Professional Documents
Culture Documents
1. Version Control
Version Number Created by Date Reviewed by Date
1 DA 2012-10-16 MF 2012-10-17
2 DA 2014-05-07 KT 2014-05-12
3 FC 2017-12-13
Pre-Requisites
• Cross-Screwdriver.
• Flat-Screwdriver.
• Pipette.
• Di-Water
• Allen key 2.5mm and 3mm
• Multimeter
3. Procedure Part I
Loosen the screws and open the analyzer.
Page 1 of 13
Technical Support Bulletin
Open right backdoor and adjust lamp Voltage to 11.8V by turning the plastic screw on the 12V DC
power supply.
Remove RCT Cover and the section of RCT to get access to the lamp housing.
Page 2 of 13
Technical Support Bulletin
Procedure Part II
Adjust Lamp housing and Lens
Take out the Cuvette No. 39 and check whether the light beam is in the middle of the photometer
hole.
Page 3 of 13
Technical Support Bulletin
Unscrew the two screws of the lens and move the lens plan to the lamp housing (to the left side),
tighten the screws after moving the lens.
Unscrew the two screws of the lamp housing and check whether the light beam is in the middle of
the photometer hole.
Page 4 of 13
Technical Support Bulletin
Page 5 of 13
Technical Support Bulletin
Please put a paper inside the light beam and perform an Autospan. We want to have all the gains out
of range.
Page 6 of 13
Technical Support Bulletin
Page 7 of 13
Technical Support Bulletin
Open [Maintenance], select „Manual Span“ and select 700nm filter and press [Start]. Voltage and
absorbance will be displayed and values will be refreshed every 5sec.
Page 8 of 13
Technical Support Bulletin
Align the photometer manually by moving it gently to all directions in order to achieve the maximum
voltage. Please mind the 5sec. delay!
If the highest Voltage (max 2.5V) is achieved, fix only one screw as shown in the picture below.
Page 9 of 13
Technical Support Bulletin
Page 10 of 13
Technical Support Bulletin
Align the photometer manually by moving it gently to achieve the maximum voltage (max 2.5V).
Verify the center of the light path using the tool, provided with the analyzer, as below:
Das Bild k ann zurzeit nicht angezeigt werden.
Page 11 of 13
Technical Support Bulletin
In case the light is not in the center of the tool, please perform the Procedure Part III.
Page 12 of 13
Technical Support Bulletin
Perform an „AutoSpan“
If the gain values match, perform “Cuvette rinse” two times consecutively and observe whether the
graphs are congruent respectively parallel with a shift less then 0,003abs. (see screenshot below)
Page 13 of 13