C A L I B R A T I O N C E R T I F I C A T E

Reader SN: 20242 Process Defn: TLD100_1 Status: PASS

Calibration Date: 06/24/14 10:13:54 Operator: A1


Edit Date: 06/24/14 09:52:52 Operator: A1

Exposed Samples: 6 Min Exposed Samples: 6 Exposure: 10 mSv


Annealed Samples: 6 Min Annealed Samples: 6

Exposed weighted raw counts avg: 2,2157


Exposed weighted raw counts CV: 2,99
Exposed weighted raw counts CV max: 10,00

Annealed weighted raw counts avg: 0,0715


Annealed weighted raw counts CV: 12,37
Annealed weighted raw counts CV max: 15,00

Light source average: 1024,3


Light source CV: 0,17
Light source CV max: 5,00