Technical Publication

SMS-M-002 (Rev. 1.0)

Service Manual

Model: PX-300HF

Important!
Please read this manual thoroughly and use the system correctly before using the
system. This manual desires to keep in promised place to use always again after
reading.

EcoRay Co., Ltd.


621-14, Docheon-Dong, Gwangsan-Gu, Gwangju, Korea 506-301
Tel: +82-62-952-3251 / Fax: +82-62-951-5261

Seoul Office: #402, Cygnus B/D, 115-12, Nonhyun-Dong, Kangnam-Ku, Seoul, Korea 135-821
Tel) +82 70 7510 3402 Fax) +82 70 8630 3420 E-mail: sales@ecoray.kr
EcoRay PX-300HF

MPORTANT! X-RAY PROTECTIONS!

X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS SAFETY


INSTRUCTIONS ARE STRICTLY OBSERVED IN USE!

Accordingly, when use this product should be thoroughly read the "Radiation Safety Control Regulation"
and then compliance with it in use. This product is serious injury's danger if do not keep use guideline
or regulation therefore, read and understood this manual before in use.

This product was designed by safety level (SL) presenting in Radiation Safety Control Regulation.
However, if do not keep using method and Radiation Safety Control Regulation, will not be able to
avoid serious injury's danger from radiation. Therefore, to prevent accident from radiation, everyone who
work connected with radioactive emission must understand radiation safety control regulation and
understand product manual fulfill thorough education and formality.

Our company recognizes enough about danger of excessive exposure to operator, installation people,
and maintenance people. Therefore, our company strongly advises that operator, installation people,
maintenance people must be educated about radiation safety control regulation and should be used this
product when understand fluently about all knowledge of operation this product.

Also, our company informs that operator, installation people, and maintenance people have a
responsibility to be educated fluently about radiation safety control regulation and understood about
product use method.

Also, our company has no responsibility about injury or calamity that occurs does not keep use method
of this product and radiation safety control regulation.

Therefore, our company strongly advises that to be used this product after that person in charge of
installation, maintenance and operator should understood radiation safety control regulation of domestic
or foreign.

Thanks for purchasing product of EcoRay.


Before using the system, please read this manual thoroughly and use the system correctly. The
precautions and prohibitions are seen though the manual is classified as follows.

(Rev. 1.0) 1 Service Manual


EcoRay PX-300HF

NOTICE AND CONTRAINDICATION!

States a direct danger that may cause death or serious injury if it is not avoided.

States an indirect or potential danger that may cause death or serious injury if it is not
avoided.

State a danger that may cause slight or medium injury or may cause damage in
equipment or fire if it is not avoided.

NOTE ☞
States the information, which helps to use the system correctly.

CONTRAINDICATION!
This x-ray system is not suitable for therapeutic applications.

(Rev. 1.0) 2 Service Manual


EcoRay PX-300HF

Precautions in the use of X-ray system


(For the Safety and Prevention of Danger)

1. Only one experienced technician should operate the equipment.


2. When installing the equipment, pay attention to the following items.
2.1 Do not install it near water faucet or similar equipment.
2.2 Install it away from potential sources of problems such as abnormal pressure, temperature or
humidity, drafts, direct sunlight, dust, chlorine or sulphur gas.
2.3 During transportation and operation of the equipment, avoid tilting, vibration and sharp impact
against it.
2.4 Keep the equipment away from the areas where chemicals or gases are stored.
2.5 Use only the correct electrical power source with matching frequency, voltage and current (or
wattage).
2.6 Check the condition of the battery power source (Power Supply Unit). (Power and polarity before
operating the equipment.)
3. Before operating the equipment, pay attention to the following items:
3.1 Check the conditions of switch contacts, polarity, and make sure the equipment performs correctly.
3.2 Confirm that the ground is connected properly.
3.3 Check all wiring for proper and correct connections.
3.4 Check the condition of the external electric circuit breaker.
3.5 Circuit breaker will be directly connected to the system and do not connect with another system.
3.6 When the equipment has been leaved for a long time (more than 10days), You must warm-up
equipment and X-ray tube sufficiently before use.
4. While operating the equipment, pay attention to the following items:
4.1 Do not ever-exceed time or the amount of equipment use needed for diagnosis.
4.2 Observe the equipment and the patient continuously for detection of problems.
4.3 When a problem is detected with the equipment or the patient, take proper action to stop the
equipment without harming the patient.
4.4 Do not let the equipment touched by the patient.
5. After operating the equipment, pay attention to the following items:
5.1 Turn of the switch and then, turn off the main power switch.
5.2 Clean the part that contacted with body.
6. When the equipment is out of order, call a certificated engineer. (Do not try to repair it.)
7. Do not modify any part of the equipment.
8. The equipment must be checked periodically for good operation condition without troubles.
9. Do not use x-ray system connecting with other system.

(Rev. 1.0) 3 Service Manual


EcoRay PX-300HF

1. The owner/user has a responsibility for management and maintenance of this equipment.
2. This equipment must be operated by only authorized personnel.
3. Contact authorized service engineer when repair or inspection of the inside of the equipment.
4. In general, modifications are strictly prohibited by the Regulations. Please contact our agency if it is
needed to modify the device.
5. Do not use this system in conjunction with the other machines because seriously danger could be
happened unexpectedly.

Periodic Inspection

1. Maintenance is required to maintain safety and performance of this system for a long time.
2. Maintenance must be done periodically with maintenance plan.
3. This system must be inspected from the inspection office per every 3 years.

Warranty

1. This warranty shall apply only to defects that arise within 12months of purchase.
2. This product could have a free of charge service if troubles are occurred during the warranty
period.
3. The warranty does not cover defects caused by abuse, mishandling, accidents or improper
installation.
4. The following will not be covered under the warranty.
4-1. Modifications to the product, or repairs made by unauthorized person but manufacturer or
authorized person by manufacturer.
4-2. Acts of God. Including fire, flood, storm, power or voltage surges.
4-3. Improper use or installation.
4-4. Defects that arise in conjunction with the use of other machines.
4-5. Use of improper parts or modification the circuit.
4-6. Improper use
5. In case of repaired during the warranted period, User shall advise the result of test to manufacture.

(Rev. 1.0) 4 Service Manual


EcoRay PX-300HF

Precautions in Use

When using this equipment, please observe the following precautions for the safety of the operator
and the patient.

1. Improper use of the X-ray equipment might cause the operator or the patient to be accidentally
exposed to X-ray radiation.
2. During X-ray radiation, any person other than the subject the patient should not stay in the room.
3. X-ray room must have radiation protection facility by "Radiation Safety Control Regulation"

1. Only qualified personnel should handle the equipment


2. Electric shock or leakage of electricity can happen unless ground is not installed necessarily.
3. Keep the equipment dry
4. Otherwise it may cause an electric shock to the operator of the patient and if any liquid or water
flew into electric device turn off the power and contact company or service agency.

NOTE ☞

For the safety of operator and patient, opening the X-ray generator cover, X-ray tube,
Collimator and other equipment cover are strictly prohibited. Only qualified engineer
could open the cover.

(Rev. 1.0) 5 Service Manual


EcoRay PX-300HF

Scope

1. This manual contains information of system operation, components, control keys, switches and
hazards and technical reference information.
2. This manual may not be reproduced in whole or in part without the written permission of EcoRay Co.,
Ltd.
3. The text of this manual is a translation based on the Korean alphabet (Hangŭl) original.

EcoRay Co., Ltd.


621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea

Revision History

Revision No. Revision Date Change Description

Rev. 1.0 2008-03-10 Release version

(Rev. 1.0) 6 Service Manual


EcoRay PX-300HF

Table of Contents

Chapter 1
1. Pre-installation ……………………………………………………………….……….………………………….……….…. 10
1.1 An Overview ……………………………………………………………………………………………………….…………………….……..… 10
1.2 Responsibility of Customers ……………………………….………………………………………….………….…………….….……. 10
1.3 Constructions and Function .…………………………………………….………….…...………………………………………..….…. 10
1.4 General details …………..……………….………………………….……………………………..…………………………………….….…. 11
1.5 Service Numbers …………..……………………………………………...…………………………….…………………………….………. 11
1.6 Product Pre-installation Data ..……….…………………………………………………………………………………………………. 12
1.7 X-ray room Requirements ……...……………………………..………………………………………………………………….………. 13

Chapter 2
2. Installation …………………….…………………………………..…………………………………………………………… 18
2.1 An Overview ……………………………………………………………………………………………………………………………….………. 18
2.2 Pre-installation checks points and Tools and Testers ………………………………………..….…………………………... 18
2.3 Pre-installation Check ………………………………..……………..…………………………………………..…….……………………… 19
2.4 Installation of Generator ……….………………………………………………………..……………………….….……………………... 20
2.5 Installation of Tube Stand ………………………………………………………………………………….……….……………………... 27
2.6 System Cable Wiring ……………………………………………………………….………………………………………..……………... 35

Chapter 3
3. Technical Information …………….………………………………..………………….……………….……………..……. 37
3.1 An Overview …………………………………………………………………………………….……………………………..…………..……… 37
3.2 Installation information ……………………………………………………….…………………………..…………….………....…..…… 37
3.3. Exposure condition data ………………………………………………………………..……………………………..………...………… 37
3.4 Maintenance History ……………………………………………………………………………..…………………………….…..………… 38

Chapter 4
4. Calibration …………….…………………………………………….………………………….……………………………… 40
4.1 An Overview ..……………………………………………………………………………………………….…….……………………………… 40
4.2 Introduction ………………………………………….……………………………………………….…………………………………...……… 40
4.3 Needed tool and testers when Calibration …..……………………………………………………………………………..…… 40
4.4 Check point before calibration …..………………………………….………………………………………………….………...…… 40
4.5 Consist of Calibration Program …..…………………………………………………………………………..……………….……..… 41
4.6 How to use Calibration program ………………………………………………………………………………….….…….…….…… 41

(Rev. 1.0) 7 Service Manual


EcoRay PX-300HF

4.7 Time of Calibration …………………………………………………………………………………………………….………...…….…… 42


4.8 Must calibrated when Main Parts are repaired or replaced ………………………….…………….…………..….…… 43

Chapter 5
5. Trouble Shooting ………………………………..……………………………..………….….………………….…………. 45
5.1 An Overview ………………………………………………….……………………………………………….……..……………..……….…… 45
5.2 Error type ……………………………………………………………………………………………………………...………………..……… 45
5.3 Trouble Shooting ……………………………………………………………………………………….………....……………………...…… 51
5.4 Handle Bar (SMS-TS-340 Board) …………………………….…..…………………………………..……………………...……..… 73

Chapter 6
6. Circuit Diagram ……………………………………………………………………………….…………………………...…. 75

(Rev. 1.0) 8 Service Manual


EcoRay PX-300HF

Chapter 1

Pre-installation

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 9 Service Manual


EcoRay PX-300HF

1. Pre-installation

1.1 An Overview

1.1.1 This manual contains operation instructions for radiological technicians, radiologists and other
medical personnel who are trained to operate radiation equipment.
1.1.2 These instructions are not suitable as a replacement for components or exchange of circuit or
medical training.
1.1.3 This system was made for trained personnel of medical examination, diagnosis and before
surgical operation and after.
1.1.4 This manual contains of PX-300HF product information, its associated components, needed
environment and electrical requirements.

1.2 Responsibility of Customers

1.2.1 Customer has a responsibility for planning and preparation of installation.


1.2.2 The following items ((1.2.3) are must be checked in pre-installation site by customer and
additional work may be needed by depending on the specific site circumstances.
1.2.3 Install required material before delivery of system components.
1. Complete room floor, ceiling, wall and cleaning.
2. Electric power facility (Circuit Breaker, ground, duct, raceway, etc.)
3. Provide current room dimension, hallway and entry door sizes.
4. Cost of alteration and modifications when not specifically provided in sales contract.

NOTE ☞ Complete and proper pre-installation will avoid delays and confusion.

1.3 Constructions and Function

This X-ray system is consist of Tube Stand which support X-ray tube and moveable, Console, X-ray
generator, Patient table, and X-ray tube unit.
1.3.1 Tube Stand is consist of main power supply unit, electric circuit, handle bar and mechanical
devices and could move Tube stand with floor rail.
1.3.2 Console is consist of each switch and could adjust exposure values and display it.
1.3.3 X-ray generator is consists of control unit and high voltage transformer unit and control x-ray
exposure values and make high voltages by console.
1.3.4. X-ray tube unit consist of X-ray tube, Collimator and Handle bar and could move Tube Stand

(Rev. 1.0) 10 Service Manual


EcoRay PX-300HF

and adjust x-ray exposure area.


1.3.5 X-ray radiation will be made by pushing of X-ray switch of console or pushing of hand switch.
1.3.6 X-ray radiation will permeate patient from x-ray tube to bucky device and after permeation,
could archive x-ray images by x-ray film or digital detector.

1.4 General details

1.4.1 Manufacturer
EcoRay Co., Ltd.
621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea

1.4.2 Place of Manufacture of the PX-300HF


EcoRay Co., Ltd.
621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea

1.4.3 Representative in European Union

1.5 Service Numbers

1.5.1 If the system does not work reliably, or if errors occur in terms of control functions which are
described in the manual, the owner should contact the nearest EcoRay Service Center.
1.5.2 The EcoRay Service Center can be reached at the following Phone and Fax numbers:
1. Service for Europe

2. Service for Outside of Europe, Worldwide


Tel) +82 70 7510 3402 Fax) +82 70 8630 3420 E-mail: sales@ecoray.kr

(Rev. 1.0) 11 Service Manual


EcoRay PX-300HF

1.6 Product Pre-installation Data

1.6.1 Generator

1. Physical characteristics

Dimension (mm). Approx. deviation ±10% Weight (Kg)


Item
Length Width Height Approx.

Mobile Cart
(Including PSU, Generator, 1300 640 1980 300
Tube Stand)
Generator
(Including H/V Transformer) 590 600 440 105
(Installed inside of Mobile Cart)

2. Dimension

Note!
Console and X-ray generator, PSU was already installed inside of Mobile Cart!

1.6.2 Tube Stand (SMS-TS-3)

1. Physical characteristics

Size (mm) deviation ±10% Weight(Kg)


Configurations Deviation less
width length height than ±5 Kg

Tube Stand 180 150 1,690 Approx .120

(Rev. 1.0) 12 Service Manual


EcoRay PX-300HF

2. Dimension

1.7 X-ray room Requirements

1.7.1 Transportation and Storage Requirements


1. An ambient temperature: +10°C ~ +60°C
2. A relative humidity: 10% ~ 90% (Without dew condensation)
3. An atmospheric pressure: 700hPa ~ 1,060hPa

1.7.2 Electrical requirements


1. An ambient temperature: +10°C ~ +35°C
2. A relative humidity: 30% ~ 65% (Without dew condensation)
3. An atmospheric pressure: 700hPa ~ 1,060hPa
4. This system must be used in the without corrosion and explosion gas.

Troubles are could be occurred in the use or storage under unsuitable environment
condition.

(Rev. 1.0) 13 Service Manual


EcoRay PX-300HF

1.7.3 Electrical requirements


1. This generator contains advanced circuitry which will maintain the selected X-ray values during
irregular input voltage condition. If input voltage condition is over or less, then could not get a
good X-ray radiation and have a good electric facility to be supplied with good regulation
condition.

2. Supply specific electric power to the system.


(1) Do not under size distribution transformer.
(2) Insure the suitable electric facility for the system.

3. Generator Power line requirements.


(1) Power line voltage variation must not exceed ± 10%.
(2) Maximum line regulation for the maximum kVA demand is 10%.

Model PX-300HF

Input Power Max. 1 kVA

Input Voltage Single phase, 230V

4. Recommended wire size


(1) Correct sizing of the feeder wire is critical to proper generator operation.
(2) Wire size is depend on the kW rating of the generator, line voltage and the distance from the
generator and electric circuit breaker.

(3) Minimum wire size from generator and electric circuit breaker.

Wire size
Generator Line Voltage
9m
1 kVA 230 VAC 2 mm2

(Rev. 1.0) 14 Service Manual


EcoRay PX-300HF

1.7.4 Wiring connection of main devices.

Cable 4
X-ray tube Handle bar Tube Stand

Cable 3

Cable 5
Main Power

Cable 1

Generator
H/V Transformer
(Mobile Cart) Cable 2
Console

1.7.5 Main wiring connection (Refer to System Cable Wiring. P43)

Cable No Q’ty Cable length Function Remarks

Single phase power


Cable 1 1 9,000mm Connect line with indicated.
(120VAC)

Cable 2 1 Assembled Signal line Connect generator with console

Cable 3 1 4,000mm Power, Tube Stator line Connect generator with handle bar

Cable 4 1 650mm Tube Stator Including temperature relay signal

Cable 5 1 pair 2.,500mm High Voltage Cable Connect H/V transformer with X-ray tube

NOTE ☞
Use recommended ground wire size by local regulation. And ground terminal is located
in beside of generator main fuse.

(Rev. 1.0) 15 Service Manual


EcoRay PX-300HF

1.7.6 Fuse capacity of main parts.

No Fuse Name Position Capacity Quantity

1 F1,F2 Main Fuse 500V, 63A 2

2 F3,F4 Trans Fuse 250V, 15A 2

3 F701 ECO-XR5-MiO 250V, 10A 1

4 F901 ECO-XR5-MiO 250V, 3.15A 1

5 F801 ECO-XR5-CnD 250V, 10A 1

6 F501 ECO-XR5-FnD 250V, 3.15A 1

7 F701 ECO-XR5-GC 250V, 20A 1

8 F700,F704,F705,F708 ECO-XR5-GC 250V, 5A 4

9 F702,F703,F706 ECO-XR5-GC 250V, 3.15A 3

10 F707 ECO-XR5-GC 250V, 10A 1

1.7.7 Interconnection and grounding requirements.


1. Main line of the generator must be connecting with main electric circuit breaker.
(1) 415VAC 30A circuit breaker must be used.
2. Connected external device with generator must have electric circuit breaker or fuse.
3. Ground terminal of the generator must be grounded with ground line of the main power line.
4. Use recommended ground wire size by local regulation.

(Rev. 1.0) 16 Service Manual


EcoRay PX-300HF

Chapter 2

Installation

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 17 Service Manual


EcoRay PX-300HF

2. Installation

2.1 An Overview

2.1.1 This installation depends on generator’s type and configuration and must be installed by
instruction of this manual.

2.2 Pre-installation checks points and Tools and Testers.

2.2.1 Pre-installation checks points


1. Before installation, engineer must check and notice as follows.
(1) Check input voltage, temperature, humidity of installation place.
(2) Read and understand installation manual before installation.
(3) Install X-ray system with patient safety first.

2.2.2 Tools and Testers.


1. Described installation tools and test equipments are necessary and could be changeable
installation method by optional items and then engineer should prepare tools and tester in
needed.
(1) Standard and generally tools and tester set used by engineer.
(Multi meter, Screwdriver, Nipper, Long nose, Wrench set, Solder wire, Soldering iron, Hex
Wrench)
(2) Electric Drill and assorted bits set.
(3) Grinding machine for cutting steel.
(4) Cleaning things for parts.
(5) kV meter
(6) mA meter
(7) Radiation Dose meter
(8) Operation & Installation Manual

Only qualified (must trained by EcoRay) engineer could install this X-ray system!

(Rev. 1.0) 18 Service Manual


EcoRay PX-300HF

2.3 Pre-installation Check

2.3.1 Prior to beginning installation, it is recommended to inspect the site and verify that the X-Ray
room complies with Pre-installation requirements, such as:
1. Incoming line of X-ray room (Voltage etc.)
2. Main circuit breaker

2.3.2 Precaution when connecting electric power.


1. First, check the input voltage with tester and then connect the power cable with circuit breaker.
2. The input power must be single phase 120V AC and could check voltage with tester.

Power Cable
Connected with
230V AC

Power cable must be connected with 230V AC wall outlet!


If connected with low or high voltage wall outlet, it cause of troubles!

(Rev. 1.0) 19 Service Manual


EcoRay PX-300HF

2.4 Installation of Generator


(Actually, X-ray Generator was installed inside of Mobile Cart!)

2.4.1 Appearance
1. Console

2. Generator

3. High Voltage Transformer

2.4.2 Consist of Main Parts and Functions


1. Consist of Main Parts
(1) Console
Membrane Switches

External terminal

Console PCB

Note!
X-ray Generator, High Voltage Transformer, PSU, Console was already installed inside
of Mobile Cart!

(Rev. 1.0) 20 Service Manual


EcoRay PX-300HF

(2) Generator (Assembled inside of Mobile Cart)

ECO-XR5-FnD Board

Communication Terminal

Socket Receptacle

High Voltage Transformer

IGBT Block

ECO-XR5-Mio Board

ECO-XR5-CnD Board

2. Function of Main Parts


(1) Console
● Membrane Switches: Select and control functions.
● Console PCB: Control each function after receive signal from Membrane switches.
● External terminal: Connect Console PCB with Generator Main CPU.

(2) Generator
● Communication Cable: Signal cable which connects Console PCB with Generator Main CPU.
● ECO-XR5-FnD Board: Control PCB for driver circuit of IGBT and Tube Filament.
● ECO-XR5-CnD Board: Control PCB for charging and discharging condition of Condenser.
● ECO-XR5- Mio Board: Main PCB assembled Microprocessor.
● Emergency Stop switch: Power off the electric power quickly in emergency situation.
● IGBT Block: Semiconductor switch unit for high voltage to control input and output power to
be supplied High Voltage Transformer.
● High Voltage Transformer: It is a kind of transformer that supply high voltage to the X-ray
tube after received voltage from generator and boost voltage with
maximum 150,000V.
● Socket Receptacle: Connecting socket for High Voltage Cable

(Rev. 1.0) 21 Service Manual


EcoRay PX-300HF

2.4.3. Unpacking
1. X-ray generator was assembled with mobile cart to be installed easily and easy move.
2. Check X-ray generator external condition (damaged or not) after unpacking.
3. If generator was damaged, stop installation and notify this fact to carrier and distributor
immediately.
4. After unpacking, have to check all parts are exactly same as like order and do not dump any of
packing parts such as envelope, box, documents until confirmation is completely.

2.4.4 Start Installation


1. Installation procedure of X-ray generator
(1) Check serial no on the attached label with enveloped certificate’s serial no.
(2) Remove the generator cabinet.
(Remove bolts beside of generator cover)
(3) Check external condition and internal cable wiring condition. If damage parts are found, order
that parts immediately to distributor to prevent delayed installation.
(4) If released connector is found, should connect it in the installation site.
(5) Connect cables by wiring diagram. (Refer to Wiring diagram)
(6) Check cable wiring condition again after wiring.
(7) Connect main electric power after checking wiring.
(8) Must check input voltage with tester before connect it with generator.
(9) Power ON the circuit breaker after checking input voltage.
(10) Check input voltage of each part.
(11) Check functions after Power ON of Console.
(12) Engineer never leave installation site until installation is completely finished.

(Rev. 1.0) 22 Service Manual


EcoRay PX-300HF

X-ray Generator was assembled inside of Mobile cart for easy installation!

Never use re-connected communication cable! If cut communication cable, it cause of


trouble!

2 Generator Terminal for External Devices


(1) Generator Terminal for External Devices

High Voltage Sockets

Connect Console

Connect Tube Stand

Generator

(2) Consist of External Devices


Console

Picture 1

Tube Stand

NOTE ☞
X-ray Generator, Console, Tube Stand was already assembled with Mobile Cart!

(Rev. 1.0) 23 Service Manual


EcoRay PX-300HF

(3) Images after installation

3. Connect High Voltage Transformer with X-ray Tube


(1) Wiring High Voltage Cable (Connect High Voltage Transformer and X-ray tube)

Socket Receptacle X-ray Tube Connected Image

(2) Connect High Voltage Cable


Insulation silicon oil for
Cable Plug Receptacle Socket
reinforcement

The terminal pins of the high voltage cable are extremely delicate and easily
damaged. And then, handle them with very carefully.

(Rev. 1.0) 24 Service Manual


EcoRay PX-300HF

(3) Manufacturer’s Instructions


1) Prepare receptacle socket of high voltage transformer that will be installed and high voltage
cable. (High voltage cable must be handling with always clean condition)
2) Apply enough insulation silicon to the entire of surface of Cable plug (High voltage
transformer) and near of the pin. (Put approximately 1cm of insulation silicon into the High
Voltage Transformer receptacle socket.)
3) Fix cable nut with firmly after connecting of high voltage cable with high voltage transformer.
4) Apply enough insulation silicon to the entire of surface of Cable plug (X-ray tube) and near of
the pin.
5) Connect opposite side of high voltage cable to the X-ray tube receptacle socket.
6) Fix cable nut with firmly after connecting of high voltage cable with X-ray tube.

High voltage cable wiring must be done in dryness place without humidity.

4. Connect Tube Stator


(1) Wiring of X-ray tube Anode Stator
External terminal of Generator
Back side of Handle (ECO-XR5-GCboard)

NOTE ☞
Anode side of

X-ray Tube When connecting of Tube Stator,


never use “Temperature Relay” point.

SC: Stator Common (Terminal 1)


SM: Main Winding of the Stator (Terminal 2)
SS: Shift of the Stator (Terminal 3)
NC: Not use (Terminal 4)
ET: Connect Earth (Terminal E)
Tube 1 Temp: Temp. Relay (Terminal 5)
Tube 1 Temp: Temp. Relay (Terminal 6)

(Rev. 1.0) 25 Service Manual


EcoRay PX-300HF

(2) Specification of X-Ray Tube Anode Stator

X-ray tube Anode Stator wiring is must correctly with recommended by X-ray tube
manufacturer. Never connect terminal No.1 to No. 5 and No. 6.

The terminal pins of the high voltage cable are extremely delicate and easily
damaged. And then, handle them with very carefully.

5. Power Cable Connection


(1) The power cable cord was assembled with X-ray generator and located bottom side of Mobile
cart.
(2) The power cable of this generator must be connected correctly with wall outlet (120V AC).
(3) Before power cable connecting, please check input power voltage.
(4) Connect power cord when input voltage for generator is correct, and push the Power On switch
if want to run X-ray generator.
(5) Main power will be supplied to the generator if power cord is connected with wall outlet.
Remove the power cord from wall outlet, if break electric power completely.

After wiring, adjust voltage level sensor circuit of CnD Board.


How to adjust: Refer to Charge & Discharge circuit error (Check ECO-XR5-CnD Board).

(Rev. 1.0) 26 Service Manual


EcoRay PX-300HF

6. Specification

Item Functions PX-300HF


Generator Type High Frequency Inverter
KVp range 40 ~ 125kVp ±5%
mA range 50 ~ 300mA ±5%
1mSec ~ 6Sec
Exposure Time
(Below 100mSec: inside ±1mSec. Over 100mSec: inside ±10%)
Generator
APR 288 (Standard)
AEC Optional item
Output Frequency 40kHz
Maximum Power 32kW
Line Voltage Single phase, 230V ±10% , 50/60Hz
Rating 125kV / 200mA, 100kV / 300mA at 100mSec
High Voltage
Maximum kVp 125kVp
Transformer
Maximum mA 300mA

2.5 Installation of Tube Stand

2.5.1 Appearance of Tube Stand (SMS-TS-3)

Front view Side view

(Rev. 1.0) 27 Service Manual


EcoRay PX-300HF

2.5.1 Consist of Main parts

7. Electromagnetic lockup device


for vertical travel

6. Electromagnetic lockup device for tube


1. Stand rotation (Tube Arm)
5. X-ray tube
2. Vertical move device
4. Handle Bar

3. Stand base

3. Functions of main components


(1) Stand: Device helps the X-ray tube travelling up & down
(2) Vertical move device: holding X-ray tube and to be connected to Tube Stand to support up &
down travel.
(3) Stand base: holding Tube Stand and to be connected to Floor rail for longitudinal travel.
(4) Handle Bar: The device has a Angle display panel, emergency PWR-off button, electromagnetic
lockup buttons and user holding handles for X-ray tube position control.
(5) X-ray tube device: where it has X-ray tube, Handle Bar and collimator assembled.
(6) Tube Arm: Helping X-ray tube rotation from Handle Bar.
(7) Electromagnetic lockup device for vertical travel: Helping X-ray tube up/down travel from
Handle Bar.

2.5.2 Unpacking
1. For easy installation, Mobile X-ray system almost assembled. And after unpacking, engineer just
connects X-ray tube, Handle, Collimator.
2. Check if all the devices are well condition after unpacked.
3. In case any of the devices are broken or missed, please contact your local agent or manufacturer
quickly.
4. They are packed as a set of Stand, a set of Handle Bar, a set of balance weight.
5. Do not scrap the packing boxes, envelops or inside papers before you make sure all about the
contents.
6. Sometime, Tube Stand separated with Mobile cart by buyer’s asking.

(Rev. 1.0) 28 Service Manual


EcoRay PX-300HF

It is necessary to have 2 or more people to move and install the Tube Stand devices.

Box 1

Tube Stand

Front view Side view

2.5.3 Installation procedure of Tube Stand


(Actually Tube Stand already installed with Mobile Cart!)

1. Check if the serial No. on the label is match with the serial No. on the quality certificate.
2. During the unpacking, check if all the devices are in well condition, especially at cables. In case
any broken or damaged parts, please contact your local agent for replacements.
3. Bring the devices to the installation place. (Do not assemble the devices in advanced) 2 or more
people are needed to move them because the devices are very heavy.
4. Fix (install) the Floor rail on a right position.
5. Connect Tube Stand into the Stand base.
6. Press down the tube head (tube arm) to the level where you can put the holding pin. It is
necessary to pull it down without weight balance.
7. Put the holding pins around the balance weight box position.
8. Then install X-ray tube, Handle Bar and collimator.
9. And put balance weights into the tube stand (in the empty space of balance weights). The
balance weights are very heavy, so please be careful when you move them into the space.
10. Check the cable connections and make them tight if necessary.
11. Make all the cable connections according to the cable diagram.
12. Check if all the connections are correct.
13. If the cable & wire connections are O.K, put main AC input power to the Generator. (Check AC
input power level before you make Power ON)
14. One person holds the X-ray tube and the other removes the holding pin at behind of Tube
Stand.
15. Check if the weight balance is O.K, if the balance is not perfect, add or remove the weight

(Rev. 1.0) 29 Service Manual


EcoRay PX-300HF

balances.
16. PWR ON and check if electromagnetic lockup functions are all O.K – test it from Handle Bar.
17. The engineers should stay in the room until the installation works are all over.

Holding pin position

1. Set up the stand on Mobile base.

2. Fix the stand with bolts.

3. Press the tube arm down to the position where you can put

the holding pin.

Balance weights

Holding pin position


4. Check the holding pin position and put it in.

5. Insert balance weights into the stand.

6. Holding the Arm well and remove the holding pin

then check if the weight balance is O.K

(Rev. 1.0) 30 Service Manual


EcoRay PX-300HF

2.5.4 Starting of installation.

1. Understanding Wiring
Tube Support part of
(1) Understanding of ECO-XR5-GC Board. PIN
ECO-XR5-GC Board
NO
(Upper side of Generator) (Upper side of Generator)
1 Stator Com (SC1)
2 Stator Shift (SS1)
3 Stator Main (SM1)
4 AC Out
5 AC 0V
6 NC
7 DC 0V
8 DC24V
9 EMG
10 EMG
11 NC
12 Tube1 Temp.
Tube Support Unit
13 Tube1 Temp.
14 Door
15 Door
16 GND

(2) Understanding of SMS-TS-340 Board (Handle bar of Tube Stand)

P6 Connector of
PIN
SMS-TS-340 Board
NO.
(Handle bar of Tube Stand)
1 Stator Com (SC1)
P6 Connector
2 Stator Main (SM1)
3 Stator Shift (SS1)
4 AC 0V
5 AC 24V
6 EMG
7 EMG
8 Tube1 Temp.
9 Tube1 Temp.
10 GND

(Rev. 1.0) 31 Service Manual


EcoRay PX-300HF

(2) Connect ECO-XR5-GCof Tube Support Unit and P6 Connector of SMS-TS-340

Tube support unit Tube Support unit of P6 Connector of

ECO-XR5-GCBoard SMS-TS-340

(Upper side of Generator) (Handle bar)

1 Stator Com (SC1) 1 Stator Com (SC1)

2 Stator Shift (SS1) 2 Stator Main(SM1)


P6 Connector
3 Stator Main (SM1) 3 Stator Shift (SS1)

4 AC Out 4 AC 0V

5 AC 0V 5 AC 24V

6 NC NC NC

7 DC 0V NC NC

8 DC24V NC NC

9 EMG 6 EMG

10 EMG 7 EMG

11 NC NC

12 Tube1 Temp. 8 Tube1 Temp.

13 Tube1 Temp. 9 Tube1 Temp.

14 Door NC NC

15 Door NC NC

16 GND 10 GND

(4) Connect P7 of SMS-TS-340 with X-ray Tube

SMS-TS-340

SMS-TS-340의

P7 Connector Anode side of


Handle bar (Inside of Handle bar) X-ray Tube

1 Stator Com (SC1) 1 Stator Com (SC1)

2 Stator Main(SM1) 2 Stator Main(SM1)

3 Stator Shift (SS1) 3 Stator Shift (SS1)

4 Tube1 Temp. 5 Tube1 Temp.

5 Tube1 Temp. 6 Tube1 Temp.


Anode side of
6 GND E GND
X-ray tube

(Rev. 1.0) 32 Service Manual


EcoRay PX-300HF

(5) Connect P8 connector of SMS-TS-340 with Emergency Switch (There is no electric polarity)
(6) Connect P4 connector of SMS-TS-340 with Collimator (There is no electric polarity)
(7) Connect P9 connector of SMS-TS-340 with X-ray Tube rotation electromagnetic rock.
(Take notice of electric polarity)
(8) Connect P12 connector of SMS-TS-340 with X-ray Tube rotation electromagnetic rock.
(Take notice of electric polarity)
(9) Connect P5 connector of SMS-TS-340 with SID display unit (Optional)

2. Assemble of X-ray tube device


(1) X-ray tube device consists of X-ray tube, tube arm (rotation), Handle Bar and collimator.

X-ray tube

Tube arm

Handle Bar

Collimator
X-ray tube device

3. Verification of cables.
(1) Compare the cable configuration is matching to the circuit diagram or cable diagram.
(2) Connect PWR input to Tube Stand
(3) Main PWR ON
(4) Test the electromagnetic locker functions by controlling Handle Bar.

(Rev. 1.0) 33 Service Manual


EcoRay PX-300HF

4. Cabling between Tube Stand and HFG


(1) Power line connection
→ The PWR cables are coming from HFG and it is necessary to measure input PWR level before
the actual PWR engages to Tube Stand.
→ The PWR level is 24VDC and should be measure by a multi-meter to make it sure.

Tube stand (side view) SMS-GC-430 (Upper side of Generator

(Rev. 1.0) 34 Service Manual


EcoRay PX-300HF

2.6 System Cable Wiring

SMS-TS-340 Board
X-ray Tube
(Handle bar of Tube Support)

Electromagnetic lock for Tube rotation

ECO-XR5-GCBoard (Upper side of Generator)

NOTE!
If do not use Door and EMG functions, then never remove the attached pin of tube
support connector pin (Door and EMG). If needs these functions, use these pin connectors

(Rev. 1.0) 35 Service Manual


EcoRay PX-300HF

Chapter 3

Technical Information

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 36 Service Manual


EcoRay PX-300HF

3. Technical Information

3.1 An Overview

This technical information contains that maintenance history and calibration data.

3.2 Installation information

After installation, enter the installation information as below form.

Hospital Name

Install Engineer Install Date

3.3. Exposure condition data

3.3.1 Main Parts.

Install Date Hospital Name

Model X-Ray
Model
Generator
Tube
S/N. S/N.

H/V Cable Length 6m 8m 10m 12m 15m

3.3.2 Calibration Data

Small Focal Large Focal

50mA 75mA 100mA 150mA 200mA 300mA

40kV 40kV

60kV 60kV

80kV 80kV

100kV 100kV

125kV 125kV

(Rev. 1.0) 37 Service Manual


EcoRay PX-300HF

3.4 Maintenance History

3.4.1 Enter the following information after each data modification in this book or Periodic

Maintenance Service

Date Engineer Name Service Note

(Rev. 1.0) 38 Service Manual


EcoRay PX-300HF

Chapter 4

Calibration

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 39 Service Manual


EcoRay PX-300HF

4. Calibration

4.1 An Overview

This manual contains that adjustment method of exposure condition and how to calibrate exposure
data. Calibration data will be memorized at memory chip at the ECO-XR5-Mio Board with digital
value.

4.2 Introduction

4.2.1 This calibration tool is specified program supplied by EcoRay and it will be operated at the
personal computer.
4.2.2 Control exposure value (Ex. kV, mA, etc)
4.2.3 This calibration tool must be operated by qualified engineer by EcoRay.
4.2.4 This calibration tool was designed by many experience data of manufacturer and this tool will
be operated by specified program supplied manufacturer.

4.3 Needed tool and testers when Calibration


Following tools and testers are needed in generally and another tools and testers must be
prepared if needed.
4.3.1 Calibration program
4.3.2 Generally testers and tools.
4.3.3 Insulation silicon for high voltage cable reinforcement.
4.3.4 Digital voltage meter.
4.3.5 kVp meter
4.3.6 mA meter
4.3.7 Time meter
4.3.8 Radiation Dose meter
4.3.9 Al filter set for half-valuelayer test.
4.3.10 Phantoms set for Fluoroscopy test.
4.3.11 Another tester for calibration.

4.4 Check point before calibration


4.4.1 Input voltage
4.4.2 Ground condition
4.4.3 Prepare calibration tools and tester set.
4.4.4 X-ray tube model (technical data) and external equipment.

(Rev. 1.0) 40 Service Manual


EcoRay PX-300HF

4.5 Consist of Calibration Program

4.5.1 Consist of Calibration Program

2. Calibration mode

1. Controller mode
5. Communication state display

& Memory display window

3. X-ray Tube

selection window
4. Bucky selection window

4.5.2 Main Functions.


1. Controller mode: Select radiographic, fluoroscopic and AEC function and display.
2. Calibration mode: After running of calibration, adjust calibration data and save it.
3. X-ray Tube selection window: Select radiographic x-ray tub and fluoroscopic x-ray tube and set
tube stator working condition.
4. Bucky selection window: Select Bucky position, Bucky type and change data.
5. Communication state display & Memory display window: Communicate generator with PC and
read or save setting data.

4.6 How to use Calibration program

4.6.1 Function of Icons


1. Execution Icon. If click this icon on the PC screen, program will be running
automatically.

2. Power Icon. If click this icon, program will be activated.


(Running: Green lamp, Stop: Yellow lamp condition)

3. Communication port: Only COM1 port supplied.

4. Select Radiographic, Fluoroscopy, AEC function

5. Display window for kV, mA, Time

6. Microprocessor information window: memory mA data, kV data, exposure


count, product serial No., microprocessor program version.

7. Calibration mode ON/OFF switch.

8. Select calibration area and display input data.

(Rev. 1.0) 41 Service Manual


EcoRay PX-300HF

9. Calibration data input window

10. Compensation mode of mA data

11. Used compensate mA data if mA data tolerance is too high. Click

“Advanced..” icon if want to compensate mA data.

12. After running, could see mA graphic and could adjust it.

13. Initialize the compensated mA data.

14. Save the compensated mA data.

15. Read the compensated mA data.

16. Read the compensated mA data from PRAM of Microprocessor.

17. Save the compensated mA data from PRAM of Microprocessor.

18. Save the compensated mA data to another computer.

19. Read the compensated mA data from another computer.

4.6.2 Start Calibration


1. Click Execution Icon on the PC screen.
2. Select Radiographic, Fluoroscopic, AEC functions on the controller mode.
3. Select X-ray tube position and Tube stator on the Tube selection window.
4. Select Bucky position and type on the Bucky selection window.
5. Click ON icon of Calibration mode and then program will be activated.
6. Adjust calibration data after running of calibration program.
(Calibration program must be operated by only qualified engineer by manufacturer.)

4.7 Time of Calibration

4.7.1 In installation: After installation, this x-ray system must be inspected by radiation inspection office
even though this x-ray system was calibrated at the factory before delivery. And
must be calibrated when needs calibration.
4.7.2 In periodic inspection: This x-ray system must be inspected by radiation inspection office per
every 3 years. And must be calibrated when needs calibration.
4.7.3 In repaired or replaced: If main parts are repaired or replaced, x-ray system must be inspected by
radiation inspection office. And must be calibrated when needs calibration.
4.7.4. In re-installation or move: If re-install or install it another place then x-ray system must be
inspected by radiation inspection office. And must be calibrated
when needs calibration.

(Rev. 1.0) 42 Service Manual


EcoRay PX-300HF

4.8 Must calibrated when Main Parts are repaired or replaced

4.8.1 Repaired or Replaced of X-ray Tube


4.8.2 Repaired or Replaced of ECO-XR5-MiO Board
4.8.3 Repaired or Replaced of ECO-XR5-CnD Board
4.8.4 Repaired or Replaced of ECO-XR5-FnD Board
4.8.5 Repaired or Replaced of High Voltage Transformer
4.8.6 Replaced of High Voltage Cable or replaced with another length of high voltage cable.

(Rev. 1.0) 43 Service Manual


EcoRay PX-300HF

Chapter 5

Trouble Shooting

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 44 Service Manual


EcoRay PX-300HF

5. Trouble Shooting

5.1 An Overview

5.1.1 This generator has a self-diagnosing function and display error code on the console display if
system has an error.

5.2 Error type


5.2.1 Error type
No Error type Cause
1. Charging to Main Capacity is not normal.
Charge Circuit
1 2. Input voltage is out of boundary. (Too high or too low)
problem
(Standard input voltage: Single phase, 220VAC ±10%)
Tube stator 1. Cable connection to Tube stator is wrong.
2
Circuit problem 2. ECO-XR5-MiO in Stator circuit is out of order.
1. Tube Filament circuit connection is wrong.
Tube Filament
3 2. Filament circuit is out of order.
Circuit problem
3. Tube Filament is short.
HV-DRV-OLP (ECO-XR5-FnD) signal or
Over current to IGBT Driver IC may occur
circuit is wrong.
the error.
(See circuit diagram)
IGBT-OCP (ECO-XR5-FnD) signal or Over current to IGBT Module may occur the
Generator
circuit is wrong. (See circuit diagram) error
4 Protection Circuit
DRV-PWR-UVP (ECO-XR5-FnD) signal
problem Voltage from the IGBT Driver IC is lower
or circuit is wrong.
than standard.
(See circuit diagram)
FIL-OCP (ECO-XR5-FnD) signal or
Over current on the Tube Filament
circuit is wrong. (See circuit diagram)
Communication 1. Wrong connection of communication cable.
5
program 2. Communication circuit in the Main board or Console board is out of order.
1. Wrong connection of communication cable.
6 AEC problem 2. Wrong cable connection between AEC Board and Ion Chamber.
3. AEC Board or Ion Chamber is out of order.
Irregular Exposure 1. Actual exposure time is shorter than set time.
7
Time problem 2. Exposure switch is bad condition.

(Rev. 1.0) 45 Service Manual


EcoRay PX-300HF

5.2.2 Error type and display


Error code will be displayed on the console when occurred error.

Error displayed as like image.

1. Error type 1: Displayed E r r C r F P at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC Exposure
Stator
Circuit Circuit Filament nication Circuit time
Circuit HV-DRV IGBT-OCP DRV-PER
error error - OLP error error error
error
To be checked
To be checked Normal Normal Normal
(Any one of 4 is possibly faulty)

2. Error type 2: Displayed E r r C r F at kVp, mA, Time section

Refer to ECO-XR5-CnD Circuit


Tube
Charge Filament Commu AEC
Stator Exposure
Circuit Circuit IGBT- DRV- Filament nication Circuit
Circuit HV-DRV time error
error error OCP PER - OLP error error
error

To be checked Normal Normal Normal Normal

3. Error type 3: Displayed E r r C r P at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be checked
To be checked Normal Normal Normal Normal
(Any one of 4 is possibly faulty)

4. Error type 4: Displayed E r r C r at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error

To be checked Normal Normal Normal Normal Normal

(Rev. 1.0) 46 Service Manual


EcoRay PX-300HF

5. Error type 5: Displayed E r r C FP at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be To be To be checked
Normal Normal Normal Normal
checked checked (Any one of 4 is possibly faulty)

6. Error type 6: Displayed E r r C F at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be To be
Normal Normal Normal Normal Normal
checked checked

7. Error type 7: Displayed E r r C P at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be To be checked
Normal Normal Normal Normal Normal
checked (Any one of 4 is possibly faulty)

8. Error type 8: Displayed E r r CHA at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal
checked

9. Error type 9: Displayed E r r r FP at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be checked
Normal To be checked Normal Normal Normal
(Any one of 4 is possibly faulty)

(Rev. 1.0) 47 Service Manual


EcoRay PX-300HF

10. Error type 10: Displayed E r r r F at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error

Normal To be checked Normal Normal Normal Normal

11. Error type 11: Displayed E r r r P at kVp, mA, Time section

Tube Refer ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be To be checked
Normal Normal Normal Normal Normal
checked (Any one of 4 is possibly faulty)

12. Error type 12: Displayed E r r roto at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal
checked

13. Error type 13: Displayed E r r FP at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be To be checked
Normal Normal Normal Normal Normal
checked (Any one of 4 is possibly faulty)

14. Error type 14: Displayed E r r FILA at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal
checked

(Rev. 1.0) 48 Service Manual


EcoRay PX-300HF

15. Error type 15: Displayed E r r 232 at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal
checked

16. Error type 16: Displayed E r r AEC at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal
checked

17. Error type 17: Displayed E r r SEC at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal
checked

18. Error type 18: Displayed E r r Pro----O at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal Normal Normal Normal
checked

19. Error type 19: Displayed E r r Pro--I-- at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal Normal Normal Normal
checked

(Rev. 1.0) 49 Service Manual


EcoRay PX-300HF

20. Error type 20: Displayed E r r Pro---P- at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal Normal Normal Normal
checked

21. Error type 21: Displayed E r r Pro----F at kVp, mA, Time section

Tube Refer to ECO-XR5-CnD Circuit


Charge Filament Commu AEC
Stator Exposure
Circuit Circuit Filament nication Circuit
Circuit HV-DRV IGBT-OCP DRV-PER time error
error error - OLP error error
error
To be
Normal Normal Normal Normal Normal Normal Normal Normal Normal
checked

22. Error type 22: Displayed E r r Inte at kVp, mA, Time section
1. Inter Lock signal trouble and circuit trouble
2. Inter Lock is Door Open, Tube Temperature, Bucky start signal, Emergency Stop switch.

(Rev. 1.0) 50 Service Manual


EcoRay PX-300HF

5.3 Trouble Shooting

5.3.1 Charge & Discharge circuit error (Check ECO-XR5-CnD Board)


1. Consist of part

(Rev. 1.0) 51 Service Manual


EcoRay PX-300HF

2. Check Procedures 1
(1) Board Name: ECO-XR5-CnD
(2) Functions: This board controls charge/discharge of Main Condenser.
(3) Followings are the check procedures for ECO-XR5-CnD board when charge/discharge trouble.

Under normal condition Under defective How to maintain

Provide main input power to 1. Check Connector (P803) and


generator and check if LED
No light at LED measure the voltage (Normal: 220VAC)
(804) is ON.
*Before the power ON at the (D804) 2. Check the Fuse F801.
console. 3. Measure the voltage TP801 and

U801 Heat sink (Gnd)

(Normal: 21VDC ±0.5V)

Console Power ON.

Standby lamp 1. Check LED (D816) lightening.

flickering then (Normal: Red light)


2. Check the Connector (P805) and
Standby lamp is flickering of turns yellow and
measure the voltage on it.
Console. error code
(Normal: 310VDC between pin 1 & 4)
displayed on the
3. Check the Connector (P804) and
FND measure the voltage on it.
Standby lamp turned to (Normal: 220VAC between pin 1 & 2)
green and LCD display goes 4. Check the Connector (P802).

enabled. 5. Measure the voltage between Test


point TP803 & TP804
(Normal: +5VDC ±0.2V)
In case of Relay 6. Measure the voltage between Test

(RL801, 802 do point TP803 & TP805


(Normal: +12VDC ± 0.2V)
Operating is ready. not work
7. Check the function of Relay (RL801)
correctly, check
Normal: it works when OP console
the frequency at
PWR is ON and stops when main
Test point (TP802,
capacitor charges about 90%
803) and make 8. Check the function of Relay (RL802)
Chapter 4.1.3.1 is describing
sure if it is 75kHz Normal: RL802 starts to work
how to adjust the frequency. or not. immediately when RL801 stops.

(Rev. 1.0) 52 Service Manual


EcoRay PX-300HF

(4) Board Name: ECO-XR5-CnD


(5) Functions: This board controls charge/discharge of Main Condenser.
(5) Adjust frequency between Test point TP803 and TP802.

Under normal condition Under defective How to maintain

OP console Power On. In order to readjust input power freq....


1. Short circuit of P801 pin1 & pin2.
2. Tester probe connection to TP803
(Gnd) and TP802.
Study the function of P801
3. Read the detected value.
at ECO-XR5-CnD Board. Freq. readjustment
4. Adjust VR801 to meet 75KHz
(deviation allowed: 75KHz ±0.5KHz)
5. with PSU: 76-78KHz
P801 function of Connector (deviation:±1KHz )
P801 pin 1: Voltage input
freq. selective terminal (L.
volt)
P801 pin 2: Common ground
P801 pin 3: freq. Selective
Charging freq. check Measure the charging power frequency.
terminal for charge (C. volt)

In order to adjust charging power freq..


Measuring input power freq.
1. Short circuit of P801 pin2 & pin3.
1. Connect pin 1 & 2 of
2. Tester probe connection to TP803
P801.
(Gnd) and TP802.
2. Tester probe connection Freq. readjustment 3. Read the detected value.
to TP803 (GND) and TP802.
4. Adjust VR801 to meet 76KHz
3. Read the detected value.
(deviation allowed: 76KHz ±0.5KHz)
4. Normal value: 75KHz
5. with PSU: 76-78KHz
(with PSU: 76-78KHz)
(deviation:±1KHz)

Check charge frequency after adjustment of


Measuring charging power
freq. input power frequency.
1. Connect pin 2 & 3 of
P801.
2. Tester probe connection
to TP803 (Gnd) and TP802
3. Read the detected value. 1. Check input power and charging power freq after adjustment of
4. Normal value: 76KHz charging frequency.
(with PSU: 76-78KHz) 2. System turn off and on again, and measure the frequency.
3. Open P801 pin 1, 2, 3 if input power frequency is normal.

(Rev. 1.0) 53 Service Manual


EcoRay PX-300HF

5.3.2 Tube Stator Circuit error (Check ECO-XR5-Mio Board)


1. Consist of part

2. Consist of Tube Stator


Temperature Relay

Tube Anode Socket Receptacle

Anode Stator wiring terminal

Anode Stator Coil terminal

M: Main wiring of X-ray tube stator.

A: Assistance wiring of X-ray tube stator.

NC: Not use

ET: Earth

Note! Never connect terminal 1 to terminal 5 or 6.

(Rev. 1.0) 54 Service Manual


EcoRay PX-300HF

3. Check procedure
(1) Board Name: ECO-XR5-Mio Board
(2) Functions: Controlling generator including Tube Stator movement.
(3) Under the tube stator trouble, check tube stator wiring condition and then check ECO-XR5-
Mio Board and check points are as follows;

Under normal condition Under defective How to maintain

Provide main power to generator 1. Check the cable connection between


HFG and X-ray tube (Tube Stator lines)
and Power ON at console.
(7-5. Tube Stator circuit error sub 2, 3)
2. Check the Connector (P901) and
voltage on it -- 220VAC is normal
3. Check the Fuse (F901, F701).
4. Measure the voltage output between
Press ‘Ready’ button on OP
Test point (TP109) and Test point
console. (TP105); 3VDC ±0.2V is normal.
5. If above voltage is not normal, adjust
VR701 to fix it at 3VDC.
6. Measure the voltage between Test
point (TP109) and Test point (TP106)
Ready lamp lightening of No Ready lamp ; 3VDC ±0.2V is normal.
console. light on of console 7. In case the voltage output is not
good, adjust VR702 to fix it at 3VDC.
8. Check the output voltage from
Diode (D701, 702) -- 7~8 VDC is
normal.

Functioning well. 9. In case the output voltage is not O.K,


check the coil winding status on
transformer T701, 702 and measure the
If you still have coil value. (see the circuit diagram for

the same problem check points)


Normal: Input pin 1 & 3 = 1
Contact service dept. (No LED light ON
output pin 4 & 5 = 1.5
with Ready)

(Rev. 1.0) 55 Service Manual


EcoRay PX-300HF

5.3.3 Tube Filament Circuit error (Check ECO-XR5-FnD Board)


1. Consist of part

2. High Voltage Cable

Measuring method and Procedure


- Must power off before checking X-ray tube filament short condition.
- Remove Cathode ( - ) from the high voltage transformer.
- Check resistance value C with S or C with L of cable plug.
- If short condition, resistance value could not check.

(Rev. 1.0) 56 Service Manual


EcoRay PX-300HF

3. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling Tube Filament
(3) Under Tube Filament circuit trouble, check high voltage and wiring condition of circuit and
then check ECO-XR5-FnD board.

Under normal condition Under defective How to maintain

Turn on the Main power and 1. Check Connector (P501) and measure
the voltage on it ; 220VAC is normal
then turn on the Console Power
2. Check Fuse (F501).
3. Fix the Relay (RL501) position and check
LED (D516, 517, 518)
its functioning.
ECO-XR5-FnD (Driver Board) Not lighting
Normal: Relay working and Ready lamp ON
Normal: All LED lighted
LED (D516, 517, 518) 4. Measure the voltage between Heat sink
with green
(U406) and Test point (M5V) Nomal: 5VDC
±0.1V
5. Measure the voltage between Heat sink
Set exposure values
(U406) and KA324 (U518 pin 4)
; 12VDC ±0.1V is normal
6. Measure the voltage between Heat sink
Error when Ready (U406) and KA324 (U518 pin 11)
Push the “Ready” button
; -12VDC ±0.1V is normal
7. In case above voltages are not good,
check input PWR of ECO-XR5-Mio (Main
Board).
8. Fix the Connector (P403) well.
X-ray switch could use after

Ready lamp lightening.


1. Check HV cables. (Anode, Cathode
directions and contact condition)
2. Tube Focal LED lightening check.
3. Check cable connections between
Could use system after repair
Connector (P401) and HV Transformer.
4. Fix the RL401 Relay of focal change, and
check its functioning
7. Voltage checks before Ready between Heat sink (U406) and (Normal: Relay On -> Large, Relay off ->
Resistor (R426). Small)
(Normal: approx.RMS 0.8VDC ±0.1V and voltage is difference at 5. Voltage check between Connector (P401)
each mA with Ready condition) pin 3 and pin 4 before Ready at Small focus
8. Before pressing Ready, measure the voltage between Heat sink (Normal: 35VDC ±2V)
(U406) and KA39 3(U519 pin3) 6. Voltage check between Connector (P401)
(Normal: 1VDC ±0.1V and this is different at each mA under Ready pin 2 & 4 before Ready at Large focus.
condition) (Normal: 35VDC ±2V)

(Rev. 1.0) 57 Service Manual


EcoRay PX-300HF

5.3.4 IGBT Device signal and Protection circuit trouble (Check ECO-XR5-FnD Board)
1. Consist of part

(Rev. 1.0) 58 Service Manual


EcoRay PX-300HF

2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check high voltage cable, filament short condition of X-ray tube, wiring condition of circuit
when IGBT Device signal and protection circuit trouble and then check ECO-XR5-FnD Board
condition as bellows.

Under normal condition Under defective How to maintain

Set exposure value 1. Power OFF before the circuit check.


2. Check resistance value between P502 pin 1 & 2, and check
Diode voltage on Driver board.
(Detection point is same and measure it at Power OFF status
only)
(Normal: resistance -> 1 ㏀ ± 50Ω,
Push the Ready button
Diode voltage -> 0.7VDC ± 0.1V)
3. Measure the resistance between P502 pin 3 & 4 and voltage
at Diode in Power OFF condition.
Error (Normal: Resistance -> 1㏀ ± 50Ω,
comes Diode voltage -> 0.7VDC ± 0.1V)
4. Measure resistance value between P503 pin 1 & 2 and Diode
Push the X-ray button when X-
voltage under Power OFF status.
ray (Normal: Resistance -> 1㏀ ± 50Ω,
Diode voltage -> 0.7VDC ± 0.1V)
exposure
5. Measure resistance between P503 pin 3 & 4 and Diode
voltage at Power OFF status.
(Normal: Resistance
X-ray exposure is OK
Diode voltage -> 0.7VDC ± 0.1V)

(Rev. 1.0) 59 Service Manual


EcoRay PX-300HF

5.3.5 HV-DRV-OLP trouble or Error Pro O- - - error (ECO-XR5-FnD Board)


1. Consist of PCB

2. Check procedure about Board


(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and check waveforms at the HV-DRV-OLP
check point with oscilloscope when IGBT Device signal and protection circuit trouble and
checked waveforms are as follows.

Error code display Normal waveform Abnormal waveform

Displayed on console

(Error Pro o---)

(Rev. 1.0) 60 Service Manual


EcoRay PX-300HF

5.3.6 IGBT_OCP trouble or Error Pro - I- - error (ECO-XR5-FnD Board)


1. Consist of PCB

2. Check procedure about Board


(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and then check IGBT_OCP trouble when IGBT
Device signal and protection circuit trouble. Check procedure is as follows.

Normal Operation Under defective How to check and repair

1. Fix the Connector (P504)


Set exposure value
(connected to IGBT current sensor)

2. Detect waveform between U519 pin 7 and

Heat sink (U406).


Error comes
3. If the waveform is nor normal, adjust
Push Ready button when X-Ray
exposure VR702 to make proper waveform.

4. This setting is always completed in

factory, so do not adjust VR702 except

educated engineers.
Push X-ray button 5. In case IGBT_OCP circuit is out of order,

please change IGBT or call Service center.

X-ray exposure is done

(Rev. 1.0) 61 Service Manual


EcoRay PX-300HF

5.3.7 IGBT_OCP trouble or Error Pro - I- - error (ECO-XR5-FnD Board)


1. Consist of PCB

2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and check IGBT_OCP trouble when IGBT
Device signal and protection circuit troubles. And check procedures are as follows.

Error code display Normal waveform Abnormal waveform

Displayed on console

(Error Pro -I--)

(Rev. 1.0) 62 Service Manual


EcoRay PX-300HF

5.3.8 DRV _ PWR _ UVP trouble or Error Pro - - P – error (ECO-XR5-FnD Board)
1. Consist of PCB

2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and the check IGBT_OCP trouble when IGBT
Device signal and protection circuit trouble. And check procedures are as follows.

Normal Operation Under defective How to check and repair

1. LED light ON check at D516, 517, 518


Set exposure value
(Normal: All LEDs are light ON)

2. Fix the Connector (P501) and check

voltage (normal : 220VAC)


Error comes
3. Check the Fuse (F501).
Push Ready button when X-ray
exposure 4. Fix the Relay (RL501) and check its

functioning (normal: Relay working with

lamp ON)

5. Measure the voltage between Test point


Push X-ray button (TP500, 501) (normal: 18VDC ±0.2V)

6. If above voltage is not O.K, adjust VR501

to meet the proper range at 18VDC.

7. In case there is no voltage out, check the

X-ray exposure is done Resistor (R502).

(Rev. 1.0) 63 Service Manual


EcoRay PX-300HF

5.3.9 DRV _ PWR _ UVP trouble or Error Pro --P- error (ECO-XR5-FnD Board)
1. Consist of PCB.

2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and then check IGBT _ OCP troubles when
IGBT Device signal and protection circuit troubles. And check procedures are as follows.

Normal Operation Under defective How to check and repair

Error comes when X-ray


Set exposure value 8. See next page about repair
exposure

(Rev. 1.0) 64 Service Manual


EcoRay PX-300HF

Normal Operation Under defective How to check and repair

Set exposure value 9. Check voltage between Heat sink (U504) and

Diode Cathode (D504) (Normal: 18VDC ±0.3V)

10. Check voltage between Heat sink (U504) and


Error comes
Diode Anode (D505) (Normal: -13VDC ±0.3V)
Push Ready button when X-ray
11. Check voltage between Heat sink (U504) and
exposure
Diode Anode (D524) (Normal: -10VDC ±0.3V)

12. Check voltage between Heat sink (U504) and

Resister (R526) (Normal: 15VDC ±0.3V)

Push X-ray button 13. Check voltage between Heat sink (U506) and

Diode Cathode (D506) (Normal: 18VDC ±0.3V)

14. Check voltage between Heat sink (U506) and

Diode Anode (D507) (Normal: -13VDC ±0.3V)

15. Check voltage between Heat sink (U506) and


X-ray exposure is done
Diode Anode (D519) (Normal: -10VDC ±0.3V)

16. Check voltage between Heat sink (U505) and

Resister (R521) (Normal: 15VDC ±0.3V)

17. Check voltage between Heat sink (U508) and

Diode Cathode (D508) (Normal: 18VDC ±0.3V)

18. Check voltage between Heat sink (U508) and

Diode Anode (D509) (Normal: -13VDC ±0.3V)

19. Check voltage between Heat sink (U508) and

Diode Anode (D521) (Normal: -10VDC ±0.3V)

20. Check voltage between Heat sink (U508) and

Resister (R523) (Normal: 15VDC ±0.3V)

(Rev. 1.0) 65 Service Manual


EcoRay PX-300HF

5.3.10 FIL _ OCP trouble or Error Pro - - F - error (ECO-XR5-FnD Board)


1. Consist of PCB

2. Check procedure
(1) Board Name: ECO-XR5-FnD Board
(2) Functions: Controlling IGBT and Tube Filament
(3) Check wiring condition of ECO-XR5-FnD Board and then check FIL _ OCP trouble when IGBT
Device signals and protection circuit trouble. And check procedures are as follows.

Normal Operation Under defective How to check and repair

Set exposure value 1. The function of FIL_OCP is to protect tube


focal spots.
2. Check voltage between Test point (TP402)
and Heat sink (U406)
Error comes
(Normal: 3.6VDC ±0.1V)
Push Ready button when X-ray
exposure 3. If the voltage is not O.K, adjust VR402 to
fix the correct voltage.
4. It is strongly recommended that VR402
is strictly prohibited from uneducated
persons. (Keeping factory setting is very
Push X-ray button important)
5. In case FIL_OCP circuit is defective, X-ray
tube or Driver board should be replaced.
Call to Service center for the further
actions.
X-ray exposure is done

(Rev. 1.0) 66 Service Manual


EcoRay PX-300HF

5.3.11 Communication error between generator and console or Err 232 error
(ECO-XR5-Mio Boar & SMS-CONB-340 Board)
1. Consist of PCB

2. Check procedure
(1) Board Name: ECO-XR5-Mio Boar and SMS-CONB-340 Board
(2) Functions: Communication transfer between generator and console.
(3) Check communication cable condition when communication problem between console and
ECO-XR5-Mio Board. And check procedures are as follows.

Normal Operation Under defective How to check and repair

1. Communication cable connection check.


Turn ON power of console
2. LED (D107, 108) light check. Normal: D107 flickering
when parameters are being set (green light)
Abnormal: D108 light ON (Red color)
3. LED (D906, 911, 912) check. Normal: D906 light on
before Power On .D911, 912 light ON after Power ON.
Self-diagnosing automatically by Err 232 appeared
Abnormal: No LED light ON after Power ON.
program on the console 4. Connector (P901) checks and measures the voltage.
(Normal: 220VAC)
5. Check the Fuse (F901)
6. Check voltage between Test point (TP109) and Test
point (TP110) (Normal: 7.5VDC ± 0.1V)
System is ready to work 7. If the voltage (TP110) is not correct, adjust VR901 to
fix it at 7.5VDC.
8. Check voltage between Test point (TP109) and Test
point (TP111) (Normal: 15VDC ± 0.1V)
9. In case the voltage from TP111 is wrong, adjust
Err 232 appeared
VR902 to fix it at 15VDC.
Set exposure value
on the console 10. Check RX signal between MAX232 (U101) pin 13 and
Test point (TP109). Normal: see the next page
(Oscilloscope waveform)
11. Check TX signal between MAX232 (U101) pin 14 and
Test point (TP109). Normal: see the next page
(Oscilloscope waveform)
X-ray exposure

(Rev. 1.0) 67 Service Manual


EcoRay PX-300HF

5.3.12 Error AEC error (ECO-XR5-Mio Boar and SMS-CONB-340 Board)


1. Connect part

2. Check procedure
(1) Check Ion Chamber connector condition.
(2) Chamber position and Bucky position must be same. (Same position)
(3) Check Number flickering or not on the Time display of console after X-ray exposure with
‘beep’ sound (4 times).
(Normal: Changed normal display condition after Number flickering on the Time display)

(Rev. 1.0) 68 Service Manual


EcoRay PX-300HF

3. Consist of part 1

4. Check procedure 1
(1) Board Name: SMS-CONB-340 Board
(2) Functions: Controlling Ion Chamber
(3) Check communication cable condition and then check circuit when communication troubles
occurred between console and ECO-XR5-Mio Board and Ion Chamber. And check procedures
are as follows.

Normal Operation Under defective How to check and repair

Turn ON power of console 1. Fix the Communication cable between OP console and

HFG

2. Fix the Connector (CCON7).


Self-diagnosing automatically by Err AEC comes 3. Fix the Connector (CCON6).
program
on the display 4. Fix the Connector (PCON1).

5. Check LED (PD1, 2, 3)

Normal: LED (PD1, 2, 3) light ON.


System is ready to be used.
6. Check LED (MD2)

Normal: MD2 is flickering with data communication.

7. Check the voltage between Connector (PCON1) pin 1

Select AEC function & 4 Normal: 7.5VDC ±0.2V

8. Check the voltage of Capacitor (PC6)-both end

Normal: 5VDC ±0.1V


Set exposure value 9. Check the voltage of Capacitor ( PC17)-both end

Normal: 3.3VDC ±0.1V

10. Check the voltage of Capacitor (PC19)-both end


X-ray exposure
Normal: 1.9VDC ±0.1V

(Rev. 1.0) 69 Service Manual


EcoRay PX-300HF

5. Consist of part 2

6. Check procedure 2
(1) Board Name: SMS-AEC Board
(2) Functions: Controlling Ion Chamber
(3) Check communication condition when communication trouble occurred between
input/output signal of Ion Chamber and circuit error. And check procedures are as follows.

Normal Operation Under defective How to check and repair

Turn ON power of console 1. Fix the communication cable between OP console and

AEC board.
2. Check the Connector (CCON1).
Self-diagnosing automatically by Err AEC comes
3. Check Connector (CCON3, 4, 5).
program on the display
4. Check the console and Connector (CON1) cable
connections.

5. Check LED (LED1, 2, 3)


System is ready to be used. Normal: LED (LED1, 2, 3) light ON

6. Check LED (LED4)

Normal: LED4 flickers during data communication

Select AEC function 7. Measure the voltage of Capacitor (PC1) -both end
Normal: 5VDC ±0.1V

8. Measure the voltage of Capacitor (PC2)-both end

Set exposure value Normal: 1.9VDC ±0.1V

9. Measure the voltage of Capacitor (PC4)-both end

Normal: 3.3VDC ±0.1V

X-ray exposure 10. Measure the voltage of Capacitor (PC17)-both end

Normal: 3.3VDC ±0.1V

(Rev. 1.0) 70 Service Manual


EcoRay PX-300HF

7. Consist of part 3

8. Check procedure 3
(1) Board Name: ECO-XR5-Mio Board
(2) Functions: Controlling Ion Chamber

Normal Operation Under defective How to check and repair

Turn ON power of console 1. Check communication cable between OP console and


Connector (P101).
2. Check cable connection between SMS-AEC
(AEC Board) and Connector (P102).
Self-diagnosing automatically by Err AEC comes
3. Check LED (D107, 108)
program on the display Normal: D107 flickers during data setting (green)
Abnormal: D108 light ON (Red)
4. Check LED (D906, 911, 912)
Normal: D906 lights ON before Power On
System is ready to be used.
D911, 912 light ON after Power On
Abnormal: No LED light ON after Power On
5. Check the Connector (P901) and voltage.
(Normal: 220VAC)
Select AEC function 6. Check the Fuse (F901)
7. Measure the voltage Test point (TP109) and
Test point (TP110)
(Normal: 7.5VDC ±0.1V)
Set exposure value
8. In case the TP110 voltage is not normal, adjust VR901
to fix 7.5VDC.
9. Measure the voltage Test point (TP109) and Test point
(TP111)
X-ray exposure
(Normal: 15VDC ±0.1V)
10. In case Test point (TP111) voltage is not normal,
adjust VR902 to meet 15VDC.

(Rev. 1.0) 71 Service Manual


EcoRay PX-300HF

9. Consist of working signal

10. Followings are standard sensitivity value by program.


Sensitivity range Value from Test point(DC Out) & Test point(CCON2 pin3)-GND
Sensitivity switch in OP console (Density) is operated from -8 to
-8 ~ +8 +8, and the voltage between Test point(DC Out) & Test point
(CCON2 pin3)-GND will move from -1.5V to -9.5V (deviation: ±0.1V)

5.3.13 Error SEC error (SMS-CONB-340 Board)


1. Error type: Caused by SMS-CONB-340 (Console Board) trouble, Membrane switch trouble,
Hand switch trouble
1. Refer to clause 4 of 4.1.3.12 about SMS-CONB-340 Board checking.
2. Check hand switch working condition.
3. Check membrane switch condition.
4. The error comes when you release the exposure switch earlier than set exposure time,
which it is not a failure.

(Rev. 1.0) 72 Service Manual


EcoRay PX-300HF

5.4 Handle Bar (SMS-TS-340 Board)

5.4.1 Compositions

Rectification diode D1 Fuse F1, F2

5.4.2 How to check and verify P6 Connector (ECO-XR5-GCBoard in HFG)

(1) Board name: ECO-XR5-GCBoard


(2) Function: Transmit and receive the signal & voltage of external devices to HFG.
(3) Followings are check sequence when ECO-XR5-GCBoard is out of order

Normal Function Faulty status Failure details and trouble shooting

Handle Bar functioning at Maintain Handle Bar 1. SMS-TS-340 Board F1, F2 fuse check

Tube stand PCB (SMS-TS-340) 2. SMS-TS-340 Board D1 rectification diode

check

3. Input power check on SMS-TS-340 Board

P6 Connector

4. SMS-TS-340 Board의 P8 wire connection

check (disconnected or not)

5. ECO-XR5-GCBoard function check

1. SMS-TS-340 Board F1, F2, F3, F4, F5


Check HFG external
Electromagnetic Fuse check
interface board
functioning 2. SMS-TS-340 Board D1, D2 rectification
(ECO-XR5-
diode check
GCBoard)
3. SMS-TS-340 Board’s P6 Connector input

power check

4. SMS-TS-340 Board의 P8 wire connection

check (if disconnected or not)


Repair
Normal functioning 5. ECO-XR5-GCBoard function check
completed

(Rev. 1.0) 73 Service Manual


EcoRay PX-300HF

Chapter 6

Circuit Diagram

High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0) 74 Service Manual


EcoRay PX-300HF

6. Circuit Diagram
6.1 Block Diagram

(Rev. 1.0) 75 Service Manual


EcoRay PX-300HF

6.2 Console Board

(Rev. 1.0) 76 Service Manual


EcoRay PX-300HF

(Rev. 1.0) 77 Service Manual


EcoRay PX-300HF

(Rev. 1.0) 78 Service Manual


EcoRay PX-300HF

(Rev. 1.0) 79 Service Manual


EcoRay PX-300HF

6.3 ECO-XR5-Mio Board

(Rev. 1.0) 80 Service Manual


EcoRay PX-300HF

(Rev. 1.0) 81 Service Manual


EcoRay PX-300HF

6.4 ECO-XR5-CnD Board

(Rev. 1.0) 82 Service Manual


EcoRay PX-300HF

6.5 ECO-XR5-FnD Board

(Rev. 1.0) 83 Service Manual


EcoRay PX-300HF

6.6 Handle Bar (ECO-XR5-HANDLE)

(Rev. 1.0) 84 Service Manual