Professional Documents
Culture Documents
Calibration of Medical
Infusion Pumps
Guidance Note:
Calibrating Medical Infusion Pumps
buoyancy and the thermal expansion of the disposable It is important to identify how the error is calculated as
syringe [6,7]. The volumetric flow rate is determined from the otherwise false corrections may be applied. More detailed
quotient of the total volume and time taken for the delivery of information on these conflicting definitions is provided in [11].
the liquid.
Common Issues with the Calibration of
The volumetric flow rate (Q) can be expressed as:
Infusion Pumps
1 1
1 1 20
Air bubbles – a sufficient volume of water should be
passed to purge any air bubbles.
Flow stability and pump response time – there is a
Where tf is the final time, ti is the initial time, If is the final need to wait until the target infusion flow rate has been
balance reading, Ii is the initial balance reading, ρw is the reached, as large errors will occur before this. The
density of water used for the calibration, ρA is the density of time taken before the required target flow rate is
air, ρB is the density of the mass standards used to calibrate achieved is often referred to as the start-up delay or
the balance, ϒ is the coefficient of thermal expansion of the delayed onset. These factors are illustrated in the
disposable syringe and T is the temperature of the water diagram below:
used. More information on the coefficients is provided in ISO
4787.
This publication is to provide outline information only which (unless agreed by TÜV SÜD National Engineering Laboratory in writing) may not be reproduced for any
other purpose or form part of any order or contract or be regarded as representation relating to products or services concerned.