Professional Documents
Culture Documents
XAN® 222
XAN® 215
XAN® 220
FISCHERSCOPE® X-RAY XAN®
X-ray fluorescence (EDXRF) measuring instrument
For coating thickness measurements and materials analysis according to the energy dispersive X-ray fluo-
rescence method
Manufacturer
Helmut Fischer GmbH Phone: +49 (0) 70 31 3 03 - 0
Institut für Elektronik und Messtechnik Fax: +49 (0) 70 31 3 03 - 710
Industriestraße 21 www.helmut-fischer.com
D-71069 Sindelfingen mail@helmut-fischer.com
On our home page www.helmut-fischer.com you will find the addresses of our sole agencies and subsidiary
companies around the globe.
Quality Assurance System of the Helmut Fischer GmbH Institut für Elektronik und Messtechnik
DIN EN ISO 9001:2015 Management system certified by Swiss Association for Quality and Man-
agement Systems (SQS)
DIN EN ISO/IEC 17025 Calibration lab accredited for certified mass per unit area standards
© 2022 by Helmut Fischer GmbH Institut für Elektronik und Messtechnik, Germany.
This document is protected by copyright. All rights reserved. This document may not be reproduced by any
means (print, photocopy, microfilm or any other method) in full or in part, or processed, multiplied or distrib-
uted to third parties by electronic means without the written consent of Helmut Fischer GmbH Institut für
Elektronik und Messtechnik.
Subject to correction and technical changes.
FISCHERSCOPE®, XAN® and WinFTM® are registered trade marks of the Helmut Fischer GmbH Institut für
Elektronik und Messtechnik in Germany and other countries.
Note: Designations not marked with ® or ™ may also be protected by law.
2
Table of Contents
1 Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.1 Read the Operator’s Manual Carefully Before Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2 Warning Notices Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.4 Limitation of Liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.5 Safety of the Electrical Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.6 X-Radiation Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.7 Repeat Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.8 Safety Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.9 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.10 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.11 Maintenance and Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.12 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.13 Requirements for the Operating Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.14 Frequently Asked Questions Concerning X-Radiation Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2 Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.1 Areas of Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 Standard-free Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.3 X-Ray Fluorescence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.4 Functional Principle of the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3 Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.1 Cover of the Measurement Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Manually Operable XY Stage (XAN 222) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 Aperture (Collimator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4 Shutter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5 Specimen Lighting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.6 Key Switch for the X-Ray Energy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.7 Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.8 Focus Adjustment Knob on the Right Side of the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4 Set up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.1 Set-up Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.2 Setting up the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3 Establishing Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.4 Installing WinFTM software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.5 Switching on the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.6 Switching off the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.7 Communication between Instrument and WinFTM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3
Table of Contents
5 Position Specimen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.1 Flat Specimens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5.2 Cylindrical or Curved Specimens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
5.3 Measurements on Inclines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
5.4 Position of Jewelry Items. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
5.5 Positioning of Specimen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.6 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
6 Performing Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4 FISCHERSCOPE® X-RAY
Table of Contents
5
Table of Contents
12 DefMA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
12.1 The DefMA Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
12.2 Creating a new DefMA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
12.3 Entering Compensation Spectra. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
12.4 Display the Measurement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
6 FISCHERSCOPE® X-RAY
Table of Contents
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
7
Table of Contents
8 FISCHERSCOPE® X-RAY
1 Safety Information
ATTENTION Indicates a risk that can lead to damage or destruction of the product.
CAUTION Indicates a risk that can lead to light or moderate injuries or material damage.
DANGER Indicates a risk that will lead immediately to death or serious injuries.
IONIZING RADIATION Danger of radiation. The device generates X-rays when you start a
measurement.
9
Safety Information
Modifications, repairs, maintenance and service work on the instrument and its accessories must be per-
formed by authorized Fischer service personnel only.
There is no other intended use. The risk for damage from unintended use lies solely with the user.
10 FISCHERSCOPE® X-RAY
Safety Information
1.9 Start-up
1.10 Operation
Switch off the instrument at least once a month. If you then switch the instrument on again, all components
are reinitialized. This ensures that your instrument works optimally and stably. If you leave the instrument
permanently on, then after approx. one month this message will appear: To ensure optimal performance and
stability, we recommend to restart the instrument regularly. Please turn off the instrument for 10 seconds and
restart it afterwards!
11
Safety Information
1.12 Disposal
The beryllium window on the detector can break and release dust if handled improper-
ly. Beryllium is toxic if swallowed and very toxic if inhaled. Beryllium can cause skin and
eye irritation.
► Never damage the beryllium window.
12 FISCHERSCOPE® X-RAY
Safety Information
Trained Specialists
Trained specialists can utilize the full bandwidth of the technical capabilities of the FISCHERSCOPE X-RAY.
In addition, trained specialists have the following knowledge:
Knowledge of the physical principle of x-ray fluorescence and of the functional principle of the
FISCHERSCOPE X-RAY instrument
The aforementioned physics knowledge is important for defining the measuring application and for other
settings. Only the most important facts are explained in this Operators Manual. This does not replace detailed
training.
The practice-oriented seminars of Fischer are a worthwhile opportunity for obtaining the required knowl-
edge. Several times a year, user seminars that provide insights into the physical fundamentals of x-ray fluo-
rescence and prepare users for their work with the measuring instruments are conducted at various loca-
tions.
Seminar dates can be found on our website www.helmut-fischer.com.
13
Safety Information
Question Answer
Is there a risk that my health will No. The X-RAY instrument is a fully protected instrument. During no
be impaired by X-radiation due to phase of your work at the instrument is it possible for X-radiation to
my work on the X-RAY instru- escape from the measurement chamber. This is also true when the
ment? cover of the measuring head is opened. In such cases, the radiation
source will be switched off immediately through the security devices.
How can I check if the safety de- You can recognize its proper function via the red control LED on the
vice of the measuring head func- front side of the measuring head. Everything is OK as long as the
tions properly? CONTROL LED is not flashing.
Will the specimens that have been No. The fluorescence radiation of the specimen subsides fully al-
exposed to X-radiation them- ready a split-second after the primary X-radiation is switched off. You
selves turn into radiating material can touch the specimen after the measurement procedure without
due to the X-ray fluorescence ef- any problems.
fect after the primary radiation
has been switched off?
Will the primary X-radiation dam- No. The X-radiation has absolutely no lasting influence on the materi-
age or alter the material structure al; it fully retains its quality. The X-ray fluorescence method is a non-
of the specimen? destructive measurement method. Live animals or plants must not
be exposed to the X-radiation in the measurement chamber.
Is the safety or the accuracy of the No. Magnets influence neither the measurement results nor the
measurement results influenced safety. You can use magnets without problems.
by magnets on customer-supplied
fixtures in the measurement
chamber?
14 FISCHERSCOPE® X-RAY
2 Instrument Overview
The FISCHERSCOPE® X-RAY is a high performance energy dispersive x-ray fluorescence (EDXRF) spectrome-
ter.
The WinFTM® (Fischer Thickness Management Software for Windows®) software controls the instrument
and handles the evaluation of the signals supplied by the instrument.
The measured values (coating thickness, material compositions, mass per unit area) are stored and dis-
played on the monitor. Using WinFTM, you can conveniently design the measurement results as a print form
for printout and export them to other applications.
WinFTM runs under Windows®. WinFTM version 6.50 or higher runs only on Windows 10. From version 6.50
on, the older Windows versions are no longer supported.
15
Instrument Overview
Primary X-radiation
Electron
Shell M L K
16 FISCHERSCOPE® X-RAY
Instrument Overview
X-ray tube
Anode
Cathode
Optics WinFTM
Shutter main window
Primary x-radiation
Video camera
Mirror
Spectrum
Detector
Aperture (Collimator)
Coating layer
Base material
Functional principle
1. The x-ray tube generates the primary x-radiation (primary radiation). The electrically heated cathode
emits electrons. Accelerated by the applied high voltage to very high speeds, the electrons bombard the
anode material. This generates the primary x-radiation.
2. The shutter serves as a safety device. It is opened during a measurement only. In case of a malfunction
or when opening the measurement chamber, it will be closed automatically. This ensures, that the user
is never exposed to x-radiation.
3. A light source (not shown in Fig. 2-2) illuminates the sample. A mirror and lens direct the image of the
measurement location to a color video camera. The mirror has a hole in its center for the primary radia-
tion to pass through.
17
Instrument Overview
4. The aperture (collimator) limits the cross-section of the primary beam in order to excite a measurement
spot of a defined size.
5. The primary x-radiation impacts the atoms on the sample surface (coating layer and base material) and
in the process knocks electrons from the inner electron shell. Electrons from outer electron shells fill the
resultant voids emitting a fluorescence radiation that is characteristic in its energy distribution for a par-
ticular material.
6. The energy dispersive detector measures the energy distribution of the fluorescence radiation. A multi-
stage electronics circuit processes the measurement signals.
7. The measured spectrum shows lines or peaks that are characteristic for the chemical elements in the
sample.
8. The WinFTM Software computes the thickness of the coating(s) and/or the analysis result. The video im-
age of the sample is shown in the WinFTM window. The precise position of the measurement location and
the measurement spot is possible due to the special design of the optical and the x-ray guidance sys-
tems.
18 FISCHERSCOPE® X-RAY
3 Components
3.4 Shutter
The shutter is used to keep the primary x-rays from entering the measurement chamber and the environ-
ment.
The shutter is part of the safety system. The shutter closes immediately, when you open the cover of the
measurement chamber or when a malfunction occurs. X-rays can then no longer reach the measurement
chamber and the environment.
19
Components
XRAY On Control
Power
Start Stop On / Off
Shutter Shutter
open closed
LED Function
X-RAY ON Lights red when the X-radiation is ready.
CONTROL Flashes red when the safety system discovers an error or when the cover is not
fully closed, measurements are not possible.
No malfunction message of the safety system is present if the LED does not light.
LED at the START Lights when a measurement is in progress. The shutter is opened.
button
LED at the STOP button Lights when no measurement is in progress. The shutter is closed.
20 FISCHERSCOPE® X-RAY
4 Set up
The set-up location influences the product life and maintenance effort directly.
Corrosive environments cause corrosion on mechanical and electronic components. This can lead to the fol-
lowing problems:
Rough-running or total breakdown of mechanical adjustment mechanisms
Contact problems on plug connectors or total breakdown of electronic components
Salt acid steams or salt aerosols (salt spray test) in the ambient air cause defects on x-ray tubes, proportion-
al counters and silicon detectors in short intervals. Salt acid corrodes the Beryllium window of these compo-
nents. Thus, the components will be damaged rapidly.
If the instrument is located in a room, separated from the electroplating, make sure that the exit air of the
electroplating cannot reach the room via doors, windows or extraction of air systems.
21
Set up
Heavy ambient air pollution causes the ventilation ducts getting clogged. Thus, the maximum permissible
temperature of components can be exceeded. Additional damage can be caused by deposits of aggressive
dirt.
► Set up the instrument in such a way that you always have easy access to the power
cable to disconnect the instrument from the power supply in case of a malfunction
(disconnecting device).
1. Open all containers of the shipment and take out the components.
before all components of the measuring system are connected to one another.
► Do not yet connect the instrument and the PC to the line voltage.
22 FISCHERSCOPE® X-RAY
Set up
4
1 Diagnostic interface, for service
purposes only
I 1 2 Key switch for high voltage
0 3 USB interface, to PC
3 2 4 Main switch
5 Voltage supply for instrument
Figure 4-1: Connectors and switches at the left side of the instrument
1. USB: The USB cable has two different connectors: a quadratic one and a flat one. Connect the quadratic
connector to the instrument and the flat connector to a USB 2 socket at the back panel of the PC. If no free
USB 2 socket is available, connect the mouse or the keyboard to a USB 3 socket.
23
Set up
The following table helps to distinguish between USB 2 and USB 3 sockets. Please note that not all char-
acteristics need to apply, e.g. a USB 3 socket may also be black.
2. Connect the following components to the mains supply using the extension socket supplied. Use the sup-
plied power cables (10 A, 250 V, 3 x 0.75 mm2 ) to connect the instrument and monitor. Never use power
cables with insufficient specifications!
PC
Instrument
Monitor
3. Connection between PC and monitor, depending on the equipment with display port or VGA cable
4. Mouse and keyboard
24 FISCHERSCOPE® X-RAY
Set up
ter about half an hour of measuring time:
► Calibration/Normalization
► Monitoring measurement devices including corrective measures
25
Set up
performance and stability, we recommend to restart the instrument regularly. Please turn off the instrument
for 10 seconds and restart it afterwards!
26 FISCHERSCOPE® X-RAY
5 Position Specimen
There are some basic rules for positioning specimens. For each measurement it must be ensured that the x-
ray fluorescence radiation can reach the detector without obstruction. Otherwise, the measurement result
may be falsified.
Figure 5-1: Positioning of the specimen for cylindrical or curved specimens
Regardless of the positioning, cylindrical or curved specimens must always have a minimum diameter or a
minimum curvature diameter. If the diameters are smaller than this minimum value, then a greater coating
thickness than the actual value will be displayed. This is caused by the effectively inclined radiation through
the coating at the measurement location, which causes a higher fluorescence intensity.
The following applies to simple measuring applications (e.g., single coating applications) and in one plane cy-
lindrical or curved objects):
This means that the diameter of the specimen must be at least four times as large as the diameter of the
measurement spot.
Correspondingly, the following applies to the curvature radius:
For multi-coating measuring applications that are curved in more than one direction (many curved and bent
contacts), this minimum value may be increased by a factor of up to 10 for top coatings, depending on the
measuring application. This factor may be even higher for interim coatings.
In addition, when making such measurements ensure that the position of the measurement spot corre-
sponds to the position in the video image in order to ensure accurate positioning and measurements. If you
have questions in this regard, please contact Fischer.
27
Position Specimen
Information about measurements on thin wires can be obtained from your authorized dealer or from Fischer.
Figure 5-2:
Figure 5-3: Correct positioning of a ring, the measuring spot is fully covered
Figure 5-4: Correct positioning of a chain, the measuring spot is fully covered
28 FISCHERSCOPE® X-RAY
Position Specimen
Observe the rules for positioning different specimen in Chap. 5 on page 27.
5.6 Procedure
29
Position Specimen
30 FISCHERSCOPE® X-RAY
6 Performing Measurements
To perform measurements you need a product file. This file contains all settings for a specific measuring ap-
plication.
When the instrument is delivered by Fischer, products that cover your measurement tasks are usually avail-
able. With these products files you can perform measurements immediately.
If you need new measurement tasks, you can request products files from Fischer. The product files are then
send to you by email or on a data carrier. Before you can use the new product files, you need to import and
then normalize or calibrate them.
For further information see the following chapters:
Procedure
3. Click OK.
4. Position the specimen, see Chap. 5.5 on page 29.
5. To start the measurement,
click Start.
The measurement will be performed.
31
Performing Measurements
32 FISCHERSCOPE® X-RAY
7 User Interface of the WinFTM Software
The WinFTM® (Fischer Thickness Management Software for Windows®) software controls the instrument
and handles the evaluation of the signals supplied by the instrument.
The measured values (coating thickness, material compositions, mass per unit area) are stored and dis-
played on the monitor. Using WinFTM, you can conveniently design the measurement results as a print form
for printout and export them to other applications.
WinFTM runs under Windows®. WinFTM version 6.50 or higher runs only on Windows 10. From version 6.50
on, the older Windows versions are no longer supported.
Title bar
Menu bar
(WinFTM main menu)
Tool bar
Product bar
Statistics field
Measurement
value field
Video window
Status bar
Instruction field
Item Function
Title bar Shows the current folder.
Menu bar Contains all menus. Many of the functions available from the menu bar are also
available from the tool bar.
Tool bar The tool bar offers quick access to the most frequently used functions. You can adapt
the tool bar according to your needs.
33
User Interface of the WinFTM Software
Item Function
Product bar Here you can select a product by clicking the button Select. In the center the number
and name of the current product are displayed, here No. 60 Au/Pd/Ni/Cu.
Video window Here you can see the specimen, see Chap. 7.2 on page 34
Instruction field Here you can start a measurement. Hints and error messages are displayed in this
field (e.g. messages during calibration).
Statistics field Here the statistical results of the last block are displayed, see Chap. 7.5 on page 35.
Calibration status If the measurement application is calibrated, the name of the calibration standard
line set is displayed here. If the measurement application is not calibrated, the message
Calibration: Standard-free is displayed.
Measurement value The measurement readings are displayed here. The field displays either the last
field measurement reading or a list of all readings of the current product, see Chap. 7.6 on
page 36.
Status bar Here important information is displayed, e.g. measurement distance, XYZ coordi-
nates, and the mq-value of the last measurement.
Figure 7-2: Video image, 1) only available at instruments equipped with motor-driven measuring stage
34 FISCHERSCOPE® X-RAY
User Interface of the WinFTM Software
Item Function
Measurement unit Displays the measurement unit for the crosshairs and the measurement spot.
Crosshairs To scale presentation of the crosshairs.
To scale presentation Displays the form and size of the actual measurement spot. Aperture shape and
of the measurement size as well as the measurement distance are considered.
spot
Brightness slide con- Changes the brightness of the video image.
trol
Button for changing the Changes the color of the crosshairs in order to optimize the contrast between
color of the crosshairs specimen and crosshairs.
1 only available at instruments equipped with motor-driven measuring stage
35
User Interface of the WinFTM Software
List Presentation
The list of measurement values are presented in the order of the measurements. In addition to the measure-
ment values, the date and time, the name of the operator and the tolerance limits are displayed as well. The
name of the operator and the tolerance limits are displayed, if they were previously entered.
If you have set up tolerance limits, the measurement value is displayed with colored background. If the mea-
surement value is within the tolerance limits, the measurement value is displayed on green background, oth-
erwise on red background.
36 FISCHERSCOPE® X-RAY
User Interface of the WinFTM Software
Figure 7-3: SPC chart presentation, the upper image shows the mean value over time,
the lower image shows the standard deviation over time
37
User Interface of the WinFTM Software
Display Explanation
Help text If you hover with the mouse pointer over a command button, additional help text
will be displayed here.
Measurement dis- Measurement distance between the specimen measurement spot and collimator
tance minus 15 mm, when the video image is focused.
XYZ coordinates1 Coordinates of the current position of the measurement stage.
mq value Quality factor of the last measurement value. The mq value is a measure for the
agreement of the measured spectrum with the theoretical spectrum, see
Chap 7.9 on page 39.
CAP The caps lock key of the computer keyboard was pressed.
NUM The number pad of the keyboard is activated.
1) only available at instruments equipped with motor-driven measuring stage
38 FISCHERSCOPE® X-RAY
User Interface of the WinFTM Software
7.9 mq Value
The mq value (measurement quality value) is a quality factor of the measurement value. It is displayed in the
status bar.
The mq value is a measure for the agreement of the measured spectrum with the theoretical spectrum. The
mq value is an indication for the confidence level and the trueness of the measurement value. The smaller
the mq-value of a reading is, the more correct is the measurement result. The optimal value for mq is 0.
General recommendation:
39
User Interface of the WinFTM Software
40 FISCHERSCOPE® X-RAY
User Interface of the WinFTM Software
Cursor
Spectral lines
The spectrum window shows the channels [Chan] at the X axis and the count rates [cps, counts per second]
at the Y axis. Each channel corresponds to an energy value.
The spectrum of the X-ray fluorescence radiation is displayed as a distribution curve of count rate contents
(in cps) of a certain number of channels.
At your instrument, this number is between:
Channel # 0 (corresponding to an energy of 0 keV) and
Channel # 1023 (corresponding to an energy of approx. 33 keV).
You can change the display of the X axis to show the energies directly.
41
User Interface of the WinFTM Software
42 FISCHERSCOPE® X-RAY
User Interface of the WinFTM Software
Procedure
1. To start a measurement, click .
The measurement runs until you terminate it.
2. To terminate a measurement, click
Procedure
1. Select in the Spectrum window View > Elements.
You can display the extended elements list. To do so, select in the Spectrum window
View > Full element list.
2. Click into the spectrum window, keep the mouse button pressed and move the cursor to a peak of the
spectrum.
A list of radiation components of one or more elements is displayed on the left next to the spectrum win-
dow.
3. Click in the elements list on one radiation component.
The energy lines associated with this element appear as colored lines in the spectrum window. You can
compare the peaks of the measured spectrum with the superimposed lines to identify the elements that
comprises the sample.
4. Repeat steps 2 to 3 to determine further elements of the sample.
5. If the specimen contains multiple elements, you can determine the composition of the specimen and au-
tomatically create a DefMA for analysis.
43
User Interface of the WinFTM Software
To determine the composition of the specimen, click on in the toolbar. The composition of the
sample is then displayed in the window to the left of the spectrum.
To automatically create a DefMA based on the calculated analysis values, click on the toolbar on .
You can change the colors of the spectra and the energy lines. To do so, select in the spectrum
window View > Color….
7.20 Energy
The energy distribution contained in the spectrum informs about the composition of the sample, because
each occurring element of the sample reflects the qualitative composition through its peak position (at the
horizontal position specified for this element).
The higher the X-ray fluorescence energy of an element, the further to the right the corresponding line will be
in the spectrum.
According to Moseley’s Law, the energy increases as the atomic number of the respective element increases,
i.e., the lines of the “heavier” elements (those with a higher atomic number) are further to the right in the
spectrum.
The energy is measured in keV = kilo electron volts. 1 eV is the energy that an electron gains as it moves
through a field with a potential difference of 1 V.
7.21 Intensity
The intensity of a radiation is the amount of radiation that strikes a certain unit of area in a specified time
period.
In X-ray spectroscopy, the intensity of the radiation is measured as the count rate in counts per second (cps).
Each peak in the spectrum corresponds to a relative maximum referenced to its surrounding. This maximum
correlates to the amount of radiating atoms of the corresponding element.
The higher the peak, the more atoms of the respective element are present in the sample, i.e., the higher the
concentration of the respective element in the sample.
Thus, the intensity of the spectral line is a measure for the quantitative composition of the sample.
44 FISCHERSCOPE® X-RAY
8 WinFTM File Structure
To ensure correct measurements, a set of specific files needs to be set up. The process flow for setting up the
files needed for measurements, is the same on all FISCHERSCOPE X-RAY instruments:
Product
Measurement
The essential content of the files product, measurement application and DefMA and their relations among
each other is shown in the following figure:
Product
Measurement results and their presentation
Image of the specimen
Measurement time and number of measurements
Export masks, block protocol etc.
Measurement Application
Calibration data (Calibration standard set)
Measurement time during calibration
Number of measurements during calibration
Anode current, aperture
DefMA
Measurement mode
Type and order of the coatings
Tube voltage
Type and composition of the calibration standards
Treatment of interfering spectra
Special evaluation methods
Figure 8-2: Content of the files DefMA, measurement application and product
45
WinFTM File Structure
8.1 DefMA
The DefMA Definition of the Measurement Settings of the Application) is a file, in which the basic settings for
a measurement are saved, like:
Measurement mode
Tube voltage used
Type and order of the coatings
Type and composition of the substrate material of the specimen and of the calibration standards
Treatment of interfering spectra
Special evaluation methods
To create a DefMA you need the WinFTM extension SUPER. The usage of the SUPER software requires knowl-
edge about the physics of x-ray fluorescence and the functionality of the instrument.
If you don’t have the knowledge, do not use the SUPER software, but request a new measurement application
from Fischer instead. Fischer offers a professional service for the creation of customer’s DefMA.
8.3 Product
The product is a file, which serves for the data organization and the measurement control. Everything you
define and enter for solving a measurement task is saved in the product file:
Measurement results and their presentation
Image of the product
Measurement time and number of measurements
Export masks
46 FISCHERSCOPE® X-RAY
WinFTM File Structure
Product
1. In this context a calibration standard set is a particular data set, which contains the characteristics of
the used calibration standards.
47
WinFTM File Structure
48 FISCHERSCOPE® X-RAY
9 Measurement Product Administration
The instrument needs certain basic information to generate measurements. This information is contained in
the product file (hereinafter referred to as product), see also Chap. 8 on page 45. For this reason a product
that fits your specimen must be selected or created before performing a measurement.
C D
E
2. In the A field change to the directory that contains the product you want to copy.
3. In the B field select the product you want to copy.
4. In the C field change to the directory where the new product is to be created.
If you want to create a new directory:
In the D field click the icon to create a new directory.
In the C field change to the newly created directory.
5. In the E field enter a name and number for the new product.
6. Click Copy (Field F).
7. To close the window click Cancel.
8. To measure with the new product, you must select the new product. To do so click .
49
Measurement Product Administration
Procedure
1. In the WinFTM main window select Product > New… .
The window Select an application appears.
2. Shall the new product base on an existing measuring application or on a new one?
Existing measuring application:continue with Step 8.
New measuring application:continue with Step 3.
3. In the window Select an application click New.
The window Select a DefMA appears.
4. Select the DefMA and click OK.
The window Input of the application name appears.
5. Enter in the fields the Name and No. of the measuring application and click OK.
The window Input Calibr. Std. Set appears.
6. Do not enter anything in the window Input Calibr. Std. Set and click OK.
The window End of calibration. Accept? appears.
7. Click Yes. Continue with Step n
8. Select the measuring application and click OK.
The window Create new product appears.
Enter in the fields the Name and No. of the product. Further options:
To rename the folder, click .
To create a new folder, click .
9. Click OK.
The new product is now created and active.
50 FISCHERSCOPE® X-RAY
Measurement Product Administration
Procedure
1. In the WinFTM main window select Product > Modify… .
The window Modify product appears.
2. You can change many properties of the product, e.g.:
Procedure
1. If you have received the product file by email, save the file in a directory that does not belong to WinFTM,
e.g. C:\temp.
2. If you have received the product file by CD, insert the CD into the CD drive.
3. Optional, if you have installed the WinFTM extension PDM: You can create a new directory for the new
product as follows:
a In the WinFTM main menu, select Product > Select ...
The window Select Product appears.
b Click in the upper right .
4. In the WinFTM main window select Product > Copy to/from File… .
The window Copy from/to file appears.
5. Select the tab File >>> Product.
6. Click Select.
51
Measurement Product Administration
Procedure
1. In the WinFTM main window select Product > Modify… .
The window Modify product appears.
2. Select the tab Application.
3. Click Other Application.
The window Modify application appears.
4. Select the measuring application and click OK.
52 FISCHERSCOPE® X-RAY
Measurement Product Administration
Procedure
1. In the WinFTM main window select Calibrate > Delete application… .
The window Delete application appears.
2. Select the measuring application you need to delete and click OK.
If the measuring application is assigned to one or several products, the following message appears for
each assigned product: Application is used in: <Folder Name>, Product No. <Product Number> <Product
Name> Delete this product?
To delete the product assigned to the measuring application, click Yes.
53
Measurement Product Administration
Procedure
1. If you have received the DefMA file by email as a file attachment, save the file in a directory that does not
belong to WinFTM, e.g. C:\temp.
2. If you have received the DefMA file by CD, insert the CD into the CD drive.
3. In the WinFTM main window select Product > Copy to/from File….
The window Copy from/to file appears.
4. Select the tab File >>> DefMA.
5. Click Select.
The window Select file appears.
6. Select the DefMA file and click Open.
The selected path appears in the field File.
7. Click Start copy.
The file will be copied. After this the report The DefMA was copied successfully appears.
8. Click OK.
9. To close the window, click cancel.
54 FISCHERSCOPE® X-RAY
10 Handling of Measurement Readings and
Statistical Evaluations
10.1 Block
A block is a test series that combines several readings that have been made under the same conditions or
that belong to the same order.
All single readings of a block are numbered consecutively and each block, in turn, has a consecutive number
as well.
55
Handling of Measurement Readings and Statistical Evaluations
To label a block
1. In the WinFTM main window select Evaluation > Order No. / Operator….
The window Order No. / Operator appears.
2. Fill in the fields. All blocks will be labeled with your entered values from now on. In doing so, the values
are taken as long as you enter new values.
56 FISCHERSCOPE® X-RAY
Handling of Measurement Readings and Statistical Evaluations
57
Handling of Measurement Readings and Statistical Evaluations
58 FISCHERSCOPE® X-RAY
Handling of Measurement Readings and Statistical Evaluations
4. Click Yes.
If you click No, then the measurement reading appears at the end of the list.
Procedure
1. In the WinFTM main window select Evaluation > Evaluate Single Block….
The window Block Result appears.
2. You can now read the statistical data, move quickly between blocks or delete blocks.
Procedure
1. In the WinFTM main window select Evaluation > Evaluate All Blocks….
The window EVALUATION appears.
59
Handling of Measurement Readings and Statistical Evaluations
60 FISCHERSCOPE® X-RAY
11 Setting Up the Print Form Using PDM
Functions
Insert variable
Enter any desired text
Insert Bitmap graphics (e.g., company logo)
61
Setting Up the Print Form Using PDM
62 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
@ANB Order number of the selected block or the information that has been entered into the first entry
field of the dialog window Order No./ Operator.
@ANR Number of the current product
@ART Name of the current product
@BDT Date of the selected block, i.e., the date, when the last measurement of the block was taken
@BM1 Lot number of the selected block or the information that has been entered into the second entry
field of the dialog window Order No./ Operator
@BNR Block number of the selected block
@BTM Time of the selected block. That is, the time when the last measurement of a block was taken.
@CK1 Critical process capability index Cpk
Prerequisites for computing and inserting the process capability indexes:
The control box with Cp and Cpk from the product tab Presentation (Product > Modify product) is
checked and the control box Specification limits is checked
@CP1 Process capability index Cp
Prerequisites for computing and inserting the process capability indexes:
The control box with Cp and Cpk from the product tab Presentation (Product > Modify product) is
checked and the control box Specification limits is checked
@DAT Current date for the print form
Default format: DD/MM/YYYY, e.g., 24.05.2021
@DF1 Density of the coating that has been entered into the current product. This applies only to coat-
ings with a fixed composition
@DIR Name of the directory (folder), where the current product is saved
@DM1 Unit of measurement (e.g., µm, mils or %) of the quantity to be measured
63
Setting Up the Print Form Using PDM
@FIS Inserts the header with the instrument name into the print form. If this variable is not used in the
print form template, it will automatically be placed at the beginning of the print form template,
when saving. The header will then appear at the beginning of the print form; i.e., the header can-
not be deactivated.
@MA1 Largest single reading in the block = Maximum
@MAB Name of the measurement application that the current product accesses
@MI1 Smallest single reading in the block = Minimum
@MNR Number of the measurement application that the current product accesses
@MW1 Mean value of the single readings of the block
@N<1 Number of readings in the block that fall below a preset lower specification limit (LSL)
@N>1 Number of readings in the block that exceed a preset upper specification limit (USL)
@NB1 Number of readings present in the block that contribute to the statistical block result
@NCAL Number of single readings that are taken on the calibration standard during the calibration of
the current measurement application (the current meas. application is that meas. application
that the current product accesses), see also @TCAL
@NTZ Text content of the notepad of the product. Enter/edit notes in the product: Product > Modify
product, Tab Notepad
@OG1 Upper specification limit USL (upper tolerance) for the presentation and computation of the up-
per limit violations in a block
@P<1 Percentage of single readings in a block that fall below the lower specification limit LSL
@P>1 Percentage of single readings in a block that exceed the upper specification limit USL
@PRF Operator of the selected block, or the information entered into the third entry field of the dialog
window Order No./ Operator
@RG1 Range, difference between the largest (maximum) and smallest (minimum) single reading of the
selected block
@SD1 Standard deviation of the single readings of the selected block. The standard deviation s is a
measure for the distribution of the single readings of the measurement series around their
common mean value. It equals the mean square deviation of the single readings from the mean
value and is computed according to the equation found in the appendix
@STF The text Calibration standard-free appears in the print form at that location where the variable
@STF has been inserted in the print form template.
Prerequisite: The current product accesses a measurement application that has been calibrated
standard-free, that is, no calibration standards have been entered into the calibration standard
set of this measurement application and used for the calibration.
If this variable is not used in the print form template, it will automatically be placed at the begin-
ning of the print form template. Then the text Calibration standard-free appears at the beginning.
@TCAL Measurement time (in seconds) for the single readings that are carried out during the calibra-
tion of the current measurement application on every calibration standard. The current measur-
ing application is that meas. application that accesses the current product, see also @NCAL
@TIM Current time at the time the print form is printed
Format: hh:mm:ss, e.g., 13:45:30
64 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
@ANB Order number of the last block of the selected blocks or the information that has been entered
into the first entry field of the dialog window Order No./ Operator.
@BM1 Lot number of the last block of the selected blocks or the information that has been entered into
the second entry field of the dialog window Order No./ Operator.
@EDB Number of the last block of the selected blocks.
@EDT Date of the last block of the selected blocks.
@ETM Time of the last block of the selected blocks.
@HG1 Histogram of all single readings of the selected blocks in the current product, see also @WN1
@OE1 Upper control limit UCL(d) of the mean values of an SPC chart.
@OG1 Upper specification limit USL (upper tolerance) for the presentation and computation of the up-
per limit violations in the selected blocks of the current product.
@OR1 Upper control limit UCL(s) or UCL(R) of the standard deviation s or the range R of an SPC chart
@P<1 Percentage of single readings in the selected blocks of the current product that fall below the
lower specification limit LSL
@P>1 Percentage of single readings in the selected blocks of the current product that exceed the up-
per specification limit USL
@PRF Operator of the last block of the selected blocks or the information entered into the third entry
field of the dialog window Order No./ Operator
65
Setting Up the Print Form Using PDM
@RG1 Range = Difference between the largest (maximum) and smallest (minimum) single readings of
the selected blocks of the current product.
@SA1 Standard deviation s_a.
This standard deviation expresses the additional distribution of the mean values of fixed blocks
that cannot be explained by the standard deviation s of the single readings within the blocks.
Thus, it is an additional distribution that is based on causes different from the distribution of the
single readings. This characteristic statistical quantity is available only with fixed block sizes.
@SD1 Standard deviation s of the single readings of the selected blocks in the current product.
The standard deviation s is a measure for the distribution of the single readings of the measure-
ment series around their common mean value. It equals the mean square deviation of the single
readings from the mean value and is computed according to the equation found in the appendix
For fixed block sizes @SD1 is the standard deviation of the block mean values.
@SDT Date of the first block of the selected blocks in the current product. The date of the last measure-
ment of the block is used as the date for this block.
@SP1 Statistical process control chart (SPC chart) for the attribute 1 of the sample
@STB Number of the first block of the selected blocks in the current product.
@STM Time of the first block of the selected blocks in the current product. The time of the last mea-
surement of the block is used as the time for this block
@UE1 Lower control limit LCL(d) of the mean values of an SPC chart.
@UG1 Lower specification limit LSL (lower tolerance) for the presentation and computation of the vio-
lations of the lower limit in the selected block of the current product.
@VK1 Coefficient of variation of the single readings of the selected block. The coefficient of variation is
the standard deviation in percent based on the mean value and is computed according to the
equation found in the appendix.
For fixed block sizes, @VK1 is the coefficient of variation of the block mean values.
@WN1 Normal probability chart of the single readings of the selected blocks in the current product. See
also @HG1
@BDT, @BTM and @BNR:
Are available for the block print form template
Refer to the individual block
Are not available for the final print form template
Select block:
Menu command Evaluation > Block selection
66 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
Notes
The variables must always be entered in upper case letters
The character @ is part of the variable, i.e., it must be entered, must not be deleted and it must not be
followed by space
Images that are located in the clipboard can be inserted in the print form template
Edit print form or print form template > Edit > Insert
Images saved in the BMP format can be inserted in the print form template with the variable @VIF.
67
Setting Up the Print Form Using PDM
11.8 Printing the Print Form Using the Supplementary Software PDM
Print forms can be printed very easily when using the supplementary software PDM.
Select the print form using the menu commands OR the command buttons:
Call using the menu command: Call using the menu command:
Evaluation > Print form block evalua- Evaluation > Print form final evalua-
tion tion
The final print form evaluates only those blocks that have been selected before for the final print form using
Evaluation > Block selection or the button Block selection.
68 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
69
Setting Up the Print Form Using PDM
PTB [German Physical and Technical Institute] uses the term "Überdeckungswahrscheinlichkeit" [overlap-
ping probability], which is identical with the term "confidence interval" used in statistics literature.
Assuming a normal distribution probability density for the difference to the true value, one identifies its dis-
tribution parameter s with the standard measurement uncertainty u, and using multiples thereof, can specify
well-defined measurement uncertainties. The often recommended expansion factor k shall be 2, such that a
range X + 2*u contains the true value with a 95% probability.
For several reasons, k=1 is meaningful as well, because here the standard deviation reaches the random
portion (see below) of the measurement uncertainty for a large number of repeat measurements. However,
here the hit rate is only 68%.
Safety conscientious companies and people prefer to allow less than 1% probability for errors - and select
k=3. In the end, however, it does not matter in what unit the measurement error is specified; important is that
it is known.
To meet different requirements WinFTM allows for the input of k. The default value is set to k=2. If you need
to change k, contact your Fischer Service Representative.
70 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
Figure 11-1: Calibration standard set. Input of the nominal values of the standards. The uncertainties (errors)
must be entered as well. If the pre-set Zero is not changed then the program will automatically
use 1% of the nominal values for coating thicknesses and area masses, and 1 mass % for concen-
trations as the uncertainty, referenced to k=1 (68% confidence interval).
The example applies to the measurement application Au/AuCuCd/Ni/CuZnPb.
71
Setting Up the Print Form Using PDM
For program-technology reasons, the systematic portion of the error is available only via On-line-Export.
Figure 11-2: Setting of the error display for screen output and export in the
"Calibration standard set". Default setting: Systematic portion.
The systematic portion cannot be determined if no calibration has been carried out. It will be set to Zero by
the program.
Switch Measurement uncertainty Error specification for single Error specification for mean val-
display readings ues
Random portion Theoretic error estimate from the Confidence interval of the mean
measured spectrum value according to t-distribution
(student-V.), computed from the
standard deviation of the block.
Systematic portion Computed from the random error Uses the error value of the last
of the calibration measurement measurement of the block
and the uncertainties of the nomi-
nal values of the standards, ac-
cording to the input in the "Cali-
bration standard set"
Systematic + random portion Root of the squared sum of the Root of the squared sum of the
random and the systematic por- random and the systematic por-
tions tions
72 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
73
Setting Up the Print Form Using PDM
74 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
Measurement
Product handling
Block handling
Evaluation
75
Setting Up the Print Form Using PDM
p24 Exports all results of block No. n. If n=0, then the current block will be exported
f62 Online Export ON
f63 Online Export OFF
f82 Exports the spectrum from ROI1 to ROI2 to the PC
Accessories
f42 S Creates a text box on the screen with Text = S, returns TRUE or FALSE, (Button [OK] or
[Cancel])
f50 - Starts a reference measurement
f64 N Sets a delay (pause) of N sec.
f90 Closes a window if it is not the main window. If no window can be closed then return code
= ERROR
f91 Resets X-RAY, new PC-Communication, new HV control etc.
p25 Saves the current spectrum
p98 Single step mode: Task is carried out with single steps by pressing F4
p99 Single step mode (p98) OFF
p80 S,L Symbolizes a logical relationship according to the table below. In case of TRUE (the corre-
sponding relation is fulfilled) the task jumps to label L
76 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
S Meaning
„i“ represents the number of the measurement data column
> Ki For reading i (i=1,2,3) of the current set of readings applies: Reading > USL
< Ki For reading i (i=1,2,3) of the current set of readings applies: Reading < LSL
> KX For at least one (any) reading of the current set of readings applies: Reading > USL
< KX For at least one (any) reading of the current set of readings applies: Reading < LSL
?Ki The reading i (i=1,2,3) of the current set of readings is not measurable (exactly the reading
designated with the consecutive number i is meant)
?KX At least one reading of the current set of readings is not measurable (any reading of the
current set readings is meant)
77
Setting Up the Print Form Using PDM
The number of atoms or the mass per unit area can be determined from the fluorescence spectrum. Howev-
er, this area-related mass AM (e.g., in g per m² or mg per cm2) is not the typically preferred unit of measure-
ment in coating technology - rather, the coating thickness Th (e.g., in µm or mils) is normally used for the
designation of the coating process or for the specification of the coated product and is usually requested in
the supplier’s document.
Thus, the mass per unit area needs to be converted into the coating thickness.
The correlation between mass per unit area AM and the geometric coating thickness th is:
AM
(1) th =
ρ
where ρ designates an effective density (in g/cm³) that represents the mass of the coating element per unit
volume.
In combination with a thickness measurement per cross cut, i.e., at a known coating thickness th, the density
ρ of the coating can be computed from it using equation (1). Other coating thickness measurement methods
for determining the geometric coating thickness may be used as well.
It is then possible to adjust the two methods through proper selection of the density such that they provide
the same measurement result for the geometric coating thickness.
However, a prerequisite is that the density does not fluctuate significantly during the coating process. This
would lead to deviating coating thickness readings when using the X-ray fluorescence method.
Physical background
An XRF coating thickness test instrument evaluates the intensities of the radiation of the coating and sub-
strate materials. The intensity of the characteristic radiation of the coating material is determined from the
number of atoms within the excited sample volume.
The intensity of this radiation is higher as more atoms of the sample volume are excited by the primary radi-
ation and emit their radiation to the detector.
However, the following applies to the radiation of the substrate material:
The higher the intensity of the characteristic radiation of the substrate material, the lower is the coating
thickness.
Reason: The absorption of the substrate material radiation penetrating the coating is lower through a “thin-
ner” coating.
The same applies to multiple coatings and coatings comprised of several elements.
78 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
The following assumption is made for the quantitative computation of these effects in the defined test area
(sample volume) under the surface to be measured (measurement spot):
The excited atoms are distributed homogeneously across the entire volume, i.e., the specific density of the
material is constant in the entire volume that is to be measured.
Essentially, the measurement of the x-ray fluorescence is an indirect count of the excited atoms under a sur-
face with a precisely defined size, the so-called “measurement spot”.
11.22 Correlation between Mass per Unit Area and Coating Thickness
If a coating is comprised of several elements, the instrument initially measures the mass per unit area of
each individual element.
Then the respective masses per unit area of all elements are added taking into account their weight propor-
tions, and a mean value is computed.
The division of the mean value of the mass per unit area through the weighted mean density provides the
sought coating thickness th:
AMmean
(2) th =
ρmean
AMmean Mean value of the mass per unit areas of all elements
1 1 1 1
ρmean Mean specific density = + + + ...
ρmean ρ1 c1 ρ2 c2 ρ3 c3
1, 2, ... Continuous number of the elements in the analyzed coating
c Element concentration
th Coating thickness
79
Setting Up the Print Form Using PDM
The computation assumes a homogeneous composition of the sample. When preparing the sample, ensure
that a representative density is stated. The analysis may provide a wrong result if samples are not homoge-
nous and only a small portion of the sample is used for the analysis.
Important: The thickness measurement using X-ray fluorescence assumes a well-defined and constant den-
sity ρ .
The two schematic drawings in the figure below show that the X-ray fluorescence method is actually blind to
the (geometric) thickness.
Two coatings with the same geometric thickness are presented here (prerequisite: same material, i.e., coat-
ing element).
The presentation of process 1 shows the atoms of the coating being stacked “loosely”. However, this cannot
be measured as the spectrum does not show a measurable line. The reason may be that an additional “ma-
trix element” is present in the coating.
The presentation of process 2 shows that the atoms are significantly closer to one another.
Correspondingly, the density is greater than the one shown in Figure 1. The XRF method can provide the cor-
rect (geometric) coating thickness if the measurement application that is used in process 1 uses a lower den-
sity for converting the mass per unit area into coating thickness according to equation 1 than the one of pro-
cess 2. Otherwise, the measurement result would not correspond to that of a method that measures the geo-
metric thickness directly.
I
Figure 11-7: Illustration of the density issue: Although the linear expansion of the two coatings is the same, the
density (and with it the mass per unit area) of process 1 (left) is significantly lower than that of
process 2 (right).
In practical applications, this issue often becomes apparent and is the cause of deviating results between
supplier and customer. This is most often the case if the coating contains elements other than the main ele-
ment and the instrument or measurement application does not or can not measure these elements. In such
cases, the density must be lower than the one that was to be used if only the main element was present.
Examples
Precious metal coatings, e.g., gold: Here, instead of the density of pure gold of 19.3 g/cm³ found in tables,
a density of, for example, 17.5 g/cm³ as stated by the manufacturer of the gold bath or as agreed upon
between the parties is used.
The reason for this is that the bath contains elements or compounds (e.g., cyanide) that are included in
the coating but not visible in the X-ray fluorescence spectrum. However, the used density should be pro-
vided in the supply documents and should be part of the specifications; after all, it determines or not the
products meet the supply specifications.
80 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
Hard material coating, e.g., TiN or TiC: Here, C or N is incorporated in the coating but remains invisible in
the fluorescence spectrum. The (effective) density can be determined according to the method of com-
parison with a cross cut as mentioned above.
All stored density values within the measurement application of WinFTM are values taken from tables
(source: Periodic Table of Elements, Heräus GmbH/Hanau, Reference 5C 3.81\N Ko). However, they can be
changed by the user.
Figure 11-8: Dialog window “DefMA”. The current density setting for each element is shown in the left window.
81
Setting Up the Print Form Using PDM
Figure 11-9: Dialog window “Definition of elements“. The density of the current element can be entered or ed-
ited in the entry box “g/cmˆ3”.
Calibrating
In the context of this recommendation, calibrating in the spirit of the instrument technology refers to an ”ad-
justment”. In the instrument technology, re-measuring the calibration standards in order to verify that the
readings obtained with the measurement application are still correct is referred to as ”calibration”. Still, for
practical and historical reasons and for the necessary agreement with the Standards DIN 50987 and ISO
3497 (Coating Thickness Measurement Using X-ray Fluorescence) we will continue to use these once select-
ed designations, especially since these designations are also used in the software of the instruments.
Each measurement application is calibrated with calibration standards that should in their structures be as
close as possible to the sample but may, however, differ from the samples in their coating thickness and to a
limited degree in the composition of the coatings and of the substrate material. The calibration standards
must exhibit known coating thicknesses, and for the analysis, known compositions (nominal values).
82 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
To this end, the supplier of the calibration standards measures them against higher-value master standards
with secured nominal values.
A certificate confirms the traceability, describes the method how the nominal values have been determined
and may also indicate the test conditions, the nominal values and their measurement uncertainties.
Through the calibration, the nominal values of the standards are transferred to the measurement applica-
tion. Thereafter, the measured coating thicknesses and concentrations are correct, as long as the calibration
standards are labeled correctly. After each calibration, the calibration is checked by re-measuring the stan-
dards to ensure that all entries have been correct and that the calibration indeed leads to correct results.
These check measurements should be documented in the same manner as all other test measurements.
Trueness
The instrument will then measure the coating thicknesses of other specimens that correspond in coating
structure and substrate material to the measurement application correctly as well. The trueness of mea-
surements is a quality that is particularly important because supply contracts for coated products often re-
quire the observance of specification limits. Wrong or incorrect measurements can then lead to unwarranted
acceptance or rejection. The trueness aids in avoiding non-conforming measurement results both within the
production and between supplier and customer. A measurement must be based on calibration standards and
be traceable. Only then is it correct. If calibration standards do not exist or are not available for certain mea-
surement applications, then reference samples that have been measured using a calibrated measurement
application or have been measured standard-free shall assume their roll.
Strictly speaking, a correct measurement is secure only for such samples that correspond to the calibration
standards in their coating structure and composition. WinFTM allows for certain differences between the
structure of the sample and that of the standard, e.g., differences in the substrate material. Although these
differences are physically taken into account correctly, for formal reasons they require documentation.
83
Setting Up the Print Form Using PDM
ment device monitoring for the calibration standards. We recommend to treat the standards gently, to store
them according to instructions and to have them tested or re-certified regularly, e.g., once a year. Certified
calibration standard sets or so-called application kits are available for most of the typical measurement ap-
plications. They are necessary to verify the trueness of the measured values in cases of dispute and to mini-
mize product liability risks.
Measurement Uncertainty u
The measurement uncertainty u of the mean values is derived from the standard deviation s and the number
n of the individual readings included in the test series:
(4) u = c(n)s
where c(n) is a factor that is dependent on the number of single readings and the so-called confidence level
1-. For n = 5, 10, 20, 30 and 50 (and a confidence level of 95%) the resultant factors are:
c(5) = 1.24 ---- c(10) = 0.71 ---- c(20) = 0.48 ---- c(30) = 0.37 ---- c(50) = 0.28.
Confidence Interval
The measurement uncertainty forms the confidence interval when deducted as difference from and added as
sum to the mean value (plus/minus). It is an interval around the mean value of the test series that contains
with a 95% probability the true (overall) result that one was to obtain if one were to measure infinite times. In
measurement technology, the results are always displayed graphically in the same manner by drawing the
mean value as a circle or point and the confidence interval as an ”error bar” from this point (”bar-bell”). The
confidence interval becomes narrower as the number of single readings that form the mean value increases
and the better the repeatability precision of the instrument, or respectively, the smaller the standard devia-
tion of the single readings is (see above equation for the measurement uncertainty u and the factors c(n)).
Variations
The variations of an X-ray fluorescence instrument depend on the measurement parameters measurement
time, distance between the detector and the specimen, intensity of the primary beam and other characteris-
tics of the instrument as well as on the physics of the X-ray fluorescence with regard to the measurement
application. It is interesting that with the X-ray fluorescence method, the variations of the readings can be
calculated theoretically from the spectrum of a single reading such that a measure for the variation can be
84 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
specified with one single reading. The standard deviation is used to evaluate the repeatability precision of the
instrument with regard to the measurement application, the feature to be measured and the setting param-
eters of the instrument. The better the repeatability precision of the instrument, the less the measurement
results will be ”falsified” by the instrument. Thus, a good repeatability precision is considered an important
quality feature of an instrument and is of particular importance when evaluating the suitability of a measure-
ment device.
Guaranteed Error
The calibration standards are subject to a ”measurement uncertainty” as well because their nominal values
are obtained through measuring. For this reason, the nominal values of the standards also include an error,
the guaranteed error (e.g., 5% of the nominal value) that includes the mean value, and its interval should in-
clude the true value.
Accuracy
Trueness Precision
Repeatability Reproducibility
The repeatability precision of an instrument must not be confused with the trueness of a reading. The true-
ness of a measurement result is determined by comparing the mean value B of a test series of a sample, for
example a calibration standard or a reference sample, with its known value A that has been recognized as
being correct, the nominal value. The trueness is represented by the difference B – A between the mean val-
ue and the nominal value. The measurement result is correct if the difference B – A is near Zero.
Please note that we expressly speak of the trueness as a property of the measurement readings. However,
many believe the trueness to be a property of the instrument! Unprofessionally, this is expressed in the ques-
tion for the ”accuracy of the instrument”. However, the trueness is obviously very dependent on the settings
(e.g., the calibration) and the monitoring of the instrument. These in turn depend on the competency and de-
pendability of the user. For this reason, the instrument manufacturer is not able to guarantee the trueness of
the values measured with this instrument or even apply a sticker to the instrument stating the trueness.
Each measurement result is as correct or wrong as allowed by the user. Thus, the trueness cannot be a cri-
terion for the quality of an instrument, at best it can be an indication of the stability (trueness during long-
term operation of the instrument without necessary interim corrective measures).
In measurement technology, the term ”accuracy” is essentially meaningless and of little relevance. In con-
trast to it, the terms trueness and precision that are unambiguous (contrary to the term accuracy) are pre-
ferred. On the international scene of standardization, the situation is similar: the term accuracy can be ex-
plained only qualitatively; it is split into the terms trueness and precision. The term precision in turn is divid-
85
Setting Up the Print Form Using PDM
ed into repeatability (repeatability precision = precision under repeatable conditions) and reproducibility
(comparative precision = precision under comparable conditions).
86 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
number of measurements) and in this manner the confidence interval made narrower. If this establishes a
systematic deviation, it must be researched which of the two results is correct and which one is not or even
if both are false. How to do this will be shown in the next section.
Precision
In contrast to trueness, precision is an instrument property. It can be easily characterized, e.g., by repeated
measurements on a sample (whose true value must not even be known) at one and the same location under
repeatability conditions. The mean value produces the requested measurement reading and the standard de-
viation of the test series is a measure for the instrument’s precision. If the measurement application has
been calibrated correctly, then the mean value of the test series is also correct and corresponds to the true
value (which is the mean value that results from infinitely repeated measurements) within the confidence
interval.
However, if one knows the correct measurement result of a sample at a marked location and measures it
with a different instrument and by a different user, the measurement result is typically different. This can be
observed, for example, with round robin tests on the same sample with a marked reference area, even if
each instrument is correctly calibrated and checked. The reasons are the random deviations of each single
reading, and thus of each mean value of a test series, the random deviations at each calibration, and also the
fact that each calibration standard is labeled with a nominal value that through its measurement is also af-
flicted with an unavoidable deviation from the true value (expressed by the uncertainty of the standard). The
sum of these deviations is expressed in the comparative precision of the measurement method. The repeat-
ability precision characterizes the random deviations of the results that are obtained at one instrument by
one person with one and the same sample at the same measurement location at repeated measurements,
i.e., under repeatability conditions. Comparative precision describes the random deviations between different
instruments, calibrated with different calibration standards, at repeated measurements performed by differ-
ent persons at the same sample using the same measurement locations (measurements under comparable
conditions).
Theoretically, there should be trust between the supplier and customer. However, the reality is: ”Trusting is
good, inspecting is better!”. One of the most important prerequisites for the inspections is correct measure-
ment readings. However, different readings between supplier and customer should generally not be taken as
a cause to continuously doubt the measurement results or measurement methods of the other party or to
demand arbitration measurements with different methods that are to verify that one’s own measurement is
correct. First one should investigate whether the measurement results have been obtained at the same mea-
surement spots and the same specimens. To answer this question, supplier and customer should exchange
and measure the same samples with marked measurement spots. If the results are still different, the in-
87
Setting Up the Print Form Using PDM
volved instruments should be calibrated using certified calibration standards. This determines which instru-
ment measures correctly and which instrument requires corrective measures. If the differences remain and
the problem cannot be resolved, only a well-founded consultation through experts of the instrument manu-
facturer may help in the end.
Each user who monitors the measurement devices as described herein, can verify the trueness of his own
measurements at any time. He can also trust his measurement results and will not be troubled by unwar-
ranted complaints. In practical situations, the systematic deviations, which almost always arise from inade-
quate or infrequent monitoring of the measurement devices, occur more significantly and more frequently
and above all are of a greater magnitude than the random deviations between the instruments of different
parties. Typically, the random deviations are hardly noticed and if they are noticed, they are accepted due to
their insignificance. Thus, random deviations as opposed to systematic ones can usually be neglected as the
reason for measurement reading differences. It is not without reason that the X-ray fluorescence method is
considered the most accurate measurement method for coating thicknesses due to its high comparative pre-
cision. Because the most common and grave deviations are of a systematic nature, they must be found and
corrected. This is the task of measurement device monitoring.
Reference samples
Reference samples are considered derivative or secondary calibration standards. There are no alternatives
to reference samples if no primary standards are used or use of an annual calibration service is made. When
checking the standards, a potentially existing difference between the substrate materials of the sample and
the substrate material of the calibration standard must be taken into account: prior to the check measure-
ment, the measurement application must be set to the base material of the calibration standard set, and after
the check it must be reset to the substrate material of the sample. Otherwise, the standard or the sample
88 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
would not be measured correctly. This is not necessary with a reference sample because its substrate mate-
rial is identical to that of the measurement application. A prominent example would be the check measure-
ment of the measurement application Sn/CuSn6 with Sn/Ni or Sn/Cu standards. Here, a measurement with-
out taking into account the fact that the calibration standard does not contain Sn in the substrate material
would result in a reading with a value that is up to 2-3 µm too low.
The check measurement is carried out by making an equal number of measurements (e.g., 5) under repeat-
ability conditions within the reference area of one and the same sample at regular intervals. Each of these
test series is called a group. The same measurement time and the same base setting of the measurement
application to be checked are used as during the regular measurements. Each group provides as its result
one mean value and a measure for the variation, e.g., the standard deviation. For each feature, these mea-
surement results are entered into a test protocol or even better into a quality control chart (SPC chart). Doc-
umentation of the measurement results is extremely important because only the documentation provides
the proof whether measurement readings are correct using the current calibration and whether the instru-
ment is okay. Measurement device monitoring is unthinkable without documentation and would not deserve
its name.
Time intervals
The time intervals of the check measurements are determined by the respective measurement application.
We recommend verifying simple measurement applications (e.g., Zn/Fe, Au/Ni, Sn/Cu etc.) once a week.
However, the user should verify more complex measurement applications (e.g., Cr/Ni/Cu/Zn) every other day
or even once a day. The measurement application is to be re-verified after every correction. The intervals
may be increased if the repeated verification never or only rarely shows a significant deviation from the nom-
inal value.
These recommended intervals indicate that annual checks with standards by a calibration service of the
manufacturer cannot replace measurement device monitoring. If the manufacturer performs a calibration
service, for example, because no calibration standards are available, the user only has the option to monitor
the measurement device using reference standards. Thus, establishing reference samples is an obligatory
consequence of using a calibration service, because continued access to standards by the user is not possi-
ble.
To determine the actual situation of the measurement application at the beginning of the control chart, the
characteristics of the reference sample or of the calibration standard, respectively, are determined with the
secured measurement application in the form of a larger random sampling (n > 50) without break. This is
called a pre-run. The reason for measuring the random sample in an as short as possible period is that the
instrument should remain stable during this time such that systematic changes of the instrument can be ne-
glected during this time span. The result of the pre-run is the mean value of all group mean values (nominal
value) and the mean value of the standard deviation from the single readings within the groups. The limits for
the mean value and for the standard deviation that are not to be violated by the random deviations are calcu-
lated from the nominal value in relation to the mean standard deviation and the number of single readings
per group. These limits are called statistical control limits. With subsequent measurements, upper and lower
violations of these control limits are therefore caused by systematic changes to the measurement applica-
tion.
Thus, the control limits characterize when subsequent measurement results are random or ”natural” (mean
value and/or standard deviation fall within the limits) or when they deviate systematically or ”unnaturally”
from the nominal values determined during the pre-run (measurement results are outside the limits) and in
doing so indicate a change in the measurement application that must be corrected. With subsequent regular
checks of the measurement application, using upper or lower violations of the control limits, it is then very
easy to discover on time deviations of the measurement application from the nominal values and to make
control corrections.
89
Setting Up the Print Form Using PDM
Pre-run
The pre-run consist of a minimum of m = 10 groups each with n = 5 single readings, i.e., a total of 50 mea-
surements. The measurement application and the identity designations must be documented.
The single readings for the pre-run should be carried out successively using the setting parameters (mea-
surement time, measurement distance, collimator, etc.) typically used for the measurement application and
should be concluded within a short period for the instrument to remain stable during this time. Systematic
changes during this period are then negligible. After each group (not after each measurement!), the refer-
ence sample should be taken from the instrument and re-positioned in order to simulate as good as possible
the subsequent procedure during the check measurement. With instruments featuring a variable setting of
the measurement distance, such as the Fischerscope XRAY, the measurement distance must be retained
during the pre-run and the subsequent check measurement because the standard deviation correlates to the
measurement distance. This distance can be noted and viewed in the ”notepad” of the product that is used for
measurement device monitoring.
1
(2) x̄total = x̄i = x̄est
m
This overall mean value forms the estimated value x̄est for the mean value of the quality control chart.
90 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
Standard deviations si
The standard deviations si of groups 1 to 10 are used to build the arithmetic mean value sI :
1
(3) sI = si
m
The index i runs across the number m of the groups.
The estimated value sI serves as the ”current” standard deviation of the individual readings. The variations
between the group mean values due to drift, interference or changes in the position of the reference samples
during the pre-run are taken into account by calculating the standard deviation of the mean values sII from
the group mean values x̄i of the m groups 1 to 10:
(4a) (x̄i − x̄total )2
sII =
m−1
sII takes into account the ”extraordinary” influences (potentially present during the pre-run) on the varia-
tions of the group mean values that may come into existence due to sample re-positioning for a new group or
by changing the measurement location or the operator – as will also occur during subsequent measure-
ments in the control chart.
The two limit values U CL(x) (upper control limit for the mean value x̄ ) and LCL(x) (lower control limit for
the mean value x̄ ) are the product of the nominal value x̄est and the standard deviation s with:
x̄est + 3s
(5a) U CL(x) = √
n
and
x̄est − 3s
(5b) LCL(x) = √
n
n is the group size. It has been taken into account that for every control limit an acceptance probability of
99.5% applies, i.e., an upper and lower violation of the control limits occurs randomly only at a total of 1% of
all cases.
An upper control limit U CL(s) applies also for the standard deviation that is calculated from the estimated
standard deviation s as follows:
The limiting factor BAO is a value that corresponds to the size of the random sample n, in this case for n = 5,
BAO = 2.09 .
BAO = 2.09 is arithmetically determined as follows:
1 − C(n)2
(7a) BAO =1+3
C(n)
91
Setting Up the Print Form Using PDM
with
Example of a pre-run
10 groups, each with 5 measurement readings, have been measured under repeatability conditions using a
nickel-plated and copper-plated reference sample (substrate material iron). For this purpose, the number of
single readings per group (in the ”Figure”) is set to a fixed number (here n=5) in the product ”Ni/Cu/Fe refer-
ence sample 10-245”.
We recommend entering the measurement data of the above pre-run in a quality control chart (SPC chart) for
each feature (here for the nickel and the copper thicknesses). The purpose of the pre-run is to calculate the
control limits. They are graphically drawn into the control chart as lines and correspond to the limit values
that lead to a correction if they are violated during subsequent measurements. The final protocol of the pre-
run generated with the WinFTM-PDM software at the actual reference sample Ni/Cu/Fe 10-245 shows be-
neath the group numbers (designated with "nbl"), the group results (mean values Ni. and Cu. per group, stan-
dard deviation s per group), the final result, the control limits as well as the control charts for the nickel coat-
ing and for the copper coating:
Helmut Fischer GmbH Office West
Danziger Straße 10 50374 Erftstadt
Tel.: 02235 41447 Fax: 02235 922990
92 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
93
Setting Up the Print Form Using PDM
All printed values are available in the print form of WinFTM PDM as variables (including the control charts).
They can re-calculate the printed control limits by using the original values (mean value x. and standard de-
viations s) and the equations specified in the previous section. Please note that the pre-run has been conclud-
ed within a short period (here: 17 minutes) in order to ensure the stability of the instrument during the pre-
run.
The protocol shows the video image of the reference sample that the camera of the Fischerscope XRAY has
recorded. The measurement area is identified by an orange-colored ring, the measurement spots where the
instrument made individual measurements during the pre-run are identified by yellow crosses. Even though
the reference area is small compared to the dimensions of the sample, the distribution of the coatings within
the reference area can be non-uniform. Because during the check measurement any measurement spot may
be selected within the reference area, a potential inhomogeneity of the coating thicknesses must already be
taken into account during the pre-run.
If the lower and upper limits are too far apart in reference to the confidence interval, then the reference sam-
ple is not suitable due to a non-uniform coating thickness within the reference area or because the sample
cannot be positioned and measured in a reproducible fashion. It is then recommended to use a different
specimen as a reference sample.
94 FISCHERSCOPE® X-RAY
Setting Up the Print Form Using PDM
95
Setting Up the Print Form Using PDM
96 FISCHERSCOPE® X-RAY
12 DefMA
If you have activated the SUPER software, you can create a DefMA by yourself.
The usage of the SUPER software requires deep knowledge of the physics of x-ray fluorescence and the func-
tionality of the instrument.
Alternatively, you can request a new measurement application from Fischer. Fischer offers a professional
service for the creation of customer’s Def.MA.
Mode Application
Material analysis For the analysis of bulk material (massive specimens).
Analysis with bal- For the analysis of bulk material (massive specimens), if the specimen cannot be de-
ance fined completely by 100 percent.
Typical applications: recycling, RoHS, soil specimens, biological specimens
Thickness mode For coating thickness measurements. The measurement results for the coating
thickness are displayed in the typical units of length such as µm. The analysis results
are shown in percent units.
The density values of the elements are taken from internal tables. You do have the
option to change these values.
Mass per unit area For coating thickness measurements. The overall area mass is displayed as well as
(relative) the percentage portion of the individual elements.
Mass per unit area For coating thickness measurements. The overall area mass is displayed as well as
(absolute) the individual area masses of the respective elements.
Solution analysis You fill the measuring cell with the solution to be analyzed, see Chap. 14.2 on
Page 111.
ROI-Mode ROI = Region of interest
Only a defined portion of the measures spectra is evaluated. You define one or more
regions (ROI) of energy channels. The portion of the spectra that is located in the de-
fined region is evaluated only.
Counting rates are issued, the fundamental parameter method is not used.
Used for special applications, where the fundamental parameter method is not appli-
cable.
Component mode For the analysis of specimens, which are not only composed of pure elements, but
also contain chemical compounds (e. g. oxides). WinFTM offers a list of components,
which can be used like elements.
The result display shows the measurement readings of the complete components,
not only the elements the components are composed of.
97
DefMA
Mode Application
Component mode For the analysis of bulk material (massive specimens), if the specimen cannot be de-
with balance fined completely by 100 percent and the specimen is not only composed of pure ele-
ments, but also contain chemical compounds (e. g. oxides).
The result display shows the measurement readings of the complete components,
not only the elements the components are composed of.
Typical applications: recycling, RoHS, soil specimens, biological specimens
Procedure
1. In the WinFTM main menu select Calibrate > DefMA new… > <Mode, see
see Table 12-1 on page 97>.
The window Definition of measuring conditions… appears.
2. In the field Label enter an explicit name for the new DefMA file.
3. In the field located right above select the high voltage and the primary filter (if present). The high voltage
and the primary filter has to fit your measurement task.
98 FISCHERSCOPE® X-RAY
DefMA
Enter the base material (substrate material) like another layer with the following options:
a In the area Thickness check the option The thickness of the layer is fixed (known).
b Check the option Massive (Sat.)
10. Click OK to close the window Definition of measuring conditions….
Procedure
1. In the WinFTM main menu select Calibrate > DefMA modify….
The window DefMA modify appears.
2. Select the DefMA and click OK.
The window Definition of measuring conditions… appears.
3. In the window Definition of measuring conditions… click Compens.Spectra.
The window Spectrum Background appears.
4. You can select one or more of the following options:
Compensation Elements
In the area Compensation Elements you can enter up to four elements. The entered compensation el-
ements are filtered out of the spectrum. If the relevant element area not shown in the list field, check
full Element list.
Example tin on glass: The glass contains Strontium (Sr) in variable concentration. If you enter Sr as
compensation element, the spectral lines caused by Sr are filtered out. The variable amount of Sr
does not influence the measurement reading.
Compensation Spectrum
If a suitable compensation spectrum is available, you can activate the option yes in the in the area
Compensation Spectrum and enter a name for the spectrum, e. g. glass or plastics. At a later moment
during the calibration you will be prompted to position the compensation spectrum. The name en-
tered here will then be displayed.
Example tin on glass: The glass causes a scattering spectrum (also referred to as background spec-
trum or base spectrum). If a sample is available that is comparable with the specimen and has no tin
coating, you can enter here glass. At a later moment during the calibration you will be prompted to
position the glass sample. WinFTM measures then the compensation spectrum and takes it into ac-
count when evaluating the spectrum.
Base Spectrum
If a suitable base spectrum is available, you can activate the option yes in the in the area Compensa-
tion Spectrum and enter a name for the spectrum, e. g. glass or plastics. At a later moment during the
99
DefMA
calibration you will be prompted to position the base spectrum. The name entered here will then be
displayed.
Background correction
If the theoretical characteristics (measured or calculated) of the compensation spectrum differs
from the actual characteristics, you can reach a correction by using this option. You can choose from
three options: no correction, with supporting points and by polynomial.
5. To close the window Spectrum Background click OK.
6. To close the window Definition of measuring conditions… click OK.
Depending on the used DefMA mode (see Table 12-1 on page 97), this field has the name Mode, Analysis
Mode, Thickness Mode, Mass per unit area (relative/absolute), Components‘s Mode or Solution Analysis. The
meaning of the individual symbols is:
d The coating thickness of the top layer (consisting of one element) is being measured and
Ccc the concentrations of the substrate material, which contains three elements, are being analyzed.
13.1 Normalization
Purpose of the Normalization
A normalization is required in the following situations:
The technical properties of the instrument's measuring head were modified, e.g., aperture, anode cur-
rent, primary filter, high voltage, ROI, changes in the base material (when base material unknown is
defined), measuring distance.
During a reference measurement, a deviation was encountered that is greater than two to three times the
standard deviation s. A reference measurement is a measurement on a reference sample with known
properties to check the normalization/calibration of a product.
A product was imported.
The normalization adapts a measuring application to the current technical properties of the instrument's
measuring head.
During the normalization, the spectra that are defined in the Definition of the Measuring Application (DefMA)
as pure elements and as base material are re-established.
The following data are stored in the software during the normalization:
The zero point for the coating thickness measurement
The influence of the base material on the measurement of the coatings
A normalization always refers to a specific measuring application. This means that every measuring applica-
tion must be normalized separately.
Procedure
1. Select the product you want to normalize, e.g. the product you just imported. In the WinFTM main menu,
select Product > Select....
The window Select Product appears. Select the product file and click OK.
2. To start the normalization, select Calibrate > Normalize....
The following procedure depends on the measuring application:
Simple measuring application Zn/Fe
For a simple measuring application, such as
Zn/Fe, everything is calculated automati-
cally, you do not need to do anything else.
Then the window End of normalization. Ac-
cept? will appear.
Click Yes.
101
Calibration, Normalization and Base Correction
Pure elements
If you are prompted to measure pure ele-
ments, then use the supplied pure element
plate, see picture on the right. Pure element
calibration standards from Fischer ("Dom-
ino") are also available for certain elements.
If a pure element is not available, then load
it from the calculated spectra library. To do
so, select Measure > Element from library.
3. When the normalization is finished, then the window End of normalization. Accept? will appear.
4. Click Yes.
The normalization is finished.
Au/Ni/(CuSn6)
The Au coating and the Ni coating on bronze shall
be measured.
Measurement mode: d d -
Dual coating
The base is CuSn6
103
Calibration, Normalization and Base Correction
Procedure
1. To start the base correction, select Calibrate > Base correction.
The window Element: Measure product‘s base material appears.
2. Position the base material, then click OK.
13.3 Calibration
Purpose of the Calibration
A calibration is required in the following situations:
A new measuring application has been set up.
A product was imported.
If during a reference measurement (a measurement on a reference sample with known properties) a
deviation was encountered that is greater than two to three times the standard deviation s.
The calibration sets the difference between nominal value and actual value to zero.
Actual value refers to the reading displayed by the instrument.
The nominal value is the value engraved on the calibration standard (engraved value).
Calibrating enables secured measurements: If the nominal values of the used standards can be traced to rec-
ognized material measures or reference materials, then the measurement is secured through the calibration.
Depending on the measuring application, it may be difficult to evaluate the result of a calibration in individual
cases. If you have any questions, please contact Fischer via the website www.helmut-fischer.com.
Fischer can carry out a calibration service for you. In this case, you will receive a corresponding sheet with
calculations and implications for evaluating the calibration.
Calibration standard set Consists of a calibration standard set with several calibration standards
(Cal-NS) of a specific coating/base material combination.
Individual calibration standard Individual calibration standard of a certain coating/base material com-
(Cal-N) bination
Calibration foil Calibration foil consisting of a free carrying layer of a certain element.
(Cal-F) Universally usable on many different base materials.
Customer-specific standards
105
Calibration, Normalization and Base Correction
Procedure
1. In the WinFTM main window select Calibrate > Calibration Standard Set….
7. To enter additional calibration standards, repeat steps 3 to 6 for every calibration standard. Before you
enter an additional calibration standard, make sure that no line is selected in the window Input Calibr. Std.
Set. Otherwise the values of the selected standard will be changed.
8. Once you have entered all values of a calibration standard, click OK in the window Input Calibr. Std. Set.
The calibration standard set is now entered and stored.
9. To check how many measurements will be performed for each cali-
bration standard, click Info.
107
Calibration, Normalization and Base Correction
Depending on the measuring application, different prompts will appear that have the following meaning:
Scatt A scatter spectrum shall be measured. As a rule, it is already saved in
the software and can be opened automatically from the menu
General > Load Spectrum and Evaluate…. Measure a scatter spectrum
(e.g., plastic) if the spectrum does not exist.
Element: Ni The pure element Ni from the pure element board shall be measured.
Base material Cal. Std. Set The base material of the calibration standard set shall be measured.
(Analysis) Example: If you calibrate using the calibration standard set
Ni/Cu alloy, Cu is the base material of the calibration standard set.
Base material specimen The base material of the real part shall be measured. An uncoated re-
(Analysis) al part must be used for this purpose.
Base (specimen) The base material of the real part shall be measured. An uncoated re-
al part must be used for this purpose.
Ni-Zn/Fe 5.0 15.0 0.0 A calibration standard made of NiZn/Fe with a coating thickness of
5 µm and a Ni content of 15 % shall be measured.
13. Position the standard, close the measuring chamber, adjust the focus and click Start.
The instrument will perform the measurement.
After the measurement, you will be prompted either to measure a new standard or to measure the same
standard one more time. If you measure the same standard again, move it slightly to measure it at a dif-
ferent location.
1 2 3 1 2 3
109
Calibration, Normalization and Base Correction
The calibration was successful if all re-measurements are within the respective ranges and you are able to
use the measuring application.
If a standard is outside the range, check to see of this measurement range is relevant for your measuring
application. If the measurement range is not relevant for your measuring application, you can still use the
measuring application.
If several re-measurements are outside the range, check the data and potential error sources. Then perform
a new calibration.
For special measuring applications, other conditions may apply for a successful calibration. However, these
will be stated separately.
For the solution analysis you need a measuring cell. You fill the measuring cell with the solution to be ana-
lyzed, position the measuring cell and perform the measurement.
The measuring cells are available with different reference plates. When selecting the measuring cell, note
that the characteristic x-ray lines of the sample to be measured may not be too close to the x-ray lines of the
reference plate material. Otherwise, problems may arise during the evaluation of the spectrum.
111
Solution Analysis
Procedure
2. Fill the cup with the sample. Fill with the maxi-
mum possible amount of liquid, i.e., the surface
of the liquid should be slightly convex.
Procedure
1. Place the measuring cell into the measuring chamber. Depending on the instrument type, the cup open-
ing must face upwards or downwards:
Instrument with measuring direction from top to bottom: Cup opening must face upwards.
Instrument with measuring direction from bottom to top: Cup opening must face downwards.
To make the focusing easier, position the measuring cell such that the measurement beam does not
point to the center, but to the edge curvature of the measuring cell, see the following figure:
Procedure
113
Solution Analysis
Procedure
1. If you have received the product file by email, save the file in a directory that does not belong to WinFTM,
e.g. C:\temp.
2. If you have received the product file by CD, insert the CD into the CD drive.
3. Optional, if you have installed the WinFTM extension PDM: You can create a new directory for the new
product as follows:
a In the WinFTM main menu, select Product > Select ...
The window Select Product appears.
b Click in the upper right .
4. In the WinFTM main window select Product > Copy to/from File… .
The window Copy from/to file appears.
5. Select the tab File >>> Product.
6. Click Select.
Procedure
1. If you have received the Def.MA file by email as a file attachment, save the file in a directory that does not
belong to WinFTM, e.g. C:\temp.
2. If you have received the Def.MA file by CD, insert the CD into the CD drive.
3. In the WinFTM main window select Product > Copy to/from File….
The window Copy from/to file appears.
4. Select the tab File >>> Def.MA.
5. Click Select.
The window Select file appears.
6. Select the Def.MA file and click Open.
The selected path appears in the field File.
7. Click Start copy.
The file will be copied. After this the report The Def.MA was copied successfully appears.
8. Click OK.
9. To close the window, click Cancel.
115
Solution Analysis
Procedure
1. In the WinFTM main menu select Product > New….
The window Select an application appears.
2. Click New.
The window Select a DefMA appears.
3. Select the DefMA and click OK.
The window Input of the application name appears.
4. Enter in the field Name an identifier for the new measurement application and click OK.
If your instrument is equipped with changeable apertures (collimators), the window Select collim. + meas.
dist. appears. Select the standard collimator No. 2 and click OK.
The window Input Calibr. Std. Set appears.
5. Do not enter anything in this window and click OK.
6. If your are prompted to measure scatt, do not perform a measurement, but perform the following steps
instead:
a In the WinFTM main menu select Tools > Load Spectrum and Evaluate….
b If the window Open appears, select the file scatt and click Open.
The window End of calibration. Accept? appears.
7. Click Yes.
The window Create new product appears.
8. Enter a name, directory and number for the new product and click OK.
9. To check the measurement time and the number of measurements click Info.
The window Modify product appears.
10. Click the tab Product.
11. Check the values for the measurement time and the number of measurements. Change the values, if nec-
essary.
The longer the measurement time, the better the repeatability precision. Do not select a measuring time
that is too short; 30 or 60 seconds is a typical value.
12. Click OK.
The new product file is set up and the related measurement application is normalized.
117
Solution Analysis
You can export the measured values, their XYZ coordinates, statistical evaluations and other data to other
applications like spreadsheets, statistical evaluation tools or process control systems.
All settings for the measurement data export are assigned to one product. For each product, you can define
different settings for the data export.
Basically there are two modes for exporting measurement data: online export and export on demand.
Procedure
1. To define the export settings, select in the WinFTM main menu Evaluation > Export > Export Setup….
The window Exporting Data appears.
119
Measurement Data Export
Control Explanation
Online-Export ON The measurement data is exported continuously when measurement data is ac-
quired by the instrument.
The reading is exported directly after a measurement is made.
The block statistics are exported upon block closure.
With product selection, the header is exported.
Template and No. The name and the number of the export template are displayed if an export tem-
plate is assigned to the current Product.
► To assign an export template, click Select….
Export destination
append to file Data is successively written to a file.
Complete file-hand- WinFTM waits until the handshake file NET_EXPT.END is deleted. This control box
shake (NET_EXPT.END) should be enabled only, if the technology of the target computer is capable of han-
dling it.
overwrite file The export file contains only the last exported values, i. e. the old data will be over-
written.
Destination file Displays the destination file, if already assigned.
► To assign a destination file for the data export, click Select….
Filename variable By clicking the list box you can select a rule, how the file name of the exported data
will be created.
send to serial port Data is sent directly to the RS232 interface.
The parameters of the RS232 interface and of the WinFTM software of the receiv-
ing instrument must be the same. Otherwise the receiving instrument will not be
able to read the data exported via the RS232 interface.
► To set the parameters of the RS232 interface select
Evaluation > Export > Define export interface….
to OLE-Automation WinFTM can be remote controlled via OLE and can transfer data via OLE. A Visual-
Basic (Version 6) program example is included on the WinFTM installation CD in the
directory Service\OLE_Import_Sample_VisualBasic_6.0\AutomationDemo. This sam-
ple program can be used as a template for the customers own remote control soft-
ware.
121
Measurement Data Export
Control Explanation
to Excel The data will be directly sent to an Excel file.
A new Excel workbook (Workbook 1) opens with the first exported value. Even if Ex-
cel is already opened, a new workbook will open with the first exported value. In
the field Startposition Column Row enter the cell, where the table shall start, e.g. C4.
@TAB must be inserted in the export template as column separators. In addition, a
period (.) and not a comma (,) must be selected as the decimal character. Depend-
ing on the country setting, Excel might still display the comma (,) as the decimal
separator character. At the end, the Excel workbook (Workbook 1) must be saved.
► Do not close Excel as long as WinFTM is still open.
to OPC-DA1..3 For the connection of the PC to a programmable logic control via Profibus, Profinet
to TCP-IP For the connection of the PC via TCP/IP
Procedure
Do one of the following:
To create a new export template select Evaluation > Export > new template.
The window New export template appears.
In the field Name enter a name for the new export template and click OK.
To modify an existing export template select Evaluation > Export > modify template.
The window Select export template appears.
Select the template you want to change and click OK.
The three sub-windows on the left define the different sections of data to be exported: single readings, block
statistics and header info. With the check boxes in the area Export Mode you can activate or deactivate the
sections.
In the figure above only Single reading is checked, that is, only the section for single readings is exported, the
code in the other sub-windows will be ignored.
Table 15-3: The meaning of the sub-windows for Online Export and Export on demand
The sub-window on the right shows all variables with a short explanation.
123
Measurement Data Export
In the window Definition of Export Template the template must be written in one line to
avoid unwanted carriage returns and line feeds.
For more clarity the syntax is shown here in three lines. In the window Definition of Ex-
port Template the template must be written in one line to avoid unwanted carriage re-
turns and line feeds.
Variable Explanation
Block # The text Block # will be displayed.
@TAB A tabulator will be inserted.
@BLK Contains the number of the block
@ANB Contains the value of the field Order No.
@LOT Contains the value of the field Lot No.
@BM1 Contains the value of the field Label 1.
@BM2 Contains the value of the field Label 2.
@CRX A carriage return will be inserted.
@LFX A line feed will be inserted.
x= The text x = will be displayed.
@MW Contains the mean value, must always be supplemented with the number of the respective
reading, e. g. @MW1
s= The text s = will be displayed.
@S_ Contains the standard deviation, must always be supplemented with the number of the re-
spective reading, e. g. @S_1
@END Ends the template
125
Measurement Data Export
For more clarity the syntax is shown here in three lines. In the window Definition of Ex-
port Template the template must be written in one line to avoid unwanted carriage re-
turns and line feeds.
Variable Explanation
@CAL Contains the product name.
@TAB A tabulator will be inserted.
# The character # will be displayed.
@PRN Contains the product number.
@CRX A carriage return will be inserted.
Operator The text Operator will be inserted.
@LFX A line feed will be inserted.
@EL Contains the name of the result channel (element, coating thickness, ppm etc.), must al-
ways be supplemented with the number of the respective channel, e. g. @EL1
@DM Contains the unit of the result channel (mils, nm, % etc.), must always be supplemented
with the number of the respective channel, e. g. @DM1
@END Ends the template
With a service contract you get highest security and reliability. Therefore please contact your Fischer repre-
sentative.
► Use only gentle cleaning agents. Never use alcohol or similarly aggressive
substances.
► Use a soft brush to clean the measuring chamber. Never use a vacuum cleaner to
clean the measuring chamber.
► Clean the instrument with a damp cloth gently. For stubborn dirt use a small
amount of dishwashing detergent.
bration standards can be changed, if the standards are cleaned. This can lead to mea-
surement errors.
► Never clean calibration standards.
127
Cleaning and Maintenance
Rotary knob
M7
10. Slide the filter foam pad to the right as shown in the direction of the arrow in Fig. 16-2.
11. Depending on the degree of dirt, blow out the pad with compressed air or clean it with a mild liquid deter-
gent.
If the filter foam pad can no longer be used, replace it.
Dry the cleaned filter foam pad at ambient temperature. Put the filter foam pad back into place only as
soon it is completely dry.
12. Mount the side cover in reverse order.
warranty claims.
► Before transporting the instrument, re-install the shipping lock.
► Transport the instrument in the original shipping container. If you no longer have
access to the original shipping container, rent a container from Fischer.
Procedure
1. Switch off the instrument, disconnect and remove all electrical connections.
2. Open the cover of the measurement chamber and re-install the shipping lock made of foam.
3. Remove the key for turning on the x-ray energy and place it in the case with the accessories.
4. Store the instrument in its original shipping container. If you no longer have access to the original ship-
ping container, rent a container from Fischer.
16.5 Troubleshooting
129
Cleaning and Maintenance
xi : Single readings
n: Number of measurements per block
Confidence inter- Each measurement device produces ran- q is calculated according to the equation
val q dom measurement errors. With a certain below assuming a normal distribution for a
probability (confidence level), the true val- confidence level of 95 %:
ue of the measured parameter lies in an
s
interval (double-sided confidence interval) q = t f ;1 – 0025 -------
around the measured mean value x of a n
measurement series (block/group). The
borders of this interval are at a distance q t: Quantile of the Student’s t-distribu-
from the true value . tion (Student’s factor), can be ob-
tained from tables in common publi-
x–qx+q
cations [1]
s: Standard deviation
n: Number of the measurements per
block
Standard The standard deviation is a measure of the
n
deviation s dispersion of the single readings of a mea- 1 - 2
surement series about the mean value. s = ----------- x – xi
n–1
i=1
131
Description of the Characteristic Statistical Parameters
s
V = -- 100
x
s: Standard deviation
x: Mean value of the block
For sufficiently large samples it can be
said:
V =
---
: Standard deviation
: Mean value
Range R Difference between the maximum and the
minimum values of the block R = x max – x min
R
R p = --- 100
x
R: Range in percent
x: Mean value of the block
n (< LSL) Number of measurements that are below Specification limits (USL, LSL) can be en-
the lower specification limit tered in the menu Evaluation > Final evalua-
tion…, tab SPC, button Parameter.
n (> USL) Number of measurements that are above
the upper specification limit
[1] Graf, Henning, Stange und Wilrich: Formeln der angewandten mathematischen Statistik (Equations in
Applied Mathematical Statistics); Springer Verlag (Springer Publishing Company)
n Bl
1
----------------
2
s = x – xi
n Bl – 1
i=1
n Bl : Numbers of blocks
x: Mean value of the block mean val-
ues
xi : Block mean values
Mean standard The mean standard deviation is a measure
n Bl
deviation s of the dispersion of the individual block 1
standard deviations of a measurement se- s = ------- si
n Bl
ries about their common mean value x . i =1
133
Description of the Characteristic Statistical Parameters
s
V = -- 100
xx
n Bl
1
R = ------- R i
n Bl i =1
n Bl
1
̂ = ------- s i2
n Bl
i =1
n Bl : Number of blocks
si: Standard deviations of the blocks
Cpk Process capability index C pk = Min C pu ,C po
Computed and displayed only if
specification limits are enabled
x – LSL
C pu = ------------------
-
3̂
USL – x-
C po = -------------------
3̂
n Bl
1
̂ = ------- s i2
n Bl
i =1
n Bl : Number of blocks
si: Standard deviations of the blocks
135
Description of the Characteristic Statistical Parameters
[1] Graf, Henning, Stange und Wilrich: Formeln der angewandten mathematischen Statistik (Equations in
Applied Mathematical Statistics); Springer Verlag (Springer Publishing Company)
n
1
x = n--- xi
i =1
137
Description of the Characteristic Statistical Parameters
s
V = -- 100
x
s: Standard deviation
x: Mean value of the single readings of
the selected blocks
Range R Difference between the highest and the
lowest reading of the selected blocks R = x max – x min
R
R p = --- 100
x
R: Range
x: Mean value of the single readings of
the selected blocks
LSL Lower Specification Limit Specification limits (USL, LSL) can be en-
tered in the menu
USL Upper Specification Limit Evaluation > Final evaluation…, tab SPC,
button Parameter.
1 m
s I = ---- s i
m i =1
m: Number of blocks
si: Standard deviation of the block i
sII Standard deviation of the mean values
1 - m 2
s II = ------------ xi – x
m–1
i=1
m: Number of blocks
xi : Mean value of the block i
x: Mean value of the block mean val-
ues
n (> USL) Number of measurements that are above
the upper specification limit USL
n (> USL) [%] Number of measurements that are above
the upper specification limit USL in per-
cent Displayed only if specification limits are
enabled
n (< UGW) Number of measurements that are below
the lower specification limit LSL
n (> UGW) [%] Number of measurements that are below
the lower specification limit LSL in percent
[1] Graf, Henning, Stange und Wilrich: Formeln der angewandten mathematischen Statistik (Equations in
Applied Mathematical Statistics); Springer Verlag (Springer Publishing Company)
139
140
17.4
1 1.01 Periodic table of the Elements X-Ray Properties 2 4.00
Hydrogen Helium
H He
3 6.94 4 9,01 Atomic Number 79 197,0 Atomic weight 5 10.81 6 12.01 7 14.01 8 15.99 9 18.99 10 20.18
Lithium Berylium Gold Name ( IUPAC ) GAS LIQUID SOLID Boron Carbon Nitrogen Oxygen Fluorine Neon
Li Be Au B C N O F Ne
0,05 0,11 K-alpha 68,79 9,71 L-alpha 0,19 0,28 0,39 0,52 0,68 0,85
K-beta 77,97 11,44 L-beta
M-alpha 2,12 13,38 L-gamma
11 22.99 12 24.31 13 26.98 14 28,09 15 30,97 16 32,07 17 35,45 18 39,95
Sodium Magnesium Aluminium Silicon Phosphorus Sulfur Chlorine Argon
Na Mg Al Si P S Cl Ar
1,04 1,25 1,49 1,74 2,02 2,31 2,62 2,96
1,55 1,84 2,14 2,46 2,82 3,19
19 39,10 20 40,08 21 44,96 22 47,88 23 50,94 24 52,00 25 54,94 26 55,85 27 58,93 28 58,69 29 63,55 30 65,39 31 69,72 32 72,61 33 74,92 34 78,96 35 79,90 36 79,90
Potassium Calcium Scandium Titanium Vanadium Chromium Manganese Iron Cobalt Nickel Copper Zinc Gallium Germanium Arsenic Selenium Bromine Krypton
K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
3,31 3,69 4,09 4,51 4,95 5,41 5,90 6,40 6,93 7,48 8,05 8,64 1,01 9,25 1,10 9,89 1,19 10,54 1,28 11,22 1,38 11,92 1,48 12,65 1,59
3,59 4,01 4,46 4,93 5,43 5,95 6,49 7,06 7,65 8,26 8,90 9,57 1,03 10,26 1,12 10,98 1,21 11,73 1,32 12,50 1,52 13,29 1,53 14,11 1,64
37 85,47 38 87,62 39 88,91 40 91,22 41 92,91 42 95,94 43 97,91 44 101,1 45 102,9 46 106,4 47 107,9 48 112,4 49 114,8 50 118,7 51 121,8 52 127,6 53 126,9 54 131,3
Rubidium Strontium Yttrium Zirconium Niobium Molybdenum Technetium Ruthenium Rhodium Palladium Silver Cadmium Indium Tin Antimony Tellurium Iodine Xenon
Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe
13,39 1,69 14,16 1,81 14,96 1,92 15,77 2,04 16,61 2,17 17,48 2,29 18,41 2,42 19,28 2,56 20,21 2,70 21,18 2,84 22,16 2,98 23,17 3,13 24,21 3,29 25,27 3,44 26,36 3,61 27,47 3,77 28,61 3,94 29,80 4,11
Description of the Characteristic Statistical Parameters
14,96 1,75 15,83 1,87 16,74 2,00 17,67 2,12 18,62 2,26 19,61 2,40 20,59 2,54 21,66 2,68 22,72 2,83 23,82 2,99 24,94 3,15 26,09 3,32 27,27 3,49 28,48 3,66 29,72 3,84 30,99 4,03 32,29 4,22 33,64 4,42
2,30 2,46 2,62 2,79 2,96 3,14 3,33 3,52 3,72 3,92 4,13 4,35 4,57 4,80 5,04
55 132,9 56 137,3 72 178,5 73 181,0 74 183,9 75 186,2 76 190,2 77 192,2 78 195,1 79 197,0 80 200,6 81 204,4 82 207,2 83 209,0 84 209,0 85 210,0 86 222,0
Periodic Table of the Elements with X-Ray Properties
Caesium Barium Hafnium Tantalum Tungsten Rhenium Osmium Iridium Platinum Gold Mercury Thallium Lead Bismuth Polonium Astatine Radon
Cs Ba 57-71 Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
30,97 4,29 32,19 4,47 Lanthanides 55,76 7,90 57,52 8,15 59,31 8,40 61,13 8,65 62,99 8,91 64,89 9,19 66,82 9,44 68,79 9,71 70,82 9,99 72,86 10,27 74,96 10,55 77,10 10,84 79,30 11,13 81,53 11,42 83,80 11,72
34,98 4,62 36,38 4,83 63,21 9,02 65,21 9,34 67,23 9,67 69,30 10,01 71,40 10,35 73,54 10,71 75,74 11,07 77,97 11,44 80,26 11,82 82,56 12,21 84,92 12,61 87,34 13,02 89,81 13,44 92,32 13,87 94,88 14,32
5,28 5,53 1,65 10,51 1,71 10,89 1,78 11,28 1,84 11,68 1,91 12,09 1,98 12,51 2,05 12,94 2,12 13,38 2,20 13,83 2,27 14,29 2,35 14,76 2,42 15,24 15,74 16,25 16,77
87 223,0 88 226,0 89 227,0 90 232,0 91 231,0 92 238,0 93 237,1 94 244,1 95 243,1 96 247,1 97 247,1 98 251,1 99 252,1 100 257,1 101 258,1 102 259,1 103 260,1 104 261,1
Francium Radium Acitinium Thorium Protactinium Uranium Neptunium Plutonium Americium Curium Berkelium Californium Einsteinium Fermium Mendelevium Nobelium Lawrencium Rutherford.
Fr Ra Ac Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr Rf
86,12 12,03 88,49 12,34 90,89 12,65 93,33 12,97 95,85 13,29 98,43 13,61 101,0 13,95 103,7 14,28 106,4 14,62 109,1 14,96 111,9 15,31 114,7 15,66 117,7 16,02 120,6 16,38
97,48 14,77 100,1 15,23 102,9 15,71 105,6 16,20 108,4 16,70 111,3 17,22 114,9 17,74 117,2 18,28 120,2 18,83 123,2 19,39 126,4 19,97 129,5 20,56 132,8 21,17 136,1 21,79
17,30 17,85 18,41 3,0 18,98 3,08 19,56 3,17 20,16 3,26 20,78 3,35 21,42 22,07
57 138,9 58 140,1 59 140,9 60 144,2 61 144,9 62 150,4 63 152,0 64 157,3 65 158,9 66 162,5 67 164,9 68 167,3 69 168,9 70 173,0 71 175,0
Lanthanum Cerium Praseodym. Neodymium Promethium Samarium Europium Gadolinium Terbium Dysprosium Holmium Erbium Thulium Ytterbium Lutetium
57-71 La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Lanthanides 33,44 4,65 34,72 4,84 36,02 5,03 37,36 5,23 38,65 5,43 40,12 5,64 41,53 5,85 42,98 6,06 44,47 6,28 45,99 6,50 47,53 6,72 49,10 6,95 50,73 7,18 52,36 7,41 54,06 7,65
37,80 5,04 39,26 5,26 40,75 5,49 42,27 5,72 43,95 5,96 45,40 6,21 47,03 6,46 48,72 6,71 50,39 6,98 52,18 7,25 53,93 7,53 55,69 7,81 57,58 8,10 59,35 8,40 61,28 8,71
0,83 5,79 0,88 6,05 0,93 6,32 0,98 6,60 6,89 1,08 7,18 1,13 7,48 1,19 7,79 1,24 8,10 1,29 8,42 1,35 8,75 1,41 9,09 1,46 9,42 1,52 9,78 1,58 10,14
FISCHERSCOPE® X-RAY
Data Sheet
142
General Specification
Intended use Energy dispersive X-ray measuring instrument (EDXRF) to analyze precious
metals and their alloys in composition and coating thickness.
Repeatability ≤ 1 ‰ for gold, measurement time 60 sec
Design Bench top unit with upwards opening hood
Measurement direction Bottom up
X-Ray Source
X-ray tube Tungsten tube, thermally stabilized
High voltage Three steps: 30 kV, 40 kV, 50 kV, max. anode current: 1 mA
Aperture (Collimator) Ø 1 mm (39 mils), optional Ø 2 mm (79 mils)
Measurement spot size Ø 1.2 mm (47 mils) with aperture Ø 1 mm (39 mils) and flat lying sample
(measurement distance 0 mm)
X-Ray Detection
X-ray detector Silicon PIN detector with peltier cooling
Element range Sulfur S (16) to Uranium U (92) – up to 24 elements simultaneously
Resolution (fwhm for Mn-Kα) ≤ 180 eV
Measuring distance 0 … 25 mm (0 … 1 in)
Distance compensation with patented DCM method for simplified measure-
ments at varying distances. For particular applications or for higher demands
on accuracy an additional calibration might be necessary.
Sample Alignment
Sample positioning Manually
Video microscope High-resolution CCD color camera for optical monitoring of the measurement
location along the primary beam axis, Crosshairs with a calibrated scale (ruler)
and spot-indicator, Adjustable LED illumination, Laser pointer (class 1) to sup-
port accurate specimen placement
Zoom factor Digital 1x, 2x, 3x, 4x
Sample Stage
Design Fixed sample support
Usable sample placement
area 310 x 320 mm (12.2 x 12.6 in)
Max. sample weight 13 kg (29 lb)
Max. sample height 90 mm (3.5 in)
143
FISCHERSCOPE® X-RAY XAN® 215
Electrical Data
Power supply AC 100 – 240 V ±10 % / 50 – 60 Hz max. 120 VA, without evaluation PC
Protection class IP40
Dimensions
External dimensions Width x depth x height [mm]: 403 x 588 x 365 mm, [in]: 15.9 x 23.1 x 14.4
Weight Approx. 45 kg (99 lb)
Environmental Conditions
Operating temperature 10 °C – 40 °C (50 °F – 104 °F)
Storage/Transport
temperature 0 °C – 50 °C (32 °F – 122 °F)
Admissible air humidity ≤ 95 %, non-condensing
Evaluation Unit
Computer Windows®-PC
Software Standard: Fischer WinFTM® BASIC including PDM®
Optional: Fischer WinFTM® SUPER
Standards
CE approval EN 61010, EN 61326
X-Ray standards DIN ISO 3497 and ASTM B 568
Approval Fully protected instrument with type approval according to German radiation
protection law.
Order
FISCHERSCOPE X-RAY XAN 215 605-083
605-692, contains all calibrated measurement applications necessary for the
analysis of jewelry, coins and precious metals
Special XDV product modification and XDV technical consultation on request
2022-09-30
952-097
FISCHERSCOPE®, XAN®, WinFTM®, PDM® are registered trademarks of Helmut Fischer GmbH Institut für Elektronik und Messtechnik,
Sindelfingen - Germany. Windows® is a registered trademark of Microsoft Corporation in the United States and other countries.
www.helmut-fischer.com
Data Sheet
XAN® 222
XAN® 220
FISCHERSCOPE® X-RAY XAN® 220/222
Main Features
The FISCHERSCOPE X-RAY XAN 220 and XAN 222 are optimized X-ray fluores-
cence measuring instruments for non-destructive analysis of jewelry, coins and
precious metals.
They are particularly suited for the analysis of precious metals and their alloys
in composition and coating thickness. Up to 24 elements can be determined si-
multaneously.
Typical fields of application are the analysis of:
Jewelry, precious metals and dental alloys
Yellow and white gold
Platinum and silver
Rhodium
Alloys and coatings
Multi layer coatings
Outstanding accuracy and long-term stability are characteristics of all
FISCHERSCOPE X-RAY systems. The necessity of recalibration is dramatically
reduced, saving time and effort.
The modern silicon PIN detector achieves high accuracy and good detection
sensitivity.
The fundamental parameter method by Fischer allows for the analysis of solid
and liquid specimens as well as coating systems without calibration.
Design
The XAN 220 and XAN 222 are designed as user-friendly bench-top instruments.
They differ in the support stage and the housing size:
XAN 220: Fixed sample support
XAN 222: Manually operable XY stage for accurate positioning of small parts
and larger measuring chamber
For quick and easy sample positioning, the X-ray source and semiconductor de-
tector assembly is located in the instrument‘s lower chamber. The measuring
direction is from underneath the sample, which is supported by a transparent
window.
The integrated video-microscope with zoom and crosshairs simplifies sample
placement and allows precise measuring spot adjustment.
The entire operation and evaluation of measurements as well as the clear pre-
sentation of measurement data is performed on a PC, using the powerful and
user-friendly WinFTM® software.
The FISCHERSCOPE X-RAY XAN 220 and XAN 222 fulfill DIN ISO 3497 and
ASTM B 568. The XAN 220 is a fully protected instrument with type approval
according to German radiation protection law.
146
General Specification
Intended use Energy dispersive X-ray measuring instrument (EDXRF) to analyze precious
metals and their alloys in composition and coating thickness.
Design Bench top unit with upwards opening hood
Measurement direction Bottom up
X-Ray Source
X-ray tube Micro-focus tungsten tube with beryllium window
High voltage Three steps: 30 kV, 40 kV, 50 kV; max. anode current: 1 mA
Aperture (Collimator) Ø 1 mm (39 mils), optional Ø 2 mm (79 mils) or Ø 0.6 mm (23.6 mils)
Measurement spot size Ø 1.2 mm (47 mils) with aperture Ø 1 mm (39 mils) and flat lying sample (mea-
surement distance 0 mm)
X-Ray Detection
X-ray detector Silicon Drift Detector (SDD), peltier-cooled
Element range Sulfur S (16) to Uranium U (92) – up to 24 elements simultaneously
Resolution (fwhm for Mn-Kα) ≤ 135 eV
Measuring distance 0 … 25 mm (0 … 1 in), Distance compensation with patented DCM method for
simplified measurements at varying distances. For particular applications or for
higher demands on accuracy an additional calibration might be necessary.
Repeatability ≤ 0.5 ‰ for gold, measurement time 60 sec
Sample Alignment
Sample positioning Manually
Video microscope High-resolution CCD color camera for optical monitoring of the measurement
location along the primary beam axis, crosshairs with a calibrated scale (ruler)
and spot-indicator, adjustable LED illumination
Zoom factor Digital 1x, 2x, 3x, 4x
Electrical Data
Power supply AC 100 – 240 V ±10 % / 50 – 60 Hz max. 120 VA, without evaluation PC
Protection class IP40
147
FISCHERSCOPE® X-RAY XAN® 220/222
Dimensions XAN 220 XAN 222
External dimensions
Width x depth x height 403 x 588 x 365 mm 403 x 588 x 444 mm
(16 x 23.2 x 14.4 in) (16 x 23.2 x 17.5 in)
Weight ca. 45 kg (99 lb)
Environmental Conditions
Operating temperature 10 °C – 40 °C / 50 °F – 104 °F
Storage/Transport temperature 0 °C – 50 °C / 32 °F – 122 °F
Relative humidity ≤ 95 %, non-condensing
Evaluation Unit
Computer Windows® PC
Software Standard: Fischer WinFTM® BASIC including PDM®
Optional: Fischer WinFTM® SUPER
FISCHERSCOPE®, WinFTM®, XAN® and PDM® are registered trademarks of Helmut Fischer GmbH Institut für Elektronik und Messtechnik, Sindelfin-
gen- Germany. Windows® is a registered trademark of Microsoft Corporation in the United States and other countries.
www.helmut-fischer.com
Feature Overview
WINFTM®
Software for FISCHERSCOPE® X-RAY Fluorescence Measuring Instruments
Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
1.1 WinFTM Versions: BASIC and LIGHT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .151
1.2 WinFTM Extensions: PDM and SUPER. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .151
1.3 FISIM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .151
3 WinFTM PDM: Expanded Product Data Administration and Result Documentation . . . . . . . . . . . . 165
3.1 Comparative Overview with/without WinFTM PDM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .166
The WinFTM® (Fischer Thickness Management Software for Windows®) software controls the instrument
and handles the evaluation of the signals supplied by the instrument.
The measured values (coating thickness, material compositions, mass per unit area) are stored and dis-
played on the monitor. Using WinFTM, you can conveniently design the measurement results as a print form
for printout and export them to other applications.
WinFTM runs under Windows®. WinFTM version 6.50 or higher runs only on Windows 10. From version 6.50
on, the older Windows versions are no longer supported.
The usage of the SUPER software requires deep knowledge of the physics of x-ray fluorescence and the func-
tionality of the instrument.
Alternatively, you can request a new measurement application from Fischer. Fischer offers a professional
service for the creation of customer’s Def.MA.
1.3 FISIM
The different versions and extensions of WinFTM are activated by different FISIMs (Fischer Software Identifi-
cation Module). A FISIM is a dongle, which is plugged into the computer’s USB socket, before WinFTM is start-
ed. Look at the data sheet of your instrument to find out, which WinFTM version and extension is running on
your system by default. If you need to extend the features of your WinFTM software, contact Fischer.
151
Overview
The following section describes the features of the software versions BASIC and LIGHT. If there are restric-
tions in the LIGHT version, they are indicated in the text.
153
Features of WinFTM BASIC and LIGHT
Analysis with balance: It is not required to completely define the composition of the specimen by 100 per-
cent. The analysis is performed by matching the peak areas, without the need for the sum of concentra-
tions amounting to 100 percent. Applicable for recycling, analysis of trace elements in plastics, soil
specimens, biological specimens and non-homogeneous composites.
Function Description
Video image of the specimen The specimen is shown in the video window inside the WinFTM main win-
dow.
Scaled crosshairs Scaled crosshairs are electronically inserted into the video image. They
show the real size and position of the measurement spot on the surface of
the specimen. The color of the crosshairs is selectable.
Auto-focus (on models with The detector unit moves up and down automatically. The auto-focus func-
motor-driven z-axis) tion allows a significantly more precise setting of the measurement dis-
tance than subjective manual focusing. Thus, the reproducibility of the
measurement result is increased.
Measurement of the distance Focusing the video image provides the correct measuring distance. Win-
between the test spot and the FTM takes into account these distance values, when computing the coating
detector / DCM method thickness measurement data, using the DCM Method (Distance Controlled
Measurement).
This allows for automated testing of complex surface geometries and mea-
surements in recesses.
Displaying programmed You can display programmed points in the video image.
points in the video image (on
models with motor-driven XY-
stage)
Software-based video zoom You can zoom the video image in steps of 100, 200, 300 and 400 percent.
function Additionally, you can display the video image in full screen mode.
Saving video images together You can edit the video image with a graphic program of your choice e.g. for
with the measured values adding comments or dimensions. When you save the video image, it will be
assigned to your measured values. Additional text notes will be stored,
linked to the video image.
155
Features of WinFTM BASIC and LIGHT
Function Description
Recognition of vertical or hor- The instrument can recognize strip structures on specimens and readjust
izontal strips (on models with the XY position along this structure. Thus, you can make sure that on spec-
motor-driven XY-stage) imens with geometric tolerances, the measurement is always performed
at the correct position.
Image/Pattern Recognition The instrument can recognize a predefined structure and readjust the XY
(on models with motor-driven position of the XY-stage along this structure automatically. Thus, you can
XY-stage and WinFTM exten- compensate deviations in positioning during measurements on more spec-
sion SUPER activated) imen (e. g. for routine tests in production).
Function Description
Standard Windows® user in- Users with basic knowledge of Windows® will very quickly be familiar with
terface the program’s user interface.
Product functions You can set up, select, delete, edit and copy products. Additionally, you can
search for products, using different search criteria.
Definition of the spectral In the DefMA (Definition of the Measurement Settings of the Application)
background the spectral background can be defined. This function increases the accu-
racy of the analysis as long as a suitable underground spectrum is avail-
able.
Multiple Excitation For each application, the excitation parameters “high voltage” and “prima-
ry filter” are set to produce the best possible results. For some applica-
tions, however, it may be necessary to work with different excitations in
order to measure all parameters optimally. You can use multiple exci-
tations within a single measurement, so that all parameters are measured
under the best possible conditions; the collected results are then present-
ed in one combined evaluation.
Application example: Determination of the thickness of a NiP coating with
simultaneous determination of the Phosphorus concentration
Control field and joystick on In addition to the operation via the Software WinFTM you can control pro-
the measuring head (joystick gram and instrument functions with the control field and the joystick on
on models with motor-driven the measuring head.
XY-stage only)
Automatic searching and se- You can search for products by name (designation) using a bar-code read-
lecting of products (option) er. An optional bar-code keyboard (order no. 602-296) with a bar-code
reader pen (order no. 602-292) is available.
Short menu The supervisor can reduce the menu functions available for the operator.
In this manner, sensitive data such as measuring time and calibration pa-
rameters can be protected from overwriting, and the operator is present-
ed with a simplified menu. The long menu with full functionality is accessi-
ble only with a password.
Function Description
Quick selection of products You can select products and tasks quickly by using command buttons (on
and tasks the monitor), which can be edited.
Additionally, you can allocate products or tasks to the keys of the control
field on the measuring head.
mq value (measurement The mq value is a measure for the quality of a measurement result. It is
quality) displayed for every measurement and allows for a plausibility check of the
result. The mq value can be displayed or printed for every individual result.
A warning appears in case of a limit violation.
Calculating the random mea- The measurement uncertainty, which is based on natural count statistics,
surement uncertainty u of a is an important quantity in measurement technology.
measured quantity The measurement uncertainty u of the current single reading can be dis-
played.
The anticipated measurement uncertainty can be calculated individually
for each measured quantity.
A measured quantity can be varied in a wide range.
The calculated result for u is presented graphically as a chart: either as
theoretically anticipated standard deviation (SD) or as relative standard
deviation (RSD). This allows for an estimation of the measurement range.
If several measurement quantities are to be measured independent of one
another, the calculation will require specifications for the size of the other
variables that will not have the measurement uncertainty calculated. For
example, when determining the measurement uncertainty of a buried
coating (for example Ni in Au/Ni/Cu), the thickness of the top coating(s)
must be known.
The final result chart can be printed.
Traceability of the measurement result is possible by calculating the ran-
dom and systematic portions of the measurement uncertainty.
Random and systematic portion can be presented separately (both the sin-
gle readings and the mean values).
Uncertainties of the calibration standards can be entered.
The display of the measurement uncertainty(ies) can be shown or hidden.
Convenient copying of DefMA Copying of DefMA files is supported by a convenient data administration
tool.
Two overview windows (source directory and target directory) allow for a
convenient and clear control of the copy procedure. A DefMA file can be
opened and checked prior to copying.
Available DefMA files can be taken from one folder.
The Search/Sort function is helpful in finding the desired DefMA file.
Easily viewable DefMA notepad contents support searching.
157
Features of WinFTM BASIC and LIGHT
Function Description
Programmed measurement You can simply program routinely repeating measurement sequences and
sequences (Tasks) execute them automatically.
Even very complex measurement sequences can be simplified.
Dialogs for operators can be shown on the screen (e.g. information about
positioning of the specimen).
The sequence of a Task is started with the click of a soft key (icon) on the
monitor or with the push of a button on the control panel of the instrument.
Can be used even by semiskilled laborers.
Remote control You can operate WinFTM by remote control via your company network.
Network operation Beside the operator at the measuring station one or several other employ-
ees with linked computers can monitor the readings or can do evaluations.
On each of the client computers, WinFTM has to be installed in the same
version (e.g. BASIC or BASIC+PDM) like on the host computer.
International units of mea- You can display the readings in international units of measurement (μm /
surement mils …) and define special individual units of measurement.
Multitasking Many software functions can be operated simultaneously to the measure-
ment e.g. toggling the SPC channel, scrolling the list scroll field, perform-
ing an evaluation, changing the current product, working on the video
menu, help functions.
However, some other menu areas are locked during the measurement
(e.g. product selection, XY programming, calibration …).
Operating manual and con- The operating manual explains all essential facts of the program in a clear
text-sensitive online help manner.
The context-sensitive online help offers fast, direct access to easily under-
standable operational information.
Language selection You can select different languages for the menus and dialogues of the
WinFTM software. Available languages: English, German, French, Italian,
Spanish, Turkish, Polish, Czech, Swedish, Finnish, Portuguese. Selected
Asian languages on request.
Function Description
Block function A block is a concluded series of measurements. Measurement data is
stored and evaluated as blocks, including the date and time of the mea-
surement, a notepad, the continuous number of the block as well as block-
related data. The content of additional data fields such as order number,
batch number, and operator name can be defined freely.
Function Description
Tagging and commenting Blocks can be tagged with event marks and can receive comments. For
blocks example, in the course of a process that is to be controlled, you can tag
certain events by adding event marks and adding comments to the related
block.
Deleting measurement data Measurement data and blocks can be deleted, overwritten and undeleted.
and blocks
Function Description
Displaying the measurement You can display the measurement results in list form, as SPC chart or as a
results large number display (with optional display of the measurement uncer-
tainty). The resolution (Number of decimal places) can be set.
BASIC: Up to 24 result chan- You can control the representation for each channel individually.
nels,
LIGHT: Up to 3 result chan-
nels
Editing the video image and You can edit the video image with a graphic program of your choice e.g. for
storing it together with the adding comments, dimensions and explanatory text. The edited video im-
product age is stored together with the product.
Print functions With black and white printers, color values are printed in grayscale auto-
matically. For larger page widths, the printer is switched automatically to
landscape format and reduction of the type size.
SPC presentation of the The SPC chart can directly be displayed on the monitor as x-bar/s chart or
measurement results x-bar/r chart. For automated measurement sequences with large batch
sizes, you can easily recognize long-term changes of the coating thickness
and of other process parameters through the presentation of a trend line.
This enables, among other things, the testing of the measurement device
capability.
Specification limits: Graphi- BASIC: You can define specification limits for each of max. 24 result chan-
cal distinction (in color) for nels.
exceeding or falling short LIGHT: You can define specification limits for each of max. 3 result chan-
nels.
Single readings and statisti- Display and printout of single readings as well as statistical evaluation of
cal evaluation the measurement data in the form of block results or final results.
Displaying statistical evalua- Histograms, probability charts, statistical process control charts (SPC), Cp
tions and Cpk values can be displayed and printed.
Print forms for the block and You can print the block and final evaluation with editable print form head-
final evaluation er, measurement data, statistical results, histogram, probability chart, etc.
Exporting data You can export measurement results via the USB interface or via your
company network. You can define the format via export templates.
159
Features of WinFTM BASIC and LIGHT
Function Description
Two- and three-dimensional You can display the coating thickness distribution in two- or three-dimen-
display of the coating thick- sional charts.
ness distribution (on models
with motor-driven XY-stage)
Storing and displaying of ad- You can store and display the following additional information: date and
ditional information time of the last change to a product, the last measurement, normalization
or calibration of the current product, the last reference measurement, the
time of operation etc.
Storing XY(Z)-coordinates, The XY(Z)-coordinates, video images and spectra are stored together with
video images and spectra the measurement data for future use and documentation.
(XY(Z) coordinates on models
with motor-driven XY-stage)
Copying products including You can copy products including related measurement data e.g. for data
measurement data backup.
Automatic, time-controlled You can select folders, which will be backed up automatically. Thus, you
backup can back up your measurement applications, products including measure-
ment data, spectra etc. automatically.
Function Description
Calibration of new or existing The obtained calibration data are stored.
products
Standard-free calibration You can calibrate without calibration standards by means of the funda-
mental parameter method.
Standard-based calibration You can correct the standard-free calibration by calibrating with calibra-
with the use of up to 64 stan- tion standards. Thus, the accuracy of the analysis and thickness measure-
dards ments is improved.
Matrix-specific calibration standards and pure element calibration stan-
dards are available. Up to 64 calibration standards can be used for each
measurement application.
Calibration standard set Input and storing of nominal values of the used calibration standards is
performed in the Calibration standard set in the measurement application
of the product. After the calibration has been performed, the Calibration
standard set contains, in addition to the nominal values, the calculated ac-
tual values including the correction. You can use this data to create a doc-
ument for monitoring the measuring equipment.
Calculating and saving mea- Measurement parameters like measuring time, calibration time, correc-
surement parameters and tion data, data of the last normalization and calibration, etc. can be calcu-
calibration results lated and saved.
Library of pure element The library of pure element spectra makes measurement of pure ele-
spectra ments unnecessary when performing a calibration. This is important in
cases, where pure elements are not available.
Spectra library During the installation of the instrument the spectra of 12 or 14 elements
(depending on the detector type) are measured and saved in the spectra
library. Using these spectra all further spectra are calculated by the in-
strument.
You can control whether the elements are measured or the spectra of the
spectra library are used during calibration.
Reference measurement The Reference measurement checks and adjusts the drift of the instru-
ment. It is also used for monitoring the measuring equipment.
Stability test of the tube and The instrument performs automatic long-term self tests to check the pri-
of the detector mary radiation, the resolution and gain of the spectrometer (detector and
electronics).
Automatic reduction of the x- After several minutes of instrument idle the x-ray tube current is reduced
ray tube current automatically. Thus, the service life of the tube is increased and energy is
saved.
Drift compensation A shift of the peak position in the spectrum is compensated by the soft-
ware. This avoids wrong results especially when there are critical peak
overlaps.
161
Features of WinFTM BASIC and LIGHT
Function Description
Clear display of the sum For the evaluation of measured spectra you can overlay the spectrum
spectra with colored element lines of up to 24 elements in different colours.
spectra You can compare spectra in order to identify materials, e.g., for material
sorting.
You can compare the spectrum of an unknown material with the refer-
ence spectra of known materials, which are stored in a spectra library.
The number of stored spectra in the library can be increased or reduced.
Saving spectra You can save measured spectra for subsequent re-use (e.g., for subse-
quent evaluation when the measurement application changes). The spec-
tra can be stored manually or automatically.
Loading of spectra into the You can load spectra into the background of the spectra window to com-
background pare measured and stored spectra.
Zooming spectra ranges You can zoom into an area of the spectra window by using the right mouse
button.
Function Description
Moving the XY(Z) stage You can control the move of the XY(Z) stage accurately by clicking the
mouse button in the video image.
Joystick function of the The right mouse button offers additional joystick functionality in the video
mouse field.
Travel to absolute or relative You can enter absolute or relative coordinates to travel the stage. The zero
coordinates point of an operation can be specified individually.
Laser light pointer The red laser point points to the subsequent measurement position in the
retracted condition and thus significantly simplifies a quick sample posi-
tioning
Point & Shoot You can travel to the desired position by clicking the left mouse button.
Point & Measure You can travel to the desired position and perform a measurement at this
position by clicking the middle mouse button.
Point & Save You can travel to the desired position and save the related coordinates by
clicking the right mouse button.
Joystick at measuring head You can easily control the XY(Z) stage with the joystick at the measuring
head.
In addition, you can control the XY(Z) stage with the mouse.
Function Description
Measurement point overview The measurement point overview windows offer you a full overview of all
programmed points of the current measurement coordinates and of the
entire travel range.
Programming individual coor- You can program arbitrary individual coordinates, on which measurements
dinates will be performed.
Programming lines You can define a start point and an end point of a line and a number interim
points. On these points measurements will be performed.
Programming arrays You can define two corner coordinates of an array and the number of
points within the array (columns and rows). Furthermore you can define
the sequence, in which the points will be travelled. On these points, mea-
surements will be performed.
Programming patterns You can define a freely programmable geometric pattern of measurement
points. You can replicate this pattern as often as you want. Thus, you can
travel the same points on different specimens or on different parts of the
same specimen.
Programming offset In addition to the measuring points (line, array or pattern) you can define
one or two reference points. Thus, you can change specimens and make
sure the measurements are performed at the same coordinates, while the
specimen is not located at the same position.
163
Features of WinFTM BASIC and LIGHT
PDM (Product Data Management) is a supplementary software module to WinFTM BASIC or LIGHT. The PDM
software features the following functions:
Function Description
Setting up new products A measurement application may be linked to any number of products
linked to an existing (1:n assignment). Thus, each edit of a measurement application affects
measurement application all products that are linked to it.
With PDM, each product (a file that contains, among other things, the
measurement results) is saved in a freely selectable folder. You can
copy your products into different folders and sort them according to
supplier, customer and part number.
The operator only needs the product designation for the test order, and
not necessarily knowledge of the measurement application.
PDM saves measurement data product-related and not measurement
application related.
Convenient administration of You can set up new measurement applications based on an existing
measurement applications definition file (DefMA).
You can edit, copy and delete measurement applications. With PDM,
these procedures affect all products that are linked to a measurement
application. When copying a product, the measurement application is
not duplicated, instead it is retained.
You can normalize and calibrate measurement applications. These pro-
cedures affect all products that are linked to a measurement applica-
tion.
Result documentation You can assign individual print form templates to each product for
block and final evaluation.
You can evaluate the measurement readings of several or even all
products of a measurement application simultaneously in one tem-
plate.
You can freely set up, copy and edit product-related print forms (print
form templates).
A portion of the print form template may include special dummies
(variables) for measurement results and field contents. These dum-
mies take the place of user-entered data.
You can present your measurement results in a clear manner with
graphical effects. Additionally, various information elements can be
integrated like the image of the specimen, your company logo etc.
You can create and insert statistical characteristics.
You can save the print form in data files for future use or for documen-
tation purposes.
165
WinFTM PDM: Expanded Product Data Administration and Result Documentation
Function Description
Free block selection (Search Enables final evaluation and the subsequent export of selected mea-
function with filter) surement data blocks. The filter criteria applied may concern, for
example, date and time, block number, order no., batch no., operator
name.
You can combine arbitrary blocks to a final evaluation and export the
data of arbitrary blocks.
Overview image (on models You can create an overview image of the specimen automatically.
with motor-driven XY-stage) If you have programmed measurement points, they are displayed in the
overview image.
You can simply navigate and measure using the overview image (see Point
& Measure on page 162).
XY-Scan mode (on models You can travel the XY-stage while a measurement is running. Thus, you can
with motor-driven XY-stage) measure the mean value over a defined area of the specimen.
167
WinFTM PDM: Expanded Product Data Administration and Result Documentation
For both WinFTM versions BASIC and LIGHT the extension SUPER is available. With WinFTM SUPER you can
change the following parameters:
Measurement mode
Tube voltage used
Type and order of the coatings
Type and composition of the substrate material of the specimen and of the calibration standards
Treatment of interfering spectra
Special evaluation methods
These parameters are stored in the DefMA file (Definition of the Measurement Settings of the Application).
The usage of the SUPER software requires deep knowledge of the physics of x-ray fluorescence and the func-
tionality of the instrument.
Alternatively, you can request a new measurement application from Fischer. Fischer offers a professional
service for the creation of customer’s Def.MA.
In addition, the SUPER software features the following functions:
Function Description
Convenient entering of ele- Convenient entering of elements, e.g., in the DefMA using the window Peri-
ments odic System.
Pool of DefMA files You can establish a pool of DefMA files. Several users can access to this
pool.
Image/Pattern Recognition The instrument can recognize a predefined structure and readjust the XY
(on models with motor-driv- position of the XY-stage along this structure automatically. Thus, you can
en XY-stage and WinFTM ex- compensate deviations in positioning during measurements on more
tension SUPER activated) specimen (e. g. for routine tests in production).
169
www.helmut-fischer.com
HELMUT FISCHER GMBH INSTITUT FÜR ELEKTRONIK UND MESSTECHNIK
Industriestraße 21 Tel: +49 (0) 70 31 / 3 03 – 0
71069 Sindelfingen, Germany Fax: +49 (0) 70 31 / 3 03 – 710
Training Protocol
Installation and Training of an X-ray fluorescence measuring instrument
FISCHERSCOPE® X-RAY with WinFTM® software
Participants
Topics
No Topic ✔
www.helmut-fischer.com | mail@helmut-fischer.com
No Topic ✔
Base Correction
Density correction
Documentation
18 Creating an application via copy
19 Measuring and evaluating the spectrum
20 Deleting products etc.
A a product 50
Curved specimens 27
Administration, Product 49 Cylindrical specimens 27
Analysis with balance 97
Aperture 19
D
Areas of application 15
ASTM B568 9 Data export 119
Autom. Block closing, variable 55 DefMA 45, 46
Creating of 98
B Importing 54
Modes 97
Background correction 100 Viewing 53
Base correction 103 Deleting
Base spectrum 99 Measuring application 53
BASIC 151 Product 52
Block 55, 158 Determining the elements of a sample 43
Definitions 55 Diagnostic interface 23
Evaluation 68 DIN 50 987 9
Block size Disable the communication 26
Fixed 55 Dispose
variable 55 of the instrument 12
Bolts 27 Dongle 151
Dual Function of a Product 47
C
Calibration 104 E
Changing EN 61140 10
Measuring application 53 EN ISO 3497 9
Product 51 Enable the communication 26
Characteristic statistical parameters 131 Energy 44
Class of Materials 154 Environmental conditions 21
Cleaning 127 Establishing connections 23
Calibration standards 127 Evaluations
Instrument 127 statistical 55
Measuring cell 113 Event marks 159
Collimator 19 Export template 122
Communication
between instrument and WinFTM 26
F
disable 26
enable 26 FAQ x-radiation safety 14
Compensation spectra 99 File Structure of WinFTM 45
Component mode 97 Final evaluation 68
with balance 97 FISIM 25, 151
Component mode with balance 98 Fixed block size 55
Components 19 Flat Specimens 27
Connections 23 Fluorescence, x-ray 16
Copying Focus Adjustment Knob 20
Measuring application 53 Fully protected instrument 10
Product 49 Functional principle of the instrument 17
Cover of the Measurement Chamber 19
Creating
a DefMA 98
177
Index
H Deleting 53
Measuring cell 111
Handling of measurement readings 55 Cleaning of 113
Humidity, relative 21 Message
ERROR Helium Difference...ERROR Helium
I Difference... 31
Importing mq value 39
DefMA 54
Product 51, 114 N
Inclines, measurements on 28 Normalization 101
Instrument
disposal 12
O
Instrument, functional priciple of 17
Intended use 9 Operating personnel, requirements 12
Intensity 44 Operation temperature 21
Operators with introductory training 12
K Overwriting a measurement reading 58
Key switch 19, 23
P
L PDM 151, 165
Supplementary software 61
Large number presentation 36 Performing Measurements 31
Liability limitation 10 Periodic table of the elements 140
LIGHT 151 Position Specimen 27
Lightning, specimen 19 Preparing
Limitation of liability 10 Instrument for transport 129
Limited operating mode 40 Measurements, process flow 45
List presentation 36 Primary x-radiation 16
Location for set-up 21 Print Form Template Editor 61
Process flow for preparing measurements 45
M Product 45, 46
Main switch 23 Administration 49
Main window WinFTM 33 Assigning a new measuring application 52
Maintenance 127 Changing 51
Repair 10 Copying 49
Manual measurements 31 Creating 50
Mass per unit area 77, 97 Deleting 52
Material analysis 97 Importing 51, 114
Measurement
Application 45, 46 R
Chamber cover 19 Ratio method 155
Data export 119 Reference measurement 101
On inclines 28 Relative humidity 21
Uncertainty 36 Repair and maintenance 10
Value field 36 Repeat inspection 10
Measurement reading Requirements for the operating personnel 12
Saving 58 Residual 154
Measurement readings Rods 27
Handling of 55 ROI-mode 97
Measuring application
Assigning Product 52
S
Changing 53
Copying 53 Safety 9, 10
T
Temperature 21
Thickness mode 97
Trained specialists 13
Transporting the instrument 129
Troubleshooting 129
U
Uncertainty of the measurement 36
Unpacking the measuring system 22
USB
cable 23
USB interface 23
User interface of the WinFTM software 33
V
Variable block size 55
Video
Image 34
Viewing
DefMA 53
Voltage supply 23
W
Warning notices 9
WinFTM
File structure 45
Main window 33
179
Index