Technical Publication

SMS-M-001 (Rev. 1.0)

Operation Manual

Model: PX-300HF

Important!
Please read this manual thoroughly and use the system correctly before using the
system. This manual desires to keep in promised place to use always again after
reading.

EcoRay Co., Ltd.


621-14, Docheon-Dong, Gwangsan-Gu, Gwangju, Korea 506-301

Tel: +82-62-952-3251 / Fax: +82-62-951-5261 / E-mail: x-ray@eco-ray.co.kr

Seoul Office: #402, Cygnus B/D, 115-12, Nonhyun-Dong, Kangnam-Ku, Seoul, Korea 135-821
Tel) +82 70 7510 3402 Fax) +82 70 8630 3420 E-mail: sales@ecoray.kr
EcoRay PX-300HF

IMPORTANT! X-RAY PROTECTIONS!

X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS SAFETY


INSTRUCTIONS ARE STRICTLY OBSERVED IN USE!

Accordingly, when use this product should be thoroughly read the "Radiation Safety Control Regulation"
and then compliance with it in use. This product is serious injury's danger if do not keep use guideline
or regulation therefore, read and understood this manual before in use.

This product was designed by safety level (SL) presenting in Radiation Safety Control Regulation.
However, if do not keep using method and Radiation Safety Control Regulation, will not be able to
avoid serious injury's danger from radiation. Therefore, to prevent accident from radiation, everyone who
work connected with radioactive emission must understand radiation safety control regulation and
understand product manual fulfill thorough education and formality.

Our company recognizes enough about danger of excessive exposure to operator, installation people,
and maintenance people. Therefore, our company strongly advises that operator, installation people,
maintenance people must be educated about radiation safety control regulation and should be used this
product when understand fluently about all knowledge of operation this product.

Also, our company informs that operator, installation people, and maintenance people have a
responsibility to be educated fluently about radiation safety control regulation and understood about
product use method.

Also, our company has no responsibility about injury or calamity that occurs does not keep use method
of this product and radiation safety control regulation.

Therefore, our company strongly advises that to be used this product after that person in charge of
installation, maintenance and operator should understood radiation safety control regulation of domestic
or foreign.

Thanks for purchasing product of EcoRay.


Before using the system, please read this manual thoroughly and use the system correctly. The
precautions and prohibitions are seen though the manual is classified as follows.

(Rev. 1.0) 1 Operation Manual


EcoRay PX-300HF

NOTICE AND CONTRAINDICATION!

States a direct danger that may cause death or serious injury if it is not avoided.

States an indirect or potential danger that may cause death or serious injury if it is not avoided.

State a danger that may cause slight or medium injury or may cause damage in equipment or fire
if it is not avoided.

NOTE ☞

States the information, which helps to use the system correctly.

CONTRAINDICATION!

This x-ray system is not suitable for therapeutic applications.

(Rev. 1.0) 2 Operation Manual


EcoRay PX-300HF

Precautions in the use of X-ray system


(For the Safety and Prevention of Danger)

1. Only one experienced technician should operate the equipment.


2. When installing the equipment, pay attention to the following items.
2.1 Do not install it near water faucet or similar equipment.
2.2 Install it away from potential sources of problems such as abnormal pressure, temperature or
humidity, drafts, direct sunlight, dust, chlorine or sulphur gas.
2.3 During transportation and operation of the equipment, avoid tilting, vibration and sharp impact
against it.
2.4 Keep the equipment away from the areas where chemicals or gases are stored.
2.5 Use only the correct electrical power source with matching frequency, voltage and current (or
wattage).
2.6 Check the condition of the battery power source (Power Supply Unit). (Power and polarity before
operating the equipment.)
3. Before operating the equipment, pay attention to the following items:
3.1 Check the conditions of switch contacts, polarity, and make sure the equipment performs correctly.
3.2 Confirm that the ground is connected properly.
3.3 Check all wiring for proper and correct connections.
3.4 Check the condition of the external electric circuit breaker.
3.5 Circuit breaker will be directly connected to the system and do not connect with another system.
3.6 When the equipment has been leaved for a long time (more than 10days), You must warm-up
equipment and X-ray tube sufficiently before use.
4. While operating the equipment, pay attention to the following items:
4.1 Do not ever-exceed time or the amount of equipment use needed for diagnosis.
4.2 Observe the equipment and the patient continuously for detection of problems.
4.3 When a problem is detected with the equipment or the patient, take proper action to stop the
equipment without harming the patient.
4.4 Do not let the equipment touched by the patient.
5. After operating the equipment, pay attention to the following items:
5.1 Turn of the switch and then, turn off the main power switch.
5.2 Clean the part that contacted with body.
6. When the equipment is out of order, call a certificated engineer. (Do not try to repair it.)
7. Do not modify any part of the equipment.
8. The equipment must be checked periodically for good operation condition without troubles.
9. Do not use x-ray system connecting with other system.

(Rev. 1.0) 3 Operation Manual


EcoRay PX-300HF

1. The owner/user has a responsibility for management and maintenance of this equipment.
2. This equipment must be operated by only authorized personnel.
3. Contact authorized service engineer when repair or inspection of the inside of the equipment.
4. In general, modifications are strictly prohibited by the Regulations. Please contact our agency if it is
needed to modify the device.
5. Do not use this system in conjunction with the other machines because seriously danger could be
happened unexpectedly.

Periodic Inspection

1. Maintenance is required to maintain safety and performance of this system for a long time.
2. Maintenance must be done periodically with maintenance plan.
3. This system must be inspected from the inspection office per every 3 years.

Warranty

1. This warranty shall apply only to defects that arise within 12months of purchase.
2. This product could have a free of charge service if troubles are occurred during the warranty period.
3. The warranty does not cover defects caused by abuse, mishandling, accidents or improper
installation.
4. The following will not be covered under the warranty.
4-1. Modifications to the product, or repairs made by unauthorized person but manufacturer or
authorized person by manufacturer.
4-2. Acts of God. Including fire, flood, storm, power or voltage surges.
4-3. Improper use or installation.
4-4. Defects that arise in conjunction with the use of other machines.
4-5. Use of improper parts or modification the circuit.
4-6. Improper use
5. In case of repaired during the warranted period, User shall advise the result of test to manufacture.

(Rev. 1.0) 4 Operation Manual


EcoRay PX-300HF

Precautions in Use

When using this equipment, please observe the following precautions for the safety of the operator
and the patient.

1. Improper use of the X-ray equipment might cause the operator or the patient to be accidentally
exposed to X-ray radiation.
2. During X-ray radiation, any person other than the subject the patient should not stay in the room.
3. X-ray room must have radiation protection facility by "Radiation Safety Control Regulation"

1. Only qualified personnel should handle the equipment


2. Electric shock or leakage of electricity can happen unless ground is not installed necessarily.
3. Keep the equipment dry
4. Otherwise it may cause an electric shock to the operator of the patient and if any liquid or water
flew into electric device turn off the power and contact company or service agency.

1. This system is made radiation and must shield completely X-ray room with lead shield. After that,
must be inspected by authorized inspection office and then have to use system.
2. In the radiation exposure, operator have to adjust exposure area with suitable exposure point and
try to expose it with not over exposure.
3. The most effective method to secure from first and second radiation defilades perfectly and
should be established x-ray room by lead etc... And, to minimize radiation exposure from radiation
exposure to operator, lead partition, lead gloves, apron, cover, Gonad Shields, Thyroid collar, Ovary
Shields must be used.

NOTE ☞

For the safety of operator and patient, opening the X-ray generator cover, X-ray tube,
Collimator and other equipment cover are strictly prohibited. Only qualified engineer could open
the cover.

(Rev. 1.0) 5 Operation Manual


EcoRay PX-300HF

Scope

1. This manual contains information of system operation, components, control keys, switches and
hazards and technical reference information.
2. This manual may not be reproduced in whole or in part without the written permission of ECORAY
CO., LTD.
3. The text of this manual is a translation based on the Korean alphabet (Hangŭl) original.

© September, 2007
EcoRay Co., Ltd.
621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea

Revision History

Revision No. Revision Date Change Description

Rev. 1.0 2008-03-10 Release version

(Rev. 1.0) 6 Operation Manual


EcoRay PX-300HF

Abbreviation

kV Kilo-voltage
mA milliampere
mAs milli-ampere Second mA*Sec
mS milli-second 1/1000 second
APR Automatic Programming Radiography
AEC Automatic Exposure Control

Summary of Functions

Symbol Function Description

Switching OFF

Switching ON

When yellow lamp ON, generator is


standby condition for operation

X-ray Exposure Switch

When x-ray exposure, lamp ON

Error display

Reset switch

Filament Focal indication lamp

(Rev. 1.0) 7 Operation Manual


EcoRay PX-300HF

Table of Contents

Chapter 1
1. Introduction and Safety ……………………….……...…………………………………………………………………… 11
1.1 An Overview …………………………..……..……………………………………………………………………………………………………. 11
1.2 Safety Information ………………………………………………………………………………………………..……………………………. 11
1.3 Electrical Safety ………...……………………………………………………………………..…………………………………….…………… 12
1.4 Protection from Radiation …………………………………………………………………….…………………….…………..…………. 13
1.5 Maintenance ………………………….………….………….……………………………………………..………….……………….…………. 14
1.6 Responsibilities ……………..……………………………………………..………………………………………………………….…………. 15
1.7 System Error ………………………….………………………………………………………………………………………………...…………. 16
1.8 Non-Approved Modifications ……………….…………………………………………………………………………………………… 16
1.9 External Devices ………………………………………....………………………………….………………………………..…………………. 16
1.10 Use According to Regulations ……………………...…………………………………………………………………….……………. 17
1.11 Contraindication ………………………………….…...……...……………………………………..………………………….….…………. 17
1.12 User profile ……………………………….…………………………………………………………………..…………………………….……. 17
1.13 Training ……....…………………….……………………………………………………………………………….………………….…….……. 17
1.14 Directive for User ……………………………………………………….……………………………………………………………….……. 18
1.15 Specification ……………………………..………………………………………………..……………………………………………….……. 18
1.16 Option …………………………………………………..……………………………………………………………………………………….…. 18
1.17 Labels ……….……………………..………………...………………………………………………………….…………………………….……. 19
1.18 Certifications ………..…………………………….………………..……………………….……………………………….…………….……. 20
1.19 Using environment condition ………………………………………………………...…………………..…………………..….……. 20
1.20 Construction and Function ………………………………………………………….…………………………………………..………. 21
1.21 General details ………………………………..……………………...………………………………………………………………..………. 22
1.22 Service Numbers ……..……………………………………………………...……………………...………………………………..………. 22
1.23 Disuse of Product …………………………………..………………………...……………………….……………………………..………. 22
1.24 EMERGENCY STOP ……………………………..………………………...………...………………………………………………..………. 23
1.25 Powering OFF the System ………………………………...……………………………………………………………………..………. 24
1.26 Potential Hazards ……………………………………………………….……….……………………………………………………………. 25

Chapter 2
2. Introduction and Using Method ……………………………………………………..…………………………………. 31
2.1 Introduction ………………….……………………………..………………………………………………………………….…………………. 31
2.2 How to use Generator (SIG-30-CON) ………………….…..…………………………….…………………………………………. 32
2.3 How to use Tube Stand (SMS-TS-3) ………….……………………….…………………………………………………………….. 42

(Rev. 1.0) 8 Operation Manual


EcoRay PX-300HF

Chapter 3
3. X-ray room Requirements ………………………………………………………………………………..………………. 50
3.1 Environmental Requirements ……………..………………………………………………………………………………….……….…. 50
3.2 Electrical requirements …………………..…….……………………………………………………………………………….….…….…. 50

Chapter 4
4. Maintenance ……………………………………..………….…………………………………………………………………. 54
4.1 An Overview ……………..…………………………………………………………………………………………………………………….…. 54
4.2 Periodic Maintenance Procedures ……………..……………………………………………….………………………………….…. 54
4.3 Cable check ……………..……………………………………………………………………………..…..………………………….……….…. 55
4.4 Console check ……………..………………………………………………………………………………………………………………….…. 55
4.5 High voltage transformer check ……………..………………………………………………………….………………………….…. 55
4.6 X-Ray Tube check ……………..……..……………………………………………………………………………………………….…….…. 55
4.7 Periodic Maintenance check point …………………………………………………………………….……………..………………. 56

(Rev. 1.0) 9 Operation Manual


EcoRay PX-300HF

Chapter 1

Introduction and Safety

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 10 Operation Manual


EcoRay PX-300HF

1. Introduction and Safety

1.1 An Overview

This manual contains operation instructions for PX-300HF and also product description, safety and
operation method.
This X-ray system was designed for diagnosis’s human beings and could get good regulations by
controlled digital value by microprocessor.
This X-ray system has a self-diagnosing function and then could improve service quality, save time
and decrease customers complain.

1.2 Safety Information

1.2.1 General Safety Information


1. This manual provides guideline to operate and use the system safely.
2. The owner of this x-ray system has to get a using permission of authorized government office.
3. The owner of this x-ray system has a responsible to get a using permission from government office.
4. The owner of this x-ray system is responsible for verifying continued compliance with all applicable
regulations and standards and has to keep specific requirements or regulations that each country
made rules
5. All users of this x-ray system must know about potential hazards may occur when using.
6. All users of this x-ray system are responsible and understanding about continued compliance with
all applicable regulations and standards and safety, emergency instructions.
7. All users must know how potential hazards can be avoided and must also know how they can
protect.
8. EcoRay confirms that PX-300HF is manufactured in accordance with the appropriate legal
regulations.
9. EcoRay assumes no responsibility or liability for after-sale operating and safety practices. (Nor can
it be responsible for personal injury or damage resulting from misuse of its systems)

1.2.2 Safety Information about X-ray exposure


1. It is very danger when x-ray over exposed to body.
2. Patient and user must observe the related radiation regulation and especially it will be more
dangerous when radiation is exposed around genitals.
3. This x-ray system must be operated by only authorized operator.
4. Radiation shields is must necessary for protection of patient and user when x-ray exposed and
radiation shields must be equipped by related regulations.

(Rev. 1.0) 11 Operation Manual


EcoRay PX-300HF

5. If X-ray was exposed to body, could be seriously injured and be careful do not exposed first,
second X-ray radiation.
6. Specially, radiant ray can be accumulated in body and can be continued from several months to
several years. Therefore, safest method for operator avoids for the first, second beam.
7. Also, because the second radiate beam can be stronger than reach to film, must take safeguard by
this.
8. The most effective method to secure from first and second radiation defilades perfectly and should
be established x-ray room by lead etc... And, to minimize radiation exposure from radiation
exposure to operator, lead partition, lead gloves, apron, cover, Gonad Shields, Thyroid collar, Ovary
Shields must be used.

1.3 Electrical Safety

Electrical circuits inside the equipment may use voltages which are capable of causing serious
injury or death from electric shock. To avoid this hazard, never remove any of the cabinet
covers.

Observe the following safety procedures to avoid electric shock or serious injury to operators and
patients and to avoid system malfunction.

1. Do not remove any of the assembly covers.


2. Do not touch all electric circuit and components to be connected this x-ray system.
3. Do not bypass, jumper or otherwise disable the safety interlocks.
4. Do not place food or beverage containers on any part of the equipment. If spilled they can cause
short circuits and power off immediately.
5. Always remove power to the equipment before cleaning.
6. When cleaning the cover, use a slightly damp cloth or sponge for cleaning.
7. When cleaning the inside of system, remove the cover then remove the dust with duster. Only
qualified service engineer could clean the inside of system. (It is danger when touched electric
parts)
8. Only qualified service engineers are allowed to service or repair a system.

(Rev. 1.0) 12 Operation Manual


EcoRay PX-300HF

This X-ray system is not rated for water-tight operation. Water, soap or other liquids which
get inside the may cause short-circuits, electrical shocks or fire. If liquid should enter the
system, do not connect the unit to the power supply. If liquids drip into the equipment,
disconnect the power cord and do not operate the system until it can be cleaned and
inspected by a qualified service engineer (EcoRay engineer or Representative’s engineer).

1.3.1 Electrical Safety


1. This x-ray system was classified with “Class 1, Type B” and was manufactured to be stand with
safety level or characteristics mentioned with maximum leakage current.

1.3.2 Electric danger in product fire


1. In case of fire caused by electricity
→ The owner/operator should prepare a plan for the area in which the unit is used, containing
the following safety measures:
→ If fire is occurred by electric, must do it as follows.
● Switching off the unit and removing the main power supply cable
● Evacuation of all persons in the vicinity
● Call for assistance
● In an emergency, use only a fire extinguisher which is suitable for putting out electrical fires.

1.4 Protection from Radiation

1.4.1 Radiation Exposures Safety


1. The owner/operator must read and understood relative regulation before use.
2. After turn on the power, if push console switch or hand switch, X-ray will be exposed and
especially, must handle with care to prevent unintended X-ray exposure.
3. Use X-ray protection cloth, X-ray protection plate before X-ray exposure.
4. The owner/operator must install the lead shield facility at the X-ray room and have to check and
it is necessary for patient and operator’s safety.
5. The owner/operator has a responsibility for radiation monitoring.
6. The owner must have the periodic inspection about X-ray system by qualified expert or company.
7. X-ray radiation protection report must be made before using X-ray system and this report must
confirm the technical requirements and relative regulations are correct matched.

(Rev. 1.0) 13 Operation Manual


EcoRay PX-300HF

(Also, must complied with each state or local regulations)


8. If X-ray is exposed to body, it cause of serious injury and then especially take care do not
exposed to first, second X-ray beam.
9. Especially, X-ray will be accumulated to the body and accumulated X-ray maintain during few
months or few years. It is the best safety and certain way to avoid the X-ray exposures is avoid
first, second X-ray beam.
10. Also, second X-ray beam is stronger than first X-ray beam to be reached the film and especially
take of this beam. And the owner or operator must be prepared this safety action.
11. Best protection way from the X-ray first beam and second X-ray beam is to cover the X-ray
room with lead shield after that have to use X-ray system.

1.4.2 Maximum permissible Dose (MPD)


1. For occupational exposure, the ICRP (International commission on radiological protection)
recommends that the MPD accumulated.

Shall not exceed of the following limits:


(1) Annual effective does: 50 mSv
(2) Annual dose for the lens of the eye: 150 mSv
(3) Annual does for all other areas: 500 mSv
(e.g. red bone marrow. breast. lung. gonads and extremities)

1.4.3 Monitoring of personnel


1. The operator must do not use it with over exposure condition and must monitored continuously of
operator’s exposed.
2. The most effective method is the using the radiation leakage meter determine whether or not
existence the MPD.
3. These measurements shall be taken at all locations where the operator or any portion of his body
may be exposed. Exposure must never exceed the accepted tolerable dose of ICRP.

The operator / patient will be seriously body injured or death by potential radiation hazards
used x-ray system without maintenance plan, using procedure, operation method.

1.5 Maintenance

1.5.1 Maintenance must be done by qualified engineer by manufacturer.


1.5.2 This X-ray system must be maintained periodically by local or state regulations.

(Rev. 1.0) 14 Operation Manual


EcoRay PX-300HF

1.6 Responsibilities

1.6.1 Ensure that all authorized person to use the equipment are aware of the danger of excessive
exposure to X-Ray radiation.
1.6.2 The responsibility herein described that is well acquainted with sufficient training of the
radiation safety control regulation and instruction of the equipment by definitely and accurately.
1.6.3 The manufacturer is not liable for injury and damage which may result from violation of the
radiation safety control regulation and instruction of the equipment.
1.6.4 The owner/operator must confirm that X-ray room must comply with relative regulations.
1.6.5 The owner/operator must ask relative regulations to be used correctly complied with regulations
to government office or independent institute. And the owner/operator has a responsibility
about this.
1.6.6 This X-ray system must be used in the suitable X-ray room to be complied with relative
regulations and technical requirements. It is the responsibility of owner/operator to keep this
rule
1.6.7. The owner of this system has a responsibility to control strictly exposed radiation of operator
and must order to operator to be used “Pocket Dose meter”.

The owner/operator must know that it is very dangerous if radiation over exposed to patient.

The manufacturers don’t have a responsibility that if seriously bodily injured or accident
occurred by over X-ray exposed.

It is a responsibility of operator/owner when accident occurred during operates X-ray system.


The manufacturer doesn’t have a responsibility about that case. Also, if X-ray system is troubled
by unauthorized amendment or adjustment, the manufacturer doesn’t have a responsibility.

(Rev. 1.0) 15 Operation Manual


EcoRay PX-300HF

1.7 System Error

1.7.1 This X-ray system has a self-diagnosing function and could be checked automatically and if
system has an error, call service or push the reset button and then initialize the main CPU data.
1.7.2 After Reset, when displayed error continuously, it is meaning of system error.
1.7.3 In this case, do not uses X-ray system before service engineer check the X-ray system.
1. Power off the console power switch.
2. Powering off the main circuit breaker.
3. Contact authorized service engineer immediately.
4. Do not use X-ray system before service engineer checked the X-ray system completely.
5. Refer to Generator service manual for more error information.

1.8 Non-Approved Modifications

1.8.1 Unauthorized changes or modifications to any parts of the system can have dangerous
consequences and changes or modifications should be done after permission of EcoRay Co., Ltd.
1.8.2 Do not use with changes or modifications without permission of manufacturer.

1.9 External Devices

1.9.1 Do not connect non-authorized devices to this system.


1.9.2 Additional equipment or external devices should only be connected by an authorized service
engineer of manufacturer and all additional equipment connected to external interfaces of
this system.
1.9.3 External interfaces of the unit must fulfill the requirements of Standard IEC 60601-1-1.

If the system is operated together with other electrical devices from a multiple socket outlet,
current leakage can increase; the operator/user in this case is responsible for ensuring that the
requirements of IEC 60601-1-1/IEC 60601-1 are still met.

(Rev. 1.0) 16 Operation Manual


EcoRay PX-300HF

1.10 Use According to Regulations

1.10.1 This X-ray system must be used by regulations after reading and understanding relative
radiation regulations.
1.10.2 EcoRay’s PX-300HF system is intended for use on human beings.
1.10.3 Regulations on its use cover the area of only diagnostic radiography applications.
1.10.4 It is suitable for daily use in a clinical or hospital environment by various users.

1.11 Contraindication

This x-ray system is not suitable for therapeutic applications.

1.12 User profile

This x-ray system was designed suitable for diagnosis’s human beings and must be used by
qualified personnel for using this system.

1.13 Training

After installation, EcoRay engineer or EcoRay’s agent will carry our training of its use and if you
have any further questions on the system, please contact address as shown in this manual.

Obey the instructions of operation manual for both patient and operator’s continuously safety
and read and understood the operation manual and keep the manual in accessible place.

Radiation Safety plan must be established and advice strongly understand and observe the
radiation protection regulations

(Rev. 1.0) 17 Operation Manual


EcoRay PX-300HF

1.14 Directive for User

1.14.1 The user, physician must not touch the patient and the interface connectors simultaneously.
1.14.2 Communication must be easily with patient in X-ray exposure.

1.15 Specification

Item Functions PX-300HF

Generator Type High Frequency Inverter

Maximum kVp 40~125kV ±5%

Maximum mA 50~300mA ±5%

1mSec ~ 6Sec
Exposure Time
(Below: 100mSec: ±1mSec Over 100mSec: ±10%)
Generator
APR 288 (Standard)

AEC Optional item

Output Frequency 40kHz

Maximum Power 32kW

Line Voltage Single phase, 230V. 50/60Hz

Rating 100kV / 300mA, 125kV / 200mA at 100mSec


High Voltage
Maximum kVp 125kVp
Transformer
Maximum mA 300mA

1.16 Option

1.16.1 General ionization could be used at AEC function.

(Rev. 1.0) 18 Operation Manual


EcoRay PX-300HF

1.17 Labels

1.17.1 Main parts of generator has a labels contains information as follows.


1.17.2 Labeling: Product Name, Model, Main power, manufactured date, Manufacturer & Address,
Serial No. are printed and attached at the generator.

Tube Stand
Generator

Note!

Other labels are located in inside of Mobile Cart.

(Rev. 1.0) 19 Operation Manual


EcoRay PX-300HF

1.18 Certifications

1.18.1 EcoRay have certificate as below about production for this product.
1. ISO 9001
2. ISO 13485

1.19 Using environment condition

1.19.1 Transportation and Installation, storage condition:


1. An ambient temperature: +10°C ~ +60°C
2. Humidity: 10% ~ 90% (Must be without condensation)
3. Pressure gauge: 700hPa ~ 1,060hPa

1.19.2 Working condition:


1. An ambient temperature: +10°C ~ +35°C
2. Humidity: 30% ~ 65% (Must be without condensation)
3. Pressure gauge: 700hPa ~ 1,060hPa
4. This system must be used in the without corrosion and explosion gas.

Troubles are could be occurred in the use or storage under unsuitable environment condition.

(Rev. 1.0) 20 Operation Manual


EcoRay PX-300HF

1.20 Construction and Function

1.20.1 Construction

1.20.2 Function
1. This X-ray system consists of Tube Stand which support X-ray tube and moveable, Console, X-ray
generator and X-ray tube unit.
2. Tube Stand: Consist of main power supply unit, electric circuit, handle bar and mechanical devices
and assembled with mobile cart.
3. Console: Consist of each switch and could adjust exposure values and display it.
4. X-ray generator: Consists of control unit and high voltage transformer unit and control x-ray
exposure values and make high voltages by console.
5. X-ray tube unit: Consist of X-ray tube, Collimator and Handle bar and could move Tube Stand
and adjust x-ray exposure area.
6. X-ray radiation will be made by pushing of X-ray switch of console or pushing of hand switch.
7. X-ray radiation will permeate patient from x-ray tube to bucky device and after permeation, could
archive x-ray images by x-ray film or digital detector.

(Rev. 1.0) 21 Operation Manual


EcoRay PX-300HF

1.21 General details

1.21.1 Manufacturer
EcoRay Co., Ltd.

1.21.2 Place of Manufacture of the / PX-300HF


EcoRay Co., Ltd.
621-14, Docheon-dong, Gwangsan-gu, 506-301, Korea

1.21.3 Representative in Europe Union

1.21.4 Information
Please contact if have any query about product or safety.
EcoRay Co., Ltd.
621-14, Docheon-dong, Gwangsan-gu, 506-301, Korea
Tel: +82-62-952-3251
Fax: +82-62-951-5261

1.22 Service Numbers

1.22.1 If the system does not work reliably, or if errors occur in terms of control functions which are
described in the manual, the owner should contact the nearest EcoRay Service Center.
1.22.2 The EcoRay Service Center can be reached at the following Phone and Fax numbers:
1.22.3 Service for Europe
1.22.4 Service for Outside of Europe, Worldwide
Tel) +82 70 7510 3402 Fax) +82 70 8630 3420 E-mail: sales@ecoray.kr

1.23 Disuse of Product

1.23.1 Please observed the each national waste treatment regulations when disuse system or parts

and must be observed the.

1.23.2 It cause of environmental pollution if discard it without permission.

(Rev. 1.0) 22 Operation Manual


EcoRay PX-300HF

1.24 EMERGENCY STOP

1.24.1 EMERGENCY STOP switch

1. Emergency stop switch is located on the upper side of generator and


upper side of handle bar.
2. It serves to switch off the system in a dangerous situation as quickly as
possible.
3. If pressed the EMERGENCY STOP switch, disable system to prevent any
further X-ray radiation.

EMERGENCY STOP EMERGENCY STOP

Handle bar of Tube Stand


Upper side of Generator

If pressed the EMERGENCY STOP switch, the unit is immediately cut off from the main
voltage. Only use the EMERGENCY STOP switch in real emergencies critical situations. Do not
use the EMERGENCY STOP switch to power off the X-ray system during normal operation. By
using the EMERGENCY STOP switch, current information cannot be saved.

1.24.2 EMERGENCY STOP Function Test


1. To ensure that the EMERGENCY STOP switch is working properly, perform this test before use.
2. Working procedure of EMERGENCY STOP switch.

1. Press the EMERGENCY STOP switch: cut off the main power
2. Released the Switch (Turn switch clockwise): Power ON
3. Start Test: Press EMERGENCY STOP switch, then push the Power ON switch of console
with over 3 times.
4. In Normal: Not supplied main power even if push the Power ON switch of console.
5. In case of power off with EMERGENCY STOP switch: Any lamp of console is not light up.
6. In case of power off with Console Power Off switch: The green LED in the Standby icon
light up to yellow. The green LED in the Power Off button must off.

(Rev. 1.0) 23 Operation Manual


EcoRay PX-300HF

1.25 Powering OFF the System

1.25.1 Powering OFF Normally


1. Normally the unit should be powered off after use with power off switch of console.

1.25.2 Complete separation from the power supply


1. If the unit has been powered off, but there are still indications that parts of the system are still
powered on, off the circuit breaker switch if want to completely powered off.
2. In spite of the power off with circuit breaker, power supplied to system, call EcoRay Service
engineer and do not use the system until it has been inspected by service engineer.

If the unit has been powered off, but there are still indications that parts of the system are
still powered on, off the circuit breaker switch if want to completely powered off.

1.25.3 Procedure when Powering OFF

1. Powering OFF Normally (Power On / Off of Console)

● Check Standby lamp condition of Console: Yellow lamp lighting

● Push the Power ON of Console: Standby lamp on and off with yellow and green

during 20 seconds, power supplied when green lamp lights up.

● Push the Power OFF of Console: When push the Power OFF switch, green lamp light
up on the Power OFF, at the same time Standby lamp light up with yellow.

2. Complete Separation from the Power Supply (Inside of Generator. PSU breaker)

● Switch OFF PSU breaker the circuit breaker inside of Generator.


● Standby lamp condition of Console: OFF condition
OFF
● Power OFF condition of Console: OFF condition
● In spite of the power off with circuit breaker, power supplied to system, call
EcoRay Service engineer and do not use the system until it has been
inspected by service engineer.

(Rev. 1.0) 24 Operation Manual


EcoRay PX-300HF

3. Procedure when Emergency Stop

● When emergency situation occurred during operation, push the


EMERGENCY STOP button on the upper side of generator or upper side of
handle bar.
● Standby lamp condition of Console: OFF Condition
● Power ON lamp condition of Console: OFF Condition
● To recover the system, turn the EMERGENCY STOP button with clockwise.

1.26 Potential Hazards

1.26.1 Medical electrical units and X-ray systems represent potential sources of danger.

1.26.2 User must know how they can detect the conditions for possible hazards, and how they can
appropriately protect themselves and others from injury.

1.26.3 The operator must take account of possible dangers and make plans to prevent them from
occurring.

1.26.4 The emergency procedure and also the operational safety procedure for the system, as well
as the safety, emergency and operational procedures of the institution must be observed.

1.26.5 PX-300HF system has no source of danger unusual for this kind, and meets all the safety
regulations for this type of medical electrical equipment.

(Rev. 1.0) 25 Operation Manual


EcoRay PX-300HF

1. Overheating, Explosion and Crash Hazard


Hazards Cause Precaution
Excessive use of X-ray tube causing it The temperature of the X-ray tube is
to overheat, can result in a rupture of continuously monitored and if certain
the pressure compensating bladder in temperature levels are reached, the anode
X-ray tube
the housing. This can cause hot oil to current is limited or radiation is totally
escape, potentially resulting in scalds disabled.
or burns.
Draping the X-ray generator may 1. Never install x-ray generator near the
inhibit airflow, resulting in cooling patient.
inefficiencies and may result in 2. Never place the x-ray generator against
X-ray Generator
reaching electric parts heat sooner and the patient’s skin.
if electric parts are heated over, it 3. Never put anything on the x-ray
could be exploded. generator.
Flammable and explosive anesthetics, 1. If any of these products are near the
liquids, vapors and gases can cause an system, DO NOT SWITCHES IT ON.
explosion when used around electrical 2. If any of these products are discovered
equipment. near the system while it is on, DO NOT
TOUCH ANY CONTROL OR SWITCH THE
Explosion hazard SYSTEM OFF.
3. Evacuate the facility immediately.
4. Have the facility specialists remove the
substances and ventilate the area before
having patients and personnel return to
the facility.
1. If improperly moved or use it 1. When moving Tube stand or patient
improperly, it could collide with table, watch around to prevent Crash.
other system. 2. Do not install the system not level with
floor.
3. Operator asks patient doesn’t move
Crash Hazard 2. If electronic magnetic lock system is while using electronic magnetic lock
trouble during use, Tube stand may fall system.
down and cause injury to the patient 4. After fixing exposure position, ensure
or medical personnel. the electronic magnetic lock system to be
locked.

(Rev. 1.0) 26 Operation Manual


EcoRay PX-300HF

2. Mechanical Movement Hazard

Hazards Cause Precaution


Tube stand and Patient table are used When move Tube stand and Patient table,
electronic magnetic lock system and it have to look around continuously to
Mechanical cause of patient of operator’s injury if prevent Crash with patient or things.
Movement Hazard use it with misuse or malfunction.

If used with another equipments, its 1. Do not use equipments not mentioned
cause of injury by unexpected danger in this manual or not mentioned in the
situation. EcoRay’s printed matters.
Equipment Safety 2. This system must be installed by
authorized personnel by EcoRay.

3. Radiation Safety

Hazards Cause Precaution


1. If this X-ray system is used without 1. It is the responsibility of the owner to
observe the relative regulations, it ensure that all personnel wear protective
cause of death or seriously injury. X-ray clothing and use radiation safety
2. Observe the relative regulations, precautions when operating the system.
make maintenance plan, operating 2. Operators should avoid placing their
instructions are observed. If not, it has hands or any part of their bodies in the
a possible danger to patient and path of the X-ray beam.
operator. 3. It is the responsibility of the owner to
ensure that owner or operator must
Radiation Safety understand with relative regulations.
3. This X-ray system makes an audio
warning or audio signal to operator when
radiation exposed.
Audio warning 1: Sounds when pushs
switch.
Audio warning 2: Sounds when radiation is
activated. (During radiation time)
Audio warning 3: Sounds when error
occurred.

(Rev. 1.0) 27 Operation Manual


EcoRay PX-300HF

4. Electrical Safety

Hazards Cause Precaution


Electrical circuits inside the system can Safety covers have been designed for this
cause serious injury or death. system to mitigate the hazards caused by
If repaired with unauthorized parts, it electrical circuits and high voltages. Never
cause of operator or patient’s serious remove the covers or operate the system
injury or death. while the covers are off. If any cover is off,
This system operates on high voltages. request service technician before using the
High voltages are present throughout systems.
the system. When repairing, must be repaired by
High voltage transformer and X-ray authorized engineer and never modify
Electrical Safety tube: Max. 150,000V circuit.
X-ray Generator: Max. 350 V DC Don’t touch the electric parts thoughtlessly.

WARNING: Removing the power cable


from the wall outlet is the only way to
eliminate the power supply voltage from
the entire system.

WARNING: Never any adjust the X-ray This x-ray tube is produced suitable for
generator and the X-ray tube. If adjust regulations. Never make any changes inside
against this warning, it makes cause of of circuit or parts. If needed change,
Unauthorized X-ray
serious injury of death. contact the EcoRay Service center before
tube change
change or modify it.

Fire from near or other equipments, it 1. Install X-ray system in good air flow
makes a same damage as like this condition.
product is fired. 2. If there is a fire in the area or system in
on fire, to do as follows.
Electrical Fire z Switch the circuit breaker off, stop supply
electric power.
z Escape all people near the system.
z Ask assistance.
z Put the fire out with fire extinguisher.

(Rev. 1.0) 28 Operation Manual


EcoRay PX-300HF

5. Handle with Care

Hazards Cause Precaution


X-ray switch is located on the console. 1. Be aware of X-ray switches of console
All of console switches are membrane are broken by using and operator has to
switch. And do not push switches with pay attention when using X-ray switches.
sharp things or wetted things. These
using is cause of trouble or 2. Hand switch works as like X-ray switch,
malfunction. take of using hand switch.
3. To install holder to prevent falling of
If hand switch is fallen or stepped, it hand switch.
X-ray Switch Safety cause of damaged or malfunction. 4. After using hand switch, return it to the
holder.
5. Be careful lest hand switch cable not
press or cut. It cause of system trouble.

Almost any liquid spilled into electronic 1. This system is not water proof, it is very
equipment automatically becomes a dander if liquid is spilled on it.
conductor of electricity. This can cause 2. Do not put liquid on system.
short circuit or fire by this short circuit. 3. It must be switched off immediately if
liquid is spilled on it.
4. Do not switch the system on until the
liquid has dried completely.
5. Do not splash or spray cleaning
solution on the system.
Cleaning Safety
6. Clean the system by unplugging the
power cord from the wall outlet and using
a damp cloth or sponge to wipe the
system clean.
7. When cleaning the inside of system,
using duster after remove the cover. This
cleaning must be done by an authorized
service engineer. (Because, it is very
danger when mishandle.)

(Rev. 1.0) 29 Operation Manual


EcoRay PX-300HF

Chapter 2

Introduction and Using Method

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 30 Operation Manual


EcoRay PX-300HF

2. Introduction and Using Method

2.1 Introduction

2.1.1 An Overview
1. This mobile manual contains of operation, safety and specification of this X-ray system.
2. This X-ray system was developed by high frequency inverter circuit base.
3. This X-ray system was designed for suitable for diagnosing human being with radiation.
4. This X-ray system was designed to control radiation with digital method and calibrated radiation
before delivery could checkable by radiation meter.
5. This X-ray system was designed to do self diagnostic by electric circuit and program and
therefore, it’s easy to check the system’s condition and it indicates the type of trouble on the
display.

2.1.2 Special Features


1. This X-ray system use high frequency inverter type system.
2. This X-ray system have APR (Anatomical Programming Radiographic) and AEC (Automatic
Exposure control – Optional item).
3. Maximum voltage is 125kVp.
4. Maximum power consumption is 32kW.
5. Maximum output frequency is 40kHz.
6. This X-ray system has an APR function in basic and it could save exposure condition in advance
and could use it easily when exposure.
7. APR function is changeable by operator’s need.
8. This X-ray system has a self diagnosing function by electric circuit and program and could check
system condition when trouble occurred.
9. This X-ray system has a protection function that limit operation when abnormal voltage supplied
to system.
10. This X-ray system has an X-ray tube stator’s protection function that it checks tube stator
condition by electrical sensing and if tube stator condition is abnormal, limits x-ray exposure.
11. This X-ray system has an X-ray tube Filament’s protection function that it checks tube filament
current condition by electrical sensing and it tube filament condition is abnormal, limits x-ray
exposure.
12. This X-ray system has a High voltage Feed Back circuit to check high voltage condition and
could archive good exposure dose reproducibility.
13. This X-ray system has a circuit to control inrush current when radiation exposed and it makes
system stabilized.

(Rev. 1.0) 31 Operation Manual


EcoRay PX-300HF

14. This X-ray system has quick rise time as 1mSec and could have good radiation efficiency.
15 This X-ray system could use low electric condition with PSU (Power Storage Unit – Battery type).

2.2 How to use Generator (SIG-30-CON)

2.2.1 Function of Console

9. Ready / X-Ray lamp 1.Tube Focal lamp


10. Generator Standby lamp 2. Reset switch
11. Over load lamp 3. Power Switch

4. kV / mA / Time
Display window

12. Anatomical mode


5. Bucky switch

6. Ready / X-Ray switch

13. AEC mode 7. Chamber density switch

14. Chamber Detector 8. Selected Chamber


Display lamp

2.2.2 Explanation of Console Function

1. Tube Focal lamp / Reset switch


(1) Tube Focal is divided with small and large focal and lamp will be lighted when selected.
(2) Reset switch is used in system error condition and if pushed this switch, it makes microprocessor
initialization.

Reset switch

Small Focal

Large Focal

2. Power Switch
(1) This switch controls electric power and if push ON switch, power supplied to system and if OFF
switch, disconnect power. Green lamp lighted
when power off

Power ON
Power OFF

(Rev. 1.0) 32 Operation Manual


EcoRay PX-300HF

3. kV / mA / Time display window

(1) Display set exposure value with numeral.

kV, mA, mAs/Time display window

kV adjustment switch mAs / Time selection switch

mA adjustment switch mAs / Time adjustment switch

4. Bucky selection switch


(1) If push bucky icon and then will be selected and lamp will be lighted.

Wall Bucky Fluoroscopy Bucky

Table Bucky Tomography Bucky

5. Ready / X-Ray switch


(1) This switched are used in X-ray exposure. Ready switch is preparing radiation and X-ray switch is
radiation exposure.

X-Ray exposure switch


Ready switch

6. AEC selection switch


(1) AEC is abbreviation of Automatic Exposure Control and it controls exposure time automatically.

AEC ON/OFF switch Select Chamber position

Chamber Detector Display Chamber density

switch
Selected Chamber Display
lamp

(Rev. 1.0) 33 Operation Manual


EcoRay PX-300HF

7. Selection of APR
(1) APR is abbreviation of Anatomical Programming Radiographic and it could memory exposure
value to the microprocessor of PCB and could use it with easily.
(2) Memorized exposure value of microprocessor could change by operator’s need.
(3) APR function will be activated automatically by Powering ON of Console.

Select body size


Select body point

Select exposure method


Memory switch

8. Generator status display lamp


(1) Display Overload, Standby, Ready and X-ray working condition and could check system condition
by lighted lamp status.

Power status display Ready status display lamp

Overload display lamp X-Ray exposure display lamp

(Rev. 1.0) 34 Operation Manual


EcoRay PX-300HF

2.2.3 Supply electric power

1. Supply electric power to the system


(1) Must check the input voltage by tester before Powering ON of Circuit breaker
(2) Single phase 120VAC must supply to the system and could check voltage by tester.

2. Supply electric power to the generator


(1) If push power ON switch of Console, electric power will be supplied to the generator and
disconnect electric power if push OFF switch of Console.
(In OFF condition: green lamp lighting above OFF switch)
(2) Could check generator electric condition by Standby lamp lighting condition. (Standby: Yellow,
Supplied power: Green)
(3) External devices will be activated by Console power ON.

Power ON switch

Electric power status lamp Power OFF switch

Normal: Yellow

(Rev. 1.0) 35 Operation Manual


EcoRay PX-300HF

2.2.4 EMERGENCY STOP

1. EMERGENCY STOP
(1) Safety devices for patient or operator to disconnect electric power when patient or operator are
placed in the electrical or mechanical emergency situation.
(2) Emergency stop switches are located at the top of the generator and top of the handle bar and if
push this switch and then disconnect electric power, and turn it clockwise electric power will be
supplied.

EMERGENCY STOP EMERGENCY STOP

2.2.5 Turn on the Power ON

1. Introduction of Power ON
(1) To operator the system, push the power ON switch of Console.
(2) After power on, system will diagnosing automatically during 15 seconds.

● Push the power ON switch

● After power ON, Standby lamp lighting with yellow and green.

In normal condition, green lamp lighting after 15 seconds later.

● After power on, system will diagnosing automatically during 15 seconds.

● Display progress at the mA window by 0 to 100.

● In the normal condition, 100 will be displayed at the mA window and could use

X-ray system.

● In abnormal condition, error will be displayed at the mA window before 100 and

then could not use X-ray system.

(Rev. 1.0) 36 Operation Manual


EcoRay PX-300HF

2. Procedure of Power ON
(1) Push power ON switch of Console
(2) Standby lamp lighting with yellow and green if system is normal green lamp will be lighting.
(3) ”Set 0 r 1.00” will be displayed at the Console window and if system is normal condition,
“Set 100 r 1.00” display after 15 seconds later.
(4) “Set 100 r 1.00” displayed at the window, could use X-ray system.
(5) If abnormal condition, Error code will be displayed at the Console window and then could not
use X-ray system.

2.2.6 Check System Condition

1. In normal condition

(1) After power ON, check system condition automatically.

(2) After power ON, will be changed to the Radiography mode after
15 seconds later.

2. In abnormal condition

(1) After power ON, check system condition automatically.

(2) During system check condition, error code will be displayed


when system troubled.

(3) If error displayed, turn ON/OFF repeatedly, but error code not
appeared, system must be checked by authorized engineer.

2.2.7 How to Use


1. APR (Anatomical Programming Radiographic) function
(1) APR function will be activated automatically after power ON of console.
(2) Memorized APR data is saved by manufacturer’s experience but operator could change APR data
by need
(3) APR could use without specific procedure.
(4) APR is consists of 4 function switches.
(See next page images)

(Rev. 1.0) 37 Operation Manual


EcoRay PX-300HF

1. Select Body size


3. Select Body point

2. Select Exposure method


4. Memory switch

(5) Select Body size

Thin Body size Auxiliary

Middle Body size Obese Body size

(6) Select Exposure method

A-P exposure method APR data memory switch

Lateral exposure method Oblique exposure method

(7) Select Body point


● Body point is consists of 21 anatomical mode and 3 types of exposure method.
● If push switch and then selected and lamp lighting.

2. Adjust exposure condition


(1) If change exposure data and memory it, old exposure data will be deleted and could not use it
again.

APR Function Adjust kV, mA, Time Memory

Select APR point


Memorized new data
(Body size, Exposure method,
Body point)

(2) Adjust Exposure value and display


● kV adjustment: Increase →No. 1, Decrease → No. 2

Fast adjustment → No. 1 + No. 2 or No. 3 + No. 2


1 2 3

● mA adjustment: Increase → No. 1, Decrease → No. 2

1 2
● mAs / Time select: If push No. 3 switch and then selected and selected switch will be
lighting.
● mAs / Time adjustment: Increase → No. 1, Decrease → No. 2
1 2 3

(Rev. 1.0) 38 Operation Manual


EcoRay PX-300HF

(3) Display Exposure values


● Adjusted exposure values are displayed at the FND window by number.

Display window

Adjustment switches

3. Select Bucky (Optional Item)


- Do not connect non-authorized devices to this system.
- Additional equipment or external devices should only be connected by an authorized service
engineer of manufacturer and all additional equipment connected to external interfaces of this
system.
- External interfaces of the unit must fulfill the requirements of Standard IEC 60601-1-1.

If the system is operated together with other electrical devices from a multiple socket outlet,
current leakage can increase; the operator/user in this case is responsible for ensuring that the
requirements of IEC 60601-1-1/IEC 60601-1 are still met.

(1) Bucky will be activated automatically after Console power ON.


(2) Bucky signal was designed to operate only EcoRay’s bucky devices. If want to use another
company’s bucky devices, please contact authorized distributor or manufacturer and ask services.

1 2 3 4

(3) Select Bucky position


● No. 1→ Table Bucky, No. 2 → Wall Bucky, No. 3 → Table Bucky (Fluoroscopy)
No. 4 → Auxiliary Bucky
(4) How to use Bucky
● Select Bucky
● If selected, lamp lighting above the selected Bucky switch.
● Bucky device will be operated automatically by program when push Ready and X-Ray switch

(Rev. 1.0) 39 Operation Manual


EcoRay PX-300HF

4. AEC Functions

1. AEC ON/OFF switch 2. Select Chamber position

4. Display Chamber density

3. Chamber Detector switch


5. Density switch

6. Selected Chamber display


lamp
(1) How to use AEC
● Ionization chamber for AEC will be installed by Bucky position and if select Bucky, chamber will
be selected automatically by program.
→ If push AEC ON/OFF switch, AEC mode will activate.
→ If turn ON AEC mode, lamp lighting above the AEC switch.
→ At the same time, density display and selected indicating lamp will be lighting.
→ Select wanted function and then adjust density with density switch.
→ Adjusted density value will be displayed by number (Density: -8 ~ +8. 16 levels))
→ If beep sound or error code occurred from Console in the using AEC function, push the
“Reset” switch of Console to initialize the system or push the OFF switch of Console and
push the ON switch. Even though, could not use AEC function, ask services by authorized
engineer. (Refer to Maintenance)

(Rev. 1.0) 40 Operation Manual


EcoRay PX-300HF

5. X-ray Exposure
(1) Beep sound and lamp is will be activated during X-ray exposure to announce X-ray radiation to
operator.
(2) X-ray switch is located at the bottom of console and indicating lamp is located at the upper of
console.
(3) X-ray switch is designed to be assembled it at the external and use it. (Hand Switch)
(4) 1 second later after pushing the “Ready” switch, in normal condition, Ready lamp lighting and
then could use X-ray exposure.
(5) If push the "Ready" switch, "Ready" lamp will be lighting with green.
(6) If push the "X-Ray" switch, beep sound will be made and “X-ray” lamp will be lighting with yellow
during X-ray exposure.
(7) X-ray exposure values are different by body point and could adjust it by operator’s need.
(8) Radiation will be exposed by only setting exposure value and could not over expose by circuit
about the X-ray tube capacity.

Ready / X-Ray

Indicating lamp

X-Ray switch

Ready switch

Hand Switch

(Rev. 1.0) 41 Operation Manual


EcoRay PX-300HF

2.3 How to use Tube Stand (SMS-TS-3)

2.3.1 Consist of Tube Stand


6. Electromagnetic (Vertical travel)

5. Electromagnetic

1. Body (Tube rotation)

4. X-ray Tube Arm

2. Up/Down part

3. Tube Stand

Rotation button

2.3.2 Function of Tube Stand


1. Function of Main Parts.
(1) Body: Column for X-tube moving up/down
(2) Up/Down part: Assembled with Tube arm and could move up/down.
(3) Tube Stand rotation button: If push the button, could rotate the Tube stand per 90°.
(4) X-ray Tube Arm: Assemble Electron Tube, Handle, and Collimator.
(5) Electromagnetic: For tube rotation, it was controlled by Handle switch.
(6) Electromagnetic: For up/down movement of X-ray tub, it was controlled by Handle switch.
(7) Handle: Angle, Emergency Stop switch and electromagnetic control switch was assembled.

2.3.3 Turn on the Power ON


1. Tube Stand power will be supplied automatically by Power ON of console.
2. Turn ON the Console Power.
(1) To operator the system, push the power ON switch of Console.
(2) After power on, system will diagnosing automatically during 15 seconds.

(Rev. 1.0) 42 Operation Manual


EcoRay PX-300HF

2.3.4 How to use Handle

1. Consist of Handle 1. EMERGENCY STOP

2. Electromagnetic lock switch

6. Protractor 3. X-ray Tube

4. Handle grip
7. Iris knobs
5. Collimator

8. Lamp Switch

2. Each name of Handle

No Description
① Tube rotation
② Tube Support Unit Transverse movement
③ All magnetic lock will be release without tube rotation.
④ Tube Support Unit vertical movement
⑤ Front and back movement of Handle Unit
⑥ Same function of “③”
⑦ Display Tube Protractor
⑧ Display 42inch with Table top <-> X-ray Tube
⑨ Display 42inch with Cassette <-> X-ray Tube
⑩ Display 72inch with Wall Bucky <-> X-ray Tube
⑪ Display 42inch with Wall Bucky <-> X-ray Tube
⑫ Emergency Switch
⑬ Hand grip

(Rev. 1.0) 43 Operation Manual


EcoRay PX-300HF

3. EMERGENCY STOP switch


(1) It serves to switch off the system in a dangerous situation as quickly as possible.
(2) Emergency stop switch is located on the upper side of Handle, and if pressed the
EMERGENCYSTOP switch, disable system to prevent any further X-ray radiation and if turn it
clockwise, electric power will be supplied to normally again.

EMERGENCY STOP

NOTE ☞

In the EMERGENCY STOP switch was pressed condition, all equipments are could not use
including generator.

4. How to use electromagnetic lock


(1) This product used electromagnetic lock.
(2) Electromagnetic lock will be operated by Handle switches.
(3) If push the Handle switch, electromagnetic lock will be released and if leased the Handle switch,
electromagnetic operate.
(4) If push the Handle switch, beside lamp of switch lighting.

5. Movement of Tube Stand


(1) When electric power supplied to Tube Stand, Tube Stand was locked by electromagnetic lock
and if push the Handle switch, electromagnetic lock will be released and could moves up and
down and could rotate tube arm.
(2) Do not move Tube Stand without release the electromagnetic lock.

If try to move Tube Stand by force in lock condition, it cause of troubles.

(Rev. 1.0) 44 Operation Manual


EcoRay PX-300HF

(3) Moving Forward & Backward: Press Stopper Pedal (Brake) and then brake will be release, if
released, could push or draw the mobile X-ray system with
hand grip.
(4) Parking: If want to X-ray Exposure, mobile X-ray unit must be parked. Move the Tube Stand
with little power in the condition of pushing the Handle switch.

Hand grip

Stopper Pedal

Lock: If you press this pedal, mobile

Unlock Lock x-ray system will be parked.

Unlock: If you want to move, press


this pedal.

1. Mobile X-ray System must be parked in flat floor.


If parked in lean floor or bumpy floor, mobile x-ray could moves.
It cause of crash accident!
Warning! 2. X-ray must be exposed in the parking condition!

(Rev. 1.0) 45 Operation Manual


EcoRay PX-300HF

6. How to use Protractor


(1) Protractor of Handle displays of X-ray tube rotation protractor.
(2) When rotate Handle with pushing switch with wanted direction, protractor will be moved by
the center of gravity and display protractor with scale.
(3) Reduce speed and released the switch to stop the X-ray tube rotation

If X-ray exposed in X-ray tube protractor is not correct condition, could not get right X-ray
images.

7. Operation of Collimator

(1) Function of Collimator


- X-ray irradiation angle is adjusted by manual.
- Function and description of Collimator is as below.

No Description
① X-Ray filed adjustable knobs
② Momentaly push switch of lamp
③ Dial for film size
④ Back panel
⑤ Lock Knobe
⑥ Mounting flange for X-Ray tube
⑧ Center Cross marked plate
⑨ Cone rail

(Rev. 1.0) 46 Operation Manual


EcoRay PX-300HF

(2) Incident rays of Collimator


1) X-ray radiation incident with vertically with table top. (Refer to below image)

(3) How to use Collimator


- Move Tube Stand to the patient exposure point.
① Minimum distance should be 45cm patient with X-ray tube.
② X-ray tube should be set with long as possible in normal condition.
- Push the Collimator lamp switch.
(After turn ON lamp, 30 seconds later collimator lamp will be turn off automatically.)
- After turn ON lamp, adjust exposure area with adjusting of iris knobs.
- If collimator lamp turns off during adjusting exposure area, turn ON the collimator lamp again.
- If setting is finished, push the X-ray switch.
- There is no problem when X-ray exposed in the lighting condition.

X-ray radiation is dangerous for both patient and operator and try to be exposed in
minimum exposure area as soon as possible.

(Rev. 1.0) 47 Operation Manual


EcoRay PX-300HF

2.3.5 P.S.U. (Power Storage Unit – It’s located in inside of Mobile Cart)

Caution!

First, this mobile x-ray unit should be charged during 1 hour before use.
If not charged, Mobile X-ray system could not operated in normally.

Notice!
Charging is necessary after installation

(Rev. 1.0) 48 Operation Manual


EcoRay PX-300HF

Chapter 3

X-ray room Requirements

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 49 Operation Manual


EcoRay PX-300HF

3. X-ray room Requirements

3.1 Environmental Requirements

3.1.1 An ambient temperature: +10°C ~ +40°C


3.1.2 A relative humidity: 30% ~ 75% (Without dew condensation)
3.1.3 An atmospheric pressure: 700hPa ~ 1,060hPa
3.1.4 This system must be used in the without corrosion and explosion gas.

3.2 Electrical requirements

3.2.1 This generator contains advanced circuitry which will maintain the selected X-ray values during
irregular input voltage condition. If input voltage condition is over or less, then could not get a
good X-ray radiation and have a good electric facility to be supplied with good regulation
condition.

3.2.2 Supply specific electric power to the system.


1. Do not under size distribution transformer.
2. Insure the suitable electric facility for the system.

3.2.3 Generator Power line requirements.


1. Power line voltage variation must not exceed ± 10%.
2. Maximum line regulation for the maximum kVA demand is 5%.

Model PX-300HF

Input Power Max. 1 kVA

Input Voltage Single phase, 120V

3.2.5 Recommended wire size


1. Correct sizing of the feeder wire is critical to proper generator operation.
2. Wire size is depend on the kW rating of the generator, line voltage and the distance from the
generator and electric circuit breaker.
3. Minimum wire size from generator and electric circuit breaker.

Wire size
Generator Line Voltage
9m
1 kVA 208-240 VAC 2 mm2

(Rev. 1.0) 50 Operation Manual


EcoRay PX-300HF

3.2.6 Interconnection and grounding requirements.


1. Main line of the generator must be connecting with main electric circuit breaker.
(1) Must use 415VAC 30A circuit breaker.
2. Connected external device with generator must have electric circuit breaker or fuse.
3. Ground terminal of the generator must be grounded with ground line of the main power line.
4. Use recommended ground wire size by local regulation.
5. Wiring connection of main devices.

Cable 4
X-ray tube Handle bar Tube Stand

Cable 3

Cable 5
Main Power

Cable 1

Generator
H/V Transformer
(Mobile Cart) Cable 2
Console

(Rev. 1.0) 51 Operation Manual


EcoRay PX-300HF

6. Connected Main Devices

Cable No Q’ty Cable length Function Remarks

Single phase power


Cable 1 1 9,000mm Connect line with indicated.
(120VAC)

Cable 2 1 Assembled Signal line Connect generator with console

Cable 3 1 4,000mm Power, Tube Stator line Connect generator with handle bar

Cable 4 1 650mm Tube Stator Including temperature relay signal

Cable 5 1 pair 2.,500mm High Voltage Cable Connect H/V transformer with X-ray tube

NOTE ☞

Use recommended ground wire size by local regulation. And ground terminal is located in beside of
generator main fuse.

7. Fuse capacity of main parts.

No Fuse Name Position Capacity Quantity

1 F1,F2 Main Fuse 500V, 63A 2

2 F3,F4 Trans Fuse 250V, 15A 2

3 F701 Cue-XR5-MiO 250V, 10A 1

4 F901 Cue-XR5-MiO 250V, 3.15A 1

5 F801 Cue-XR5-CnD 250V, 10A 1

6 F501 Cue-XR5-FnD 250V, 3.15A 1

7 F701 ECO-XR5-GC 250V, 20A 1

8 F700,F704,F705,F708 ECO-XR5-GC 250V, 5A 4

9 F702,F703,F706 ECO-XR5-GC 250V, 3.15A 3

10 F707 ECO-XR5-GC 250V, 10A 1

(Rev. 1.0) 52 Operation Manual


EcoRay PX-300HF

Chapter 4

Maintenance

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 53 Operation Manual


EcoRay PX-300HF

4. Maintenance

4.1 An Overview

4.1.1. The purpose of this Periodic Maintenance is to assure continued safe performance of the X-
Ray generator, to increase serviceability, to reduce the cost (save time, repairs, etc.) and to
assure the safety (personal risk).
4.1.2 The following checks and maintenance procedures, together with the suggested intervals, are
the manufacturer’s recommendation for the most effective periodic maintenance schedule for
this generator.
4.1.3 Service task here described must be performed exclusively by service personnel specifically
trained on medical X-Ray generator.

4.2 Periodic Maintenance Procedures

4.2.1 To use X-ray system without troubles, periodic maintenance is the most effective method.
4.2.2 If X-ray system is out of order, contact local distributor or manufacturer ask service.
4.2.1 General cleaning

Never attempt to clean or handle any part of the X-Ray generator when it is turn on. Switch
off the generator main disconnect before cleaning or inspecting.

1. External surfaces cleaning


→ Clean external covers and surface frequently (without cloth).
→ If chemical is present on the surface, clean it quickly to protect rust.
→ Use dry clothe when cleaning the surface.
→ Do not use cleaners or solvents of any kind.
→ If external surface is scratched, touch up if need.

2. Internal generator cleaning

Keep the external protection cover in place all the time, only remove them to perform service
operations by qualified engineer.

(Rev. 1.0) 54 Operation Manual


EcoRay PX-300HF

4.3 Cable check

4.3.1 Check that all electrical connections are firm and secure and that all cable clamps and strain

relief are in place.

4.3.2 Also check that connectors do not have exposed wires and check cable Sheaths for wear and

fraying.

4.4 Console check

4.4.1 Check hand switch wiring condition.

4.4.2 Check correct operation of console switches, displays and indicators by performing the

following test:

1. Check console standby lamp lighting with yellow.

2. Push the “Power On” button of console and supply electric power to generator.

3. After system checking, changed normal mode to use or not.

4. Check all switches working condition.

5. Set over X-ray exposure value and check “Over Load” message display or not.

6. Check “Ready” and “X-Ray” switches working condition.

4.5 High voltage transformer check

4.5.1 Check oil leakage or not.

4.5.2 If oil leakage was found, have to check oil level with oil fill plug.

(Normal: within 15mm of the top surface of the transformer)

4.5.3 Put silicon oil to high voltage cable plug after cleaning with dry cloth.

4.5.4 Check arc evidence of high voltage cable plug and Receptacle Socket or not.

4.5.5 Check high voltage cable fixed tightly or not.

4.6 X-Ray Tube check

4.6.1 Check oil leakage or not from the X-ray tube.

4.6.2 If oil leakage was found, have to refill or exchange X-ray tube.

(Rev. 1.0) 55 Operation Manual


EcoRay PX-300HF

4.6.3 Put silicon oil to high voltage cable plug after cleaning with dry cloth.

4.6.4 Check arc evidence of high voltage cable plug and Receptacle Socket or not.

4.6.5 Check high voltage cable fixed tightly or not.

4.7 Periodic Maintenance check point

4.7.1 Make a periodic maintenance plan to use X-ray system safety and operator has a responsibility

about this.

4.7.2 Periodic maintenance must be done by qualified service engineer by manufacturer.

4.7.3 Periodic maintenance must be done after installation and must be inspected by authorized

inspection office per every 3 years.

1. Periodic Maintenance check point

Item Check Point Daily Monthly Annually

1. Check external surface condition (Damaged or Broken) ○

2. Check segment or LED display condition is normally or not. ○

3. Check cable condition. (Pressed or Twisted or Disconnected) ○


Appearance

4. Check hand switch working condition. ○

5. Check all buttons function, when push the button, displayed



it normally or not on the display.

1. Check X-ray tube rotor rotating or not when push the



“Ready” button.

Working 2. Check X-ray exposed or not when push the “X-ray” button. ○

3. Check selected Bucky working or not when pushes the



“Ready” and “X-ray” button.

Periodic 1. X-ray performance inspection must be done in periodically


ㅡ ㅡ ○
Inspection by relative regulations.

1. X-ray room condition is suitable for relative regulations. ○

Environment 2. Check around the X-ray system that dangerous factors for

patient to be or not.

(Rev. 1.0) 56 Operation Manual