You are on page 1of 1

Scale parameters and functions

6.3 DR 3 calibration parameters

Variable Note Type Length Read Write Default Min. Max. Modbus
(bytes) write protec- registers
tion
Pro-
tection
Bit 13 Bit 13: Reserve BIT 0 rw r 0 0 1 1036.3
Bit 14 Bit 14: Reserve BIT 0 rw r 0 0 1 1036.2
Bit 15 Bit 15: Reserve BIT 2 rw r 0 0 1 1036.1
Reserve USHORT 2 rw r 0 0 6 1033
Maximum tare Tare maximum FLOAT 4 rw r 100 0 250 1034
load (Page 73) [in % of WRmax ]
Reserve FLOAT 4 rw r 0 0 100.0 1036
Reserve FLOAT 4 rw r 0 0 100.0 1038
Maximum Negative range of FLOAT 4 rw r 1 0 100.0 1040
negative zero the semi-
setting limit automatic zeroing
(semi- [in % of maximum
automatically) weighing range
(Page 73) WRmax]
Maximum Positive range of FLOAT 4 rw r 3.0 0 100.0 1042
positive zero the semi-
setting limit automatic zeroing
(semi- [in % of maximum
automatically) weighing range
(Page 73) WRmax]
Standstill Standstill range FLOAT 4 rw r 1 0 9.999.999 1044
range (in d)
(Page 74)
Standstill time Standstill time 1 in TIME 4 rw r 2000 10 10000 1046
(Page 74) ms
Standstill wait- Waiting time until TIME 4 rw - 0 0 10000 1048
ing time standstill.
(Page 74) 0: standstill-
dependent scale
command, if there
is no standstill,
immediately re-
jected.
> 0: Maximum
waiting time until
command is exe-
cuted
Low-pass filter Low-pass filter 1 - FLOAT 4 rw r 2 0 50 1050
limit frequency cutoff frequency:
(Page 75) 0: Filter disabled
Low-pass filter Filter order USHORT 2 rw r 4 1 4 1052
number Low-pass filter 1
(Page 75)
Reserve Reserve USHORT 2 rw - 0 - - 1053

SIWAREX WP521/WP522
70 Manual, 04/2016, A5E36695225A