GQ or Method (2) Ge
likely to occur due to hull flexing. There is the added problem of shaft
couplings for which the stiffness may be difficult to determine. If the displace-
ment is to be measured with reference to two points on the shaft, then 1 will
probably be limited to a maximum of 3 ft. This is due to the conflicting require-
ments of producing a light easily portable system with sufficient rigidity not to
be influenced by shaft rotation. The whole system would normally be calibrated on
a dummy shaft subjected to known angles 0 and then clamped to the ship's shaft.
The method of clamping which seems to be preferred is by means of half rings
bolted together. Although this system gives rise to doubts as to the effective
clamping position, and hence the length 1, it does provide a very rigid method
of attachment.
Method (2) (Ge) The torsional shear stress produces principal strains on helices
which are at +5° to the shaft axis. It is therefore necessary to attach some form
of extensometer or strain measuring device. Since the measurement of strain is
along a curved path, the conventional mechanical extensometer is unsuitable and
the most obvious choice for such a measurement is the resistance strain gauge
Accurate positioning of the gauge is important, but if strain is being measured
the length over which the measurement is made is not important. However the use
of the strain gauge involves one further variable quantity, namely gauge factor,
which is the ratio of electrical to mechanical strain.
Electrical Measurement
The mechanical. to electrical signal conversion may take a variety of forms,
the most common being voltage, frequency or phase dependent. Although the latter
two methods may appear to be preferable in that the signal amplitude is unimportant,
modern electronic technology has made possible the accurate measurement of
extremely small voltages.
14_
1. Neasuring Equipment
This was a strain indicator of commercial make using a carrier system of strain
gauge bridge excitation, a 5 volt supply having a frequency of 2,500 or 1,000 cycles
per second. The system operates either as a direct reading meter device or as a
null balance system. A sensitivity switch changes the value of the full scale
deflection of the meter in ranges of 30e, 300te, 1,000i.te and 10,000e steps. The
null balance is achieved from two switch positions of 1O,000e and 1,000e steps
plus a continuous dial adjustment totalling 2,000.ie marked in 10e intervals. This
latter dial can be read to better than ±4te. The strain steps referred to assume
one active strain gauge having a gauge factor of 2. The readings for a four-arm
torsional strain gauge bridge must be modified as follows:-
Reading 2
Strain = x
No. of Gauges Gauge factor
-6
The absolute accuracy claimed by the manufacturer for this measuring system is as
follows: -
Experience with a number of these strain indicators has shown that the order of
repeatability and accuracy is certainly higher than quoted. Provision is also
made for a rectified signal to be available from two output terminals, voltage
being related to the full scale deflection of the meter. Internal calibration
control and drift monitoring facilities are also provided. During calibration the
null balance system is used, but during measurement it is more convenient to use a
combined technique whereby balance is achieved in 100te steps, the residual being
read directly from the meter.
2. Slip Rings
Both commercial and "homemade" slip rings have been used successfully, the
choice is one of convenience, the commercial ring being easier to fit, the home-
made or constructed ring being about a quarter of the cost. Both types of ring
mounted on a shaft insert are shown in Fig.5. Wherever these units have been used
they behaved well and have shown no detectable wear. All four cases referred to
in Fig.k used. commercial slip rings and in case III a constructed unit on one shalt
was used for comparison purposes. In case II, the rings and brushes were
subjected to an estimated 1,000,000 revolutions without any visible deterioration.
In cases II and III the same commercial rings were used, the difference in shaft
diameter being accommodated by semi-circular packing pieces.
In case III, because of the gauge position it was possible to take zero torque
readings both when the shaft was stationary 3nd rotating at 380 r.p.m. The
difference between the zeros, stationary and rotating was not measurable and must
therefore have been less than 1 microstrain. This "rotating" zero was recorded
during single engine trials, and the trailing propeller caused the disengaged shalt
to rotate.
In cases II, III and IV the rings, whilst in use1 were subjected to
unintentional contamination by various fluids. Case II was positioned under a
fueld pipe leak, case III having one slip ring assembly splattered with grease
from a gear coupling, and in case IV, the drive gear box developed a leak which
spilled oil over the slip ring whilst running. In no instance was there a
noticeable effect on the readings, there being no erratic change of signal level
nor change of shaft zero from the original clean state.
-7
3. Brush Gear
In all cases commercial brushes made of silver/graphite were used. These
brushes, two per ring, were supported on beryllium copper canti].ever springs. The
brushes were attached at the connection end to a common insulating block, arid then
angled out to the shaft. The brush contact points were thus at about 600
separation around the shaft. This arrangement made for very easy setting up and
removal. Positioning, to obtain the correct spring pressure, was not critical and
the unit could be disengaged whilst the shaft was rotating.
k. Strain Gauges
In case I the gauges used were paper-backed, grid wound iridium platinum wire
gauges. These were chosen for the high gauge factor (6). This trial was part of
a research into propeller excited vibration (Ref.k). Under these circumstances it
was more important to record the fluctuating strains with good signal to noise ratio
than to achieve a measurement of mean torque to a high order of accuracy. However,
as an AEI torsionmeter, fitted by the British Ship Research Association was
available, it was decided to compare results obtained from both meters. The type
of gauge used was very sensitive to temperature change and to reduce this effect
careful gauge resistance selection was necessary. Two torque bridges were fitted
in the same axial position, but separated by 90° in the radial direction. The slip
ring assembly was mounted over the gauge position. The gauges were bonded with
rapid setting cement. Due to the extreme cold, local heating was introduced by
means of a butane gas burner in the shafting compartment of the ship. Whilst this
raised the shaft temperature from 0°C to a more acceptable temperature (about 15°C),
the humidity became extremely high, any unprotected steel rusting rapidly. In
case II, paper-backed Karma wire grid wound gauges were used. These gauges were
bonded on to the shafts out in the open alongside the ship's slipway. Electric
heaters were installed near the gauge positions and the whole covered with a
tarpaulin. During the 2k hour curing time it rained, again giving extremely high
humidity conditions. The gauges were then waterproofed and calibrated, and the
shafts installed in the ship. Approximately 2 months elapsed between calibration
and use.
In cases III and 1V1 epoxy-backed, etched foil self-temperature-compensated
gauges were used and bonded under laboratory conditions.
5. Gauge Glue
These are listed in the case histories but special mention should be made
about Araldite Twin Pack. This seems to be an. ideal strain gauge glue for field
work. Experiments in the laboratory have shown it to be as good as the conventional
-8
epoxy strain gauge resins, and it does not seem to be too critical as to the ratios
of the two part mix. Small quantities can be mixed easily and it can be purchased
at any local ironmongers. One word of warning; difficulty has been experienced on
diameters below 1 in. Under these circumstances the gauged units experience very
undesirable hysteresis. It is suspected that this trouble is due to a thick
deposit of glue over the centre section of the gauges If this is so it probably
stems from the gauge bonding technique, where little or no bonding pressure is
applied.
6. Revolution Counting
For case I, in conjunction with the slip ring manufacturer, NPL devised a
method of pulse counting. This consisted of a segmented ring composed of alternative
silver and epoxy resin segments, the segments being incorporated into the slip ring
construction. The silver segments were joined together electrically during
manufacture and earthed to the shaft during installation. A capacitance probe was
then used as a pulse generator, the pulses being fed to an electronic counter via a
suitable amplifier system.
The system has proved to be very reliable. It has a number of advantages over
reflective light, photoelectric or inductive systems, since it is not affected by
oil or dirt and the pulse size is not related to the shaft rotational speed. In
case II, NPL were not required to measure power, only torque, and so no tachometer
was fittede
In case III, the high speed diesel engines were fitted with shaft pick-up
points suitable for hand-held mechanical tachometers. For this particular trial,
where the engines were speed governed, a hand-held tachometer was considered more
convenient and was used.
In case IV, the engines were supplied 'ith electronic counters.
Case II
The shafts for this ship were calibrated by means of a very simple calibration
rig; the results of the calibration are shown in Fig.8.
To calibrate, each shaft in turn was laid on a concrete floor and restrained
at each end in the vertical direction by two metal straps grouted to the floor.
Single torque arms were bolted to each end of the shaft, through the normal shaft
coupling flanges. One arm was bolted to the floor, the other arm being attached at
its outer end to a tripod via a tension screw jack and a measuring hydrostatic
capsule (accuracy ±5 of full scale deflection). The torque applied was equivalent
to the capsule reading times the torque arm distance of 10 ft. To reduce the
errors due to friction, the rig was hit at the metal strap positions with a heavy
mallet at each load condition.
Case III
A simple calibration rig was constructed at NPL and the constructional details
are shown in Fig.9. The load measuring capsule was similar to that used in
Case II. The torsionmeter units were calibrated twice and used for ship trials on
three occasions over a period of eleven months. Note was taken of zero shift and
overall repeatability and these are discussed later. The calibration was made at
Ship Division, Feltham, whilst all trials work was carried out on the Solent.
Calibration results are shownin Fig.10.
Although no other torsionmeter was available, the engine manufacturer had
calibrated the engines on a brake at the factory, arid, during the trial, records
of r.p.m. and various engine parameters were taken. From this information
horsepower figures were derived. A comparison of manufacturer's and NPL values of
Case IV
In this series of trials the shipbuilder and NPL produced 12 shaft torsion-
meters, each shaft being fitted with two strain gauge bridges. The shafts were
calibrated in both the port and. starboard turning directions in the rig mentioned
- 10 -
for Case III. Three shafts selected at random, as far as the strain gauge bridges
were concerned, were then sent to the engine manufacturer to take part in engine
and gear box trials. This enabled a comparison to be made between a calibration
of a simple static type and results obtained froi a Heenan and Froude water brake
under running conditions. The results of these calibrations and trials are shown
in Figs.12 and 13.
Discussion of Results
Case I
The results, Fig.6, show good agreement between the two measuring systems. It
was to be expected that the two strain gauge bridges would be in better agreement
with each other than with the AEI torsionmeter, since they were measured at closer
intervals of time. However, there is no doubt that the best smooth line, as
defined by the strain gauge system, was confirmed by BSRA readings. The results of
the AEI torsionmeter static calibration are shown in Fig.7, and it is interesting
to note the change of calibration factor with respect to output reading; this
variation is approximately 3%. Although the readings were converted to torque by
use of this correction curve and calibration formula, it is preferable to have a
torsionnieter which has a more linear relationship between torque and output reading.
As will be seen in Cases II, III and IV, strain gauges meet this requirement.
Case II
The object of including the results of this trial is to show the possibilities
of calibration of smaller shafts by simple means. The calibrations in Fig.8 show
the repeatability that is possible on two separate shaft systems.
The general shape of the calibration curve is influenced by a number of
factors; friction within the calibration system, load capsule calibration errors,
and the inaccuracies caused by the displacement of the torque arm through a large
angle under load. This displacement was caused by the twisting of the shaft under
torsional loading and the flexibility of the torque arm. Since the jacking anchor
point was fixed, there was an effective change of length of the torque arm. (See
Fig.lk).
The friction losses almost certainly produce the steeply falling portion of
the curve in the lower part of the calibration, whereas the rising portion of the
curve is due mx4inly to the angular displacement error mentioned above.
Case III
This set of results (Figs.10 and ii) is excellent, both the repeatability
achieved over a period of months and the stability of the zero indicate that the
strain gauge system offers long term stability for torsionmeter systems. The zero
scatter was of the order of ±1 i/'+% of Full Scale Reading spread over a period of
8 months. This period involved two trials on separate ships, two calibrations in
the laboratory, and the use of two similar strain indicators. Using the shaft
diameters and gauge factors, the calculated values of modulus of rigidity of the
two shafts are 11.91 x 1O6 lbf/inch2 and 11.97 x lbf/inch2, which are
satisfactory values for steel, lying within the spread of modulus measurement made
by the BSRA (Reference 3). Again, there is the steeply falling curve at the lower
end of the calibration due to friction, but even accepting this possible source of
error the dynamometer has a linearity of better than ±1% over half of the torque
range and better than ±2% over four fifths of the range. It is now normal
practice to ignore the lower portion of the calibration curve and to establish the
calibration factor by extrapolation of the upper part of the calibration towards
zero. In this particular case this would give calibration factors of approximately
6.25 and 6.28 lbf ft/division. The correctness of such a decision was confirmed in
Case IV when the calibration curve of similar shape was treated in the same way.
The results are discussed below.
Case IV
During the gauging and calibration o± these shaft inserts, the opportunity was
taken to determine the effect of the proximity of the flange upon the gauge output.
These particular shaft inserts were 8 inches between flanges and five strain gauge
bridges were bonded to one shaft, the gauge positioning being at equal intervals
along the shaft The shaft was loaded torsionally and. each bridge output recorded.
Only the bridge nearest to the flange gave a significantly different output of the
order of 2%. The others gave readings which were within the tolerance of the
gauge factors for this particular set of gaiges.
The twelve shafts were calibrated in the rig described for Case III and then
three of the shafts were used during engine trials on a test bed.. The calibration
results for these three shafts and the comparison of brake and torsionmeter
readings are shown in Fig.12 and tabulated in Fig.13.
The shaft static calibration shows the now expected shape for this type of
calibration, and it was decided to extrapolate the best fit straight line obtained
from the higher portion of the calibration, to derive the lower portion. This
calibration line was used to derive the torque values for the brake/torsionmeter
tests. In this comparison only four torque values were more than 1% in disagreement
with the brake results.
The biggest difficulty with this type of comparison is the difference in
response of the two systems. The brake recording dial is deliberately damped to
give a stea&y reading and this damping produces a significant delay in recording
- 12 -
changes in torque. Since the strain indicator is read at the same time as the
brake dial, errors must occur, This is the probable reason for the apparent
scatter of results even at apparently constant brake torque conditions.
Conclusions
Methods of measuring ship shaft torque have always been a controversal
subject, each method having its own champions. However a number of systems
produce equally accurate results when installed by skilled technicians. The
general accuracy of all the systems is in the order of ±2% of maximum torque and
under favourable conditions ±1% is possible. This is probably the limit if the
ship shaft is used as the spring element. Greater accuracies could be achieved by
special calibrated shaft inserts, but the production of such units for large
torque values would be costly.
The advantages of the strain gauge system lie in other directions and the
more important can be listed as follows:-
Shaft size
Many torsionmeter systems are not suitable for attachment to shafts of small
diameter. The strain gauge does not suffer from this limitation. Slip rings of
the commercial type can be made or adapted to suit a specific shaft size, but the
temporary ring constiuction is limited to a minimum diameter of approximately
3 inches unless pre-formed during manufacture.
Installation time
At first sight it would appear that a "clamp on" torsionnieter would be quicker
to install than a strain gauge system. However in the author's experience this is
not the case. On one occasion the author and one assistant fitted two complete
strain gauge torsionmeter systems in less time than it took to install a "clamp on"
torsionmeter fitted to the same shaft.
If slow drying glues are used the installation time will take 2k hours
although the actual time "on site" is only about 6-8 hours. üth quick drying
glues or impact adhesives the total "on site" is, obviously, 6-8 hours, These
Skilled labour
All torsionmeter systems require skilled labour to ensure a proper
+. Cost
The price of a torsionmeter installation will depend upon the sophistication
of the system used. However, in its simplest form, the strain gauge torsionmeter
is cheaper than any other system.
Such a system would employ D.C. voltage bridge excitation with the output
measured using a D.C. microvolt meter calibrated in iv/v or direct strain. With
temporary slip rings such a system would have a total cost of less than 1OO plus
installation costs. Such a system was adopted by the White Fish Authority and
described by them in Reference 5.
Rotational speed
The gauges, having negligible mass, are neither affected by high rotational
speed nor influence the shaft. The shaft speed is limited only by the behaviour
of the slip ring assembly. Since the ring mass is small and balanced about the
shaft axis, the ring speed is limited only by the rubbing speed between ring and.
brush. Modern techniques are such that "noise free" running is possible at speeds
far in excess of any likely ship shaft speed.
Reliability
It is likely that there have been more experiments abandoned due to strain
gauge system failures than any other type of electrical strain or displacement
measuring device. Because of this the strain gauge has received much bad
publicity. This is unfortunate since the majority of failures have occurred when
no other system could have been used to perform the required task. These failures
are usually caused by installation damage, mainly wiring, and it is this type of
failure which has caused much of the criticism of strain gauge systems. In fact
the author cannot remember an occasion where failure could. be directly attributed
to gauge failure unless due to damage from some external source.
A strain gauge torsionmeter suitably protected rtfl give a high degree of
reliability, with the additional advantage that a second gauge bridge can be
bonded to the shaft to be available in the event of damage. This extra strain
gauge bridge takes little extra time to install and one bridge can be used as a
measuring check on the other to confirm satisfactory installation of both systems.
The strain gauge system, like all others, has a number of disadvantages, and
for satisfactory operation the effects of these disadvantages must be reduced to a
minimum:-
(a) Damage during installation
A strain gauge is a fragile device and is easily damaged during the
installation period. Every care must be taken to ensure adequate protection at
all times.
- 1k -
Earth resistance
The strain gauge bridge will cease to measure accurately or even have a
linear calibration if the bridge earth resistance is allowed to fall below 100
megohms. This can lead to difficulty during installation unless adequate
precautions are taken.
Brush gear damage
The recommended use of cantilever springs for the brush gear requires that
adequate protection be given to them since they are fragile.
Machining accuracy of the shaft
In the case of shafts which are oval or for hollow shafts where the bore is
not concentric, errors occur in trying to establish the values of torque from the
strain data. Under these circmtances calibration is the only suitable solution
in obtaining accurate results.
Cable resistance
If long runs of cable are used due allowance must be made for the voltage
drop along the bridge input supplies. This was necessary in Case IV where the
recording station was situated on the bridge of the ship. This is done by
comparing the gauge bridge resistance to cable resistance and correcting the strain
reading proportionately.
Acknowledgement
The work described above has been carried out at the National Physical
Laboratory.
The author wishes to acknowledge the co-operation and assistance of the
following organisations -
The author is also indebted for the help of staff of Ship Division, NPL; in
particular, to members of the Equipment Group.
- 15 -
References
(i) COOK, R. 1951 Marine torsiozimeters and thrust meters. Trans. I. Mar. E.,
Vol.63, p,ll5.
(1,) SIL1TERL]AF, A.., MARWOOD, W.J., BOYLE, H.B. 196k Some ship and model
measurements of unsteady propeller forces. Trans. R.I.N.A., Vol.106,
p.klS.
FIG. 3.
CASE NUMBER 1 2 3 4
OCEANOGRAPHIC HIGH SPEED HIGH SPEED HIGH SPEED
TYPE OF SHIP RESEARCH ROUND BILGE liOft. MOTOR PLANING
VESSEL CRAFT BOAT CRAFT
SHAFT DIA. 9 InS 51/4 InS 49x 40(HOL) 51/4X 31/4(HOL)
Z MATERIAL STEEL. STEEL STEEL STEEL
0
MAX. R.P.M. 190 42.0 1,000 1,700
4
o MODULUS OF
RIGIDITY 11B6x106 UNKNOWN UNKNOWN UNKNOWN
o (lbf/1n2)
w MAX. LINEAR
a.
STRAIN 220/..Le 530ue 570,u.e 25o,e
-AU GE
I-
LI. MAX. SHEAR
4 STRESS 5,300 12,700 13,700 6,000
U)
(Ibf/Ifl2
'
2
N OF SHAFTS 1
PAPER
2
PAPER
= 3
7,500
Ui
±10
7,000
Lii
0-i
0>
6,500
6,000
0 10 ao 30
SHAFT REVOLUTIONS/ SEC (n)
CASE HISTORY I
z
0
lii
1O5
uI<
'czW
a:
I04
tow
ow
103
I-U,
jg
wo
0
0
I-
101
100
30 40 50 60 70 80 90 100 110
TORSIONMETER READING (DIVISIONS)
CASE HISTORY I
X SHAFT. 1.
-I- SHAFT. a.
1700
1650
1.00
1550
1500
200 400 600 800 1,000 1,200 1,400 1,600 1,800
READING (DivisioNs)
CASE HISTORY II
LIFTING JACK.
- HYDROSTATIC
CAPSULE.
LOADING GUIDES.
////////////////////////////////// ///
FIG 9A SIDE ELEVATION
640 0
0
I-
x
0 + SHAFT.a.
LL
630
9 0 1g
620 x 0
SHAFT.1.
2.00 400 600 800 1,000 1,200 1,400 1,600 1,800 2.,000
BRIDGE OUTPUT (DIVISIONS)
1240
+
x
o
- BRIDGE A. SHAFT 3
BRIDGE
BRIDGE
B.
B.
SHAFT 3
SHAFT 12
1230
0
ta20 0
w
0 D
0
I-
1210
tL
0
w
z
0 --0- J
0
1200 a-
U-
-J 0
4-)
iiqo 0'
'-4
.1-I
100 200 300 400 500 600 700 600 900 10O
BRIDGE OUTPUT (DIVISIONS)
CASE HISTORY
(THESE SHAFTS USED FOR ENGINE TEST BED TRIAL)
FIG. 13
R.P.M.
(R TORQUE
TORQUE TORQUE
BRAKE TOR W
x ioo)%
lbf ft lbf ft
SHAFT 3 BRIDGE A
SHAFT 3 BRIDGE B
SHAFT 12 BRIDGE B
SHAFT 2 BRIDGE A