WaveLight® FS200

Quick Startup Manual

Step 1

Main
Laser Power Up
Switch
• Activate the power-supply using the “Main Switch” located
at the right side of the main unit or check for “On” (I)
position.
• Rotate the “Key Switch” clockwise to end position and back
to middle position.
WaveLight FS200 User Manual carefully, because it is more detailed in reference to the device functions.

• Wait for Windows and the WaveLight® FS200 Application to


complete boot-up.
• The login window will appear.
This Quick Startup Manual does not substitute the detailed User Manual. Read the

• Login using the master username and password or the


individual login.
• After completing the login process, select “Treatment” in
the “Popup Window”.
Note: To disable the selection “Microscope / Treatment
mode” please un-tick “Device Mode Selection” under
“Setup” (F10) and “System”.

Step 2
Calibration
• The system will automatically enter the calibration dialogue.
®

Note: The link to a possibly connected server is indicated by the below mentioned symbols.
The WaveLight® FS200 is connected to the WaveNet™ Server.
The WaveLight® FS200 is stand-alone or connection to the WaveNet™ Server failed.

Quick Startup Manual WaveLight FS200 en-us / Rev.05 / 17-02-02 Item No.: 6662 2014 Page 1 of 6
WaveLight® FS200
Quick Startup Manual

Prepare The Test Interface For Calibration


• Remove the transparent cover and clean the test disk using the cleaning pad.

Cleaning Pad
(Reverse Side
Of Test Disk
Holder)
WaveLight FS200 User Manual carefully, because it is more detailed in reference to the device functions.
This Quick Startup Manual does not substitute the detailed User Manual. Read the

• Place test disk in the “Test Disk Holder” with the black side down (WaveLight readable) and
mount “Test Disk Holder” to the “Adjustment Cone”.

Test Disk Holder

Test Disk

• Insert test interface and connect its vacuum tubing to the main unit.

Application Tubing Test


Interface Interface

Test Interface

• Confirm message “Insert test interface and PMMA ” with “OK”.


• Confirm message “Attention: Reference movement in all directions will be started” with “OK”.
®

The system will now carry out the reference movement and the system will continue into a
Vacuum Check automatically.

Note: The patient bed motion is blocked to all directions during the “reference run”.

Page 2 of 6 Quick Startup Manual WaveLight FS200 en-us / Rev.05 / 17-02-02 Item No.: 6662 2014
WaveLight® FS200
Quick Startup Manual

Step 3
Vacuum Check
• A successfully passed "Vacuum Check" is indicated with the associated green symbol in
the task bar and a green dot next to Vacuum Check in the initialization window.
WaveLight FS200 User Manual carefully, because it is more detailed in reference to the device functions.

Test in progress
Test passed
Test failed
This Quick Startup Manual does not substitute the detailed User Manual. Read the

Note: The above mentioned indications refer to the status of the currently active calibration /
test routine.

Step 4
Warm-Up
• At the task bar “Scanner Unit is running” will appear for a view minutes and later change to
“Warming up Laser”.

• After completing the “Warm-Up” sequence, the system continues automatically with the
“Energy Check”.

Note: The warm-up time depends to the ambient temperature and might last a minimum of
20 minutes but might take more than 30 minutes in a cold room environment.
If the system is restarted after initial warm up is completed the warm up time will last
(depending to the room temperature) a minimum of 4 minutes.

Step 5
Energy Check
• A successfully passed "Energy Check" is indicated with the associated green symbol in
the task bar, and a green dot next to the Energy Check in the initialization window.

Shows the current


maximum pulse energy in
”µJ”
®

Note: In “Treatment Mode” the pulse energy might be limited according to the selected
application. The maximum power used for treatments is not higher than 2.4 µJ.

Quick Startup Manual WaveLight FS200 en-us / Rev.05 / 17-02-02 Item No.: 6662 2014 Page 3 of 6
WaveLight® FS200
Quick Startup Manual

Step 6
Ablation Check
• Must be initiated by the user.
• Confirm message “Confirm with OK and proceed with ablation check” with “OK”.
WaveLight FS200 User Manual carefully, because it is more detailed in reference to the device functions.

• Click on the “Ablation Check” button to start.

Note: In case the optical arm was moved out of “Home Position” the following message will
This Quick Startup Manual does not substitute the detailed User Manual. Read the

appear: “Please move the optics arm to home position”.

• Please press the "Home Position Key" at the control panel.


• Confirm the message “Insert test interface with new PMMA” with “OK”.
• “Perform ablation check by pressing the laser foot pedal” will be indicated in the Info &
Warnings box.
• Keep the laser pedal pressed until the below mentioned
message appears. Laser Pedal

Note: Progress bar might stop at a value of 99%.

Ablation Check Evaluation


• To meet the criteria of passing the “Ablation Check” the following items have to be fulfilled:
- The last complete visible step of the stairway pattern must be parallel between the
orientation lines.
- The tolerance between the intersection line and the orientation lines is
± 5 µm in height.
- The ablated edge ring must be constant circular and between the two green circles of
the OPMI video image (diameter 10.0 ± 1.0 mm).

Test passed  the last complete step of stairway pattern ends inside the orientation lines
®

Orientation Lines

Ablated Edge Ring

Stairway Pattern

Last Step

Page 4 of 6 Quick Startup Manual WaveLight FS200 en-us / Rev.05 / 17-02-02 Item No.: 6662 2014
WaveLight® FS200
Quick Startup Manual

Does not count as


stairway pattern
(missing sectors)
WaveLight FS200 User Manual carefully, because it is more detailed in reference to the device functions.
This Quick Startup Manual does not substitute the detailed User Manual. Read the

Last complete stairway


pattern (no missing sector)

Test failed  The last complete step of Stairway pattern ends outside orientation lines

• Confirm that the criteria are fulfilled by using the microscope and selecting “ Yes”
• Remove the test interface, disconnect the vacuum tubing and confirm with "OK".
• The progress bar will now turn green and show 100% complete.

A successfully passed "Ablation Check" is indicated with the associated green symbol in the
task bar and a green dot next to Ablation Check in the initialization window:
®

Quick Startup Manual WaveLight FS200 en-us / Rev.05 / 17-02-02 Item No.: 6662 2014 Page 5 of 6
WaveLight® FS200
Quick Startup Manual

Step 7

• After having completed the calibration the system requires to remove the test interface.
Please remove the test interface and disconnect the vacuum tubing and confirm with "OK".
• The WaveLight® FS200 will automatically return to the “Standby Mode”.
WaveLight FS200 User Manual carefully, because it is more detailed in reference to the device functions.
This Quick Startup Manual does not substitute the detailed User Manual. Read the

Note:
• If one of the tests fails, the individual test can be repeated by pressing the related “Start-
button”.
• The laser is only ready for application on human eyes after having successfully passed all of
the required test and calibration procedures.

Step 8
Treatment Card
• To perform treatments, a valid WaveCard™ must be inserted into the WaveLight® FS200
user panel.
CORRECT: WRONG
Little Arrow on top
®

At the task bar the status of the WaveCard™ and the amount of procedures available is
displayed.

17 treatments available

No valid WaveCard™ inserted

No treatments available

Page 6 of 6 Quick Startup Manual WaveLight FS200 en-us / Rev.05 / 17-02-02 Item No.: 6662 2014