Professional Documents
Culture Documents
GE OEC® 9800
Operator Manual Update
and Supplement
5431291-1EN
Rev 1
© 2011
© GE OEC Medical
Systems, Inc.
All Rights Reserved
GE OEC 9800 Operator Manual Update and Supplement
Revision history
CAUTION: U.S. federal law restricts this device to sale by or on the order of a physician, veterinarian,
or other designated licensed practitioner as appropriate for its clinical use.
IMPORTANT: SAVE THESE INSTRUCTIONS. THIS SUPPLEMENT PROVIDES ALL CAUTIONS AND
WARNINGS RELATED TO THE OPERATION OF THE SYSTEM. PLEASE READ THIS MANUAL BEFORE
USING EQUIPMENT.
NOTE: The information provided in this supplement does not include all information regarding the
operation of the system. Please refer to the system operator manual(s) for complete
information regarding the safe and effective use of the system. For additional copies of the
operator manual(s), please contact GE customer service.
This manual may not be reproduced, in whole or in part, without the written permission of GE
Healthcare.
Other product and company names mentioned herein are the property of their respective owners.
The contents of this document are accurate at the time of publication. However, changes in design
and additional features can, at any time, be incorporated in the hardware and software and may
not be reflected in this version of the document. Contact GE Healthcare Technical Support for
clarification, if discrepancies arise.
This document was originally written and verified in English, part number 5431291-1ENDEV.
GE OEC Medical Systems, a General Electric company, going to market as GE Healthcare.
ii
GE OEC 9800 Operator Manual Update and Supplement
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as
unsorted municipal waste and must be collected separately. Contact an authorized representative
of the manufacturer for information concerning the decommissioning of your equipment.
The separate collection symbol is affixed to a battery, or its packaging, to advise you that the
battery must be recycled or disposed of in accordance with local or country laws. The letters below
the separate collection symbol indicate whether certain elements (Pb=Lead, Cd=Cadmium,
Hg=Mercury) are contained in the battery. To minimize potential effects on the environment and
human health, it is important that all marked batteries that you remove from the product are
properly recycled or disposed.
For information on how the battery may be safely removed from the device, please consult the
service manual or equipment instructions. Information on the potential effects on the environment
and human health of the substances used in batteries is available at this url:
http://www.gehealthcare.com/euen/weee-recycling/index.html.
iii
GE OEC 9800 Operator Manual Update and Supplement
iv
GE OEC 9800 Operator Manual Update and Supplement
Table of contents
Revision history.......................................................................................................................................................................ii
Medical device directive....................................................................................................................................................iii
v
GE OEC 9800 Operator Manual Update and Supplement
Section 4: Diagnostics
Workstation messages..................................................................................................................................................4-1
Overview ......................................................................................................................................................................4-1
Error recovery steps ...............................................................................................................................................4-1
Right monitor messages.......................................................................................................................................4-2
Left monitor messages ...................................................................................................................................... 4-10
C-Arm messages ........................................................................................................................................................... 4-12
vi
GE OEC 9800 Operator Manual Update and Supplement
vii
GE OEC 9800 Operator Manual Update and Supplement
viii
Compliance, labels, and symbols
Section 1
Compliance, labels, and symbols
1.1 Overview
This document provides all cautions and warnings related to the operation of the system as well as
additional information about your mobile imaging system. The information in this document is
provided in addition to the information found in your Operator Manuals and is provided to help you
better understand the operation of the system. This document covers a variety of items related to
upgrades made to the hardware and/or software of the Workstation and C-Arm.
1-1
Compliance, labels, and symbols
1-2
Compliance, labels, and symbols
WARNING: Electrical circuits inside the equipment may use voltages which are capable of causing
serious injury or death from electric shock. To avoid this hazard, never remove any of the
cabinet covers.
WARNING: The use of the wrong type of fire extinguisher presents electrical shock and burn hazards.
To avoid these hazards, a fire extinguisher which meets applicable regulations and
standards must be available in the room where the equipment is being used. Remember
that equipment that is equipped with batteries is a source of electrical current, even when
AC power is disconnected.
WARNING: This equipment either produces or is used in the vicinity of ionizing radiation. Observe
proper safety practices during operation.
CAUTION: The GE OEC 9800 C-Arm should only be used in conjunction with a compatible and
appropriately configured GE OEC 9800 Workstation. 9800 C-Arms and Workstations are
not compatible with components of 2600, 9600, 9900, or any other systems. Damage may
result to the system if incompatible components are connected.
1-3
Compliance, labels, and symbols
该标志表明本产品含有超过中国标准 SJ/T11363-2006 《
电子信息产品中有毒有害物质的限量要求》 中限量的有毒有害物质。
标志中的数字为本产品的环保使用期,表明本产品在正常使用的条件下,有毒有
害物质不会发生外泄或突变,用户使用本产品不会对环境造成严重污染或对其人
身、财产造成严重损害的期限。单位为年。
为保证所申明的环保使用期限,应按产品手册中所规定的环境条件和方法进行正常使用,
并严格遵守产品维修手册中规定的定期维修和保养要求。
产品中的消耗件和某些零部件可能有其单独的环保使用期限标志,并且其环保使用期限有
可能比整个产品本身的环保使用期限短。
应到期按产品维修程序更换那些消耗件和零部件,以保证所申明的整个产品的环保使用期
限。
本产品在使用寿命结束时不可作为普通生活垃圾处理,应被单独收集妥善处理。
This symbol indicates the product contains hazardous materials in excess of the limits established
by the Chinese standard SJ/T11363-2006 Requirements for Concentration Limits for Certain
Hazardous Substances in Electronic Information Products. The number in the symbol is the
Environment-friendly Use Period (EFUP), which indicates the period during which the toxic or
hazardous substances or elements contained in electronic information products will not leak or
mutate under normal operating conditions so that the use of such electronic information products
will not result in any severe environmental pollution, any bodily injury or damage to any assets. The
unit of the period is “Year.”
In order to maintain the declared EFUP, the product shall be operated normally according to the
instructions and environmental conditions as defined in the product manual, and periodic
maintenance schedules specified in Product Maintenance Procedures shall be followed strictly.
Consumables or certain parts may have their own label with an EFUP value less than the product.
Periodic replacement of those consumables or parts to maintain the declared EFUP shall be done in
accordance with the Product Maintenance Procedures.
This product must not be disposed of as unsorted municipal waste, and must be collected
separately and handled properly after decommissioning.
1-4
Compliance, labels, and symbols
产品中有毒有害物质或元素的名称及含量
Table of hazardous substances’ name and concentration.
部件名称 有毒有害物质或元素
(Component Name) (Hazardous substances’ name)
X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出 SJ/T11363-2006
标准规定的限量要求
1-5
Compliance, labels, and symbols
1.5 Labels
Depending on your system configuration, the following labels and symbols may have been added
to your system.
Located as near as possible to pinch points, these labels identify the pinch points. Keep hands and
feet away from pinch points when the C-Arm is in motion. These labels are located as shown in
Figure 1-1.
1-6
Compliance, labels, and symbols
Tables 1-1 through 1-15 identify the locations and descriptions of other labels that may appear on
your system.
Table 1-1
Label
Table 1-2
Label
Table 1-3
Label
• Warns about not touching the connectors on the back of the Workstation.
• Identifies the need to keep connectors covered when not in use.
Description • Identifies the standard that medical devices connected to the Workstation must meet.
This label applies to the connectors on the rear of the Workstation as well as to any
connectors not in use on other medical devices such as printers, etc. that are connected to
the Workstation.
1-7
Compliance, labels, and symbols
Table 1-4
Label
Table 1-5
Label
1-8
Compliance, labels, and symbols
Table 1-6
Label
Table 1-7
Label
1-9
Compliance, labels, and symbols
Table 1-8
Label
Table 1-9
Label
1 - 10
Compliance, labels, and symbols
Table 1-10
Label
Table 1-11
Label
Table 1-12
Label
1 - 11
Compliance, labels, and symbols
Table 1-13
Label
Table 1-14
Label
Table 1-15
Label
Location C-Arm rear cover label panel and on the back of the Workstation.
Indicates that the waste of electrical and electronic equipment must not be disposed as
unsorted municipal waste and must be collected separately. Contact an authorized
Description representative of the manufacturer for information concerning the decommissioning of
your equipment.
1 - 12
Compliance, labels, and symbols
1.6 Symbols
Table 1-16 through Table 1-60 identify the locations and descriptions of symbols that may appear
on your system.
Table 1-16
Symbol
Table 1-17
Symbol
Table 1-18
Symbol
1 - 13
Compliance, labels, and symbols
Table 1-19
Symbol
Table 1-20
Symbol
Table 1-21
Symbol
Table 1-22
Symbol
1 - 14
Compliance, labels, and symbols
Table 1-23
Symbol
WARNING: Removing the skin spacer may result in increased radiation exposure to
the patient. The rate of exposure increases exponentially as the anatomy
is positioned closer to the X-ray tube.
Table 1-24
Symbol
1 - 15
Compliance, labels, and symbols
Table 1-25
Symbol
Table 1-26
Symbol
Table 1-27
Symbol
Table 1-28
Symbol
1 - 16
Compliance, labels, and symbols
Table 1-29
Symbol
Table 1-30
Symbol
Table 1-31
Symbol
Table 1-32
Symbol
Table 1-33
Symbol
1 - 17
Compliance, labels, and symbols
Table 1-34
Symbol
Spot Film Label, HV Gen X-ray Control label, Image Intensifier label, Beam Limiting Device
Location Label, system nameplate/rating label (C-Arm and Workstation).
Description Indicates model/type of the system.
Table 1-35
Symbol
Spot Film Label, HV Gen X-ray Control label, Image Intensifier label, Beam Limiting Device
Location Label, system nameplate/rating label (C-Arm and Workstation).
Description Indicates date of manufacture of the system.
Table 1-36
Symbol
Spot Film Label, HV Gen X-ray Control label, Image Intensifier label, Beam Limiting Device
Location Label.
Description Indicates part number of the component.
Table 1-37
Symbol
Spot Film Label, HV Gen X-ray Control label, Image Intensifier label, Beam Limiting Device
Location Label, system nameplate/rating label (C-Arm and Workstation).
Description Indicates serial number of the component/system.
Table 1-38
Symbol
1 - 18
Compliance, labels, and symbols
Table 1-39
Symbol
Location C-Arm rear cover label panel and on the back of the Workstation.
Description See Medical Device Directive section.
Table 1-40
Symbol
Table 1-41
Symbol
Table 1-42
Symbol
Table 1-43
Symbol
1 - 19
Compliance, labels, and symbols
Table 1-44
Symbol
Table 1-45
Symbol
Table 1-46
Symbol
Table 1-47
Symbol
Table 1-48
Symbol
1 - 20
Compliance, labels, and symbols
Table 1-49
Symbol
Table 1-50
Symbol
Table 1-51
Symbol
Table 1-52
Symbol
Table 1-53
Symbol
1 - 21
Compliance, labels, and symbols
Table 1-54
Symbol
Table 1-55
Symbol
Table 1-56
Symbol
Table 1-57
Symbol
Table 1-58
Symbol
1 - 22
Compliance, labels, and symbols
Table 1-59
Symbol
Location C-Arm rear cover label panel and on the back of the Workstation.
Description See Medical Device Directive in this document.
Table 1-60
Symbol
1 - 23
Compliance, labels, and symbols
1 - 24
Workstation updates
Section 2
Workstation updates
CAUTION: A dedicated AC power line is recommended to avoid conflicts with the power requirements
of other equipment.
The GE OEC Medical Systems, Inc. authorized field representative shall provide the owner of the
system with information as to which rooms, in the area designated to him by the owner, meet the
requirements of line voltage, line current, and line regulation for the system. It is the owner's
responsibility to operate the system only in those rooms which meet these requirements.
This information must be obtained at the time of system installation on the site. If this information is
not available for the room in question, it can be obtained by a qualified technician or service
representative.
Failure to meet proper power requirements results in a violation of the U.S. Federal Performance
Standards 21CFR 1020.30-32.
CAUTION: Power to the system is not removed until the Workstation is unplugged from the electrical
outlet or the main breaker is turned off. To make sure power is disconnected, verify that
the green light on the lower left back panel of the Workstation is off.
2-1
Workstation updates
CAUTION: At initial setup or when removing from storage, stabilize for 24 hours at ambient
temperature and humidity before applying power. Failure to observe this warning will
result in damage to equipment.
CAUTION: Data loss and system failure can occur if the system startup sequence is interrupted.
Use this checklist to prepare the GE OEC 9800 System for transport to another area for use or for
storage.
NOTE: Refer to “Setup and Positioning” in the GE OEC 9800 Workstation Operator Manual and
“Mechanical Positioning” in the GE OEC 9800 C-Arm Operator Manual for detailed information
on moving the Workstation and C-Arm.
CAUTION: Familiarize yourself with the location and mechanical operation of all controls prior to
moving the system.
WARNING: Cords and cables connected to the Workstation and C-Arm can become trip and/or snag
hazards in the work area. Make sure that all cords and cables are routed to eliminate trip
and/or snag hazards. Personnel working around the Workstation and C-Arm should be
aware of all cords and cables whenever they move around the system or the system is
moved.
WARNING: Two people should maintain control of the Workstation or C-Arm when moving the it up
or down an incline. Do not move the Workstation or C-Arm over inclines greater than 10
degrees. Do not move the Workstation or C-Arm up or down stairs or steps. Do not lock
the Workstation or C-Arm in place on an incline greater than 5 degrees.
CAUTION: To avoid losing control of the C-Arm, always reduce transport speed before moving the
steering handle out of the 0 degree position.
2-2
Workstation updates
1. Make sure all cables and connectors are properly stored on the Workstation and C-Arm.
2. Ensure that the system is properly shut down by implementing the procedures detailed in
the Shutdown sequence in this section.
3. Disconnect and store the Workstation power cord and C-Arm interconnect cable by
implementing the procedures detailed in the Shutdown sequence in this section.
4. On the C-Arm, lock all movable mechanical assembly brakes including:
• the C-Arm orbital rotation brake
• the flip flop brake
• the wig wag brake
• the L-arm and horizontal cross arm brakes
5. Make sure all handles on the Workstation and C-Arm are secure.
6. Make sure the brakes on the C-Arm and Workstation are working correctly and set for
transport prior to moving the system.
7. Make sure the steering handles on the C-Arm are operating correctly.
8. Move the C-Arm by unlocking the wheel brakes and pushing on the steering handles.
Make sure that the wheels roll smoothly.
9. Move the Workstation by placing the Workstation wheel brakes in transport position and
pushing on the Workstation handles. Make sure that the wheels roll smoothly.
10. Once you have reached your destination, set the wheel brakes on the Workstation and
on the C-Arm.
CAUTION: Data loss and system failure can occur if the system startup sequence is interrupted.
CAUTION: Do not shut the system down or unplug the power cord during startup. Interrupting the
system startup sequence may result in corrupted data. Wait for the system to finish startup
before shutting down or unplugging the system.
2-3
Workstation updates
CAUTION: Data loss can occur if the system startup sequence is interrupted. Do not press the power
switch or unplug the Workstation until system startup is complete. System startup is
complete when the log on screen or the Patient Information screen is displayed on the
right monitor. Wait for the system startup sequence to complete before pressing the power
switch to shut down the system.
Use the following checklist when preparing the GE OEC 9800 System for use.
1. Place the C-Arm and Workstation in a position that leaves the various cables and
connectors easily accessible and from which the system can easily be moved into
position for use.
2. Check the cables and connectors for breaks, wear, bent pins, or bent prongs, or any other
damage. Do not use worn or damaged cables.
3. Plug the interconnect cable from the Workstation into the interconnect socket on the
C-Arm.
CAUTION: If the connector does not lock in place, unreliable system operation may result.
CAUTION: Plugging the interconnect cable into the Workstation after the Workstation power cord is
plugged into an outlet, or after the Workstation has been powered up can cause loss of
functionality or system damage. Always follow the startup checklist sequence.
4. Plug the handswitch and/or footswitch into the appropriate socket(s) on the C-Arm
interface panel.
5. If you are using a 9800 C-Arm with motorized drive, plug the Remote User Interface (RUI)
into the RUI connector on the C-Arm interface panel. Refer to “Connecting the Remote
User Interface” and “Orientation” in the GE OEC 9800 Motorized Drive C-Arm Operator
Manual Supplement for detailed instructions on connecting the C-Arm and RUI.
6. Plug the Workstation power cord into an appropriate wall socket.
WARNING: Cords and cables connected to the Workstation and C-Arm can become trip or snag
hazards in the work area. Make sure that all cords and cables are routed to eliminate trip
or snag hazards. Personnel working around the Workstation and C-Arm should be aware
of all cords and cables whenever they move around the system or the system is moved.
2-4
Workstation updates
7. Press the power switch on the Workstation to start the system (see Figure 2-1). Depending
on the system software version, the system may display the initialization message
(Figure 2-2). The C-Arm initializes simultaneously.
2-5
Workstation updates
Should system initialization fail, an initialization failure message is displayed (Figure 2-3).
In this event, the system is unusable. Call for service.
CAUTION: To avoid potential data loss or hard disk failure, do not shut down the system while the
initialization message is displayed.
8. Following system initialization, either a log in screen is displayed if the system is password
protected, or the Patient Information screen appears.
NOTE: If the Workstation fails to start normally, turn the Workstation off and check all the cable
connections and the circuit breakers on the rear bottom of the Workstation. Restart the
Workstation. If the Workstation still fails to start, call for service.
9. Position the C-Arm
10. Check C-Arm operation.
A) Move the C-Arm through its complete orbital and rotational range of motion. Make
sure that binding or rough motion does not impede smooth operation.
B) If you are using a 9800 Motorized Drive C-Arm, squeeze the collision detector and
verify that a collision message is displayed and that all motion is stopped.
11. If you are using an RUI with a 9800 Motorized Drive C-Arm, move the system into
position for use and orient the RUI to the C-Arm and the patient. Refer to “Orientation” in
the GE OEC 9800 Motorized Drive C-Arm Operator Manual Supplement for detailed
instructions on orienting the C-Arm and RUI.
2-6
Workstation updates
2-7
Workstation updates
CAUTION: If a Cine run is playing back, this step is highly recommended to prevent the potential hard
drive failure while the system shuts down.
1. Prior to powering off the system, exit all screens until the Main Screen (Figure 2-4) is
displayed. Wait one minute.
2. Return all moving assemblies to their most compact positions. Lower the C-Arm vertical
column and retract the cross-arm.
3. Systems with L-Arm only: Orient the L-arm vertically and down.
4. Systems with a motorized C-Arm only: Move the Image Intensifier (II) to the 0,0 position
(vertical with no orbital or rotational deflection).
5. Press the power switch on the front of the system. The switch's green LED may flash one
or more times, for up to 10 seconds, during the shutdown sequence.
6. Disconnect the Workstation power cord from the AC outlet, coil it, and store it on the
Workstation.
7. RUI with a 9800 Motorized Drive C-Arm only: Disconnect the RUI from the system before
you move the C-Arm to another location. Failure to disconnect the RUI can result in
damage to the RUI and the RUI connector.
8. Store the cables and switches on the C-Arm.
9. Disconnect the interconnect cable, coil it, and store it on the Workstation.
Shutdown sequence is complete.
2-8
Workstation updates
CAUTION: If the system fails to shut down normally, call for service.
CAUTION: Unplugging the interconnect cable from the Workstation before the Workstation has been
powered down and unplugged from the outlet can cause loss of functionality or system
damage. Always follow the shutdown sequence.
NOTE: The shutdown sequence cannot be aborted. You must wait for it to finish before you can restart
the system.
CAUTION: When you turn the system off, always wait at least 10 seconds before powering the system
back on. This will help prevent operational problems. Startup failures can result when
restarting the Workstation if you do not wait 10 seconds after shut down is complete before
restarting the Workstation.
If the 10 second interval is not allowed before restart, the system may fail to restart
properly. Several error messages or other symptoms may occur during the restart
including an inoperable right monitor touch screen, or disabling of X-rays.
If these symptoms occur, allow the system to startup completely, displaying either the log
on or the Patient Information screen, and then press the power switch to allow the system
to shut down properly. Wait the required 10 seconds after shut down and restart the
system using the power switch.
After any anomalous power event, such as a power surge, momentary power outage, etc.,
the system should be completely shut down and rebooted to ensure that it is in normal
operating condition.
Should the system reboot automatically due to an anomalous power event, allow the
system to initialize fully, shut the system down, wait 10 seconds and reboot.
1. Close all Workstation screens until the Main Screen is displayed. Wait one minute to
ensure that all Workstation processes have been completed.
2. Press the power switch on the front of the system.
3. Wait 10 seconds.
4. To restart the Workstation, press the power switch on the front of the system.
CAUTION: To ensure proper reboot and function of the system, wait 10 seconds before turning power
back on. If problem persists, call service.
2-9
Workstation updates
WARNING: Site specific procedures should be in place for handling the patient in case of imaging loss
or the loss or other Workstation or C-Arm functions during an exam.
2 - 10
Workstation updates
System lockups
Should the system unexpectedly become unresponsive or if it locks up, restart the Workstation to
return it to its normal operating condition. Refer to “Restarting the system” in this supplement for
detailed instructions. If the system fails to respond or continues to lock up, call for service.
Communication failure
If a COMM FAILURE message displays on the C-Arm control panel, press the Fast Stop button on the
back of the Control Panel Housing to disable X-rays. Press the power switch to shut down the
system. Wait 10 seconds before pressing the power switch again to restart the system. Should this
condition persist, shut down the system and call GE OEC service.
Image freeze
If the image on the left monitor “freezes,” press the power switch to shut down the system, wait 10
seconds and press the power switch again. Should this condition persist, shut down the system and
call GE OEC service.
Initialization failure
For systems equipped with software version 30 and above (press the Help key to display the
software versions screen), the Workstation displays a system initialization screen showing the
initialization progress. Once the initialization message closes, either the Patient Information screen
or the Login screen displays, and the system is ready for use. Should initialization fail, an
Initialization Failure message displays, indicating that patient data and image loss may have
occurred. If this Initialization Failure message displays, shut down the system and call GE OEC
service.
No boot
Do not interrupt system power until the system has fully completed the boot cycle. If the system
fails to boot within 3 minutes, shut down the system, wait 10 seconds, and then press the power
switch again. Should this condition persist, shut down the system and call GE OEC service.
No X-ray
If the system fails to take X-rays, turn off the system following the shutdown procedure. Wait 10
seconds, and then turn on the system. Should this condition persist, shut down the system and call
GE OEC service.
2 - 11
Workstation updates
2.7 Keyboard
The following information is in addition to the descriptions of the keyboard keys found in the GE OEC
9800 Workstation Operator Manual. For complete descriptions of the keys and their functions refer
to the Operator Manual.
2 - 12
Workstation updates
2.8.2 Customization
Customizations or changes to the Workstation can take up to one minute to save. If the system is
rebooted before the changes have been saved, the changes must be re-applied.
1. ARCNET connection
2. Parallel printer port (standard 25-pin D type)
connector
3. RS-232 serial communication (standard 9-pin D
type) connector
4. Left monitor standard video EIA 170 or CCIR
output (standard resolution)
5. Left monitor fast scan video (high
resolution)
6. Right monitor fast scan video (high
resolution)
7. Direct digital printer interface (DDPI 37-pin D
type) connector
8. Room control connector (Input/Output switch/
relay control for injector, room, door)
9. Ethernet connection
Figure 2-6 Workstation external connections
WARNING: Do not touch connectors and patient simultaneously. Electric shock could occur.
2 - 13
Workstation updates
WARNING: Connector covers must be in place when connectors are not in use to avoid inadvertent
connector contact.
WARNING: Connect only medical devices that comply with UL/IEC 60601 Standards. Contact GE
Healthcare for a list of compliant devices. Connecting a non-compliant device could cause
injury to the operator, the patient, or damage to the equipment. Connect only shielded
cables or cables supplied by GE OEC Medical Systems. When connecting accessories verify
normal operation in the configuration the accessory will be used. Clean the external
connector area with dry cloths only. Do not use liquid cleaners on this area.
2 - 14
Workstation updates
2.9 Footswitches
CAUTION: The basic two- and three-pedal footswitches used with the 9800 Workstation may not
operate as expected when paired with a Workstation model for which they were not
intended.
The three-pedal footswitch is intended for use with cardiac systems.
When a three-pedal footswitch is attached to a non-cardiac system, it
functions as a two-pedal footswitch using the first two switches and the
mode switch. In this case, the third switch will cause a beep to sound
when pressed but will have no functionality.
The two-pedal footswitch is intended for use with non-cardiac systems.
When a two-pedal footswitch is connected to a cardiac system,
functions usually assigned to the third pedal of a three-pedal footswitch
(e.g., Digital Cine Pulse) are unavailable. However, the second footswitch
can be manually configured for Digital Cine Pulse using the mode
screen. For information on setting switch modes refer to “Imaging” in the
GE OEC 9800 Workstation Operator Manual.
2.10 Handswitch
CAUTION: Do not stretch the hand control cord further than 3.048 m (10 feet). This may result in
damage to the cord. If the cord is damaged and touches the floor, it is a safety hazard. Call
the Communications Center for assistance.
2 - 15
Workstation updates
CAUTION: Simultaneously pressing the Fluoro ( ), HLF ( ), and Mode ( ) switches on a three-
pedal footswitch causes a digital cine shot to begin. Take precautions to avoid
simultaneous pressing of these switches by keeping the footswitch area clear of objects
that could cause one or more of these switches to be pressed accidentally.
Occasionally, multiple switches are pressed during a single shot. Pressing of multiple switches
during a single shot is not recommended. However, the system accommodates multiple switch
presses as shown in the following table. Multiple switch presses can happen using the foot or hand
switches or X-ray switch on the C-Arm control panel housing, or by pressing a combination of those
switches. The switches are marked as follows:
Fluoro Switch on both foot and hand switches.
• If a fluoro switch is pressed (on the footswitch, handswitch or C-Arm) while another fluoro
switch is being pressed, the fluoro shot will continue until both fluoro switches have been
released.
• If an HLF switch is pressed (on the footswitch or handswitch) while another HLF switch is
being pressed, the HLF shot will continue until both HLF switches have been released.
• If an HLF switch is pressed while a fluoro switch (or digital cine pedal on a 3-pedal
footswitch) is being pressed, an HLF shot will not begin until the fluoro switch has been
released.
• If a fluoro switch is pressed while an HLF switch (or digital cine pedal on a 3-pedal
footswitch) is being pressed, a fluoro shot will not begin until the HLF switch has been
released.
• If the digital cine pedal is pressed (on a 3-pedal footswitch) while a fluoro or HLF switch is
being pressed, a digital cine shot will begin.
2 - 16
Workstation updates
1. The time required to obtain the proper X-ray technique. This time is up to 1 second on the
first exposure of a particular view only. For subsequent exposures of the same view, this
portion of the delay is typically less than .25 second.
2. The Fluoro Noise Reduction (FNR) filter setting. Higher FNR settings will result in longer
extensions of the exposure. Default FNR setting results in approximately 0.7 seconds of
extension for processing.
The total exposure extension is limited unconditionally to 1.25 seconds.
WARNING: This feature can be disabled by a Field Engineer but could result in poor image quality and
increased exposure to patient.
Attempting to take very short exposures in quick succession may not operate as expected.
This could result in degraded image quality of the initial or subsequent exposures.
With Extended Exposure disabled, X-rays may continue to be produced following the
release of a momentary X-ray switch press until the desired technique has stabilized.
With Extended Exposure enabled, X-rays may continue to be produced until the system
finishes processing the image to produce a stabilized image.
2.14 Cleaning
WARNING: Always turn the system off and disconnect power before cleaning. Use a slightly damp
cloth or sponge for cleaning.
Follow these guidelines when cleaning the system:
1. Clean the covers and panels periodically using a lightly dampened cloth or sponge.
2. Do not use solvents which may damage or discolor paint finishes or plastic components.
3. Take care not to drip liquids where they can enter electronic assemblies through panel or
cover seams.
4. Clean the external connector area with dry cloths only. Do not use liquid cleaners on this
area.
2 - 17
Workstation updates
WARNING: The X-ray system is not rated for water-tight operation. Water, soap, or other liquids, if
allowed to drip into the equipment, can cause electrical short circuits leading to electric
shock and fire hazards.
If liquids spill into the Workstation or C-Arm electronics, properly shut down the system
and then disconnect the power cord. Do not operate the system until it can be cleaned and
inspected by a qualified service engineer.
Workstation
The following cleaning agents are approved for cleaning the Workstation and C-Arm:
• Virex II 256
• Precise® Hospital Foam Cleaner Disinfectant
• MadaCide-FD
• Bleach and Water in a 1:25 mixture
• Isopropyl alcohol
NOTE: Misty Clear Lemon 10 Disinfectant, Misty Multi-Purpose Disinfectant, and Misty Multi-Purpose
Disinfectant II are no longer approved cleaning agents.
Monitors
The following cleaning agents are approved for cleaning the Workstation monitors:
• MadaCide-FD
• Precise® Hospital Foam Cleaner Disinfectant
• 5.25% sodium hypochlorite diluted between 1:10 and 1:100 with water
• Isopropyl alcohol
2 - 18
Updated Workstation screens and functions
Section 3
Updated Workstation
screens and functions
3.1 Overview
Since publication of the GE OEC 9800 Workstation Operator Manual, several screens have been
changed or updated. This section describes screens that may appear on the Workstation monitors,
depending on the system software version.
3.2 Screens
3-1
Updated Workstation screens and functions
2. Enter the password assigned to the Workstation and select OK. The Workstation finishes
starting up or closes the screen saver and is ready for use.
NOTE: Until the correct password is entered and accepted by the Workstation, X-rays and all switches
and remotes are disabled.
When you have finished using the Workstation you can log off by pressing the log off key ( ) or
simply by shutting down the Workstation. When the log off key is pressed, a confirmation dialog box
appears. Select OK to log off or Cancel to return to work on the Workstation.
NOTE: If the log off key is pressed multiple times while the log off in Progress screen is displayed,
system functionality is disabled.
NOTE: If the Workstation is not configured to use a password, the log off key is disabled.
3-2
Updated Workstation screens and functions
If you begin taking X-rays: The Patient Information screen closes automatically and the
Main screen is displayed. The patient you have just entered becomes the current patient.
If you select Exit: The patient you have just entered becomes the current patient, the
Patient Information screen closes, and the Main screen is displayed. The system is ready
for you to begin taking X-rays.
If you select New Exam: The patient information you have just entered is saved in the
Scheduled Exams list and the Patient Information screen is cleared.
NOTE: Each time you choose New Exam or Scheduled Exams, the C-Arm settings are reset to the
system default parameters. This includes the cine overwrite warning, fluoro timer, Contrast/
Brightness, noise filtering, and mode settings. The collimator leaves and iris adjust to the
position selected on the Customize C-Arm screen.
C-Arm settings are not retained with the exam information. When the current patient is
changed, any imaging parameters that are set will be reset to the system default values. When
returning to re-image the patient, any parameters you were using for that patient must be re-
specified. See “System Setup” in the GE OEC 9800 Workstation Operator Manual for any
settings that can be pre-determined for default.
Current patient
Before images are taken you must select a current patient record to receive the images. The system
creates an unnamed file. The current patient is selected in one of the following ways.
• Press the Scheduled Exams button to select a patient record from the Scheduled
Exams screen.
• Resume an exam by selecting the Resume Exam (or Saved Exams) button and
selecting a patient record from the Select Current Patient screen.
3-3
Updated Workstation screens and functions
3-4
Updated Workstation screens and functions
If some images are moved from an original study to a new study, all dose summary information
remains with the original study. If all images are moved from an original study to a new study, all
dose summary information moves to the new study and then the original study is deleted.
The Scheduled Exams screen displays an alphabetical list of patient records that have
been entered in the Patient Information screen and saved with the New Exam button and
for whom no images have been stored on the system. To sort the list by physician, select
the By Physician button on the screen. If the list is longer than can be displayed on a
single screen, use the arrow keys on the screen to scroll through the list.
3. Select the desired patient from the list.
4. Select the OK button to return to the Patient Information screen and to make the selected
patient the current patient. The system is now ready for you to begin taking X-rays.
3-5
Updated Workstation screens and functions
NOTE: Each time you choose New Exam or Scheduled Exams, the C-Arm settings are reset to default
parameters. This includes the cine overwrite warning, fluoro timer, Contrast/Brightness, noise
filtering, and mode settings. The collimator leaves and iris adjust to the position selected on
the Customize C-Arm screen.
C-Arm settings are not retained with the exam information. When the current patient is
changed, any imaging parameters that are set will be reset to the system default values. When
returning to re-image the patient, any parameters you were using for that patient must be re-
specified. See “System Setup” in the GE OEC 9800 Workstation Operator Manual for any
settings that can be pre-determined for default.
CAUTION: Do not shut down the system while DICOM query is in process. Interrupting the DICOM
query may result in corrupted data.
NOTE: Taking an X-ray while transferring data to a DICOM server will cause the transfer to fail, and
the DICOM feature may be disabled. The image will be taken correctly, but it may be necessary
to reboot the system to re-enable DICOM functionality.
3. The Update Schedule operation can retrieve up to 500 records. You can use the Schedule
Filter button to limit the number of records that are displayed. To display the Schedule
3-6
Updated Workstation screens and functions
Filter screen, select the Schedule Filter button from the Patient Information screen. A
screen similar to the screen in Figure 3-5 is displayed.
4. Enter information into the Schedule Filter screen to search for patient records that meet
the filter criteria. Choose one of the following Modalities as designated by your IT
requirements:
–RF: X-ray Radio Fluoroscopic Image Storage
–XA: X-ray Angiographic Image Storage
–SC: Secondary Capture Image Storage
–ALL: Select ALL unless querying for a specific modality
NOTE: No error checking is done for date or time values, so results from the server may vary.
5. Check the Get Today's Exams box, or enter patient and/or physician information for a
refined query.
The lower half of the Schedule Filter screen provides a way to refine or filter your search
of the DICOM work list. Multiple information entries may be used to refine your filter.
6. Enter the specific filtering criteria needed.
–Patient Name: Enter patient's name according to DICOM Format: Last^First^MI.
Use a ^ between the names, not a comma. (i.e.: jones^jason^h^jr.).
If supported by the server being queried, use * for a general
search of a name. (i.e.: Jones*) Check the server's DICOM
Conformance Sheet for more information.
–Patient ID: Enter the patient ID.
–Accession #: Enter the accession number.
3-7
Updated Workstation screens and functions
NOTE: The Accession field is always enabled on the DICOM query directory screen. The accession field
may be enabled or disabled on the Patient Information screen. This is set up in the Customize
Patient Information screen.
–Study Date: YYYYMMDD. You must use this format. No spaces or other
characters other than numbers are allowed. You can enter a date
range using the following formats:
–YYYYMMDDT retrieves all studies before and including this date.
–YYYYMMDD-T retrieves all studies after and including this date.
–YYYYMMDD-YYYMMDDT retrieves all studies between and
including these dates.
–Study Time: HHMMSS. Use this format to enter the time. Use the same syntax
as that used for Study Date. Time is entered based on 24 hours.
(i.e.: 2:00 pm is entered 140000.)
–Physician: Enter the physician's name according to DICOM format:
Last^First^MI. Refer to Patient Name above for more details.
–Patient DOB: Enter all or part of the patient’s date of birth using the same
formatting as the Study Date entry.
Select the modality used with the scheduled patient(s) you wish to see.
7. Select the OK button and the Schedule Filter screen is closed. The Scheduled Exams
screen is displayed.
8. Select the Update Schedule button and the patient names, filtered according to the
criteria entered in the filter screen, are displayed.
9. Select a patient from the list.
10. Select the OK button to return to the Patient Information screen and to make the
selected patient the current patient. The system is now ready for you to begin taking
X-rays.
CAUTION: Always ensure the correct patient name is displayed on the patient information screen
before producing X-rays; otherwise, images may be stored under an incorrect patient’s
name.
NOTE: Each time you choose New Exam or Scheduled Exams, the C-Arm settings are reset to default
parameters. This includes the cine overwrite warning, fluoro timer, Contrast/Brightness, noise
filtering and mode settings. The collimator leaves and iris adjust to the position selected on the
Customize C-Arm screen.
C-Arm settings are not retained with the exam information. When the current patient is
changed, any imaging parameters that are set will be reset to the system default values. When
returning to re-image the patient, any parameters you were using for that patient must be re-
specified. See “System Setup” in the GE OEC 9800 Workstation Operator Manual for any
settings that can be pre-determined for default.
3-8
Updated Workstation screens and functions
The Select Current Patient screen displays an alphabetical list of patient records for
whom images have been stored on the system. To sort the list by date or by physician,
select the By Date or the By Physician button on the screen. If the list is longer than can
be displayed on a single screen, use the arrow keys on the screen to scroll through the
list.
3. Select the desired patient from the list.
4. Select the OK button to return to the Patient Information screen and to make the selected
patient the current patient. The system is now ready for you to begin taking X-rays.
CAUTION: In the Image Directory screen, when a patient is selected from the Saved Exams screen,
that patient does not become the current patient. Any new images taken will be added to
the current patient selected from the Patient Information screen.
The Saved Exams screen, accessible from the Image Directory screen, is used to view images from a
saved patient exam.
3-9
Updated Workstation screens and functions
1. To view images from a saved patient exam, select the Image Directory button ( ) from
the keyboard. A screen similar to the screen in Figure 3-7 is displayed.
2. Select the Saved Exams button on the Image Directory screen to display the Saved Exams
screen similar to the screen in Figure 3-8.
This screen lists patient exams that have been saved on the system. Use the arrow keys
on the touchscreen to scroll through the list.
3. Select the patient whose images you wish to review, select OK, and the system returns to
the Image Directory screen.
4. The patient information for the selected patient is displayed along with the images
associated with the selected patient.
3 - 10
Updated Workstation screens and functions
3 - 11
Updated Workstation screens and functions
3 - 12
Updated Workstation screens and functions
NOTE: The Mode Status Bar shows the setting of the footswitch buttons. See “Footswitches and
Modes” in this section for more information.
3 - 13
Updated Workstation screens and functions
3 - 14
Updated Workstation screens and functions
3 - 15
Updated Workstation screens and functions
WARNING: When the patient is positioned more than 30cm from the Image Intensifier, the dose
received by the patient is higher than the indications displayed on the Workstation.
The Air Kerma display option on the 9800 Workstation provides two indications of Air Kerma, the
Cumulative Air Kerma and the Air Kerma Rate.
• The Cumulative Air Kerma (CAK) is an indication of the accumulated Air Kerma at the
IRP for the duration of the exam. The number is at 0 at the beginning of the exam, and
increases for each X-ray exposure. The CAK value is automatically reset to 0 at the
beginning of each new exam. CAK is displayed in terms of mGy or Gy.
• The Air Kerma Rate (AKR) is an indication of the instantaneous rate of radiation.
During an X-ray exposure, the AKR displays in units of mGy/min. Otherwise, the 9800
Workstation displays the AKR as 0 mGy/min.
• The displayed AKR and CAK have an accuracy error of less than 35% for exposure
times greater than 3 seconds.
NOTE: Air Kerma is useful in predicting short-term deterministic effects from radiation. Examples of
deterministic effects include skin injury and hair loss. The United States Food and Drug
Administration, in “Recording Information In The Patient's Medical Record That Identifies The
Potential For Serious X-ray-Induced Skin Injuries,” (1995), suggests a threshold of 1 Gy, beyond
which deterministic effects might appear.
3 - 16
Updated Workstation screens and functions
NOTE: The 9800 system does NOT disable X-ray production at any radiation dose threshold.
The 9800 Workstation can display CAK and AKR using different units:
Measurement Units Conversion
CAK or Entrance Dose Gy or rads 1Gy corresponds to 100 rads.
AKR or Entrance Exposure Rate mGy/min or R/min 1Gy corresponds to 114 Roentgens (R).
3 - 17
Updated Workstation screens and functions
1. Exposure timer.
2. Air Kerma Rate (AKR)
3. Cumulative Air Kerma (CAK)
4. Dose Area Product (DAP)
5. X-ray technique factors
NOTE: The system can be configured to display AKR plus CAK, DAP, or all three dose values on the
X-ray image screen depending on local regulatory requirements and customer preferences
where allowed. If you wish to change your current settings, please contact your service
representative.
3 - 18
Updated Workstation screens and functions
1. Patient information.
2. Accumulated Dose Area Product
for the exam.
3. Cumulative Air Kermas (CAK).
NOTE: If the system is configured to
display only Air Kerma, line 2
will be empty and line 3 will
display Cumulative Air Kerma
for the exam. If the system is
configured to display both Air
Kerma and DAP, line 2 will
show DAP and line 3 will
display Air Kerma.
4. Percentage of total dose area
product in each mode or field of
view.
5. Time of exposure, in seconds, for
each imaging mode.
NOTE: In Pulsed modes, the total time that X-rays were generated displays in the Time column, rather
than the length of time that the X-ray switch was pressed.
NOTE: The Workstation can be set to display Air Kerma, Dose Area Product, or both, in this summary
screen. Contact service to change this setting.
NOTE: A printer must be connected and configured on the System in order to print patient summary
information. Refer to Select a Device in Chapter 7 of the GE OEC 9800 Workstation Manual
to select a printer, if necessary.
1. Touch Dose Summary on the Image Directory screen. The dose summary displays on the
left monitor.
NOTE: Wait 2-3 seconds for the image of the Dose Summary to stabilize on the left monitor before
proceeding with step 2.
2. At the left of the Image Directory screen on the right monitor, touch the area beneath the
printer name. This sends a screen capture of the dose summary to the printer.
3. Touch the Copy button. The image prints on your printer.
3 - 19
Updated Workstation screens and functions
The following audible alarms are provided as alerts that X-rays are being generated.
• Standard fluoro generates a pulsed audible alarm.
• High level fluoro generates a pulsed audible alarm at twice the rate as the standard
alarm.
• The Reset Alarm generates a constant alarm until Alarm Reset is pressed.
• The beginning of Film Mode is signaled by a beep. The end of the exposure is signaled
by three quick beeps.
• Digital spot generates a single audible beep followed by a second beep with a
different tone to indicate when the image is automatically saved.
3 - 20
Diagnostics
Section 4
Diagnostics
4.1.1 Overview
This section describes messages that may appear on the Workstation monitors and C-Arm control
panel display, depending on the system software version. Some of the messages are informational
messages displayed to advise the user that the system is operating correctly. Other messages are
displayed when the system fails during the startup sequence or fails during system operation.
Workstation and C-Arm failures may result in impaired system operation or automatic system
shutdown. The messages are listed in alphabetical order.
Table 4-1 through Table 4-3 catalog messages that may appear on the right or left monitor, or on
the C-Arm. Messages in the tables are followed by their meanings and the recommended actions to
be taken. Some messages listed here may not apply to your system.
WARNING: Site-specific procedures should be in place for handling the patient in case of the loss of
imaging or other Workstation or C-Arm functions during an exam. Workstation and
C-Arm functions and imaging may be recovered using the following error recovery
procedure.
Perform the following error recovery procedure if you encounter problems during start-up or
operation:
1. Read the displayed message carefully. Some messages only require the user to
acknowledge the error by selecting OK and allow the system to recover from the error
automatically.
Other messages contain instructions to shut down and restart the system. Refer to the
“Starting, Shutting Down, and Restarting the System” sections in this document for
information on properly shutting down and restarting the system.
4-1
Diagnostics
If shutting down and restarting the system fails to restore normal operation, proceed with
step 2.
2. Remove all power, quarantine the system, and call for service.
WARNING: Ignoring error and warning messages may result in equipment damage and
personal injury.
4-2
Diagnostics
4-3
Diagnostics
4-4
Diagnostics
4-5
Diagnostics
4-6
Diagnostics
4-7
Diagnostics
4-8
Diagnostics
4-9
Diagnostics
4 - 10
Diagnostics
4 - 11
Diagnostics
4 - 12
Diagnostics
4 - 13
Diagnostics
4 - 14
Diagnostics
4 - 15
Diagnostics
4 - 16
Diagnostics
4 - 17
Diagnostics
4 - 18
Diagnostics
4 - 19
Diagnostics
4 - 20
Diagnostics
4 - 21
Diagnostics
4 - 22
Diagnostics
4 - 23
Diagnostics
4 - 24
Fluoro imaging
Section 5
Fluoro imaging
5.1 Overview
This section describes the basic fluoro features of all systems. All systems provide standard
fluoroscopy modes.
Fluoroscopy mode is the only mode available on the GSP and ESP models of the Workstation.
Vascular and cardiac systems provide basic fluoro as well as vascular features. For an explanation
of vascular imaging features, refer to Section 6, “Vascular imaging” in this document.
NOTE: If the system defaults have been configured to operate at lower doses, the low dose button on
the C-Arm control panel may not provide further dose reduction. To determine if your system
defaults are configured for low dose operation, contact your field service representative to
determine if the International B ABS table is configured on your system.
5-1
Fluoro imaging
A screen similar to the screen in Figure 5-2 is displayed on the right monitor.
5-2
Fluoro imaging
Status bar
The status bar is located on the bottom right monitor display. During an exam, the configured active
mode pair is displayed on the status bar.
The status bar (Figure 5-3) shows:
1. A Fluoro shot can be initiated by pressing the left X-ray switch.
2. A High Level Fluoro (HLF) shot can be initiated by pressing the X-ray switch.
All information next to the icon describes what imaging and pulse or cine rate is applied using
the X-ray switch with that label.
All information next to the icon describes what imaging and pulse or cine rate is applied using
the X-ray switch with that label.
If film mode is selected, the film icon is displayed in the status bar (Figure 5-4).
NOTE: For producing film shots, refer to your C-Arm operator manual.
Fluoro mode displays live fluoroscopic images on the left monitor.
5-3
Fluoro imaging
During continuous HLF, the mA can increase to as much as 20 mA. During pulsed HLF the
mA can increase to as much as 40 mA. During pulsed digital cine HLF, the mA can increase
to as much as 150 mA. This can expose the patient and those working around the X-ray
field to a significantly larger dose of radiation than they would receive during fluoro or
pulsed fluoro operation. To minimize X-ray exposure hazards, use HLF with discretion.
HLF mode decreases noise level and improves image quality by increasing fluoroscopic mA.
Continuous HLF uses mA levels up to 20 mA.
Pulsed HLF ranging from 1-8 pps provides pulsed fluoro. When using HLF mode, mA can increase up
to 40 mA.
Automatic termination of HLF and pulsed HLF exposure occurs after each 30-second interval of
continuous use. Automatic exposure termination reduces unnecessary X-ray exposure and is
designed to comply with X-ray tube ratings.
5-4
Fluoro imaging
5-5
Fluoro imaging
CAUTION: Combined use of both a pulsed fluoro mode and the low dose mode may have a
compounding effect in degrading image quality.
CAUTION: Use of the low dose mode can significantly reduce the dose received by the patient;
however, image quality may be degraded in low dose mode.”
Pulsed mode generates a preset number of X-ray pulses each second while the X-ray switch is
pressed. While pulsed HLF increases radiation, use of a pulsed fluoro mode can significantly reduce
the radiation dose received by the patient; image quality, however, may be degraded in pulsed
fluoro modes.
Please refer to “Vascular Imaging” in the GE OEC 9800 Workstation Operator Manual for a
description of digital cine pulse.
When continuous mode is active, no pulse rate (pps) is displayed on the status bar (Figure 5-5).
To enable pulsed operation at the currently selected pulse rate, press the Pulse key on the C-Arm or
select the Pulse button on the Mode screen.
When pulse is enabled, the Pulse button on the Mode screen is highlighted, Pulse LED on C-Arm
control panel is illuminated, and the pulse rate (pps) is displayed on the status bar (Figure 5-6).
5-6
Fluoro imaging
2. Select a pulse rate from the pulse rate screen. The selected pulse rate displays
automatically on the status bar and the screen closes.
3. Select the Exit button on the Mode screen to return to the Main screen.
5-7
Fluoro imaging
Figure 5-10 Fluoro pulse rates for 50Hz and 60Hz systems
5-8
Fluoro imaging
2. Select a pulse rate from the pulse rate pop-up screen. The pulse rate will display on the
status bar and the pop-up screen will automatically close.
3. Select the EXIT button on the Mode screen to return to the Main screen.
For information on using the digital cine pulse rate box (Figure 5-11), refer to “Vascular Imaging” in
this document.
Save key
To save the last image held on the left monitor display, press the Save key ( ) on the Workstation
keyboard, the C-Arm control panel, or the handswitch.
An auto save feature may be enabled to automatically save the image displayed on the left monitor
when the X-ray switch is released.
See “Customize” in the GE OEC 9800 Workstation Operators Manual for information on enabling the
auto save feature.
5-9
Fluoro imaging
5 - 10
Vascular imaging
Section 6
Vascular imaging
6.1 Overview
This section explains how to produce images using vascular imaging modes. Vascular and Cardiac
systems include all standard fluoroscopy features.
This section describes vascular imaging features including:
• Subtraction
• Roadmapping
• Digital Cine Pulse
Refer to the “Fluoro Imaging” section in this document for basic fluoro information. The Fluoro
Imaging section should be read before beginning vascular imaging procedures.
6-1
Vascular imaging
6-2
Vascular imaging
With vascular active, the status bar displays Roadmap and Subtract or Fluoro and Subtract.
• Press the left X-ray pedal to initiate a Roadmap or fluoro shot.
• Press the right X-ray pedal to initiate a subtraction run.
For more information on Roadmap or Subtraction, see “Roadmap” or “Subtraction” in this section.
6-3
Vascular imaging
6-4
Vascular imaging
If the highlighted center bar is displayed next to the HLF button, the active fluoroscopy pair is Fluoro/
HLF.
If the highlighted bar is next to DIG.SPOT, select the HLF button to activate HLF in the standard
fluoroscopy group.
4. Complete the fluoro part of the exam to view the patient anatomy.
5. Press the Mode switch to display Roadmap and Subtract on the status bar.
6. Complete the Roadmap procedure. For more information on completing a subtraction procedure,
please refer to “Subtraction imaging” in this section.
6-5
Vascular imaging
WARNING: Use caution while Workstation is in use. The cine disk is always running during operation
of the Workstation. Any sudden, intense impact can damage the cine disk and could result
in loss of previously recorded cine runs and prevent further imaging.
When acquiring a cine run, the Workstation will not save the run until 5 frames have been captured.
Runs shorter than 5 frames will be discarded. At lower cine acquisition rates, it may take up to 5
seconds to record a cine run successfully. A recording has begun when a frame counter appears on
the left monitor. See “Dynamic Recording” in the 0EC 9800 Workstation Operator Manual for more
information on cine recording and reviewing.
If the CINE DISK IS NOT AVAILABLE message appears, the cine disk has failed to initialize correctly.
Turn the Workstation off, wait ten seconds, and then restart the Workstation.
NOTE: Cine runs will not playback automatically if the Cine disk is full. To play the Cine run, select
Image Directory and then select the last Cine run recorded. The Cine run will begin to play back.
If the HLF 'Acquire On/Off' checkbox is marked (Figure 6-7), the HLF acquire can be disabled by
navigating to the 'Mode' screen and 'unselecting' the 'Pulse' button on the 'Mode' screen.
NOTE: If a Cine run is reviewed and paused on the 9800 Workstation before it is transferred to DICOM,
the paused frame, rather than the first acquired frame, is the first frame displayed during
playback on the DICOM review Workstation.
6-6
Vascular imaging
6-7
Vascular imaging
CAUTION: Do not use pulsed fluoro modes for subtraction or Roadmap imaging. Use of pulsed fluoro
modes for subtraction or Roadmapping may result in poor image quality.
NOTE: The noise filtering level may only be modified during playback of a 30 FPS (25 FPS on 50 Hz)
Cine run.
X-RAY TECHNIQUE
MASK SUBTRACTION
1 2 3 4 5 6 7
TIME
CONTRAST
CONCENTRATION
The left monitor displays a uniform gray result and the system begins subtracting images
and displaying them. As long as the subtraction continues, new images are subtracted
from the mask obtained at the beginning of the cine run. As images are acquired, a
syringe icon is displayed indicating that the system is ready for contrast to be injected,
3. Contrast is injected, and the subtracted images are displayed on the left monitor. Vessels
are outlined with contrast.
4. The contrast concentration peaks.
5. Contrast concentration decreases.
6-8
Vascular imaging
6. When the footswitch pedal is released, the last subtracted image is displayed on the left
monitor.
NOTE: Releasing the X-ray pedal at any time during this procedure terminates the subtraction process.
Cine is set automatically. To change the cine frame rate, use the Cine Setup screen. Refer
to “Dynamic Recording” in the 9800 Workstation Operator Manual for more information.
3. Press and hold the ( ) X-ray pedal to begin the subtraction process and create a
mask automatically. Acquiring and creating the mask image can take up to 2 seconds.
4. After the mask is acquired, continue pressing the ( ) X-ray pedal to subtract
subsequent images from the mask. Subtraction begins when a subtracted image and a
cine frame number are displayed on the left monitor.
5. An injection icon appears on the left monitor, signaling the operator to inject a contrast
media. If an approved injector is properly connected to the Workstation, a signal is sent to
the injector.
Subtracted images are displayed on the left monitor as the contrast concentration peaks
and decreases.
6. Release the ( ) X-ray pedal to end the subtraction.
The subtraction cine run plays back automatically on the left monitor unless auto cine playback has
been disabled.
NOTE: Pressing the Swap key ( ) replaces the temporary image on the right monitor with the image
on the left monitor.
Refer to Reviewing a Cine Run in the 9800 Workstation Operator Manual for details on auto
playback. Refer to “Dynamic Recording” in the 9800 Workstation Operator Manual for a description
of the Cine Playback screen.
6-9
Vascular imaging
Registration
If the patient moves during a Subtraction run or Roadmap fluoro, misalignment of the mask and
incoming images could cause a poor image. Registration allows you to move the mask image to
produce an accurate registration by aligning the mask with the incoming images. You can realign
the mask during the procedure, or during playback if the images are recorded.
2. Use the mask Registration arrow buttons to align the mask image with the incoming
images.
3. Select reset (target button at center of Registration arrows) to restore the mask to its
original position.
6 - 10
Vascular imaging
Landmarking
In some subtracted images, it may be necessary to see the position of vessels in relation to certain
background anatomies. Landmarking allows you to vary the percentage of background anatomy
displayed in a subtracted image. This function is available for Subtraction and Roadmapping.
To vary the percentage of background anatomy:
1. Select Adjust Mask on the Main screen or the on the Cine Playback screen.
2. Select the Landmarking arrow buttons (see Figure 6-11).
3. Touch the Exit button to close the Adjust Mask screen.
NOTE: The mask is adjusted in 5% increments with each press of the Landmark button.
CAUTION: Do not use pulsed fluoro modes for Roadmapping. Use of pulsed fluoro modes for
Roadmapping may result in poor image quality.
6 - 11
Vascular imaging
6 - 12
Vascular imaging
3. Press the Mode key to activate Roadmap mode. Verify Roadmap is displayed on the
status bar.
NOTE: Use the Cine Setup screen to enable cine acquisition. Refer to “Dynamic Recording” in the
GE OEC 9800 Workstation Operator Manual for more information.
4. Press the left ( ) X-ray pedal to initiate the first phase of Roadmapping. Roadmap – 1 is
displayed on the left side of the status bar.
5. A fluoro image is displayed on the left monitor.
6. Inject contrast media when the inject icon displays on the left monitor.
7. Release the left ( ) X-ray pedal when a satisfactory image is displayed on the left
monitor and the contrast is at, or near, peak concentration in the area of interest.
8. The Roadmap mask is created automatically and displayed on the left monitor for
phase 2.
NOTE: No further motion of patient, the C-Arm, or table should occur, or the mask will become
unusable.
9. Press the left ( ) X-ray pedal again to begin Roadmapping. Roadmap – 2 is displayed
on the left side of the status bar. The saved mask is subtracted from the new fluoro
images. The resulting Roadmap is displayed on the left monitor. The X-ray technique is
locked.
10. Continue the Roadmap subtractions as many times as necessary by pressing the left
pedal. The original contrast image is used as the mask for each subtraction.
11. To exit Roadmap mode, press the Mode key.
6 - 13
Vascular imaging
6 - 14
Vascular imaging
6 - 15
Vascular imaging
WARNING: During Digital Cine Pulse, the mA can increase to as much as 150 mA. This can expose the
patient and those working around the X-ray field to a significantly larger dose of radiation
than they would receive during normal pulsed fluoro operation. To minimize X-ray
exposure hazards, use Digital Cine Pulse with discretion.
Cardiac systems
The digital cine footswitch pedal is always available to invoke digital cine
operation during a procedure, regardless of other preset mode pairs. The
digital cine rate box on the Mode screen allows the user to display and
change the preset rate. This information is not displayed on the status
bar.
To verify the selected pulse rate, display the Mode screen. Check the rate information in the Digital
Cine Rate screen (Figure 6-14). Default is 15 pps (12 pps on 50 Hz systems). Change the rate, if
necessary.
Press the right ( ) footswitch pedal to create a Digital Cine Pulse image.
6 - 16
Vascular imaging
Figure 6-15 Digital Cine Pulse rate pop-Up (50Hz and 60Hz systems)
Vascular systems
1. Verify HLF is displayed on the status bar. If Subtract is displayed, press the Mode key to
activate HLF mode. If DIG.SPOT is displayed, use the Mode screen to select HLF.
2. Select the Pulse button on the Mode screen (Figure 6-16) to enable pulsed operation. The
Pulse button will be highlighted when active.
6 - 17
Vascular imaging
3. Select the Rate button to set a pulse rate of 15 or 30 (12 or 25) pps from the Digital Cine
box on the pulse rate screen (Figure 4). Verify the selected pulse rate is displayed on the
status bar.
The frame rate displayed is the same as the pulse rate. Acquisition occurs at that rate.
Figure 6-17 Vascular pulse rate screen (50Hz and 60Hz systems)
4. Press the right ( ) X-ray pedal. The digital cine image is acquired to the cine disk
automatically.
NOTE: Pulse rate screens may vary slightly depending on the product that the Workstation was
purchased with.
6 - 18
Vascular imaging
6 - 19
Vascular imaging
6 - 20
Image directory
Section 7
Image directory
The damaged thumbnail icon is displayed when a patient image thumbnail has been damaged or
otherwise become inaccessible to the system. Even though the thumbnail is damaged, the image
itself is still available and can be displayed by selecting the damaged thumbnail icon displayed in its
place.
7-1
Image directory
CAUTION: Do not attempt to take a shot during a DICOM transfer or when copying/printing to external
media. System function is blocked during transfer procedures and any attempt to use the
system while transfer is underway may require a system reboot.
Refer to GE OEC 9800 DICOM supplement for detailed instructions on archiving images and patient
information to a DICOM server.
For more information on available storage options, please call the Communications Center or your
local sales representative. Refer to “Introduction and Safety” in the GE OEC 9800 Workstation
Operator Manual for telephone numbers.
You can archive static images from the system disk to DICOM storage. You also can archive cine
runs to the same storage device.
7-2
Image directory
1. To copy patient information and images to a storage device, press the Image Directory
key ( ) on the Workstation keyboard. An image directory screen similar to the one
shown in Figure 7-2 is displayed.
If images have been acquired for the current patient, those images are displayed. Use the
Saved Exams… button to select a different patient.
NOTE: Selecting a patient from the Saved Exams screen does not change the Current Patient. See
“Current Patient” in this document for information about changing the Current Patient.
2. Select the Copy To button from the Image Directory screen. The Copy To screen is
displayed.
3. Select the DICOM server or the medium to which you wish to copy the patient images and
select OK to return to the Image Directory screen.
4. If you are copying to a DICOM Server, options for storing patient information and images
are set when the server is defined using the DICOM Configuration screen. Refer to “Image
Review, Hardcopy, and Archive” section in the GE OEC 9800 DICOM supplement for
detailed instructions.
5. Select the images you wish to copy.
6. Select the Copy button. A message is displayed while the information is copied.
NOTE: When the copy is complete, verify that the information was correctly written to the storage
device. You can verify the information by viewing the images copied to the storage device in
the Image Directory, see “Reviewing and Copying Images” in this section for more information.
7-3
Image directory
NOTE: A printer must be connected and configured on the System in order to print patient summary
information. Refer to Select a Device in Chapter 7 of the GE OEC 9800 Workstation Manual
to select a printer, if necessary.
1. Touch the text box located at the top center of the Image Directory screen, containing
the patient 's last name. If no patient name has been specified, or patient names are
suppressed, the patient last name display s as "Unnamed. "
7-4
C-Arm notes, cautions, warnings
Section 8
C-Arm notes, cautions, warnings
8.1 Overview
This section contains information about C-Arm functions and hardware in addition to that
contained in the 9800 C-Arm Operator Manual and the 9800 Motorized Drive C-Arm Operator
Manual Supplement.
Figure 8-2 Display during Film Mode showing kVp (left) and mAs (right)
See C-Arm messages section for more information on system messages and their meanings.
8-1
C-Arm notes, cautions, warnings
8.3 Contrast/Brightness
You can use the Contrast/Brightness controls on the C-Arm control panel to select Auto Contrast/
Brightness, or manually select the contrast and brightness levels.
The Contrast/Brightness controls serve a dual function. They specify the Contrast/Brightness setting
that will be applied to a live exposure, and they allow you to modify the Contrast/Brightness setting
on a swapped or recalled image (including a Cine run).
When starting a new exam, the Contrast/Brightness setting defaults to Auto mode, and the Auto
LED is illuminated.
CAUTION: When using the film cassette holder, ensure that the film cassette is centered. Improper
centering may result in misalignment with the X-ray beam and an incomplete film
exposure may occur.
8-2
C-Arm notes, cautions, warnings
WARNING: Verify that the cassette holder is securely attached to the image intensifier. Unsecured
cassette holders may fall, injuring patients or personnel.
WARNING: Verify that the cassette is held securely within the cassette holder. Unsecured film cassettes
may fall, injuring patients or personnel.
Figure 8-4 C-Arm interface panel; non-motorized (left) and motorized (right)
The C-Arm batteries are charging whenever the C-Arm is connected to the Workstation and the
Workstation is plugged into a wall socket. The Workstation does not have to be turned on for the
C-Arm to begin charging the batteries.
Under normal use the C-Arm batteries will maintain their charge. More charging time may be
required in the following situations.
• Initial setup or return to service after an extended period of storage. Batteries may
require several hours of charging before using the C-Arm.
• Heavy use for an extended period of time. Batteries may require an overnight charge
to ensure sufficient charge for continued use.
• Whenever the C-Arm control panel displays a “low battery” message or a message
requiring additional charging time.
8-3
C-Arm notes, cautions, warnings
WARNING: Pinch points exist between the C-Arm support assembly and the C-Arm. Do not place
fingers or allow clothing in between these assemblies when positioning the C-Arm.
CAUTION: Use the handles provided on the C-Arm to position mechanical assemblies. The handles
are provided for your safety.
CAUTION: Guide the C-Arm carefully with both hands until the C-Arm is in the position you want and
then lock the brake.
CAUTION: A possible pinch point exists between the C-Arm and the tip of the front cover. Do not place
your foot on the tip of the front cover while operating the vertical column or while
positioning the C-Arm.
CAUTION: Grasp one of the C-Arm positioning handles to prevent uncontrolled C-Arm movement
whenever you release the brake.
CAUTION: If the Wig-Wag brake is set to the extreme right, use caution not to injure your knuckles
when turning the steering handle 90° to the left.
8-4
C-Arm notes, cautions, warnings
CAUTION: When positioning the C-Arm, observe moving assemblies to prevent patient contact and
ensure the safety of anyone near the C-Arm.
CAUTION: When accessories (e.g. film cassette) are attached to the C-Arm, releasing the orbital or
lateral rotational brake could cause the C-Arm to move.
Use the vertical column to raise or lower the C-Arm to the desired height. The vertical column
switches are found on the top of the C-Arm control panel housing (see Figure 8-5 ).
After the C-Arm is switched on, one of the vertical column switches must be pressed to initialize the
lift column. This initialization takes approximately five seconds after which lift motion is enabled. If
the column does not move after the system is turned on, wait approximately 5 seconds after
releasing the key and then press the key again. This wait should give the lift motor time to initialize.
A 0.4 second delay is built into the up and down switch operations. If the vertical column switches
are pressed rapidly, the lift column may not move. If the column does not move after a switch press,
release the switch, wait a second, and then press the switch again to resume movement.
WARNING: In the event of uncommanded vertical column movement, immediately move the C-Arm
out of the patient environment until the problem has been corrected.
8-5
C-Arm notes, cautions, warnings
WARNING: If pressing a fast stop button fails to stop motor movement or X-rays, place the Workstation
power switch in the off position, or disconnect the power plug from the AC receptacle. Call
for service if the problem persists.
NOTE: Pressing either fast stop button disables X-rays and disables all motorized C-Arm motion.
NOTE: C-Arm motorized motion can be used after releasing a Fast Stop button.
CAUTION: When either fast stop button is pressed, the system must be rebooted to regain full system
functionality including X-ray generation.
8-6
C-Arm notes, cautions, warnings
WARNING: Pressing the Motion Stop button on the RUI immediately stops all C-Arm motion but does
not disable X-rays. Press one of the Fast Stop buttons on the C-Arm control panel housing
to immediately disable X-rays.
The RUI has a Motion Stop button located above and to the right of the joystick (Figure 8-7 ).
Pressing the Motion Stop button stops all motorized C-Arm and vertical lift but does NOT
disable X-rays.
To re-enable motorized C-Arm motion, reset the RUI Motion Stop button by turning it to the left and
allowing it to pop up. C-Arm motion is disabled until the RUI Motion Stop button is reset. If you
receive a “Motion Disabled” message on the Workstation, check the state of the Motion Stop button
to make sure it has been reset.
NOTE: When the C-Arm motion has been stopped using the Motion Stop button, you can move the C-
Arm manually by using the Clutch Handle to disengage the C-Arm clutch. See the “Operation”
section in the 9800 Motorized Drive C-Arm Operator Manual Supplement for more information.
8-7
C-Arm notes, cautions, warnings
CAUTION: The patient should be positioned as far away from the X-ray source as possible in order to
keep the absorbed dose to the patient as low as reasonably achievable.
CAUTION: Additional objects in the beam, such as tables and patient supports, may cause the
imaging system to increase the amount of radiation, in order to obtain an optimal image.
Prevent additional exposure by removing unnecessary objects from the path of the beam.
8-8
C-Arm notes, cautions, warnings
The skin spacer should only be removed on the instructions of a physician. The skin spacer should
be reattached to the collimator assembly immediately following the procedure.
Attach the skin spacer by pushing it onto the X-ray tube cover.
WARNING: Removing the skin spacer may result in increased radiation exposure to the patient. The
rate of exposure increases exponentially as the anatomy is positioned closer to the X-ray
tube.
8-9
C-Arm notes, cautions, warnings
8 - 10
Planned maintenance
Section 9
Planned maintenance
WARNING: Circuits inside the equipment use voltages capable of causing serious injury or death from
electrical shock. Only GE OEC trained and qualified personnel should remove the covers
or perform any type of service task.
WARNING: The planned maintenance procedure produces X-rays. Take the appropriate precautions.
9-1
Planned maintenance
9-2
Updated technical specifications
Section 10
Updated technical
specifications
CAUTION: A dedicated AC power line is recommended to avoid conflicts with the power requirements
of other equipment.
The GE OEC Medical Systems, Inc. authorized field representative shall provide the owner of the
system with information as to which rooms, in the area designated to him by the owner, meet the
requirements of line voltage, line current, and line regulation for the system. It is the owner's
responsibility to operate the system only in those rooms which meet these requirements.
This information must be obtained at the time of system installation on the site. If this information is
not available for the room in question, it can be obtained by a qualified technician or service
representative.
Failure to meet proper power requirements results in a violation of the U.S. Federal Performance
Standards 21CFR 1020.30-32.
10 - 1
Updated technical specifications
WARNING: To avoid software corruption, ensure that you wait for a full boot up and shutdown before
performing a leakage test.
WARNING: Do not toggle the Line switch from Forward to Reverse or vice versa with the system
powered on.
WARNING: Power must be off on the system before and during any tests. Do not interrupt power to
the system at any time while the system is booting or shutting down to avoid file system
corruption.
10 - 2
Updated technical specifications
10 - 3
Updated technical specifications
Standard systems
Line voltage (VAC) Maximum current Maximum power Standby current Standby power
(+/- 10%) (Amps) (kW) (Amps) (kW)
100 18 6.06
120 15 5.05
200 10 3.03
2.4 0.61
220 10 2.75
230 10 2.63
240 10 2.52
10 - 4
Updated technical specifications
10 - 5
Updated technical specifications
Horizontal plane
The following illustration shows measurement points in a plane perpendicular to the beam axis at a
height indicated by reference point C1. C1 is the 30-inch SSD, and is illustrated in the Vertical Plane
illustration, below. The measurement points are 25 cm apart. The table following the Figure 10-1
shows expected scatter radiation at each point.
10 - 6
Updated technical specifications
Vertical plane
The following illustration shows measurement points in a plane through the beam axis at a height
indicated by reference point C1, which represents the 30-inch SSD. The measurement points are 25
cm apart. The table following Figure 10-2 shows expected scatter radiation at these points.
10 - 7
Updated technical specifications
Horizontal plane
The following illustration shows measurement points in a plane perpendicular to the beam axis at a
height indicated by reference point C1. C1 is the 30-inch SSD, and is illustrated in the Vertical Plane
illustration, below. The measurement points are 25 cm apart. The table following Figure 10-3 shows
expected scatter radiation at each point.
10 - 8
Updated technical specifications
Vertical plane
The following illustration shows measurement points in a plane through the beam axis at a height
indicated by reference point C1, which represents the 30-inch SSD. The measurement points are
25 cm apart. The table following Figure 10-4 shows expected scatter radiation at these points.
10 - 9
Updated technical specifications
10.7 Dimensions
10 - 10
Updated technical specifications
10 - 11
Updated technical specifications
10 - 12
Updated technical specifications
10.7.4 Workstation
10 - 13
Updated technical specifications
10.8.2 Training
Training on system operation for clinicians and technical service training for biomedical engineers is
available. Use the contact information in this section to obtain more information.
10.8.3 Phone
1. Call the number that corresponds to your geographical location.
US: 800-874-7378
Canada: 800 668-0732
China: 800 810-8188
All other locations: Contact your local sales representative.
2. Establish an account with the customer service representative.
3. Place your order.
10.8.4 Fax
If you have a catalog and a GE Fax Order form, complete the form and fax it to 801-355-8114. If you
do not have a catalog, and would like one you can obtain them by dialing the number listed for your
geographical location.
10.8.5 On-line
You can order Disposables on the GE Healthcare Web site from any geographical location but prior
to ordering you must register on-line to open an account. Registration is free.
1. Go to www.gehealthcare.com/usen/index.html.
2. Register to open an account.
NOTE: Record your user name and password for future account access.
3. Log on using the user name and password that you established when you registered.
4. Browse through the on-line catalog and locate the GE OEC supplies that you want to
purchase.
10 - 14
Updated technical specifications
NOTE: If you have difficulties, dial the telephone number listed previously for your geographical
location or contact your local sales representative.
You may also call these numbers to order circuit diagrams, component part lists, calibration
instructions, arrange training or obtain other information which will assist qualified service
engineers to repair the system.
10 - 15
Updated technical specifications
10 - 16
Index
I-1
Error recovery I
Communication failure 2-11 Image directory
Image freeze 2-11 Copy patient information 7-2
Images not saved 2-11 Damaged thumbnail 7-1
Initialization failure 2-11 Review subtracted cine runs 7-2
No boot 2-11 Imaging modes summary 3-15
No X-ray 2-11 Intended use 1-1
Power loss and system restart 2-10 K
System lockups 2-11 Keyboard 2-12
Error recovery steps 4-1 Diacritic keys 2-12
Essential performance criteria 1-2 Save key 2-12
Exposure control Swap key 2-12
Automatic 3-12
L
Manual 3-12
Exposure timer 3-17 Labels
Extended exposure feature 2-16 Beam Limiting Device certification 1-9
External connections, Workstation 2-13 Certification labels location 1-7
Electric shock hazard 1-7
F Film cassette certification 1-11
Facility current leakage tests 10-2 HV generator/X-ray
Fast stop button 8-5 control certification 1-10
Film cassette holder 8-2 Image intensifier certification 1-9
Fluoro imaging Ionizing radiation 1-11, 1-12
Cardiac systems 5-8 IPX8 footswitch protection 1-11
Change pulse rate 5-7, 5-8 Manufacturing information 1-8
Digital spot imaging 5-5 Nameplate/rating 1-8
Enable pulsed X-ray 5-6 Pinch points 1-6
High level fluoro (HLF) 5-4 Power off/restart 1-7
Mode screen 5-2 WEEE label 1-12
Overview 5-1 Workstation connector ports 1-12
Producing digital spot image 5-5 X-ray tube assembly certification 1-10
Producing HLF image 5-4 Landmarking 6-11
Producing image 5-3 Left monitor messages 4-10
Saving images 5-9
M
Status bar 5-3
Vascular systems 6-2 Mag mode keys 2-13
Workstation setup 5-1 Maintenance
X-ray switch mode setup 5-3 Planned 9-1
Footswitches Manual exposure control 3-12
Modes and 3-14 Medical device directive iii
Three-pedal 2-15 Messages
Two-pedal 2-15 C-Arm 4-12
Left monitor 4-10
H Right monitor 4-2
Handswitch 2-15 Workstation 4-1
High Level Fluoro (HLF) 5-4 Mode display 3-13
I-2
Mode pairs 12 Inch (30.5 cm) II 10-8
Vascular imaging 6-4 9 Inch (22.8 cm) II 10-6
Modes of operation 3-12 Scheduled exams screen 3-5
Activating 3-12 Screens
Deactivating 3-12 Change current patient 3-9
Imaging modes summary 3-15 Customize time and date 3-11
Mode display 3-13 Dose Summary 3-19
Motorized C-Arm drive Login 3-1
C-Arm lockups 8-7 Mode screen 3-13
Fast stop button 8-5 Mode screen, Cardiac 6-7
Maximum rotational speed 8-7 Mode screen, Fluoro 5-2
Stop motion button 8-5 Mode screen, Vascular 6-3
Unexpected power loss 8-7 Patient information 3-2
Move images to new study 3-4 Saved exams 3-9
Multiple switch presses 2-16 Scheduled exams 3-5
O Update schedule - Schedule Filter 3-6
Ordering disposable and Shutting down the system 2-8
accessory items 10-14 Shutdown checklist 2-8
Source to skin distance 8-8
P
Specifications
Patient information 3-2 Classification type 10-1
Patient summary System input power 10-4
Printing 7-4 X-ray source assembly 10-5
Pinch points 1-6 Starting the system 2-3
Positioning the C-Arm 8-4 Restarting the system 2-9
Power requirements 2-1, 10-1 Startup checklist 2-3
Workstation 2-1 Status bar
R Modes displayed on 3-15
Registration, image Stop motion button 8-5
in Roadmap and subtraction 6-10 Subtraction imaging 6-8
Restarting the system 2-9 Adjusting mask 6-10
Review subtracted cine runs 7-2 Landmarking 6-11
Right monitor messages 4-2 Procedure 6-9
Roadmap imaging Registration 6-10
1st phase 6-12 Sequence 6-8
2nd phase 6-12 Swap key 2-12
Clearing a mask 6-14 Symbols
Mask from subtraction cine run 6-15 Anesthetic proof 1-16
Procedure 6-12 Battery 1-21
Selecting and clearing a mask 6-13 C-Arm transport position 1-16
Sequence 6-12 CE Mark 1-19
Using a saved mask 6-14 Circuit breaker 1 1-13
S Circuit breaker 2 1-13
Save key 2-12 Circuit breaker 3 1-14
Saved exams 3-9 Circuit breaker 4 1-14
Scatter radiation 10-6 Circuit breaker toggle OFF 1-20
I-3
Circuit breaker toggle ON 1-20 Digital Cine Pulse 6-16
CSA-NRTL 1-16 Cine acquisition 6-6
Dangerous voltage 1-17 Digital Cine Pulse 6-15
Date of manufacture 1-18 Mode pairs 6-4
Digital Cine Pulsed 1-23 Mode screen 6-3
EFUP 50 China 1-17 Overview 6-1
Equipotential terminals 1-15 Roadmap imaging 6-11
Fast stop 1-19 clearing a mask 6-14
Fluoro/Roadmap 1-19 mask from subtraction cine run 6-15
Focus 1-13 procedure 6-12
Foot crush points 1-16 selecting, clearing mask 6-13
Footswitch port 1-22 using a saved mask 6-14
Handswitch port 1-22 Setting up modes 6-5
High res out-left monitor 1-20 Subtraction imaging 6-8
High res out-right monitor 1-21 adjusting mask 6-10
HLF/Dig Spot/Subtract 1-19 landmarking 6-11
Ionizing radiation 1-17 procedure 6-9
Low res out-left monitor 1-20 registration 6-10
Mode toggle key 1-17 sequence 6-8
Parallel printer port 1-21 Switching mode pairs 6-4
Part number 1-18 Vascular systems
Pinch points 1-14 Digital Cine Pulse 6-17
Power switch 1-20 X-ray pedal setup 6-1
Printer port 1-21 Vascular systems
RUI port 1-22 Fluoro imaging 6-2
Serial number 1-18 W
Smiley face 1-19 Workstation
Switch/relay control 1-21 Connecting a device 2-14
System model/type 1-18 Customization 2-13
Type B equipment 1-18 External connections 2-13
VDE certification 1-17
X
Warning 1-15
X-ray equipment certification update 1-1
WEEE 1-19, 1-23
X-ray indication
X-ray key lock OFF 1-22
Audible 3-20
X-ray key lock ON 1-22
Visual 3-20
System input power 10-4
X-ray switch security error 8-8
T
Time and date, set 3-11
Transport and storage checklist 2-2
U
Update Schedule - Schedule Filter 3-6
Updated safety information 1-2
V
Vascular imaging
Cardiac systems 6-7
I-4