MODEL: M2LR

Super-mini Signal Conditioners Mini-M Series available.)

RTD TRANSMITTER [2] OUTPUT


(field- and PC-configurable) Current
Functions & Features Z1: Range 0 – 20 mA DC
• Accepts direct input from an RTD and provides a standard Voltage
process signal V1: Range -2.5 – +2.5 V DC
• Easy ‘One-Step Cal’ calibration using the front three V2: Range -10 – +10 V DC
control buttons without needing a PC; PC software is also Output ranges are fully programmable out of the range
usable. specified when ordering by setting up the DIP switch and
• Input range, output type and range are configurable using ‘One-Step-Cal’ Calibration. PC Configurator is also
available.
23 (.91)

POWER INPUT
AC Power
76
(2.99) M: 85 – 264 V AC (Operational voltage range 85 – 264 V,
47 – 66 Hz)

124
(4.88)
[3] OPTIONS
mm (inch) blank: none
/Q: With options (specify the specification)

MODEL: M2LR-[1][2]-M[3]
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to M-System's web site.)
ORDERING INFORMATION /C01: Silicone coating
• Code number: M2LR-[1][2]-M[3] /C02: Polyurethane coating
Specify a code from below for each [1] through [3]. /C03: Rubber coating
(e.g. M2LR-4Z1-M/Q) TERMINAL SCREW MATERIAL
• Temperature range (e.g. 0 – 100°C) /S01: Stainless steel
• Output range (e.g. 4 – 20 mA DC)
• Specify the specification for option code /Q
(e.g. /C01/S01)
RELATED PRODUCTS
• PC configurator software (model: M2LRCFG)
Downloadable at M-System’s web site.
[1] INPUT RTD (2- or 3-wire) A dedicated cable is required to connect the module to the
1: JPt 100 (JIS'89) PC. Please refer to the internet software download site or
(Usable range: -200 to +500°C, -328 to +932°F; min.span: 20°C, 36°F) the users manual for the PC configurator for applicable
3: Pt 100 (JIS'89) cable types.
(Usable range: -200 to +650°C, -328 to +1202°F; min.span: 20°C, 36°F)
4: Pt 100 (JIS'97, IEC)
(Usable range: -200 to +850°C, -328 to +1562°F; min.span: 20°C, 36°F)
GENERAL SPECIFICATIONS
Construction: Plug-in
5: Pt 50 Ω (JIS'81)
Connection: M3 screw terminals (torque 0.8 N·m)
(Usable range: -200 to +649°C, -328 to +1200°F; min.span: 20°C, 36°F)
Housing material: Flame-resistant resin (black)
7: Pt 1000
Isolation: Input to output to power
(Usable range: -200 to +850°C, -328 to +1562°F; min.span: 20°C, 36°F)
Overrange output: Approx. -15 to +115 %
9: Cu 10 @25°C
Zero adjustment: -15 to +15 % (front)
(Usable range: -50 to +250°C, -58 to +482°F; min.span: 20°C, 36°F)
Span adjustment: 85 to 115 % (front)
(Consult M-System for users’ tables.)
Burnout protection: Upscale, downscale or no burnout
(Input range are fully programmable out of the range
selectable with DIP switches
specified when ordering by setting up the DIP switch and
Status indicator LED: Tri-color (green/amber/red) LED;
using ‘One-Step-Cal’ Calibration. PC Configurator is also
Flashing patterns indicate operation status of the

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
MODEL: M2LR
transmitter. INSTALLATION
Configuration Power Consumption
‘One-Step Cal’ calibration: With I/O type and the full-scale •AC Power input:
range configured via the side DIP switches, precise 0 % and Approx. 4 VA at 100 V
100 % ranges are calibrated via the front control buttons Approx. 5 VA at 200 V
with a help of LED. Approx. 6 VA at 264 V
PC configurator: Via Windows PC connected to the front Operating temperature: -5 to +55°C (23 to 131°F)
jack. Operating humidity: 30 to 90 %RH (non-condensing)
Programmable features include: Mounting: Surface or DIN rail
•I/O type and range Weight: 120 g (0.26 lbs)
•Zero and span adjustments
•Burnout
•User’s linearization table PERFORMANCE
Configurator connection: 2.5 dia. miniature jack; Overall accuracy: Input accuracy + output accuracy
RS-232C level Inversely proportional to the span.
See CALCULATION EXAMPLES OF OVERALL ACURACY.
• Input accuracy: (whichever is greater)
INPUT SPECIFICATIONS ±0.1 % of FS or ±0.15°C (Pt and JPt)
Maximum leadwire resistance: 20 Ω per wire (3-wire) ±0.1 % of FS or ±1°C (Cu 10)
Sensing current: ≤ 1.2 mA • Output accuracy: ±0.04 % of max. output range
If not specified, the input range is 100°C. Temp. coefficient: ±0.015 %/°C (±0.008 %/°F) of max. span
Response time: ≤ 1 sec. (0 – 90 %)
OUTPUT SPECIFICATIONS Burnout response time: ≤ 1 sec.
• DC Current Line voltage effect: ±0.1 % over voltage range
Maximum range: 0 – 20 mA DC Insulation resistance: ≥ 100 MΩ with 500 V DC
Conformance range: 0 – 24 mA DC Dielectric strength: 2000 V AC @1 minute (input to output
Minimum span: 1 mA to power to ground)
Offset: Lower range can be any specific value within the
output range provided that the minimum span is CALCULATION EXAMPLES OF OVERALL ACCURACY
maintained. [Example] Pt 100, 0 – 100°C, Output Type -10 – +10 V,
Load resistance: Output drive 12 V max. Output Range 1 – 5 V
(e.g. 4 – 20 mA: 600 Ω [12 V/20 mA]) Input Accuracy* (0.15°C**) / Span (100°C) × 100 %
If not specified, the output range is 4 – 20 mA DC. + Max. Output Range (20 V) / Span (4 V) × 0.04 %
• DC VOLTAGE = ±0.35 %
Code V1 (narrow spans) *Calculate the accuracy in °C.
Conformance range: -3 – +3 V DC **100°C × 0.1 % = 0.1°C ≤ 0.15°C. 0.15°C is used as input
Output range: -2.5 – +2.5 V DC accuracy value.
Minimum span: 250 mV
Code V2 (wide spans)
Conformance range: -11.5 – +11.5 V DC
Output range: -10 – +10 V DC
Minimum span: 1 V
Offset: Lower range can be any specific value within the
output range provided that the minimum span is
maintained.
Load resistance: Output drive 1 mA max.
(e.g. 1 – 5 V: 5000 Ω [5 V / 1 mA])
If not specified, the output range is shown below.
V1: 0 – 1 V DC
V2: 1 – 5 V DC

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
MODEL: M2LR
EXTERNAL VIEW
■ FRONT VIEW ■ SIDE VIEW
LED1 (LD1)

LED2 (LD2) Configuration DIP SW

LED3 (LD3)
1 2 3 4 5 6 7 8
ON
Configurator Jack
OFF
MODE Button
SW2

UP Button 1 2 3 4
ON SW1
DOWN Button
OFF

For detailed information on the configuration and calibration, refer to the instruction manual.

EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch)


6 (.23)

DIN RAIL
35mm wide 4 1
8 5
2–4.2x5
(.17x.20)
70.5 (2.78)

59 (2.32)
72 (2.83)
MTG HOLE
6 (.24) deep

8–M3 12 9
SCREW 14 13

21.5 (.85) 10 (.39) 84 (3.31) 15 (.59)


4 (.16)

114 (4.49) [4 (.16)] 23 (.91)

• When mounting, no extra space is needed between units.

SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM

STATUS CONTROL SW
2.5V DIP SW Isolation
RTD
Digital Output
A 5 Computation Driver
9 +
OUTPUT
Low Drift
B 1 12 –
Amplifier
B 8 13 U
2.5kΩ POWER
14 V

CONFIGURATOR JACK 4

Base Socket

Specifications are subject to change without notice.

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
M2LR
INSTRUCTION MANUAL
RTD TRANSMITTER MODEL M2LR
(field- and PC-configurable)

BEFORE USE .... COMPONENT IDENTIFICATION


Thank you for choosing M-System. Before use, please check Body Base Socket
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.

Fixing Screw
■ PACKAGE INCLUDES:
Signal conditioner (body + base socket)....................... (1)
Connection
■ MODEL NO. Diagram
Confirm Model No. marking on the product to be exactly Specifications
what you ordered.

■ INSTRUCTION MANUAL
This manual describes necessary points of caution when INSTALLATION
you use this product, including installation, connection and
Loosen the fixing screw at the front of the unit in order to
basic maintenance procedures.
separate the body from the base socket.
The M2LR is programmable using the PC configurator soft-
ware. For detailed information on the PC configuration, re-
■ DIN RAIL MOUNTING
fer to the M2LRCFG instruction manual. The M2LRCFG
Set the base socket so that
PC Configurator Software is downloadable at M-System’s
its DIN rail adaptor is at
web site: http://www.m-system.co.jp.
the bottom. Position the
upper hook at the rear side DIN Rail
of base socket on the DIN 35mm wide
POINTS OF CAUTION rail and push in the lower.
■ POWER INPUT RATING & OPERATIONAL RANGE When removing the socket,
• Locate the power input rating marked on the product and push down the DIN rail
confirm its operational range as indicated below: adaptor utilizing a minus
85 – 264V AC rating: 85 – 264V, 47 – 66 Hz, approx. 4 – 6VA screwdriver and pull.

Spring Loaded
■ GENERAL PRECAUTIONS ■ WALL MOUNTING
DIN Rail Adaptor
• Before you remove the unit from its base socket or mount Refer to “EXTERNAL DI-
it, turn off the power supply and input signal for safety. MENSIONS.”

■ ENVIRONMENT
• Indoor use TERMINAL CONNECTIONS
• When heavy dust or metal particles are present in the air,
Connect the unit as in the diagram below.
install the unit inside proper housing with sufficient ven-
tilation.
• Do not install the unit where it is subjected to continuous RTD
A 5 9 +
vibration. Do not subject the unit to physical impact. OUTPUT
• Environmental temperature must be within -5 to +55°C B 1 12 –
(23 to 131°F) with relative humidity within 30 to 90% RH
B 8 13 U
in order to ensure adequate life span and operation. POWER
• Be sure that the ventilation slits are not covered with ca- CONFIGURATOR JACK 14 V
bles, etc.

■ WIRING
• Do not install cables (power supply, input and output)
close to noise sources (relay drive cable, high frequency
line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.

■ AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
sheet.

EM-5057-A Rev.2 P.  / 6

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
M2LR

EXTERNAL VIEWS
■ FRONT VIEW ■ SIDE VIEW
LED1 (LD1)

LED2 (LD2) Configuration DIP SW

LED3 (LD3)
1 2 3 4 5 6 7 8
ON
Configurator Jack
OFF
MODE Button
SW2

UP Button 1 2 3 4
ON SW1
DOWN Button
OFF

CONFIGURATION MODE & DIP SW SETTINGS


When you program the transmitter module, two configura- ■ CONFIGURATION MODE (SW2) Table 1
tion modes are available: Field Configuration using DIP SW MODE SW2-8
/ control buttons, and PC Software. Configuration mode can be
DIP SW OFF
confirmed with the front LED.
The internal DIP switches are used to configure input and PC (*) ON
output type. Once the module is configured, precise ranges
(*) Factory setting
are set up with the front control buttons using a simula-
tor connected to the input terminals and a multimeter con- ■ INPUT TYPE (SW2) Table 2
nected to the output terminals as a reference. INPUT SW2-7 SW2-6 SW2-5
The calibrated input and output ranges are stored in the in- JPt 100 (JIS ’89) OFF OFF OFF
ternal memory. The module reads the DIP-switch-calibrat- Pt 100 (JIS ’89) ON OFF OFF
ed configuration only once after the power supply is turned Pt 100 (JIS ’97, IEC) OFF ON OFF
on. Set the switches with the power supply removed. Pt 50Ω (JIS ’81) ON ON OFF
Selectable I/O type and ranges are listed in Table 6 and 7. Pt 1000 OFF OFF ON
Cu 10 @25°C ON OFF ON
■ DIP SW CONFIGURATION MODE User’s table ON ON ON
Turn the SW2-8 OFF to enable the DIP SW (Field Configu-
■ BURNOUT (SW2) Table 3
ration) mode as shown in Table 1.
See Table 2 to configure the input and Table 4 for the out- BURNOUT SW2-1 SW2-2
put. No burnout OFF OFF
Upscale (*) OFF ON
■ PC CONFIGURATION MODE Downscale ON ON
Turn the SW2-8 ON to enable the PC Configuration mode as (*) Factory setting
shown in Table 1. All programmable features can be set up
■ OUTPUT TYPE (SW2 & 1) Table 4
on a PC regardless of other DIP SW setting except that the
OUTPUT SW2-4 SW2-3 SW1-4 SW1-3 SW1-2 SW1-1
output type must be selected with the DIP SW1-1 through
SW1-4 (See Table 5). 0 – 20mA OFF OFF OFF ON OFF OFF
For detailed information on the PC configuration, refer to -2.5 – +2.5V OFF ON ON OFF OFF ON
-10 – +10V ON OFF ON OFF ON OFF
the M2LRCFG instruction manual.
■ OUTPUT TYPE / PC CONFIG (SW1) Table 5
OUTPUT SW1-4 SW1-3 SW1-2 SW1-1
0 – 20mA OFF ON OFF OFF
-2.5 – +2.5V ON OFF OFF ON
-10 – +10V ON OFF ON OFF

EM-5057-A Rev.2 P.  / 6

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
M2LR
■ INPUT TYPE & RANGE Table 6
°C °F
MIN. SPAN MAX. RANGE MIN. SPAN MAX. RANGE
JPt 100 (JIS ’89) 20 -200 to +500 36 -328 to +932
Pt 100 (JIS ’89) 20 -200 to +650 36 -328 to +1202
Pt 100 (JIS ’97, IEC) 20 -200 to +850 36 -328 to +1562
Pt 50 (JIS ’81) 20 -200 to +649 36 -328 to +1200
Pt 1000 20 -200 to +850 36 -328 to +1562
Cu 10 @25°C 20 -50 to +250 36 -58 to +482

■ OUTPUT TYPE & RANGE Table 7


OUTPUT TYPE MINIMUM SPAN MAXIMUM RANGE CONFORMANCE RANGE
DC Current 1mA 0 to 20mA 0 to 24mA
DC Voltage, Narrow Spans 0.25V -2.5 to +2.5V -3 to +3V
DC Voltage, Wide Spans 1V -10 to +10V -11.5 to +11.5V

■ INPUT & OUTPUT RANGING


CHECKING [Example]
1) Run Mode: Confirm that the LD1 green light is blink-
1) Terminal wiring: Check that all cables are correctly con-
ing.
nected according to the connection diagram.
2) Input Configuration Mode: Press MODE button for long-
2) DIP SW setting: Check that the switches are set to ap-
er than 5 seconds until the LD1 red light is ON and the
propriate positions.
LD2 red light is blinking.
3) Power input voltage: Check voltage across the terminal
3) 0% Input Ranging: Apply the desired minimum input
13 – 14 with a multimeter.
level (e.g. 0°C) from the simulator and push DOWN but-
4) Input: Check that the input signal is within 0 – 100% of
ton until the LD1 flashes for approx. 2 sec. and then turns
full-scale.
OFF. When you release the button, the LD1 is returned
If the RTD or its extension wires are broken, the output
to ON.
goes over 100% (below 0% with downscale burnout) due
The flashing LD1 means that the value is stored in the
to burnout function. Confirm the status indicator LED
memory. If the LED does not change, the entered level
pattern and check leadwires in such a case.
may be inappropriate: too small a span, or out of usable
5) Output: Check that the load resistance meets the de-
range (same for all steps).
scribed specifications.
4) 100% Input Ranging: Apply the desired maximum input
level (e.g. 100°C) from the simulator and push UP button
until the LD1 flashes for approx. 2 sec. and then turns
I/O RANGING & FINE ADJUSTMENTS OFF. When you release the button, the LD1 is returned
After the DIP SW setting is complete, set up the precise to ON.
input and output range using the front control buttons. 5) Output Configuration Mode: Push MODE button and
The front LEDs’ colors and flashing patterns help you to confirm that the LD3 red light instead of LD2 is blink-
easily identify the transmitter’s status and confirm the set- ing.
up actions in each step of Calibration Modes. Please read 6) 0% Output Ranging: Increase or decrease the simulated
the following explanations referring to “Calibration Flow input until the meter shows the desired minimum out-
Chart” in the next page. put level (e.g. 1V). Push DOWN button until the LD1
flashes for approx. 2 sec. and then turns OFF. When you
■ PREPARATION (e.g. Input Pt 100: 0 – 100°C, Output 1 – 5V) release the button, the LD1 is returned to ON.
1) Mount the DIP-SW-configured M2LR on to a DIN rail. 7) 100% Output Ranging: Increase or decrease the simu-
2) Connect the M2LR to a simulator and a multimeter and lated input until the meter shows the desired maximum
to a DC power source as shown below. output level (e.g. 5V). Push UP button until the LD1
3) Turn the power supply on and wait for 10 minutes. flashes for approx. 2 sec. and then turns OFF. When you
release the button, the LD1 is returned to ON.
M2LR Multimeter 8) Run Mode: Programming complete, push MODE button
Simulator
+ and confirm that only the LD1 green light is blinking.
A 5 9 –

B 1 12
B 8 13
AC Power Source
14

EM-5057-A Rev.2 P.  / 6

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
M2LR

■ CALIBRATION FLOW CHART

POWER ON

MODE A Amber LED OFF


LD1
Control
G UP G Green LED ON
Buttons
Fine Adjustments ? LD2
RUN MODE I/O Ranging ?

LD3
DOWN R Red LED Blink
G

PUSH MODE 1 – 2 s PRESS & HOLD MODE > 5s


Apply simulated 0% input signal.
LD1 Once the transmitter LD1
enters the fine zero 0% INPUT
R R PUSH DOWN
FINE ZERO calibration mode, INPUT CONFIGURED
LD2 LD2
CALIBRATION MODE the output goes to CALIBRATION
G preadjusted 0% R
LD3 LD3
MODE 100% INPUT
UP or DOWN value regardless PUSH UP
of the present input CONFIGURED
signal.
Apply simulated 100% input signal.

PUSH MODE PUSH MODE


Adjust simulated input until the output meter
shows desired 0% output.
LD1 Once the transmitter LD1
enters the fine span 0% OUTPUT
R R PUSH DOWN
FINE SPAN calibration mode, OUTPUT CONFIGURED
LD2 LD2
CALIBRATION MODE the output goes to CALIBRATION
preadjusted 100%
LD3 LD3
MODE 100% OUTPUT
UP or DOWN value regardless PUSH UP
G of the present input R CONFIGURED
signal.
Adjust simulated input until the output meter
shows desired 100% output.
PUSH MODE PUSH MODE
When you set 0% or 100% input/output ranges,
keep pressing UP or DOWN button until the
LD1 flashes for approx. 2 seconds and turns off,
LD1
which indicates the setup is complete.
G When you release the button, the LD1 is returned
LD2 to ON.
RUN MODE If the LED does not change, the entered level may
LD3 be inappropriate: e.g. out of usable range.

■ ZERO & SPAN ADJUSTMENTS Note 1: Calibration steps can be skipped when not needed
After the transmitter is installed and operational, fine zero by repeating pushing MODE buttons.
and span tuning can be performed as explained below. Both Note 2: There is no stated order of setting 0% and 100%
zero and span are adjustable within ±15%. levels or no limitation of entering values for multiple times
1) Run Mode: Confirm that the LD1 green light is blink- within one step of Calibration Mode. Signal level is stored
ing. each time the respective UP or DOWN button is pressed.
2) Fine Zero Calibration Mode: Press MODE button for 1
or 2 seconds until the LD1 red light is ON and the LD2
green light is blinking.
3) Use UP (increase) and DOWN (decrease) buttons to ad-
just the output to 0%.
4) Fine Span Calibration Mode: Push MODE button and
confirm that the LD3 green light instead of LD2 is blink-
ing.
5) Use UP (increase) and DOWN (decrease) buttons to ad-
just the output to 100%.
6) Run Mode: Push MODE button and confirm that only the
LD1 green light is blinking.

EM-5057-A Rev.2 P.  / 6

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
M2LR

STATUS INDICATOR LED


Combinations of the three front LEDs (LD1, LD2, LD3) indicate the transmitter’s operating status by different flashing pat-
terns.
Examples are shown below. A Amber LED OFF

G Green LED ON

R Red LED Blink

NORMAL OPERATION (RUN) MODE


LD1 LD1
PC Configuration RUN Mode DIP SW Configuration RUN Mode
G G
LD2
The transmitter is configured via PC and LD2
The transmitter is configured via DIP SW and
is in normal operating conditions. is in normal operating conditions.
LD3 LD3

ERROR MODE
LD1 LD1
Parameter Error Output Saturated in PC Configuration Mode
A G
LD2
Error occurred when reading parameters from the LD2
The output is below -15% or above +115%.
internal memory. Turn the power suppy off and on.
LD3 If the error still exists, reset the parameters to default LD3
and program the unit again. A
LD1 LD1
Communication Error Output Saturated in DIP SW Configuration Mode
*1 G
LD2
Internal CPU’s communication error. LD2
The output is below -15% or above +115%.
A
LD3 *1. LD1 may be one of the G G A . LD3
A
LD1 LD1
AD Conversion Error Burnout in PC Configuration Mode
*1 G
LD2
Internal AD conversion error. LD2
R
LD3 *1. LD1 may be one of the G G A . LD3
R
LD1 LD1
DIP SW Error Burnout in DIP SW Configuration Mode
R G
LD2
DIP SW configuration is inappropriate. LD2
R Check the DIP SW setting referring to Tables 2 – 10. R
LD3 LD3
R

HOW TO RESET TO DEFAULT STATUS


1) Set SW2-1 to ON position, SW2-2 to OFF position.
2) Turn on the power supply to the transmitter while press-
ing MODE button.
3) Wait until green light turns on at the LD1, LD2 and
LD3.
4) Reset SW2-1 to OFF position. Turn off and on the power
supply.

EM-5057-A Rev.2 P.  / 6

SENTRONIC AG Produkte, Support und Service


Rugghölzli 2
CH - 5453 Busslingen
Tel. +41 (0)56 222 38 18
Fax +41 (0)56 222 10 12
mailbox@sentronic.com
www.sentronic.com
M2LR

EXTERNAL DIMENSIONS unit: mm (inch)


6 (.23)

DIN RAIL
35mm wide 4 1
8 5
2–4.2x5
(.17x.20)
70.5 (2.78)

59 (2.32)
72 (2.83)
MTG HOLE
6 (.24) deep

8–M3 12 9
SCREW 14 13

21.5 (.85) 10 (.39) 84 (3.31) 15 (.59)

4 (.16)
114 (4.49) [4 (.16)] 23 (.91)

• When mounting, no extra space is needed between units.

M-SYSTEM WARRANTY
M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System’s sole liability, and purchaser’s exclusive
remedies, under this warranty are, at M-System’s option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this
warranty. To submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original
sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: info@m-system.co.jp

EM-5057-A Rev.2 P.  / 6