Live Vitals Access Device: - Accuracy Reports

BLOOD PRESSURE & HEART RATE BODY TEMPERATURE


Measuring method: Oscillometric method Measuring method: Infra-red
Cuff perimeter: 22-35cm Body part: Forehead
Measuring range: Blood pressure: 0-300mmHg Measuring range: 32 -42 ℃
Systolic pressure: 60-260mmHg Tolerance: 36-39℃ +/- 0.2℃, >39℃ or <36℃ +/- 0.3℃
Diastolic pressure: 40-199mmHg
Heart Rate: 40-180bpm
Tolerance Rate: Blood pressure ±3 mmHg
BLOOD OXYGEN
Heart rate ±5%
Measuring method: Reflective
Measuring range: 90%-99%
ECG Tolerance: +/- 2%
Measurement mode: Single Lead (supports multi-lead)
Input impedance: >10MΩ
Input circuit current: <0.1μA BLOOD GLUCOSE
Calibration voltage: 1mV +-0.05mV Measurement range: 1.1mmol/L~33.3mmol/L
Sensitivity tolerance: <10% Measurement time: 5S±1S
50Hz interference suppression filter: >20dB When the sample concentration < 5.5mmol/L
Maximum input range: +/- 4.7mV Tolerance: (100mg/dL), the tolerance <0.42mmol/L (<7.7mg/dL)
Polarization resistance voltage: +/-300mV, Amplitude variation <10%
Frequency response: 0.5-40Hz When the sample concentration ≥5.5mmol/L
Read overshoot < 10%-20% (≥100mg/dL), the tolerance <7.5%.