Professional Documents
Culture Documents
Introduction ers placed directly on skin will experience non-rigid body move
ment. Another approach, described by Cappello et al. (1996),
Knee movement during normal function involves a complex
used a cluster of points to correct for deformation in the limb
set of coupled translations and rotations. The movement of the
segment by a double calibration technique. The cluster of points
knee is dependent on the type of activity, the muscle firing is recorded in full extension and in full flexion; the body is
patterns as well as the passive anatomy of the knee joint. A assumed to deform linearly between these two positions. This
knowledge of the in vivo movement of the knee is important linearity assumption, along with the assumption that the dy
for understanding normal function as well as addressing clinical namic deformation that will be recorded during an actual activ
problems. For example, it has been well established that follow ity is equivalent to the static deformation, limit the applicability
ing rupture to the anterior cruciate ligament of the knee joint, of this technique. As yet none of these methods provide an
there is an increase in the anterior translation of the tibia demon absolute standard to identify nonrigid body movement of the
strated during clinical testing. However, the relationship be limb segment.
tween passive clinical instability and dynamic instability re Skeletal movement can be measured using alternative ap
mains unclear. There is a need for improved methods of in vivo proaches to a skin-based marker system. These approaches in
testing that can be used to assess the complete six-degree-of clude stereoradiography (Jonsson and Karrholm, 1993), bone
freedom movement of the knee joint. pins (LaFortune et aI., 1992; Murphy, 1990; Reinschmidt et aI.,
The most frequently used method for measuring human 1995), external fixation devices (Cappozza et aI., 1996) or
movement involves placing markers or fixtures on the skin sur single plane fluoroscopic techniques (Sati et aI., 1996; Banks
face of the segment being analyzed. The movement of the mark and Hodge, 1996; Stein et aI., 1993; Stiehl et aI., 1995). All of
ers or fixtures is typically used to infer the underlying relative these methods are invasive or expose the test subject to radia
movement between two adjacent segments (e'-g., knee joint) tion. Therefore, the wide spread applicability of these methods
with the goal of precisely defining the movement of the joint. is limited.
Skin movement relative to the underlying bone is a primary Recently a new approach using a cluster of points has been
factor limiting the resolution of detailed joint movement using described (Andriacchi et aI., 1994; Andriacchi and Toney,
skin-based systems (Holden et aI., 1997). The scope and clini 1995), which has the potential to address the nonrigid body
cal relevance of in vivo measurement of human motion would movement of the limb segment in a systematic manner. The
be substantially improved if reliable measurements of skeletal method has been summarized in general terms and tested against
movement could be obtained from skin-based marker systems. results from two subjects wearing Ilizarov devices. However,
Spoor and Veldpas ( 1980) have described a technique using the details of the methodology were not fully described.
a minimum mean square error approach that lessens the effect The purpose of this paper is to present the details of a new
of deformation between any two time steps, time 1 to time 2. method based on a point cluster technique (PCT) to reduce the
This method assumes the markers are fixed to a segment that nonrigid body motion artifacts during in vivo human motion
moves as a rigid body between any two time steps. This assump testing and to evaluate its applicability to in vivo testing. The
tion limits the scope of application for this method, since mark- PCT method was combined with an optimization approach and
tested using a simulation model. The in vivo study was designed
to evaluate the PCT method and to compare with a previously
COnlributed by the Bioengineering Division for publication in the JOURNAL
OF BIOMECHANICAL ENGINEERING. Manuscript received by the Bioengineering
reported study (La Fortune et aI., 1992) on the three-dimen
Division September 24. 1997; revised manuscript received July 22, 1998. Associ sional kinematics of the human knee during walking acquired
ate Technical Editor: R. C. Haul. using intra-cortical traction pins in the femur and tibia.
Journal of Biomechanical Engineering Copyright © 1998 by ASME DECEMBER 1998, Vol. 120 / 743
•
.-------------------- --------- ----
Analytical Methods We can now define the local coordinates L (I), for each of the
points by
Point Cluster Equations. The PCT is based on a cluster
of points unifonnly distributed on the limb segment. Each point L(/), = RT(t)'(G(t); .. C(t» i = I ... n (7)
(i) is assigned an arbitrary mass m(t), , which can be varied at
R(t) is orthogonal since it is constructed from three orthogonal
each time step. The center of mass and the inertia tensor of
the cluster of points are calculated as described below. The eigenvectors as shown in Eq (6).
eigenvalues and the eigenvectors of the inertia tensor arc the Optimization Method and Simulation Study. In conjunc
principal moments of inertia and the principle axes of the point tion with the point cluster method describe above, there is the
cluster. The eigenvectors establish a transformation between the potential to enhance the ability to isolate rigid body motion
segment and the global coordinate system. The eigenvalues are using optimization techniques. The eigenvalues form a basis for
invariant to movement if the segment is behaving as a rigid an optimization technique since they remain invariant if the
body. If the eigenvalues change from their value in the rest segment moves as a rigid body. The mass weighting factors
position, then the segment where the markers are placed has m (t), can be assigned a unit value if the cluster segment moves
deviated from rigid body movement. By adjusting the individual primarily as a rigid body. Typically, markers placed on the skin
weighting m(t), of each point, the artifact due to nonrigid body will undergo both rigid body motion and motion due to the skin
movement can be reduced. deformation, If nonrigid body movement occurs, the eigenval
The point cluster technique was implemented by placing a ues will change from their values in the rest position (to). An
group of markers (n > 3) on a limb segment. At any time t, algorithm was developed to minimize the eigenvalue changes
denote the location of marker i in the global coordinate system by redistributing the weight factors m(t), at each time step. The
by G(t)" where algorithm begins by calculating the sum of the squares of the
three eigenvalues (Eigenvalue norm) for times to and t, as
g(t),,x]
G(t), = g(t)"y i=I ... n (1) ! 2
A o = \10'0)' +
2 2
O'nh + (\"h and
[ g(t),.,
A, = leA,); + (As)~ + (As)~ (8)
Each point G(t), was arbitrarily assigned a mass m(t)i' The
As previously noted, the mass m(/)i was used as a weighting
mass serves as a weighting factor, which can be adjusted to
factor to reduce variations in eigenvalues. The eigenvalue norms
minimize changes in eigenvalues. The center of mass C (t) of
A, and Ao were written as a function of each cluster point mass
the points is given by
m(tJi' Each Ao and A., was calculated from Eg. (5) as a function
C(t) = 2:j-) G(t), . m(t), of the mass distribution m (t,); using Eqs. (1) - ( 4 ) to establish
(2) the matrix given in Eq. (5). Thus the difference between A,
2:;'~1 m(t),
and A o expressed in Eq. (8) could be calculated in terms of
At time t, the location of G (t), is given by the n cluster point masses (m(t,),) for each time step f,. This
difference was implicitly defined as a function F(mU,» in the
G(t), = C(t) + R(t)· L(t)i i = I ... n (3)
following form:
where R(t) is the orientation matrix, and L(t), is the location
F(m(tJ" ., . , m(t,)n) = (A, ~ A o )2 (9)
of the points in a local coordinate system.
To define the orientation matrix R(t) and local coordinate A mass distribution for n cluster points m(tJi~ln at each time
system in the cluster, the points G(t), were translated to a step ts that minimized F(m(t,),) was solved numerically. Min
coordinate system located at the mass center C (t) as follows: imizing F(m(tJ,) took the form of a nonlinear minimization
P(t)i = G(t)i - C(t) i = I ... n (4) of n unknowns m(ts)i for i = I .. n, which was solved at each
time step. For computational efficiency the mass distribution
The inertia tensor I(t) for the discrete cluster points was given was defined in terms of a single distribution parameter E. The
by parameter E was calculated by redefining the mass distribution
i=l
The eigenvectors E; of I(t) were used to form the transforma such that the point with the largest displacement from the rest
tion matrix R(t) as follows: position in the local coordinate system will have the lowest
weight factor. Note that for rigid body movement each point
(6)
would have zero displacement from its rest position. E was
where the three eigenvectors have the form calculated by first determining the displacement 6L( t s )' of each
point (i) relative to its position in the reference position L(t,),
e(t»),x] at time to as
E(t») = e(t»).y j = 1,2,3
[ (10)
e (t) ),l
The local coordinate L(U, at time step t, for each point was
Associated with each of these eigenvectors is an eigenvalue calculated using Eq. (7). The relative movement (nonrigid body
'A(t»),j = 1,2,3. movement) of each coordinate relative to the rest position (at
time = to) is given by f?L (ts ). ). A trial redistribution of m(t)i Simulation Results. The systematic error introduced into
was calculated such that the point with the largest displacement four of the eight cluster points was linearly increased over 20
from the reference position (max (!::,.L(tS)i) was assigned the time steps while the random error was applied to all eight cluster
lowest weighting factor and the remaining distribution was de points (Fig. 1 (A». Eight markers were selected to match our
termined by the following: in vivo model. The error was calculated as the distance that
each cluster point moved from its initial position. As the system
atic error for some of the cluster points increased to 6 cm, the
change in the eigenvalue norm (Fig. I(B» approached a 20
Thus, the mass distribution was expressed in terms of a single percent increase from the initial rest position. In spite of the
parameter E. The problem of minimizing F(m(ts)J, ... , m(t,)n)
large error in half of the cluster points, the mass center error was
for n variables was reduced to minimizing
approximately 3 cm at its maximum. The error in the location of
F (m. ( E( ts ) U < tol (12) the trial point provides an important evaluation of the influence
of the cluster point errors because its calculation is based on
the position of the mass center as well as the angular position
Equation ( II) provides the initial trial solution to minimize determined from the eigenvectors as given by Eq. (7). This is
F(m(t.')i)' F(mU,)i) is then minimized with respect tOE using the type of calculation that is ultimately used to locate the
the Levenberg-Marquardt technique (Levenberg, 1944; Mar location anatomical coordinate system fixed in the underlying
quardt, 1963) to within a nonzero convergence tolerance given bone when the cluster method is applied to in vivo movement
as tol in Eq. (12). The parameter E is determined for each time studies. Again, in spite of the large error in four of the eight
step to provide a redistribution of the weighting parameters cluster points, the calculation of the trial point error was less
m(t)i that will minimize the effect of markers contributing to than 3.5 cm prior to the application of the mass optimization
the non-rigid body motion. . algorithm. Thus, the cluster method provided a substantial im
provement in the calculation of the location of the trial point
Simulation Study. To test the algorithm, error was intro without employing the optimization algorithm. In fact, if one
duced into a fixed cluster of eight points (Fig. I) in the form anticipates cluster point error on the order of 2 cm or less, the
of both random and systematic error. Eight points were selected cluster point algorithm can be used quite efficiently without the
to replicate the number of markers used for the in vivo study. introduction of the optimization algorithm. However, if nonrigid
The random error represented the error primarily due to digitiza body movement becomes large, the redistribution mass algo
tion noise associated with the optoelectronic system. The ran rithm substantially reduces the error in the location of the mass
dom noise assigned to each cluster point coordinate had a mag center as well as the ultimate calculation of the trial point posi
nitude ranging between 0 and 5 mm. The systematic error was tion. By minimizing the change in the eigenvalue norm, one
assigned to a subset of the cluster points to represent the move can achieve a substantial improvement in the systematic error
ment of skin or underlying muscle during contraction. The sys associated with simulated skin movement. While the simulation
tematic error was assigned in the longitudinal direction of the study provided the only absolute test of the algorithm's ability
cluster to represent skin movement tangential to the surface of to correct for nonrigid body motion in cluster points, it remained
the segment. Four of the eight cluster points were assigned an difficult to assess the applicability of this analytical technique
error that increased from 0 and 6 cm over 20 time steps. The to true in vivo testing. Therefore, an in vivo study was conducted
corrected and uncorrected cluster position was compared in to examine the applicability of point cluster methodology to its
terms of position of the cluster center of mass and orientation. intended goal, in vivo testing.
In addition, the error in the position of a trial point not included
in the cluster was used to evaluate error correction. Calculations In Vivo Tests. The in vivo testing protocol was designed
were based on setting tol = 0.1 in Eq. (12). to follow the protocol described by LaFortune et aL (1992)
with several exceptions. In that study, five healthy male subjects
were tested while walking with intracortical pins placed in the
femur and tibia to quantify six-degree-of-freedom movement
of the knee joint. In the present study, five male subjects and
, ., ,
""
" , ~, '
,.
five female subjects were also tested during level walking using
A a protocol approved by the Institution Review Board. The aver
. ~ Non-optimized ,- --:-_--, age age of the population was 31.7 ::'::: 10.5 years, the weight
Optimized~ was 574 ::'::: 120.9 N, and the height was 1.7 ::'::: 0.1 m. For
t Systematic
Error
analysis, a single trial was randomly selected from six trials
E Random
acquired on each patient.
~
~1II 20b:.on-OPlim.iz.ed .
Error
Trial Point
~~ \ B
~i3 ." ..... '... ~OptJmlzed
Instrumentation. Reflective markers on the limb segment
l-
&~
W
II .....
o· t were tracked using a four camera video-based motion measure
'''< +
Clusler /1>_ 5
ment system (GaitLink TM, Computerized Functional Testing
Corporation, Chicago, IL). The cameras were positioned in an
Points
arc approximately 1.5 m from the center of the measurement
volume on one side of a 10m walkway. A modified direct linear
~
transform technique (Hatze, 1988) was used to reconstruct the
I three-dimensional locations of the markers. The transform pa
Non'optimized .
~ •• Optimized
0
L 1 2
~_~ _
._J _
E
rameters were calculated using 88 markers in known locations,
/ \"c, chosen to span the volume of interest completely.
The system accuracy was tested using a rigid test device
TimuStl:lp fabricated to represent the thigh, shank segment and knee joint.
The test device knee could be flexed at predetermined angles.
Fig. 1 The simUlation study involved introducing random and systematic Markers were placed on this device, then data were acquired
error into a fixed set of eight cluster points (E). The systematic error in while the device was rolled along the walkway. From these
four cluster points was linearly increased with time (A). The errors
(dashed line) in eigenvalue, mass center, and trial points increased with
tests, root mean square angular errors of 0.3, 0.6, and 1.0 deg
increasing systematic error. Following optimization (solid tine), there were measured for flexion, adduction, and internal rotation at
was substantial error reduction in each parameter. the knee respectively. For translational errors, the lateral, ante-
• ---------------
rior, and superior rms errors were 1.5, 1.5, and 1.0 mm, respec
tively.
Subjects and Protocol. The motion of the knee joint was Trans-epicondylar
quantified in terms of flexion, rotation, and displacement. A total Reference 1\1/111/111/1111/1
---- ----------
70 There are several observations that can be derived from the
60
simulation study. The cluster method provides a substantial
LaFortune et al
correction in the location of the origin of a cluster coordinate
c: iii 50 (1992) system (mass center) as well as the orientation axes (eigenvec
';o( e 40
QI
and PCT '- tors) of the cluster system. Even with large errors introduced
.! g' 30 into half of the cluster points, the ultimate calculation of an
11.0
- 20 arbitrary trial point was substantially improved even prior to
10 the application of the optimization algorithm (Fig. 1). Thus, ip
situations where cluster points are strategically located such that
skin movement is minimized, the point cluster system can be
-10
efficiently used without the recruitment of the optimization al
gorithm. While the mass redistribution algorithm can substan
c tially improve the calculation of the mass center location and
o
n
:::l
"tl
optimization at each time step. Convergence, numerical condi
c:iilc:( 5 LaFortune et al
o CIl tioning, and the appropriate selection of the optimization algo
+if (1992)~ PCT
.Bel
o CIl
~e.c:
--. - --- . - -- . . --
O+-------==----------/ll-----~__::l=_
rithm are important considerations. At this point, the algorithm
was not selected for efficiency, but rather to demonstrate and
o evaluate the analytical procedure associated with the point clus
~ ter. The ultimate evaluation of this type of methodology must
:::l -5
"tl
.c
c:(
-1
LaFortune et al
(1992)
1 - LaFortune et al
(1992) I #
, ... . ..
Heel
Strike
Off
Ei'ii (1992)
~~ f ..
Fig. 4 A comparison of averaged rotation of the femur with respect to c:CIl..l
-; 0 "~-¥--=------.I---4,-----"-",,,,,---,,,,,,;i<I--''''-
the tibia. The measurements derived from the skin-based PCT markers E
(solid line) had comparable temporal patterns (location of maxima and
minima) to those obtained with intracortical pins (LaFortune et al., 1992).
Note that LaFortune's results were inverted to a fixed tibial frame.
.CIl
u
lc;
i:3 'C -1
..I CIl
:!:!
ment showed greater error during swing phase than during
stance phase (Fig. 6).
-2
Discussion
The point cluster method in this paper addresses the problem
of deriving limb segment motion from markers placed on the -.!2
..
skin. The method provides a basis for determining the artifact E"
u.s
-c: LaFortune at a/I
associated with nonrigid body movement in a manner that per (1992)
c:c:(
CIl
mits an improvement in the approximation to the rigid portion E
of the segment movement.
The results of the simulation study provided an important
.
u
CIl
"ii ..
III 0
basis to evaluate the point cluster methodology as well as the c'i:CIl
optimization algorithm against an absolute standard. The sys a. ...
c:(:g
tematic error introduced into the cluster was designed to repre a.
sent skin movement associated with a portion of the cluster Stance Swing
segment moving on a contracting muscle or any other source
Heel Toe
of skin surface movement. The random noise was designed to be Strike Off
associated with the influence of the optoelectronic measurement
systems typically used to acquire the movement of markers in Fig. 5 A comparison of averaged translation of the femur with respect
three-dimensional space. These errors were selected to replicate to the tibia. The translation is based on the displacement of the femoral
reference point (Fig. 3) with respect to a fixed tibial system. Note LaFor
what is anticipated from in vivo studies (Cappozzo et aI., 1996; tune used a different reference point (Fig. 3) and normalized translation
Maslen and Ackland, 1994; Holden et aI., 1997; LaFortune et to zero at high strike. LaFortune's results were converted to a fixed tibial
aI., 1992). system.
-I
Table 1 A comparison of the maximum rotational and translational mo 80
tion between measurements obtained with the point cluster method and 70
those reported by LaFortune. The peak values were defined as shown 60
in Figs. 4 and 5. 1
\
50
Strike
Toe
Oft
Axial 4.\l 0.70 2.13 -0.32 1.98 1.02 Fig. 6 An illustration of the temporal change in the eigenvalue norm
(em) (0.63) (0.38) (0.71) (0.29)
dUring the gait cycle
be determined through in vivo studies. However, absolute evalu can be attributed to a number of factors. These factors include
ation even in highly controlled in vivo studies is difficult. Even the precise definition of the transepicondylar reference point,
a study with rigid pins fixed into a rigid portion of the segment the differences in gait that may be associated with patients
such as bone will have errors associated with the instrumenta walldng with the intracortical pins versus those walking nor
tion. Thus, any physical study will have unknown errors that mally as well as possible differences in axis locations.
will contaminate the absolute analytical evaluation of the ap The physical interpretation of the anterior-posterior position
proach. At present, the best physical. data availa.ble, for eval.ua of the femur with respect to the tibia provides some interesting
tion of the methodology are reported m a study With mtracortlcal insights into the relative tibiofemoral contact movement over
pins placed in the femur and the tibia (LaFortune et aI., 1992). the gait cycle. At heel strike the results of this study demonstrate
While there are limitations to direct comparisons between pa that the tibia is anteriorly displaced with respect to the femur.
tients measured in LaFortune's study compared to those mea This finding is reasonable in the context of the loads that occur
sured in this study, a comparison of the similarities of the results at heel strike. Typically, as the foot strikes the ground there is
as well as a physical interpretation of the results provides an an anteriorly directed force at the heel. Simultaneously at heel
important basis for understanding the in vivo application .of this strike, the quadriceps are firing, which will also produce an
methodology. There has been a paucity of in vivo testmg re anterior force on the tibia with the knee at full extension. As
ported with similar methods (Spoor and Veldpaus, 1980; Cap the knee goes past midstance phase it again reaches extension
pello et aI., 1996). prior to toe-off. The results of this study demonstrate that the
Comparison of this approach to the work of either Spoor tibia displaces posteriorly with respect to the femur during this
and Valdpaus or Capello et al. is problematic. The Spoor and portion of the gait cycle. This finding is again reasonable consid
Valdpaus work provided only a detailed description of their ering that at this time in the gait cycle the quadriceps are not
approach; no results of computer simulations or in vivo testing firing and the external forces are tending to push the tibia poste
were presented. The work of Cappello et al. do not provide the riorly as the weight is transferring to the contralateral limb.
mathematics of their technique; results were limited to a single These two points in the walking cycle with the knee at full
individual performing a cycling activity. The results of the in extension produced extreme differences in the anterior-posterior
vivo comparison to previous studies where intracortical pins position of the transepicondylar reference point. These findings
were placed also demonstrated that the point cluster method provide a basis for the dynamic evaluation of anterior-posterior
produces a reasonable approximation to in vivo knee joint move stability of the knee joint. In addition, it points out the impor
ment. The internal rotation calculations were quite comparable. tance of relating knee kinematics to the particular dynamic func
The differences in the anterior-posterior displacement patterns tion when attempting to describe tibiofemoral kinematics.
There are practical limitations of PCT. In practice, having
Table 2 The average and standard deviation (sd.) of the percent eigen this overabundance of markers on each segment can lead to
value norms change from the reference position for the thigh and shank difficulties in tracking and labeling the individual markers. Ad
segments over one walking cycle. The eigenvalue changes provide a ditionally, care must be taken in marker placement to ensure
metric for each segment's deviation from rigid body motion. Also shown the markers are positioned such that the point clusters are not
are the demographic measurements for each subject.
symmetric; that is, the cluster of markers possess no axes of
% Eigenvalye Cbunge •
~-----------
a
Hoppenfeld, S., and Huton, R., 1976, "Physical examination of the knee,"
AR39421-05A2: "Anterior Cruciate Ligament-Functional Physical Examination of the Spine and Extremities, 171-196.
Biomechanics.' , Jonsson, H., and Karrholm, J., 1994, "Tlu'ee-dimensional knee joint movements
during a step-up: evaluation after anterior cruciate ligament rupture," J. Or/hop.
References Res., 12(6), 769-779.
Andriacchi, T. P., Sen, K., Toney, M. K., and Yoder, D., 1994, "New develop LaFortune, M. A., Cavanagh,P. R., Sommer, H. J., and Kalenak, A., 1992,
ments in musculoskeletal functional testing," Proc. Canadian Soc. ofBiomechan "Three dimensional kinematics of the human knee during walking," J. Biome
ics, VIIIth Biennial Conf.. pp. 12-13. chanics, 25, 347-357.
Andriacchi, T. P., and Toney, M. K., 1995, "In Vivo Measurement of Six Levenberg, K., 1944, "A method for the solution of certain problems in least
Degrees-of-Freedom Knee Movement During Functional Testing," in: Trans. 41 squares," Quart. ApI. Math., 2, 164-168.
Ann. Meet. Orth. Res. Soc., Orlando, FL, p. 698. Marquardt, D., 1963, "An algorithm for least-squares estimation of nonlinear
Banks, S. A., and Hodge, W. A., 1996, "Accurate measurement of three-dimen parameters," SIAM J. Appl. Math., 11,431-441.
sional knee replacement kinematics using single-plane fluoroscopy," IEEE Trans. Maslen, B. A., and Ackland, T. R., 1994, "Radiographic study of skin displace
Biomed. Eng., 43(6), 638-649. ment errors in the foot and ankle during standing," Clinical Biomechanics, 9,
Benedetti, M. G., and Cappozza, A., 1994, "Anatomical landmark definition 291-296.
and identification in computer aided movement analysis in a rehabilitation context Murphy, M. C, 1990, "Geometry and the Kinematics of the Normal Human
II (Internal Report)," Universita Degli Studi La Sapienza, 1-31. Knee," Ph.D. thesis, Massachusetts Institute of Technology, Cambridge, MA.
Cappello, A., Cappozza, A., La Palombara, P. F., Leardini, A., and Bertani, Reinschmidt, c., van den Bogert, A. J., Lundberg, A., Murphy, N., and Nigg,
A., 1996, "Skin artifact compensation by double calibration in bone motion B. M., 1995, "Tibiofemoral and calcaneus-tibia motion during running: skin vs.
reconstruction," Paper No. 2.6.3-2, presented at the 18th Annual Int. Conf. of bone markers," in: Conference Proceedings. American Society of Biomechanics,
the IEEE Engineering in Medicine and Biology Society.
Stanford, 41-42.
Cappozza, A., Catani, E, Leardini, A., Benedetti, M. G., and Della Croce, U.,
Sati, A., de Guise, J. A., Larouche, S., and Drouin, G., 1996, "Quantitative
1996, "Position and orientation in space of bones during movement: experimental
assessment of skin-bone movement at the knee," The Knee, 3, 121-138.
artifacts," Clinical Biomechanics, II, 90-100.
Spoor, C. W., and Veldpaus, E E., 1980, "Rigid body motion calculated from
Dyrby, C. 0., Toney, M. K., and Andriacchi, T. P., 1997, "Relation between
spatial co-ordinates of markers," J. Biomechanics, 13,391-393.
knee flexion and tibial-femoral rotation duri"ng activities involving deep flexion,"
Stein, L. A., Endicott, A. N., Sampalis, J. S., Kaplow, M. A., Patel, M. D., and
Gait and Posture, 5, 179.
Hatze, H., 1988, "Higb-precision three-dimensional photogrammetric calibra . Mitchell, N. S., 1993, "Motion of the patella during walking: A video digital
tion and object space reconstruction using a modified DLT-approach," J. Biome fluoroscopic study in healthy volunteers," Amer. 1. Roentgenology, 161, 618
chanics, 21,533-538. 620.
Holden, J. P., Orsini, J. A., Siegel, K. L., Kepple, T. M., Gerber, L. H., and Stiehl, J. B., Komistek, R. D., Dennis, D. A., Paxson, R. D., and Hoff, W, A.,
Stanhope, S. J., 1997, "Surface movement errors in shank kinematics and knee 1995, "Fluoroscopic analysis of kinematics after posterior-cruciate-retaining knee
kinetics during gait," Gait and Posture, 3,217-227. arthroplasty," 1. Bone Joint Surg. Br.. 77, 884-889.
a !