You are on page 1of 4

Engström Carestation TRM 1505-1018-000

Corrective Repair Checklist


.

Customer Name: Service Record


Number:
System ID or Serial Number: Equipment ID:
Software version: Date:
Performed by:

Inspect the system


CAUTION Before performing the tests:

• Complete all necessary calibrations and sub-


assembly tests. Refer to the individual procedures for
a list of necessary calibrations.
• Completely reassemble the system. If a test failure
occurs, make appropriate repairs and test for correct
operation.
.

Procedure Expected results Outcome Comments


The equipment is not Pass / Fail
• Components attached.
damaged.
• Gas supplies connected.
• Components are • Casters are not loose and brakes are
correctly attached. operational.
• The casters are not • Mains indicator turns "ON."
loose and the brakes
are set and prevent
movement.
• Power cord is
connected to a wall
outlet. The mains
indicator turns on when
AC Power is
connected.

Checkout procedure
.

Procedure Expected results Outcome Comments


Perform System Checkout. All checks pass. Pass / Fail

1
Backlight test
.

Procedure Expected results Outcome Comments


Access the Calibration Backlights trun off and screen brightness may Pass / Fail
menu. change.

• In the standby mode,


push the System Setup
key.
• on the System Setup
menu, select Install/
Service. (23-17-21).
• On the Install/Service
menu, select
Calibration.

Power failure test


.

Procedure Expected results Outcome Comments


System displays the 'On Battery' message Pass/Fail
1. Connect the power cord to a and cancels when AC is applied.
wall outlet. The mains
indicator on the front panel of
the display unit comes on
when AC Power is
connected.
2. Set the system swtich to On,
and then start a case.
3. Unpllug the power cord with
the system turned on.
4. Make sure the 'On battery'
message is displayed.
5. Connect the power cable
again.
6. Make sure the alarm
cancels.

2 1505-1018-000-CRC 04 2019
Electrical safety test
.

Procedure Expected results Actual results Outcome Comments


Measure the ground The measured Pass / Fail
resistance. ground resistance is
less than 0.2 Ohms.
Measure the leakage The measured Pass / Fail
current. leakage current is
less than 300
μAmps at 120 / 100
Vac and is less than
500 μAmps at 220 /
240 Vac.

1505-1018-000-CRC 04 2019 3
4 1505-1018-000-CRC 04 2019