You are on page 1of 106

CDS 08

Central Delivery System


Technical Manual
Software 0.996 and higher

Revision: 1/01.04

Part no. 632 502 1

0123
TECHNICAL MANUAL
CENTRAL DELIVERY SYSTEM CDS 08
The Technical Manual contains all necessary information on the use of
the concentrate delivery system CDS 08. Before operating the device,
the operator must have been instructed by the manufacturer on how to
use the system and must be thoroughly familiar with the contents of the
Operating Instructions.

The device may only be operated by persons who have been instructed
on the proper operation and handling of the unit.

The Operating Instructions must be carefully studied before attempting


to operate the system.

Explanation of symbols used in the Operating Instructions:

Note
Informs the operator that if the steps are not followed as described,
a specific function will be executed incorrectly or will not be executed
at all, or will not produce the desired effect.

Caution
Advises the operator against certain procedures or actions that could
cause damage to the equipment or may have adverse effects on
operators and patients.

The parameters entered must be verified by the operator, i.e. the


operator must check that the values entered are correct. If the
verification reveals a deviation between the desired parameters and the
parameters displayed, the setting must be corrected before activating
the function.

The actual values displayed must be compared with the desired values
specified!

The system reflects the latest state of technology and complies with the
requirements of EN 60601-1 (IEC 601).

The system may only be operated in rooms reserved for medical use
according to VDE 0170.

It is classified as class IIb (MDD) equipment.

Assembly, extensions, adjustments, modifications or repairs may only


be carried out by the manufacturer or persons authorized by him.

Fresenius Medical Care CDS 08 TM 1/01.04 0-1


Adresses

Please address any inquiries to:

Manufacturer Fresenius Medical Care AG


61346 Bad Homburg
Germany
Phone: +49 6172–609–0
www.fmc-ag.com

Service Fresenius Medical Care


Central Europe Deutschland GmbH
Geschäftsbereich Zentraleuropa
Kundendienst / Servicecenter
Steinmühlstraße 24b
61352 Bad Homburg
Germany
Phone: +49 6172–609–7100
Fax: +49 6172–609–7102
E-mail: ServicecenterD@fmc-ag.com

International Fresenius Medical Care


service Deutschland GmbH
Service Support International
Hafenstraße 9
97424 Schweinfurt
Germany
Phone: +49 9721–678–333 (Hotline)
Fax: +49 9721–678–130

Local service

0-2 Fresenius Medical Care CDS 08 TM 1/01.04


HOW TO USE THE TECHNICAL MANUAL
Purpose This manual is intended for:
– first studies (to acquire a basic knowledge)
– reference purposes (for start-up, maintenance, repair)

Qualification This manual is intended for service technicians


– who are familiar with the current Operating Instructions
– who have the necessary background experience in mechanics,
electrical and medical engineering
– who have access to the necessary auxiliary and measuring
equipment
– who have been authorized by the manufacturer to perform
maintenance and repair

Limitations The study of this manual does not represent an alternative to the
training courses offered by the manufacturer.

Manual changes Manual changes will be released as new editions, supplements or


product information.
In general, this manual is subject to to change without notice.

Representation Unless specified otherwise, potentials indicated in circuit diagrams and


calibration instructions are related to the pertinent ground.
For example:
24 means ground for the 24 V voltage.

Note When repairing the CDS 08 and replacing spare parts, observe the
protective measures (e.g. EN 100 015-1) to be taken to avoid
electrostatic discharge.

Technical Specifications For the technical data of the Concentrate Delivery System CDS 08:
(see chapter 1 on page 1-1)

Fresenius Medical Care CDS 08 TM 1/01.04 0-3


0-4 Fresenius Medical Care CDS 08 TM 1/01.04
TABLE OF CONTENTS

1 General Description of the System


1.1 Intended Use .............................................................................................................................. 1-1
1.1.1 Field of Application ...................................................................................................................... 1-1
1.1.2 General Warnings........................................................................................................................ 1-1
1.1.3 Patient Safety .............................................................................................................................. 1-2
1.1.4 Operator Safety ........................................................................................................................... 1-2
1.1.5 Protection of Buildings................................................................................................................. 1-2
1.1.6 Service Procedures ..................................................................................................................... 1-3
1.2 Design and Function of the Individual Components of the System ..................................... 1-4
1.2.1 CDS 08 Front View...................................................................................................................... 1-4
1.2.2 CDS 08 Rear View ...................................................................................................................... 1-5
1.2.3 CDS 08 Hydraulic Unit................................................................................................................. 1-6
1.3 System Description ................................................................................................................... 1-7
1.3.1 General Description..................................................................................................................... 1-7
1.3.2 Functional Description ................................................................................................................. 1-7
1.3.3 CDS 08 Flow Diagram................................................................................................................. 1-8
1.3.4 Valves.......................................................................................................................................... 1-9
1.3.5 Sensors ....................................................................................................................................... 1-9
1.4 Installation and Start-Up ......................................................................................................... 1-10
1.5 Accessories ............................................................................................................................. 1-11
1.5.1 Remote Control Indicator........................................................................................................... 1-12
1.5.2 External Tank Sensors and Traffic Light ................................................................................... 1-12
1.5.3 Ultrafilter Unit............................................................................................................................. 1-12
1.6 Technical Specifications......................................................................................................... 1-13
1.6.1 Machine Data ............................................................................................................................ 1-13
1.6.2 Operating Conditions................................................................................................................. 1-14
1.6.3 Disposal..................................................................................................................................... 1-14
1.6.4 Power Supply ............................................................................................................................ 1-14
1.7 Abbreviations and Symbols ................................................................................................... 1-15
1.8 Storage and Transport ............................................................................................................ 1-16
1.9 CE Declaration of Conformity ................................................................................................ 1-17
1.10 Bibliography............................................................................................................................. 1-20

2 Operating Modes and Functions


2.1 Controls...................................................................................................................................... 2-1
2.2 Program Description ................................................................................................................. 2-2
2.3 Operating Programs.................................................................................................................. 2-3
2.3.1 STANDBY ................................................................................................................................... 2-3
2.3.2 SUPPLY ...................................................................................................................................... 2-4
2.3.3 RINSE.......................................................................................................................................... 2-4
2.3.4 FILL ............................................................................................................................................. 2-4

Fresenius Medical Care CDS 08 TM 1/01.04 0-5


2.3.5 DISINFECTION ........................................................................................................................... 2-4
2.3.6 PRE-VENTILATION .................................................................................................................... 2-5

3 Disinfection and Preservation

4 Error Messages
4.1 Tank 1 empty !............................................................................................................................ 4-1
4.2 Tank 2 empty !............................................................................................................................ 4-1
4.3 Tank empty !............................................................................................................................... 4-1
4.4 Both Tanks empty ! ................................................................................................................... 4-2
4.5 RO-Tank empty ! ........................................................................................................................ 4-2
4.6 Disinf.-Tank empty ! .................................................................................................................. 4-2
4.7 Wrong concentrate !.................................................................................................................. 4-2
4.8 Pressure high !........................................................................................................................... 4-3
4.9 Only in STANDBY ! .................................................................................................................... 4-3
4.10 Only after RINSE !...................................................................................................................... 4-3
4.11 Only by Service !........................................................................................................................ 4-4

5 Technical Safety Checks and Maintenance Procedures


5.1 General Notes ............................................................................................................................ 5-1
5.2 TSC Report and Maintenance Procedures .............................................................................. 5-2
5.3 Electrical Test ............................................................................................................................ 5-6
5.3.1 CDS 08 Protective Conductor Resistance.................................................................................. 5-6

6 Adjustments and Calibration


6.1 Settings Menu ............................................................................................................................ 6-1
6.1.1 Concentrate ................................................................................................................................. 6-3
6.1.2 Tank............................................................................................................................................. 6-3
6.1.3 Pressure (reg.)............................................................................................................................. 6-3
6.1.4 Pressure (hot) .............................................................................................................................. 6-3
6.1.5 Ventilation Time ........................................................................................................................... 6-3
6.1.6 Drain Time ................................................................................................................................... 6-4
6.1.7 Disinfection Time ......................................................................................................................... 6-4
6.1.8 Tube Length................................................................................................................................. 6-4
6.2 System Parameters ................................................................................................................... 6-5
6.2.1 Tank Swapping ............................................................................................................................ 6-6
6.2.2 Tank-Empty Detector................................................................................................................... 6-7
6.2.3 Supply Mode................................................................................................................................ 6-7

0-6 Fresenius Medical Care CDS 08 TM 1/01.04


6.2.4 Pump Speed................................................................................................................................ 6-7
6.2.5 Language..................................................................................................................................... 6-7
6.2.6 Brightness.................................................................................................................................... 6-7
6.2.7 KI / KP ......................................................................................................................................... 6-8
6.2.8 Tank SWCondBic – Bicarbonate Conductivity Limit.................................................................... 6-8
6.2.9 Tank SWCondACID – Acid Concentrate Conductivity Limit........................................................ 6-8
6.2.10 External tank sensors .................................................................................................................. 6-8
6.2.11 Masch.hour.................................................................................................................................. 6-8
6.2.12 Circulation Pause ........................................................................................................................ 6-8
6.2.13 Circulation Time........................................................................................................................... 6-8
6.2.14 Ventilation Standby...................................................................................................................... 6-9
6.3 Calibrations.............................................................................................................................. 6-10
6.3.1 Pressure Transducer Zero-Point Calibration ............................................................................. 6-10
6.3.2 External tank sensors ................................................................................................................ 6-10
6.3.3 Remote control indicator............................................................................................................ 6-11
6.4 Installation and Start-Up ......................................................................................................... 6-12
6.4.1 Preparation ................................................................................................................................ 6-12
6.4.2 Filling and Deaerating the Tank Connectors ............................................................................. 6-12
6.4.3 Filling and Deaerating the Ring Main ........................................................................................ 6-12
6.4.4 Setting the Ring Main Pressure................................................................................................. 6-12
6.4.5 Checking the Tank Swapping Unit ............................................................................................ 6-13
6.4.6 Start-Up Report ......................................................................................................................... 6-14

7 Circuit Diagrams and Circuit Descriptions


7.1 CDS 08 Wiring Diagram ............................................................................................................ 7-1
7.2 EMA Power Supply Unit Rev 2.1 .............................................................................................. 7-3
7.2.1 Circuit Description ....................................................................................................................... 7-3
7.2.2 Circuit Diagram and Component Layout of EMA Power Supply Unit Rev 2.1............................. 7-4
7.3 EMA Power Supply Unit Rev. 3 1/1 .......................................................................................... 7-7
7.3.1 Circuit Description ....................................................................................................................... 7-7
7.3.2 Circuit Diagram and Component Layout of EMA Power Supply Unit Rev. 3 1/1......................... 7-8
7.4 Control P.C.B. LP 795.............................................................................................................. 7-11
7.4.1 Circuit Description ..................................................................................................................... 7-11
7.4.2 Circuit Diagram and Component Layout of Control P.C.B. LP 795 ........................................... 7-14
7.5 Keypad P.C.B. LP 796 ............................................................................................................. 7-21
7.5.1 Circuit Description ..................................................................................................................... 7-21
7.5.2 Circuit Diagram and Component Layout of Keypad P.C.B. LP 796 .......................................... 7-22
7.6 P.C.B. LP 797 ........................................................................................................................... 7-24
7.6.1 Circuit Diagram and Component Layout of P.C.B. LP 797........................................................ 7-24
7.7 Remote Control Indicator ....................................................................................................... 7-27
7.7.1 Circuit Description ..................................................................................................................... 7-27
7.7.2 Circuit Diagram and Component Layout of Remote Control Indicator ...................................... 7-30
7.8 Status Indicator Control.......................................................................................................... 7-33
7.8.1 Circuit Description ..................................................................................................................... 7-33
7.8.2 Circuit Diagram and Component Layout of Status Indicator Control......................................... 7-34
7.9 External Tank Sensor.............................................................................................................. 7-37
7.9.1 Circuit Description ..................................................................................................................... 7-37
7.9.2 Circuit Diagram and Component Layout of External Tank Sensor............................................ 7-40

Fresenius Medical Care CDS 08 TM 1/01.04 0-7


0-8 Fresenius Medical Care CDS 08 TM 1/01.04
Chapter 1: General Description of the System

1 GENERAL DESCRIPTION OF THE SYSTEM

1.1 Intended Use

1.1.1 Field of Application

The CDS 08 (Central Delivery System) is a system intended for the delivery of bicarbonate and acid
concentrates from tanks. The concentrates are delivered from the tanks to the individual media supply
centers (MSC) via a ring main. The dialysis systems use the concentrates to prepare the dialysate.

Two tanks, which are monitored by an internal or external tank-empty detector and activated accordingly
by the CDS 08, can be connected to each type of concentrate.

The CDS 08 is operated in the immediate vicinity of the concentrate containers outside the treatment
rooms.

1.1.2 General Warnings

The CDS 08 may only be operated under the specified operating conditions.

The control unit must be protected against humidity (splash water, condensation water, etc.) and
moisture.

In the event of a defect of the control unit, the type of problem (effect of the malfunction) should be
recorded before dismantling the system. A system that has been dismantled can only be repaired if a
detailed description of the problem is available.

The total output of the system may not be exceeded.

The device may only be operated by persons who have been instructed on the proper operation and
handling of the unit.

Start-up of the system must be performed by a trained technician.

Caution
Disconnect the power plug before opening the system.

Caution
Do not confuse bicarbonate concentrate with acid concentrate.
Strong chemical reactions would lead to precipitation in the tubing,
which, in turn, would become obstructed.

Fresenius Medical Care CDS 08 OP 1/01.04 1-1


Chapter 1: General Description of the System

1.1.3 Patient Safety

When connecting, reconnecting or replacing the containers always disinfect the connectors.

Avoid contamination of the connectors caused by contact with skin or non-sterile objects.

The CDS 08 must be disinfected by authorized personnel in accordance with the disinfection
specifications.

Caution
Before the start-up of the CDS 08 take appropriate measures to
verify that there is no disinfectant in the ring main.

1.1.4 Operator Safety

When adding the disinfectant, care must be taken not to spill any disinfectant concentrate. To prevent
burns these fluids must always be handled with care:
– Use rubber gloves to avoid contact of the disinfectant with the skin.
– Wear goggles when working with acidic substances!

Note
The system may only be disinfected after consulting the
manufacturer of the unit or by personnel authorized by the
manufacturer.

Caution
Observe the safety precautions of the disinfectant used!
In the event of contact with acid:
– Eye:
Immediately flush with gently flowing water for 15 minutes.
– Skin:
Use soap under flowing water for neutralisation.
– Ingestion:
Do not induce vomiting, but have victim drink plenty of still water.

1.1.5 Protection of Buildings

Improper handling of the CDS 08 may cause leakage. Uncontrolled leakage of fluid may result in damage
to equipment, installations and buildings.
– Tubings must be protected against possible mechanical damage.
– Regular visual inspection of all tubings, connectors and piping containing fluid and of the CDS 08 is
required to avoid damage caused by leaking fluid.

1-2 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.1.6 Service Procedures

The applicable ESD precautions (EN 1000 1.5/1) must be observed in the event of all technical
interventions in the system.

Fresenius Medical Care CDS 08 OP 1/01.04 1-3


Chapter 1: General Description of the System

1.2 Design and Function of the Individual Components of the


System

1.2.1 CDS 08 Front View

2
3

Legend:

1 Ultrafilter unit
2 Status indicators
3 Display
4 Operating keys

1-4 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.2.2 CDS 08 Rear View

1 13
2 12

3 11
4 10

5 9
8
6
7

Legend:

1 ON/OFF switch
2 RS 232 interface
3 Port for remote control indicator
4 Tank level port (tank 1 + 2)
5 Nurse call port
6 Type label
7 Power cable
8 Tank 2 port
9 Tank 1 port
10 Deaeration port
11 Drain port
12 Ring main – inlet port
13 Ring main – outlet port

Fresenius Medical Care CDS 08 OP 1/01.04 1-5


Chapter 1: General Description of the System

1.2.3 CDS 08 Hydraulic Unit

20 1
19
2
18
17 3

16
15
4
14
13 5
6

12

11 7

10 8

Legend:

1 P – Pressure sensor 11 P1 – Pump 1


2 V3 – Deaeration valve 12 R1 – Pump 1 check valve
3 Motor for pump 1 13 V1 – Concentrate inlet valve for tank 1
4 Motor for pump 2 14 V2 – Concentrate inlet valve for tank 2
5 V5 – Circulation valve 15 Tank 1 port
6 V4 – Drain valve 16 Tank 2 port
7 CD/T – Conductivity/temperature sensor 17 Drain port
8 P.C.B. LP 797 18 Deaeration port
9 P2 – Pump 2 19 Ring main - outlet port
10 R2 – Pump 2 check valve 20 Ring main - inlet port

1-6 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.3 System Description

1.3.1 General Description

The central delivery system CDS 08 is a system for the delivery of dialysis concentrates and is intended
for medical use only.

The CDS 08 is provided with two tank ports each of which is controlled by a solenoid valve. Both the
temperature and the conductivity of the fluid are determined by a sensor. Two parallel gear pumps deliver
the fluid to the ring inlet and ultrafilter unit via a pressure sensor. The ultrafilter unit prevents particles,
bacteria, endotoxins and air from entering the ring main.

The microprocessor control permits fully automatic monitoring and control of the central delivery system
CDS 08.

All controls and status indicators required to operate the system are located on the front of the CDS 08.
The Power ON/OFF switch is located on the rear of the CDS 08.

1.3.2 Functional Description

The dialysis concentrate is delivered from the connected container to the conductivity/temperature sensor
CD/T via the concentrate inlet valve V1 or V2. Two parallel gear pumps deliver the fluid to the ultrafilter
unit via the pressure sensor P.

The dialysis concentrate flows through two parallel ultrafilters. The ultrafilters prevent germs and bacteria
from entering the ring main and purge air from the system.

After having passed the ultrafilter unit, the dialysis concentrate is pumped via the ring main to the
connectors of the media supply centers (MSC).

The pressure holding valve R3 of the ring main is located in the ultrafilter unit at the end of the system. It
is adjusted manually during the installation of each system. The dialysis concentrate returns from the
pressure holding valve R3 of the ring main via the circulation valve V5 and the conductivity/temperature
sensor CD/T to the feed line of the pumps.

Opening the deaeration valve V3 purges all air from the system.

The drain valve V4 diverts the fluid directly to the drain.

Fresenius Medical Care CDS 08 OP 1/01.04 1-7


Chapter 1: General Description of the System

1.3.3 CDS 08 Flow Diagram

Legend:

V1 Concentrate inlet valve for tank 1


V2 Concentrate inlet valve for tank 2
V3 Deaeration valve
V4 Drain valve
V5 Circulation valve
CD/T Conductivity/temperature sensor
P1 Pump 1
P2 Pump 2
R1 Pump 1 check valve
R2 Pump 2 check valve
R3 Pressure holding valve of ring main
P Pressure sensor

1-8 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.3.4 Valves

V1 – Concentrate inlet The concentrate inlet valve V1 for tank 1 is open if tank 1 is the currently
valve for tank 1 selected tank.
It is possible to activate the tanks manually or by automatic tank
swapping.

V2 – Concentrate inlet The concentrate inlet valve V2 for tank 2 is open if tank 2 is the currently
valve for tank 2 selected tank.
It is possible to activate the tanks manually or by automatic tank
swapping.

V3 – Deaeration valve The deaeration valve V3 is open in the programs during the deaeration
phase and purges any air present from the system.

V4 – Drain valve The drain valve V4 is open during the drain phase and diverts fluid
directly to the drain.

V5 – Circulation valve The circulation valve V5 is open during the circulation phase and diverts
the dialysis concentrate in the return line directly to the feed line of the
pumps.

R1 – Pump 1 check valve The pump 1 check valve R1 prevents fluid from flowing back to pump 1.

R2 – Pump 2 check valve The pump 2 check valve R2 prevents fluid from flowing back to pump 2.

R3 – Pressure holding The adjustable pressure holding valve R3 of the ring main is used to
valve of ring main generate the correct back pressure in the concentrate ring main
required by the dialysis machines.

1.3.5 Sensors

CD/T – The conductivity/temperature sensor CD/T is used to measure the


Conductivity/temperature conductivity and temperature of the delivered dialysis concentrate. The
sensor message Wrong concentrate ! will be displayed if the conductivity of
the selected dialysis concentrate is below the minimum value.

P – Pressure sensor The pressure sensor P is used to measure the operating pressure of the
dialysis concentrate downstream of the pumps. Based on the
programmed theoretical values, the operating pressure of the system is
regulated by the delivery capacity of the gear pumps.
The pressure sensor P has no function in the deaeration program
phase.

Fresenius Medical Care CDS 08 OP 1/01.04 1-9


Chapter 1: General Description of the System

1.4 Installation and Start-Up

Start-up of the CDS 08 must only be performed by technicians authorized by Fresenius Medical Care.

1-10 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.5 Accessories

10 9 8 7 6

1 2 3 4 3 5

Legend:

1 Tank 1
2 Drain
3 External tank sensors
4 CDS 08
5 Tank 2
6 Traffic light
7 Remote control indicator
8 Ring main return line
9 Ring main feed line
10 Ultrafilter unit

Fresenius Medical Care CDS 08 OP 1/01.04 1-11


Chapter 1: General Description of the System

1.5.1 Remote Control Indicator

The remote control indicator is provided for the indication and detection of possible malfunctions from a
distance, the indication of operating conditions and for switching between the STANDBY and SUPPLY
programs when the system is connected.

The STANDBY and SUPPLY LEDs indicate the current program mode.The program key P is used to
switch between the programs. When pressing the key, the selected program will be displayed flashing and
will be activated after 2 seconds.

The indicators with the water/aggregate symbols alert the operator to a conductivity alarm and a
malfunction of the CDS 08.
– Green status indicator: Operation
– Yellow status indicator: Warning/information
– Red status indicator: Alarm

The audible alarm can be suppressed by pressing the P key for 1 minute. The remote control indicator is
connected to the sub D9 connector on the rear of the CDS 08.

1.5.2 External Tank Sensors and Traffic Light

The external tank sensors are provided for the detection of empty tanks. Depending on the measured
conductivity a signal is transmitted to the CDS 08 via the traffic light.

O External tank sensors

External tank sensors are used when the length of the tubes between the connected tanks and the
CDS 08 exceeds 5 meters. They detect an empty tank earlier than internal tank sensors, thus permitting
shorter swapping/rinsing times.

O Traffic light

The traffic light visually indicates the fluid level of two connected tanks filled with the same type of
concentrate. When the tank is empty a signal will be transmitted to the CDS 08. This signal will trigger
tank swapping or an alarm message if both tanks are empty.

The status indicators are controlled by means of a conductivity sensor.


– Red indicator: Connected tank is empty.
– Green indicator: Connected tank is ready for operation.

1.5.3 Ultrafilter Unit

The ultrafilter unit prevents particles, bacteria, endotoxins and air from entering the ring main. The system
is provided with either two or three filter units.

The integrated pressure holding valve R3 of the ring main is used to adjust the fluid pressure in the ring
return line.

1-12 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.6 Technical Specifications

1.6.1 Machine Data

Hydraulic data Pressure:


0.2 – 2.6 bar ± 0.2 bar (adjustable)
Conductivity:
0 – 300 mS/cm ± 5 mS/cm
Temperature:
0 – 40 °C ± 3 °C
Fluid temperature:
max. 95 °C

Drain connections Unobstructed flow to the drain

Optional Serial interface RS 232


Connection for external tank sensors
Connection for nurse call system
Connection for remote control unit

Dimensions Height:
300 mm
Width:
500 mm
Depth:
320 mm (+ 150 mm distance to wall)
Weight:
21 kg

Ultrafilter unit Height:


600 mm
Width:
306 mm
Depth:
220 mm
Weight:
3.5 kg
Housing colour:
White RAL 9001
Materials:
Special steel V4A, polyethylene (PE), polypropylene (PP), silicone (SI)

Fresenius Medical Care CDS 08 OP 1/01.04 1-13


Chapter 1: General Description of the System

1.6.2 Operating Conditions

Humidity max. 70 – 80 %

Ambient temperature 15 – 30 °C

1.6.3 Disposal

The CDS 08 and its components are to be disposed of at the end of their service life in accordance with
applicable national regulations.

1.6.4 Power Supply

Line voltage ~ 230 V ± 10 %


30 mA ground fault interrupter

Frequency 50 / 60 Hz

Operating current approx. 2.0 A

1-14 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.7 Abbreviations and Symbols

The abbreviations and symbols on the type label are described on the
example of a CDS 08 as follows:
Type label CDS 08:

Symbol (degree of
protection)

Type B equipment, e.g. suitable for external and internal use on the
patient, except for use on the heart.

Safety class I Equipment with protective conductor connection to prevent high contact
voltages.

IPX 1 Drip-water proof equipment

Year of manufacture The year of manufacture is included in the serial number printed on the
type label on the rear of the system.
e.g. serial number 841 V 0153
The first digit indicates the year of production, in the above example:
year of manufacture 1998.

CE mark

0123

The CE mark confirms that the basic requirements specified by the


Medical Device Directive (Guidelines for Medical Products 93/42/EEC)
are complied with and that the product has been subjected to a
conformity procedure.

Fresenius Medical Care CDS 08 OP 1/01.04 1-15


Chapter 1: General Description of the System

1.8 Storage and Transport

Relative humidity max. 70 – 80 %

Ambient temperature 0 – 50 °C
Protect from frost

Transport Keep upright, handle with care!

Storage Store upright!

1-16 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

1.9 CE Declaration of Conformity

Fresenius Medical Care CDS 08 OP 1/01.04 1-17


Chapter 1: General Description of the System

1-18 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 1: General Description of the System

Fresenius Medical Care CDS 08 OP 1/01.04 1-19


Chapter 1: General Description of the System

1.10 Bibliography

The following is a list of further literature available on the CDS 08:


– Installation Requirements
– Installation guidelines
– Inspection list
– Test report
– Documentation book
– Operating data record
– Machine card
– Check list
– Setup report
– Disinfection report
– Instruction report

The following is a list of further literature available on the dialysis concentrate ring main:
– Installation Requirements
– Installation guidelines
– Inspection list
– Acceptance report of the ring main leakage test

1-20 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 2: Operating Modes and Functions

2 OPERATING MODES AND FUNCTIONS

2.1 Controls

1 2 3 4 5 6 7

13 12 11 10 9

Legend:

1 Green status indicator: Operation


2 Yellow status indicator: Warning/Information
3 Red status indicator: Alarm
4 ACID concentrate LED
5 BIC concentrate LED
6 Tank 1 LED
7 Tank 2 LED
8 Display (1 line with 20 characters)
9  key (Enter - Confirm)
10 VW keys (Edit)
11 E key (Settings)
12 I key (Information menu)
13 P key (Program)

O I/O power switch

The I/O power switch, which is located on the rear of the system, is used to turn the control of the CDS 08
on and off.

If the power supply to the CDS 08 is interrupted for more than 60 seconds, the program must be restarted.

Fresenius Medical Care CDS 08 OP 1/01.04 2-1


Chapter 2: Operating Modes and Functions

2.2 Program Description

O Operation and Program Structure

P key (Program) Pressing the P key will switch between the STANDBY, SUPPLY,
RINSE, FILL, DISINFECTION and PRE-VENTILATION programs, and
will also abort the current program.

 key (Enter) Pressing the  key starts the selected program, confirms a display
message or saves a value.

E key (Settings) Pressing the E key will display the settings menu at any time to change
the alarm limits.

VW keys The V and W keys can be used to retrieve parameters and change
values.

I key (Information menu) Pressing the I key will display the information menu.

Fig.: Information menu

Note
In the event of a power failure the entered data and settings will be
retained in the memory. The system will be deaerated for a short time
in the SUPPLY program after a power failure.

2-2 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 2: Operating Modes and Functions

2.3 Operating Programs

The STANDBY, SUPPLY, RINSE, FILL, DISINFECTION and PRE-VENTILATION programs are the
operating programs of the CDS 08.

2.3.1 STANDBY

If no dialysis is performed, the CDS 08 must be made ready for operation in the STANDBY program. This
program is used to circulate the fluid in the entire ring main automatically for a short time at hour intervals.

Fig.: Program sequence

Fresenius Medical Care CDS 08 OP 1/01.04 2-3


Chapter 2: Operating Modes and Functions

2.3.2 SUPPLY

The SUPPLY program is used to supply the connected media supply centers with the necessary
concentrate. The message SUPPLY (circ.) will be displayed.

The SUPPLY program is used to check the tank level and readjust the set operating pressure.

The VENTILATE/DRAIN program is activated every hour to prevent contamination of the ring main.

Pressing the P key will terminate the SUPPLY program. Pressing the P key once again will additionally
deaerate the ultrafilter (deaeration time = 10 s).

After dialysis use the P key of the SUPPLY program to switch to the STANDBY program.

2.3.3 RINSE

The RINSE program is used to rinse the ring main with permeate before changing the type of concentrate.

Both tank connectors are connected to a permeate tank to rinse the ring main. The CDS 08 collects
permeate alternately from tank 1 or 2. After the set drain time has elapsed, the CDS 08 checks the
conductivity at the ring outlet. If this value falls below 2 mS/cm, the RINSE program will be terminated.

The fluid will be rinsed from the ring main and diverted to the drain.

2.3.4 FILL

The FILL program is used to fill the entire system for the first time.

The CDS 08 alternately collects the corresponding concentrate via both tank feed lines, and checks the
conductivity during this phase. If the conductivity exceeds 2 mS/cm, the FILL program will be terminated.

The fluid will be rinsed from the ring main and diverted to the drain.

2.3.5 DISINFECTION

The CDS 08 can disinfect the entire ring main semi-automatically.

Note
Disinfection may only be performed by authorized personnel or
trained technicians.

2-4 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 2: Operating Modes and Functions

2.3.6 PRE-VENTILATION

The PRE-VENTILATION program of the CDS 08 consists of two phases:


– Deaeration phase
The CDS 08 requires approx. 5 sec/m for the deaeration phase which is determined by the tube lenth
from the tank to the CDS 08.
– Deaeration time
The deaeration time is the time used for deaerating the ultrafilter. It is set using the setting menu
(standard time: approx. 10 sec.).

The PRE-VENTILATION program allows to perform the first deaeration phase at an earlier stage during
a dialysis-free time. The CDS 08 memorizes this and will deaerate only for the second deaeration time
during the next automatic tank swapping.

O Operation of the PRE-VENTILATION program

If an empty tank is detected during dialysis, swapping to the other tank takes place. The red traffic light
indicator of the empty tank is on.

On completion of dialysis the empty tank will be replaced. The red traffic light indicator is still illuminated.

After completion of dialysis the SUPPLY program will be aborted. The PRE-VENTILATION must now be
started to switch the traffic light indicator from red to green.

The CDS 08 deaerates the new-connected tank until the deaeration time has elapsed and switches to the
STANDBY program.

If there is a tank swapping during dialysis, deaeration takes place only for the set deaeration time.

Fresenius Medical Care CDS 08 OP 1/01.04 2-5


Chapter 2: Operating Modes and Functions

2-6 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 3: Disinfection and Preservation

3 DISINFECTION AND PRESERVATION

Caution
Disinfection and preservation of the CDS 08 must only be performed
by technicians authorized by Fresenius Medical Care.
Disinfection and preservation of the CDS 08 must be performed in
accordance with the applicable disinfection and preservation
regulation.

When adding the disinfectant, care must be taken not to spill any
disinfectant concentrate. To prevent burns these fluids must always be
handled with care:

– Use rubber gloves to avoid contact of the disinfectant with the skin.
– Wear goggles when working with acidic substances!

Note
The system may only be disinfected after consulting the
manufacturer of the unit or by personnel authorized by the
manufacturer.

Caution
Observe the safety precautions of the disinfectant used!
In the event of contact with acid:
– Eye:
Immediately flush with gently flowing water for 15 minutes.
– Skin:
Use soap under flowing water for neutralisation.
Ingestion:
Do not induce vomiting, but have victim drink plenty of still water.

Fresenius Medical Care CDS 08 OP 1/01.04 3-1


Chapter 3: Disinfection and Preservation

3-2 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 4: Error Messages

4 ERROR MESSAGES

Note
In the event of an error, the CDS 08 will display the corresponding
error message.

4.1 Tank 1 empty !

Status indicator: Yellow

Cause: Automatic tank swapping, tank 1 empty

Remedy: Replace tank 1with a full tank and start the PRE-VENTILATION
program on completion of dialysis.

4.2 Tank 2 empty !

Status indicator: Yellow

Cause: Automatic tank swapping, tank 2 empty

Remedy: Replace tank 2 with a full tank and start the PRE-VENTILATION
program on completion of dialysis.

4.3 Tank empty !

Status indicator: Red

Cause: Manual tank swapping, the selected tank is empty.

Remedy: Replace the tank with a full tank and press the  key to confirm the
displayed message.

Fresenius Medical Care CDS 08 OP 1/01.04 4-1


Chapter 4: Error Messages

4.4 Both Tanks empty !

Status indicator: Red

Cause: Both tanks empty

Remedy: Replace the empty tanks with full tanks and start the PRE-
VENTILATION program.

4.5 RO-Tank empty !

Status indicator: Red

Cause: An empty permeate tank has been detected in the RINSE program.

Remedy: Refill permeate tank.

4.6 Disinf.-Tank empty !

Status indicator: Red

Cause: Disinfectant tank empty

Remedy: Refill disinfectant tank.

4.7 Wrong concentrate !

Status indicator: Red

Cause: Air mixture before the conductivity sensor

Remedy: Start the FILL MODULE and RINSE MODULE programs.

Status indicator: Red

Cause: Concentrate and bicarbonate have been mixed up.

Remedy: Check the concentrates.

4-2 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 4: Error Messages

4.8 Pressure high !

Status indicator: Red

Cause: Pressure after the pumps exceeds the limit of 2.6 bar.

Remedy: Unkink the tubing.

Status indicator: Red

Cause: Tubing to the ultrafilter is kinked.

Remedy: Unkink the tubing.

Status indicator: Red

Cause: Connectors blocked.

Remedy: Check the connectors.

Status indicator: Red

Cause: Ultrafilter is blocked.

Remedy: Replace the ultrafilter.

4.9 Only in STANDBY !

Cause: Manual tank swapping during a program.

Remedy: Terminate the program in progress.

4.10 Only after RINSE !

Cause: Attempt to change from acid to bicarbonate or vice versa.

Remedy: Rinse ring main and fill with new type of concentrate.

Fresenius Medical Care CDS 08 OP 1/01.04 4-3


Chapter 4: Error Messages

4.11 Only by Service !

Cause: Attempt to change system values

Remedy: If required, call the service.

4-4 Fresenius Medical Care CDS 08 OP 1/01.04


Chapter 5: Technical Safety Checks and Maintenance Procedures

5 TECHNICAL SAFETY CHECKS AND


MAINTENANCE PROCEDURES

5.1 General Notes

This chapter includes the Technical Safety Checks (TSC) and the maintenance procedures to be
performed.

The TSC are included in the maintenance procedures.

These checks must be performed every 12 months.

Fresenius Medical Care CDS 08 TM 1/01.04 5-1


Chapter 5: Technical Safety Checks and Maintenance Procedures

5.2 TSC Report and Maintenance Procedures

Manufacturer: Date:

Device: Technician:

Scope and intervals of the Technical Safety Checks


INTERVALS: annually (every 12 months)

The following inspections must be carried out every 12 months at the latest by persons who are qualified
to properly perform the specified maintenance procedures owing to their educational background and
training, their knowledge and experience gained in practice and who are not bound to any directions with
regard to their activity. (see also the BMA announcement of Apr. 02, 1987 and the MPBetreibV of Aug.
29, 1997)

TSC MA No. Description Theoretical Value/Function OK

1 General Inspection

TSC 1.1 Labels and identification Must be present and clearly legible. T

TSC 1.2 Cables and connections of all No damage or apparent changes T


modules and power supply systems

TSC 1.3 Controls / indicators and their May not be damaged, labelling T
labeling clearly legible.

TSC 1.4 Mechanical condition Must permit further safe use. T

MA 1.5 Tubing / tank feeding lines May not be damaged. T

MA 1.6 CPC couplings May not be damaged. T

TSC 1.7 Machine card Must be present. T

TSC 1.8 Medical Device Register Must be present. T

TSC 1.9 Operating Instructions Must be present. T

MA 1.10 Ring main flow diagram Must be present. T

MA 1.11 Clean all components T

TSC 1.12 Check for leaks inside the unit. T

5-2 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 5: Technical Safety Checks and Maintenance Procedures

TSC MA No. Description Theoretical Value/Function OK

2 Functional test of operating programs

TSC 2.1 RINSE program Function; see Operating T


Instructions

TSC 2.2 FILL program Function; see Operating T


Instructions

TSC 2.3 SUPPLY program Function; see Operating T


Instructions

TSC 2.4 PRE-VENTILATION program Function; see Operating T


Instructions

3 Electrical test
Measurement points (see chapter 5.3.1 on page 5-6) for electrical test

TSC 3.1 Current leakage to ground < 1 mA mA

TSC 3.2.1 Protective conductor resistance < 0.3 Ohm Ω

TSC 3.2.2 Protective conductor resistance < 0.3 Ohm Ω


hydraulic module press-fit bolt
upper left

TSC 3.2.3 Protective conductor resistance < 0.3 Ohm Ω


electronic module flat head screw
upper left

TSC 3.2.4 Protective conductor resistance < 0.3 Ohm Ω


front panel mounting screw

4 Setting values

TSC 4.1 Pressure bar

TSC 4.2 Deaeration time sec

TSC 4.3 Tube length m

TSC 4.4 Tank swapping Manual / automatic T

TSC 4.5 Tank-empty detector Internal / external T

TSC 4.6 Concentrate type Complies with the connected tank T


type.

5 Pressures

MA 5.1 Pressure at first media supply bar


center

MA 5.2 Pressure at last media supply bar


center

TSC 5.3 Pressure display Complies with the (nominal) T


pressure.

Fresenius Medical Care CDS 08 TM 1/01.04 5-3


Chapter 5: Technical Safety Checks and Maintenance Procedures

TSC MA No. Description Theoretical Value/Function OK

6 Temperature

6.1 Temperature sensor Displays correct values (room T


temperature).

7 Conductivity

7.1 CD sensor Displays correct values: T


BIC: 50 – 55 mS/cm
Acid: 180 – 220 mS/cm

8 Tank sensors

TSC 8.1 Mechanical condition Must permit further safe use T

TSC 8.2 Leakages No leaks present T

TSC 8.3 Function Proper sensor function T

9 Status indicators

TSC 9.1 Mechanical condition Must permit further safe use T

TSC 9.2 Function All indicators okay T

10 Ultrafilter

TSC 10.1 Mechanical condition Must permit further safe use T

MA 10.2 Replacement once a year Replaced T

11 Remote control

TSC 11.1 Mechanical condition Must permit further safe use T

TSC 11.2 Switching between Standby / Check T


Supply

TSC 11.3 Alarms Displayed visually and indicated T


audibly

12 Carriage

12.1 Mechanical condition Must permit further safe use T

13 Ring main

13.1 Mechanical condition Must permit further safe use T

13.2 Leakages Check the connections to the media T


supply centers for tightness

5-4 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 5: Technical Safety Checks and Maintenance Procedures

The technical safety checks are to be entered in the Medical Device Register and the results of the check
must be documented.
Corrective action must be taken to restore the safe function of any device which is not in proper or safe
service and/or operating condition, or the operator is to be informed of the risk which the use of this
device may present.

I herewith confirm proper performance of any maintenance work as specified above, and the data to be
true and correct.

Technician’s signature: Customer’s signature:

................................................., Date: ......................... ................................................., Date: .........................

............................................................................. .............................................................................

Fresenius Medical Care CDS 08 TM 1/01.04 5-5


Chapter 5: Technical Safety Checks and Maintenance Procedures

5.3 Electrical Test

5.3.1 CDS 08 Protective Conductor Resistance

3.2.2
Protective conductor resistance on hydraulic module
(press-fit bolt on upper left)

3.2.3
Protective conductor resistance on electronics module
(flat head screw on upper left)

5-6 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 5: Technical Safety Checks and Maintenance Procedures

3.2.4
Protective conductor resistance on front panel
(mounting screw)

Fresenius Medical Care CDS 08 TM 1/01.04 5-7


Chapter 5: Technical Safety Checks and Maintenance Procedures

5-8 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 6: Adjustments and Calibration

6 ADJUSTMENTS AND CALIBRATION

6.1 Settings Menu

– Press the E key to access the settings menu.


– The settings menu can be exited at any time by pressing the P and I keys.
– Press the E key or the VW keys to access the individual menu items and then the  key to alter them.
– Press the  key to store the set values. The display will then move automatically to the next menu item
of the settings menu.

Fresenius Medical Care CDS 08 TM 1/01.04 6-1


Chapter 6: Adjustments and Calibration

Fig.: Settings menu

6-2 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 6: Adjustments and Calibration

6.1.1 Concentrate

Define the fluid to be pumped into the connected ring main using the Concentrate menu item.

Whilst the FILL and SUPPLY programs are in progress, the CDS 08 checks at the end of the deaeration
phase whether the connected concentrate is identical to the concentrate preset in the Concentrate menu.
The concentrate check prevents any non-preset concentrate from entering the ring main.

O Possible settings

– Acid: acid bicarbonate concentrate or acetate concentrate


– Bicarbonate: 8.4 % bicarbonate

6.1.2 Tank

The menu option Tank can be used to skip automatic tank swapping and initiate manual tank swapping
from tank 1 to tank 2 and vice versa. This however, can be done only when the CDS 08 is in the
STANDBY program.

6.1.3 Pressure (reg.)

The menu option Pressure (reg.) can be used to set the CDS 08 output pressure for pressure regulation.
This parameter sets the pressure at the user points of the ring main.

In the SUPPLY, RINSE, FILL, and DISINFECTION (cold) programs, pressure control is active in the
Drain and Circulation phases.

6.1.4 Pressure (hot)

No function

6.1.5 Ventilation Time

The ventilation time, i.e. the dearation time is composed of the time interval required to deaerate the line
running from the tank to the CDS 08 and an additional deaeration time for deaerating the ultrafilter:

Total deaeration time


= line deaeration (tank to CDS 08) + additional deaeration time

The deaeration time does not have any effect on the brief deaeration during the SUPPLY program which
can be performed every hour, nor on the Pre-ventilation which can also be performed. The default
deaeration time has been set to 10 seconds in the settings menu.

Fresenius Medical Care CDS 08 TM 1/01.04 6-3


Chapter 6: Adjustments and Calibration

6.1.6 Drain Time

The drain time parameter is required in the RINSE and FILL programs during the draining phase. It is
dependent on the length and the cross-section of the ring main and the number of user points in the ring
main. While the drain time is elapsing, the entire contents of the ring main are replaced by a new fluid.
The drain time must be set on the basis of empirical values.

6.1.7 Disinfection Time

The disinfection time parameter is required in the DISINFECTION program during the circulation phase;
it defines the time of exposure to the disinfectant. During the hot rinse phase, where a temperature of 85
°C is to be reached at the end of the ring main, the disinfection time depends on the length of the ring main
and the number of user points in the ring main.

6.1.8 Tube Length

The menu option Tube length is provided to enter the length of the tube running from the tank via the
external tank sensor to the CDS 08. The tube length parameter is taken to calculate the first phase of the
deaeration time (5 sec/m).

Note
During the deaeration phase, the CD/T conductivity and temperature
sensor is not activated.

6-4 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 6: Adjustments and Calibration

6.2 System Parameters

Fresenius Medical Care CDS 08 TM 1/01.04 6-5


Chapter 6: Adjustments and Calibration

In order to set specific operating parameters, the DIP switch IC 23, Pos. 2 / P.C.B. LP 795, must be reset.
This will enable the menu option Edit system parameters in the settings menu.

Caution
System parameters may only be changed by technicians authorized
and trained by Fresenius Medical Care.

Fig.: P.C.B. LP 795 detail

6.2.1 Tank Swapping

Tank swapping can be done either automatically or manually.

O Manual Tank Swapping

The CDS 08 detects the empty tank and switches to the alarm position.

O Automatic tank swapping

The CDS 08 automatically switches to the filled tank.

6-6 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 6: Adjustments and Calibration

6.2.2 Tank-Empty Detector

O Internal tank swapping unit

The internal tank swapping unit is used for systems where the distance between the CDS 08 and the
storage tanks is very short (< 5m).

The internal tank swapping unit is provided with three redundant procedures for detecting an empty tank.
Monitoring of the pump pressure, the pump speed and the conductivity. As soon as certain limits are
exceeded or fallen below, an empty tank is detected; if necessary, the system will switch to the other tank.
The limits for the pressure drop and the pump voltage are preset. The conductivity limit must be set in the
System parameters menu.

Although the internal tank swapping unit is used, a tank level indicator can be connected; the signals of
the external tank swapping unit will then however, not be detected.

O External tank swapping unit

The external tank swapping unit consists of two external tank sensors (conductivity sensors detecting
whether the tanks are empty or not empty) and a controller with a signal traffic light.

The signal traffic light indicates the tank levels "not empty" (green) and "empty" (red).

The controller transfers the current tank level to the CDS 08. The supply line of the traffic light controller
must be connected to the tank level connector on the rear of the CDS 08.

6.2.3 Supply Mode

The system parameter Supply Mode can be used to set the type of ring main installed.

6.2.4 Pump Speed

The system parameter Pump speed can be used to define the pump speed during deaeration. This set
value only controls the operating speed of the pump during the deaeration of the ultrafilter.

6.2.5 Language

The system parameter Language can be used to set the display software to the following languages:
German, English, French, Italian, Spanish and Swedish.

6.2.6 Brightness

The display brightness can be set in four steps.

Fresenius Medical Care CDS 08 TM 1/01.04 6-7


Chapter 6: Adjustments and Calibration

Note
If no key is pressed for a prolonged time, the display brightness is
automatically reduced to 25 %. If a key is then pressed , the
messages are again displayed with the set brightness.

6.2.7 KI / KP

The pumps are pressure-controlled by means of a PI controller. The two system parameters KI and KP
are used to set the control characteristics. These two parameters effect the rate at which the gear pumps
respond to a change in pressure in the ring main.

6.2.8 Tank SWCondBic – Bicarbonate Conductivity Limit

The system parameter Tank SWCondBic can be used to set the bicarbonate conductivity limit at which
the interal tank-empty detector of the CDS 08 detects an empty tank.

6.2.9 Tank SWCondACID – Acid Concentrate Conductivity Limit

The system parameter Tank SWCondACID can be used to set the acid concentrate conductivity limit at
which the interal tank-empty detector of the CDS 08 detects an empty tank.

6.2.10 External tank sensors

It is not possible to detect permeate or disinfectant if conductivity sensors are used. The CDS 08 would
always detect an empty tank in the corresponding program. The setting for this system parameter is
therefore fixed to No. The CDS 08 automatically switches back to the internal tank-empty detector,
suppressing the detection of an empty tank by the conductivity sensors in this phase.

6.2.11 Masch.hour

The system parameter Masch.hour can be used to display the operating hours of the CDS 08.

6.2.12 Circulation Pause

The system parameter Circulation pause is used to set the time period between the individual circulation
phases. Circulation is only performed in the STANDBY program.

6.2.13 Circulation Time

The system parameter Circulation time is used to set the duration of circulation. Circulation is only
performed in the STANDBY program.

6-8 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 6: Adjustments and Calibration

6.2.14 Ventilation Standby

The system parameter Ventilation Standby is used to set the deaeration time after the hourly circulation
of the system.

Fresenius Medical Care CDS 08 TM 1/01.04 6-9


Chapter 6: Adjustments and Calibration

6.3 Calibrations

6.3.1 Pressure Transducer Zero-Point Calibration

– Turn the CDS 08 power on.


– Select the STANDBY program (the pump must be stopped).
– Press the I key.
– Select the option Pressure from the information menu.
– The concentrate supply system must be depressurized.
– Set the pressure to 0 mbar using the potentiometer P1 / LP 795.

Note
With the 50 PSI ENVEC pressure transducer the potentiometer for
the zero-point calibration on P.C.B. LP 795 is no longer required.
A trimming potentiometer is therefore located on the printed circuit
board of the pressure transducer (calibration tolerance: +115 mbar).

6.3.2 External tank sensors

– The switching threshold of the external tank transducer is set to approx. 900 ohms.
– If the external tank-empty detector switches too early or too late, the operating point can be reset using
the potentiometer P1.
– Turn the potentiometer P1 clockwise: this permits better detection of fluids.
– Turn the potentiometer P1 counter-clockwise: this permits better detection of air.

Signal (WH)
+Vs (GRY)
GND (GN)
Pressure sensor (YL)
Pressure sensor (BRN)

Installation position

n.c.
Pressure sensor
Pressure sensor
CD cell
CD cell

6-10 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 6: Adjustments and Calibration

6.3.3 Remote control indicator

The remote control indicator is connected to the sub D9 connector on the rear of the CDS 08.

Remote Control

AQUASAFE 08

CDS 08
?

I/O
Manual

Fresenius Medical Care

The printed circuit board of the remote control indicator is provided with a jumper which must be used to
program the remote control indicator for the CDS 08.

Setting CDS 08:


Jumper to the left

Fresenius Medical Care CDS 08 TM 1/01.04 6-11


Chapter 6: Adjustments and Calibration

6.4 Installation and Start-Up

Start-up of the CDS 08 must only be performed by technicians authorized by Fresenius Medical Care.

6.4.1 Preparation

– Connect all lines to the CDS 08.


– Connect the signal traffic light.
– Mount the ultrafilter unit and insert the Diasafe.
– Open the ring main pressure holding valve R3.

6.4.2 Filling and Deaerating the Tank Connectors

– Connect and open tank 1 (left tank).


– Remove the coupling of tank 1 from the CDS 08, fit the connector, and deaerate the tank supply line.
– Connect and open tank 2 (right tank).
– Remove the coupling of tank 2 from the CDS 08, fit the connector, and deaerate the tank supply line.

6.4.3 Filling and Deaerating the Ring Main

– Turn the CDS 08 power on.


– Make the following settings in the Settings menu:
Drain time = 30 min
Ventilation time = 240 sec
Pressure (reg.) = 800 mbar
– Start the FILL program.
– During the Ventilation time, the ultrafilter unit is filled and deaerated.
– In the subsequent Drain time, the ring main is filled and deaerated.
– Increase the Pressure (reg.) in increments up to 2000 mbar during the drain time.
– Deaerate all user points at the media supply centers (MSC).
– The ring main is deaerated as soon as air or acid-air mixture no longer escapes from the outlet of the
CDS 08. If the drain time of 30 minutes is insufficient, the FILL program must be restarted.

6.4.4 Setting the Ring Main Pressure

– Set the standard values in the Settings menu.


– Start the SUPPLY program.
– Set the Pressure (reg.) such that a pressure of 0.4 bar is present at the first media supply center.
– Using the ring main pressure holding valve R3, set the pressure such that 0.15 bar is present at the
last media supply center.

6-12 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 6: Adjustments and Calibration

6.4.5 Checking the Tank Swapping Unit

– Select tank 1.
– Attach a jumper across terminal 1 and terminal 2 (conductivity electrodes) at the tank sensor of tank 1.
– The tank 1 tank sensor detects conductivity; the LED is lit; and the green indicator of the signal traffic
light (to the left) is illuminated.
– Release the jumper from terminal 1 and terminal 2 (conductivity electrodes) of the tank sensor of tank
1.
– The tank 1 tank sensor does not detect any conductivity; the LED is not lit; and the red indicator of the
signal traffic light (to the left) is illuminated.
– The CDS 08 must detect that tank 1 is empty and must switch to tank 2.
– Confirm the Tank 1 empty message.
– Select tank 1.
– Attach a jumper across terminal 1 and terminal 2 (conductivity electrodes) at the tank sensor of tank 1.
– The tank 2 tank sensor detects conductivity; the LED is lit; and the green indicator of the signal traffic
light (to the right) is illuminated.
– Release the jumper from terminal 1 and terminal 2 (conductivity electrodes) of the tank sensor of tank
2.
– The tank 2 tank sensor does not detect any conductivity; the LED is not lit; and the red indicator of the
signal traffic light (to the right) is illuminated.
– The CDS 08 must detect that tank 2 is empty and must switch to tank 1.
– Confirm the Tank 2 empty message.

Note
The tube length must be set correctly in the Settings menu of the
CDS 08.
If the tanks are connected using the white fabric-reinforced tube, an
additional 2 meters must be added to the actual tube length for safety
reasons.
If the tanks are connected using the black tube, the tube length must
be set to twice the length in the Settings menu because of the larger
cross-section.
Select the PRE-VENTILATION program after each tank replacement
to ensure that the tank connector is deaerated correctly.

Fresenius Medical Care CDS 08 TM 1/01.04 6-13


Chapter 6: Adjustments and Calibration

6.4.6 Start-Up Report

Manufacturer: Date:

System: Serial number:

Operating hours: Serial number:

Location: Service report. no.:

1. Tank swapping

Manual T

Automatic T

2. Tank-empty detector

Internal sensors T

External sensors T

3. Desired pressure mbar

4. Concentrate T

5. Tube length m

6. Outlet pressure at the first media supply center without dialysis machines mbar

I herewith confirm proper performance of any maintenance work as specified above, and the data to be
true and correct.

Technician’s signature: Customer’s signature:

................................................., Date: ......................... ................................................., Date: .........................

............................................................................. .............................................................................

6-14 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7 CIRCUIT DIAGRAMS AND CIRCUIT


DESCRIPTIONS

7.1 CDS 08 Wiring Diagram

external tank sensors

ext. tank-empty detector


(status indicators)

Remote control

Heater (option)
Power supply

Nurse call
unit

RS 232

Pump
Pump
Valve (tank 1)
Valve (tank 2)
Valve (deaeration)
Valve (drain)
Valve (circ)
Temperature
Pressure
Conductivity

Display

Fresenius Medical Care CDS 08 TM 1/01.04 7-1


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-2 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.2 EMA Power Supply Unit Rev 2.1

7.2.1 Circuit Description

O Direct voltage generation, + 24 volts

Approx. 19 V are supplied from the transformer via the fuse S1 3.15 A MT to the rectifier GL 1. Four
capacitors C4 to C7 are connected in parallel as charging electrolytic capacitors on the DC side of the
rectifier. The smoothed voltage is available at the connector strip X3, terminal KL2. LED 5 serves only to
indicate whether voltage is available.

O Direct voltage generation, ±12 volts

Two 15-volt circuits are supplied from the transformer via the two fuses SI3 2A MT and SI2 1A MT to the
rectifier GL 2. On the DC side of the rectifier, the charging electrolytic capacitors C1 for the positive voltage
and C2 for the negative voltage are connected as smoothing capacitors. The smoothed DC voltage is
supplied to the voltage regulator IC 1 for the positive voltage and IC 2 for the negative voltage. The voltage
regulators generate voltages of +12 volts and –12 volts; C 8 for the positive voltage and C 9 for the
negative voltage. C12 and C13 are blocking capacitors for high-frequency oscillation.

LD 2 and LD 3 are pure indicating diodes which require the presence of ±12 volts.

The 5-volt generation is not required in the CDS 08 and is, therefore, not fitted.

O Power-on logic

The line power on and off logic is supplied by a separate power supply unit, consisting of the transformer
TR 1, the rectifier GL4, and the smoothing capacitor C 14. The voltage regulator IC 4 stabilizes the supply
voltage at +12 V. LED 4 serves only to indicate whether the internal supply voltage is available.

Power on is executed by pressing the I/O key at the input connector X3, pin 2, against ground. This causes
C 21 to discharge via R 14. When the switching threshold of gates IC 7, pins 1 and 2, is reached, output
pin 3 becomes high. With the rising slope, the data flip-flip IC 5, pin 3, is triggered. The data flip-flop which
is connected as a toggle changes its state at output pin 1 each time the I/O key is pressed. In this manner,
the line power relays RL 1 and RL 2 are switched on and off via the transistor T 1. D 1 serves as a free-
wheeling diode. Switching the system on or off by pressing the I/O-button unintentionally is prevented by
a time delay of approx. 1 second, consisting of C 21 and R 14.

Fresenius Medical Care CDS 08 TM 1/01.04 7-3


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.2.2 Circuit Diagram and Component Layout of EMA Power Supply Unit Rev 2.1

O Component layout of EMA Power Supply Unit Rev 2.1

7-4 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram of EMA Power Supply Unit Rev 2.1

Fresenius Medical Care CDS 08 TM 1/01.04 7-5


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-6 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.3 EMA Power Supply Unit Rev. 3 1/1

7.3.1 Circuit Description

O Direct voltage generation, + 24 volts

Approx. 19 V are supplied from the transformer via the fuse S1 3.15 A T to the rectifier GL 1. Four
capacitors C4 to C7 are connected in parallel as charging electrolytic capacitors on the DC side of the
rectifier (C22 is not fitted). The smoothed voltage is available at the connector strip X3, terminal KL1. LED
5 serves only to indicate whether voltage is available.

O Direct voltage generation, ±12 volts

The two 15-V supplies from the transformer via the fuses SI 2 (1Am) and SI 3 (2Am) to the rectifiers GL
2 and GL 3. On the DC side of the rectifiers, C1 (for the positive voltage) and C2 (for the negative voltage)
are connected as smoothing electrolytic capacitors. The smoothed DC voltages are supplied to the
voltage regulator IC 1 for the positive voltage and IC 2 for the negative voltage. The voltage regulators
generate voltages of +12 volts (3 A) and -12 volts (1 A). Used as blocking capacitors for the high-
frequency oscillation tendency are C8 and C12 for the positive voltage and C9 and C13 for the negative
voltage.
LED 2 and LED 3 are pure indicating diodes for the presence of the ±12-V voltage.

O Power-on logic

– Jogging operation (with ON/OFF delay)


The line power on and off logic is supplied by a separate power supply unit, consisting of the
transformer TR 1, the rectifier GL4, and the smoothing capacitor C 14. The voltage regulator IC 4
stabilizes the supply voltage at +12 V. LED 4 serves only to indicate whether the internal supply
voltage is available.
Power on is executed by pressing the I/O key at the input connector X3, pin 2, against ground. This
causes C 21 to discharge via R 14. When the switching threshold of gates IC 7, pins 1 and 2, is
reached, output pin 3 becomes high. With the rising slope, the data flip-flip IC 5, pin 3, is triggered. The
data flip-flop which is connected as a toggle changes its state at output pin 1 each time the I/O key is
pressed. In this manner, the line power relays RL 1 and RL 2 are switched on and off via the transistor
T 1. D 1 serves as a free-wheeling diode. Switching the system on or off by pressing the I/O key
unintentionally is prevented by a time delay of approx. 1 second, consisting of C 21 and R 14.
– Switching operation
If a power switch is used, the on and off logic can be bypassed by refitting the bridge B8. This, however,
requires a power switch.
Power on after a failure of the line voltage is, thus, not required any longer.

Fresenius Medical Care CDS 08 TM 1/01.04 7-7


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.3.2 Circuit Diagram and Component Layout of EMA Power Supply Unit Rev. 3 1/1

O Component layout of EMA Power Supply Unit Rev. 3 1/1

7-8 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram of EMA Power Supply Unit Rev. 3 1/1

Fresenius Medical Care CDS 08 TM 1/01.04 7-9


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-10 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.4 Control P.C.B. LP 795

7.4.1 Circuit Description

O General notes

The P.C.B. LP 795A comprises the entire control for the CDS 08, consisting of analog section,
microcontroller and actuator activation.

Circuit diagram 1/3 comprises the reference voltage generation as well as the measurement circuits for
temperature, pressure and conductivity. Circuit diagram 2/3 comprises the microcontroller including its
peripheral equipment. Circuit diagram 3/3 comprises the power output stages for the DC motors and the
solenoid valves.

O Description

Page 1 comprises the reference voltage generation as well as the analog circuits for temperature,
pressure and conductivity.

The +10-V reference voltage is generated by IC 13. This voltage is used to obtain +5REF and +COND by
means of voltage division with R30/R32/R36. +COND is used to generate the -COND (identical in
amount) by means of the inverting amplifier IC 17/3.

The temperature measuring circuit consists of the OP IC 17/1, whose non-inverting input is applied to half
the reference voltage (+5REF). The NTC is connected in series with the linearizing resistor R25 in the
inverting branch. The load resistor R26 and the feedback resistor R24 have been selected so that, across
the desired range of measurement of 0 – 80 degrees, an output voltage of 0 – 5 V is achieved. C21 serves
to smooth the output voltage.

The pressure measuring amplifier consists of the OPs IC 18, which are connected as instrument
amplifiers. P1 can be used to calibrate the offset of the pressure sensor. The circuit delivers an output
voltage of 0 – 5 V at a pressure of –0.2 bar to 2.8 bar.

The conductivity measurement circuit is an oscillator whose conductivity cell serves as a frequency-
defining resistor. A voltage ramp is compared with the cell voltage which serves as a comparator
threshold. The oscillator consists of a buffer amplifier (IC 12/1) with a voltage divider (R44/CD cell) on its
output side, the integrator IC 24/1, the comparator IC 21, and two analog converters. The -COND
reference voltage is supplied to the amplifier IC 24/1 and the integrator IC 24/2 via the analog switch IC
20-14/15. A positive voltage which reaches the CD cell via R44 and the relay contact RL2 develops at IC
24/1. The partial voltage developing at the CD cell is supplied to the comparator IC 21 via IC 20-3-4. The
inverting input of IC 21 is connected to the output of the integrator which, at its output, delivers a voltage
ramp with constant slope. The comparator output remains high as long as the integrator voltage is less
than the cell voltage. As soon as the switching threshold is reached, the comparator changes its state and
reverses IC 20. As a result, +COND is applied to IC 24/1 and IC 24/2. The polarity of the cell voltage as
well as the slope of the integrator ramp are reversing. At the same time, the comparator threshold is
applied to ground, so that this switching state is preserved until the integration capacitor C30 has
discharged. Then the comparator changes its state again, and the cycle is repeated. At a conductivity of

Fresenius Medical Care CDS 08 TM 1/01.04 7-11


Chapter 7: Circuit Diagrams and Circuit Descriptions

0 mS/cm (infinite cell resistance), the entire output voltage of IC 24/1 drops at the cell. C30 must then be
charged until it has reached this voltage, before the comparator changes its state. In this case, this results
in the lower frequency limit of the circuit. This frequency can be defined via the integrator resistor R39. If
the conductivity rises, only a partial voltage drops at the cell; the switching threshold is reduced and, thus,
the frequency increased. The slope (frequency / conductivity) can be determined using the voltage divider
Rxx.

The two relays RL2 and RL3 are used to calibrate the circuit whenever the system is switched on. The
controller first activates RL2 and, with open RL3 (conductivity = 0), measures the zero-point frequency of
the circuit. Then, with RL3 in the stead of the CD cell, the reference resistor R47 is connected. This
resistance is dimensioned such that it approximately corresponds to the minimum cell resistance for the
desired range. These two values can be used to determine the slope. Subsequently, RL2 is used to
reconnect the CD cell.

The feedback capacitor C29 improves the switching behavior of the comparator. The output signal of the
comparator is used to obtain a TTL-satisfactory level via R41 and R42.

Page 2 comprises the 80C535 microcontroller IC7 including peripheral equipment. The 80C535 is
provided with six I/O ports and 8 analog inputs. The ports serve as inputs and outputs for keys, valve
drivers, lamp drivers, etc. (see I/O definitions "CDS_IO.H").

Two of the analog inputs are used to measure temperature and pressure. The resistors R20 and R22 limit
the input current flowing into the controller inputs in case the range is exceeded (<0 V, >5 V). The output
frequency of the conductivity measuring circuit is supplied onto the counter input T1 (pin 26) of the
controller. The controller reads out the counter reading every half-second, using it to determine the
conductivity.

The interface module IC4 assumes the task of converting the level of the RS232 interface from TTL to
±10 V and vice versa.

As is typical for controllers of the 8051 series, the multiplexed address/data bus of the 80C535 is
connected to the EPROM IC15, the external RAM IC14, the digital-to-analog converter IC11, and the
buffers IC19 and IC22, via the address latch IC16.

RAM/ROM address range:


– (Battery) RAM:
0..1FFFH
– EPROM:
0..7FFFH (with type 27C256, J1: 2-3)
or:
0..FFFFH (with type 27C512, J1: 1-2)

The external peripheral modules can be addressed in the address range above the RAM. The addresses
are decoded by means of the GAL IC10. With the READ and WRITE lines of the controller included,
linking of the address lines A13..A15 results in the following I/O addresses:
– DAC_OUT (2000H):
The digital-to-analog converter IC11 is switched such that it can deliver an output voltage of 0..10 V.
This voltage serves as the rated value for the pump output stage.
– DIG_IN (4000H):
The input buffer IC19 is provided with 8 digital inputs for the DIP switch IC23, the external tank-empty
sensors, and the external heater module.
– DISPLAY (6000H):
The data for activation of the VF display is transmitted to the front panel via the output buffer IC22.

7-12 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

Page 3 comprises the circuit sections for activation of the actuators. Via the level shifters IC8/IC9 and the
FETs T4..T11, the controller port P1 switches the pump and valve outputs.

The PWM controller IC2 is supplied with an analog rated value of 0 – 10 V by the DAC IC11 (page 2). The
internal OP compares this value with the current pump voltage. If necessary, the pulse duty factor is
readjusted. Amplification is set to 2.5 by the ratio of the resistors R6/R5 to R12/R7. Thus, a rated value of
0...10 V causes a pump voltage of 0...25 V. The resistors R13 and R14 provide for an admissible common-
mode range of the internal OPs. The resistor R10 loads the DAC to such a degree that the latter's output
voltage can fall to values of almost 0 V. The open-collector outputs at the pins IC2-12/13 activate the
quasi-Darlington stage T3/T2 via R9. The output voltage for the motors is available as filtered DC voltage
after the storage inductance coil DR4. The voltage drop at the current sensing resistors R16/R19 caused
by the motor current is supplied to the current-limiting input IC2-4, limiting the output current to 4 amperes.

The dual optocoupler IC1 is intended to isolate the potentials between the tank-empty sensors and the
controller inputs.

Fresenius Medical Care CDS 08 TM 1/01.04 7-13


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.4.2 Circuit Diagram and Component Layout of Control P.C.B. LP 795

O Component layout of Control P.C.B. LP 795

7-14 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram 1/3 Control P.C.B. LP 795

Fresenius Medical Care CDS 08 TM 1/01.04 7-15


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-16 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram 2/3 Control P.C.B. LP 795

Fresenius Medical Care CDS 08 TM 1/01.04 7-17


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-18 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram 3/3 Control P.C.B. LP 795

Fresenius Medical Care CDS 08 TM 1/01.04 7-19


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-20 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.5 Keypad P.C.B. LP 796

7.5.1 Circuit Description

O General notes

The P.C.B. LP 796 comprises the keypad, the LED indicators and the VF display connection for the
CDS 08. The board is connected to the controller board LP 795 by means of a 34-pin ribbon cable.

O Description

On the anodic side, the LED indicators are connected to the +12-V voltage. They are energized by the
corresponding Darlington circuits on the P.C.B. LP 795 via variable multipliers.

The keys switch the corresponding inputs of port 4 of the 80C535 to ground.

The connections for the VF display module are connected through from X1 to X2.

Fresenius Medical Care CDS 08 TM 1/01.04 7-21


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.5.2 Circuit Diagram and Component Layout of Keypad P.C.B. LP 796

O Component layout of Keypad P.C.B. LP 796

7-22 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram of Keypad P.C.B. LP 796

Fresenius Medical Care CDS 08 TM 1/01.04 7-23


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.6 P.C.B. LP 797

7.6.1 Circuit Diagram and Component Layout of P.C.B. LP 797

O Component layout of P.C.B. LP 797

7-24 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram of P.C.B. LP 797

Fresenius Medical Care CDS 08 TM 1/01.04 7-25


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-26 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.7 Remote Control Indicator

7.7.1 Circuit Description

O Structure

The remote control indicator board is accommodated in a separate housing and is connected to the
CDS 08 via a 9-pin connector.

O Functions

When the key in the remote control indicator is actuated, the binary counter IC1 4060 is enabled by means
of a low level at the reset input. The counter is operated with a clock pulse frequency of approx. 250 Hz.
By means of the wiring at pin 14/output 8, the clock pulse is divided by 256 (flashing frequency of approx.
1 Hz), thus energizing the "Standby" LED and , inverted via T3, the "Supply" LED. After division by 2048
at the connector pin 1/Q12, a high level is applied after approx. 8 seconds. This level switches T4 and
generates a 12-V level at X1/pin 6. Using the jumper J1, this output level can be inverted by T6. This
request signal is supplied to the CDS 08.

The signals are supplied to the remote control indicator via an OC driver ULN 2003A. The audible alarm
is activated via IC3A-IC3D, in addition to the visual alarm (red). Using IC2A, the alarm level can be
reduced by actuating the key. P2 defines a "loud" level; P1 defines a "reduced" level.

Signal CDS 08 / Remote control 10-pin


9-pin D-Sub indicator connector
Terminal strip X1

Standby 9 8 8

Supply 4 7 7

Status 1, yellow 8 6 6

Status 2, yellow 3 5 5

Normal 7 4 4

Alarm red 2 3 3

+12 V 1 1 1

GND 5 9 9

Request signal 6 2 2

Fresenius Medical Care CDS 08 TM 1/01.04 7-27


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Pin assignment of terminal strip – SUB-D9 connector

red
UB (+12 V)
green
Request signal
blue LED indicator, red
and signal generator
brown
LED indicator, green
Sub-D9
pink LED indicator, yellow
connector
and status LED 2
white LED indicator, yellow
and status LED 1
gray
Supply LED
black
Standby LED
yellow
GND

violet (not used) LED indicator, yellow

7-28 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Pin assignment of remote control

Remote control RS232 (EDP)

Exposed socket
Control line, shielded Station EDP line, shielded
J-Y(ST)Y 6 x 2 x 0.6 mm2 Category 5: CAT 5 / 4x2 x 0.6
max. load: 40 mA

Connect 1:1
Color coding Connect 1:1
see table below

Remote control Exposed socket EDP / network


Equipment room

Remote control RS232 (EDP)

Pin assignment of J-Y(ST)Y control line:

1 blue (blue-red) +12 V

2 red (blue-red) LED status indicator red and signal tone


generator

3 yellow (yellow-white) LED status indicator yellow and status 2

4 white (yellow-white) Supply LED

5 green (green-white) GND

6 white (green-white) Request signal (+12 V or GND)

7 brown (brown-white) LED status indicator green

8 white (brown-white) LED status indicator yellow and status 1

9 black (black-white) STANDBY LED

Fresenius Medical Care CDS 08 TM 1/01.04 7-29


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.7.2 Circuit Diagram and Component Layout of Remote Control Indicator

O Component layout of Remote Control Indicator

7-30 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram of Remote Control Indicator

Fresenius Medical Care CDS 08 TM 1/01.04 7-31


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-32 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.8 Status Indicator Control

7.8.1 Circuit Description

The status indicator control is activated via a conductivity sensor (CD sensor). Red light indicates that the
connected tank is empty. Green light indicates that the connected tank is ready for operation.

If an external tank sensor is addressed, the "GND" signal is supplied to the connection terminal 5 (for tank
sensor 1) and connection terminal 10 (for tank sensor 2) respectively.

This signal causes a relay to pick up, whose changeover contact switches the red LED off and, because
of the jumpered pressure sensor, the green LED on via a further changeover contact.

If the signal light is red, the "GND" changeover signal for the CDS 08 is made by means of a transistor
(T1 or T2 = BC 107).

Fresenius Medical Care CDS 08 TM 1/01.04 7-33


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.8.2 Circuit Diagram and Component Layout of Status Indicator Control

O Component layout of Status Indicator Control

7-34 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram of Status Indicator Control

Fresenius Medical Care CDS 08 TM 1/01.04 7-35


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-36 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.9 External Tank Sensor

7.9.1 Circuit Description

The main components of the conductivity-controlled tank-empty detector are the two sensor electrodes
(CD cell at X2/1 and 2), which are separated by a glass cuvette, and the comparator circuit OP2.

The output of an oscillator circuit (pin 7 OP1) feeds a transformer L1 which, on the one hand, DC-
decouples the test voltage at the sensor electrodes and, on the other, makes available a better-to-
evaluate signal by transforming the impedance.

If, for instance, the tank is reaching empty, the impedance of L1 rises (by the transformation ratio of L1 =
12 V / 220 V). The voltage at R6 drops. C7 is discharging. Now, the signal is supplied to the comparator
via a rectifier. This desired-and-actual-value comparator now generates a signal for differentiating the
following two states:
– Fluid present (L) (i.e. T1 is disabled, LD1 is off).
– Fluid not present (H) (i.e. T1 is conducting, LD1 is on).

Calibration: The switching threshold is set to approx. 900 ohms; any further calibration is not required. If,
however, the tank still switches over too early or too late, the operating point can be set using the
potentiometer P1. Clockwise adjustment: permits better detection of fluid. Counter-clockwise adjustment:
permits better detection of air.

Fresenius Medical Care CDS 08 TM 1/01.04 7-37


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-38 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Block diagram of External Tank Sensor

Fresenius Medical Care CDS 08 TM 1/01.04 7-39


Chapter 7: Circuit Diagrams and Circuit Descriptions

7.9.2 Circuit Diagram and Component Layout of External Tank Sensor

O Component layout of external tank sensor

Signal (WH)
+Vs (GRY)
GND (GN)
Pressure sensor (YL)
Pressure sensor (BRN)

Installation position

n.c.
Pressure sensor
Pressure sensor
CD cell
CD cell

7-40 Fresenius Medical Care CDS 08 TM 1/01.04


Chapter 7: Circuit Diagrams and Circuit Descriptions

O Circuit diagram of external tank sensor

Fresenius Medical Care CDS 08 TM 1/01.04 7-41


Chapter 7: Circuit Diagrams and Circuit Descriptions

7-42 Fresenius Medical Care CDS 08 TM 1/01.04

You might also like