Operation Amp Service Manual Model PXP 60hf Portable X Ray Unit
Operation Amp Service Manual Model PXP 60hf Portable X Ray Unit
,
Ltd.
CONTENTS
TABLES
1 Specifications ---------------------------------------------------------------- 19
ILLUSTRATIONS
1 Operation Console Panel and Display ---------------------------------------------------------------- 12
2 Dimensions ---------------------------------------------------------------- 20
3 Labels ---------------------------------------------------------------- 21
1 Notice ----------------------------------------------------------------------- 22
2 Operation Sheet ----------------------------------------------------------------------- 22
3 Block Diagram ----------------------------------------------------------------------- 23
4 Calibration Process ----------------------------------------------------------------------- 24
5 Calibration Code and Data ----------------------------------------------------------------------- 28
6 Error code & Countermeasure ----------------------------------------------------------------------- 28
7 Schematics ----------------------------------------------------------------------- 29
TABLES
1 mA Calibration Code ----------------------------------------------------------------------- 28
2 Filament Preheat Calibration ----------------------------------------------------------------------- 28
3 Error Code ----------------------------------------------------------------------- 28
2
Product Manual: PXP-60HF Poskom Co.,
Ltd.
PART I. OPERATION
1. GENERAL DESCRIPTION
1.1. OVERVIEW
PXP-60HF is one of PXP-HF portable family generators. It consists with the highest power inverter in
PXP-HF family of X-ray generators. Likewise of others, the PXP-60HF is most compact, ultra-light
weight and powerful HF inverter type of x-ray systems in this portable category. It is designed
especially for nursing home care or home care medical market and small medical clinics as well as
veterinary practitioner or battle field use, All PXP-HF family x-ray units are the ripple free HF type of
generator in producing 40~50% more powerful radiation output than conventional x-ray units in its
class.
Therefore, PXP-60HF is the perfect x-ray portable unit for an indoor and outdoor medical diagnostic
practitioner in the radiographic application where requires an easy of operation and mobility.
Features are;
-Ultra light weight and compact size
-High frequency type of generator
-High output 100kV / 60mA
-APR pre-set (8) memory PROM
-Remote control software
-Built in collimator and laser pointer
-Auto line compensation
-Digital readout reverse switch
-Accessories: metal carrying case, Hand switch, power cord
-Power requirement: 220 Volt. 50 / 60 Hz
This consolidated manual consists of operating and service instruction. This operator’s manual
provides PXP-60HF operating instructions for X-ray technicians, radiologists and other medical
personal trained in the use of X-ray equipment. Service manual provides installation instruction,
calibration and routine maintenance for field service personnel.
This manual is not designed to replace or substitute for certified training in the radiological or medical
field. In reading this manual, the operator should be qualified in the use of equipment of this type. This
portable X-ray system is only to be used by licensed practitioners in diagnostic, operative and post-
operative procedures.
3
Product Manual: PXP-60HF Poskom Co.,
Ltd.
PXP-60HF portable X-ray can be used in variety of diagnostic radiographic systems such as mobile
stand or radiographic table integrated with X-ray tube mounting devices. Although many of the
features of these portable systems were developed with radiographic application in mind, operation is
described in general terms to permit the use of those features in radiographic application
This manual provides guideline for the safe use and operation of the portable system. The operator
should be in trained in the use of equipment of this type and should refer to this manual in conjunction
with that training.
It is the responsibility of the owner to ensure that the system is operated only by properly trained,
qualified personnel who have obtained credentials from local, state, and federal authorities where
required.
Potential hazards exist in the use of medical electronic devices and x-ray systems. All users and
operators must understand the safety and emergency procedures, and the operating instructions given
in this manual
Poskom Co., Ltd. certifies that each system complies with the Radiation Control for Health Safety Act
in Title 21, Chapter I, subchapter J of the Code of Federal Regulation in the USA and EN 60601,
general requirements for radiation protection in diagnostic X-Ray equipment for IEC 601-1-3 in
European Nations.
Poskom and distributors assume no responsibility or liability for after-sale operating and safety
practices, nor can it be responsible for personal injury or damage resulting form misuse of its portable
x-ray systems.
Please address a question and comments regarding safety issue to Poskom service organization
and/or each distributor in your region.
Possible hazards are associated with unsafe operation of this type of portable X-ray equipment. These
hazards are described on the following warning or caution signs with the safety precautions which
should be taken.
CAUTION
The X-ray radiation could be dangerous for patients and operators
unless following safety regulations are strictly observed.
4
Product Manual: PXP-60HF Poskom Co.,
Ltd.
2. OPERATIONAL SAFETY
User’s safety, easy of operation and product reliability are our major concern in designing this PXP-HF
family of generators. However, it is advisable to observe the following instruction and guideline for your
additional safety in operating this portable.
A. This product should be operated only by or under the supervision of legally qualified people.
B. PXP-60HF is designed for the radiographic uses and not for fluoroscopy or other associated
applications.
C. PXP-60HF should be used for the diagnosis, not for the therapy.
D. PXP-60HF is specified as Class type B equipment under the IEC 601-1, 2 standards.
E. Do not modify the equipment at your discretion and in case any modification is required
unavoidably, ask the help of Poskom or its authorized dealer for the service.
F. This system has been calibrated for optimal operations.
G. If you have any troubles with the equipment, power it off immediately, and make contact with
Poskom or its authorized dealers for their service and assistance.
H. If you want to connect any devices made by other companies to this equipment, ask Poskom or
its authorized dealers for their instructions.
5
Product Manual: PXP-60HF Poskom Co.,
Ltd.
3. ENVIRONMENTAL SAFETY
3.1. Avoid following places for the normal operation and safe storage.
A. Where the equipment is exposed to water vapor.
B. Where the equipment is exposed to direct sunlight.
C. Where the equipment is exposed to dust.
D. Where the equipment is exposed to high humidity.
E. Where there is a ventilation problem.
F. Where the equipment is exposed to salty atmosphere.
G. Where the equipment is exposed to chemicals or gas.
3.2. For sound operation of the machine, it should keep away from the place with strong vibration and
maintain proper environment and conditions.
A. Operation Environment
Temperature range 10°C ~ 40°C,
Relative Humidity Range 30% ~ 75% RH
B. Optimal Temperature and Humidity
Temperature range 17°C ~ 23°C,
Relative Humidity Range 40% ~ 60% RH
3.3. For safe storage and transportation, you must keep following range of temperature, humidity and
atmosphere.
Environment for storage and transportation
Temperature range -25°C ~ +60°C
Relative Humidity Range 10% ~ 95% RH
4. RADIATION SAFETY
CAUTION
The ionized radiation could be dangerous for patients and operators
unless following safety regulations are strictly observed.
A. Users and operators should wear appropriate protecting devices and clothes.
B. Stay distance from the radiant sources and all the possible secondary radiation zones.
C. Eliminate all unnecessary objects near the exposure zones.
D. The distance from the focus to skin should be kept at least 20 cm.
E. For the experimental uses, please apply the lowest possible values about kV, mAs and
exposure time.
F. Be careful not to exceed the limited radiograms in the exposure area.
6
Product Manual: PXP-60HF Poskom Co.,
Ltd.
The unit must in reference to the mentioned EMC-hints in the accompanying documents be installed
and operated.
Portable and mobile RF – communicating systems (such as cell phones) can have influence to
medical electrical equipment.
RF – emission
Class A
Acc. CISPR 11 The PXP-60HF is suitable for use in all establishments,
Harmonic emissions including domestic establishments other than domestic and
Not applicable
Acc. IEC 61000-3-2 those directly connected to the public low-voltage power
Voltage fluctuations/ Voltage fluctuations supply network that supplies buildings used for domestic
Flicker emissions complies Flicker not purposes.
Acc. IEC 61000-3-3 applicable
7
Product Manual: PXP-60HF Poskom Co.,
Ltd.
8
Product Manual: PXP-60HF Poskom Co.,
Ltd.
Conducted RF 3 Vrms
3 Vrms d = 1,2 √P
Acc. IEC 61000-4-6 150 kHz to 80 MHz
Radiated RF 3 V/m
3 V/m d = 1,2 √P 80 MHz to 800 MHz
Acc. IEC 61000-4-3 80 MHz 50 2,5 GHz
NOTE 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
9
Product Manual: PXP-60HF Poskom Co.,
Ltd.
a : Field strengths from fixed transmitters, such as base stations for radio ( cellular/ cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To access the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the PXP-60HF is used exceeds the
applicable RF compliance level above, the PXP-60HF should be observed to verify normal operation. If abnormal
performance is observed, additional measurements may not necessary such as reorienting or relocation the PXP-
60HF.
b : Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
10
Product Manual: PXP-60HF Poskom Co.,
Ltd.
6. SYSTEM CONFIGURATION
6.1. PXP-60HF is a radiological device for medical uses. It should be applied for the radiographic
diagnosis and operated by qualified practitioners. Users have to comply with safety and
health regulations concerning the ionizing protection and the electrical and mechanical safety
of the medical devices.
6.2 PXP-60HF is composed of the following parts and devices: They should be firmly fixed and
connected with each other for the sound and normal operation.
A. Operation Board
B. Inverter
C. HV Tank including X-ray Tube
D. Collimator
E. Exposure Switch
F. Power Cord
6.3. Supplied outfit and documents
A. Aluminum Carrying Case
B. Operation Manual
C. Service Manual
D. Power Plug
E. Hand Switch (When you push hand switch in order to make radiography, it should be
pressed by two steps: step one is ready and step two is the exposure.)
11
Product Manual: PXP-60HF Poskom Co.,
Ltd.
5
3 4 13
10
8
12
11
1
2 7
9
3 Ready lamp Lighted when you push the hand switch one time for
preparation
4 Exposure lamp Alert lamp during x-ray exposure
5 Wait lamp Lighted until the second exposure started
6 kV up/down switch Select kV value by up or down buttons
10 Display reversal switch Help to read LED value of kV, mAs reversed
11 Collimator lamp on Switch for operation of the lamp inside the collimator
12 Laser pointer To adjust exposure focus
12
Product Manual: PXP-60HF Poskom Co.,
Ltd.
WARNING
Please check the input voltage and the power requirement are in accordance with
the system labels, which attached on the body of the machine. The input voltage
tolerance within + 10% is acceptable for normal operation.
CAUTION
Verify thoroughly a power cable and other cable interconnection before operating
the portable and making x-ray exposure.
CAUTION
When the portable is first used after purchase or x-ray tube is cool, x-ray tube is
needed for warming up to secure long life of the tube. Following method of pilot
test will relieve the damage on the x-ray tube. Operate the machine with low kV
values (50kV, 5mAs) three times consequently and execute with higher values
[90kV, 50mAs].
13
Product Manual: PXP-60HF Poskom Co.,
Ltd.
kV mAs
D. Select the kV values using the [6] kV up/down switch.
E. Select the mAs values using the [7] mAs up/down switch.
F. Positioning the machine at the ready condition by one time pushing of the hand switch.
G. When the [3] ready LED lighted, press the hand switch again and make exposure. During the x-
ray exposure, x-ray LED will be lighted on.
H. After the exposure is ended, [5] “wait LED” sign will be lighted on and off for the waiting time.
7.3. APR MODE
A. Please open the carrying case.
B. Connect power plug into the PXP-60HF device.
C. Turn on the line switch and the display will be shown up as follows.
kV mAs
D. Select the [8] APR switch out of 1~8.
E. Select the kV values using the [6] kV up/down switch.
F. Select the mAs values using the [7] mAs up/down switch.
G. Save the selected kV and mAs figures with pushing the [9] “store” button.
H. When the [3] ready LED sign appears, press the hand switch again and make exposure.
During the x-ray exposure, the [4] X-ray LED will be lighted on.
I. After the exposure is ended, [5] “wait LED” will be lighted on and off for the waiting time.
NOTE
As it was illustrated in the front pages, the exposure switch should be pressed by
two steps. However, you can make immediate exposure in case the filament of the
tube is hot enough.
CAUTION
If you want to make exposures several times in row, wait until the [5] wait lamp is
lighted off. If the number of consecutive exposure exceeds three times (at higher
than 50mAs), wait for 3 minutes in order to prevent overheating of the machine.
14
Product Manual: PXP-60HF Poskom Co.,
Ltd.
9.1. INTRODUCTION
Currently, mAs, kVp and Pre-set memory techniques (PROM memory) can be selected by the
operator by depressing push buttons or by scrolling displays on a control panel. Each technique or
mode selection depends on an anatomic region or view for radiographic examination. The collimator
lamp and laser pointer is turned “on” by the lamp push-button and turned “off” by a 30 second
collimator timer run off.
Many portable operators and field users have suggested we find a better way to select an exposure
technique and parameters along with mAs, kVp and pre-set PROM memory values. Most field
operators hold portable X-Ray units with one hand, while holding the exposure hand switch with the
other hand. This makes selecting technique factors on the control panel awkward without dropping the
portable unit or hand switch. We have modified our X-ray portable software and functions of the Prep.
and Exposure hand switch to allow selection of all exposure parameters (mAs & kVp) and control of
the collimator lamp and laser beam.
The exposure hand switch consists of a two step hand switch. The first (1) stage of the hand switch is
called the “Preparation (Prep. or Ready)” stage. The second (2) stage of the hand switch is called the
“Exposure (Exp or X-Ray out)“ stage. (See drawing 1.)
All functions of the remote control described above are executed by clicking the first stage on the hand
switch which is the “Prep.” mode of operation. Each remote control function and technique value
selection is described below.
(Drawing 1)
1st stage switch = Prep. Switch
2nd stage switch = Exp. Switch
15
Product Manual: PXP-60HF Poskom Co.,
Ltd.
The kVp and mAs setting, memory selection and collimator lamp and laser pointer illumination
functions are controlled remotely by the hand switch by pressing the “Prep.” stage push-button as
follows:
ATTENTION
Click the “Prep.” hand switch three (3) times consecutively. This will allow you
to enter in the “remote hand control” menu mode for operation.
<TRIPLE CLICK>
NOTE
All pre-set memory data should be entered and stored at each memory station (PXP-60HF: 1-8),
prior to performing the above mode of operation.
One (1) click of the “Prep” switch will cause the kVp LED readout to blink. You will now be in the kVp
selection menu mode. The kVp values increase by one (1) kVp unit per click of the “Prep.” switch. If
you depress and hold the prep switch for a few second, the kVp readout will increases rapidly up to
the maximum value (PXP-60HF: 100 kVp). The kVp readout will circulate from lowest to highest
values as long as you hold and press the button. When the desired kV readout appears on the control
16
Product Manual: PXP-60HF Poskom Co.,
Ltd.
display panel, release the prep hand switch and wait for a few seconds. The kVp value will set up
automatically.
ATTENTION
Click the Prep. hand switch rapidly twice. The collimator lamp and laser
pointer beam will illuminate for X-ray exposure positioning.
<DOUBLE CLICK>
With the exposure technique (mAs & kVp) set with your remote hand switch, you are ready to make a
radiographic exposure for examination. It is very important to position the right region of interest on a
patient’s anatomy. Therefore, the collimator beam must be turned on to confirm the size of region and
size of film covered. The laser beam must also be turned on for checking the precise center of the
target prior to making a final X-ray exposure. When the Prep switch is rapidly clicked twice, the
collimator lamp and laser beam will turn on simultaneously until the lamp timer (30 seconds) runs off.
However, it will turn off automatically when x-ray exposure sequence is completed.
WARNING
When an X-ray exposure is made in a room or outdoors, lower radiation will scatter throughout the
surrounding area. An authorized operator is expected to take precautions and is highly recommended
to ware a lead apron and gloves if necessary for your patient’s and your own protection.
After the x-ray exposure technique is selected and collimator light beam size is adjusted with the
collimator blade knob, an x-ray exposure can be made by an authorized individual of the medical clinic
or hospital.
18
Product Manual: PXP-60HF Poskom Co.,
Ltd.
10. SPECIFICATIONS
[Table 1. Specifications]
Output Power 3.2 kW
Input Power Voltage 200V-240V
Phase & Frequency Single / 50/60 Hz
Current Short time 18A
Radiography kV Range in 1kV steps Tube voltage current mAs
40kV ~ 50kV 60mA 0.4 ~ 100
51kV ~ 70kV 40mA 0.4 ~ 64
61kV ~ 70kV 40mA 80 ~ 100
71kV ~ 80kV 40mA 0.4 ~ 32
71kV ~ 80kV 35mA 40 ~80
81kV ~ 90kV 35mA 0.4 ~ 40
81kV ~ 90kV 30mA 50 ~ 80
91kV ~ 100kV 30mA 0.4 ~ 50
91kV ~ 100kV 25mA 64 ~ 80
MAs Range 0.4mAs – 100mAs, 25steps
Max. kV Deviation ±7 %
Max. mAs Deviation ±10 %
Indication kV (error code) /mAs: 7-segment LED
X-ray Tube Model Name D-205 TOSHIBA
19
Product Manual: PXP-60HF Poskom Co.,
Ltd.
1) PXP-60HF
20
Product Manual: PXP-60HF Poskom Co.,
Ltd.
Illustration 3. Labels
PXP-60HF Portable Label Sticker and location
1.1. If the output calibration is needed after periodical inspections or repairs, please observe following
rules and methods.
CAUTION
The ionized radiation is dangerous for the operator if the following
safety measures are not strictly observed.
1.2. This manual is designed to ensure correct use and operation of PXP-60HF. Please read all the
lines thoroughly before you use this equipment.
1.3. Incorrect use and operation exceeding described conditions in this manual may occur damage of
the machine and shorten its life. Particular attention must be paid to all the warnings, cautions and
notes incorporated herein.
1.4. This equipment should be used only by the legally qualified persons and practitioners.
1.5. Regarding the AC cable and software, only the one developed and supplied by Poskom should be
used.
1.6. No responsibility is taken by Poskom for any infringement of patients or other rights of the third
parties which many results from the use of this manual.
1.7. Always keep the manual at hand for your reference.
2. OPERATION SHEET
22
Product Manual: PXP-60HF Poskom Co.,
Ltd.
LASER SWITCH
23
Product Manual: PXP-60HF Poskom Co.,
Ltd.
3. BLOCK DIAGRAM
READY
1
PXP60HF-CTL-001
CON1 3V/35mA
2
X-RAY LASER POINTER
HAND SWITCH
3
35312-03
CONTROL BOARD CON2 1
2
* EXPORT : NORMAL
* DOMESTIC : NOT USED
OP PANEL 35312-02 LASER
1 CON6
2
& HT CONT. TANK
35312-02 T
2
CON5 CON4 35312-05
35312-11 CON3 THERMAL
35312-05 EP_FB(-)
10
11
EP(-)
1
2
3
4
5
1
2
3
4
5
6
7
8
9
1
2
3
4
5
1
AGND
GND
LAMP_CONT
FIL_CONT1
FIL_CONT2
EP_CONT1
EP_CONT2
/100V_SEL
IP_FB(+)
IP(+)
FIL_FB
AGND
+12V
DGND
-12V
+5V
10
11
1
2
3
4
5
1
2
3
4
5
6
7
8
9
35312-05 35312-11
1
2
1
2
1
2
1
2
35312
1
CON7 2
35312
PXP60HF-PWR CON10
1
2
FG
35312
1
CON8 2
35312
1
PWR & DRIVE CON9 2
CON4 T2
1 C1 3
FILAMENT DRIVE 2 2
3 +12.2
CB1 4 SHIELD GND SHIELD LINE 3
53258-04 C2
FIL XFMR
5
4 53258-05
3
2 CON1
1
COLLIMATOR DRIVECON6
250V/16A 1
250V/20A 2 COLLIMATOR
CN1 SHIELD LINE
PLUG AC MALE LAMP
53258-02 12V/50W
PXP-60HF-Main Block
CHECKED BY DATE
Size FSCM NO. DWG NO Rev
A3 A
<Cage Code>
Sheet
1 of 1
24
ProductManual: PXP-60HF Poskom Co., Ltd.
-PREPARATION
Following tools and instruments are needed.
a. Tools: Philips screwdriver No.1 & 2, Straight slot screwdriver
b. Instruments: DVM, oscilloscope (Tektronics TDS 3032 and more than Equivalent)
4. CALIBRATION PROCESS
kV mAs
-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the
oscilloscope channel 1.
-If it is not about 9V, adjust mA by VR3 on the OP board. Then observe the kV waveform of the
oscilloscope channel 2.
-If it is not 2V, adjust kV by VR2 on the OP board. After adjustment, set the tube voltage kV to 80KV
by the kV up/down switch.
-Set the kV and the mAs at 80kV and 0.4mAs(40mA).Then observe the mAs, kV waveform of the
oscilloscope channel 1, 2.
25
ProductManual: PXP-60HF Poskom Co., Ltd.
kV mAs
-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the
oscilloscope channel 1.
-If it is not 7V, adjust mA by the calibration data on the OP board.
-Turn off the line switch, and wait until the display switch the light off.
-DIP switch 1 on the PCB OP board is turned on.
-Turn on the line switch and the display indicates as follows.
-Then push the mAs up switch (see operation sheet).
The display indicates as follows.
kV mAs
Calibration code Calibration data
26
ProductManual: PXP-60HF Poskom Co., Ltd.
kV mAs
Calibration code Calibration data
kV mAs
Preheat calibration code Preheat calibration data
27
ProductManual: PXP-60HF Poskom Co., Ltd.
kV mAs
Preheat calibration code Preheat calibration data
-Adjust data by the mAs up/down switch ⑥ (on the control panel).
-After it is adjusted, push the Store switch ⑨ (on the control panel).
-DIP switch 1 on the PCB OP board is turned off then exposure X-ray.
-If overshoot or undershoot occur in the waveform, set the calibration data again.
-Table. 2 shows the calibration code and the calibration data against an each tube current.
The calibration data is possible to be changed in accordance with the tube condition.
28
ProductManual: PXP-60HF Poskom Co., Ltd.
29
ProductManual: PXP-60HF Poskom Co., Ltd.
30
Product Manual: PXP-60HF Poskom Co.,
Ltd.
7. SCHEMATICS
(2F-5)
SEG_A (2G-5)
SEG_B (2H-5)
SEG_C (2J-5)
SEG_D (2K-5)
SEG_E (2L-5)
SEG_F (2N-5)
SEG_G (2O-5)
SEG_DP
R21
15 1/4W
1 2
R22
15 1/4W
1 2
10
10
10
10
10
10
7
5
U8 R23
DP
DP
DP
DP
DP
DP
A
G
C
F
(1B-13) 1 16 15 1/4W
SEG01_OUT I1 Q1
(1B-13) 2 15 1 2
SEG02_OUT (1B-12) 3 I2 Q2 14
SEG03_OUT I3 Q3
(1B-12) 4 13
SEG04_OUT I4 Q4
(1B-12) 5 12 R24
SEG05_OUT I5 Q5
CC
CC
CC
CC
CC
CC
CC
CC
CC
CC
CC
CC
(1B-12) 6 11 15 1/4W
SEG06_OUT (1B-12) 7 I6 Q6 10 1 2
SEG07_OUT 8 I7 Q7 9
GND VCC
3
8
3
8
3
8
3
8
3
8
3
8
SEG1 uD_SR105K SEG2 uD_SR105K SEG3 uD_SR105K SEG4 uD_SR105K SEG5 uD_SR105K SEG6 uD_SR105K
TD62304 R25
15 1/4W
1 2
+5V R26
15 1/4W
1 2
LED2 READY LED3 X-RAY LED4 WAIT LED5 M1 LED6 M2 LED7 M3 LED8 M4 LED9 M5
R27 2 1 2 1 2 1 2 1 2 1 2 1 2 1 2 1
15 1/4W GREEN YELLOW RED GREEN GREEN GREEN GREEN GREEN
1 2
APPROVED BY DESIGNED BY
HTP-24 PORTABLE X-RAY
31
Product Manual: PXP-60HF Poskom Co.,
Ltd.
+5V
1
RA7
4.7K
2
3
4
5
CON6
ISO_PWR (1B-10),(7G-13)
1 /FAULT
(1B-10),(5N-13)
IP_FB_IN
2 R29
ISO_GND 270
3 1 2
2
35312-0310
D33
1N4148 PC1
1 8 (1B-10)
HS_READY
1
CON1
2 7
1
2 3 6 (1B-9)
3 HS_X-RAY
R30
4 270
35312-0410 1 2
2
4 5
D34
1N4148 TLP620-2
+5V
D35 D36
CON2
2 1 2 1
1
1N4148 1N4148
(1M-7)
2 /LASER
35312-0210
APPROVED BY DESIGNED BY
HTP-24 PORTABLE X-RAY
32
Product Manual: PXP-60HF Poskom Co.,
Ltd.
(7B-13)
IP_FB
R31 10K 1%
(1J-5)
IP_FB_ADC
R32
+12V
10K 1%
2
D37 U9A
4
LM358 U10 C21 BOX332
2
1N4148 2 - R33 39K 14 5 1 2
40mA(0 ~ 8V) R34 4.7k 1 1 3 1 +IN 1SC 10
-IN CPI
1
3 + 11
VR1 +12V COM
2
10K 12 7 (1B-10),(4J-11)
IP_FB_IN
2
+VS FOUT
8
+ D38 4 13
R35 C62 -VS VOUT
1/2W 200 1N4148 -12V VFC32
1
C22 MN104
1
C24 BOX103
6.8uF/16V tantal 1 2
CON3
IP_FB(+)
1
2
EP_FB(+) C60 39P
3 2 1
4
R75 100K 1%
5
35312-0510
(1A-10) (6B-14),(7B-11)
/HT_CHECK
EP_FB
+12V +12V C28 MN104
2 1
2
D39
8
100KV(0 ~ -5V) 1N4148 5 R36 1K
+
R37 100K 1% 7 (1J-5)
EP_FB_ADC
1
6
-
2
2
C25 C26 C27 LM358 R84
2
R
4
D40 U9B
BOX223 BOX223 BOX223 C58 MN104
1N4148
1
R76 2 1
1K
2
1
1
-12V C63
-12V
1
MN104
2
VR2 2K
1 3 R38
10K 1%
1 2
C29 MN101
2 1
TP3 C30 MN101
2 1 TP4
+12V
C31 MN104
1 2 EP_REF
1
IP_REF
1
R39
8
8
(1F-7) 1 2 3 10K 1%
EP_CONT +
1 (6B-14) (1F-7) 1 2 5
EP_REF IP_CONT
1
+
2 7 (6C-5)
IP_REF
2
1
-
C32 U11A 6
1
1
-
R41 LM358
1
4
4
10K 1% 1 2 10M MN101 R44
2
2
2 10K 1%
-12V
2
2
APPROVED BY DESIGNED BY
HTP-24 PORTABLE X-RAY
33
Product Manual: PXP-60HF Poskom Co.,
Ltd.
+12V
+12V
R45
5.6K U12
(5J-9),(7B-11) 1 2 1 12 R47 910 CON4
EP_FB +IN1 VCC
(5F-4) 1 2 1 2
EP_REF 1
R46 1 2 2 8 EP_CONT1
5.6K 2 1 1 2 -IN1 C1 9 2
1
C36 3 E1 R50 910
C34 R48 R49 F.B. 1 2
BOX102 BOX682 470K 680 EP_CONT2 3
4
2
2 1 5 11
C37 BOX102 CT C2 10
1 2 6 E2 FIL_CONT1
R51 RT FIL_CONT2 5
6.2K 1% 4 6
DT
7
15 GND 14 LAMP_CONT 7
16 -IN2 VR 13 8
+IN2 OC
(6B-7)
TL494 FIL_FB 9
10
1
1
C38 + R52 /100V_SEL 11
47K
U13A 1uF/25V TANTAL
2
14066 35312-1110
2
1 2
1
R53
13
2K
2
(1M-6)
/HV_ON
2
+5V
2
R57
-12V 10K 1%
1
U13B
U14 +12V 14066
4
8
OP07 R59 11 10
R60 2 - 5.6K U15
4.7K 6 1 2 1 12
1
(6K-11) 1 2 3 + 1 2 +IN1 VCC
2
FIL_FB 1 2 2 8
12
R80 R61 2 1 1 2 -IN1 C1 9 R58
2
E1
7
1
2
2 1 BOX473 470K 10K
1
2 1 5 11
C42 BOX102 CT C2 10
+12V 3 1 2 6 E2
1 2 2 U17 R64 RT
1
PN2222 DRV1 1PN2222 5K 1% 4
2
2
R79 5.6K R70 470K
1
1 2 2 1
TL494
(5L-4) R78 30K C61 BOX104
IP_REF
1 2
IP_FB
R77 5.6K
1
1
C43 R67
BOX473 47K
U13C
2
14066
2
4 3
1
R68
5
2K
2
APPROVED BY DESIGNED BY
(1M-6)
/PREP HTP-24 PORTABLE X-RAY
34
Product Manual: PXP-60HF Poskom Co.,
Ltd.
+12V +12V
TP5
+12V
2
C44
R69 R85 1 2
100k IP_FB
64K MN104
1
U16A
8
1
3
+
1
(5I-15) 1 2 2
IP_FB -
R71 LM393
1
100K
4
R86 C45
100k BOX473
(1B-10),(4J-11)
2
TP6 /FAULT
EP_FB
1
U16B +5V
8
5
+
7
1 2 6
EP_FB -
(5J-9),(6B-14)
R72 LM393 74HC573 27512 TD62304AP TD62304AP 14066
1
100K
4
2
2
10u/16V 56K 10u/16V 100K
1
TP7
+5V
+5V
1
C52 + C53
MN104 100uF/50V DGND HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE HOLE
2
1
2
CON5 TP9
1
2 +12V +12V
3
1
4
5 TP11
TP10
1
35312-0510
1
C54 +
100uF/50V C55
MN104 AGND AGND
2
TP12
1
C56 + C57
100uF/50V MN104
-12V
2
2
35
Product Manual: PXP-60HF Poskom Co.,
Ltd.
CAPACITOR BANK BOARD
53258-05
CON1 224/AC250V 224/AC250V TB1 TB3
SC-10-D050 330/400V 68K 1W
5 DC+ OUT 1 1
4 R7 C53 C54 C55
3 1/2W470K L2 C8
2 R8 C13 102/AC250V C9 C10 C15 C16 C11 C14
1 C12 - + + + +
102/AC250V + + + +
C17
R9 FET GATE DRIVE CIRCUIT
R12 NTC10D22 BD1
D25XB60 TB2 TB4
DC- OUT S1
68K 1W 1 1 +24V
330u/400V/30mm*40mm GND 1
BTA26-600 SIG1(+) 2
Q11 R13 F1 50T2A 3
10K 4
5
L3 6
SQE20 7
8
CON7
DRV2A G
1
S 2
S1
+24V 35312
C18 100/400V GND 1
2
D14 SIG1(-) 3
1N4937 T1 PQ32/30 4
1 5
C19 6
M104/100V + D6 1N4937 7
R14 1W100 C20 47/25V 8 CON10
2 DRV2A G
4 1
S 2
C21 S1 35312
C47 R15 222/1KV TP1 +5V 0.5A TP3 +24V
R16 2W24K GND 1
1/400V 2W 100K +5V 2
D8 S2L60 U2 LM7805CT GND 3
D7 7 1 3 4
1
TP2 1N4937 IN OUT +5V 5
GND 1 6
GND
3 C24 +12.2V 2 7
Q21 C2235 R29 + C45 + GND 3 8
CON8
27K C22 HS3 C46 100/16V -12V 4 G
DRV2A
1
2
100/16V CON2 +24V S 2
S1
C48 HS1 35312-5 +24V 35312
1
2
22/25V U7 UC3844 TP4 8 GND
CON3 2
Q12
7
8 3 D15 S2L60 + 3 6 6 8
CON9
1
4 5 7
4
D10 +12.2V
5 C35 Q14 M472/100V C37
+ + + PN2222 R27 10 Q15 +15V
+ C36 C33 C34 R23 R41 1N4746 IRF640 D12
1N4937 2200/16V(13pie) 2200/16V 1W680 47K R28 1
47/25V 2200/16V R24 +15V Q17 R30 Z6 2
U4 6 1/2W470 PN2907 1N4937 1W1K 3
D11 1N4937 78L05 12 + C40 20K 4
TG 3 1 R25 Q18 22/25V CON4
IN OUT R26 100 5.1K PN2222 R32 10 1N4746 Q19 53258-4
GND
IRF640
1
1N4937
4
TH 1
ISO_5V
2
ISO_GND R39
1%1.8K
C44 222/AC250V
3
2
1
2
1
36
Product Manual: PXP-60HF Poskom Co.,
Ltd.
37