Professional Documents
Culture Documents
eBike PC
Firmware Version 3
Operator’s Manual
The information in this manual only applies to the eBike PC, firmware version 3. It does not apply to earlier firmware
versions.
Due to continuing product innovation, specifications in this manual are subject to change without notice.
CardioSmart, CardioSys, CASE, and MAC are trademarks owned by GE Medical Systems Information Technologies, a
General Electric Company going to market as GE Healthcare.
4 Preparation 14
7 Configuration 27
8 Error Messages 33
10 Technical Specifications 35
11 Order Information 38
12 Appendix 39
General Information
The product eBike PC bears the CE marking CE- The safety information given in this manual is
0123 (notified body TÜV Süd Product Service classified as follows:
GmbH, Ridlerstr. 65, 80339 München, Germany)
indicating its compliance with the provisions of the Danger
Council Directive 93/42/EEC about medical devices indicates an imminent hazard. If not avoided, the
and fulfills the essential requirements of Annex I of hazard will result in death or serious injury.
this directive. It is an MDD class IIa product.
The device fulfills the requirements of standard EN/ Warning
IEC 60601-1 "Medical Electrical Equipment, Part 1:
indicates a hazard. If not avoided, the hazard can
General Requirements for Safety" as well as the
result in death or serious injury.
interference protection requirements of standard EN/
IEC 60601-1-2 "Electromagnetic Compatibility –
Caution
Medical Electrical Devices" and all applicable
amendments. indicates a potential hazard. If not avoided, this
hazard may result in minor personal injury or
It is protection class II equipment.
product/property damage.
The device is in radio-interference protection class B
in accordance with CISPR11/EN 55011.
To ensure patient safety and interference-free
The CE marking covers only the accessories listed in
operation and to guarantee the specified measuring
the "Order Information" chapter.
accuracy, we recommend only original equipment
This manual is an integral part of the equipment. It accessories as available through GE Healthcare
should be available to the equipment operator at all distribution. The user is responsible if accessories
times. Close observance of the information given in from other manufacturers are used.
the manual is a prerequisite for proper equipment
GE Healthcare is responsible for the effects on safety,
performance and correct operation and ensures
reliability, and performance of the device, only if
patient and operator safety. Please note that
– assembly operations, extensions, readjustments,
information pertinent to several chapters is given
modifications, or repairs are carried out by GE
only once. Therefore, carefully read the manual
Healthcare or by persons authorized by GE
once in its entirety.
Healthcare
If unauthorized personnel opens the control terminal,
– the device is used in accordance with the
damaging the calibration sticker, any warranty claim
instructions given in this manual.
shall become void.
This manual reflects the equipment specifications and
applicable safety standards valid at the time of
printing. All rights are reserved for devices, circuits,
techniques, software programs, and names appearing
in this manual.
On request GE Healthcare will provide a Service
Manual.
ergoline GmbH
Lindenstraße 5
D-72475 Bitz, Germany
Tel. +49 7431 9894-0
Fax +49 7431 9894-127
Distributor:
GE Medical Systems
Information Technologies, Inc.
8200 West Tower Avenue
Milwaukee, WI 53223 USA
Tel. +1 414 355 5000
1 800 558 7044 (USA only)
1 800 668 0732 (Canada only)
Fax +1 414 355 3790
PVC-free.
2.2 Equipment Symbols
NIBP
With this button on the control terminal, a
blood pressure measurement can be
initiated before or during the exercise Small size.
test. The same button can be used to stop
a blood pressure measurement. Standard size.
Symbol ’type B applied part’.
Large size.
Catalog number.
Serial number.
Danger Warning
Explosion Hazard — Equipment Failure —
The device is not designed for use in areas where an – Cellular telephones must not be used in the
explosion hazard may occur. An explosion hazard immediate vicinity, because they may interfere
may result from the use of flammable anesthetics, with the proper functioning of the ergometer.
skin cleansing agents and disinfectants. Electromagnetic interference most probably
exists when the watt reading changes frequently
even though the speed is above 30 RPM.
Warning
Patient Hazard, Equipment Damage — Shock Hazard —
– Protect the eBike from direct sunlight. – When the ergometer is connected to other
Otherwise equipment components may reach equipment or if a medical system is created, it
inadmissibly high temperatures. The eBike is a must be ensured that the added leakage currents
medical device and must NEVER be used do not present a hazard. If you have questions
outdoors. Furthermore, the device has no about this issue, please contact your GE
additional protection against ingress of Healthcare dealer or GE Healthcare Service.
humidity. Humidity that has entered the device – For use, the ergometer may only be connected to
may cause equipment failure and can increase electric installations that fulfill the local
the risk of electric shock. requirements.
Also, do not operate the eBike in the vicinity of Patient Hazard —
power installations, because they may impair
The German Medical Device Operator Ordinance
equipment functions.
(MPBetreibV, § 5) demands that users
Risk to Persons —
– must have been trained in the use of the
Before using the ergometer, the operator must
ergometer
ascertain that it is in correct working order and
– must be familiar with the routines for the
operating condition. In particular, the cables and
handling and assembly of the ergometer
connectors must be checked for signs of damage.
Damaged parts must be replaced immediately, – must be familiar with and observe the safety
before use. rules and regulations for the operation of this
type of equipment
Equipment Failure —
– The special shielded cables supplied by GE – must be informed about any other pertinent
Healthcare must be used to connect the device to rules and regulations (e.g. safety instructions)
other pieces of equipment. – must be informed about the potential hazards
arising from the use of this type of equipment.
Warning Note
Patient Hazard — – The blood pressure module and the approved
– Do not use a cuff to take blood pressure accessories are defibrillation-proof and can
measurements on patients suffering from sickle remain attached to the patient during
cell anemia or where skin lesions are likely to defibrillation.
occur. – If the cuff pressure exceeds the maximum value
– The cuff may cause hematomas in patients with of 300 mmHg during inflation, the inflation
severe blood coagulation disease. In these procedure will be aborted and the cuff will
instances, the user must make a decision for or deflate. As a redundant safety precaution, the
against automatic blood pressure cuff is immediately deflated when the cuff
measurements. pressure exceeds 310 mmHg.
You can check the proper functioning of this
safety precaution by abruptly bending your arm
while the cuff is being inflated, causing a brief
Caution overpressure in the cuff. The cuff must
Compromised Measuring Accuracy — immediately deflate.
– Arrhythmias occurring frequently during a – Measurements that do not yield a valid
measurement may compromise the accuracy of measurement will not be repeated during the
the measuring values. In certain cases, a valid exercise test.
measurement will not be possible. – If the inflation phase takes longer than 40
– Electromagnetic fields may also impair the seconds or if an adequate pressure does not
measuring accuracy. build up in the cuff within a reasonable period
of time, the measurement will be aborted and
the cuff will deflate.
– If a valid measurement cannot be completed
within 120 seconds, the measurement will be
aborted and the cuff will deflate.
– If the cuff pressure remains constant for some
time, the measurement will also be aborted and
the cuff will deflate.
Setup
Set up the ergometer where you intend to use it and adjust
the two leg levellers 12 for ergometer stability (Figure 2-
1). Protect the ergometer and, in particular, the saddle
cover from direct sunlight
Figure 3-1 Leg levellers
Turning the Control Terminal
The control terminal can be installed with the display
facing the patient or the operator (Figure 3-2). It is
recommended to install the control terminal with the
display toward the operator and the speed indication
toward the patient (See “Speed Readout for Patient”).
Mains Connection
Stand in front of the ergometer and grasp both handles
Caution
of the handlebar.
Equipment Damage — Before connecting the
ergometer to the power line, make sure that the Tip the ergometer forward all the way and carefully
voltage rating of your local power line matches the lay it on its handlebar (Figure 3-3).
value indicated on the device nameplate. The Now you can access the connector panel on the underside
nameplate is located on the back of the ergometer, of the ergometer.
at the bottom.
Connect the power cord to socket a (Figure 3-4).
Mount the strain relief b as shown in Figure 3-4.
Connection to GE Electrocardiographs
electrocardiograph connection cable connector
on eBike
CardioSys 22336203 port 1
CardioSoft *) 22336203 port 1
2017911-131 USB
CASE 22336203 port 1
CardioSmart ST**) 2006795-002 port 1
MAC 1200 ST 2006795-002 port 1
Figure 3-3 Mounting position MAC 1600 ST 2006795-002 port 1
MAC 5000 ST/ 2008110-001 analog
5500 ST (connection 2008114-001 port 3
only via COM
module
(Figure 3-6)***
Transport
Disconnect all connected cables.
Stand in front of the ergometer and grasp both handles
of the handlebar.
Tip the ergometer forward until it rests only on the
castors.
Wait until you have achieved an equilibrium before
moving the ergometer.
When you have reached the new location, lower the
ergometer very carefully. Otherwise, severe damage
may occur.
Caution
Figure 3-7 Transport Equipment Damage — Avoid strong vibrations of
the ergometer during transport.
4 Preparation
Ergometry
Manual The ergometer runs an automatic selftest (Figure 4-1).
Settings Then the main menu appears (Figure 4-2). When the
main menu appears, the ergometer is ready for operation.
Select Note
The device may be configured to default to one of
Figure 4-2 Main menu
the operating modes. In this case you will see the
initial screen of the selected operating mode instead
of the main menu. You can display the main menu
with the button (see section 7
"Configuration").
a NIBP
b
If an error code (e.g. “Internal fault NIBP defect
c
0010”) appears after the selftest, please refer to
“Error Messages” on page 33 for troubleshooting
d eBike e
information.
It is important to switch on the eBike before
applying the blood pressure cuff to the patient.
Figure 4-3 Control terminal buttons In addition to the three softkeys c, d, e (Figure 4-3) the
a Starts and stops a blood pressure measurement control terminal also accommodates buttons NIBP
a and
b Brings up the main menu or returns to the
b.
previous menu level
c, d, e Softkeys activating the function described above With button NIBP
a blood pressure measurement can be
initiated before or during the exercise test. The same
button can be used to stop a blood pressure measurement.
7
should be a 10° angle between the axis formed by the
3
~10˚ upper body and the thigh when the pedal is in its lower
15
position.
8
5 Display 8 indicates the saddle height as a numeric value
(eBike comfort only).
Watt
0 min
0
/min
Adjusting the saddle height of the eBike comfort
/
Ask the patient to sit on the saddle.
mmHg /min
In the main menu, select the desired mode of
Saddle operation, e.g. PC Mode (Figure 4-8).
Adjust the saddle height with the right and left
f g softkeys f, g.
Figure 4-8 PC Mode menu
a Blood pressure readings (systolic, diastolic) or cuff Note
pressure during measurement The patient may remain seated on the saddle while
b Current load the height is being adjusted. In PC Mode, the saddle
c Duration of exercise test (in "Manual" and
height adjustment buttons appear automatically
"Ergometry" modes only)
d Pulse rate (configurable) when the electrocardiograph start signal is
e Speed (revolutions/minute) detected.
f Saddle up
g Saddle down
Caution
2017911-010 1,3 m
2017911-011 2,0 m
23
-33 cm
Erroneous Measurements —
14 cm
LATEX If the cuffs are not the correct size, the measured
values may be incorrect.
Always use the cuff size suitable for the patient's arm.
The maximum arm circumference is indicated on each
cuff.
Figure 4-9 Correct cuff size
When closing the Velcro strap, check that the index line
(a, Figure 4-10) at the metal clasp is inside the INDEX
range (b) identified by the arrow, and not outside.
a b
2017911-010 1,3 m
2017911-011 2,0 m
-33 cm 2017911-010 1,3 m
23 2017911-011 2,0 m
INDEX -33 cm
23
14 cm
LATEX
LATEX
Slip the microphone deep into the pocket and be sure not
to twist the microphone cable.
Cuff Application
The accurate placement of the microphone is the primary
condition for reliable pressure measurements during
exercise tests.
Warning
Patient Hazard — If, by mistake, excessive pressure
builds up in the cuff, either remove the cuff
immediately from the patient’s arm or disconnect
Figure 4-18 Cuff tubing the cuff tubing at the control terminal. The same
applies when the cuff does not deflate correctly.
Watt
0 min
0
/min
page 27).
Now the ergometer waits for the start signal from the
/ electrocardiograph.
mmHg /min When you have finished preparing the patient, initiate
the exercise test on the electrocardiograph.
Saddle
The ergometer will emit three beeps and the speed
display will blink three times to indicate the
f g beginning of the test. Now the patient should start
pedaling.
Figure 5-3 PC mode initial menu
a Blood pressure readings (systolic, diastolic) or cuff
pressure during measurement
b Current load
c Duration of exercise test (in "Manual" and
"Ergometry" modes only)
d Pulse rate (configurable)
e Speed (revolutions/minute)
f Saddle up
g Saddle down
Note
To quit the PC mode, turn the ergometer off and on
again.
If the patient’s pedal speed drops below 30 revs/
min, the load indication (b, Figure 5-3) starts
blinking. Should this condition persist for more
than 20 seconds, the displayed watt reading will
change to "0".
The load indication will also blink when patients do
not pedal within the range corresponding to the
family of characteristics shown in Figure 10-1,
because in this case their speed is below the
required load.
25 0
Watt min
0
/min
When you have finished preparing the patient, Start
the test with the softkey in the middle.
The ergometer will emit three beeps and the speed
display will blink three times to indicate the beginning of
/ the test. The exercise test clock (c, Figure 5-5) will start
mmHg /min at the same time. When, after approximately 15 seconds,
Start the basic load appears at b, the patient should begin
pedaling.
Figure 5-5 Ergometry mode
a Blood pressure readings (systolic, diastolic)
b Basic load (appears 15 seconds after activation of
Start) Note
c Duration of exercise test The protocol details are listed in the Appendix.
d Pulse rate (configurable)
e Speed (revolutions/minute) Refer to section “Configuration” on page 27 for
protocol editing instructions.
25 1
Watt min
58
/min
the load with the right and left softkeys (increments
configurable between 1 and 25 W) or advance to the
recovery phase with the softkey in the middle.
/
mmHg /min
+5W Recovery -5W
Figure 5-6 Ergometry mode Ending the Test
The test ends automatically when all stages, including the
recovery phase, have been completed. It is also possible
25 16 58
Watt min /min
to end the test manually at any time. To abort the test,
press the softkey in the middle to advance to the recovery
phase (Figure 5-7).
5.4 Manual
Selecting and Starting the Operating Mode
In the manual operating mode, you have to control the
ergometer manually, i.e., basic load, load increase, etc.
25
Watt min
0 0
/min
have to be entered manually. Blood pressure
measurements, too, have to be started with the NIBP
button.
25 15 58
Watt min /min
with the
page 27).
button (see “Configuration” on
Adjust the basic load with the right and left softkeys.
/ The increment is configurable between 1 and 25 W.
mmHg /min When you have finished preparing the patient, Start
+5W End -5W the test with the softkey in the middle.
The ergometer will emit three beeps and the speed
Figure 5-9 Manual mode
display will blink three times to indicate the beginning of
the test. The exercise test clock will start at the same time.
Now the patient should begin pedaling.
7 Configuration
Settings Note
Default Mode The device may be configured to default to one of
Protocols the operating modes. In this case you will see the
Contrast initial screen of the selected operating mode instead
Load Change of the main menu. You can display the main menu
Language with the button.
Select
Note
Figure 7-2 Configuration menu – Upon delivery, the ergometer is configured for
communication with digitally controlled
electrocardiographs.
– When you operate the ergometer in conjunction
with the MAC 5000 ST/5500 ST (analog
control), select "PC Mode" in the "Default
Mode" menu and "Analog/Digital" in the
"EKG Type" menu.
Default Mode
Default Mode In this menu you choose the default mode activated when
the ergometer is turned on.
Menu
Using the arrow buttons, choose an operating mode,
PC Mode
then Select the option to exit the menu.
Ergometry
Manual Quit the configuration mode with .
Select
Protocol
6. Note
Basic Load 25W Chapter 12 in the Appendix provides an overview of
Stage Time 2 min the exercise test protocols stored in the ergometer.
Load Stage 25 W
Select
Contrast
The contrast can be adjusted from 0 to 100% in steps of
5%.
Contrast
50 %
Select
+/- 1 Watt
+/- 5 Watt
+/- 10 Watt
+/- 25 Watt
Select
Language
Choose the language for the displayed texts.
Language
Deutsch
English
Français
Español
Italiano
Select
Beep
On this screen you enable and disable the QRS beep for
Beep blood pressure measurements.
On
Off
Software Version
Select to display the installed software version.
Select
Date
22.10.2003
Time
12:44:49
Select
Figure 7-12 Adjusting the date
EKG Type
In this menu you toggle between digital and analog
control of the ergometer. Access to this menu is protected
EKG Type with a password.
EKG Type
Analog / Digital
Digital
Select
RPM
In this menu you choose the limit values for an "RPM
RPM alert", when the LED for high or low speed (RPM) lights
up on the speed display.
Min. 0 . . 70:
54 /min Use the arrow buttons to choose the minimum or
maximum RPM parameter and confirm with Select.
Max. 50 . . 130:
65 /min Change the RPM limit value with the arrow buttons
and confirm with Select.
Select The minimum value can be adjusted between 0 and
70 RPM, the maximum value between 50 and 130. With
Figure 7-15 Adjusting RPM limit values the limit values shown in Figure 7-15, the LED for the
low speed lights up if the patient pedals below 54 RPM,
the LED for the high speed lights up above 65 RPM.
451 - 550 88 - 95
Figure 7-16
Pulse Display
With the menu item "Pulse Display" you enable and
disable indication of the pulse rate.
8 Error Messages
Internal Fault NIBP defect 0010 The pressure sensor for blood Turn off the ergometer with the power switch,
pressure measurements could not wait approx. 5 seconds, then switch it on again.
be initialized. Be sure to switch on Should the error message persist, have the
the eBike before applying the cuff ergometer checked by trained service
to the patient. technicians before using it again. This is a
general recommendation and applies whenever
an error message is displayed.
Cleaning Cables
9.3 Disposal of the Product
Disconnect cables from the device before cleaning
them. Do not dispose the product described in this
Operator Manual as unsorted municipal
Use a cloth moistened with soap water to wipe the
waste. It must be collected separately.
cables clean. Do not immerse the cables in liquid.
Please contact an authorized representative
of the manufacturer for information con-
cerning the decommissioning of your
equipment.
Consult operating instructions.
– 160 kg max.
Saddle Height Adjustment
Measuring Principle
40 45 45 49 51 49
50 47 47 48 49 56 58 58
60 48 48 49 50 53 54 59 61
70 50 50 51 52 53 54 55 57
80 52 52 53 54 57 57 59 60
90 53 52 52 52 53 56 57 59
100 53 52 52 52 53 54 57 59
110 54 53 53 53 56 56 57 58
120 55 55 54 54 55 56 58 58
11 Order Information
2017911-105 Bicycle ergometer eBike comfort with 2017911-019 Child's pedal, adjustable
control terminal PC, incl. power cord, 22336203 Connection cable for CardioSys,
operator's manual, without connection CardioSoft, CASE, CASE 8000,
cable to electrocardiograph MicroLab, length 5 m
2017911-107 Bicycle ergometer eBike comfort with 2006795-002 Connection cable for CardioSmart ST,
control terminal PC and blood pressure MAC 1200 ST, MAC 1600 ST, length
module, incl. blood pressure cuff 5m
(standard), power cord, operator's
2008110-001 Analog connection cable for
manual, without connection cable to
MAC 5000 ST/5500 ST, length 4.6 m
electrocardiograph
2008114-001 Digital connection cable for MAC 5000
Add-ons ST/5500 ST, length 4.6 m
12 Appendix
1. WHO stage 25 2 25 60 25 99
2. BAL stage 50 3 50 60 25 99
3. Hollmann stage 30 3 40 60 25 99
4. STD.France stage 30 3 30 60 25 99
5. Standard stage 20 1 25 60 25 99
6. stage 25 2 25 60 251) 99
7. stage 25 2 25 60 251) 99
8. stage 25 2 25 60 251) 99
9. stage 25 2 25 60 251) 99
Warning
The equipment or system should not be used
adjacent to, or stacked with, other equipment. If
adjacent or stacked use is necessary, the equipment
or system should be tested to verify normal
operation in the configuration in which it is being
used.
The eBike PC is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that the eBike PC is used in such an environment
RF emissions to Group 1 The eBike PC uses RF energy only for its internal
EN 55011 function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The eBike PC is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that the eBike PC is used in such an environment
Electrical fast transient/ ± 2 kV for power supply ± 2 kV Mains power should be that of a typical
burst to EN 61000-4-4 lines commercial or hospital environment.
passed
± 1 kV for input/output lines
Voltage dips, short < 5 % UT < 5 % UT Mains power should be that of a typical
interruptions and voltage (> 95 % dip in UT) for commercial or hospital environment. If
variations on power 0.5 cycles the user of the eBike PC requires
supply input lines to EN continued operation during power mains
40 % UT 40 % UT
61000-4-11 interruptions, it is recommended that the
(60 % dip in UT) for 5 cycles
eBike PC be powered from an
70 % UT uninterruptible power supply or a battery.
(30 % dip in UT) for 70 % UT
25 cycles
< 5 % UT
(> 95 % dip in UT) for < 5 % UT
5 seconds
Power frequency (50/ 3 A/m passed Power frequency magnetic fields should
60 Hz) magnetic field to be at levels characteristics of a typical
EN 61000-4-8 location in a typical commercial or
hospital environment. The eBike PC has
no components susceptible to magnetic
fields.
The eBike PC is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that the eBike PC is used in such an environment
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a ) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the equipment is used exceeds the applicable RF
compliance level above, the eBike PC should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the eBike PC.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The eBike PC is intended for use in the electromagnetic environment specified below on which radiated RF
disturbances are controlled. The customer or the user of the eBike PC can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the eBike PC as recommended below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Warning
The use of accessories, transducers and cables
other than those specified may result in increased
emissions or decreased immunity performance of
the equipment or system.
Note
Any supplied accessories that do not affect EMC
compliance are not included
A L
Accessories 38 Language Selection 29
Adjusting Saddle and Handlebar 16 Language, Select 15
Applying the Blood-Pressure Cuff 17 Load Increments, Selection 29
B M
Biocompatibility 8 Mains Connection 12
Blood pressure cuff, cleaning and disinfection 34 Maintenance 34
Manual Mode 24
C Manual Purpose 6
MDD 4
Cables, Cleaning 34
MENU button 14
Caution 4
Microphone position 17
CE marking 4
Checks before each use 34
Configuration 27 N
Contrast adjustment 29 NIBP button 14
Control terminal PCplus 7 Noise level 36
Cuff application 18
Cuff size 17 P
PC Mode 20
D Power switch 14
Danger 4, 9 Preparing the Patient 16
Date, Adjust 30
Dimensions 35 Q
Disposal of the Product 34
QRS beep, on/off 30
E R
eBike basic 7
eBike comfort 7 Revision History 6
EKG Type 32
Electrocardiographs, connection of 12 S
Electromagnetic Compatibility (EMC) 40 Saddle height 16
Environment 35 Safety Information 9
Equipment symbols 8 Selftest 14
Error messages 33 Setup and Mains Connection 11
Exercise Test 22 Softkeys 15
Exercise Test Protocols, Internal 39 Software version, view 30
Exercise tests, configuration 28
Explosion hazard 9
T
Taking the Blood Pressure 25
F Technical safety inspections 34
Family of characteristics 36 Technical Specifications 35
Time, Adjust 30
G Transport 13
General information 4
W
I Warning 4
Weight 35
Intended Audience 6
Intended Use 7
GE Medical Systems
Information Technologies, Inc. GE Medical Systems Asia Headquarters
8200 West Tower Avenue Information Technologies, GmbH GE China Co., Ltd.
Milwaukee, WI 53223 USA Munzinger Straße 5 No1 Huatuo Road,
Tel: +1 414 355 5000 79111 Freiburg Zhangjiang Hi-Tech Park Pudong,
1 800 558 7044 (US only) GERMANY Shanghai, P.R.China 201203
1 800 668 0732 (Canada only) Tel: +49 761 4543 - 0 Tel: +86 21 38777888
Fax: +1 414 355 3790 Fax: +49 761 4543 - 233 Fax: +86 21 38777402
www.gehealthcare.com