G2 & G2 PLUS
OPERATION &
MAINTENANCE MANUAL
EASYDOSE G2 .................................................................................................. 3
INSTALLATION & MOUNTING ......................................................................... 3
UTILITY CONNECTIONS ................................................................................. 4
SYSTEM PARTS............................................................................................... 5
EASYDOSE G2 INTRODUCTION
SAFETY NOTES
This system is equipped with special devices and alarms to permit continuous
safe operation. However, it is possible that individuals may at some time
encounter liquid nitrogen in the open.
The user should be aware that liquid nitrogen has the following properties:
Clear
Colorless
Non-flammable
Extremely cold (-320F, -195C)
Will freeze skin on contact
May cause severe burns on contact
Extreme care must be taken to avoid liquid nitrogen splashing on clothing, into
shoes, or even into gloves.
Pressure relief valves are provided in critical zones to avoid damage from
excessive pressure build-up in confined areas. If additional plumbing is added,
include safety relief valves where needed.
EASYDOSE G2 INTRODUCTION
SPECIFICATIONS
EASYDOSE G2 SPECIFICATIONS
INPUT VOLTAGE 110 VAC-60 Hz/220 VAC-50 Hz +/- 10%
CURRENT 1.0 AMPS – 110 VAC / 0.5 AMPS – 220 VAC
POWER 110 WATTS
OPERATING ENVIRONMENT TEMPERATURE 5°C - 45°C
HUMIDITY RATING 0% - 100% RELATIVE HUMIDITY (NON-CONDENSING)
ALTITUDE RATING CAPABLE OF OPERATION AT ALT. TO 3050M (10000 FT)
NOISE LEVEL MAX CONTINUOUS SOUND PRESSURE LEVEL 78 dB (A)
WEIGHT (DOSER WITH CONSOLE) 66 lbs. (30 kg)
OPERATION PRESSURIZE OR INERT CONTAINERS AT ALL LINE SPEEDS
YEAR OF MANUFACTURE 2013
UNITED STATES PATENT NUMBER 5,743,096
EASYDOSE G2
INSTALLATION & MOUNTING
INCREMENTAL ENCODER
300 – 1,500
pulses/container
VBC P/N: 60005 (500 pulses) OPTIONAL
VBC P/N: 60050 (5,000 pulses) OPTIONAL
Support Bracket
By Customer
6” (152mm)
5” MAX (127mm)
EASYDOSE G2
UTILITY CONNECTIONS
3 1 To CIP
equipped OPTIONAL PREVENTATIVE MOISTURE
heater block INGRESS KIT:
DEWAR FED EASYDOSER (P/N 60302)
2
Port
A
Port B
Port A Port B
EASYDOSE G2
SYSTEM PARTS
LN2 SUPPLY
LN2 OUTLET
Nozzle:41577 Seal:32606
EASYDOSE G2 OPERATION
CONTROLLER – SCREEN NAVIGATION
LOGIN TO CONTINUE
PAST THIS POINT
USER: OPERATOR
PASSWORD: 12345
EASYDOSE G2 OPERATION
CONTROLLER SETTINGS
OFFSET DISTANCE OFFSET DISTANCE OFFSET DISTANCE
FINE TUNING FINE TUNING
Standard 3D/CIP Dosed too late
& CIP & LDV Decrease offset
Sensor Container
triggered dosed
at this at this
LINE TRAVEL
point point
3D/CIP
STANDARD/CIP
HOT FILL COLD FILL
≥160°F (71°C) <160°F (71°C)
The centerline spacing is the
Product filling temperature. distance from the center of a Selecting the appropriate block
Select "Hot" or "Cold" based container to the center of an will allow the controller to set
on product temperature adjacent container the correct system delay
Values corresponding to distance can be entered in any units, but must remain consistent for all similar values
US Patent No. 5,743,096 VACUUM BARRIER CORPORATION 7
MAY 2014 EASYDOSE G2
EASYDOSE G2 OPERATION
CONTROLLER SETTINGS
LINE
TRAVEL
CAPPER / SEAMER
FILLER
TRANSFER
PRODUCT
SENSOR
EASYDOSER
EASYDOSE G2 OPERATION
STARTUP PROCEDURE
1) Ensure all LN2 and vacuum connections are mounted/installed per the “INSTALLATION & MOUNTING” section.
2) Turn on power switch to control panel.
3) Enter parameters as outlined in “CONTROLLER SETTINGS”.
4) Open LN2 supply valve.
5) Allow Easydoser to fill with LN2.
6) Turn “Dosing” to “ON”
7) Begin conveying containers under the Easydoser. Observe for proper synchronization of the dose into the container.
4
LN2 SUPPLY
2 3
EASYDOSE G2 OPERATION
SYSTEM SHUTDOWN PROCEDURE
3 LN2 SUPPLY
1
4
`EASYDOSEG2 OPERATION
WARM-UP/PURGE PROCEDURE
1) Close liquid nitrogen supply valve
2) From the “Main Menu”, turn “Dosing” to “OFF”
3) Drain the Easydoser reservoir completely by pressing the “Drain” button on the home screen. Liquid Nitrogen will
drain through the nozzle for 6 minutes unless “Drain” is pressed again, ceasing the drain function. If one 6-minute
drain cycle does not empty the Easydoser, press “Drain” again to start a second 6-minute cycle.
4) Remove safety relief valve and attach dry nitrogen gas regulated to 10-40 psig (0.69-2.74 bar). Verify that purge
supply valve is open for 8 hours.
5) Close purge supply valve, remove regulated gas, and replace safety relief valve.
6) Remove the heater block with 3mm hex key (P/N 48070)
7) Remove the nozzle with 3/8” wrench (P/N 36296) and inspect nozzle to ensure that it is clean and dry. Replace
immediately with new nozzle, torque to 100 in/lb (11.3 Nm)
8) Re-attach heater block
PURGE CONNECTION 2
Remove safety relief valve from
Port A and attach 10-40 psig (0.7-
2.8 bar) N2 gas
1 LN2 SUPPLY
3 2
Port A
4 5
3
6 7 8 Port A Port B
3
US Patent No. 5,743,096 VACUUM BARRIER CORPORATION 11
MAY 2014 EASYDOSE G2
EASYDOSE G2 OPERATION
DRAIN / NOZZLE CHANGE PROCEDURE
1) Close liquid nitrogen supply valve
2) From the “Main Menu”, turn “Dosing” to “OFF”
3) Drain the Easydoser reservoir completely by pressing the “Drain” button on the home screen.
Liquid Nitrogen will drain through the nozzle for 6 minutes unless “Drain” is pressed again, ceasing the drain function. If one
6-minute drain cycle does not empty the Easydoser, press “Drain” again to start a second 6-minute cycle.
4) Remove the heater block with 3mm hex key (P/N 48070)
5) Remove the nozzle with 3/8” wrench (P/N 36296) & replace immediately with new dry nozzle, torque to 100 in/lb (11.3 Nm)
6) Re-attach heater block
7) Open liquid nitrogen supply valve
LN2 SUPPLY
1 7
2 3
4 5 6
EASYDOSE G2 OPERATION
TROUBLESHOOTING GUIDE
EASYDOSE G2 OPERATION
TROUBLESHOOTING GUIDE
SYMPTOM D: Ice or frost on vent of EASYDOSER
POSSIBLE CAUSE ACTION
1. Vent heater not operating 1. Ensure heater is properly connected. Check for 24VDC at
heater. If 24VDC not present check wiring to control panel. Check
resistance of heater, if not approximately 26 ohms replace heater.
See wiring diagram.
EASYDOSE G2 OPERATION
TROUBLESHOOTING GUIDE
SYMPTOM I. Programmable controller alarms/error messages
PROBLEM ACTION
1. Control panel reads: “Dosing Off” 1. On Home Screen, set “Dosing” to “ON”
5. Control panel reads: “Dose duration too long” 5. Reduce dose duration. See “EASYDOSE G2
OPERATION” section. Use larger size nozzle if
needed. Consult VBC
6. Control panel reads: “Dose duration too 6. Increase dose duration. See “EASYDOSE G2
short” OPERATION” section. Use smaller size nozzle if
needed. Note: minimum dose duration below
EASYDOSE G2 PLUS: 6 ms
EASYDOSE G2 STD: 25 ms
7. Control panel reads: “Offset distance too 7. Increase offset distance. See “EASYDOSE G2
short” OPERATION” section. Ensure “Timing Sensor” is
positioned correctly according to the EASYDOSE
G2 INSTALLATION & MOUNTING schematic
8. Control panel reads: “Service dosing valve” 8. Replace dosing valve solenoid (VBC P/N:
58707 OR 59707) & dosing valve stem (VBC P/N:
47319). Reset counter after valve has been
replaced(See “EASYDOSE G2 OPERATION”
section)
9. Control panel reads: “No active nodes found 9. Ensure data cable from PLC to touch screen is
on network”, “connector disconnected”, “Error properly connected inside control console.
612” or “####” on touch screen See wiring diagram.
11. Check encoder signal 11. Alarm is triggered if no encoder signal is
detected. Check encoder connection. See wiring
diagram.
12. Control panel reads: “Fixed Speed Dosing 12. Lower line speed below steady stream
Inhibited, Line Speed too High!” switchover value or increase steady stream switch
over value.
EASYDOSE G2 OPERATION
ELECTRICAL TROUBLESHOOTING GUIDE FOR HIGH SPEED VALVE DRIVER
(EASYDOSE G2 PLUS ONLY)
YES NO
Is Dosing Valve Output LED indicator blinking on PLC controller? Check timing sensor: Is sensor LED blinking?
NO YES YES
Disable product sensor and Check if driver input LED is Check cable wiring
recheck blinking on Dosing blinking inside control console? from sensor to control
Valve Output LED console. See wiring
YES NO NO
Set switch on driver to Check fuses in driver Check voltage for sensor: +24VDC
Test mode and see if output terminal blocks between terminals DC+ & DC- inside
dosing valve opens (2 AMP 5x20mm) control console
YES NO YES NO
Check wiring to Check fuses in output Check for 24VDC between Check wiring from
dosing valve terminal blocks PLC Timing Sensor Input & power supply to
(2 AMP 5x20mm) DC- with nothing in front of terminal strip
sensor. Check for 24VDC with inside control
object in front of the sensor. console. See
wiring
EASYDOSE G2 MAINTENANCE
ROUTINE PROCEDURES
Prior to each use: Removed and inspect heater block. For best results, remove heater block when
the dosing unit is warm. If necessary, clean block surfaces as deemed
appropriate.
25 Million Cycles: Remove and check valve stem, gland seal and operator ass’y for wear.
Lubricate shaft of valve operator with Dow 111 (VBC P/N 58590).
Bearing should slide freely over shaft of valve operator.
(See re-lubrication procedure)
50 Million Cycles: “SERVICE DOSING VALVE” message appears on touch screen. Replace
dosing valve solenoid (VBC P/N: 58707 OR 59707) & dosing valve stem
(VBC P/N: 47319). Reset counter in Advanced Settings menu after valve has
been replaced.
EASYDOSE G2 MAINTENANCE
VALVE STEM RELUBRICATION PROCEDURE
EASYDOSE dosing valve stem assembly (Standard: 58707 & PLUS: 59707)
FOR ALL ELECTRIC OPERATOR STEM ASSEMBLIES, USE VBC P/N: 58590 (DOW CORNING 111)
APPLY HERE
MOVE THE BUSHING UP AND DOWN OVER THE LUBRICANT. THIS WILL SPREAD THE
LUBRICANT EVENLY WHILE COATING THE BUSHING SEALS. REINSTALL THE VALVE
STEM IN THE EASYDOSER UNIT.
26 27
15
6&7
8&9
20
00
23
10
LDV
16
12
18
19
13
3D/CIP
15 14
15
STANDARD
21
20
22
24 STD. CIP MAINT. KIT
25 3D CIP MAINT. KIT 17 20 22
28
29 30
32
31
33 11
34
41 & 42
37
36
43
35
48
49
47
46
45 50
39
44
38
40