Professional Documents
Culture Documents
Publication information
Where to find information The User Assistance contains all information about the
product, including the following:
• Safety
• Routine operation
• Configuration information
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
3
General attention
Open Source and Commercial Software cobas® infinity central lab may include components or
modules of commercial or open-source software. For
further information on the intellectual property and other
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
4
Contact addresses
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
5
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
6
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 7
Table of contents
3 Pre-analysis
Test management 107
Preparing samples for distribution 110
Preparing samples for distribution in the
sample workflow engine 113
4 Sample analysis
Monitoring orders and tubes from the work
area 121
Managing samples in the laboratory area 129
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
8
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
9
Intended use
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
10
Roche Service representative (global and Roche Service representatives configure the system and
local) master data (test, test groups, senders, instruments,
reports, interfaces, etc.) according to specific
requirements of the customers with regards to
connected hosts, instruments, and sample workflow.
Access level depends on specific user rights.
GP/Hospital doctor/Hospital nurse/ These users are not employees of the laboratory; thus
Community nurse their access is restricted to the Lab Link module only.
They enter orders manually within the Lab Link module,
release print-outs of barcode labels for the tubes for
positive patient/sample identification, release print-outs
of reports for the patient for blood drawing, and view and
print the patient result report of their patients.
Product names Except where the context clearly indicated otherwise, the
following product names and descriptors are used.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
11
Symbol Explanation
o List item
r Start of a task
c Frequency of a task.
n Duration of a task.
j Prerequisites of a task.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
12
Symbol Explanation
Abbreviation Definition
Safety classifications
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
13
WARNING!
Warning...
r ...indicates a hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION!
Caution...
r ...indicates a hazardous situation which, if not
avoided, could result in minor or moderate injury.
NOTICE!
Notice...
r ...indicates a hazardous situation that, if not avoided,
may result in damage to the system.
i Tip...
...indicates additional information on correct use
or useful tips.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
14
Security
WARNING!
Secure infrastructure
WARNING!
Enabled security parameters
r All security parameters (profiles, access restrictions,
etc.) are enabled by default. Remember that disabling
security parameters may lead to potentially major
risks, e.g. unqualified staff may change instruments or
system configurations, or unauthorized users may
access confidential information.
UPS
WARNING!
UPS usage and software shutdown
CAUTION!
UPS usage
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
15
User settings
WARNING!
Restricting access to critical functions
WARNING!
Access to the software and user accounts
r Access to the software should only be granted to
users who have been trained and assigned a user
name and password. The password must be
confidential and must comply with the usual security
principles. It must be changed periodically to prevent
unauthorized parties from gaining access.
r There must be one account per user. Do not create
generic user accounts to be used by more than one
person. Create a different account for each user, even
if they have the same access rights.
r It is recommended that you develop a Standard
Operating Procedure (SOP) to ensure that only
qualified staff access the software.
WARNING!
Passwords, access, and system shutdown
r For security reasons and for patient confidentiality, all
users must exit the software or lock their computer
before leaving their workplace.
r It is recommended to correctly configure the
automatic log-off time.
CAUTION!
System performance
i System configurations
The preconfigured settings in the system must
only be used as a reference. It is recommended
that they are not used as default settings since
they should be modified to suit the system and
the health center requirements.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
16
General
CAUTION!
Failure to understand the software icons and
symbols
WARNING!
Regional settings
r Make sure that the configuration of the software
matches the regional settings and the local
requirements.
CAUTION!
Servers and software performance
CAUTION!
Alarms on validation screens and reports
CAUTION!
Barcodes
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
17
CAUTION!
Errors in identification of orders, tubes or patients
i Reading devices
Risk of results being misinterpreted
• It is recommended that you use reading
devices (optical readers, barcode readers,
card readers, etc.) to take results. Check the
reading devices are operating correctly before
using them in real environments.
Results
WARNING!
Result format
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
18
1 Accessing the application via When accessing the software using a link or No (default value)
Direct Access without having externally, this parameter allows you to access only
to log on the screens marked as accessible without having to
enter the password and user name on the logon
screen.
2 Session timeout (min) This parameter allows you to specify how long the 15
session is kept open (in minutes) since activity was
last detected. Once this time is exceeded, the
software logon screen appears, and the user name
and password must be entered again.
3 Direct access session This parameter only applies to sessions accessed 15
timeout (min) from an external environment. This parameter
allows you to specify how long the session is kept
open (in minutes) since activity was last detected.
Once this time is exceeded, the session closes.
4 Days configuration trace is Indicates the number of days the configuration 9999
kept trace will be kept before deleting it.
5 History retention period (d) Days before the deletion the order traces. 9999
y Parameter configuration for 21 CFR Part 11 compliance
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
19
6 Delete the orders and When deleting orders and patients, the application 24 (default value)
patients from the recycle bin sends them to the recycle bin. These orders /
after (hours) patients remain in the recycle bin until they exceed
the number of hours set in this parameter. The
system then proceeds to permanently delete these
items from the database via the end of day process.
7 Number of logon attempts Indicates the number of consecutive mistakes that 5
before the user is blocked the you can make when logging in.
8 Password renewal period (d) Period of time when the system will request to 90 (default value)
change the password for a user.
9 Password expiry warning (d) Period of time when the system will start reminding 10
you to change the password before the password
renewal period.
10 Minimum number of Minimum number of uppercase letters that the 1 (default value)
uppercase letters in password must contain.
password
11 Minimum number of digits or Minimum number of digits or symbols that the 1 (default value)
symbols in password password must contain.
12 Minimum password length Minimum length required for the password. 6 (default value)
13 Request password when Indicates if the password for the user will be Yes
accessing the validation requested when entering the validation screen or
screen (positive user Rule Engine validation screen.
identification) and the rule
engine administration screen
14 Enable automatic data Indicates if the system will remove order data No (default value)
removal in database automatically from the database.
15 Add comment when editing When enabling this parameter, the application Yes
validated tests (for all required forces the user who edits validated results and
applications: General Lab, wants to save them, to enter a comment. The
Emergency Lab and callout that appears for entering the comment
Microbiology) cannot be left empty. It is possible to enter a coded
comment.
These comments do not appear by default in the
report. If you want to view them there, the check
box of the Imp./Host column on the Comments
screen should be selected.
16 Request confirmation when If this parameter is enabled, the system requests Yes
changing results (for all you to confirm whether the test result change
required applications: General should actually be saved in the database. This
Lab and Emergency Lab) option helps you avoid mistakes.
y Parameter configuration for 21 CFR Part 11 compliance
ID Configuration Procedure
18 Report configuration includes the user that validates the 1. Choose the corresponding report.
result.
2. Choose the Design button.
3. Include the Validation user name list field in the
report.
y Report configuration for 21 CFR Part 11 compliance
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
20
ID Configuration
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 23
Interface overview
In this chapter 1
Accessing the software ....................................................... 25
Logging on to the software..................................... 25
Logging off the software.......................................... 25
Changing the password ........................................... 26
About the software framework ......................................... 27
Performing basic tasks in the software ......................... 29
Changing the software module ............................. 29
Changing the user profile........................................ 29
Changing the profile location................................. 30
About location and multisite................................... 31
Viewing notifications ................................................. 33
Viewing work area and monitoring charts ........ 34
About the Notifications screen.............................. 35
Assigning and managing shortcuts ..................... 36
Navigating through tabs, menus, and
screens............................................................................ 36
Searching for general information ....................... 37
Filtering ...................................................................................... 39
Filtering drop-down list values .............................. 39
1 Interface overview
Filtering information................................................... 39
Saving/deleting a filter.............................................. 40
Sorting and exporting table information ....................... 42
Sorting table information ......................................... 42
Exporting tables........................................................... 42
About the common interface items ................................ 44
List of common function keys................................ 44
List of common icons ................................................ 44
List of Lab Flow icons................................................ 45
List of sample workflow engine icons................. 47
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
24
1 Interface overview Table of contents
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 25
j m A compatible browser
m A valid user name and password
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
26 Accessing the software
system administrator.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 27
w Software framework
Global information area From this area, you can do the following:
• Change the software module
• Change the user profile
• Change the profile location
• Export tables
• Print barcode labels
u Related topics
• Changing the software module (29)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
28 About the software framework
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 29
In this section
Changing the software module (29)
Changing the user profile (29)
Changing the profile location (30)
About location and multisite (31)
Viewing notifications (33)
Viewing work area and monitoring charts (34)
About the Notifications screen (35)
Assigning and managing shortcuts (36)
Navigating through tabs, menus, and screens (36)
Searching for general information (37)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
30 Performing basic tasks in the software
u Related topics
• Changing the profile location (30)
u Related topics
• About location and multisite (31)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 31
Selecting the profile location If your profile is associated with more than one center or
service, you can select the location you are working in.
This affects the assignment of location to orders and
tests.
Manual or automatic location assignment to When you perform manual tasks such as entering orders,
orders and tests tests, or test results, a location is assigned to orders and
tests. In some cases, you can change the order location.
1 Interface overview
Some automatic actions performed by hosts, instruments
or by the rule engine also result in the assignment of a
location to orders, tests, and test results.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
32 Performing basic tasks in the software
Creating orders You create an order from Main • The order gets the profile location.
> Order entry > Order entry. However, you can change the order location
by selecting it from the Location drop-
down list in the Demographics group box
(if displayed).
• All the tests entered at this time get the
same location as the order.
You create a pre-order from • The order gets the profile location.
Main > Order entry > [Pre- However, you can change the location by
order entry]. selecting it from the Location drop-down
list in the Order details group box (if
displayed).
• All the tests entered at this time get the
same location as the order.
You retrieve a pre-order in Main The order gets the pre-order location and
> Order entry > Collection. cannot be edited.
An instrument sends an order. The order and all its tests get the instrument Contact your Roche
location. Service representative.
A host creates an order. The order and test locations depend on the Contact your Roche
host configuration. Service representative.
The rule engine creates an order The new order and all its tests get the location Contact your system
based on an existing order. of the existing order. administrator.
Adding tests You add a test to an existing The added test gets the profile location,
order in Main > Order entry > regardless of the order location.
Order entry.
You add a test to an existing The added test gets the profile location, u Adding/deleting
order from a validation screen. regardless of the order location. tests to/from an
order (150).
You retrieve a test affected by a The retrieved test gets the order location. u Rejecting or
rejection rule. retrieving tests (108).
The rule engine adds a test to an The added test gets the order location. Contact your system
existing order. administrator.
Entering or You enter or edit the results of a The test gets the profile location. u Entering and editing
editing test test from a validation screen. results manually
results
i If you relate the test result to an
instrument, the test gets the location of
(135).
the instrument.
You enter or edit results of a test The test location does not change and you can
from a worklist. only enter or edit results from tests with the
same location as your current location.
The rule engine reruns a test The test location does not change. Contact your system
with dilution or assigns an administrator.
automatic result to a test.
The rule engine enters or edits The test location does not change. Contact your system
the results of a test. administrator.
y Manual or automatic location assignment to orders and tests
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 33
An instrument sends test results The test gets the instrument location. Contact your Roche
or makes a query. Service representative.
A host sends the results of a The test location depends on the host Contact your Roche
test. configuration. Service representative.
y Manual or automatic location assignment to orders and tests
Filtering work areas by location Since all the orders and tests have an assigned location,
you can use this demographic to filter your work area
items. This way, you can narrow down the number of
items displayed.
Filtering and printing worklists Since all worklists are assigned to at least one location,
you can use this demographic to filter your worklists. The
software considers the locations assigned to the
worklists and your current location to display or print
worklists. If you have more than one location assigned,
you can choose the location to consider when adding
tests in the worklist.
Filtering QC results by location Since all analyzers are assigned to one location, you can
use this demographic to filter your QC results. The
software considers the locations assigned to the
analyzers and the locations you can access.
Viewing notifications
In the overview area, you can view the alarms and
warnings issued by the system when the events
configured as alarm triggers take place (orders are not
validated, instruments get disconnected, etc.). 1 Interface overview
j m Notification severity conditions configured by your
system administrator
r To view notifications
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
34 Performing basic tasks in the software
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 35
1 Interface overview
A Number of unread notifications of each type: C Number of items affected by the corresponding
- Red: high-severity notifications notification.
- Orange: medium-severity notifications
- Gray: low-severity notifications
B Check box indicating that the notification has been
read.
w Notification screen
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
36 Performing basic tasks in the software
r To sort shortcuts
u Related topics
• List of common icons (44)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 37
Keyboard shortcuts for tab and menu Use the following keyboard shortcuts to navigate through
navigation tabs and menus:
u Related topics
• Searching for an order (65)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
38 Performing basic tasks in the software
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 39
Filtering
Several screens in the software include tables displaying
orders, patients, and tests, among others. Also, some
drop-down lists contain long lists of items to choose
from. You can filter the items on such tables or drop-
down lists to find the information you need more easily.
Wild-card characters in filter fields When entering text in filter fields, you can use the
following wild-card characters:
• * matches one or more characters. For example, in the
Order ID filter field, enter GL201307* to obtain the
available Order IDs starting with this sequence
number, such as GL20130710 and GL20130711.
• ? matches a single character. For example, in the
First name filter field, enter J?an if you do not know
if the patient’s name is Joan or Juan.
In this section
Filtering drop-down list values (39)
Filtering information (39)
Saving/deleting a filter (40)
Filtering information
You can reduce the information displayed on tables to
see only the items that you need.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
40 Filtering
Saving/deleting a filter
You can save, edit, and delete frequently used search
configurations to easily apply them when you access the
screen again.
area:
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 41
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
42 Sorting and exporting table information
In this section
Sorting table information (42)
Exporting tables (42)
descending order.
Exporting tables
From the global information area, you can export the
information displayed on a table to different file formats.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 43
r To export tables
1 Interface overview
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
44 About the common interface items
Common function keys You can use these function keys to perform tasks
throughout the software.
Module icons These icons indicate that the adjacent item belongs to a
specific software module.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 45
Icon Description
Microbiology module
y Module icons
Icon Description
Enabled area
Disabled area
Enabled group
Disabled group
Enabled buffer
Disabled buffer
Enabled analyzer
Disabled analyzer
y Target icons
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
46 About the common interface items
Icon Description
Enabled archive
Disabled archive
y Target icons
Tube icons These icons indicate the types of tube or their status.
Icon Description
Classified aliquot
Aliquoted tube
Closed target
Distribution error
y Tube icons
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Interface overview 47
Icon Description
Aliquot error
y Tube icons
Rack icons These icons indicate the position of a tube on the rack
on the following screens:
• Main > Tube management > Manual distribution
• Main > Tube management > Manual aliquot
• Main > Tube management > Manual archive
• Main > Workflow > Manual archive
• Main > Workflow > Archive query
Icon Description
y Rack icons
Workflow icons The following icons indicate the type of target in the
Sample tracking screen.
1 Interface overview
Icon Description
Instrument buffer
Instrument
Manual archive
Post-analytic
Pre-analytic
Virtual target
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
48 About the common interface items
Tube icons The following icons indicate the type of tube and its
status.
Icon Description
Aliquot created
Sample centrifuged
Sample retrieved
Sample disposed
Closed node
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Routine tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 51
General tasks
In this chapter 2
Viewing and assigning comments ................................. 53
Assigning coded comments ................................... 53
Assigning rich-text comments............................... 54
List of comment icons............................................... 55
Printing barcode labels manually .................................... 57
Printing order labels manually ............................... 57
Printing tube labels .................................................... 60
Printing tray labels...................................................... 61
List of manual print options for barcode
labels................................................................................ 61
Checking connections ......................................................... 63
Checking the host connection............................... 63
Checking the instrument connection ................. 63
Order management ............................................................... 65
Searching for an order.............................................. 65
Accessing closed orders .......................................... 65
About locked orders .................................................. 66
Viewing order traces.................................................. 67
Releasing tests or orders .................................................... 69
Masking..................................................................................... 70
Masking and unmasking manually...................... 70
About the Masking screen...................................... 72
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
52 Table of contents
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 53
In this section
Assigning coded comments (53)
Assigning rich-text comments (54)
List of comment icons (55)
• Order
• Test
or, only for Microbiology:
• Culture medium
• Sample type
• Isolate
I This option will be automatically filled according
to the icon you choose to enter the screen.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
54 Viewing and assigning comments
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 55
Displays the order comment and allows you to assign a Validation and Queries screens.
comment to an order.
Displays the patient comment and allows you to assign Validation and Queries screens.
a comment to a patient.
y Comment icons
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
56 Viewing and assigning comments
Displays the test comment and allows you to assign a Validation and Queries screens.
comment to a test.
Displays the test, sample type, culture medium, or Seeding, and Validation screens.
isolate comment when you hover over it.
Indicates that the comment comes from an instrument. Type column on the Comment, and Validation
screens.
Indicates that the comment has been entered manually Type column on the Comment, and Validation
by a user. screens.
Indicates that the comment has been entered Type column on the Comment, and Validation
automatically by the system. screens.
y Comment icons
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 57
In this section
Printing order labels manually (57)
Printing tube labels (60)
Printing tray labels (61)
List of manual print options for barcode labels (61)
i
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
58 Printing barcode labels manually
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 59
u Related topics
• List of manual print options for barcode labels (61)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
60 Printing barcode labels manually
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 61
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
62 Printing barcode labels manually
General Lab Main > Order entry > One label for the order plus one
Emergency Lab Order entry label for each pending tube
Lab Link
Main > Work areas > One label for each order tube
[Work area]
Main > Order entry > One label for the order plus one
Collection label for each collected sample
Global information area Existing orders (with order One label for each order
2 General tasks
range)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 63
Checking connections
In this section
Checking the host connection (63)
Checking the instrument connection (63)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
64 Checking connections
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 65
Order management
In this section
Searching for an order (65)
Accessing closed orders (65)
About locked orders (66)
Viewing order traces (67)
4 Enter either the order ID, patient ID, or tube ID, and
then press Enter.
I If you are unsure of the data, enter an asterisk (*)
in any of these fields to search for all information.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
66 Order management
u Related topics
• Entering and editing results manually (135)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 67
u Related topics
• Searching for an order (65)
• Monitoring the workflow in the sample workflow
engine (87)
• Entering and editing results manually (135)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
68 Order management
u Related topics
• About location and multisite (31)
• About test rejection (107)
• List of result entry and validation screen items (177)
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 69
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
70 Masking
Masking
For lab maintenance, lack of reagent, or cost control, etc.
you may want to avoid performing certain tests or sorting
tubes to certain instruments or targets at specific times.
To do so, a masking actor must mask or unmask tests,
instruments, targets, or a combination of all.
CAUTION!
Unintended instrument unmasking
In this section
Masking and unmasking manually (70)
About the Masking screen (72)
r To mask manually
1 Choose Monitoring > Masking.
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 71
r To unmask manually
1 Choose Monitoring > Masking.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
72 Masking
u Related topics
• About the Masking screen (72)
Masking actors
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 73
Direct masking The masking actor masks an item. The actor's icon is
displayed in black on the tab corresponding to the type
of item masked.
Indirect masking The masking actor does not mask the item itself, but
another item which is related. The actor's icon is
displayed as gray in all the tabs where the item appears.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
74 Masking
Tabs
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 75
Right panel
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
76 Masking
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 77
In this section
Reviewing AON results (77)
Rejecting AON results (78)
Displaying AON results on the Levey-Jennings
chart (79)
Resetting an AON profile (81)
Configuring the AON Result review screen (82)
List of AON Result review screen items (82)
review.
2 In the Filter settings area, from the drop-down lists,
choose the relevant options, and the choose the
Apply button.
• Alternatively, from the Available filters drop-
down list, choose a saved filter.
I If you already have a Default filter, it is
automatically applied. The Parameter time
option in the Date field depends on configuration.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
78 Monitoring Average of Normals results
u Related topics
• Displaying AON results on the Levey-Jennings chart
(79)
• Rejecting AON results (78)
• Configuring the AON Result review screen (82)
• List of AON Result review screen items (82)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 79
u Related topics
• Reviewing AON results (77)
• List of AON Result review screen items (82)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
80 Monitoring Average of Normals results
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 81
u Related topics
• Reviewing AON results (77)
• Rejecting AON results (78)
• Configuring the AON Result review screen (82)
• List of AON Result review screen items (82)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
82 Monitoring Average of Normals results
u Related topics
• List of AON Result review screen items (82)
u Related topics
• List of AON Result review screen items (82)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 83
AON status Indicates the status of the AON profile for that result. The
status is set as correct or incorrect according to the
threshold levels defined for that profile. The AON status
affects the validation of the related patient results.
Icons Description
Icons Description
Affected patients' results Indicates the number of patient result affected by this
AON result.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
84 Monitoring Average of Normals results
Icons Description
Statistics values This table displays the statistic values calculated for a
series of results of an instrument-test-profile
combination.
Result markers If, from the Grouping period drop-down list, you choose
any grouping other than the No group option, the
software calculates an average of the following values:
results.
• Group SD: SD calculated from all the results in the
group.
Only available for symmetric profiles.
• SD index: SD calculated for the results included in
the group.
Only available for symmetric profiles.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 85
Tracking a sample
You can monitor the status and location of the samples
as they move through the different instruments you have
connected.
r To track a sample
1 Choose Main > Tube management > Sample
tracking.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
86 Monitoring the workflow
u Related topics
• Closing a target (211)
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 87
r To track a sample
1 Choose Main > Workflow > Sample tracking.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
88 Monitoring the workflow in the sample workflow engine
u Related topics
• About the Sample tracking screen (88)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 89
Date and time This column displays the date and time when the last
event of the target, node, or user was performed.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
90 Monitoring the workflow in the sample workflow engine
Instrument position It displays the position of the tube sent by the instrument.
You can only view the position of the past and current
targets or nodes.
Internal position This column displays the tube position for past and
current targets or nodes generated by cobas® infinity
central lab.
u Related topics
• List of route history items in the sample workflow
engine (92)
• List of sample workflow engine icons (47)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 91
On the Route details tab, you can only see some of the
events the tube passed through, while in the Sample
tracking result tab, the complete workflow is displayed.
For a detailed view of all the performed events, you must
go to the Route history screen.
Date and time This column displays the date and time when the event
was performed, starting with the most recent one.
• Aliquot created
• Aliquot registered
• Node reset
• Sample sorted
• Sample retrieved
• Sample disposed
u Related topics
• List of route history items in the sample workflow
engine (92)
• List of sample workflow engine icons (47)
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
92 Monitoring the workflow in the sample workflow engine
u Related topics
• List of route history items in the sample workflow
engine (92)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 93
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
94 Monitoring the workflow in the sample workflow engine
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 95
Actor Shows the target that performs the event or the software
component that triggers the event.
User ID Shows the identifier of the user that performs the event.
Profile Shows the user profile that is logged in when the action
is performed. It only applies when the event is related to
a user action.
Aliquot created The number of the aliquot ID. • The pre-analytical target that • N/A when the pre-
creates the aliquot and sends analytic creates the
the message. aliquot.
• N/A when the aliquot is created • User who triggered
manually. the manual creation
of the aliquot.
Aliquot error • Name of the targets that are in The pre-analytical target that sends N/A
the route for the aliquot. the message.
• The codes of the aliquot errors.
Aliquot registered • Tests for which the aliquot has The pre-analytical target that N/A
been created. creates the aliquot and sends the
message.
• Primary tube ID.
N/A when the aliquot is created User who triggered the
• Aliquot volume.
manually. manual creation of the
aliquot.
Aliquoting error • Name of the targets that are in The pre-analytical target that sends N/A
the route for the aliquot. the message.
• The codes of the aliquot errors.
Disabled rack-pos. link • The ID printed on the rack. The analyzer, pre-analytic, or post- N/A
analytic that deactivates the rack-
• The rack ID in the instrument.
pos. link information.
• The position of the tube (rack-
pos. link information).
Manual retrieval The name of the target from which N/A The user who chooses
requested the tube has been manually the Retrieve button.
requested to be retrieved.
y Information that can appear in some columns in the route history table
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
96 Monitoring the workflow in the sample workflow engine
Node reset • The node that has been reset. N/A The user who performs
the reset action.
• If there is a target associated to
the node, this target is also
displayed.
• For nodes with results, it shows
if the results have been kept or
the test has been repeated.
Rack-pos. link created • The ID printed on the rack. The analyzer, pre-analytical, or post- N/A
analytical target that creates the
• The rack ID in the instrument.
rack-pos. link information.
• The position of the tube (rack-
pos. link information).
Rack-pos. link deleted • The ID printed on the rack. The post-analytical target that N/A
deletes the rack-pos. link
• The rack ID in the instrument.
information.
• The position of the tube (rack-
"System workflow engine" when the
pos. link information).
end of day process or a query from
the analyzer deletes the rack-pos.
link information.
Re-enabled rack-pos. • The ID printed on the rack. N/A The user that reactivates
link the rack-pos. link
• The rack ID in the instrument.
information.
• The position of the tube (rack-
pos. link information).
Result • The name of the analyzer that The analyzer target that sends the N/A
sends the result. result when there is an on-line
record.
• The name of the test for which
the result is received. The target where the tube is stored
if the result is the archive position.
• "~SYSWFE~"(system workflow
N/A when the instrument that sends
engine), if the result contains the
a result is not associated with a
archive position. target.
N/A when you enter the result The user who entered
manually. the result.
Sample centrifugation N/A The pre-analytical target that N/A
started performs the centrifugation.
Sample centrifuged N/A The pre-analytical target that N/A
with error performs the centrifugation.
Sample centrifuged N/A The pre-analytical target that sends N/A
the centrifugation status.
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 97
Sample instrument • The status of the tube: loaded or The analyzer target that sends the N/A
location unloaded. sample instrument location.
• If applicable, the module into
which the tube has been loaded
or from which it has been
unloaded.
Sample not N/A The pre-analytical target that sends N/A
centrifuged the centrifugation status.
Preanalytic sample • If there is an error related to the The pre-analytical target that sends N/A
processing error centrifugation, it shows the error.
Centrifugation error.
• If there is an error related to the
creation of the aliquot, it shows
Aliquot creation error.
• The description of the error.
• The code of the error.
• The node in which the event has
been generated.
Sample quality entered The sample quality informed. The pre-analytical target that sends N/A
the value.
N/A when the sample quality is The user who entered
entered manually. the sample quality
manually.
Sample retrieved The target where the tube has been The post-analytical or pre-analytical N/A
retrieved. target that sends a message when it
retrieves a tube.
N/A when the tube has been The user who retrieves
retrieved manually. the tube manually.
y Information that can appear in some columns in the route history table
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
98 Monitoring the workflow in the sample workflow engine
Sample seen • For the cobas® 8100 automated The pre-analytical or post-analytical N/A
workflow series, it displays the target that sends the message.
following information if
applicable:
– Name of the module where
the event is triggered.
– Name of the tray into which
the tube has been loaded.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 99
N/A when the tube has been sorted The user who forces the
to that target manually. target.
Target forced • The node that has been forced. N/A The user who forces the
target.
• The target that has been forced.
• The workflow the forced node
belongs to.
Tube decapped • Name of the decapper target. Pre-analytic or post-analytic target N/A
that decaps the tube.
• Tube decapped.
Tube capped • Name of the recapper target. Pre-analytic or post-analytic target N/A
that recaps the tube.
• Tube capped and cap type.
Tube position migrated • The Lab Flow post-analytical The post-analytical target from The user who migrates
target from which tube positions which the tubes have been the tube position.
migrated.
have been migrated.
• The sample workflow engine
post-analytical target to which
the tube positions have been
migrated.
Tube position restored • The Lab Flow post-analytical The post-analytical target to which The user who restores
target to which tube positions the tubes have been restored. the tube position.
have been restored.
• The sample workflow engine
post-analytical target from which
the tube positions have been
restored.
Unsolicited tube N/A The pre-analytical, post-analytical or N/A for unsolicited tubes
created analyzer target that sends the received by an
information of an unsolicited tube instrument.
The workplace in which the The user who archived
unsolicited tube has been manually the tube manually.
archived.
Unsolicited tube N/A External host, if the information N/A
merged comes from a host.
Order entry, if the information is The user who added data
entered manually in the Order entry to the order manually.
screen.
y Information that can appear in some columns in the route history table
2 General tasks
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
100 Managing tube positions in the sample workflow engine
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 101
Additionally, for the tubes that still have not reached the
maximum amount of scans, you can reset its counter to
allow more scans.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
102 Managing tube positions in the sample workflow engine
Enabled rack-pos. link Lets you know if the rack-position link is enabled or not.
Date/time of rack-pos. link creation Date and time when the rack-position link was created.
• In the cobas® 8100 automated workflow series, the
rack-position link is created when the tube reaches
the reformatter.
• In the cobas® connection modules (CCM), it is
created when the tube is queried by the first cobas®
8000 modular analyzer series.
• In the cobas p 501 post-analytical unit/cobas p 701
post-analytical unit, it is created when the tube is
retrieved from the refrigerator.
2 General tasks
Date/time of last scan Date and time when the rack was scanned for the last
time.
This column is empty until the rack is scanned after the
rack-pos. link has been created.
Target of last scan Target in which the rack was scanned for the last time.
This column is empty until the rack is scanned after the
rack-pos. link has been created
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
General tasks 103
Date/time of final deletion Date and time when the rack-position link is deleted.
Queries left for deactivation Number of scans left until the rack-position link is
deactivated.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
104
2 General tasks Managing tube positions in the sample workflow engine
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 105
Pre-analysis
In this chapter 3
Test management .................................................................. 107
Test rejection ................................................................ 107
About test rejection......................................... 107
Reviewing rejected tests ............................... 108
Rejecting or retrieving tests......................... 108
Preparing samples for distribution .................................. 110
Distributing a tube manually .................................. 110
Performing manual aliquots.................................... 110
Undoing a target ......................................................... 111
Forcing a target ........................................................... 111
Preparing samples for distribution in the sample
workflow engine..................................................................... 113
Distributing a tube manually in the sample
workflow engine.......................................................... 113
Performing manual aliquots in the sample
workflow engine.......................................................... 115
Resetting a node in the sample workflow
engine.............................................................................. 116
3 Pre-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
106
3 Pre-analysis Table of contents
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Pre-analysis 107
Test management
Test rejection
In this section
About test rejection (107)
Reviewing rejected tests (108)
Rejecting or retrieving tests (108)
Test rejection overview Your administrator can configure one or more rejection
rules, that is automatic processes that determine whether
a test should be removed from an order based on certain
conditions.
Test rejection icons Those tests that have been affected by a rejection rule
display a status icon on the Rejection review screen.
Some of these icons can also be displayed on work
3 Pre-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
108 Test management
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Pre-analysis 109
3 Pre-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
110 Preparing samples for distribution
position.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Pre-analysis 111
Undoing a target
Sometimes a tube is sent to a target by mistake or to an
unnecessary target. This process only happens in rare
cases but it can be reverted.
r To undo a target
1 Choose Main > Tube management > Manual
distribution.
Forcing a target
You can force tubes containing samples to pass through
the instruments you choose.
3 Pre-analysis
r To force a target
1 Choose Main > Tube management > Manual
distribution.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
112 Preparing samples for distribution
u Related topics
• Changing a tray (209)
• Closing a target (211)
3 Pre-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Pre-analysis 113
You can print holder and tube labels at the same time
you are distributing a tube by selecting the
corresponding check box. The labels are printed when
the tube is first scanned.
CAUTION!
Failure to understand the software icons and
symbols
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
114 Preparing samples for distribution in the sample workflow engine
u Related topics
• List of sample workflow engine icons (47)
• Performing manual aliquots in the sample workflow
engine (115)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Pre-analysis 115
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
116 Preparing samples for distribution in the sample workflow engine
u Related topics
• List of sample workflow engine icons (47)
j m A tube distributed
r To reset a node
1 Choose Main > Workflow > Manual distribution.
I You can also access from the Sample tracking
screen.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Pre-analysis 117
3 Pre-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
118
3 Pre-analysis Preparing samples for distribution in the sample workflow engine
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 119
Sample analysis
In this chapter 4
Monitoring orders and tubes from the work area..... 121
Work area overview.................................................... 121
List of monitoring screen items............................. 122
Sorting work area orders or tubes........................ 124
Filtering work area items.......................................... 124
Printing work area order or tube lists ................. 126
Printing reports of work area pending tests ..... 127
Managing samples in the laboratory area.................... 129
Indicating the sample quality ................................. 129
Indicating the sample quality in the sample
workflow engine.......................................................... 129
Searching for pending tubes.................................. 130
Viewing targets by tube types................................ 130
Viewing a list of tubes by tray ................................ 131
Viewing lists of seen, not registered tubes ....... 132
4 Sample analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
120
4 Sample analysis Table of contents
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Sample analysis 121
In this section
Work area overview (121)
List of monitoring screen items (122)
Sorting work area orders or tubes (124)
Filtering work area items (124)
Printing work area order or tube lists (126)
Printing reports of work area pending tests (127)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
122 Monitoring orders and tubes from the work area
u Related topics
• List of monitoring screen items (122)
Order or test status icons The status icons represent the status of an order during
the sample processing.
Icon Status
The order has no tests or there are no tests for the test group in that order.
The order tests or the test group in that order, do not have any results.
The order tests or the test group in that order has results, but they are not validated.
The order tests or the test group in that order have technically validated results.
The order tests or the test group in that order have medically validated results.
The order tests or the test group in that order results have been printed or sent to the host.
(v) The result of the order was obtained with an old version of the test.
y Order or test status icons
4 Sample analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Sample analysis 123
Turnaround time column This column indicates the processing time of the order or
tube. The starting and final time, depend on
configuration.
4 Sample analysis
Color Description
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
124 Monitoring orders and tubes from the work area
Color Description
u Related topics
• List of comment icons (55)
• List of common icons (44)
u Related topics
• Sorting table information (42)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Sample analysis 125
Saved filters When you apply various saved filters, you see the items
that meet all the criteria set (AND filter). For example, if
you select Today’s orders and Orders pending validation,
you see the orders or tubes that have been entered today
and are pending validation.
Associative filters When you apply an associative filter, you see the items
that meet at least one of the criteria set (OR filter). If you
combine Today’s orders and Orders pending validation,
you see all the orders that have either been entered
today or that are pending validation.
Location filters You can filter your work area items to see only the
orders, or tests assigned to a specific location.
Status filters You can recalculate the status of the work area and/or
the work area orders based on the following:
• You include only the tests belonging to the selected
location/s.
• You exclude delay tests.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
126 Monitoring orders and tubes from the work area
u Related topics
• Filtering information (39)
• Saving/deleting a filter (40)
• Viewing all order tests (149)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Sample analysis 127
3 Do the following:
• Choose either the All or the Selected option and,
in the latter case, select the desired orders or
tubes.
• Choose the status of the orders or tubes to be
displayed (with results, pending validation, and so
on) by selecting one or more check boxes.
4 Choose the Tests, Tubes, or Sample types buttons
to create a list of orders or tubes with the
corresponding tests, tubes (only for orders), or
sample types.
f A PDF list is generated.
5 Optionally, choose the or the icons to print or
download the report.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
128 Monitoring orders and tubes from the work area
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Sample analysis 129
4 Sample analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
130 Managing samples in the laboratory area
2 Enter the tube ID and the entry date, and then press
Enter.
3 Select the tube whose sample quality you want to
indicate.
4 Choose the Sample quality button.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Sample analysis 131
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
132 Managing samples in the laboratory area
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 133
In this chapter 5
Entering and editing results manually ........................... 135
Entering results by test ............................................. 135
About the Query/Validation by test screen....... 137
Entering pre-configured results and
comments ...................................................................... 139
Entering urinalysis results........................................ 140
Configuring hematology counters........................ 143
Counting blood cells manually .............................. 145
About the manual cell counter .............................. 147
Verifying and validating results ........................................ 148
Checking result consistency before
validation........................................................................ 148
Viewing all order tests ................................... 149
Adding/deleting tests to/from an order .. 150
Viewing the patient’s previous orders...... 151
Viewing the patient’s evolution chart....... 153
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
134 Table of contents
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 135
WARNING!
Manual result entry
CAUTION!
Manual result input and result format
Erroneous results
r Use codified results to avoid incorrect or non-existent
data from manual or automatic input.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
136 Entering and editing results manually
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 137
Column length and sorting • The length of the table columns can be modified by
dragging them.
• The items displayed in each column can be sorted by
clicking on the column header. Tests are sorted
following the order set during their configuration.
Access from a work area Depending on your configuration, you can also access
the Query/Validation by test screen from your work
area. In that case, consider the following:
Default filter If the default filter is selected when accessing the screen,
some fields are already filled in.
Filter combination Search criteria are combined by the AND operator, that
is, only the items meeting each of the selected conditions
are displayed.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
138 Entering and editing results manually
Symbol Description
Time filter If you use the Time filter, consider the following:
• If you only fill in the From field, the To field is
automatically set to 24:00.
• If you only fill in the To field, the From field is
automatically set to 00:00.
Test result column The font and color of the Test result field change
according to the result status. If a result has several
result statuses, the field is displayed in the format
defined for the result status with the highest priority.
cobas e flow tests visualization To see the hierarchy of cobas e flow main and
embedded tests in the table, you must select the Show
5 Result entry and validation
Icon/ Description
Button
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 139
Icon/ Description
Button
u Related topics
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
140 Entering and editing results manually
u Related topics
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 141
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
142 Entering and editing results manually
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 143
u Related topics
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
144 Entering and editing results manually
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 145
u Related topics
• Counting blood cells manually (145)
• About the manual cell counter (147)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
146 Entering and editing results manually
u Related topics
• About the manual cell counter (147)
• Filtering work area items (124)
• Configuring hematology counters (143)
• Completing a morphology assessment or review
(164)
• Viewing and assigning comments (53)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 147
Manual cell counter operation A common hematology test is the blood differential test,
which quantifies the number and percentage of each
type of white blood cell (WBC). The WBC test is used to
calculate the 100% test, an absolute count of the white
blood cell types determined by the user. Additionally, you
can also count other types of blood cells, such as the
metamyelocytes, which are not ordered with the WBC
test, but may assist in particular patient pathology
reports.
The non-corrected WBC A type of cell that you may want to discount from the
100% test is the nucleated red blood cells (nRBCs). Due
to continuing differentiation these erythroblast cells
maintain a nucleus, which can result in their incorrect
classification as lymphocytes, and can be mistakenly
included in the WBC test. If the connected analyzer is
able to differentiate these unique cell types, then the cell
counter can be configured to consider this classification.
If the analyzer can differentiate for nRBCs then the WBC
count sent to the software and displayed on the
validation screen is already corrected. That is, it does not
include nRBCs in the test.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
148 Verifying and validating results
WARNING!
Using images with tests
CAUTION!
Validating results
In this section
Checking result consistency before validation (148)
Repeating and holding test results (165)
Overriding result statuses and alarms (171)
Validating results manually (173)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 149
In this section
Viewing all order tests (149)
Adding/deleting tests to/from an order (150)
Viewing the patient’s previous orders (151)
Viewing the patient’s evolution chart (153)
Viewing test details (154)
Viewing QC details (156)
Viewing Average of Normals details (157)
Viewing hemogram images (158)
Viewing tube and sample information (160)
List of tab details (161)
Requesting reviews in hematology orders (164)
Completing a morphology assessment or review (164)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
150 Verifying and validating results
u Related topics
• Filtering work area items (124)
j m An open order
5 Result entry and validation
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 151
4 Do the following:
• Choose a test supergroup
• Choose an available test group
• From the Available tests list, double click it.
• From the Tests, select the test to delete, and
choose the button.
• Choose the Confirm button
f The test is added to or deleted from the order.
u Related topics
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
152 Verifying and validating results
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 153
u Related topics
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
154 Verifying and validating results
u Related topics
• Filtering work area items (124)
5 Result entry and validation
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 155
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
156 Verifying and validating results
u Related topics
• Filtering work area items (124)
• Repeating tests (165)
• Repeating tests with dilution (166)
• List of tab details (161)
Viewing QC details
To verify that the instrument was working correctly when
the test was performed, you can view information on the
related QC and calibration.
5 Result entry and validation
r To view QC details
1 Choose Main > Work areas > [Your work area].
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 157
u Related topics
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
158 Verifying and validating results
u Related topics
5 Result entry and validation
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 159
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
160 Verifying and validating results
u Related topics
• Counting blood cells manually (145)
• About the manual cell counter (147)
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 161
u Related topics
• List of Lab Flow icons (45)
• Tracking a sample (85)
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
162 Verifying and validating results
Test tab
Reference range color code The test result is displayed on a chart where colors
represent different reference ranges:
• Green: normal range
• Orange: tolerance range
• Red: panic range
• Gray: outside range
Delta check alarm icons The difference between the current and the previous test
result may trigger a delta check alarm:
Test tracking The User name column displays the name of the actor
who performed the action. However, if the software
performs the action it displays one of the following
options:
• "SYSTEM_RULENGINE": actions performed by rule
engine.
• "SYSTEM_DAILYPROC": actions performed by the
end-of-day process.
• "~SYSValDaemon~": validation actions performed
5 Result entry and validation
automatically
• "~System~": other actions performed by the software,
e.g. formula calculations
Result status
u Related topics
• List of result entry and validation screen items (177)
• List of default result statuses (181)
• List of QC Result review screen items (244)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 163
QC tab
QC OK
QC warning
QC error
y Controls icons
u Related topics
• List of result entry and validation screen items (177)
• List of default result statuses (181)
• List of QC Result review screen items (244)
• List of AON Result review screen items (82)
AON tab
AON OK
AON warning
AON error
y Controls icons
u Related topics
• List of result entry and validation screen items (177)
• List of default result statuses (181)
• List of QC Result review screen items (244)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
164 Verifying and validating results
u Related topics
• Viewing hemogram images (158)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 165
r To complete a morphology
assessment or review
1 Choose Main > Work areas > [Your work area].
• Alternatively, choose Main > Work areas >
[Your work area] > Validation.
2 Select an order or test, and choose the Task
completed button.
u Related topics
• Viewing hemogram images (158)
• Requesting reviews in hematology orders (164)
• Counting blood cells manually (145)
In this section
Repeating tests
You can repeat a test when you are unsure about the
correctness of the result.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
166 Verifying and validating results
r To repeat a test
1 Choose Main > Work areas > [Your work area].
2 Double-click the desired order.
f Your validation screen is displayed.
u Related topics
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 167
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
168 Verifying and validating results
u Related topics
• Filtering work area items (124)
• Establishing previous test results as current results
(171)
u Related topics
• Repeating tests with dilution (166)
5 Result entry and validation
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 169
u Related topics
• About holding and releasing test results (169)
• Filtering work area items (124)
About validating held test results If both technical and medical validations are configured,
and the test result is held during technical validation:
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
170 Verifying and validating results
Hold/Release icons and buttons On the Validation screen, you can hold or release a test
by choosing the buttons displayed on the test table.
u Related topics
• Holding and releasing test results (168)
• Verifying and validating results (148)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 171
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
172 Verifying and validating results
u Related topics
• Validating results manually (173)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 173
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
174 Verifying and validating results
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 175
u Related topics
• Filtering work area items (124)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
176 Sending validated results to the host from a work area
u Related topics
• Validating results from a work area (173)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 177
Test status icons Depending on the type of validation screen that you are
using, the status of a test can be represented by different
icons.
Icons/buttons Description
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
178 List of result entry and validation screen items
Icons/buttons Description
Test alarm icons Test results, as well as actions performed on tests may
trigger alarms. These are marked by icons displayed on
the result tables. Additionally, the font and color of the
Test result column change according to the result
status or alarm.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 179
Icon Description
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
180 List of result entry and validation screen items
Test val. Allows you to validate the selected tests. u Validating results manually (173).
Trace Allows you to consult the details of the various
steps the order has gone through.
Send results to Allows you to send all medically validated results, u Sending validated results to the host from a
host even if they have already been sent. work area (176)
Send unsent to Allows you to send the medically validated results u Sending validated results to the host from a
host that have not been sent yet. work area (176)
y Validation screen buttons
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 181
AONE Indicates that the AON result is out of the error threshold
defined.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
182 List of default result statuses
DCA Indicates that the previous and current results match the
configured delta check ones. Furthermore, they are
defined as alphanumeric.
DCH Indicates that the result is both within the defined range
and under the higher limit, and it is greater than or equal
to the previous result. Furthermore, it is defined as
numeric.
DCL Indicates that the result is both within the defined range
and over the lower limit, and less than the previous
result. Furthermore, it is defined as numeric.
DCT Indicates that the delta check result is within the defined
range and none of the 3 delta check alarms were
triggered. Furthermore, it is defined as numeric or
alphanumeric.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 183
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
184 List of default result statuses
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 185
IUU Indicates that the instrument unit from which the result
was received is not registered.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
186 List of default result statuses
ONC Indicates that the operator is not certified to use this type
of instrument.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 187
QCLV Indicates that the QC lot values have not been accepted.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
188 List of default result statuses
SI_WSI Indicates that the serum indices tests results have not
been received yet.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 189
SI_HR Indicates that the serum indices sensitive test run is held
until all pending serum index results are received.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
190 List of default result statuses
TPNR Indicates that there is no panic range defined for the test.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 191
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
192 List of default alarm types
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 193
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
194 List of default alarm types
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 195
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
196 List of default alarm types
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 197
RRA Indicates that a test result alarm has been added due to
the trigger of the Add result alarm action.
• Alarm name: Test result alarm added
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
198 List of default alarm types
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Result entry and validation 199
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
200 List of default alarm types
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 201
Post-analysis
In this chapter 6
Reports....................................................................................... 203
About report printing................................................. 203
Creating report filters ................................................ 205
Managing samples after analysis .................................... 207
Viewing sample results............................................. 207
Archiving samples manually................................... 207
Archiving a sample without an order
manually ......................................................................... 208
Changing a tray ........................................................... 209
Consulting an archive ............................................... 209
Removing a tube ......................................................... 210
Retrieving a sample from an automated
archive............................................................................. 210
Closing a target ........................................................... 211
Deleting a tray manually .......................................... 211
Viewing actions performed in the post-
analytical instrument ................................................. 212
Viewing tubes in the post-analytical
instrument...................................................................... 212
Managing samples after analysis in the sample
workflow engine..................................................................... 214
Viewing test results in the sample workflow
engine.............................................................................. 214
6 Post-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
202 Table of contents
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 203
Reports
In this section
About report printing (203)
Creating report filters (205)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
204 Reports
CAUTION!
Potential errors in reports due to reuse of old tests
WARNING!
Results
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 205
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
206 Reports
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 207
j m Labeled tubes
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
208 Managing samples after analysis
CAUTION!
Barcodes
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 209
Changing a tray
Trays are changed automatically when they become full.
Sometimes there is still space available in the tray from
yesterday but you prefer to start each day with a new,
empty tray. You must inform the software that you
changed the tray.
r To change a tray
1 Choose Main > Archive > Manual archive.
I Alternatively go to the Manual distribution
screen.
3 Enter the data for the new target where you want to
archive the sample.
f The next available tray appears by default. You
can accept it or delete it and enter your desired
target.
4 To print a label for the new target tray, select Print
barcode label check box.
5 Choose the Confirm button.
f Any samples you enter are now stored in the new
target.
Consulting an archive
You might have to check the status of an archived
sample because some tests are not performed daily in
the laboratory. You can check if there are any pending
6 Post-analysis
r To consult an archive
1 Choose Main > Archive > Archive query.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
210 Managing samples after analysis
Removing a tube
Once a tube reaches its maximum storage time, you must
remove it from the archive and dispose of it. This process
normally occurs automatically but you can enter this
information into the software manually so that the
position the tube occupied appears as empty.
r To remove a tube
1 Choose Main > Archive > Archive query.
WARNING!
Sample retrieval
6 Post-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 211
Closing a target
Once a tube passes through an instrument and all the
necessary processes are finished, that target can be
closed.
r To close a target
1 Choose Main > Tube management > Manual
distribution.
I Alternatively go to the Sample tracking screen.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
212 Managing samples after analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 213
6 Post-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
214 Managing samples after analysis in the sample workflow engine
j m A tube registered
In this section
Archiving a tube manually in the sample workflow
engine (215)
Archiving an unsolicited tube manually in the sample
workflow engine (216)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 215
j m A tube registered
m A manual archive target configured
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
216 Managing samples after analysis in the sample workflow engine
u Related topics
• List of Lab Flow icons (45)
• List of options in the Manual archive screen (217)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 217
u Related topics
• Archiving a tube manually in the sample workflow
engine (215)
• List of options in the Manual archive screen (217)
Tray change This button allows you to change the tray where you
archive the tubes.
6 Post-analysis
Warning for tubes with pending tests This check box warns you when the tube you want to
archive has tests that have not yet been performed.
Warning for already archived tubes This check box warns you when there is another tube
with the same tube ID archived in that target.
Accept unsolicited tubes automatically This check box allows you to directly archive unsolicited
tubes.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
218 Managing samples after analysis in the sample workflow engine
Print tray label This check box allows you to print the tray label.
u Related topics
• Archiving a tube manually in the sample workflow
engine (215)
• Archiving an unsolicited tube manually in the sample
workflow engine (216)
j m A tube registered
u Related topics
• List of Lab Flow icons (45)
• Retrieving a tube manually in the sample workflow
6 Post-analysis
engine (219)
In this section
Retrieving a tube manually in the sample workflow
engine (219)
Forcing the retrieval of a tube (220)
About automatic retrieval (221)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 219
During the process, you can also print a list of tubes that
you want to retrieve, as well as the labels of those tubes.
To print the labels, you must have at least one workplace
assigned.
WARNING!
6 Post-analysis
Sample retrieval
j m An archived tube
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
220 Managing samples after analysis in the sample workflow engine
u Related topics
• Forcing the retrieval of a tube (220)
j m An archived tube
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Post-analysis 221
u Related topics
• Retrieving a tube manually in the sample workflow
engine (219)
6 Post-analysis
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
222 Managing samples after analysis in the sample workflow engine
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC
7 QC ..................................................................................................................... 225
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
Table of contents 225
QC
In this chapter 7
About QC .................................................................................. 227
Entering control or calibrator results and reagent
lots manually............................................................................ 230
QC result review ..................................................................... 232
Reviewing control or calibrator results............... 232
Accepting and rejecting QC results..................... 234
Accepting lot values from the QC Result
review screen ............................................................... 236
Changing the QC lot usage from the QC
Review results screen................................................ 238
Changing the test status .......................................... 238
Releasing patient results when working with
a bracketing rule ......................................................... 240
Displaying QC results on the Levey-Jennings
chart................................................................................. 241
Configuring the QC Result review screen ......... 244
List of QC Result review screen items ................ 244
Displaying the cumulative chart for a control............. 253
About the cumulative chart................................................ 255
7 QC
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
226
7 QC Table of contents
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 227
About QC
Quality controls (QC) are performed daily in the
laboratory to verify whether instruments operate correctly
and produce reliable patient results.
CAUTION!
QC rules
WARNING!
QC results
WARNING!
QC testing frequency
CAUTION!
Reagent lots
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
228 About QC
CAUTION!
Automatic lot creation
WARNING!
QC lot usage
Term Definition
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 229
Term Definition
u Related topics
• Accepting and rejecting QC results (234)
• Changing the test status (238)
• Displaying QC results on the Levey-Jennings chart
(241)
• Displaying the cumulative chart for a control (253)
7 QC
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
230 Entering control or calibrator results and reagent lots manually
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 231
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
232 QC result review
QC result review
In this section
Reviewing control or calibrator results (232)
Accepting and rejecting QC results (234)
Accepting lot values from the QC Result review
screen (236)
Changing the QC lot usage from the QC Review results
screen (238)
Changing the test status (238)
Releasing patient results when working with a bracketing
rule (240)
Displaying QC results on the Levey-Jennings chart (241)
Configuring the QC Result review screen (244)
List of QC Result review screen items (244)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 233
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
234 QC result review
u Related topics
• Accepting and rejecting QC results (234)
• Changing the test status (238)
• List of QC Result review screen items (244)
• Displaying QC results on the Levey-Jennings chart
(241)
• Saving/deleting a filter (40)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 235
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
236 QC result review
u Related topics
• Filtering information (39)
• List of QC Result review screen items (244)
Icon Description
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 237
u Related topics
• Accepting and rejecting QC results (234)
• Saving/deleting a filter (40)
• List of QC Result review screen items (244)
7 QC
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
238 QC result review
j m A lot
m Work with administration screens right
QC.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 239
3 In the table, select the check box of the result for the
instrument-test-control combination whose status
you want to change.
I A correct test status is represented by the icon
in the QC status column.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
240 QC result review
3 In the table, select the check box of the result for the
instrument-test-control combination whose status
you want to change.
I A correct test status is represented by the icon
in the QC status column.
u Related topics
• Accepting and rejecting QC results (234)
• List of QC Result review screen items (244)
• Displaying QC results on the Levey-Jennings chart
(241)
• Saving/deleting a filter (40)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 241
u Related topics
• List of QC Result review screen items (244)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
242 QC result review
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 243
u Related topics
• List of QC Result review screen items (244)
7 QC
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
244 QC result review
u Related topics
• List of QC Result review screen items (244)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 245
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
246 QC result review
Icons Description
Icons Description
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 247
Icons Description
Test Indicates the test name for which the QC result was
received or calculated.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
248 QC result review
Icons Description
Result type Indicates the type of expected value for this result.
Icons Description
Qualitative result
Test type This columns indicates the test type related to the
received or calculated QC result.
• Regular test
• Hemolysis
• Icterus
• Lipemia
• Hematology
• Hematology alarm
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 249
Associated results This column indicates that patient results are linked to
this QC result and other related QC results, whereby an
icon is displayed.
Unit This column shows the units for the QC result. The units
used are chosen from the Unit drop-down lists, from the
Manual result entry and ICA > Instrument definition
> QC tests screens.
Result usage Indicates whether the result comes from a control set as
In use or as Study.
u Related topics
• Displaying QC results on the Levey-Jennings chart
(241)
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
250 QC result review
C D
E
A
B
Levey-Jennings chart icons The icons on the chart represent the status of the control
or calibration.
Icon Description
Levey-Jennings chart lines The lines represent the evolution of control results over
time. Standby bottle control results, however, are not
bounded by a line.
QC result markers If, from the Grouping period drop-down list, you choose
any grouping other than the No group option, the
software displays the lot values from the most recent QC
7 QC
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 251
Statistics table This table displays the statistic values calculated for a
series of results of an instrument-test-lot combination.
– Qualitative ( )
– Symmetric ( )
– Asymmetric ( )
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
252 QC result review
u Related topics
• Displaying QC results on the Levey-Jennings chart
(241)
7 QC
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 253
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
254 Displaying the cumulative chart for a control
u Related topics
• About the cumulative chart (255)
7 QC
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
QC 255
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
256
Glossary
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
257
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
258
patient release
Person whose sample is tested on an instrument or Final step in the result validation workflow in which
system. the evaluated test results are sent to the host.
priority sample
Property of an element on a system that describes the Representative part of a body or the substance to be
urgency with which it has to be dealt with. tested that is used in the analysis.
QC bracketing screen
Setting in which a test result must be preceded by, Component of the monitor that displays the images.
and followed by, a successful QC result before it is
released. standby bottle
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
259
test
Measuring procedure that requires laboratory
equipment and assays in a specific clinical context
and for a specific clinical purpose, in a specific
population.
test group
Tests that are listed together on an order form or
report form.
test type
Classification of tests based on how and with what a
test measures something.
tube
Container in which liquid is collected, transported,
processed, or archived, typically made of glass or
plastic.
validation
Quality assurance process in which tests are
evaluated before they are released to the host.
work area
Main user interface area for performing tasks and
viewing data.
workplace
Higher level of organization that consists of
functionally related components, and the lowest
logical unit of organization within an information
solution system below a work area.
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
260
Index
A Control or calibrator
– accepting results, 235
Addresses, 4
– changing status, 239
AON
– manual result entry, 230
– Rejecting results, 78
– rejecting results, 235
– Reset profile, 81
– result review, 232
– Result review screen, 82
Conventions used in this publication
– Results, 77, 79
– abbreviations, 12
– Results AON:Levey-Jennings chart, 79
– product names, 10
– Screen configuration, 82
– symbols, 11
B Copyright, 3
Barcode labels Cumulative chart, 253
– manual printing, 57
E
– orders, 57
Edition notice, 2
– print options, 61
– trays, 61 F
– tubes, 60
Feedback, 4
C Filter
– capture, 40
CCM AOB, 221
Checking connections H
– host connection, 63
Hematology counter, 147
– instrument connection, 63
cobas infinity central lab software L
– 21 CFR Part 11 compliance, 18 Levey-Jennings chart, 79, 241
– data privacy, 20
– intended use, 9 M
– safety classifications, 12 Masking and unmasking, 70
– symbols and abbreviations, 10
O
– system safety, 13
Order
– use environment and intended users, 9
– locked orders, 66
Comments
Order management
– coded comments, 53
– closed orders, 65
– icons, 55
– order, 65
– rich comments, 54
– order traces, 67
– rich-text comments, 54
Order release, 69
Contact, 4
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
261
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
262
Samples V
– list of seen, not registered tubes, 132
Validation
– list of tubes by tray, 131
– all order tests, 149
– pending targets, 130
– AON details, 157
– pending tubes search, 130
– checking result consistency, 148
– quality, 129
– Delta check alarm, 161
– quality in the sample workflow engine, 129
– from work area, 173
– targets by tube types, 130
– hemogram images, 158
Screenshot disclaimer, 3
– holding and releasing results, 168
Send results to host, 176
– patient´s evolution chart, 153
Software
– QC details, 156
– change password, 26
– repeating tests, 165
– common icons, 44
– result entry and validation screen items, 177
– common interface items, 44
– Result status, 154, 161, 177
– filters, 39
– sample dilution, 166
– framework, 27
– set previous results as current results, 171
– function keys, 44
– test details, 154
– Lab Flow icons, 45
– tube information, 160
– log off, 25
– viewing patient´s previous orders, 151
– log on, 25
– module change, 29 W
– monitoring charts, 34 Warranty, 3
– multisite access, 31 Workflow monitoring
– navigation, 36 – registered tubes status, 85
– notifications, 33, 35 – sample traces and tube route history, 86
– profile location, 30 – sample tracking, 85
– quick search, 37
– sample workflow engine icons, 47
– shortcuts, 36
– table information, 42
– user profile, 29
T
Table
– filtering, 39
Test release, 69
Tests
– Configuration in Hematology, 144, 145
Trademarks, 4
Roche Diagnostics
cobas® infinity central lab · Software version 3.02 · User Guide · Publication version 5.2
© 2020
Published by:
Roche Diagnostics, S.L.
E-08174 Sant Cugat del Vallès
Spain
www.roche.com