You are on page 1of 2

328 CASE REPORT VODISKAR ET AL Ann Thorac Surg

CONTACT ALLERGY TO SILICONE 2014;97:328–9

and a severe tricuspid valve insufficiency. After neonatal


biventricular repair the tricuspid valve had to be replaced
Polytetrafluoroethylene-Coated by a mechanical prosthesis (St. Jude Medical, St. Paul,
Pacemaker Leads as Surgical MN) and the right ventricular outflow tract was recon-
Management of Contact Allergy structed with a pulmonary homograft 2 1/2 years later.
After right ventricular outflow tract reconstruction she
to Silicone developed complete atrioventricular block. Therefore, an
Janez Vodiskar, MD, Heike Schn€
oring, MD, epicardial pacemaker system was implanted with right
J€
org S. Sachweh, MD, Eberhard M€uhler, MD, and upper abdominal quadrant generator placement (PICOS 01;
Jaime F. Vazquez-Jimenez, MD Biotronik Co, Berlin, Germany) and 4351M ventricular
Departments of Pediatric Cardiac Surgery and Pediatric epicardial lead (Medtronic Inc, Minneapolis, MN).
Cardiology, Rheinisch-Westfaelische Technische Hochschule At the age of 11 years she underwent another recon-
Aachen University, Aachen, Germany struction of the right ventricular outflow tract using a
mechanical prosthesis. At the same time an epicardial
We have previously reported an 18-year-old girl with a atrial electrode was placed and the pacemaker system
congenital heart defect who developed complete heart changed to a DDD system (Biotronik CYLOS DR-T;
block after one of her corrective surgeries and who Medtronic 4986-35 atrial lead). During the postoperative
needed an epicardial pacemaker implantation. She course the abdominal pacemaker pocket ceased to heal.
developed contact sensitivity to silicone compounds. In the following year she underwent multiple site-
The problem was solved by implanting a silicone-free pacing system replacements due to local healing prob-
lems. After establishing the allergic reaction to silicone
FEATURE ARTICLES

pacemaker system utilizing silicone-free transvenous


leads. The patient was readmitted 2 years later due to compounds the wound healing problem was solved by
lead failure. As no silicone-free epicardial leads were removing the complete pacemaker system and replacing
available, we decided to use standard silicone epicar- it with a silicone-free system (Medtronic generator
dial leads and enclose the whole system in Gore-Tex E2DR31 ENPLUS). The electrodes used were newly
material (W.L. Gore & Associates, Flagstaff, AZ). developed transvenous screw-in polyurethane leads
Based on our experience we would discourage the (Medtronic 904169). The electrodes were screwed deeply
use of silicone-free transvenous pacing leads for into myocardial tissue and used for epicardial pacing.
epicardial use. In March 2010 she was admitted for a routine tooth
(Ann Thorac Surg 2014;97:328–9) extraction. The pre-procedure electrocardiogram showed
Ó 2014 by The Society of Thoracic Surgeons dysfunction of the ventricular lead. An elevation of pacing
threshold to 7 volts, with stabile impendence and no signs
of a lead fracture were observed. The decision was made
to replace the ventricular lead and the generator with new
silicone-free transvenous lead and generator (Medtronic
C ontact sensitivity to various pacemaker compounds
has been reported by different authors [1–6]. Sili-
cone is often used to cover devices containing materials
ADAPTAL ADDR1PX generator and Medtronic 904169
lead). At discharge the new system showed good pace-
maker function with appropriate pacing and sensing
proven to be allergenic, such as titanium, poly-
parameters.
chloroparaxylene (parylene), or polyurethanes [3, 5]. We
In 2 months time she presented again with signs of
have previously reported the history of a patient with a
vertigo and palpitations. Clinically she was hemody-
complex heart defect, which developed complete heart
namically stabile, with a heart frequency of 45 to 60 beats
block after one of her corrective surgeries and thus
per minute. Pacemaker parameters again showed a rise in
she needed a pacemaker implantation [1]. After a
ventricular lead threshold. A probative therapy with
problematic wound healing process it was established
prednisolone was started. In the next few days we
that the patient is sensitive to silicone compounds. The
observed a further rise of stimulation threshold which
problem was than solved with the implantation of a
evolved just before surgery into a complete exit block.
silicone-free pacemaker system using newly developed
Taking into account that the previously implanted leads
silicone-free transvenous electrodes for epicardial
were originally developed for transvenous application
stimulation.
and no availability of silicone-free epicardial leads,
another method for preventing allergic reaction was
An almost 18-year-old girl was born with pulmonary
sought. Thus, the following procedure was chosen. The
atresia with intact ventricular septum, atrial septal defect,
silicone-free transvenous electrodes were exchanged for
widely used silicone-coated bipolar steroid eluting
epicardial ventricular and atrial leads (Ventricle Med-
tronic Model 4968 and Atrium Medtronic Model 4986). To
Accepted for publication April 29, 2013. minimize the tissue to pacemaker system contact both
Address correspondence to Dr Vodiskar, Department of Pediatric Cardiac
atrial and ventricular electrodes were enclosed in a
Surgery, RWTH Aachen University, Pauwelsstr 30, Aachen, D-52074, Y-shaped, tubular 5-and 6-mm Gore-Tex prosthesis
Germany; e-mail: jvodiskar@ukaachen.de. (Fig 1). The generator was completely covered by a

Ó 2014 by The Society of Thoracic Surgeons 0003-4975/$36.00


Published by Elsevier Inc http://dx.doi.org/10.1016/j.athoracsur.2013.04.135
Ann Thorac Surg CASE REPORT IRIE ET AL 329
2014;97:329–32 SPONTANEOUS WHOLE-LUNG TORSION

after 2 months. There were no signs of lead fracture and by


explantation a tissue reaction at the site of contact between
electrode head and myocardium was observed, which
could be the cause of the exit block we have described.
At the time of pacemaker system replacement there
were no silicone-free epicardial electrodes available, thus
Fig 1. Atrial and ventricular electrodes enclosed in a Y-shaped,
tubular 5- and 6-mm Gore-Tex prosthesis. we decided to use conventional epicardial electrodes and
prevent an allergy reaction by enclosing the whole system
Gore-Tex sheath. At the site of contact of the leads with in to Gote-Tex material, as previously described by Iguchi
epicardium, the prosthesis was sewn over the lead heads et al [6]. After 2 years of follow-up we have observed no
to the epicardium (Fig 2). allergic manifestations or pacemaker system or lead
Postoperatively the patient recovered fully and the dysfunction.
operative wounds showed good healing without any Based on this experience we would strongly discourage
sings of infection or allergic reaction. Pacemaker param- the use of silicone-free transvenous pacing leads for
eters at discharge showed a pacing threshold of 1.2 V and epicardial use and would recommend using either
0.7 V at 0.5 ms for the atrium and ventricle, respectively; custom made silicone-free epicardial leads if available or
impedance was stable at 1,498 Ohm. completely covering the conventional silicone leads with
At last follow-up at the end of October 2012 the patient a less allergenic material such as Gore-Tex. At this time
is doing well; postoperative wounds are fully healed and there are no silicone-free epicardial leads available on the
had, since the last operation, shown no allergic reaction. market.

FEATURE ARTICLES
The pacemaker parameters were stable with lead pacing
thresholds of 0.5 V and 1.6 V at 0.4 ms for the atrium and
ventricle, respectively. Impedance on both electrodes is References
stable and in the range of 600 Ohms. 1. Oprea ML, Schn€ oring H, Sachweh JS, Hagen O, Beitritz J,
Vazquez-Jimenez JF. Allergy to pacemaker silicone com-
pounds: recognition and surgical management. Ann Thorac
Comment Surg 2009;87:1275–7.
2. Peters MS, Schroeer AL, van Hale HM, Broadbent JC.
Silicone compounds used to cover pacemaker system Pacemaker contact sensitivity. Contact Dermatitis 1984;11:
components are a rare cause of wound healing problems 214–8.
after pacemaker implantation [4]. It has already been 3. Abdallah HI, Balsara RK, O’ Riordan AC. Pacemaker contact
demonstrated that it is important to consider contact sensitivity: clinical recognition and management. Ann Thorac
Surg 1994;57:1017–8.
allergy as an unusual cause of repetitive pacemaker 4. Maushagen E, Reichle B, Simon H. Circumscriptive erythema
wound complications and manufacturer-based skin tests after implantation of cardiac pacemaker. [Article in German]
may be required to discover a reaction to a specific system Z Kardiol 1994;83:340–2.
component [1, 5]. 5. Hayes DL, Loesl K. Pacemaker component allergy: case report
and review of the literature. J Interv Card Electrophysiol
We have proposed solving the problem of silicone-
2002;6:277–8.
contact allergy by implanting a silicone-free pacemaker 6. Iguchi N, Kasanuki H, Matsuda N, Shoda M, Ohnishi S,
system [1]. Unfortunately, the only available silicone-free Hososda S. Contact sensitivity to polychloroparaxylene-
leads at the time were leads made for transvenous coated cardiac pacemaker. Pacing Clin Electrophysiol
application. Our initial experience with these leads was 1997;20(2 Pt 1):372–3.
favorable and their application for epicardial pacing
seemed not to influence their function. We have observed
the failure of these leads after 2 years and the second time Spontaneous Whole-Lung Torsion
After Massive Pleural Effusion
and Atelectasis
Masahiro Irie, MD, Norihito Okumura, MD, PhD,
Jun Nakano, MD, PhD, Atsushi Fujiwara, MD,
Misa Noguchi, MD, Hidenao Kayawake, MD,
Akihiko Yamashina, MD, Tomoaki Matsuoka, MD, and
Kotaro Kameyama, MD, PhD
Department of Thoracic Surgery, Kurashiki Central Hospital,
Okayama, Japan

Accepted for publication April 29, 2013.


Address correspondence to Dr Irie, Department of Thoracic Surgery,
Fig 2. At the site of contact of the leads with epicardium, the pros- Kurashiki Central Hospital, 1-1-1 Miwa, Kurashiki, Okayama, 710-8602,
thesis is sewn over the lead heads to the epicardium. Japan; e-mail: masa_irie@hotmail.com.

Ó 2014 by The Society of Thoracic Surgeons 0003-4975/$36.00


Published by Elsevier Inc http://dx.doi.org/10.1016/j.athoracsur.2013.04.133

You might also like